Professional Documents
Culture Documents
DCA-1
Revision History:
Version Analyser
Date Software-
Manual Comment
(yyyy/mm/dd) Version -
(Art.-No.) Release -
Revision History 2
User Manual DCA-1
Content
1 IMPORTANT INFORMATION 8
Definition OEM / PLD 8
Liability Disclaimer 8
Software Copyrights 9
Intellectual Property 9
Warranty 9
Exclusions of manufacturer’s warranty 10
Guidelines for getting service assistance 11
2 TERMS, ABBREVIATIONS AND SYMBOLS USED IN THIS MANUAL
13
3 SAFETY 16
Duty of care of the user 17
Intended use 17
Reagents 19
Safety instructions 19
Electrical safety 19
Mechanical safety 20
Reasonable foreseeable and remaining risks 20
Emergency and first aid 21
Fire fighting 21
4 INSTALLATION 22
Name Plate (Serial no.) 22
Component overview 23
Software 23
Mounting and connecting the analyzer 24
Thermal block and User interface 26
Content 3
User Manual DCA-1
External printer 28
Connection to an external Printer (Port A) 28
Activate and deactivate the external printer 30
CuvCARDs - Load a cuvette balance (optional) 31
ChipCARDs – Store or load methods 36
Connection to a HOST (Port B) 37
Connecting an external Barcode Scanner 38
5 ANALYZER AND METHOD PARAMETERS 41
Instrument parameters 41
Method memory positions and Default Method Parameters 42
6 OPERATION 44
Measuring principle 44
User qualification 45
Switch on the analyzer 45
STANDBY 47
Method selection 48
Measuring Mode (= “cuvette in”) 48
Measuring Mode (= “Patient ID”) 49
Carry out a PT Measurement 51
Carry out an aPTT Measurement 54
Carry out a Fibrinogen Measurement g/l 58
Parameter configuration - Access WITHOUT PIN 61
Calibration curve 61
Set calibration curve points for PT 62
Set calibration curve points for Fibrinogen 64
Set MNP value 66
Save parameters 67
Parameter configuration - Access WITH PIN 68
Menu <General> 69
6.6.1.1 Method name 69
Content 4
User Manual DCA-1
Content 5
User Manual DCA-1
Content 6
User Manual DCA-1
Content 7
User Manual DCA-1
1 IMPORTANT INFORMATION
Liability Disclaimer
1 Important Information 8
User Manual DCA-1
Software Copyrights
Intellectual Property
Warranty
1 Important Information 9
User Manual DCA-1
1 Important Information 10
User Manual DCA-1
NOTE
Supporting our service department with the basic
information as detailed as possible will help us to
give you the needed assistance. Thank you in
advance.
Customer identification
✓ Customer
✓ Name
✓ Address
✓ Country
✓ Email
✓ Phone
✓ Fax
✓ Distributor / Enduser (in case of enduser, please state the
distributor)
Analyzer identification
✓ Type of analyzer
✓ Serial-Number (SN)
✓ Reference number (REF)
✓ Software version
1 Important Information 11
User Manual DCA-1
Problem description
Describe the problem as detailed as possible. This may clarify the
situation.
✓ Display Messages
✓ Printouts
✓ Screenshots
✓ Photos
Analytical data
In case of analytical problems further information will be needed.
Describe the problem as detailed as possible.
✓ Test type: eg. PT, aPTT, Fibrinogen, ….
✓ Sample (Patient sample, Control N, Control P, dilution, ….)
✓ Reagent Manufacturer, Name of reagent, Order number, Lot
Number, date of expire.
✓ Leaflets
✓ Actual measured results, raw value in seconds, converted
values.
✓ Conversion, Conversion table, ISI, …
1 Important Information 12
User Manual DCA-1
Table 2-1 Symbols used in this Manual, on the analyzer itself, and
on available accessories and consumables
Date of
Manufacturer
manufacture
Do not use if
CE
packaging is
Conformity
damaged
Refer to
Operators Do Not Reuse
Manual
Caution
Protect From
Danger
Sunlight.
Warning
Traceblility
Temperature number
Limitation. according to
customer.
Humidity Important
Limitation. Information.
Electric
Shock Biological Risk.
