Correcting astigmatism with toric intraocular


lenses: Effect of posterior corneal astigmatism
Douglas D. Koch, MD, Richard B. Jenkins, MD, Mitchell P. Weikert, MD, Elizabeth Yeu, MD,
Li Wang, MD, PhD

PURPOSE: To evaluate the impact of posterior corneal astigmatism on outcomes with toric
intraocular lenses (IOLs).
SETTING: Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA.
DESIGN: Case series.
METHODS: Corneal astigmatism was measured using 5 devices before and 3 weeks after cataract
surgery. Toric IOL alignment was recorded at surgery and at the slitlamp 3 weeks postoperatively.
The actual corneal astigmatism was calculated based on refractive astigmatism 3 weeks postoper-
atively and the effective toric power calculated with the Holladay 2 formula. The prediction error was
calculated as the difference between the astigmatism measured by each device and the actual
corneal astigmatism. Vector analysis was used in all calculations.
RESULTS: With the IOLMaster, Lenstar, Atlas, manual keratometer, and Galilei (combined Placido–
dual Scheimpflug analyzer), the mean prediction errors (D) were, respectively, 0.59 @ 89.7, 0.48 @
91.2, 0.51 @ 78.7, 0.62 @ 97.2, and 0.57 @ 93.9 for with-the-rule (WTR) astigmatism (60 to 120
degrees), and 0.17 @ 86.2, 0.23 @ 77.7, 0.23 @ 91.4, 0.41 @ 58.4, and 0.12 @ 7.3 for against-
the-rule (ATR) astigmatism (0 to 30 degrees and 150 to 180 degrees). In the WTR eyes, there
were significant WTR prediction errors (0.5 to 0.6 diopters [D]) by all devices. In ATR eyes,
WTR prediction errors were 0.2 to 0.3 D by all devices except the Placido–dual Scheimpflug
analyzer (all P<.05 with Bonferroni correction).
CONCLUSIONS: Corneal astigmatism was overestimated in WTR by all devices and underestimated
in ATR by all except the Placido–dual Scheimpflug analyzer. A new toric IOL nomogram is proposed.
Financial Disclosure: Drs. Koch, Weikert, and Wang received research support from Ziemer USA,
Inc. Dr. Koch has a financial interest with Alcon Laboratories, Inc., Abbott Medical Optics, Inc.,
Optimedica Corp., and Ziemer USA, Inc. No other author has a financial or proprietary interest in
any material or method mentioned.
J Cataract Refract Surg 2013; -:-–- Q 2013 ASCRS and ESCRS

Accurate measurement of total corneal astigmatism in minimize postoperative corneal and ocular astigma-
cataract patients is crucial for achieving optimum tism. Unfortunately, several studies suggest that cur-
postoperative uncorrected visual acuity and patient rent methodology is inadequate for achieving
satisfaction, especially with the implantation of toric optimum astigmatic outcomes. In 3 studies of results
intraocular lenses (IOLs). Traditionally, keratometers with toric IOLs,2–4 the mean postoperative refractive
and Placido-disk corneal topographers have been astigmatism after implantation of a toric IOL ranged
used to measure the curvature of the anterior surface from 0.72 diopter (D) G 0.43 (SD) to 1.03 G 0.79 D.
of the cornea and corneal power and astigmatism Traditionally, it has been assumed that the posterior
values have been calculated by assuming a fixed pos- cornea induces minimal refractive astigmatism and
terior:anterior curvature ratio. Since Jaffe and Clay- can therefore be ignored in astigmatic calculations.
man1 first analyzed the relationship between cataract However, several studies using different methodolo-
surgical technique and refractive results in individual gies5–11 recently reported that the posterior cornea
patients in 1975, ophthalmic surgeons have calculated has astigmatism that ranges from 0.26 to 0.78 D. In
surgically induced refractive change in an attempt to our recent study,11 we found that in most eyes, the

