You are on page 1of 15

International Orthodontics 2023; 21: 100753

Websites:
www.em-consulte.com
www.sciencedirect.com

Original article
Effects of various coating methods on the
mechanical, physical, and aesthetic
properties of GUMMETAL® archwires: In vitro
study

Alan Zhou 1, Steven Makowka 2, Stephen Warunek 1, Ming-Yu Chen 1, Thikriat Al-Jewair 1

Available online: 13 April 2023 1. Department of Orthodontics, School of Dental Medicine, State University of New
York at Buffalo, Buffalo, NY, USA
2. Materials Testing Facility, School of Dental Medicine, State University of New York
at Buffalo, Buffalo, NY, USA

Correspondence:
Thikriat Al-Jewair, 140 Squire Hall, Department of Orthodontics, University at
Buffalo School of Dental Medicine, 3435 Main Street, Buffalo NY 14214, USA.
thikriat@buffalo.edu

Keywords Summary
Coating
GUMMETAL® Purpose > The objective of this study was to evaluate the mechanical, physical, and aesthetic
Aesthetic archwires properties of GUMMETAL® (GM) orthodontic archwires after the application of various aesthetic
Friction coating materials.
Force deflection rate Methods > This in vitro study included 180 orthodontic wires: five experimental groups used
0.016  0.022-inch GM as the core-based wire followed by the application of epoxy, polytetra-
fluoroethylene (PTFE), clear ceramic, white ceramic, or silicone; and four control groups:
0.016  0.022-inch GM, 0.019  0.025-inch GM, 0.016  0.022-inch stainless steel (SS), and
0.019  0.025-inch SS. Frictional forces, force deflection rate, yield strength, aesthetic colour
value, corrosive and wear resistance were compared between the experimental and control
groups.
Results > Among the coated wires, white ceramic exhibited the highest frictional force (2.06
 0.20 N) and silicone showed the lowest values (0.88  0.12 N). There were significant differ-
ences in static friction between experimental and control groups (P < 0.001). PTFE coating had the
highest force deflection rate (9.03  0.12 N/mm) and yield strength (10.0  0.14 N/mm) among
coated wires and white ceramic exhibited the lowest values (6.86  0.14 N/mm and 7.74
 0.17 N/mm for force deflection rate and yield strength, respectively). Differences in force
deflection rate and yield strength between experimental and control groups were statistically
significant (P < 0.001). All coated wire groups had a clinical difference in colour when compared to
A1 shade. Coated and uncoated wires showed good corrosion resistance after one week in
corrosive saliva with no detectable loss of mass.
Conclusions > This study has shown that coating wires can improve some aspects of the wire
properties but not all when compared to uncoated GM and SS. Future investigation of the materials
used in this study is required to further characterize their properties.

tome 21 > n82 > June 2023


10.1016/j.ortho.2023.100753
© 2023 CEO. Published by Elsevier Masson SAS. All rights reserved.
A. Zhou, S. Makowka, S. Warunek, M-Y Chen, T. Al-Jewair

Original article
Introduction Material and methods
Orthodontic archwires are coated to enhance their perfor- Study design
mance, aesthetics, or both. The demand for more aestheti- Sample size estimation was conducted in reference to Albawardi
cally pleasing orthodontic appliances originated from et al. [16]. With a power of 90%, alpha of 5% and mean difference
patients who wanted a less metallic display [1–4]. The sur- in static friction of 0.7N, a minimum of 16 wires per group was
face characteristics and material composition of an ortho- needed. This in-vitro study included 180 wires divided into five
dontic wire determine properties such as its corrosive coated GM experimental groups and four uncoated GM and SS
resistance, wear resistance, aesthetic quality, and frictional control groups. The control wires (20 wires/group) were
force [5,6]. Coating orthodontic wires thus gives clinicians 0.016  0.022-inch GM (GUMMETAL®; JM Ortho corporation, Lot
the ability to manipulate these properties to improve clinical # 200109 M), 0.019  0.025-inch GM (GUMMETAL®; JM Ortho
performance and appearance. Corporation, Lot # 200109 M), 0.016  0.022-inch SS (Ormco Co,
Orthodontic archwires are commonly coated with diverse mate- Glendora, CA, Lot# 219-1408), and 0.019  0.025-inch SS (Ormco
rials, including epoxy resin, polytetrafluoroethylene (PTFE), rho- Co, Glendora, CA, Lot# 219-1412). The five experimental groups
dium, and other polymers [4,6–8]. Epoxy resin coatings have (20 wires/group) used 0.016  0.022-inch GM (GUMMETAL®; JM
demonstrated good adhesion, chemical resistance, and dimen- Ortho Corporation Lot # 200109 M) as the core wire, coated with
sional stability, as well as being electrically insulating [2]. PTFE five different medically compliant materials: epoxy, PTFE, clear
coatings are non-toxic, non-reactive, and resistant to heat, with ceramic, white ceramic, and silicone.
a low coefficient of friction [2,9,10]. However, these coatings
may deteriorate, exposing the underlying metal, and the colour Coating of wires
of the coating may change over time [4,11]. Such materials have The core wire was 0.016  0.022-inch GM and the goal was to
been applied to nickel titanium (NiTi) and stainless steel (SS) have the coatings add dimension to the wire to reach to
wires [4,6,7,9,12]. A new B-titanium archwire material has 0.019  0.025-inch. The coating of wires was performed by
recently been marketed for use in orthodontics commercially Applied Medical Coatings (AMC, Lockport, NY, USA). First, the
known as GUMMETAL® (GM; Rocky Mountain Morita Corpora- wires were ultrasonic cleaned for seven minutes. Then, they
tion, Tokyo, Japan). were cleaned with deionized water and a generic cleaning
GM has a chemical composition of titanium-niobium-tantalum- solution. The wires were air-dried, and plasma treated for five
zirconium (Ti 59%, Nb 36%,Ta 2%, Zr 3%) [13]. It is flexible, minutes with an oxygen and hydrogen mixture. At this stage,
super-elastic, possesses high yield strength, low Young's mod- the wires were ready for the coating process.
ulus, low coefficient of friction and high spring-back with no The epoxy coating (MACROPOXY® HS High Solids Epoxy; Sherwin
hysteresis [13–15]. GM is also easier to bend compared with Williams, Cleveland, OH) started with hanging the wires on both
other titanium alloy wires [13]. More recently, an aesthetic GM ends of the arch on a custom-made apparatus. A two-part (Part
plated with rhodium has been marketed. Albawardi et al. in an A B58-400 and Part B B58V400) mixture was sprayed with a
in vitro study found that rhodium-plated GM exhibited higher spray cup gun at 26 psi for 10 seconds on each side of the wires
frictional forces than uncoated GM [16]. Upon SEM imaging, in a room at 228C and allowed to dry.
rhodium-plated GM showed signs of cracking and flaking indi- The PTFE (Xlyan® 8820HR/E3216B, Whiteford Corporation,
cating that there is a need for an improved aesthetic GM wire. In Elverson, PA) coating started with hanging the wires on both
addition, the high friction of the rhodium-plated GM limits its ends of the arch on a custom-made apparatus. Then it was spray
application for sliding mechanics. treated with a spray cup gun at 26 psi for 10 seconds on each
There has not been a study that has attempted to coat GM side of the wires in a room at 228C. Subsequently, the wires
with other materials such as epoxy resin, or PTFE, two were oven baked at 4008C for 12 minutes, then air cooled.
commonly used wire coating materials. There is also limited Both the clear ceramic (Cerakote MC-156. High Gloss Ceramic Clear,
evidence on wire coating materials that are not commonly CERAKOTE®; White City, OR) and white ceramic (Cerakote H-Series.
used such as ceramic and silicone. The aim of this study was Part A, CERAKOTE®; White City, OR) coatings were applied using
to evaluate the mechanical, physical, and aesthetic proper- the same process. Both coatings were sprayed with a spray cup
ties of GM wires coated with multiple aesthetic materials in gun at 26 psi for 10 seconds on each side of the wires in a room at
comparison to uncoated GM and SS. The study hypothesized 228C. The clear ceramic coating was then allowed to air dry to
that there will be no difference in frictional forces, force complete the coat. However, the white ceramic coating was baked
deflection rate, yield strength, aesthetic properties, and for three hours at 1208C and then allowed to air dry.
corrosive and wear resistance among the coated GM wires, For the silicone (Dow Corning® MDX4-4159 Fluid 50% Medical
between coated and uncoated GM, and between coated GM Grade Dispersion, Dow Corning®, Midland, MI) coating, the
and uncoated SS. wires were dipped in a silicone mixture (3% silicone and