Warning
Separate
collection,
handling and
disposal for
Keep dry!
waste electrical
and electronic
equipment and
its components
Observe
Precautions for
handling
electrostatic
Do not throw sensitive devices
(ESD) / Warning
of damages by
electrostatic
discharge
In vitro
Diagnostica
Symbol Meaning
Chapters
3 SAFETY
3 Safety 16
User Manual DCA-1
Intended use
WARNING
For in-vitro diagnostic (IVD) use only!
3 Safety 17
User Manual DCA-1
3 Safety 18
User Manual DCA-1
Reagents
For proper coagulation analysis we recommend to use reagents,
controls and buffers supplied by PLD.
Always read the information leaflet in the package and observe
the instructions given by the reagent manufacturer.
NOTE
Utilize reagents and controls only according to
directions as provided by the reagent manufacturer
to avoid incorrect measuring results or malfunction
of the analyzer.
Safety instructions
Electrical safety
3 Safety 19
User Manual DCA-1
Mechanical safety
3 Safety 20
User Manual DCA-1
CAUTION
In case of emergency immediately interrupt the
operation of the analyzer and pull out the external
power supply from the socket.
Fire fighting
3 Safety 21
User Manual DCA-1
4 INSTALLATION
Manufacture date
Type: Analyzer name (according to customer)
CE Conformity
In vitro Diagnostica
Input: Information about voltage and current
4 Installation 22
User Manual DCA-1
Component overview
Software
4 Installation 23
User Manual DCA-1
4 Installation 24
User Manual DCA-1
NOTICE
The analyzer is delivered with an external power
supply.
The analyzer has to be connected exclusively with
the delivered external power supply. Make sure to
have a protected ground at the main socket.
Otherwise a safe operation can not be guaranteed.
We do not warrant against a defect of the analyzer
in case another plug-in power supply than the
delivered has been used.
6. Connect the cable of the power supply to the analyzer and the
external power supply itself to a socket outlet without voltage
surges produced by heavy power users, e.g. lifts and
centrifuges.
7. Ensure easy accessibility of the power plug at the analyzer
rear side and/or at the external power supply at the wall
socket to disconnect the device in case of an emergency.
1 2 3 4
1 USB Port B
2 Serial interface for Barcode-Scanner or external Printer
3 Power supply socket
4 SD card-Slot
4 Installation 25
User Manual DCA-1
Figure 4-3 Thermal block (left area), User interface (right area) and
ChipCARD / CuvCARD reader slot on the right side of the
analyzer
The operation area of the analyzer consists of the heated Thermal
block with
✓ Reagent position
✓ Cuvette positions
✓ Measuring channel
… and of the User interface with
✓ Number keys
✓ Mode / Esc / Enter keys
✓ Start / Reset key
✓ Arrow keys (for navigation)
✓ Display
The function of the keys is described in the following table.
4 Installation 26
User Manual DCA-1
4 Installation 27
User Manual DCA-1
External printer
NOTE
Please read the suppliers manual carefully.
4 Installation 28
User Manual DCA-1
NOTE
Never operate the printer without paper!
Without inserted paper or if the printer function is
deactivated you may not be able to print results and
parameters.
We recommend to purchase and use the printer
exlusively with the thermal paper offered by the
manufacturer.
Operating the printer with another paper might
result in malfunction of the printer with the following
effects:
➢ No paper transport.
➢ Paper abrasion.
➢ No print out.
NOTICE
We do not warrant against a defect of the
printing unit if the use of printer paper other than
recommended is the is the cause of the defect.
NOTE
Mind proper storage conditions for the thermal
printer paper:
✓ Dry environment.
✓ Protect from light.
✓ Protect from heat.
4 Installation 29
User Manual DCA-1
4 Installation 30
User Manual DCA-1
NOTE
Upon delivery the analyzer is preloaded with a
cuvette balance that corresponds to the quantity of
cuvette as delivered in the accessory box.
NOTE
Disposal of the CuvCARD may be done in the
domestic waste. Ensure in your own interest that no
balance remains on the CuvCARD.
4 Installation 31
User Manual DCA-1
➢ Using the CuvCARD you can select, whether the total cuvette
credit balance shall be loaded to the analyzer. Otherwise, if
more analyzers of the same type are available, you can also
load only the necessary number to each analyzer.
