Professional Documents
Culture Documents
CH 10
CH 10
Chapter 10
Non-invasive
monitoring
146 10 Non-invasive monitoring
Amplitude mV
0.5 0.5 heart is a common error, leading
to false information.
2. Electrical interference can be a
0 0 50-Hz (in UK) mains line
interference because of
capacitance or inductive coupling
–0.5 –0.5
0 0.2 0.4 0.6 0.8 0 0.2 0.4 0.6 0.8 effect. Any electrical device
t.s t.s powered by AC can act as one
plate of a capacitor and the
A B
patient acts as the other plate.
Fig.10.6 ECG filters. (A) Unfiltered signal with noise. (B) Filtered ‘clean’ signal. Interference can also be because
of high-frequency current
interference from diathermy.
b) the diagnostic mode has a Right arm lead Most modern monitors have the
wider frequency response of over manubrium facilities to avoid interference.
0.05–150 Hz. The high- sterni Shielding of cables and leads,
frequency limit allows the Indifferent lead differential amplifiers and
assessment of the ST segment, Left arm lead electronic filters all help to
QRS morphology and V5 position over produce an interference-free
tachyarrhythmias. The left ventricle monitoring system. Differential
low-frequency limit allows amplifiers measure the difference
representation of P- and between the potential from two
T-wave morphology and different sources. If there is
ST-segment analysis. interference common to the two
4. There are many ECG electrode input terminals (e.g. mains
configurations. Usually during frequency), it can be eliminated as
anaesthesia, three skin electrodes only the differences between the
are used (right arm, left arm and two terminals is amplified. This is
indifferent leads). The three limb called common mode rejection
leads used include two that are ratio (CMRR). Amplifiers used in
‘active’ and one that is ‘inactive’ ECG monitoring should have a
(earth). Sometimes five electrodes Fig. 10.7 The CM5 ECG lead high CMRR of 100 000 : 1
are used. Lead II is ideal for configuration. to 1 000 000 : 1, which is a
detecting arryhthmias. CM5 measurement of capability to
configuration is able to detect reject the noise. They should also
89% of ST-segment changes due left anterior axillary line) and have a high input impedance
to left ventricular ischaemia. In the indifferent lead is on the left (about 10 MΩ) to minimize the
CM5, the right arm electrode is shoulder or any convenient current taken from the electrodes.
positioned on the manubrium position (Fig. 10.7). Table 10.1 shows the various
(chest lead from manubrium), 5. The CB5 configuration is useful types and sources of interference
the left arm electrode is on V5 during thoracic anaesthesia. The and how to reduce the
position (fifth interspace in the right arm electrode is positioned interference.
Arterial blood pressure 149
Arterial
Absorbance
Venous
Skin
Tissue
Bone
Time
Fig. 10.12 Pulse oximeter probes. Finger probe (top) and ear probe (bottom).
Fig. 10.14 Schematic representation
of the contribution of various body
components to the absorbance of light.
Photodetector Display
Microprocessor
4. The microprocessor is
programmed to mathematically
analyse both the DC and AC
components at 660 and 940 nm
calculating the ratio of
absorption at these two
660 940 frequencies (R/IR ratio). The
nm nm result is related to the arterial
LED sequence saturation. The absorption
of oxyhaemoglobin and
On Off
Sequence repeated many times per second
deoxyhaemoglobin at these two
Off On
wavelengths is very different.
Off Off
This allows these two
Fig. 10.13 Working principles of the pulse oximeter. The LEDs operate in sequence and wavelengths to provide good
when both are off the photodetector measures the background level of ambient light. sensitivity. 805 nm is one of
the isobestic points of
oxyhaemoglobin and
constant (DC). The non-constant deoxyhaemoglobin. The OFF
Mechanism of action
absorption (AC) is the result part allows a baseline
1. The oxygen saturation is of arterial blood pulsations measurement for any changes in
estimated by measuring the (Fig. 10.14). The sensitive ambient light.
transmission of light, through a photodetector generates a 5. A more recent design uses
pulsatile vascular tissue bed (e.g. voltage proportional to the multiple wavelengths to
finger). This is based on Beer’s transmitted light. The AC eradicate false readings from
law (the relation between component of the wave is carboxy haemoglobin and
the light absorbed and the about 1–5% of the total methaemoglobinaemia.
concentration of solute in the signal. Advanced oximeters use more
solution) and Lambert’s law 3. The high frequency of the LEDs than seven light wavelengths.
