You are on page 1of 2

HumaCount 5D

| Quick Guide

Cat No. 16450/5-1


1 2
480µl of HC5D- Click “Mode&ID” in the measuring
Diluent are screen and select “Capillary Whole
automatically Blood (CTA)” mode
dispensed into
a bullet tube by
the HC5D.

3
Blood collection by finger prick, and 20µl
HUMAN capillary tube (EDTA coded). After
blood collection is completed, wipe away
excess blood on wet lint free tissue. Avoid
any contact of tissue and blood inside the
capillary tube

For measurements in the

Quick Guide capillary blood mode you


need
for measurements in - HumaCount 5D device
Capillary Blood Mode - Capillary tube (REF 16070/30)
- Bullet tube
- Moist tissue

4
Put the capillary into the bullet tube
filled with Diluent, close the lid and
invert the sample carefully, until the
suspension is homogenous.

6
Results of 29 parameters will
be shown in the measuring

< <
screen

5• Open the lid, insert the bullet tube to the


collection needle and press the big red button
to start the recording. Make sure the needle
always touches the bottom of the bullet tube.
• HC5D automatically draws demanded volume
from the bullet tube

HUMAN Gesellschaft für Biochemica und Diagnostica mbH


Max-Planck-Ring 21 · 65205 Wiesbaden · Germany Rev. 03/2018-10
Tel. +49 6122-9988-0 · Fax +49 6122-9988-100 · e-mail: human@human.de · www.human.de