NOTE
We recommend to load the total credit balance on
either one analyzer or, if more analyzer are in the
laboratory, split it and load the necessary amount.
REASON: If the CuvCARD gets lost or is damaged
the remaining balance is lost.
4 Installation 32
User Manual DCA-1
2. Insert the CuvCARD face up into the reader slot (memory chip
in the direction of insertion). Leave the CuvCARD in the
reader.
➢ The following display appears.
Display Meaning
4 Installation 33
User Manual DCA-1
Selection Meaning
NOTE
When loading a cuvette balance from the CuvCARD
to the analyzer, the remaining balance eventually
still on the analyzer will be overwritten. Make sure
the balance on the analyzer is fully used prior to
loading a new balance from the CuvCARD!
4 Installation 34
User Manual DCA-1
7. Remove the CuvCARD from the reading unit and the analyzer
automatically turns to the measuring mode “cuvette in”.
NOTE
To print a short overview of method relevand data
e.g. Lot Number, remaining cuvette balance, select
loaded to
dd:mm:yyyy, hh:mm:ss
--CuvCARD Info--
on CuvCARD = xxx
4 Installation 35
User Manual DCA-1
4 Installation 36
User Manual DCA-1
Figure 4-6 USB cable with USB-A (for HOST) and USB-B (for
analyzer)
4 Installation 37
User Manual DCA-1
1 2
4 Installation 38
User Manual DCA-1
8. Press .
9. Press .
4 Installation 39
User Manual DCA-1
4 Installation 40
User Manual DCA-1
NOTE
Before routine tests can be performed, the user
must change certain reagent- specific parameters
such as Lot Number and calibration curve.
Otherwise results have no correct relation to the
actual reagent.
Instrument parameters
<empty>
6 OPERATION
Measuring principle
6 Operation 44
User Manual DCA-1
User qualification
WARNING
It is absolutely necessary that only skilled personnel
will access the parameter menu with a PIN Code as
improper handling of the analyzer might cause
inaccurate measuring results. (see chapter 6.7.7,
page 114 <PIN Code>).
All settings in the parameter menu must be done in
accordance with manufacturer requirements.
Upon alteration a parameter protocol must be printed
in order to check all settings again.
NOTE
Communication with the analyzer is performed via
the liquid crystal display. We assume that you are
familiar with the function of the individual keys as
described in chapter 4.5, page 26.
Contact your local distributor with regard to
adjustment of display contrast if the display cannot
be correctly read.
• Connect the power supply plug to the analyzer and the socket,
respectively.
6 Operation 45
User Manual DCA-1
Warm-up time
NOTE
The analyzer requires approximately 30 minutes to
warm-up the thermal block to an operating
temperature of 37.4°C.
During the warm-up no measurements are possible.
Only switching between menus is possible and
settings can be adjusted.
• Use the warm-up phase to load the analyzer with cuvettes and
reagents for testing. Ensure each cuvette is equipped with a
mixer.
6 Operation 46
User Manual DCA-1
STANDBY
1. Press .
NOTE
Make sure that the cuvette is removed out of the
measuring channel prior to confirm the message,
otherwise the automated cuvette detection and
measuring channel adjustment may not work
properly.
3. Press any key to confirm the message
6 Operation 47
User Manual DCA-1
Method selection
STANDBY is the level from where the user can switch between
and select the necessary method or carry out settings in
UTILITES menu
NOTE
Make sure the cuvette is removed from the
measuring channel prior confirm the message and
settings are correct.
6 Operation 48
User Manual DCA-1
➢ “cuvette in” requests the user to place the cuvette into the
measuring channel.
NOTE
For all tests you can do single determination
measurement, with one result. A mean value
calculation from two single values afterwards is not
supported by the software.
2. Press .
6 Operation 49
User Manual DCA-1
NOTE
If you wait over 30 sec without entering the ID or
with entering the ID but without pressing Enter to
proceed, the following message appears:
<Read Pat-ID Time Out>
Then, the display automatically switches back to
Patient ID display.
6 Operation 50
User Manual DCA-1
NOTE
For the special preparation of Plasma and Reagents
(reconstitution, prewarming) please see the
according Plasma/Reagent leaflet.
