You are on page 1of 49

JJ Series TP Basic

JJ Series TP Solid
JJ Series TP Premium
JJ Accessories
JJ TT-Scan

Temperature calibrators

DS_Temperature_calibrators 01/2017

7
Temperature measurement and calibration instruments

Our temperature measurement and calibration instruments are designed and built with long-term reliability, innovative
calibration technology, and superior quality in mind. A comprehensive range of temperature products and services, with
standardized electrical signals are offered. Our products are engineered and manufactured to DAkkS standards (Germany‘s
equivalent to NIST).

Our temperature calibrators, simulators, controllers and indicators, transmitters and sensors help our customers to ensure the
consistent long-term quality of their products, and improve the efficiency of their production processes.
Our strength is our flexibility in developing customized solutions, based on our wealth of technical and application experience.

8
For laboratory, industry and service

Good reasons for a calibration


•• Maintain consistently high product quality
•• Meet industry standards and legal regulations
•• Optimize processes and boost productivity
•• Avoid unscheduled downtime

Temperature sensors are subject to mechanical, thermal and chemical stress. This results in a drift the longer the sensors are in use.
Only the regular calibration of the sensors provides information on the difference between the actual temperature and the measured
temperature and makes the specific drift visible. In measuring tasks, readings are often taken without regard to the fact that every
display value contains an error. In industrial applications even the smallest inaccuracy can lead to production errors.

9
Calibration with SIKA

Dry block calibrators and micro calibration baths are used to check and calibrate a wide range of temperature measuring
instruments and temperature sensors. Mechanical, electro-mechanical or electronic measurement equipment can be checked
with ease. The following sensors and instruments can be tested directly:

•• Contact-based immersion or surface temperature sensors


•• Sensors with special shapes and sizes
•• Non-contact infrared instruments and thermal imaging cameras

The compact and robust SIKA instruments are easy to transport, simple to use and offer all the features required for the specific
test. Our instruments are already standard in many development, research and testing labs, testing and inspection departments
and in the production and manufacturing sector.

Dr. Siebert & Kühn GmbH & Co. KG


Struthweg 7-9
34260 Kaufungen
Germany

akkreditiert durch die / accredited by the

Deutsche Akkreditierungsstelle GmbH


als Kalibrierlaboratorium im / as calibration laboratory in the
Deutschen Kalibrierdienst 000271

D-K-
Kalibrierschein Kalibrierzeichen 19636-01-00
Calibration certificate Calibration mark 2014-11

Gegenstand
Object
Mikrokalibrierbad
Micro calibration bath
Prüfprotokoll / Test Certificate
Dieser Kalibrierschein dokumentiert die Rück-
führung auf nationale Normale zur Darstellung
der Einheiten in Übereinstimmung mit dem
Internationalen Einheitensystem (SI).
Hersteller Dr. Siebert & Kühn GmbH & Co. KG
Manufacturer Die DAkkS ist Unterzeichner der multilateralen
34260 Kaufungen Übereinkommen der European co-operation for
Kalibriergegenstand Temperatur-Mikrobadkalibrator Die für die Kalibrierung verwendeten
Accreditation (EA)Calibration
und der International
object Labora- Temperature Liquidbath Calibrator Messeinrichtungen werden regelmäßig
Typ TP3M165E kalibriert und sind rückführbar auf die
Type tory Accreditation Cooperation (ILAC) zur ge-
nationalen Normale der Physikalisch
genseitigen Anerkennung der Kalibrierscheine. Technischen Bundesanstalt (PTB)
Fabrikat/Serien-Nr. Für die EinhaltungHersteller
einer angemessenen Frist zurSIKA Dr. Siebert & Kühn GmbH & Co. KG Deutschlands oder auf andere nationale
1411861
Serial number Wiederholung derManufacturer
Kalibrierung ist der Benutzer 34260 Kaufungen Normale. Die für diesen Vorgang erstellte
verantwortlich. Dokumentation kann eingesehen werden. Alle
erforderlichen Messdaten sind auf der(n)
Auftraggeber Mustermann AG This calibration certificate documents the
nachfolgenden Seite(n) dieses
Customer traceability to national
Mustergass 1 Typ standards, which realizeTPM165S Kalibrierscheins aufgelistet. Dieses
the units of measurement
Type according to the
-35°C / 165 °C Prüfprotokoll darf nur vollständig und
12345 Musterstadt International System of Units (SI). unverändert weiterverbreitet werden. Auszüge
The DAkkS is signatory to the multilateral oder Änderungen bedürfen der Genehmigung
Fabrikat/Serien-Nr. der ausstellenden Stelle. Prüfprotokolle ohne
agreements of the European
Serial numberco-operation for Unterschrift und Stempel haben keine
Auftragsnummer 100 741 854 Accreditation (EA) and of the International Gültigkeit.
Order No. Laboratory Accreditation Cooperation (ILAC) for
Auftraggeber The measuring devices used for calibration are
the mutual recognition of calibration certifica-
Customer
Anzahl der Seiten des Kalibrierscheines regularly calibrated and traceable to the
4 tes. national standards of the Physikalisch
Number of pages of the certificate
The user is obliged to have the object recali- Technische Bundesanstalt (PTB) or other
brated at appropriate intervals. national standards. The documents created for
Datum der Kalibrierung 2014-11-30 this procedure are available for viewing. All the
Date of calibration Anzahl der Seiten des Kalibrierscheines 2 necessary measured data can be found on the
Number of pages of the certificate following page(s) of this calibration certificate.
Dieser Kalibrierschein darf nur vollständig und unverändert weiterverbreitet werden. Auszüge oder Änderungen bedürfen der This test certificate may not be reproduced
Genehmigung sowohl der Deutschen Akkreditierungsstelle GmbH als auch des ausstellenden Datum der Kalibrierung
Kalibrierlaboratoriums. Dez 2014 other than in full except with the permission of
Date of calibration the issuing office. Test certificates without
Kalibrierscheine ohne Unterschrift haben keine Gültigkeit.
signature and seal are not valid.
This calibration certificate may not be reproduced other than in full except with the permission of both the Deutsche
Akkreditierungsstelle GmbH and the issuing laboratory. Calibration certificates without signature are not valid.