tome 21 > n82 > June 2023

2
International Orthodontics 2023; 21: 100753

Original article
97% isopropyl alcohol and mineral spirits) for 30 seconds and 10 mm/minute and static friction was recorded using Emperor
then air dried. The wires were once again dipped in the silicone Force Software (Mecmesin, West Sussex, UK). Each wire was pulled
mixture for 30 seconds and air dried. After all coatings were three times before the next sample was loaded. Brackets were
applied, the wires were inspected with a light microscope at replaced after five friction tests.
10  to check for contaminates and then repackaged. SEM was used to qualitatively examine the surface quality (wear of
All wires then underwent assessment of their frictional forces, force coating) before and after the friction test. Representative samples of
deflection rate (three-point bending), yield strength, aesthetics, each wire group were selected at random. The wire surfaces were
corrosive resistance, and wear resistance. For the friction and three- imaged and analysed with a field emission SEM (HitachiSU-70 FE-
point bending tests, the straight distal portion of the natural arch SEM, Tokyo, Japan) at 20.0 kV, at 80  , 500  ,
form was used. The wires were marked to designate left and right 2000  magnification. In addition, energy dispersive x-ray spec-
of each arch form. The left side of the arch was used for the three- troscopy (EDS) analysis was performed to characterize surface
point bending test and the right side for the friction test. The elemental composition.
dimensions of the control wires and coated wires were measured
Three-point bending test
using a Unitron BiS-3214 traveling microscope (Unitron®, Com-
The design of the three-point bending test followed ISO standard
mack, NY) and 3  magnification objective. All procedures and
15841.2014 for wires used in orthodontics [17]. The test was
data collection were performed by one investigator (A.Z.).
performed with the same universal testing machine described
previously. To adhere to the ISO standard, the indenter and fulcrum
Friction forces
radii needed to be at 0.1  0.05 mm. The bulk of the indenter and
Two types of 0.022-inch slot McLaughlin, Bennett, and Trevisi (0-
fulcrum were 3D printed using Formlabs Grey Pro Resin with a
degree torque) prescription (MBTTM) maxillary right canine twin
Formlabs Form 3 printer (Formlabs, Somerville, MA); while the
brackets were used, metal (Victory SeriesTM, 3 M Unitek, Mon-
contacting surfaces between the indenter and fulcrum and the
rovia, CA, LOT#MG5YD) and ceramic (Clarity SeriesTM, 3 M Unitek,
wires were made of square cross-section SS wire with corners of
Monrovia, Calif. LOT#MJ1AE). The brackets were oriented at a
appropriate radii The span of the wire that was tested was 10 mm
passive (0-degree) bracket-wire orientation with the use of a 3D
with a crosshead speed of 1.25 mm/min at room temperature
printed jig with 0-degree slots made with a Formlabs Form
(258C) to a deflected distance of 3.1 mm before the crosshead
3 printer (Formlabs, Somerville, MA). A 3''  1.5''  0.125''
returned to its original position. Each wire group was tested twenty
precision ground aluminium plate (McMaster-Carr, Aurora,
times using twenty different wires. A graph was plotted with the X-
OH) was aligned with the jig using indexed mating surfaces
axis as distance (mm) and the Y-axis as force (Newtons). The force
and temporarily attached to said 3D printed mounting jig with
deflection rate and yield strength were extrapolated from the
elastic bands to maintain its position. Straight SS wires were
generated graph. The force deflection rate was obtained by taking
then attached to the brackets with elastic ligatures. Then, the
the slope between 0.25 mm and 0.75 mm. The yield strength was
straight SS wire of each assembly was placed into a 0-degree
recorded by calculating the y intercept of a parallel line 0.1 mm
slot of the jig, aligning the bracket of this assembly at 0-
offset (along the x-axis) from the initial starting point.
degrees, where it was then adhered to the aluminium plate
with Henkel LOCTITE SF 713TM adhesive accelerator and Henkel Colourimetry
LOCTITE superglue Gel Control. After adhesion, the plate was The coated and uncoated wires were compared to a selected
then removed from the 3D printed jig and transferred to the tooth colour (A1) from a prosthodontic tooth shade guide
testing fixture; where a combination of the straight wires and (Chromascop®, Ivoclar, Amherst, NY) with the use of a colour-
precision ground indexed surfaces of the aluminium plate were imeter (Konica Minolta Spectrophotometer CM-2600d, Konica
used to align the brackets to the grip of the testing Minolta, Ramsey, NJ). From each group, 15 randomly selected
fixture, where the plate was clamped in place. The straight wires were cut into 60 mm lengths from one distal end of the
wires were then exchanged for the wires to be tested. wire toward the arch and stacked next to each other to fill the
The coated wires were tied to each bracket with clear elastomeric view of the spectrophotometer's aperture. Colour parameters
modules (3MTM AlastiKTM Easy-To-Tie Ligatures Ormco, Glendora, using L*a*b* colour space system where L* axis denoting bright-
CA). Afterwards, the aluminium plates with the brackets were ness, a* representing the degree of redness (+a*) or greenness
placed on a custom setup then a Dillion Quantrol TC2i universal (–a*) and b* representing the degree of yellowness (+b*) or
testing machine (Dillon, Fairmont, MN) was used for the friction blueness (–b*) of the wires [18]. Delta E values were calculated
test. Two weights (100 g and 50 g) adding up to 150 g weight according to ASTM D2244-16 [19], to determine the difference
were hung to simulate clinical forces of canine retraction. The wire, between wire colour and A1 shade guide colour1 using the
2 2 2 2
and bracket interaction was brushed with artificial saliva (Pickering formula: DE  ab ¼ ½ðDLÞ þ ðDaÞ þ ðDbÞ 
Laboratories, Mountain View, CA. LOT#2201043) three times after The colour of the coated wires and controls was gathered with
loading but before testing. The wire was pulled at a rate of specular component excluded (SCE) [20]. The settings had an