Place the PT start reagent vial and a sufficient
number of cuvettes in the thermal block at the same
time as starting the analyzer (or at least 30 min
before starting with the measurement). The
temperature has to be constantly kept at 37,4°C in
the reagent and cuvette positions.
Sample incubation
6 Operation 51
User Manual DCA-1
➢ After the adjustment the following display appears for the PT-
Measurement and requests to add 100 µl start reagent.
6 Operation 52
User Manual DCA-1
Results
PT
Patient _ _ _ _ _
dd:mm:yyyy, hh:mm:ss
No. = 1
Channel = 1
Time = 13.7 s
% = 74.4
INR = 1.18
Figure 6-1 Example print out of PT result
➢ After the print out, “cuvette out” appears in the display.
6 Operation 53
User Manual DCA-1
• Open the light protection cap, take out the cuvette, close the
NOTE
For the special preparation of Plasma and Reagents
(reconstitution, prewarming) we refer to the
according Plasma/Reagent leaflet.
Place the reagent vial (CaCl2) in the thermal block at
the same time as starting the analyzer (or at least 30
min before starting with the measurement). The
temperature has to be constantly kept at 37,4°C.
Sample incubation
6 Operation 54
User Manual DCA-1
6 Operation 55
User Manual DCA-1
Results
6 Operation 56
User Manual DCA-1
aPTT
Patient _ _ _ _
dd:mm:yyyy, hh:mm:ss
No. = 2
Channel = 1
Time = 31.5 s
Ratio = 1.13
Figure 6-2 Example print out of aPTT result
• Open the light protection cap, take out the cuvette, close the
6 Operation 57
User Manual DCA-1
NOTE
For the special preparation of Plasma and Reagents
(reconstitution, prewarming) we refer to the
according Plasma/Reagent leaflet.
Sample incubation
6 Operation 58
User Manual DCA-1
6 Operation 59
User Manual DCA-1
Results
Fib. g/l
Patient _ _ _ _ _
dd:mm:yyyy, hh:mm:ss
No. = 3
Channel = 1
Time = 5.7 s
g/l = 3.87
Figure 6-3 Example print out of Fibrinogen result
• Open the light protection cap, take out the cuvette, close the
6 Operation 60
User Manual DCA-1
Calibration curve
Although method parameters of the analyzer do include preset
values for the calibration curve (of PT and Fibrinogen; %, g/l,
mg/dl and time in sec) it is necessary to adjust the parameters
any time you change the Lot Number of the reagents.
The reason is, that the reagent has lot-specific properties, which
could lead to wrong results when the settings do not match the
reagents results.
Calibration curves are carried out with special calibration plasma
and the new reagent.
NOTE
To set calibration curves /points all relevant methods
have to be set to “ON” under column “Reference
curve” in Table 5-1, page 42.
6 Operation 61
User Manual DCA-1
NOTE
Points on the reference curve that are not to be used
must have the entries 0.0% or 0.0 s.
A minimum of 2 points must be defined. A maximum
of 9 points can be entered.
We recommend defining 4 points (100%, 50%, 25%,
12.5%).
6 Operation 62
User Manual DCA-1
6. Use number keys to enter the clotting time and to set the
decimal point for the respective activity.
12. Press to confirm and set the Lot Number with number
13. Press
6 Operation 63
User Manual DCA-1
6 Operation 64
User Manual DCA-1
6 Operation 65
User Manual DCA-1
NOTE
MNP = Mean of Normal Plasma
6 Operation 66
User Manual DCA-1
Save parameters
Generally, after finishing parameterization or when leaving the
method menus the analyzer checks the parameters and the
following sequence appears.
6 Operation 67
User Manual DCA-1
6 Operation 68
User Manual DCA-1
Menu <General>
Selection Meaning
6 Operation 69
User Manual DCA-1
2. Press to confirm.
6 Operation 70
User Manual DCA-1
NOTE
If you keep the arrow-key pressed the characters
will be displayed faster and you can speed up your
selection.
If you keep the Enter-key pressed the cursor will
move faster.