Datum Stellv. Leiter des Kalibrierlaboratoriums Bearbeiter


Date Assistant Head of the calibration laboratory Person in charge

Umgebungsbedingungen Raumtemperatur/ Amb. temperature (23 ± 2) °C


Ambient conditions Rel. Luftfeuchte/ Rel. air humidity (50 ± 20) %
Luftdruck/ Amb. pressure (990 ± 30) mbar
2014-11-30 Olaf Schilling Olaf Schilling

Dr. Siebert & Kühn GmbH & Co. KG * Struthweg 7-9 * D-34260 Kaufungen * Telefon: 0 56 05/ 8 03 - 0 * Fax: 0 56 05/ 803 - 54/ 60
Verwendete Normale Pt100, SN ASL-02, 3730 D-K-17734-01-00 / 2012-09
Used standards DMM Keithley,SN 596023, 0249 DKD-K-13901/ 2013-10

Abgleich durchgeführt mit Silikonöl 10 cSt Bechereinsatz


Adjustement carried out with Siliconoil 10 cSt tub insert

SIKA Dr. Siebert & Kühn GmbH & Co. KG • Struthweg 7-9 • 34260 Kaufungen • Germany
Phone +49 5605-803-0 • Fax +49 5605-803-54 • info@SIKA.net • www.SIKA.net

10
Calibration function for every need

Micro bath function Black body function


The use of calibration liquids offers certain advantages if A patented infrared calibration sleeve is used to calibrate IR
temperature sensors with an unusual shape and size are to pyrometers or thermal imaging cameras. The special surface
be tested. The test item is immersed directly into the liquid structure and the asymmetrical shapes create a “cavity radiator“
without an insulating air gap, resulting in direct temperature with an emission factor of 0.9994, prevent the reflection of
contact between the calibrator and the test item. The liquid, interference radiation and emit the required temperature
such as silicone oil, is chosen depending on the calibration in an ideal form. The pyrometer under test is simply held at
temperature required. The continuous adjustment of the the specified distance above the measurement opening of the
magnetic stirrer together with the removable sensor basket calibrator, thereby forming the desired measurement area on
agitates the calibration liquid to create a large measuring zone. the bottom for the calibration to be performed. A support base
Furthermore, the sensor basket guarantees unhindered stirring can be fitted directly on the unit.
and helps protect the sensor.

Dry block function Surface temperature function


The dry block adapter sleeve is used for straight temperature Surface temperature sensors are calibrated using special
sensors with almost any length and diameter. Every adapter sleeves that are fitted vertically with the required contact
sleeve can be equipped with a single diameter bore, or multi- force. Switching calibration control to the external reference
diameter bore. Bore diameters ranging from 1.5 mm to 25.5 sensor creates the best possible temperature reference point
mm are offered in 0.5 mm increments (see page 15 for our on the surface of the sleeve. The reference sensor is located
standard adapter sleeves). The optimum thermal coupling from directly beneath the abutting face of the sleeve. The sleeve is
the block to the test instrument is achieved with the appropriate designed in such a way that the best temperature homogeneity
adapter sleeve. The dry block covers the entire temperature is achieved in the centre of the abutting face. The special design
range without the need to change the calibration medium. of the abutting face enables good thermal contact. There is
no need to use a thermally conductive paste or other thermal
conduction aids.

11
Temperature ranges TP series

Temperature range Function / Tolerance


-55 °C -35 °C -10 °C RT 100 °C 165 °C 200 °C 225 °C 255 °C 400 °C 450 °C 650 °C 850 °C 1300 °C DB IR LI
Dry block Infrared Micro
bath
-55 °C ... 200 °C ±0.4 °C

±0.2 °C

±0.2 °C

-35 °C ... 165 °C ±1 °C

±0.4 °C

±0.2 °C

±0.2 °C

±0.1 °C

±0.4 °C

±0.2 °C

±0.2 °C ±0.5 °C

±0.1 °C

±0.3 °C ±0.5 °C ±0.1 °C

-10 °C ... 100 °C ±0.05 °C

RT …200 °C ±1 °C

RT …255 °C ±0.2 °C

±0.3 °C ±0.5 °C ±0.2 °C

RT …450 °C ±0.6 °C

±0.3 °C

±0.3 °C ±0.5 °C

RT …650 °C ±1 °C

±0.8 °C

±0.4 °C

±0.2 °C

RT …700 °C ±0.4 °C

RT …850 °C ±1 °C

400 °C…1300 °C ±2 °C

12
Features Block dimensions [mm]
SU Integrated External Interface Ø Diameter Depth
Model
Surface measuring reference 18 28 60 7 x 6.5 100 150 170 200
instrument sensor
ü ü Dry block TP 17 200

ü ü ü Dry block TP 17 200 S

ü ü ü ü Dry block TP 37 200 E

ü ü Dry block TP 17 165 M

ü ü Dry block TP 17 165

ü ü ü Dry block TP 17 165 S

ü ü ü ü Dry block TP 37 165 E

ü ü ü ü ü Dry block TP 38 165

ü ü Dry block TP 17 166

ü ü ü Dry block TP 17 166 S

±1 °C ü ü ü ü Multifunction block TP 37 166 E

ü ü ü Micro bath TP M 165 S

±1 °C ü ü ü ü Multifunction bath TP 3M 165 E2

ü ü ü Dry block TP 17 Zero

ü ü Dry block TP 18 200 E

ü ü ü Micro bath TP M 255 S

±1°C ü ü ü ü Multifunction bath TP 3M 255 E

ü ü Dry block TP 17 450

ü ü ü Multifunction block TP 17 450 S

±1 °C ü ü ü ü Multifunction block TP 17 450 S-U

ü ü Dry block TP 17 650 M

ü ü Dry block TP 17 650

ü ü ü Dry block TP 17 650 S

ü ü ü ü ü Dry block TP 38 650

ü ü ü ü Dry block TP 37 700 E

ü ü ü ü Dry block TP 18 850 E

ü ü ü Dry block TP 28 1300 E

13
Requirements placed on temperature calibrators

Calibration functions

 Dry block function


 Micro bath function
 Infrared function
 Surface function

Your benefit:
 General purpose use through Performance
multifunctional calibrators

Select the desired properties


depending on the series and model
 Wide temperature range
 Powerful processors
 Variable test unit mounts
Calibration order and job site  Different stabilities

Your benefit:
Handy and compact dimensions  High-performance calibrators to
and weight. match your needs
That lets SIKA calibrators meet
all requirements that frequently
changing job sites and individual
test conditions entail.

Your benefit:
 Simple and fast
on site use

Reliable system accuracy

Even during production, SIKA


calibrators are put through Connections
various tests and a measurement
uncertainty consideration in accor-
Whether standard PC interfaces,
dance with the DAkkS directive
Ethernet, USB connection or WiFi
including comprehensive documen-
functions, always the right connection
tation of the measurement results.
exactly as needed.
Your benefit:
 Function expansion possible
 Reliable and precise  Connection of accessories
calibration results  Data backup

Your benefit:
 Flexibility and future-proof

14 14
Series TP Basic

Efficiency and portability are distinguishing features of the temperature calibrators of the TP Basic series. It consists of dry block
calibrators which cover a wide temperature range and are used on-site e.g. in the marine sector.
Designed to ensure a comfortable calibration of temperature sensors, they impress with an easy operation and a thoughtful use of
different automatic functions.