tome 21 > n82 > June 2023

3
A. Zhou, S. Makowka, S. Warunek, M-Y Chen, T. Al-Jewair

Original article
observer angle at 2-degrees [20], using the daylight D65 illu- pH of the fluids was taken with Labrat Supplies universal plastic
minant. Data were collected with both UV at 0% and 100%. A pH test strips (Labrat Supplies, Parkton, NC). All tubes were then
DE* greater than 3.7 was considered a colour difference that placed in a 378C incubator for seven days. After incubating, the
could be recognized clinically [21]. pH, weight, and colour of the wires were remeasured to evaluate
changes from the initial measurements in the wires or fluids.
Corrosion resistance Statistical analysis
Three representative wire samples, 8 mm in length, one from Data were analyzed using R (v4.0.4) through RStudio (v.461)
each group (n = 27) were randomly selected. Each wire was (RStudio, Boston, MA). Data normality were calculated with
separated into three testing conditions: air, water and corrosive Shapiro-Wilk test. The static friction, force deflection rate and
saliva (ISO 10271-2011; pH 2.3; Pickering Laboratories, Mountain yield strength of the wires were analyzed with one-way analysis
View, CA. LOT#2203049). of variance (ANOVA) followed by pairwise comparison between
Using a colourimeter (Konica Minolta Spectrophotometer CM- each wire using Tukey's honestly significant differences (HSD)
200d/2500d), the wires were individually measured on a black with the significance set at P < 0.05. The correlation between
background (black light trap) that absorbed 98% of light. After- wire dimension and static friction, force deflection rate and yield
wards, the mass of each wire was measured with a Satorius strength were analyzed with multiple testing adjusted P-value
balance model H51 (Satorius, Goettingen, Germany). Each wire using the less conservative Benjamini-Hochberg method.
was placed into separate 50 mL conical centrifuge tubes labelled
air, water, and saliva. The tubes labelled water were filled with Results
50 mL of deionized water; the tubes labelled saliva were filled Static friction
with 50 mL of corrosive saliva; and the tubes labelled air were The mean static friction of each wire group is presented in table
only sealed. Two 50 mL conical centrifuge tubes did not have I. In the metal bracket group, the white ceramic coated wire
wires in them and were blanks for water and corrosive saliva. The exhibited the highest average frictional force (2.06  0.20N),

TABLE I
Mean static friction of all wire groups according to bracket type

Bracket type Wire type Mean (n) SD P-valuea

Metal Clear ceramic 1.37 0.27 < 0.001


White ceramic 2.06 0.20

Epoxy 1.22 0.10

PTFE 1.28 0.38

Silicone 0.88 0.12

GM 16  22 1.10 0.12

GM 19  25 1.32 0.11

SS 16  22 1.22 0.10

SS 19  25 1.19 0.08

Ceramic Clear ceramic 0.93 0.19 < 0.001


White ceramic 0.98 0.18

Epoxy 0.94 0.12

PTFE 0.76 0.12

Silicone 0.68 0.11

GM 16  22 0.89 0.06

GM 19  25 1.17 0.14

SS 16  22 0.94 0.12

SS 19  25 1.17 0.12

a
One-way ANOVA by bracket type; significance set at P < 0.05.