Q 2013 ASCRS and ESCRS 0886-3350/$ - see front matter 1


Published by Elsevier Inc. http://dx.doi.org/10.1016/j.jcrs.2013.06.027
2 CORNEAL ASTIGMATISM PREDICTION ERROR

posterior corneal steep meridian is aligned vertically 1. IOLMaster (Carl Zeiss Meditec AG). This PCI device
and proposed a theory that for toric IOL implantation, measures corneal power by automated keratometry and
uses data from a hexagonal array of 6 points reflected
ignoring the posterior corneal astigmatism would
off the surface of the cornea at a diameter of approxi-
result in overcorrection in eyes having with-the-rule mately 2.5 mm, depending on the corneal curvature.
(WTR) anterior corneal astigmatism and undercorrec- 2. Lenstar (Haag-Streit AG). This optical low-coherence
tion in eyes with against-the-rule (ATR) astigmatism. reflectometry (OLCR) device calculates keratometry
Validation of this theory in patients having toric IOL (K) from an array of 32 light reflections projected off the
anterior corneal surface. Sixteen measuring points on
implantation is desirable.
each eye are arranged in 2 rings at diameters of approxi-
The purposes of this study were (1) in eyes with toric mately 1.65 mm and 2.30 mm, respectively, again depend-
IOLs, to evaluate the prediction errors of total corneal ing on the corneal curvature.
astigmatism with 5 devices, 1 of which uses Scheimp- 3. Atlas corneal topographer (Carl Zeiss Meditec AG). This
flug imaging and ray-tracing analysis to measure Placido disk–based corneal topographer provides simu-
lated K values along the steepest and flattest meridians
anterior and posterior corneal astigmatism and 4 of
at the 3.0 mm zone.
which measure the anterior surface only, and (2) to 4. Manual keratometer (Bausch & Lomb). Manual keratom-
provide a nomogram for selecting the appropriate etry is the conventional method for measuring corneal
toric power for astigmatic correction that factors in power at a diameter of approximately 3.0 mm, depending
posterior corneal astigmatism. on the corneal curvature.
5. Galilei Placido–dual Scheimpflug analyzer (Ziemer
PATIENTS AND METHODS Ophthalmic Systems AG). The anterior corneal measure-
ments are made by a proprietary method of merging
Patients Placido and Scheimpflug data, and the posterior corneal
Institutional review board approval was obtained for surface is constructed using Scheimpflug data. The device
this study. Prospectively, patients who were willing to calculates the total corneal power (TCP) using ray tracing,
participate in this study and met the following inclusion which propagates incoming parallel rays and uses Snell
criteria were enrolled from July 2011 to September 2012: law to refract the rays through the anterior and posterior
(1) good quality scans with all 5 devices, (2) no previous corneal surfaces.
ocular trauma or surgery, (3) no corneal or other ocular Of the 5 devices, the first 4 systems measure anterior
disease, (4) no contact lens wear within 2 weeks before the corneal curvature only and astigmatism values were calcu-
corneal measurements, and (5) Acrysof toric IOL (models lated from the differences between the steep K and flat
SN6AT3 to SN6AT8, Alcon Laboratories, Inc.) implantation K values based on these anterior corneal data. For the
without postoperative IOL decentration or tilt under the Placido–dual Scheimpflug analyzer, the TCP astigmatism
slitlamp examination. One eye from each patient was value, which is the difference between the steep TCP and
included in this study. Any eye with more than 5 degrees flat TCP at the 1.0 to 4.0 mm central zone, was used. Corneal
of IOL rotation from the intended meridian was excluded astigmatism was measured with all 5 devices at the preoper-
from the study. Patients with oblique corneal astigmatism ative and 3-week postoperative visits.
(steep corneal meridian at 30 to 60 degrees or 120 to 150 Biometry was performed using the PCI device and the
degrees) measured by partial coherence interferometry OLCR device. The Holladay 1 formula was used for toric
(PCI) were also excluded because of the small number of IOL power calculation. Selection of the toric IOL power
eyes in that group. and alignment was determined by the surgeons based on
all data available and a previous study of posterior corneal
Corneal Astigmatism Measurements astigmatism.11 Three surgeons (D.D.K., M.P.W., E.Y.)
performed the surgeries through 2.4 mm clear corneal inci-
The following 5 devices were used for corneal astigmatism sions. The axis of the alignment of the toric IOL was recorded
measurements in this study: at the time of surgery and at the slitlamp during the 3-week
postoperative visit. Manifest refraction was performed
3 weeks postoperatively.
Submitted: April 29, 2013.
Final revision submitted: June 12, 2013. Data Analysis
Accepted: June 14, 2013.
Eyes were divided into 2 groups depending on the
From the Cullen Eye Institute, Department of Ophthalmology, anterior corneal steep meridian measured by the PCI device
Baylor College of Medicine, Houston, Texas, USA. as follows: (1) WTR group with corneal steep meridian at
60 to 120 degrees and (2) ATR group with corneal steep
Supported in part by an unrestricted grant from Research to meridian at 0 to 30 degrees or 150 to 180 degrees.
Prevent Blindness, New York, New York, USA. Vector analysis was used in all calculations.12 To account
for the impact of IOL power and anticipated effective lens
Presented in part at the ASCRS Symposium on Cataract, IOL and position (ELP), the effective toric power of the IOL at the
Refractive Surgery, Chicago, Illinois, USA, April 2012. corneal plane was calculated using the Holladay 2 Consul-
tant Program (Holladay Consulting). The assumed actual
Corresponding author: Douglas D. Koch, MD, Department of corneal astigmatism was calculated as the difference
Ophthalmology, Baylor College of Medicine, 6565 Fannin, NC205, between the postoperative manifest refraction corrected to
Houston, Texas 77030, USA. E-mail: dkoch@bcm.edu. the corneal plane and the effective toric power. Then, for