Adapter sleeve - 60 mm

The optimal thermal coupling from the block to the test item
is achieved by an exactly fitted adapter sleeve. These adapter
sleeves can be produced according to customer specifications,
enabling the simultaneous calibration of different temperature
sensors.
This results in an easy, fast and efficient test of various test
items without compromising quality.

Some of the dry block calibrators have a large 60 mm


block borehole and can hold several test items at
once.

15
TP Basic series
TP 17 200

Technical data
Type TP 17 200
Type TP 17 200
Control sensor Internal
Dry block
Temperature range* -55...200 °C -67...392 °F
Accuracy ±0.4 °C ±0.72 °F
Stability ±0.1 °C ±0.18 °F
Measurement zone 110...150 mm 4.3...5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range -60...200 °C -76....392 °F
Resolution 0.1 °C 0.1 °F
General data
Dimensions
 Width 210 mm 8.27 in
 Height 380 + 50 mm 14.96 + 1.97 in
 Depth 300 mm 11.81 in
Weight Approx. 12.5 kg Approx. 27.56 lbs
Power supply 100...240 VAC, 50 / 60 Hz
Power consumption Approx. 555 W
* At an ambient temperature of 20 °C / 68 °F Scope of delivery
•• Test certificate
•• Mains cable
•• Insert exchange tool
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• DAkkS certificate
•• Works certificate

The TP 17 200 temperature calibrator, which is also


known as TP COOL, works in a temperature range
from -55 to 200 °C.

16
TP 17 165 M

Technical data
Type TP 17 165 M
Type TP 17 165 M
Control sensor Internal
Dry block
Temperature range* -35...165 °C -31...329 °F
Accuracy ±1 °C ±1.8 °F
Stability ±0.1 °C ±0.18 °F
Measurement zone 110...150 mm 4.3...5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range -50...165 °C -58...329 °F
Resolution 1 °C 1 °F
General data
Dimensions
 Width 210 mm 8.27 in
 Height 380 + 50 mm 14.96 + 1.97 in
 Depth 300 mm 11.81 in
Weight Approx. 10 kg Approx. 22.05 lbs
Power supply 100...240 VAC, 50 / 60 Hz
Power consumption Approx. 375 W
* At an ambient temperature of 20 °C / 68 °F Scope of delivery
•• Test certificate
•• Mains cable
•• Insert exchange tool
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• DAkkS certificate
•• Works certificate

17
TP 17 165

Technical data
Type TP 17 165
Type TP 17 165
Control sensor Internal
Dry block
Temperature range* -35...165 °C -31...329 °F
Accuracy ±0.4 °C ±0.72 °F
Stability ±0.1 °C ±0.18 °F
Measurement zone 110...150 mm 4.3...5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range -50...165 °C -58...329 °F
Resolution 0,1 °C 0.1 °F
General data
Dimensions
 Width 210 mm 8.27 in
 Height 380 + 50 mm 14.96 + 1.97 in
 Depth 300 mm 11.81 in
Weight Approx. 10 kg Approx. 22.05 lbs
Power supply 100...240 VAC, 50 / 60 Hz
Power consumption Approx. 375 W
Scope of delivery * At an ambient temperature of 20 °C / 68 °F

•• Test certificate
•• Mains cable
•• Insert exchange tool
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• DAkkS certificate
•• Works certificate

18
TP 17 166

Technical data
Type TP 17 166
Type TP 17 166
Control sensor Internal
Dry block
Temperature range* -35...165 °C -31...329 °F
Accuracy ±0.4 °C ±0.72 °F
Stability ±0.1 °C ±0.18 °F
Measurement zone 110...150 mm 4.3...5.91 in
Block dimensions
 Diameter Ø 60 mm Ø 2.36 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range -50...165 °C -58...329 °F
Resolution 0.1 °C 0.1 °F
General data
Dimensions
 Width 210 mm 8.27 in
 Height 380 + 50 mm 14.96 + 1.97 in
 Depth 300 mm 11.81 in
Weight Approx. 10 kg Approx. 22.05 lbs
Power supply 100...240 VAC, 50 / 60 Hz
Power consumption Approx. 375 W
Scope of delivery * At an ambient temperature of 20 °C / 68 °F
•• Test certificate
•• Mains cable
•• Insert exchange tool
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• DAkkS certificate
•• Works certificate

The TP 17 166 temperature calibrator model has a


very large calibration volume.

19
TP 18 200 E

Technical data
Type TP 18 200 E
Type TP 18 200 E
Control sensor Internal
Dry block
Temperature range* RT...200 °C RT...392 °F
Accuracy ±1 °C ±1.8 °F
Stability ±0.1 °C ±0.18 °F
Measurement zone 60...100 mm 2.36...3.94 in
Block dimensions
 Diameter Ø 18 mm Ø 0.71 in
 Depth 100 mm 3.94 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range 0...200 °C 32...392 °F
Resolution 1 °C 1 °F
General data
Dimensions
 Width 220 mm 8.66 in
 Height 96 mm 3.78 in
 Depth 230 + 50 mm 12.99 + 2.76 in
Weight Approx. 4 kg 8.82 lbs
Power supply 230 / 240 VAC, 50 / 60 Hz,
optional 100...115 VAC, 50 / 60 Hz
Power consumption Approx. 200 W
* At an ambient temperature of 20 °C / 68 °F

Scope of delivery
•• Test certificate
•• Mains cable
•• Insert exchange tool
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• DAkkS certificate
•• Works certificate

20
TP 17 450

Technical data
Type TP 17 450
Type TP 17 450
Control sensor Internal
Dry block
Temperature range* RT...450 °C RT...842 °F
Accuracy ±0.6 °C ±1.08 °F
Stability ±0.1 °C ±0.18 °F
Measurement zone 110...150 mm 4.3...5.91 in
Block dimensions
 Diameter Ø 60 mm Ø 2.36 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range 0...450 °C 32...842 °F
Resolution 0.1 °C 0.1°F
General data
Dimensions
 Width 150 mm 5.91 in
 Height 330 + 70 mm 12.99 + 2.76 in
 Depth 270 mm 10.63 in
Weight Approx. 7.5 kg Approx. 16.53 lbs
Power supply 230 / 240 VAC, 50 / 60 Hz
Power consumption Approx. 2000 W
* At an ambient temperature of 20 °C / 68 °F
Scope of delivery
•• Test certificate
•• Mains cable
•• Insert exchange tool
•• Operating manual

Accessories
•• Transport case
•• Service transport bag
•• Adapter sleeve
•• DAkkS certificate
•• Works certificate

The TP 17 450 temperature calibrator model has a


very large calibration volume.