tome 21 > n82 > June 2023

4
International Orthodontics 2023; 21: 100753

Original article
while the lowest was the silicone coating (0.88  0.12N). In the  0.14 N/mm) and the lowest was white ceramic (7.74
ceramic bracket group, both 0.019  0.025-inch GM and SS  0.17 N/mm). Table S-IV presents the yield strength pairwise
wires had the highest average friction (1.17  0.014N and comparisons.
1.17  0.012N, respectively) while the lowest was also the
silicone coated wire (0.68  0.11N). The one-way ANOVA test
showed significant differences between the nine wire groups for Correlation analysis between wire dimensions and
both bracket types, P < 0.001. Pairwise comparisons using frictional forces, force deflection rate, and yield
Tukey's HSD are presented in tables S-I and II. strength
Table III presents wire dimensions measured on the traveling
Force deflection rate microscope. The only coating that reached the target
The highest mean force deflection rate was exhibited by the 0.019  0.025-inch was the epoxy coating. The white and clear
0.019  0.025-inch SS (31.29 N/mm  0.76N) and the lowest ceramic coatings and PTFE dimensions were approximately
by the 0.016  0.022-inch GM (6.71  0.2 N/mm) as shown in 0.018  0.024-inch. The silicone coating was very thin and
table II. Among the coated wires, PTFE showed the highest did not have a measurable difference.
values (9.03  0.12 N/mm) and white ceramic the lowest Correlation was weak and not significant between wire dimen-
(6.86  0.14 N/mm). One-way ANOVA test indicated a signifi- sion and frictional forces in metal and ceramic brackets for all
cant difference in force deflection rates between all wire groups, wire groups (table IV). There was also a non-significant weak
P < 0.001. Pairwise comparisons are presented in table S-III. correlation between wire dimensions and force deflection rate
All wire types showed a significant difference in yield strength for all wire groups except clear ceramic which showed a strong
relative to each other (P < 0.001). The highest average yield and statistically significant correlation (table V, r = 0.67,
strength exhibited among the coated wires was PTFE (10.0 P = 0.014).

TABLE II
Mean force deflection rate and yield strength of all wire groups

Force deflection rate Yield strength


a
Wire type Mean (N/mm) SD P-value Mean (N/mm) SD P-valuea

Clear ceramic 7.41 0.10 < 0.001 8.22 0.15 < 0.001

White ceramic 6.86 0.14 7.74 0.17


Epoxy 7.55 0.07 8.72 0.21
PTFE 9.03 0.12 10.00 0.14
Silicone 7.58 0.25 8.30 0.22
GM 16  22 6.71 0.20 7.60 0.14
GM 19  25 11.62 0.15 12.29 0.12
SS 16  22 19.68 0.31 19.25 0.20
SS 19  25 31.29 0.76 29.77 0.41
a
One-way ANOVA; significance set at P < 0.05

tome 21 > n82 > June 2023

5
A. Zhou, S. Makowka, S. Warunek, M-Y Chen, T. Al-Jewair

Original article
TABLE III
Mean wire cross-section area in inches

Wire type Type Mean SD

Clear ceramic Width 0.024 0.000683

Thickness 0.018 0.000800

White ceramic Width 0.024 0.000151

Thickness 0.018 0.000331

Epoxy Width 0.025 0.000566

Thickness 0.019 0.000620

PTFE Width 0.024 0.000373

Thickness 0.017 0.000419

Silicone Width 0.022 0.000183

Thickness 0.016 0.000267

GM 16  22 Width 0.022 0.000086

Thickness 0.016 0.000107

GM 19  25 Width 0.025 0.000068

Thickness 0.019 0.000112

SS 16  22 Width 0.022 0.000100

Thickness 0.016 0.000105

SS 19  25 Width 0.025 0.000185

Thickness 0.019 0.000083

tome 21 > n82 > June 2023

6
International Orthodontics 2023; 21: 100753

Original article
TABLE IV
Correlation between wire friction and cross-section area according to bracket type

Bracket type Wire type r 95% CI Lower bound 95% CI Upper bound Adjusted P-valuea

Metal Clear ceramic 0.11 –0.56 0.69 0.905

White ceramic –0.48 –0.85 0.22 0.500

Epoxy –0.22 –0.75 0.48 0.905

PTFE –0.53 –0.87 0.15 0.500

Silicone 0.21 –0.48 0.74 0.905

GM 16  22 0.25 –0.45 0.76 0.905

GM 19  25 0.57 –0.09 0.88 0.500

SS 16  22 0.11 –0.56 0.69 0.905

SS 19  25 0.06 –0.59 0.67 0.905

Ceramic Clear ceramic 0.47 –0.23 0.85 0.500

White ceramic 0.51 –0.18 0.86 0.500

Epoxy –0.09 –0.68 0.57 0.905

PTFE 0.54 –0.14 0.87 0.500

Silicone –0.23 –0.75 0.46 0.905

GM 16  22 –0.28 –0.78 0.42 0.905

GM 19  25 0.16 –0.52 0.72 0.905

SS 16  22 0.04 –0.61 0.65 0.917

SS 19  25 0.14 –0.54 0.71 0.905

a
Adjusted P-value = multiple testing adjusted P-value using less conservative Benjamini-Hochberg method.

tome 21 > n82 > June 2023

7
A. Zhou, S. Makowka, S. Warunek, M-Y Chen, T. Al-Jewair

Original article
TABLE V
Correlation between force deflection rate, yield strength and cross-section area