(relation between absorption of allows the absorption to be This has enabled the
light and the thickness of the sampled many times during each measurement of haemoglobin
absorbing layer). pulse beat. This is used to enable value, oxygen content,
2. The amount of light transmitted running averages of saturation carboxyhaemoglobin and
depends on many factors. The to be calculated many times per methaemoglobin concentrations.
light absorbed by non-pulsatile second. This decreases the 6. A variable pitch beep provides
tissues (e.g. skin, soft tissues, ‘noise’ (e.g. movement) effect on an audible signal of changes in
bone and venous blood) is the signal. saturation.
End-tidal carbon dioxide analysers (capnographs) 153
E I E
CO2 (kPa)
D D
C
AB
Time (sec)
Fig. 10.15 Diagram of an end-tidal
carbon dioxide waveform. I = inspiration;
E = expiration; A–B represents the
emptying of the upper dead space of the
airways. As this has not undergone gas
exchange, the CO2 concentration is zero.
B–C represents the gas mixture from the
upper airways and the CO2-rich alveolar
gas. The CO2 concentration rises
continuously. C–D represents the alveolar
gas and is described as the ‘alveolar
plateau’. The curve rises very slowly. D is
the end-tidal CO2 partial pressure where
the highest possible concentration of Fig. 10.16 A main-stream end-tidal carbon dioxide analyser.
exhaled CO2 is achieved at the end of
expiration. It represents the final portion concentrations during each which in turn produces heat.
of gas which was involved in the gas respiratory cycle. The heat is measured by a
exchange in the alveoli. Under certain temperature sensor and is
conditions (see text) it represents a
Components proportional to the partial
reliable index of the arterial CO2 partial pressure of carbon dioxide gas
pressure. D–A represents inspiration 1. The sampling chamber can present in the mixture in the
where the fresh gas contains no CO2. either be positioned within the sample chamber. This produces
patient’s gas stream (main- an electrical output. This means
The end-tidal CO2 is less than stream version, Fig. 10.16) or that the amount of gas present is
alveolar CO2 because the end-tidal connected to the distal end of inversely proportional to the
CO2 is always diluted with alveolar the breathing system via a amount of infrared light present
dead space gas from unperfused sampling tube (side-stream at the detector in the sample
alveoli. These alveoli do not take version, Fig. 10.17). chamber (Fig 10.19).
part in gas exchange and so contain 2. A photodetector measures light 4. In the same way, a beam of light
no CO2. Alveolar CO2 is less than reaching it from a light source at passes through the reference
arterial CO2 as the blood from the correct infrared wavelength chamber which contains room
unventilated alveoli and lung (using optical filters) after air. The absorption detected from
parenchyma (both have higher CO2 passing through two chambers. the sample chamber is compared
contents) mixes with the blood from One acts as a reference whereas to that in the reference chamber.
ventilated alveoli. In healthy adults the other one is the sampling This allows the calculation of
with normal lungs, end-tidal CO2 is chamber (Fig. 10.18). carbon dioxide values.
0.3–0.6 kPa less than arterial CO2. 5. The inspired and exhaled
This difference is reduced if the carbon dioxide forms a square
Mechanism of action
lungs are ventilated with large tidal wave, with a zero baseline
volumes. The Greek root kapnos, 1. Carbon dioxide absorbs the unless there is rebreathing
meaning ‘smoke’, give us the term infrared radiation particularly at (Fig. 10.20A).
capnography (CO2 can be thought as a wavelength of 4.3 µm. 6. A microprocessor-controlled
the ‘smoke’ of cellular metabolism). 2. The amount of infrared infrared lamp is used. This
radiation absorbed is produces a stable infrared
End-tidal CO2 < alveolar CO2 source with a constant output.