21
TP 17 650 M

Technical data
Type TP 17 650 M
Type TP 17 650 M
Control sensor Internal
Dry block
Temperature range* RT...650 °C RT...1202 °F
Accuracy ±1 °C ±1.8 °F
Stability ±0.1 °C ±0.2 °F
Measurement zone 110...150 mm 4.3...5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range 0...650 °C 32...1202 °F
Resolution 1 °C 1 °F
General data
Dimensions
 Width 150 mm 5.91 in
 Height 330 + 70 mm 12.99 + 2.76 in
 Depth 270 mm 10.63 in
Weight Approx. 7.5 kg Approx. 16.53 lbs
Power supply 230 / 240 VAC, 50 / 60 Hz,
optional 100...115 VAC, 50 / 60 Hz
Scope of delivery optional 100...240 VAC, 50 / 60 Hz
Power consumption Approx. 1000 W
•• Test certificate
* At an ambient temperature of 20 °C / 68 °F
•• Mains cable
•• Insert exchange tool
•• Operating manual

Accessories
•• Transport case
•• Service transport bag
•• Adapter sleeve
•• DAkkS certificate
•• Works certificate

22
TP 17 650

Technical data
Type TP 17 650
Type TP 17 650
Control sensor Internal
Dry block
Temperature range* RT...650 °C RT...1202 °F
Accuracy ±0.8 °C ±1.44 °F
Stability ±0.1 °C ±0.18 °F
Measurement zone 110...150 mm 4.3...5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range 0...650 °C 32...1202 °F
Resolution 0.1 °C 0.1 °C between
RT...999.9 °F,
else 1 °F
General data
Dimensions
 Width 150 mm 5.91 in
 Height 330 + 70 mm 12.99 + 2.76 in
 Depth 270 mm 10.63 in
Weight Approx. 7.5 kg Approx. 16.53 lbs

Scope of delivery Power supply 230 / 240 VAC, 50 / 60 Hz,


optional 100...115 VAC, 50 / 60 Hz
•• Test certificate optional 100...240 VAC, 50 / 60 Hz
•• Mains cable Power consumption Approx. 1000 W
•• Insert exchange tool * At an ambient temperature of 20 °C / 68 °F
•• Operating manual

Accessories
•• Transport case
•• Service transport bag
•• Adapter sleeve
•• DAkkS certificate
•• Works certificate

23
TP 18 850 E

Technical data
Type TP 18 850 E
Type TP 18 850 E
Control sensor Internal
Dry block
Temperature range* RT...850 °C RT...1562 °F
Accuracy ±1 °C ±1.8 °C
Stability ±0.1 °C ±0.18 °C
60...100 mm or 2.36...3.94 in or
Measurement zone
160...200 mm 6.3...7.87 in
Block dimensions
 Diameter Ø 18 or 28 mm Ø 0.71 or 1.1 in
 Depth 100 or 200 mm 3.94 or 7.87 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range 0...850 °C 32...1562 °F
Resolution 1 °C 1 °F
General data
Dimensions
 Width 430 mm 16,93 in
 Height 190 mm 7.48 in
 Depth 410 + 50 mm or 16,14 + 1.97 in or
510 + 50 mm 20.08 + 1.97 in
Weight Approx. 14 kg or Approx. 30.86 lbs or
16 kg 35.27 lbs
Power supply 230 / 240 VAC, 50 / 60 Hz
Power consumption Approx. 2000 W
Scope of delivery
* At an ambient temperature of 20 °C / 68 °F
•• Test certificate
•• Mains cable
•• Insert exchange tool
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• DAkkS certificate
•• Works certificate

24
Series TP Solid

With the temperature calibrators in the TP Solid series, the emphasis is on flexibility: aside from dry block calibrators, there are
also micro calibration baths, with which almost any temperature sensor regardless of its form can be tested. Both offer an easy as
well as intuitive operation and fast access to comprehensive functions.

When used as liquid bath calibrator, the temperature sensors Moreover, an external reference sensor makes it possible
are directly immersed into the calibration liquid. This results to test different temperature monitoring equipment without
in a direct temperature contact between the calibrator regard to shape, size or measurement method. You can choose
and the test item without an insulating air gap. Using between internal and external reference by using a switch on
interchangeable calibration sleeves, these calibrators can be the calibrator. All devices of this series are equipped with a
also used for calibration of infrared pyrometers or surface serial PC interface for computer-assisted monitoring of the
temperature sensors. calibration process.

This flexibility makes the temperature calibrators of the TP


Solid series ideal for the use in machine and plant construction.

25
-25-
TP Solid series
TP 17 200 S

Technical data
Type TP 17 200 S
Type TP 17 200 S
Control sensor Internal
Dry block
Temperature range* -55...200 °C -67...392 °F
Accuracy ±0.2 °C ±0.36 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.3... 5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range -60...200 °C -76...392 °F
Resolution 0.01 °C between 0.01 °F between
-9.99...99.99 °C, -9.99...99.99 °F,
else 0.1 °C else 0.1 °F
General data
Interface RS485,
RS232 or USB (optional)
Dimensions
 Width 210 mm 8.27 in
Scope of delivery  Height 380 + 50 mm 14.96 + 1.97 in
 Depth 300 mm 11.81 in
•• Test certificate
Weight Approx. 12.5 kg Approx. 27.56 lbs
•• Mains cable
Power supply 100...240 VAC, 50 / 60 Hz
•• Insert exchange tool
Power consumption Approx. 555 W
•• Operating manual
* At an ambient temperature of 20 °C / 68 °F

Accessories
•• Transport case
•• Adapter sleeve
•• PC software
•• PC cable
•• DAkkS certificate
•• Works certificate

The TP 17 200 S temperature calibrator, which is also


known as TP COOL, works in a temperature range
from -55 to 200 °C.