Measurement Wire type r 95% CI Lower bound 95% CI Upper bound Adjusted P-valuea

Force deflection rate Clear ceramic 0.67 0.32 0.86 0.014

White ceramic 0.43 –0.02 0.73 0.389

Epoxy 0.19 –0.28 0.58 0.619

PTFE 0.02 –0.43 0.46 0.937

Silicone 0.40 –0.05 0.72 0.389

GM 16  22 –0.23 –0.61 0.23 0.619

GM 19  25 0.19 –0.28 0.58 0.619

SS 16  22 –0.28 –0.64 0.19 0.619

SS 19  25 0.27 –0.20 0.63 0.619

Yield Strength Clear Ceramic –0.38 –0.70 0.08 0.509

White Ceramic 0.52 0.10 0.78 0.184

Epoxy 0.01 –0.43 0.45 0.961

PTFE 0.02 –0.43 0.46 0.961

Silicone 0.06 –0.39 0.49 0.961

GM 16  22 0.25 –0.22 0.62 0.726

GM 19  25 0.22 –0.25 0.60 0.726

SS 16  22 –0.09 –0.51 0.37 0.961

SS 19  25 0.08 –0.38 0.51 0.961

a
Adjusted P-value = multiple testing adjusted P-value using less conservative Benjamini-Hochberg method; significance set at P < 0.05.

tome 21 > n82 > June 2023

8
International Orthodontics 2023; 21: 100753

Original article
Colour analysis compared to A1 shade Corrosion resistance
Table VI presents the colour analysis of all wire groups and their Evaluations of the mass (g), pH and colour before and after
comparison to A1 shade guide. For UV at 100%, the PTFE coated the corrosion test under corrosive saliva, water and air
wires had the lowest DE* at 12.94. The highest DE* was silicone showed that after the passing of one week, the mass (g)
coated wires at 25.99. For UV at 0%, the PTFE coated wires had the of the wires and pH of the solution did not change. Similarly,
lowest DE* at 14.12 while 0.016  0.022-inch and 019  0.025- the colour of the wires did not have a noticeable
inch SS wires had the highest (DE* of 54.42 and 53.76, respec- change except for the epoxy coating which had the highest
tively). Overall, all wires had a DE* greater than 10 and none of the DE* of around 7 in both water and corrosive saliva (tables S-
coated wires were close to a tooth shade of A1. V–VII).

TABLE VI
Colour analysis of all wire groups when compared to A1 shade

UV Type Wire type L* Mean a* Mean b* Mean DE

UV100 SCE A1 72.20 0.08 11.82 NA

A2 70.36 –0.20 12.95 2.18

Clear Ceramic 46.38 0.52 3.48 27.14

White Ceramic 65.69 –2.16 –0.62 14.21

Epoxy 83.25 –1.78 –1.31 17.25

PTFE 71.69 –1.76 –0.98 12.94

Silicone 47.60 0.52 3.46 25.99

GM 16  22 51.16 2.15 4.24 22.47

GM 19  25 52.72 0.53 4.13 20.95

SS 16  22 49.15 –0.28 4.18 24.29

SS 19  25 49.15 –0.28 4.18 24.29

UV0 SCE A1 62.06 –46.12 0.01 NA

A2 61.40 –45.70 –0.39 0.87

Clear Ceramic 40.80 –30.91 –6.40 26.91

White Ceramic 56.37 –38.50 –11.88 15.23

Epoxy 74.96 –47.90 –15.23 20.05

PTFE 63.92 –42.16 –13.41 14.12

Silicone 43.64 –31.48 –6.72 24.47

GM 16  22 45.41 –33.49 –6.83 21.98

GM 19  25 46.10 –33.72 –7.17 21.45

SS 16  22 38.97 1.89 11.12 54.42

SS 19  25 39.61 2.33 6.23 53.76

tome 21 > n82 > June 2023

9
A. Zhou, S. Makowka, S. Warunek, M-Y Chen, T. Al-Jewair

Original article
Wear resistance, surface morphology, and chemical appeared to be uniform and smoother. The surface texture of the
composition silicone coated GM was almost identical to uncoated GM. The
SEM imaging of all representative wires before friction testing visually observed appearance from smoothest to roughest of
showed that the GM surface appeared rough with noticeable coated wires were, clear ceramic, PTFE, epoxy, silicone, and
pockets and fissures (figure 1). The SS wire was smooth and white ceramic.
mostly uniform with some minor fissures. The clear ceramic Figures 2 and 3 present the wires after they underwent friction
coating was uniform and smooth. The white ceramic coating had testing through metal and ceramic brackets, respectively. Over-
large pockets, was not uniform and appeared rough. The epoxy all, there was very little wear and no noticeable breaks after
coated surface appeared to be uniform, but the coating had a sliding the wires through both metal and ceramic brackets. The
slight bumpy appearance. The PTFE coating, similar to epoxy, EDS composition analysis is shown in table S-VIII.

Figure 1
SEM images (20.0 kV; 500  original magnification) of representative samples before any testing (a) clear ceramic; (b) white ceramic;
(c) epoxy; (d) PTFE; (e) silicone; (f) GUMMETAL®; (g) stainless steel

tome 21 > n82 > June 2023

10
International Orthodontics 2023; 21: 100753

Original article
Figure 2
SEM images (20.0 kV; 500  original magnification) of representative samples after friction testing with metal brackets (a) clear
ceramic; (b) white ceramic; (c) epoxy; (d) PTFE; (e) silicone; (f) GUMMETAL®; (g) stainless steel

tome 21 > n82 > June 2023

11
A. Zhou, S. Makowka, S. Warunek, M-Y Chen, T. Al-Jewair

Original article
Figure 3
SEM images (20.0 kV; 500  original magnification) of representative samples after friction testing with ceramic brackets (a) clear
ceramic; (b) white ceramic; (c) epoxy; (d) PTFE; (e) silicone; (f) GUMMETAL®; (g) stainless steel