proportional to the number of
< PaCO2
carbon dioxide molecules The current is measured
In reality, the devices used cannot (partial pressure of carbon with a current-sensing resistor,
determine the different phases of dioxide) present in the chamber. the voltage across which
respiration but simply report the 3. The remaining infrared radiation is proportional to the
minimum and maximum CO2 falls on the thermopile detector, current flowing through it. The
End-tidal carbon dioxide analysers (capnographs) 155
Fig. 10.17 The Penlon PM9000 Express which measures end-tidal CO2, oximetry and SIDE-STREAM ANALYSERS
inhalational agent concentration using a side-stream method. (Courtesy of Penlon Ltd,
Abingdon, UK (www.penlon.com).) 1. This consists of a 1.2-mm
internal diameter tube that
samples the gases (both inspired
Sample gas and exhaled) at a constant
rate (e.g. 150–200 mL/min).
The tube is connected to a
lightweight adapter near
Detector
the patient’s end of the
breathing system (with a
Light source Sample chamber Multigas filter pneumotachograph for
Fig. 10.18 Components of a gas analyser using an infrared light source suitable for spirometry) with a small increase
end-tidal carbon dioxide measurement. The reference chamber has been omitted for the in the dead space. It delivers the
sake of clarity. gases to the sample chamber.
It is made of Teflon so it is
impermeable to carbon dioxide
supply to the light source respiratory cycle, monitors are and does not react with
is controlled by the feedback designed to measure the anaesthetic agents.
from the sensing resistor respiratory rate. 2. As the gases are humid, there is
maintaining a constant 8. Alarm limits can be set for both a moisture trap with an exhaust
current of 150 mA. high and low values. port, allowing gas to be vented
7. Using the rise and fall of the 9. To avoid drift, the monitor should to the atmosphere or returned to
carbon dioxide during the be calibrated regularly with the breathing system.
156 10 Non-invasive monitoring
Sample chamber
Filter
Fig. 10.19 Principles of infrared detector: due to the large amount of infrared
absorption in the sample chamber by the carbon dioxide, little infrared finally reaches
the detector.
sample.
intubation (no or very little
6. Portable hand-held side-stream
carbon dioxide is detected).
analysers are available (Fig.
Following manual ventilation or
10.21). They can be used during
the ingestion of carbonated
patient transport and out-of-
Time (sec) drinks, some carbon dioxide
hospital situations.
Fig. 10.20 (A) An end-tidal carbon
might be present in the
dioxide waveform which does not return stomach. Characteristically, this
to the baseline during inspiration may result in up to 5–6
MAIN-STREAM ANALYSER
indicating that rebreathing is occurring. waveforms with an abnormal
(B) An end-tidal carbon dioxide waveform shape and decreasing in
which illustrates the sloping plateau seen 1. The sample chamber is positioned
amplitude.
in patients with chronic obstructive within the patient’s gas stream,
2. As a disconnection alarm for
airways disease. The normal waveform is increasing the dead space. In
superimposed (dotted line). a ventilator or breathing
order to prevent water vapour
system. There is sudden
condensation on its windows,
absence of the end-tidal carbon
it is heated to about 41°C.
3. In order to accurately measure dioxide.
2. Since there is no need for a
end-tidal carbon dioxide, the 3. To diagnose lung embolism as a
sampling tube, there is no
sampling tube should be sudden decrease in end-tidal
transport time delay in gas
positioned as close as possible to carbon dioxide assuming that
delivery to the sample chamber.
the patient’s trachea. the arterial blood pressure
3. Other gases and vapours are not
4. A variable time delay before remains stable.
measured simultaneously.
the sample is presented to the 4. To diagnose malignant
sample chamber is expected. The See Table 10.4 for a comparison hyperpyrexia as a gradual
transit time delay depends on of side-stream and main-stream increase in end-tidal carbon
the length (which should be as analysers. dioxide.