26
TP 17 165 S

Technical data
Type TP 17 165 S
Type TP 17 165 S
Control sensor Internal
Dry block
Temperature range* -35...165 °C -31...329 °F
Accuracy ±0.2 °C ±0.36 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.3... 5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range -50...165 °C -58...329 °F
Resolution 0.01 °C between 0.01 °F between
-9.99...99.99 °C, -9.99...99.99 °F,
else 0.1 °C else 0.1 °F
General data
Interface RS485,
RS232 or USB (optional)
Dimensions
 Width 210 mm 8.27 in
Scope of delivery  Height 380 + 50 mm 14.96 + 1.97 in
 Depth 300 mm 11.81 in
•• Test certificate
Weight Approx. 10 kg Approx. 22.05 lbs
•• Mains cable
Power supply 100...240 VAC, 50 / 60 Hz
•• Insert exchange tool
Power consumption Approx. 375 W
•• Operating manual
* At an ambient temperature of 20 °C / 68 °F

Accessories
•• Transport case
•• Adapter sleeve
•• PC software
•• PC cable
•• DAkkS certificate
•• Works certificate

27
TP 17 166 S

Technical data
Type TP 17 166 S
Type TP 17 166 S
Control sensor Internal
Dry block
Temperature range* -35...165 °C -31...329 °F
Accuracy ±0.2 °C ±0.36 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.3... 5.91 in
Block dimensions
 Diameter Ø 60 mm Ø 2.36 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range -50...165 °C -58...329 °F
Resolution 0.01 °C between 0.01 °F between
-9.99...99.99 °C, -9.99...99.99 °F,
else 0.1 °C else 0.1 °F
General data
Interface RS485,
RS232 or USB (optional)
Dimensions
 Width 210 mm 8.27 in
 Height 380 + 50 mm 14.96 + 1.97 in Scope of delivery
 Depth 300 mm 11.81 in
•• Test certificate
Weight Approx. 10 kg Approx. 22.05 lbs
•• Mains cable
Power supply 100...240 VAC, 50 / 60 Hz
•• Insert exchange tool
Power consumption Approx. 375 W
* At an ambient temperature of 20 °C / 68 °F
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• PC software
•• PC cable
•• DAkkS certificate
•• Works certificate

The TP 17 166 S temperature calibrator model has a


very large calibration volume.

28
TP M 165 S

Technical data
Type TP M 165 S
Type TP M 165 S
Control sensor Internal
Micro Bath
Temperature range* -35...165 °C -31...329 °F
Accuracy ±0.1 °C ±0.18 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.33...5.91 in
Block dimensions
 Diameter Ø 60 mm Ø 2.36 in
 Depth 170 mm 6.69 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range -50...165 °C -58...329 °F
Resolution 0.01 °C between 0.01 °F between
-9.99...99.99 °C, -9.99...99.99 °F,
else 0.1 °C else 0.1 °F
General data
Interface RS485,
RS232 or USB (optional)
Dimensions
 Width 210 mm 8.27 in
Scope of delivery  Height 380 + 50 mm 14.96 + 1.97 in
•• Test certificate  Depth 300 mm 11.81 in
Weight Approx. 12.5 kg Approx. 27.56 lbs
•• Mains cable
Power supply 100...240 VAC, 50 / 60 Hz
•• Operating manual
Power consumption Approx. 375 W
•• Sensor basket, suction pump, transport cover
* At an ambient temperature of 20 °C / 68 °F
•• Magnetic stirrer with magnet lifter
•• Sensor lid with 5 silicone plugs

Accessories
•• Calibration liquid
•• Transport case
•• Tube insert
•• PC software
•• PC cable
•• DAkkS certificate
•• Works certificate

29
TP M 255 S

Technical data
Type TP M 255 S
Type TP M 255 S
Control sensor Internal
Micro Bath
Temperature range* RT...255 °C RT...491 °F
Accuracy ±0.2 °C ±0.36 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.33...5.91 in
Block dimensions
 Diameter Ø 60 mm Ø 2.36 in
 Depth 170 mm 6.69 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range 0...255 °C 32...491 °F
Resolution 0.01 °C between 0.01 °F between
RT...99.99 °C, RT...99.99 °F,
else 0.1 °C else 0.1 °F
General data
Interface RS485,
RS232 or USB (optional)
Dimensions
 Width 150 mm 5.91 in
Scope of delivery  Height 330 + 70 mm 12.99 + 2.76 in
 Depth 270 mm 10.63 in
•• Test certificate
Weight Approx. 7.5 kg Approx. 16.53 lbs
•• Mains cable
Power supply 230 / 240 VAC, 50 / 60 Hz,
•• Operating manual optional 100...115 VAC, 50 / 60 Hz
•• Sensor basket, suction pump, transport cover optional 100...240 VAC, 50 / 60 Hz

•• Magnetic stirrer with magnet lifter Power consumption Approx. 1000 W


* At an ambient temperature of 20 °C / 68 °F
•• Sensor lid with 5 silicone plugs

Accessories
•• Calibration liquid
•• Transport case
•• Service transport bag
•• Tube insert
•• PC software
•• PC cable
•• DAkkS certificate
•• Works certificate

30
TP 17 450 S

Technical data
Type TP 17 450 S
Type TP 17 450 S
Control sensor Internal
Dry block
Temperature range* RT...450 °C RT...842 °F
Accuracy ±0.3 °C ±0.54 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.33...5.91 in
Infrared
Temperature range* RT...450 °C RT...842 °F
Accuracy ±0.5 °C ±0.9 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110 mm 4.33 in
Block dimensions
 Diameter Ø 60 mm Ø 2.36 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range 0...450 °C 32...842 °F
Resolution 0.01 °C between 0.01 °F between
RT...99.99 °C, RT...99.99 °F,
else 0.1 °C else 0.1 °F
General data Scope of delivery
Interface RS485, •• Test certificate
RS232 or USB (optional)
•• Mains cable
Dimensions
•• Insert exchange tool
 Width 150 mm 5.91 in
 Height 330 + 70 mm 12.99 + 2.76 in •• Operating manual
 Depth 270 mm 10.63 in
Weight Approx. 7.5 kg Approx. 16.53 lbs Accessories
Power supply 230 / 240 VAC, 50 / 60 Hz •• Transport case
Power consumption Approx. 2000 W •• Service transport bag
* At an ambient temperature of 20 °C / 68 °F •• Adapter sleeve
•• PC software
•• PC cable
•• DAkkS certificate
•• Works certificate

The TP 17 450 S temperature calibrator model has a


very large calibration volume.