Delete character
6 Operation 71
User Manual DCA-1
5. Press to confirm.
6 Operation 72
User Manual DCA-1
7. Press to confirm.
6 Operation 73
User Manual DCA-1
6 Operation 74
User Manual DCA-1
6 Operation 75
User Manual DCA-1
example e.g <1 PT> and then press to confirm and the
next display appears:
6 Operation 76
User Manual DCA-1
6 Operation 77
User Manual DCA-1
5. Insert the ChipCARD into the reader slot with the memory chip
facing up. Leave the ChipCARD in the reader, see Figure 4-5,
page 33 for correct insertion and chapter 4.8 ChipCARDs –
Store or load methods, page 36 for more information.
➢ If no ChipCARD is inserted or if the ChipCARD is not correct a
message “INVALID ChipCARD” as floating text appears:
8. Press to confirm.
6 Operation 78
User Manual DCA-1
NOTE
Before starting measurements with the new method
we recommend to set the new method name and
enter the correct Lot.-No. of the reagent.
6 Operation 79
User Manual DCA-1
6 Operation 80
User Manual DCA-1
5. Insert the ChipCARD into the reader slot with the memory chip
facing up. Leave the ChipCARD in the reader, see Figure 4-5,
page 33 for correct insertion and chapter 4.8 ChipCARDs –
Store or load methods, page 36 for more information.
then
6 Operation 81
User Manual DCA-1
6 Operation 82
User Manual DCA-1
NOTE
Evaluating the measurement results with the set 1st
conversion parameters means that the results are
calculated based on previously determined
calibration points or calibration values given in the
lot-specific PT reagent leaflet.
Selection Meaning
6 Operation 83
User Manual DCA-1
4. Press .
6 Operation 84
User Manual DCA-1
6.6.2.1 <None>
NOTE
Reference curves must be entered as follows:
Rising curves (inhibitors) start with the lowest value.
Falling curves (coagulation factors) start with the
shortest measuring times (sec) and the highest
converted value (%, g/l, mg/dl).
The software will support this procedure only!
6 Operation 85
User Manual DCA-1
➢ max.: 4 digits.
NOTE
Regardless of the language used a dot is always
used as decimal instead e.g. a comma.
to confirm.
6 Operation 86
User Manual DCA-1
You can enter the minimum and maximum values for the
calibration curve and conversion.
➢ Points in the reference curve that are not used need entries of
0.0 % and 0.0 s. As soon as e.g. the 5th point has the entry 0.0
the next point (here the 6th) will not appear.
6 Operation 87
User Manual DCA-1
5. Press to confirm the entry, the field for the next point on
the reference curve appears automatically.
6 Operation 88
User Manual DCA-1
to confirm.
6 Operation 89
User Manual DCA-1
NOTE
Results are calculated based on set MNP (= Mean
Normal Plasma).
ISI values are given in the leaflet of the lot-specific
PT-reagent.
RATIO = patient sample in sec is divided by MNP in
sec.
INR = International Normalized Ratio
ISI = International Sensitivity Index according to
package insert
NOTE
Only if <None> has been selected for <1st
conversion> an input field to enter the MNP (Mean
Normal Plasma) value will automatically be
displayed for the INR/RATIO conversion under <2nd
conversion>.
In case of the aPTT-Method, the 1st conversion is
none and the MNP has to be entered.
If the <Reference curve> is selected under <1st
conversion> and the reference curve consists of
valid points, the MNP value is taken from the
reference curve. Thus, there is no option to set the
MNP.
6 Operation 90
User Manual DCA-1
Selection Meaning
6 Operation 91
User Manual DCA-1
6 Operation 92
User Manual DCA-1
NOTE
If the 1st conversion is active, the MNP value is not
accessible in the RATIO submenu.
6 Operation 93
User Manual DCA-1
Menu <Measurement>
CAUTION
Any modifications within this menu can cause
different measuring results!
To avoid incorrect measuring results only authorized
personnel should have access to the method
parameterization!
After each entry a parameter protocol must be
printed in order to check the entry again!
Selection Meaning
6 Operation 94
User Manual DCA-1
NOTE
Alternatively, if a barcode scanner is connected:
• Scan the barcode label of the reagent vial (a
maximum of 12 alphanumeric digits is imported
as string). The Software automatically displays
the next menu after the string was imported and
displayed for 3 seconds.
6 Operation 95
User Manual DCA-1
NOTE
For the PT method, the first incubation time is set to
0 by definition. The according value is assigned to
the second incubation. If another parameterization is
set, the analyzer might change the measuring
procedure which could lead to missleading results
and interpretation.