J CATARACT REFRACT SURG - VOL -, - 2013


CORNEAL ASTIGMATISM PREDICTION ERROR 3

each device, the corneal astigmatism prediction error, or the astigmatism was 0.08  11 and 0.12  148 in the
deviation from actual corneal astigmatism, was obtained by WTR group and ATR group, respectively.
subtracting the actual corneal astigmatism from the corneal
astigmatism measured by each device.
Analysis of aggregate corneal astigmatism prediction er- Aggregate Corneal Astigmatism Prediction Errors
rors was performed. Using preoperative and postoperative With the 4 devices that measure anterior corneal
corneal astigmatism measurements, preoperative and
curvature only, the mean preoperative and postopera-
postoperative corneal astigmatism prediction errors were as-
sessed for each device. The corneal astigmatism prediction tive corneal astigmatism prediction errors in WTR
errors were further analyzed as follows: (1) WTR/ATR eyes ranged from 0.27 to 0.62 D, all aligned along
prediction errors, which are displayed along the horizontal the vertical meridian (Figure 1, A to D) and in ATR
axis in the double-angle plot of the prediction errors, with eyes ranged from 0.17 to 0.37 D, also aligned along
negative values indicating WTR prediction errors (the device
the vertical meridian (Figure 1, A to D). With the
overestimates corneal WTR astigmatism compared with
the actual amount) and positive values, ATR errors, and combined Placido–dual Scheimpflug analyzer TCP,
(2) oblique prediction errors, which are displayed along the mean preoperative and postoperative corneal
the vertical axis in the double-angle plot of the prediction astigmatism prediction errors were 0.57 @ 93.9 and
errors, with positive values indicating oblique astigmatism 0.26 @ 97.0, respectively, in the WTR eyes and 0.12 @
prediction errors along 45 degrees and negative values,
7.3 and 0.18 @ 8.7, respectively, in the ATR eyes
along 135 degrees.
(Figure 1, E).
Statistical Analysis
With-the-Rule/Against-the-Rule and Oblique
A sample-size calculation was performed to detect a
Corneal Astigmatism Prediction Errors
corneal astigmatism prediction error of more than 0.2 D.
The primary data showed that the standard deviation In the WTR group, the WTR/ATR prediction errors
of corneal astigmatism prediction error was approximately ranged from 0.60 to 0.26 D and oblique prediction
0.4 D. With a significance level of 5% and a test power errors, from 0.26 to C0.20 D (Table 2). There were
of 80%, 32 eyes were required.
To assess whether the prediction errors were WTR/ATR significant WTR prediction errors (0.5 to 0.6 D) by all
or oblique, a 1-sample t test was performed to evaluate devices except the Placido-disk corneal topographer
whether the mean vector component values were signifi- and combined Placido–dual Scheimpflug analyzer
cantly different from zero. Bonferroni correction was used postoperative measurements. There was a significant
for multiple comparisons. Statistical analysis was performed oblique prediction error of 0.26 D by the OLCR
using SPSS for Windows software (version 15.0, SPSS, Inc.).
A probability of less than 5% (P!.05) was considered statis- device for postoperative corneal astigmatism mea-
tically significant. surements (all P!.05 with Bonferroni correction).
In the ATR group, the WTR/ATR prediction errors
RESULTS ranged from 0.29 to C0.17 D and oblique prediction
errors, from 0.13 to C0.36 D. There were significant
Forty-one eyes of 41 patients were enrolled. The mean
WTR prediction errors of 0.2 to 0.3 D by the PCI device,
age of the 18 men and 23 women was 71 G 9 years
OLCR device, Placido-disk corneal topographer, and
(range 46 to 91 years). The mean IOL power was 18.0
manual keratometer and oblique prediction errors of
G 6.0 D (range 6.0 to 28.0 D). Table 1 shows the
0.3 to 0.4 D by manual keratometry corneal astigma-
mean preoperative corneal astigmatism measured by
tism (all P!.05 with Bonferroni correction). There
each device. The mean postoperative refractive
was no significant WTR/ATR or oblique prediction
error by the combined Placido–dual Scheimpflug
Table 1. Preoperative corneal astigmatism (D) measured by
analyzer.
each device.
DISCUSSION
Device WTR (n Z 17) ATR (n Z 24)
Accurate correction of astigmatism requires accurate
PCI 1.78 @ 91 1.28 @ 1 measurements. In this study, in eyes implanted with
OLCR 1.66 @ 92 1.24 @ 2 toric IOLs, we assessed the prediction errors of total
Placido-disk corneal 1.66 @ 88 1.22 @ 0 corneal astigmatism using 4 devices that estimate total
topographer TCP corneal astigmatism based on anterior surface
Combined Placido–dual 1.76 @ 93 1.56 @ 1 measurements only and 1 device that measures both
Scheimpflug
anterior and posterior corneal astigmatism.
Manual keratometry 1.91 @ 94 1.32 @ 8
To account for the ELP variations in different
ATR Z against the rule; OLCR Z optical low-coherence reflectometry; eyes, we used the Holladay 2 Consultant Program to
PCI Z partial coherence interferometry; TCP Z total corneal power;
calculate the effective toric power of the IOL at the
WTR Z with the rule
corneal plane. Based on actual postoperative refractive