End-tidal carbon dioxide analysers (capnographs) 157
Sensing O2
Amplifier Out Thin electrolyte layer Electromagnet
membrane
Cathode Mixture
out
Ag-electrode Electrolyte
Pt-electrode Anode
Electrolyte Reference in
Switched
magnetic
Teflon membrane Circular Sample in field
Out contact plate
Fig. 10.23 Different types of oxygen analysers.
Oxygen concentration analysers 159
Table 10.6 summarizes the on and off and reset the pointer
Inhalational agent methods used in gas and vapour to the zero position.
concentration analysers analysis.
● A sample of gas is used to Mechanism of action
measure the concentration of
inhalational agent using infrared 1. The Wright respirometer is a
light absorption. one-way system. It allows the
● By selecting light of the correct
Wright respirometer measurement of the tidal volume
wavelengths, the inspired and if the flow of the gases is in one
expired concentrations of the This compact and light (weighs less direction only. The correct
agent(s) can be measured. than 150 g) respirometer is used to direction for gas flow is
● An infrared light of a
measure the tidal volume and indicated by an arrow.
wavelength of 4.6 µm is used minute volume (Fig. 10.29). 2. The slits surrounding the vane
for N2O. For other inhalational are to create a circular flow in
order to rotate the vane. The
agents, higher wavelengths are Components
used, between 8 and 9 µm. vane does 150 revolutions for
● Ultraviolet absorption, mass
1. The respirometer consists of an each litre of gas passing through.
spectrometry and quartz crystal inlet and outlet. This causes the pointer to rotate
oscillation are other methods of 2. A rotating vane surrounded by round the respirometer display.
measuring the inhalational slits (Fig. 10.30). The vane is 3. The outer display is calibrated at
agents’ concentration. attached to a pointer. 100 mL per division. The small
3. Buttons on the side of the inner display is calibrated at 1 L
respirometer to turn the device per division.
MASS SPECTROMETER
Laminar
Gas flow
resistor
Heating coil
Fig. 10.31 A pneumotachograph.
Mica vane
See text for details.
Gas flow
Fig. 10.30 Mechanism of action of the Wright respirometer. Mechanism of action
1. The principle of its function is
4. It is usually positioned on the Problems in practice and sensing the change in pressure
expiratory side of the breathing safety features across a fixed resistance through
system, which is at a lower
1. The Wright respirometer tends which gas flow is laminar.
pressure than the inspiratory
to over-read at high flow rates 2. The pressure change is only a
side. This minimizes the loss of
and under-read at low flows. few millimetres of water and is
gas volume due to leaks and
2. Water condensation from the linearly proportional, over a
expansion of the tubing.
expired gases causes the pointer certain range, to the flow rate of
5. For clinical use, the respirometer
to stick, thus preventing it from gas passing through the
reads accurately the tidal volume
rotating freely. resistance.
and minute volume (±5–10%)
3. The tidal volumes can be
within the range of 4–24 L/min.
summated over a period of a
A minimum flow of 2 L/min is
Wright respirometer minute to give the minute
required for the respirometer to
● Rotating vane attached to a volume.
function accurately.
pointer. 4. It can measure flows in both
6. To improve accuracy, the
● Fitted on the expiratory limb to inspiration and expiration (i.e.
respirometer should be
measure the tidal and minute bidirectional).
positioned as close to the
volume with an accuracy of
patient’s trachea as possible.
±5–10%. Problems in practice and
7. The resistance to breathing is
● The flow is unidirectional.
very low at about 2 cm H2O at safety features
● It over-reads at high flows and
100 L/min.
under-reads at low flows. Water vapour condensation at the
8. A paediatric version exists with
resistance will encourage the
a capability of accurate tidal
formation of turbulent flow
volume measurements between
affecting the accuracy of the
15 and 200 mL.
measurement. This can be avoided
9. A more accurate version of the Pneumotachograph by heating the parallel tubes.
Wright respirometer uses light
reflection to measure the tidal This measures gas flow. From this,
volume. The mechanical causes gas volume can be calculated.