See according reagent leaflet for correct incubation
times.
6 Operation 96
User Manual DCA-1
NOTE
The mixer function is implemented to gently
and properly mix the sample (and reagent).
Information to method-specific settings are
given in chapter 5, page 41, Table 5-1.
Before modification consult your local
distributor.
6 Operation 97
User Manual DCA-1
➢ Here you can enter parameters for the optical detection of the
sample.
NOTE
Before any modification consult your local
distributor.
6 Operation 98
User Manual DCA-1
Here the minimum time after which a clot can be detected and the
upper limit of clotting detection can be set.
1. The cursor blinks on the first entry position for “min. time”.
2. Enter the “min. time” with number keys or confirm and
proceed.
6 Operation 99
User Manual DCA-1
6 Operation 100
User Manual DCA-1
UTILITIES
Selection Meaning
6 Operation 101
User Manual DCA-1
NOTE
If the number was incorrect STANDBY display
appears again. The following display appears,
provided the correct PIN Code was entered.
6 Operation 102
User Manual DCA-1
SubMenu <Printer>
In this SubMenu you can select the print-out of different data, e.g.
test results, method- or analyzer parameters as well as error
messages via the external printer.
Table 6-5 Printer selections
Selection Meaning
<On>
Printer is activated, Press for print-
out.
<Auto> Print of results automatically, as soon as the
test result is available.
<Parameter Print of Method- and global analyzer
protocol> parameter
<Actual Print of method parameter of a selected
Method> method
<All Print of all analyzer- and method parameter
Parameter>
<Global> Print of global analyzer parameter
<Off> Printer is deactivated
6 Operation 103
User Manual DCA-1
NOTE
To select and activate the actual method for the
print out, the warm-up time of the analyzer after
“switch on” must be finished. Then, select the
6 Operation 104
User Manual DCA-1
Selection Meaning
6 Operation 105
User Manual DCA-1
NOTE
If not already made available by the HOST
operating system a USB driver has to be installed
previously on the HOST in order to use the USB
port.
Selection Meaning
6 Operation 106
User Manual DCA-1
6 Operation 107
User Manual DCA-1
6 Operation 108
User Manual DCA-1
6 Operation 109
User Manual DCA-1
SubMenu <Beeper>
Selection Meaning
6 Operation 110
User Manual DCA-1
SubMenu <Language>
Selection Meaning
2. Press
6 Operation 111
User Manual DCA-1
After that:
SubMenu <Date/Time>
In this menu you can set the actual date and time of the analyzer.
2. Press
6 Operation 112
User Manual DCA-1
3. Press respective number keys to enter date and time (from left
to right).
6. Press
6 Operation 113
User Manual DCA-1
➢ After the PIN Code has been entered, the following display
appears:
6 Operation 114
User Manual DCA-1
-----------------------------
<Analyzer name>
Ser. No. B1770711 Example print-out!
PIN No. =xxxxx x = Number
-----------------------------
NOTE
If 5 times a 0 “zero” is entered and saved, the PIN
Code is not required to enter the parameter or
UTILITIES menu.
In any case, the PIN Code must have 5 characters.
6 Operation 115
User Manual DCA-1
SubMenu <Cuvette-test>
This menu checks the function of the automatic cuvette
recognition.
➢ The thermal block has to be warmed up to operating
temperature.
6 Operation 116
User Manual DCA-1
If the CUVETTE-test is carried out but the thermal block has not
reached its final temperature the following info appears:
6 Operation 117
User Manual DCA-1
NOTE
Make sure the SD-Card does not carry any
additional adhesive stickers this could harm the
card reading unit.
The write protection of the SC-Card has to be
unlocked otherwise no data can be exported.
IF the SD-Card is protected the analyzer shows a
P-FLAG. In this case remove the SD-Card, unlock it
and insert it again
6 Operation 118
User Manual DCA-1
1. Insert a SD-Card into the card reading unit located on the rear
side of the analyzer.
2. Ensure that the SD-Card latches in the card reader.
4. Press .
5. Select < Import <-SD-Card>.
6. Press .
6 Operation 119
User Manual DCA-1
• Slightly press the outer visible face of the SD-Card into the
analyzer.
