HumaLyzer PRIMUS

Settings for Multipurpose reagents


Software Version 3.2, Date 2015-06

Kit: BILIRUBIN D+T liquicolor (DPD)


REF: 10740

Procedure for Direct Bilirubin;


Note: 2 vials per sample

Test Parameters DBIL


Method: End Point
Unit: mg/dl
Temperature: 37°C
First Wavelength: 546nm
Second Wavelength: None
Blank: Sample
Factor: 13.0
Norms High: 0.25
Norms Low: 0.01
Delay Time: 10s
Test Time:
Decimal Digits: 2
Aspirate Volume: 500 µl