J CATARACT REFRACT SURG - VOL -, - 2013


4 CORNEAL ASTIGMATISM PREDICTION ERROR

Figure 1. Double-angle plots of preoperative and postoperative


corneal astigmatism prediction errors in the WTR eyes (left)
and ATR eyes (right) using the PCI device (A), OLCR device
(B), Placido-disk corneal topographer (C), manual keratometry
(D), and combined Placido–dual Scheimpflug analyzer TCP
(E). (ATR Z against the rule; PE Z prediction error; WTR Z
with the rule).

astigmatism, the actual total corneal astigmatism was 0.3 D WTR in eyes with ATR corneal astigmatism for
determined; the corneal astigmatism prediction error preoperative and postoperative corneal measure-
for each device was then calculated. ments. In a recent study,11 we found that (1) the poste-
With the 4 devices that had measurements from the rior cornea is steeper along the vertical meridian in
anterior corneal surface only, the mean corneal more than 85% of eyes and (2) the magnitude of poste-
astigmatism prediction errors were 0.5 to 0.6 D WTR rior corneal astigmatism increased with increasing
in eyes with WTR corneal astigmatism and 0.2 to amounts of anterior corneal astigmatism in WTR

J CATARACT REFRACT SURG - VOL -, - 2013


CORNEAL ASTIGMATISM PREDICTION ERROR 5

Table 2. Mean corneal astigmatism prediction errors (D): WTR/ATR and oblique by each device.