Discussion friction testing. The rhodium-plated GM is slight grey and thus,


The study explored how coating GM with various materials not considered a truly aesthetic white archwire.
affects its mechanical, physical and aesthetic properties. The Before any mechanical testing of the wires, the surface char-
most common materials used to coat orthodontic wires are acteristics and aesthetics of the wires were assessed with a
epoxy resins and PTFE [2,7,11,22,23]. There are also ceramic Spectrophotometer and with SEM. The uncoated GM SEM images
coatings that typically used on orthodontic brackets to provide showed a rough surface with pits and fissure similar to Alba-
an aesthetic appearance [24,25]. Thus, in addition to epoxy and wardi et al.'s findings [16]. SEM images of SS wires showed a
PTFE, this study explored ceramic coatings and silicone on GM. uniform and smooth surface. The clear ceramic coating, when
These coatings can affect the surface quality and therefore first viewed with the eye, had a slightly glossy appearance but
impact the aesthetics, wear resistance, corrosion resistance, otherwise showed a similar appearance as uncoated GM. Under
and mechanical properties [1,4,7]. the SEM, the clear ceramic coating had the most uniform
Currently, there is an aesthetic rhodium-plated GM that the appearance with no noticeable pits or fissures.
manufacturer released in 2020. Albawardi et al. [16] found that From the colour analysis, the white ceramic DE* had a distin-
the frictional forces of aesthetic GM was higher than uncoated guishable difference in colour from an A1 shade. In addition, the
GM and SS wires and the coating cracked and flaked upon white ceramic coating was not consistent with each wire. Each

tome 21 > n82 > June 2023

12
International Orthodontics 2023; 21: 100753

Original article
white ceramic coated wire had slightly different shades of The white ceramic coating exhibited the most frictional force
white. The colourimeter measurement was an average of the among the coated wires with both bracket types. When compared
different shades with the white ceramic. The SEM images of the to uncoated GM and SS, white ceramic increased the frictional force
white ceramic showed the least favourable surface quality of 0.016  0.022-inch uncoated GM in both bracket types. Also, its
condition because it had the largest fissures and deepest pits. frictional forces with metal brackets was almost double that of the
This likely has led to higher frictional forces with metal and SS wire groups. This can partly be attributed to the surface rough-
ceramic brackets and was the highest among coated wires. The ness of the white ceramic coating material. Both PTFE and silicone
silicone coating was very thin but was detectable by the EDS coatings exhibited lower frictional forces than both uncoated GM
analysis. The outer appearance of a silicone coated wire looked control groups with ceramic brackets. Farronato et al. found that
like an uncoated GM wire, and thus was not aesthetic. PTFE coatings reduced friction compared to their uncoated coun-
Orthodontic arch wires coated with epoxy and PTFE are com- terpart [10]. In addition, the coated wires had either similar or less
mercially available [2,6,22,26]. frictional force than both SS wires.
There is no commercially available GM coated with epoxy and The force deflection rate of the uncoated 0.016  0.022-inch GM
PTFE. From the colourimetry measurements, similar to white was slightly higher than reported by Suzuki et al. who found the
ceramic, the epoxy was very distinguishable from the A1 shade. force deflection rate of uncoated 0.016  0.022-inch GM to be
Of the aesthetic coatings, the epoxy's white colour was the most below 3 N/mm [15]. Murguma et al. showed that the elastic
consistent among aesthetic wires. The SEM images of the epoxy modulus of coated aesthetic orthodontics wires was lower than
coating showed a more uniform surface texture since there were that of uncoated wires [6]. In this study, all coated wires pro-
less fissures than uncoated GM but a slight bumpy texture, duced a higher force deflection rate than the uncoated
possibly from filler material. The PTFE's white colour was con- 0.016  0.022-inch GM. The coated wires used in this study
sistently closest to A1 shade based on DE*in all colourimeter were different from those used by Murguma et al. and the
settings. The PTFE coating added more uniformity to the surface dimensions of the wires were also variable. Both SS wires
with less deep pits and fissures than uncoated GM. However, exhibited much higher force deflection rates than any of the
flaking was seen in the PTFE coating. coated or uncoated GM wires. The increase of force deflection
In this study, static friction force was recorded for all groups. rate was about 1 N/mm for every coating material besides PTFE.
Overall, the frictional forces were low. Kopsahilis et al. reported PTFE resulted in the most increase in wire stiffness.
similar frictional force levels when testing frictional behaviour of The yield strength was calculated from the stress/strain curve
GM [27]. Contributing factors to low frictional forces relates to 0- produced by the three-point bending test with a 0.1 mm offset.
degree angulation of the bracket and wire. This made the wire A noticeable change in the wire after the three-point bending
interact more with the elastic ligatures and the base of the test was the cracking and peeling of all but silicone coatings at
bracket than with the bracket wings [27]. Some variation with the fulcrum area.
frictional forces could be related to the wire having some binding The corrosive resistance testing performed in this study was a
between the bracket wings. The wire and bracket ligation were qualitative examination. Corrosion resistance is important in
achieved with elastic ties rather than SS ligatures. The wire and orthodontics with respect to potential leaking of metal ions into
ligature interface could have experienced binding while the wire the oral cavity [31–33]. The corrosive saliva used in this study had
is being pulled through the bracket. Once static friction has been a pH of 2.3. This test was to determine how the wires performed
overcome, a snapping action may have occurred with the ligature in an extreme environment like corrosive saliva. The coated
and this process could repeat multiple times. This binding and wires did not exhibit major loss of mass or noticeable peeling of
snapping may have contributed to the noise the universal the coating layers. However, the colourimeter detected some
machine detects while recording the frictional force. colour change with the epoxy coating in saliva and water
When compared according to bracket type, surprisingly, the environments. Commercial aesthetic orthodontic wires have
wires experienced more friction with the metal brackets than been found to change colour overtime intraorally [8]. However,
with the ceramic brackets. Tanne et al. noted that ceramic limited inferences can be made with this test because it did not
brackets were rougher than metal brackets which led to canine measure potential leaked metal ions in solution. Although Kim
retraction being slowed by 30% to 50% [28]. et al. has demonstrated that epoxy coated NiTi reduces release
On the other hand, Cha et al. demonstrated that newer ceramic of nickel ions [33], future studies should investigate this process.
brackets with a coated silica layer performed similarly to conven- From this study, coating material provides benefits and limitations
tional metal brackets [29]. This study used an artificial saliva for all when compared to uncoated 0.016  0.22-inch GM (table VII). The
friction tests to simulate some intra oral conditions. Stannard et al. epoxy coatings allowed for a more aesthetic wire but increased
found that artificial saliva itself could increase the frictional forces friction and stiffness. The PTFE coating allowed for a more aesthetic
between objects because the artificial salvia is a polar liquid wire and reduced friction but increased stiffness. A thin silicone
creating more attraction between the two surfaces [30]. coating helped reduce friction, but the wire remained metallic