6 Operation 120
User Manual DCA-1
NOTE
Make sure the SD-Card does not carry any
additional adhesive stickers this could harm the
card reading unit.
The write protection of the SC-Card has to be
unlocked otherwise no data can be exported.
If the SD-Card is protected the analyzer shows a P-
FLAG. In this case remove the SD-Card, unlock it
and insert it again
1. Insert a SD-Card into the card reading slot located on the back
side of the analyzer.
2. Ensure that the SD-Card is inserted and pushed into the card
reader as far as possible.
4. Press .
5. Select < Export - > SD-Card>.
6 Operation 121
User Manual DCA-1
6. Press .
6 Operation 122
User Manual DCA-1
• Slightly press the outer visible face of the SD-Card into the
analyzer.
SubMenu <Info>
This submenu has the purpose to show the user
✓ Name of the analyzer
✓ Current software version
✓ Date of release of software version
✓ Serial number of the analyzer
✓ Parameter ID number
✓ Number of NTC-T
6 Operation 123
User Manual DCA-1
General print-outs
6 Operation 124
User Manual DCA-1
Start Reagent:
Lot = 1 Reagent-Lot. No.
Volume = 100 ul Reagent Start Volume
incubat. = 60 s Incubation time
1st convers 1. conversion, calib.
INTERPOLAT. curve / interpolation
97.0%= 11.5 s Calibration curve points
43.0%= 20.1 s
23.0%= 31.9 s
12.0%= 57.0 s
2nd convers INR 2. conversion INR
ISI = 1.05 ISI-constant
Results:
PT Method
Patient Patient name
___________
dd:mm:yyyy, Date, Time
hh:mm:ss
No. = 1 Print.-out no.
Channel = 1 Channel No. 1
Time = 12.0s Measuring time
% = 90.4 Conversion to PT %
INR = 1.06 Conversion to INR
6 Operation 125
User Manual DCA-1
7 TROUBLESHOOTING
Application errors
7 Troubleshooting 126
User Manual DCA-1
Should any of these errors occur and they are recognized in time,
solve them immediately.
7 Troubleshooting 127
User Manual DCA-1
Status message
Cause Action
(display/print)
7 Troubleshooting 128
User Manual DCA-1
7 Troubleshooting 129
User Manual DCA-1
Error message
Cause Action
(Display)
7 Troubleshooting 130
User Manual DCA-1
Error message
Cause Action
(Display)
passes the
channel)
break Measurement Intentionally
cancelled with done by user!
and
Bright value Channel is not free, Remove the
failed e.g. remaining cuvette
cuvette, dust or Clean the
other debris channel
disturbs the light
path Then press
ENTER to retry
or ESC to go to
the next channel
If necessary call
the technical
service.
Channel A measuring Load new
locked! channel is locked cuvette balance
due to no to analyzer.
remaining cuvette Switch OFF the
balance available. analyzer and
Error during ON again to
adjustment repeat the
adjustment.
CuvCard or Parameter Contact
ChipCARD checksum on technical service
Checksum error! inserted card type
is incorrect
CuvCard or Inserted card type Re-insert card
ChipCARD may be defect. and repeat
Read error! process,
otherwise
contact technical
service
7 Troubleshooting 131
User Manual DCA-1
Error message
Cause Action
(Display)
7 Troubleshooting 132
User Manual DCA-1
Error message
Cause Action
(Display)
7 Troubleshooting 133
User Manual DCA-1
Error message
Cause Action
(Display)
7 Troubleshooting 134
User Manual DCA-1
Error message
Cause Action
(Display)
problem
persists.
WARNING: PC is not 1. Switch on the
HOST not connected. HOST / PC.
connected USB driver is not 2. Switch on the
installed. analyzer.
PC and analyzer 3. Set the
are switched on in analyzer to
the wrong order. STANDBY.
4. Install a USB
driver.
7 Troubleshooting 135
User Manual DCA-1
of the
manufacturer.
Measuring cells Additional Remove liquid
contaminated pipetting of with pipette,
with liquids plasma or reagent clean with
into the measuring appropriate
cell without absorbent cloth
cuvette. and disinfectant.