Device

PCI OLCR Placido-Disk Topographer Manual K Placido–Dual Scheimpflug

Parameter Preop Postop Preop Postop Preop Postop Preop Postop Preop Postop

WTR
WTR/ATR 0.59* 0.52* 0.48* 0.48* 0.47* 0.27 0.60* 0.46* 0.57* 0.26
Oblique C0.01 0.17 0.02 0.26* C0.20 0.10 0.15 0.25 0.08 0.06
ATR
WTR/ATR 0.17* 0.29* 0.21* 0.21* 0.23* 0.29* 0.18 0.26* C0.11 C0.17
Oblique C0.02 C0.02 C0.10 0.07 0.01 0.13 C0.36* C0.26* C0.03 C0.05

ATR Z against the rule; K Z keratometry; OLCR Z optical low-coherence reflectometry; PCI Z partial coherence interferometry; WTR Z with the rule
*Significantly different from zero (all P!.05 with Bonferroni correction)

eyes; whereas in ATR eyes, the average magnitude exceeds the refractive astigmatism; however, when
of posterior corneal astigmatism was approximately the axis is closer to ATR at 180 degrees, the refractive
0.2 D and did not change with increasing amounts of astigmatism has the greater value.
anterior corneal astigmatism. This shows that the total The difference between refractive and keratometric
corneal astigmatism prediction errors from these astigmatism has been attributed to an intraocular
devices were primarily caused by the posterior corneal astigmatic component, which is believed to be pro-
astigmatism. duced mainly by the lens (ie, lenticular astigmatism).
With the Galilei TCP values, which account for both However, 2 recent studies2,23 analyzed refractive and
anterior and posterior corneal astigmatism, we found keratometric astigmatism in pseudophakic eyes and
in the WTR group a significant WTR prediction error found that the relationship described by the simplified
of 0.57 D using preoperative corneal astigmatism; Javal rule holds true. Because the posterior cornea is a
however, in the ATR group, there were no significant negative lens, refractive astigmatism with the steep
WTR/ATR prediction errors. This indicates that meridian aligned vertically on the posterior surface
the Galilei TCP may still underestimate the posterior produces net plus refractive power along the horizon-
corneal astigmatism in WTR eyes. Further study eval- tal meridian. These findings point to posterior corneal
uating the accuracy of posterior corneal astigmatism astigmatism as the link between refractive and kerato-
measurements with this device is desirable. The Penta- metric astigmatism and the explanation of the simpli-
cam corneal topography system (Oculus, Inc.) also fied Javal rule.
uses a rotating Scheimpflug camera to measure ante- There are some limitations to this study. We of
rior and posterior corneal curvatures. Studies assess- course did not actually measure the astigmatism
ing the prediction errors of total corneal astigmatism induced by the toric IOL, so we cannot rule out effects
with this device in eyes with toric IOLs are desirable. from IOL tilt or variances in the actual toricity of

In 1890, Emile Javal13 first characterized this rela- the IOLs. We did not measure actual total corneal
tionship by publishing the Javal rule, which states astigmatism but inferred it from other pertinent
that the total astigmatism in the eye, or refractive astig- clinical data. We do not have a sound explanation
matism, is equal to 1.25 (keratometric astigmatism) for the greater-than-anticipated differences between
0.50  90. In 1988, Grosvenor et al.,14 using more the actual corneal astigmatism and the values calcu-
than 600 eyes with refractive and keratometric data, lated by the Galilei device. Possible explanations for
derived a simplified version of the Javal rule as this disparity are device inaccuracy, flaws in our meth-
follows: refractive astigmatism Z (keratometric astig- odology for calculating the actual corneal astigmatism,
matism) 0.50  90. Several authors have since or some IOL tilt was present, creating an ATR refrac-
confirmed the relationship described by the simplified tive effect. However, it seems highly unlikely that tilt
Javal rule with data from hundreds of eyes from could be the source of the magnitude of prediction
diverse populations ranging from Native American errors found in our patients because tilt was not detect-
school children to aging Hong Kong Chinese.15–21 able at the slitlamp examination.
In 1997, Alpins22 made a similar observation that In summary, our study found that using the devices
when the topographic astigmatism axis is closer to that calculate total corneal astigmatism based on ante-
90 degrees, the topographic astigmatism magnitude rior corneal measurements only, WTR astigmatism