31
TP 17 450 S-U

Technical data
Type TP 17 450 S-U
Type TP 17 450 S-U
Control sensor Switchable internal / external
Dry block
Temperature range* RT...450 °C RT...842 °F
Accuracy ±0.3 °C ±0.54 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.33...5.91 in
Infrared
Temperature range* RT...450 °C RT...842 °F
Accuracy ±0.5 °C ±0.9 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110 mm 4.33 in
Surface
Temperature range* RT...400 °C RT...752 °F
Accuracy ±1 °C ±1.8 °F
Stability ±0.2 °C ±0.36 °F
Measurement zone 110...150 mm 4.33...5.91 in
Block dimensions
 Diameter Ø 60 mm Ø 2.36 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F Scope of delivery
Display range 0...450 °C 32...842 °F •• Test certificate
Resolution 0.01 °C between 0.01 °F between •• Mains cable
RT...99.99 °C, RT...99.99 °F,
•• Insert exchange tool
else 0.1 °C else 0.1 °F
General data •• Operating manual
Interface RS485, •• External calibration reference sensor TF 650-3-300
RS232 or USB (optional)
Dimensions Optional accessories
 Width 150 mm 5.91 in •• Transport case
 Height 330 + 70 mm 12.99 + 2.76 in
•• Service transport bag
 Depth 270 mm 10.63 in
•• Adapter sleeve
Weight Approx. 7.5 kg Approx. 16.53 lbs
Power supply 230 / 240 VAC, 50 / 60 Hz •• Infrared calibration sleeve

Power consumption Approx. 2000 W •• Surface calibration sleeve


•• PC software
•• PC cable
•• DAkkS certificate
•• Works certificate

32
TP 17 650 S

Technical data
Type TP 17 650 S
Type TP 17 650 S
Control sensor Internal
Dry block
Temperature range* RT...650 °C RT...1202 °F
Accuracy ±0.4 °C ±0.72 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.3... 5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range 0...650 °C 32...1202 °F
Resolution 0.01 °C between 0.01 °F between
RT...99.99 °C, RT...99.99 °F,
else 0.1 °C 0.1 °F between
100.0...999.9 °F,
else 1 °F
General data
Interface RS485,
RS232 or USB (optional)
Scope of delivery Dimensions
•• Test certificate  Width 150 mm 5.91 in
 Height 330 + 70 mm 12.99 + 2.76 in
•• Mains cable
 Depth 270 mm 10.63 in
•• Insert exchange tool Weight Approx. 7.5 kg Approx. 16.53 lbs
•• Operating manual Power supply 230 / 240 VAC, 50 / 60 Hz,
optional 100...115 VAC, 50 / 60 Hz
Accessories optional 100...240 VAC, 50 / 60 Hz
•• Transport case Power consumption Approx. 1000 W
•• Service transport bag * At an ambient temperature of 20 °C / 68 °F

•• Adapter sleeve
•• PC software
•• PC cable
•• DAkkS certificate
•• Works certificate

33
TP 28 1300 E

Technical data
Type TP 28 1300 E
Type TP 28 1300 E
Control sensor Internal
Dry block
Temperature range* 400...1300 °C 752...2372 °F
Tolerance ±2 °C ±3.6 °F
Stability ±0.5 °C ±0.9 °F
Measurement zone at 200 mm at 7.87 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 200 mm 7.87 in
Display unit
Display 2-line, 4-digit display
Red / green, unit °C / °F
Display range 0...1300 °C 32...2372 °F
Resolution 1 °C 1 °F
General data
Interface RS485,
RS232 or USB (optional)
Dimensions
 Width 510 mm 20.08 in
 Height 290 mm 11.42 in
 Depth 415 + 100 mm 16.34 + 3.94 in
Scope of delivery Weight Approx. 17,5 kg Approx. 38.58 lbs
•• Test certificate Power supply 100...240 VAC, 50 / 60 Hz

•• Mains cable Power consumption Approx. 1000 W


* At an ambient temperature of 20 °C / 68 °F
•• Insert exchange tool
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• PC software
•• PC cable
•• DAkkS certificate
•• Works certificate

34
TP ZERO

Technical data
Type TP ZERO
Type TP ZERO
Control sensor Internal
Dry block
Temperature range* -10...100 °C 14...148 °F
Accuracy ±0.05 °C at 0 °C ±0.09 °F at 32 °F
Stability ±0.05 °C at 0 °C ±0.09 °F at 32 °F
Measurement zone 110...150 mm 4.3... 5.91 in
Block dimensions
 Diameter 7 drillings with 7 drillings with
Ø 6.5 mm Ø 0.26 in
 Depth 150 mm 5.91 in
Display unit
Display 1-line, 4-digit display
Red, unit °C (°F optional)
Display range -10...100 °C 14...212 °F
Resolution 0.1 °C 0.1 °F
General data
Dimensions
 Width 160 mm 6.3 in
 Height 320 + 50 mm 12.6 + 1.97 in
 Depth 230 mm 9.06 in
Weight Approx. 7 kg Approx. 15.43 lbs
Power supply 100...240 VAC, 50 / 60 Hz Scope of delivery
Power consumption Approx. 225 W •• Test certificate
* At an ambient temperature of 20 °C / 68 °F •• Mains cable
•• Operating manual

Accessories
•• Transport case
•• PC software
•• PC cable
•• DAkkS certificate
•• Works certificate

The TP ZERO temperature calibrator is ideal for


creating the ice point.

35
Series TP Premium

Optimal performance and outstanding ease of use are A growing variety of supported temperature ranges covers more
distinguishing features of the TP Premium series calibrators. and more temperature sensors on the market. These sensors
With the help of the intuitive menu structure, all the necessary can be calibrated with a resolution of up to 0.001 °C and meet the
entries can be made quickly and easily. Whether on the two colour, highest standards e.g. of the food and pharmaceutical industry.
graphic display or on the large touch screen of the TP Touch
series – block and set temperature as well as the difference and The comprehensive range of accessories of the TP Premium
the variance of the stability can be set and displayed. series allows time-saving calibration setups. For example, test
items can be identified by a barcode scanner or a running or
already finished calibration process can be shown on a PC or
smartphone via a Wi-Fi connection anywhere on the world. With
the ability to connect an external measuring bridge directly to the
calibrators, many test items can be simultaneously calibrated ¬
including the automatic creation of individual works certificates.

36
Quickly controlled - quickly calibrated!

For economic reasons, the machine downtimes that arise during The use a special control algorithm achieves a significantly
the calibration process should be kept as short as possible. shorter calibration time and - unique for portable calibrators -
SIKA has the solution: the worldwide first temperature calibrators temperature stability in the millikelvin range.
that shorten the calibration time by as much as 30 %. “Time is money” - the benefits for the user are obvious: shorter
Experience and knowledge gained from the aerospace industry calibration times mean shorter machine downtimes. Depending
make the innovative temperature calibrators of the TP Premium on the production process, using the new TP Premium series
series the fastest and most temperature-stable calibrators on calibrators makes substantial cost savings possible.
the market. They come equipped with a temperature controller
that is based on a completely novel operating principle.
In contrast to the mode of functioning of conventional
calibrators, in the new TP Premium the properties of the test
unit are virtually modelled.

Direct control of the test unit


By taking the thermal inertia of the test unit into consideration
in the calibration volume in the frame of a model-based control,
on top of outstanding temperature stability the following goals
can also be achieved:

•• Shortening of the waiting times until thermal equilibrium


is reached from about 30 minutes down to between 25 and
15 minutes.
•• Energy savings across the service life of a temperature
calibrator of up to 5 MWh.
•• Certainty of reaching the thermal equilibrium so that
unnecessary calibration uncertainties are avoided.