Warnings
7 Troubleshooting 136
User Manual DCA-1
7 Troubleshooting 137
User Manual DCA-1
7 Troubleshooting 138
User Manual DCA-1
7 Troubleshooting 139
User Manual DCA-1
Safety
CAUTION
• Do not use organic acids to clean the analyzer.
• Do not spray or pour on any liquids, as they can
affect the correct function of the analyzer or
damage it.
• Keep the analyzer free from dust and liquid
spillages. When not used over a longer period
of time, place a dust cover over the analyzer or
place it into a cabinet.
If liquid has been spilled on the analyzer.
• Remove the contamination with a clean,
absorbent non-woven cloth considering all
applicable hygienic requirements.
If liquid has accidentally entered, or been pipetted
into a measuring channel:
• Remove the liquid with a pipette, then clean the
measuring channel with a lint-free cloth.
• Consider all applicable hygienic requirements!
Contact our technical service if subsequent control
measurements do not produce the expected result.
NOTE
The analyzer is fitted with a lithium battery type Li-
Mn CR 2430 (life approx. 5 years). It should be
replaced by an authorised service operator after 5
years at the latest. Otherwise a faultless operation
can not be guaranteed.
BIOLOGICAL RISK
The light protection cap is considered to be
potentially contaminated. Therefore, the
manufacturer recommends to replace it once a
year. Contact your local distributor for further
information.
Disinfectant
✓ 96% Ethanol, denatured.
➢ To clean and disinfect the analyzer (Thermal block, working
area)
Hand protection
✓ Disposable nitrile rubber gloves.
➢ Avoid any hand contact with potentially infectious material.
Eye protection
✓ Safety eye glasses.
➢ Splashes are likely to occur, there is a risk of infection.
Other
✓ Lint-free cloth.
9 TRANSPORT
NOTE
Please clean and disinfect the analyzer at first.
Take care when moving the analyzer.
Secure the analyzer against slipping and tilting.
There are no special signs regarding the transport
of the analyzer.
1. Remove
cuvettes incl.
mixer.
2. Remove all vials
and adapters (if
inserted).
NOTE
Do not leave any
loose parts inside the
analyzer!
9 Transport 142
User Manual DCA-1
9 Transport 143
User Manual DCA-1
9 Transport 144
User Manual DCA-1
NOTE
Please request a Return Authorization Number prior
to any return shipment.
For proper handling of a complaint please make sure
the following information is available.
• AnalyzerType (Name Plate)
• Serial Number of the analyzer (Name Plate)
• Model (Name Plate)
• Problem Description
• If available a picture of the defective part
Furthermore, we recommend to keep a service log
book to document maintenance work with information
about date, reason and operator/ responsible person.
This makes the life cycle of the analyzer clearly
visible and traceable.
If you intend not to use the analyzer over a longer period of time
follow the instructions as described.
CAUTION
CAUTION
For the safety of the operating personnel make sure
the analyzer has been disinfected before disposal.
1. Remove the power supply from the socket and remove the
power supply cable from the analyzer.
2. Remove all loose items, e.g. cuvettes or vials from the
analyzer.
3. Clean / disinfect the analyzer properly with Ethanol (96%,
denatured).
4. Let the analyzer dry for a certain time to get rid of any possible
remaining liquid.
5. Put the analyzer into a film bag.
6. Put the power supply plug and power supply cable into a film
bag.
7. Prepare the analyzer for transport as described in Chapter 9
Transport, page 142.
Do not dismount and recylcle the analyzer by yourself.
➢ All parts parts must be disposed of in accordance with
national directives for disposal
13 APPENDIX
Technical data
System related
Characteristics Description
13 Appendix 148
User Manual DCA-1
Electrical data
13 Appendix 149
User Manual DCA-1
Pollution degree 2
Means of protection Class III
Lithium battery type Li-Mn +3V
CR 2430
Environmental requirements
Parameter Specification
Disposables
13 Appendix 150
User Manual DCA-1
Materials supplied
Quantity Description
1x DCA-1
1x 10 single cuvettes with mixer in plastic bag
1x User manual on USB stick
1x External power supply.
Optional equipment
✓ Barcode scanner
✓ External printer
✓ USB driver (if not already available at HOST)
✓ Cable for printer and scanner
Supplier documentation
13 Appendix 151