J CATARACT REFRACT SURG - VOL -, - 2013


6 CORNEAL ASTIGMATISM PREDICTION ERROR

Table 3. Baylor toric IOL nomogram (target range up to 0.40 D


REFERENCES
WTR). Values in the table are the vector sum of the anterior 1. Jaffe NS, Clayman HM. The pathophysiology of corneal astig-
corneal and surgically induced astigmatism. Examples: (1) If matism after cataract extraction. Trans Am Acad Ophthalmol
the cornea has 3.70 D WTR and surgically induced astigmatism Otolaryngol 1975; 79:OP615–OP630
is 0.20 D WTR, use the value of 3.9 D to select IOL toricity. (2) If 2. Teus MA, Arruabarrena C, Herna ndez-Verdejo JL, Sales-
the cornea has 1.90 D ATR and surgically induced astigmatism is Sanz A, Sales-Sanz M. Correlation between keratometric and
0.20 D WTR, use the value of 1.70 D to select IOL toricity. refractive astigmatism in pseudophakic eyes. J Cataract
Refract Surg 2010; 36:1671–1675
Effective IOL Cylinder 3. Sun X-Y, Vicary D, Montgomery P, Griffiths M. Toric intraocular
Power at Corneal Plane (D) WTR (D) ATR (D) lenses for correcting astigmatism in 130 eyes. Ophthalmology
2000; 107:1776–1781; discussion by RM Kershner, 1781–1782
0.00 %1.69 (PCRI if O1.00) !0.39 4. Mendicute J, Irigoyen C, Aramberri J, Ondarra A, Monte s-
1.00 1.70–2.19 0.40*–0.79 Mico  R. Foldable toric intraocular lens for astigmatism correction
1.50 2.20–2.69 0.80–1.29 in cataract patients. J Cataract Refract Surg 2008; 34:601–607
2.00 2.70–3.19 1.30–1.79 5. Royston JM, Dunne CMM, Barnes DA. Measurement of poste-
2.50 3.20–3.79 1.80–2.29 rior corneal surface toricity. Optom Vis Sci 1990; 67:757–763.
Available at: http://journals.lww.com/optvissci/Abstract/1990/
3.00 3.80–4.39 2.30–2.79
10000/Measurement_of_Posterior_Corneal_Surface_Toricity.
3.50 4.40–4.99 2.80–3.29
2.aspx. Accessed September 26, 2013
4.00 5.00– 3.30–3.79 6. Dunne MCM, Royston JM, Barnes DA. Posterior corneal surface
ATR Z against the rule; IOL Z intraocular lens; PCRI Z peripheral
toricity and total corneal astigmatism. Optom Vis Sci 1991;
corneal relaxing incision; WTR Z with the rule 68:708–710. Available at: http://journals.lww.com/optvissci/Ab-
*Especially if spectacles have more ATR stract/1991/09000/Posterior_Corneal_Surface_Toricity_and_
Total.6.aspx. Accessed September 26, 2013
7. Ho J-D, Tsai C-Y, Liou S-W. Accuracy of corneal astigmatism
estimation by neglecting the posterior corneal surface measure-
ment. Am J Ophthalmol 2009; 147:788–795
was overestimated by 0.5 to 0.6 D and ATR astigma- 8. Dubbelman M, Sicam VA, van der Heijde GL. The shape of the
tism was underestimated by 0.2 to 0.3 D. A toric IOL anterior and posterior surface of the aging human cornea. Vision
nomogram accounting for posterior corneal astigma- Res 2006; 46:993–1001
tism is provided (Table 3). This nomogram takes 9. Prisant O, Hoang-Xuan T, Proano C, Hernandez E, Awwad ST,
into account the mean values of posterior corneal Azar DT. Vector summation of anterior and posterior corneal
topographical astigmatism. J Cataract Refract Surg 2002;
astigmatism that we found and aims to leave eyes 28:1636–1643
with small amounts of WTR refractive astigmatism. 10. Mo dis L Jr, Langenbucher A, Seitz B. Evaluation of normal
The rationale for the latter is the ongoing ATR shift corneas using the scanning-slit topography/pachymetry system.
that appears to occur in most eyes, even after tempo- Cornea 2004; 23:689–694
ral clear corneal cataract surgery.24 Further studies 11. Koch DD, Ali SF, Weikert MP, Shirayama M, Jenkins R, Wang L.
Contribution of posterior corneal astigmatism to total corneal
evaluating the performance of this nomogram are astigmatism. J Cataract Refract Surg 2012; 38:2080–2087
desirable. 12. Holladay JT, Moran JR, Kezirian GM. Analysis of aggregate
surgically induced refractive change, prediction error, and intra-
ocular astigmatism. J Cataract Refract Surg 2001; 27:61–79
WHAT WAS KNOWN 13. Javal E. Me moires d’Ophtalmome  trie: Annote
s et Pre
cedes
d’une Introduction. Paris, France, G. Masson, 1890; 131
 Astigmatism is present on the posterior corneal surface,
14. Grosvenor T, Quintero S, Perrigin DM. Predicting refractive
which is independent in magnitude and meridian from astigmatism: a suggested simplification of Javal’s rule. Am J
the astigmatism of the anterior surface. Optom Physiol Opt 1988; 65:292–297
15. Grosvenor T, Ratnakaram R. Is the relation between keratomet-
 Residual postoperative astigmatism occurs often in eyes ric astigmatism and refractive astigmatism linear? Optom Vis Sci
with toric IOL implantation. 1990; 67:606–609. Available at: http://journals.lww.com/opt-
vissci/Abstract/1990/08000/Is_the_Relation_between_Kerato-
metric_Astigmatism.9.aspx. Accessed September 26, 2013
WHAT THIS PAPER ADDS 16. Remo n L, Benlloch J, Furlan WD. Corneal and refractive astig-
matism in adults: a power vectors analysis. Optom Vis Sci
 In eyes with toric IOL implantation, corneal astigmatism 2009; 86:1182–1186. Available at: http://journals.lww.com/opt-
prediction errors with devices that measure anterior vissci/Fulltext/2009/10000/Corneal_and_Refractive_Astigma-
corneal astigmatism only were 0.5 to 0.6 D WTR in WTR tism_in_Adults__A.9.aspx. Accessed September 26, 2013
eyes and 0.2 to 0.3 D WTR in ATR eyes, showing the effect 17. Tong L, Carkeet A, Saw S-M, Tan DTH. Corneal and refrac-
tive error astigmatism in Singaporean schoolchildren: a
of posterior corneal astigmatism.
vector-based Javal’s rule. Optom Vis Sci 2001; 78:881–887.
 A nomogram is proposed for toric IOL selection based on Available at: http://journals.lww.com/optvissci/Abstract/2001/
12000/Corneal_and_Refractive_Error_Astigmatism_in.10.aspx.
preoperative corneal astigmatism.
Accessed September 26, 2013