Regelung Prüfling auf 50°C


50,01

Temperaturanzeige Prüfling
Temperaturanzeige T [°C]

50

49,99
0 5 10 15 20 25 30 35 40 45
Zeit t [min]

37
-37-
TP 38 165

Technical data
Type TP 38 165
Type TP 38 165
Control sensor switchable internal / external
Dry block
Temperature range* -35...165 °C -31...329 °F
Accuracy ±0.1 °C ±0.18 °F
Stability 0.01...0.05 °C 0.02...0.09 °F
Measurement zone 110...150 mm 4.33...5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display Monochrome, graphic display
Units °C / °F / K / Ω / mV / mA
Display range -50...165 °C -58...329 °F
Resolution 0.01 °C 0.01 °F
General data
Interface RS232,
USB (optional)
Dimensions
 Width 153 mm 6.02 in
 Height 347 mm 13.66 in
 Depth 348 mm 13.70 in
Scope of delivery Weight Approx. 12 kg Approx. 26.46 lbs
•• Test certificate Power supply 100...240 VAC, 50 / 60 Hz

•• Mains cable Power consumption Approx. 400 W

•• Insert exchange tool * At an ambient temperature of 20 °C / 68 °F

•• Operating manual
•• Precision measuring instrument

Accessories
•• Transport case
•• Adapter sleeve
•• PC software
•• PC cable
•• External calibration reference sensor
TF 255-3-300 or TFEE 255-3-300
•• DAkkS certificate
•• Works certificate

38
TP 38 650

Technical data
Type TP 38 650
Type TP 38 650
Control sensor Switchable internal / external
Dry block
Temperature range* RT...650 °C RT...1202 °F
Accuracy ±0.2 °C ±0.36 °F
Stability 0.03...0.1 °C 0.05...0.18 °F
Measurement zone 110...150 mm 4.33... 5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display Monochrome, graphic display
Units °C / °F / K / Ω / mV / mA
Display range 0...650 °C 32...1202 °F
Resolution 0.01 °C 0.01 °F
General data
Interface RS232,
USB (optional)
Dimensions
 Width 153 mm 6.02 in
 Height 347 mm 13.66 in
 Depth 348 mm 13.70 in
Weight Approx. 10 kg Approx. 22.05 lbs Scope of delivery
Power supply 230 / 240 VAC, 50 / 60 Hz, •• Test certificate
optional 100...115 VAC, 50 / 60 Hz
•• Mains cable
Power consumption Approx. 1000 W
•• Insert exchange tool
* At an ambient temperature of 20 °C / 68 °F
•• Operating manual
•• Precision measuring instrument

Accessories
•• Transport case
•• Adapter sleeve
•• PC software
•• PC cable
•• External calibration reference sensor
TF 650-3-300 or TFEE 650-3-300
•• DAkkS certificate
•• Works certificate

39
TP 37 200 E

Technical data
Type TP 37 200 E
Type TP 37 200 E
Control sensor Switchable internal / external
Dry block
Temperature range* -55...200 °C -67...392 °F
Tolerance ±0.2 °C ±0.36 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.3...5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display Brilliant Color-Touchscreen (7“)
Viewing angle 120...140°
Brightness 400 cd / m²
Unit °C / °F / K
Display range -60...200 °C -76...392 °F
Resolution 0.1 / 0.01 / 0.001 °C / °F / K
General data
Interface Ethernet
3 x USB
Dimensions
 Width 210 mm 8.27 in
Scope of delivery  Height 380 + 50 mm 14.96 + 1.97 in
 Depth 300 mm 11.81 in
•• Test certificate
Weight Approx. 12.5 kg Approx. 27.56 lbs
•• Mains cable
Power supply 100...240 VAC, 50 / 60 Hz
•• Insert exchange tool
Power consumption Approx. 555 W
•• PC- and network cable
* At an ambient temperature of 20 °C / 68 °F
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• External reference sensor TF 255-3-300
•• PC software
•• Network-switch, barcode-reader, WLAN-router
•• DAkkS certificate
•• Works certificate

40
TP 37 165 E

Technical data
Type TP 37 165 E
Type TP 37 165 E
Control sensor Switchable internal / external
Dry block
Temperature range* -35...165°C -31...329 °F
Accuracy ±0.2 °C ±0.36 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.33...5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.1 in
 Depth 150 mm 5.91 in
Display unit
Display Brilliant Color-Touchscreen (7“)
Viewing angle 120...140°
Brightness 400 cd / m²
Unit °C / °F / K
Display range -50...165 °C -58...329 °F
Resolution 0.1 / 0.01 / 0.001 °C / °F / K
General data
Interface Ethernet
3 x USB
Dimensions
 Width 210 mm 8.27 in
 Height 380 + 50 mm 14.96 + 1.97 in Scope of delivery
 Depth 300 mm 11.81 in
•• Test certificate
Weight Approx. 10 kg Approx. 22 lbs
•• Mains cable
Power supply 100...240 VAC, 50 / 60 Hz
•• Insert exchange tool
Power consumption Approx. 375 W
•• PC- and network cable
* At an ambient temperature of 20 °C / 68 °F
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• External reference sensor TF 255-3-300
•• PC software
•• Network-switch, barcode-reader, WLAN-router
•• DAkkS certificate
•• Works certificate

41
TP 37 166 E

Technical data
Type TP 37 166 E
Type TP 37 166 E
Control sensor Switchable internal / external
Dry block
Temperature range* -35...165 °C -31...329 °F
Tolerance ±0.2 °C ±0.36 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.33... 5.91 in
Infrared
Temperature range* -35...165 °C -31...329 °F
Tolerance ±0.5 °C ±0.9 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110 mm 4.33 in
Surface
Temperature range* -25...150 °C -13...302 °F
Tolerance ±1 °C ±1.8 °F
Stability ±0.2 °C ±0.36 °F
Measurement zone Surface
Block dimensions
 Diameter Ø 60 mm Ø 2.36 in
 Depth 150 mm 5.91 in
Display unit
Display Brilliant Color-Touchscreen (7“)
Scope of delivery Viewing angle 120...140°
Brightness 400 cd / m²
•• Test certificate
Unit °C / °F / K
•• Mains cable
Display range -50...165 °C -58...329 °F
•• Insert exchange tool
Resolution 0.1 / 0.01 / 0.001 °C / °F / K
•• PC- and network cable
General data
•• Operating manual Interface Ethernet
3 x USB
Accessories
Dimensions
•• Transport case  Width 210 mm 8.27 in
•• Adapter sleeve  Height 380 + 50 mm 14.96 + 1.97 in
•• Infrared calibration sleeve  Depth 300 mm 11.81 in
Weight Approx. 13 kg Approx. 28.66 lbs
•• Surface calibration sleeve
Power supply 100...240 VAC, 50 / 60 Hz
•• External reference sensor TF 255-3-300
Power consumption Approx. 375 W
•• PC software
* At an ambient temperature of 20 °C / 68 °F
•• Network-switch, barcode-reader, WLAN-router
•• DAkkS certificate
•• Works certificate

The Typ TP 37 166 E temperature calibrator model


has a very large calibration volume.

