You are on page 1of 3

Underwriters Laboratories (UL LLC) Manufacturer Report

Model: JCB35Q, JCB35R, JCB35T


Manufacturing Factory(ies): Same as Applicant

Applicant: ZHEJIANG JIECANG LINEAR MOTION TECHNOLOGY CO LTD


PROVINCIAL HIGH TECH PARK
XINCHANG COUNTY
SHAOXING
ZHEJIANG 312500 CHINA
Report No.: E355312-D1000-1/A2/C1
Report Issue Date: 2017-01-03; 2017-06-16(A1C1); 2018-02-28(A2C1)
Standard(s): ANSI/AAMI ES60601-1 (2005/(R)2012 + A1:2012, C1:2009/(R)2012 +
A2:2010/(R)2012) - Amendment 1 - Revision Date 2012/08/21
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 - Revision Date 2014/03
Additional Standards: ANSI/AAMI ES60601-1 (2005/(R)2012 + A1:2012, C1:2009/(R)2012 +
A2:2010/(R)2012) - Amendment 1 - Revision Date 2012/08/21
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 - Revision Date 2014/03
IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012
EN 60601 1:2006/ A1:2013/ A12:2014

Report Types: This report consists of the following report types:


[ Yes ] US Certification (UL Recognition)
[ Yes ] CAN Certification (UL Recognition)

You are receiving this report as a Manufacturing Factory for the above-referenced Applicant.
This report covers the Safety evaluation of the referenced model(s) according to the standard(s)
specified above.
This report includes portions of the full Applicant report and is designed to support any Follow-Up
inspections at the Manufacture’s facility.
Please refer to any Supplements to this document for a detailed manufacturer report, which includes
information for the Follow-Up Inspection as specified below:
- Appendix C: Follow-Up Service Documentation
- Appendix A: Enclosures
Report Modifications Summary
The following changes were made to this report. If none listed in the below table, this report is the
originally issued report.

NOTE: Some of the changes listed below were made in the full report and may not reflect the changes in
this associated Factory Report, which just includes the Appendices listed above from the main report.

Date Modified
Modifications Made (include Report Reference Number) Modified By
(Year-Month-Day)
2017-01-03 Original report. Report No.: E355312-D1000-1/A0/C0-UL Cary Hu
Amendment 01. Report No.: E353512-D1000-1/A1/C1-UL
Add one additional model: JCB35R,
1) Below tests were performed, see details on Record Details
and related Tables
4.11 Power Input
5.7 Humidity Conditioning
5.9.2 Determination of Accessible Parts
8.4.2 Limitation of Voltage, Current of Power
8.4.3 Voltage of Charge Limitation
8.5.4 Working Voltage Measurements
8.7 Leakage Current Tests
8.7.4.6 Touch leakage Current
8.7.3 e) Non-Frequency-Wighted Leakage Current
8.8.3 Dielectric Voltage Withstand
11 Temperature
11.6.5 Harmful Ingress (Water and Particulates)
11.6.6 Cleaning and Disinfection Cary Hu / Paul B
2017-06-16
11.8 Interruption of the Power Supply Zhang
13 Abnormal
15.3.2 Enclosure Force
15.3.3 Enclosure Impact
15.3.6 Mold Stress Relief

2) Change related clauses


- Clause 7.3.3 Type of battery and mode of insertion marked to
Pass
- Clause 8.2, Clause 8.2.2 Connection to an external d.c. power
source to Pass
- Clause 15.4.3 Batteries to Pass;
- Clause 15.4.3.1, Clause 15.4.3.2, Clause 15.3.4.3 and Clause
15.3.4.5 to Pass;
3) Add and update the related components
For secondary control board:
- Add Connector: 1300S (ASAHI KASEI CORP)
- Add Relay: W14 (Zhengjiang Fanhar Electronic Co.,Ltd)
- Add Varistor: 10D560K (Brightking (Shenzhen) Co Ltd)
- Add Connector: 101L (E I DUPONT DE NEMOURS & CO INC)
For Battery Package for JCB35R only :
- Add Enclosure: 3370 (SILVER AGE ENGINEERING PLASTICS
(DONGGUAN)CO LTD)
- Add Enclosure: PA-765B(+) (CHI MEI CORPORATION)
- Add Secondary wiring: 1015 (XINYA ELECTRONIC CO LTD)
- Add Fuse: 5F (XC ELECTRONICS(SHENZHEN)CORP LTD)
- Add Lead-acid cell: OT1.3-12 (ZHANGZHOU HUAWEI POWER
SUPPLY TECHNOLOGY CO LTD)
- Add Lead-acid cell: PS1.3-12 (SHENZHEN POWER KINGDOM
CO LTD)
- Add Lead-acid cell: FP1213 (FIRST POWER TECHNOLOGY CO
LTD)

Correction 01. Report No.: E353512-D1000-1/A1/C1-UL


Correct the rating of the PWB of the secondary control board
from Min.150Deg.C to Min.130Deg.C, no safety relevant
impact.
Correct the Clause 7.2.9 to Pass.
Amendment 02. report No.: E353512-D1000-1/A2/C1-UL
Vichie Chen/Angela
2018-02-28 Add one additional model: JCB35T
Li