J CATARACT REFRACT SURG - VOL -, - 2013


CORNEAL ASTIGMATISM PREDICTION ERROR 7

18. Dobson V, Miller JH, Harvey EM. Corneal and refractive astig- 23. Bae JG, Kim SJ, Choi YI. Pseudophakic residual astigmatism.
matism in a sample of 3- to 5-year-old children with a high prev- Korean J Ophthalmol 2004; 18:116–120. Available at: http://
alence of astigmatism. Optom Vis Sci 1999; 76:855–860. ekjo.org/Synapse/Data/PDFData/0065KJO/kjo-18-116.pdf. Ac-
Available at: http://journals.lww.com/optvissci/Abstract/1999/ cessed September 26, 2013
12000/Corneal_and_Refractive_Astigmatism_in_a_Sample_of. 24. Hayashi K, Hirata A, Manabe S, Hayashi H. Long-term change in
22.aspx. Accessed September 26, 2013 corneal astigmatism after sutureless cataract surgery. Am
19. Auger P. Confirmation of the simplified Javal’s rule [letter]. Am J J Ophthalmol 2011; 151:858–865
Optom Physiol Opt 1988; 65:915
20. Keller PR, Collin MJ, Carney LG, Davis BA, van Saarloos PP.
The relation between corneal and total astigmatism. Optom
First author:
Vis Sci 1996; 73:86–91. Available at: http://eprints.qut.edu.au/
3215/1/3215.pdf. Accessed September 26, 2013 Douglas D. Koch, MD
21. Lam AKC, Chan CCY, Lee MH, Wong KM. The aging effect on Cullen Eye Institute, Department of
corneal curvature and the validity of Javal’s rule in Hong Kong Ophthalmology, Baylor College of
Chinese. Curr Eye Res 1999; 18:83–90
Medicine, Houston, Texas, USA