tome 21 > n82 > June 2023

13
A. Zhou, S. Makowka, S. Warunek, M-Y Chen, T. Al-Jewair

Original article
TABLE VII
Added benefits and effects of coating materials on 0.016  0.022-inch GM wire

Coating material Friction on metal Friction on ceramic Force deflection rate Yield strength Esthetics Corrosion resistance Wear resistance

Clear ceramic + + + +   +

White ceramic + + + + +  +

Epoxy + + + + +  +

PTFE + – + + +  +

Silicone – – + +   +

"+'': Increase; "–'': Decrease; "'': stayed relatively the same.

looking with a slight increase in stiffness. The clear ceramic coating include different angulations. A quantitative corrosion resistance
provided a homogenous smooth surface but there was no aes- test should be conducted. This can be achieved by measuring
thetic benefit or reduction in stiffness and friction. The white the ions before and after coated wires are soaked in a corrosive
ceramic provided a white colour wire, but the rough surface texture solution to determine the degree of leeching. This study has
resulted in increased friction. All the coatings increased the force shown that coating wires can improve some aspects of the wire
deflection rate and yield strength when compared to the base properties but not all. Finally, orthodontists need to weigh the
0.016  0.22-inch GM. However, this study showed that the benefits of coated orthodontic wires and associated manufactur-
coated wires had similar or less frictional force than ing costs.
0.016  0.22-inch and 0.019  0.25-inch SS wires.
All coating materials provided corrosion resistance as well as
Conclusion
wear resistance to sliding against a bracket. However, these The various coatings provide benefits and limitations when
coating materials cracked and peeled when a sharp bend was compared to uncoated GM and SS. Coated wires had lower
placed in the wire with the three-point bending test. The results frictional forces than uncoated SS wires when used with ceramic
of this study reject the null hypothesis that there is no difference brackets. However, coated wires had comparable frictional
in mechanical, aesthetic, and physical properties between the forces with uncoated GM and SS wires with metal brackets.
different coating materials, between coated GM and uncoated Coated GM wires had good wear resistance after friction testing
GM, and between coated GM and SS. and provided corrosion resistance to low pH saliva. Future
This study has limitations. The coating protocols varied according investigation of the materials used in this study is required to
to the type of coating material applied. Some of the coating further characterize their properties.
methods required high intensity heat such as the PTFE and white
Ethical standard: For this study, no human participants or animals were
ceramic coatings. The high intensity heat may have impacted recruited by any of the authors.
the core-based GM wire properties such as its shape memory.
Funding: The study was supported by a grant from the Japan Association
Coated orthodontic wires have been found to reach a target wire of Adult Orthodontists.
size with the inner core metal alloy being smaller to fit both the
coating and metal alloy [6,12]. The wires used for coating in this Contribution: Alan Zhou: conceptualization, methodology, data curation,
investigation, writing- original draft preparation.
study had a core wire of 0.016  0.022-inch. The goal was to Steven Makowka: investigation, writing- reviewing and editing.
make all coatings consistent in application to increase the core Stephen Warunek: writing- reviewing and editing. ming chen: writing-
reviewing and editing.
wire to the same final dimension, but the coatings between
Thikriat Al-Jewair: conceptualization, methodology, data curation, supervi-
groups were of different sizes. Silicone produced the least sion, writing- reviewing and editing.
thickness while epoxy added the most which might have
Disclosure of interest: The authors declare that they have no competing
affected the study outcomes. The rhodium-plated GM was diffi- interest.
cult to obtain due to shipping complications. Therefore, this wire
was not used. Finally, the materials used in this study were
medically compliant but additional testing is needed before Supplementary data
intra-oral dental use. Supplementary data associated with this article
Future studies are needed to further explore coated GM wires. can be found, in the online version, at doi:10.
The study examined frictional forces when the bracket and wire 1016/j.ortho.2023.100753.
angulation was zero degrees. Future frictional studies should

