error 26.1 Thu Aug 22 2019

i have error 26.1 at Candela - GentleLASE Pro


device. it appears every day what i can do Non

reboot the device ?

" Report Post    RE: Aesthetic Laser

Reply from Laserway Sun Aug 25 2019

The manual says to turn off fo r30 secs,


then restart. Its a software issue.
Sometimes, reseating the large chip in the
main pcb can help (only do this with
machine off). Often an energy calibration can help it
go away