MANUAL
HIGH VOLTAGE BREAKDOWN TESTERS (FLASH TESTERS)
Oto SKY, 0t030mA 214/217
Oto 10KV, 0to30mA 215/218
Oto SKV,0to100mA 214A/219
High Voltage Breakdown teater is useful for
checking breakaown (tiasn-over) vortage
of motors, transformers, pumps, chokes, coils,
switches, connectors, plastics, rubber, PCB's,
insulation materials, electronic components etc.
Useful for industries, railways, electricity boards,
defence, communication centers, R&D
Organizations, lavoratorles, colleges, services.
centres ete.
Automatic tripping facility, HT ON and HT OFF
visual indications, HT ON buzzer, high voltage
safety probe with push button enable features,
provide operator's safety.
HT voltage and leakage current meter provide
continuous reading, simultaneously.
TEST VOLTAGE:
Oto SKV Ain 214/217 ond 214A/219,
O10 10KVAC in 215/218.
Test voltage is displayed on wide view moving coil
type of panel meter. Reading accuracy is within + 5%
TRIPPING CURRENT LEVELS:
Oto 30mA in 214/217 and 215/218.
Oto 100 mAin 214/219.
Tripping current levels can be set from 0
to maximum by front panel control.
SAFETY FEATURES: Whenever mains supply of
high voltage tester is turnad ON, high voltage
romaine OFF and is enabled only when pushbutton
‘on safety probe is kept pressed and only when
actual value of leakage currents less than the
tipping current level setting
HT ON indication glows and buzzer sounds when
high voltage. is on. Acrylic probe whit safety skirt
keeps operataor's hand away from high voltage.
Once high voltage is tripped, tho same romains off
until push button of the high voltage is reloased
and pressed again.
POWER REQUIREMENTS: 230 volt AC, +
10%,50160 Hertz
SIZE AND WEIGHT: 200mm (H) x 250mm (W) x
290mm (D). 12,14 and 45 kgs. respectively, for
244217,2157218 aud 2140/2109.
STANDARD ACCESSORIES :
Instruction manual,
SAFETY PRECAUTIONS:
High voltages generated in high voltage
breakdowm testers are hazardous, Every necessary
precaution should be taken to avoid hazards. Do
not by pass the safety features provided in the
tester. Do not modify high voltage probe and the
tester. Do not open covers. Do not attempt
sorvicing of the tector. Koop the operater end other
‘material away from the high voltage tip of the
safety probe and from the items under test. Make
all necessary arrancemants to ensure safety.
Always ensure the proper earth connection to the
tester.
‘TEST PROCEDURE:
1, Disconnect supply from the test insulation whose
breakdown flash over is to be tested.2. Keep safety probe of the tester in safe place and
ensure high voltage tip is away from living
creatures and all other material on the work table.
3. Bring variac to minimum position and tripping
level control to maximum position
4. Connect 3 pin plug of the tester to properly
earthed 3 pin outlet of mains supply of 230 volt AC,
50/60 Hertz.
5. Turn on power supply of the tester. Check HT OFF
indication glows, HT ON indication is off and both
meters are at near zero positions.
6. Press push button switch located on the safety
probe. Check HT OFF indication is off, HT ON
indication giows and audio buzzer sounds. HT ON
indication and audio buzzer sound are warnings for
presence of high voltace.
7. increase variac setting slowly until reading of KV
meter equals the value of required test voltage.
8. Touch black ground probe of the tester to one
end of the insulation under test and tip of the
safety probe to the other end of the insulation
under test.
Press push button switch located on the safety
probe for 1 minute or until sound of audie buzzer
goes off.
If audio buzzer sound remains of for 1 minute, the
insulation under test has passed the breakdown
test. If audio buzzer sound goes off before 1
minute, the insulation under test has failed the
breakdown test.
To test another insulation, touch the black ground
probe of the tester to one end of the insulation and
tip of the safety probe to the other end of
the insulation . Release the push button switch of the
safety prohe mamentarily and prass the came again
for 1 minuto or until the audio sound goes off.
9. After all of the test pieces are tested or the
testing is halted, turn off the mains supply of the
tester. Remove 3 pin plug of the tester from the
mains supply and bring variac to minimum location.
This step is necessary to avoid hazards.
PRECAUTIONS:
High voltage breakdown test is destructive test.
The insulation under testis damaged during
hraaledown eanditian Tha failed insiilation under
test should be scrapped and should not be used
again,
‘TRIPPING LEVEL ADJUSTMENT:
Most of the products are toster with maximum
valve of tripping level as per procedure mentioned
atove. However for the devices for which lower
tripping can be set as per the following procedure.
1. Place tripping level in maximum posit
2. Place variac in minimum position.
3. Sart black ground probe and the tip of the safely
probe.
4, Ture on the power of the tester.
'5. Press push button switch of the safety probe.
6. Increase variac setting very slowly unt reading
on mA meter equals the desired tripping level.
7. Decrease setting of tripping level very slowly
until audio buzzer sound goes of.
8. Release push button switch on the safety probe
and remove short of black ground probe and the tip,
of the safety probe.
0. Proce puch button of the safety probe increase
setting and inc:ease variae setting until reading on
KV meter reaches the required test voltage.
‘The tester Is now ready for testing trom step
number 8 onwards mentioned in TEST PROCEDURE
above.