tome 21 > n82 > June 2023

14
International Orthodontics 2023; 21: 100753

Original article
References
[1] Bandeira AM, dos Santos MP, Pulitini G, Elias orthodontic coated archwires. Am J Orthod [22] Batista DM, Faccini M, Valarelli FP, et al.
CN, da Costa MF. Influence of thermal or Dentofacial Orthop 2013;143:S85–91. Attractiveness of different aesthetic ortho-
chemical degradation on the frictional force of [13] Hasegawa S. A concept of en bloc movement dontic wires. Dental Press J Orthod
an experimental coated NiTi wire. Angle of teeth using gummetal wire. Tokyo: Quin- 2020;25:27–32.
Orthod 2011;81:484–9. tessence Publishing; 2014. [23] Elayyan F, Silikas N, Bearn D. Ex vivo surface
[2] Kravitz ND. Aesthetic archwire. Orthod Pro- [14] Nordstrom B, Shoji T, Anderson WC, et al. and mechanical properties of coated ortho-
ducts 2013 [Accessed June 2021: https:// Comparison of changes in irregularity and dontic archwires. Eur J Orthod 2008;30:661–7.
www.kravitzorthodontics.com/assets/pdfs/ transverse width with nickel-titanium and [24] Angolkar PV, Kapila S, Duncanson Jr MG,
Aesthic-archwires.pdf]. niobium-titanium-tantalum-zirconium arch- Nanda RS. Evaluation of friction between
[3] Kusy RP. A review of contemporary archwires: wires during initial orthodontic alignment in ceramic brackets and orthodontic wires of
their properties and characteristics. Angle adolescents: A double-blind randomized clin- four alloys. Am J Orthod Dentofacial Orthop
Orthod 1997;67:197–207. ical trial. Angle Orthod 2018;88:348–54. 1990;98:499–506.
[4] da Silva DL, Mattos CT, Simão RA, de Oliveira [15] Schmeidl K, Janiszewska-Olszowska J, Gro- [25] Matias M, Freitas MR, Freitas KMS, Janson G,
Ruellas AC. Coating stability and surface char- cholewicz K. Clinical features and physical Higa RH, Francisconi MF. Comparison of
acteristics of aesthetic orthodontic coated properties of gummetal orthodontic wire in deflection forces of aesthetic archwires com-
archwires. Angle Orthod 2013;83:994–1001. comparison with dissimilar archwires: a criti- bined with ceramic brackets. J Appl Oral Sci
[5] Krishnan M, Seema S, Kumar AV, et al. cal review. Biomed Res Int 2018;26:e20170220.
Corrosion resistance of surface modified 2021;2021:6611979. [26] Rongo R, Valletta R, Bucci R, et al. In vitro
nickel titanium archwires. Angle Orthod [16] Albawardi A, Warunek S, Makowka S, Al- biocompatibility of nickel-titanium aesthetic
2013;84:358–67. Jewair T. Friction forces generated by aes- orthodontic archwires. Angle Orthod
[6] Muguruma T, Iijima M, Yuasa T, Kawaguchi thetic Gummetal(R) (Ti-Nb) orthodontic arch- 2016;86:789–95.
K, Mizoguchi I. Characterization of the coat- wires: a comparative in vitro study. Int Orthod [27] Kopsahilis IE, Drescher D. Friction behavior of
ings covering aesthetic orthodontic archwires 2022;20:100683. the wire material Gummetal®. J Orofac
and their influence on the bending and fric- [17] International Organization for Standardiza- Orthop 2022;83:59–72.
tional properties. Angle Orthod 2017;87:610– tion. Dentistry – Wires for use in orthodontics. [28] Tanne K, Matsubara S, Shibaguchi T, Sakuda
7. Geneva, Switzerland: BSI Standards Publica- M. Wire friction from ceramic brackets during
[7] Iijima M, Muguruma T, Brantley W, et al. tion; 2014 [Accessed July 2021. https:// simulated canine retraction. Angle Orthod
Effect of coating on properties of aesthetic www.iso.org/standard/62223.html]. 1991;61:285–90 [discussion 91-2].
orthodontic nickel-titanium wires. Angle [18] Inami T, Tanimoto Y, Minami N, Yamaguchi [29] Cha JY, Kim KS, Hwang CJ. Friction of con-
Orthod 2012;82:319–25. M, Kasai K. Colour stability of laboratory ventional and silica-insert ceramic brackets in
[8] da Silva DL, Mattos CT, de Araújo MV, de glass-fiber-reinforced plastics for aesthetic various bracket-wire combinations. Angle
Oliveira Ruellas AC. Colour stability and fluor- orthodontic wires. Korean J Orthod Orthod 2007;77:100–7.
escence of different orthodontic aesthetic 2015;45:130–5. [30] Stannard JG, Gau JM, Hanna MA. Compara-
archwires. Angle Orthod 2013;83:127–32. [19] Subcommittee E12.04 on colour and appear- tive friction of orthodontic wires under dry
[9] Kameda T, Sato H, Oka S, Miyazaki A, ance analysis. Standard practice for calcula- and wet conditions. Am J Orthod
Ohkuma K, Terada K. Low temperature tion of colour tolerances and colour 1986;89:485–91.
polytetrafluoroethylene (PTFE) coating differences from instrumentally measured [31] Sugisawa H, Kitaura H, Ueda K, et al. Corro-
improves the appearance of orthodontic colour coordinates. ASTM-D2244-2022 EDI- sion resistance and mechanical properties of
wires without changing their mechanical TION. Accessed July 20, 2021: https:// titanium nitride plating on orthodontic wires.
properties. Dent Mater J 2020;39:721–34. www.astm.org/d2244-21.html. 2022. Dent Mater J 2018;37:286–92.
[10] Farronato G, Maijer R, Carìa MP, Esposito L, [20] Alghazali N, Preston A, Moaleem M, Jarad F, [32] Ba̧cela J, Labowska MB, Detyna J, Ziȩty A,
l

Alberzoni D, Cacciatore G. The effect of Teflon Aldosari AA, Smith P. The effects of different Michalak I. Functional coatings for orthodon-
coating on the resistance to sliding of ortho- spectrophotometric modes on colour mea- tic archwires-a review. Materials (Basel)
dontic archwires. Eur J Orthod 2012;34:410–7. surement of resin composite and porcelain 2020;13:3257. doi: 10.3390/ma13153257.
[11] Kaphoor AA, Sundareswaran S. Aaesthetic materials. Eur J Prosthodont Restor Dent PMID: 32707959; PMCID: PMC7435379.
nickel titanium wires–how much do they deli- 2018;26:163–73. [33] Kim H, Johnson JW. Corrosion of stainless
ver? Eur J Orthod 2012;34:603–9. [21] Johnston WM, Kao EC. Assessment of appear- steel, nickel-titanium, coated nickel-titanium,
[12] da Silva DL, Mattos CT, Sant' Anna EF, Ruellas ance match by visual observation and clinical and titanium orthodontic wires. Angle Orthod
AC, Elias CN. Cross-section dimensions and colourimetry. J Dent Res 1989;68:819–22. 1999;69:39–44.
mechanical properties of aesthetic

tome 21 > n82 > June 2023

15

You might also like