of inaccuracies (friction and Combined
inertia) and the accumulation of
water vapour are avoided. Other
Components pneumotachograph
designs use a semiconductive 1. A tube with a fixed resistance. and Pitot tube
device that is sensitive to The resistance can be a
changes in magnetic field. Tidal bundle of parallel tubes (Fig. This combination (Fig. 10.32) is
volume and minute volume can 10.31). designed to improve accuracy
be measured by converting these 2. Two sensitive pressure and calculate and measure the
changes electronically. An alarm transducers on either side of the compliance, airway pressures, gas
system can also be added. resistance. flow, volume/pressure (Fig. 10.33)
164 10 Non-invasive monitoring
Sample gas
flow to be known with sensors
calibrated accordingly.
3. Gas temperature: A knowledge
gas temperatures is required.
Usually, the sensors’ software
provides default values for a
Gas flow typical patient.
4. Humidity: moisture can affect
measurement and generation of
pressure drop. Have the pressure
ports directed upwards to
To pressure prevent fluid from draining into
transducers them.
Fig. 10.32 Combined pneumotachograph and Pitot tube. (Courtesy of GE Datex 5. Apparatus dead space:
Ohmeda.) Sensors need to have a
minimum dead space; <10 ml
for the adult flow sensors and
600 Vol Gas flow <1 ml for the neonatal sensors.
mL
6. Operating range of flow
A B sensor: Sensors are designed to
function accurately with a very
wide range of tidal volumes, I : E
ratios, frequencies and flow
Paw
cmH20 ranges.
To pressure transducer 7. Inter-sensor variability:
0 20
Fig. 10.34 Cross-section of a Pitot tube Individual sensors can have
Fig. 10.33 Volume pressure loops in a flowmeter. The two ports are facing in different performances. There
patient (A) before and (B) during CO2 opposite directions within the gas flow. should be no need for individual
insufflation in a laparoscopic operation.
Note the decrease in compliance and device calibration of the flow/
increase in airway pressure (Paw). pressure characteristics
Mechanism of action
The effects of the density and
The pressure difference between the
viscosity of the gas(es) can alter the
and flow/volume loops. Modern ports is proportional to the square
accuracy. This can be compensated
devices can be used accurately even of the flow rate.
for by continuous gas composition
in neonates and infants. analysis via a sampling tube.
Problems in practice and
THE PITOT TUBE safety features Pneumotachograph
The effects of the density and ● A bidirectional device to measure
Components the flow rate, tidal and minute
viscosity of the gas(es) can alter the
1. Two pressure ports – one facing accuracy. This can be compensated volume.
the direction of gas flow, the for by continuous gas composition ● A laminar flow across a fixed
other perpendicular to the gas analysis via a sampling tube. resistance causes changes in
flow. This is used to measure gas pressure which are measured by
flow in one direction only. transducers.
Factors affecting the readings in
2. In order to measure bidirectional ● Condensation at the resistance
pnuemotachograph
flows (inspiration and can cause turbulent flow and
expiration), two pressure ports 1. Location: should be placed inaccuracies.
face in opposite directions between the breathing system ● Improved accuracy is achieved
within the gas flow (Fig. 10.34). Y-piece and the tracheal tube. by adding a Pitot tube(s) and
3. These pressure ports are 2. Gas composition: nominal continuous gas composition
connected to pressure values of gas composition need analysis.
transducers.
Peripheral nerve stimulators 165
Peripheral nerve
stimulators
These devices are used to
monitor transmission across the
neuromuscular junction. The depth,
adequate reversal and type of
neuromuscular blockade can be
established (Fig. 10.36).
Components
1. Two surface electrodes (small
ECG electrodes) are positioned
over the nerve and connected via
Fig. 10.35 The Penlon pressure monitoring ventilator alarm. the leads to the nerve stimulator.
166 10 Non-invasive monitoring
A Normal
20 ms
750 ms
Fig. 10.38 The pattern of double-burst
B Total paralysis stimulation. Three impulses of 50 Hz
tetanus, at 20-ms intervals, every 750 ms
is shown.
MCQs
In the following lists, which of the statements (a) to (e) are true?
Answers