42
TP 3M 165 E2

Technical data
Type TP 3M 165 E2
Type TP 3M 165 E2
Control sensor Switchable internal / external
Micro Bath
Temperature range* -35...165 °C -31...329 °F
Accuracy ±0.1 °C ±0.18 °F
Stability ±0.010 °C ±0,018 °F
Measurement zone 110...150 mm 4.33...5.91 in
Dry block
Temperature range* -35...165 °C -31...329 °F
Accuracy ±0.3 °C ±0.54 °F
Stability ±0.010 °C ±0,018 °F
Measurement zone 123...163 mm 4.84...6.42 in
Infrared
Temperature range* -35...165 °C -31...329 °F
Accuracy ±0.5 °C ±0.9 °F
Stability ±0.010 °C ±0,018 °F
Measurement zone 110 mm 4.33 in
Surface
Temperature range* -25...150 °C -13...302 °F
Accuracy ±1 °C ±1.8 °F
Stability ±0.1 °C ±0.18 °F
Measurement zone Surface
Scope of delivery Block dimensions
•• Test certificate  Diameter Ø 60 mm Ø 2.36 in
•• Mains cable  Depth 170 mm 6.69 in

•• Insert exchange tool Display unit

•• PC- and network cable Display Brilliant Color-Touchscreen (7“)


Viewing angle 120...140°
•• Operating manual
Brightness 400 cd / m²
•• Sensor basket, suction pump, transport cover Unit °C / °F / K
•• Magnetic stirrer with magnet lifter Display range -50...165 °C -58...329 °F
•• Sensor lid with 5 silicone plugs Resolution 0.1 / 0.01 / 0.001 °C / °F / K
General data
Accessories
Interface Ethernet
•• calibration liquid
3 x USB
•• Transport case
Dimensions
•• Adapter sleeve  Width 210 mm 8.27 in
•• Tube insert  Height 380 + 50 mm 14.96 + 1.97 in
•• Infrared calibration sleeve  Depth 300 mm 11.81 in

•• Surface calibration sleeve Weight Approx. 13 kg Approx. 28.66 lbs

•• External reference sensor TF 255-3-300 Power supply 100...240 VAC, 50 / 60 Hz


Power consumption Approx. 375 W
•• PC software
•• Network-switch, barcode-reader, WLAN-router * At an ambient temperature of 20 °C / 68 °F

•• DAkkS certificate
•• Works certificate

43
TP 3M 255 E

Technical data
Type TP 3M 255 E
Type TP 3M 255 E
Control sensor Switchable internal / external
Micro Bath
Temperature range* RT...255 °C RT...491 °F
Tolerance ±0.2 °C ±0.18 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.33...5.91 in
Dry block
Temperature range* RT...255 °C RT...491 °F
Tolerance ±0.3 °C ±0.54 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 123...163 mm 4.84...6.42 in
Infrared
Temperature range* RT...255 °C RT...491 °F
Tolerance ±0.5 °C ±0.9 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110 mm 4.33 in
Surface
Temperature range* RT...200 °C RT...392 °F
Tolerance ±1 °C ±1.8 °F
Stability ±0.2 °C ±0.36 °F
Measurement zone Surface
Scope of delivery Block dimensions
•• Test certificate  Diameter Ø 60 mm Ø 2.36 in
•• Mains cable  Depth 170 mm 6.69 in
•• Insert exchange tool Display unit
•• PC- and network cable Display Brilliant Color-Touchscreen (7“)
Viewing angle 120...140°
•• Operating manual
Brightness 400 cd / m²
•• Sensor basket, suction pump, transport cover Unit °C / °F / K
•• Magnetic stirrer with magnet lifter Display range 0...255 °C 32...491 °F
•• Sensor lid with 5 silicone plugs Resolution 0.1 / 0.01 / 0.001 °C / °F / K
General data
Accessories
Interface Ethernet
•• calibration liquid 3 x USB
•• Transport case Dimensions
•• Adapter sleeve  Width 210 mm 8.27 in
•• Tube insert  Height 380 + 50 mm 14.96 + 1.97 in
 Depth 300 mm 11.81 in
•• Infrared calibration sleeve
Weight Approx. 8.5 kg 18.74 lbs
•• Surface calibration sleeve
Power supply 100...240 VAC, 50 / 60 Hz
•• External reference sensor TF 255-3-300
Power consumption Approx. 1000 W
•• PC software
* At an ambient temperature of 20 °C / 68 °F
•• Network-switch, barcode-reader, WLAN-router
•• DAkkS certificate
•• Works certificate

44
TP 37 700 E

Technical data
Type TP 37 700 E
Type TP 37 700 E
Control sensor Switchable internal / external
Dry block
Temperature range* RT...700 °C RT...1292 °F
Tolerance ±0.4 °C ±0.72 °F
Stability ±0.05 °C ±0.09 °F
Measurement zone 110...150 mm 4.33... 5.91 in
Block dimensions
 Diameter Ø 28 mm Ø 1.10 in
 Depth 150 mm 5.91 in
Display unit
Display Brilliant Color-Touchscreen (7“)
Viewing angle 120...140°
Brightness 400 cd / m²
Unit °C / °F / K
Display range 0...700 °C 32...1292 °F
Resolution 0.1 / 0.01 / 0.001 °C / °F / K
General data
Interface Ethernet
3 x USB
Dimensions
 Width 210 mm 8.27 in
Scope of delivery  Height 380 + 50 mm 14.96 + 1.97 in
 Depth 300 mm 11.81 in
•• Test certificate
Weight Approx. 8.5 kg Approx. 18.74 lbs
•• Mains cable
Power supply 100...240 VAC, 50 / 60 Hz
•• Insert exchange tool
Power consumption Approx. 1000 W
•• PC- and network cable
* At an ambient temperature of 20 °C / 68 °F
•• Operating manual

Accessories
•• Transport case
•• Adapter sleeve
•• External reference sensor TF 650-3-300
•• PC software
•• Network-switch, barcode-reader, WLAN-router
•• DAkkS certificate
•• Works certificate

45
Overview temperature calibrators

TP Basic* TP Solid*

-55…200 °C TP 17 200 TP 17 200 S


-35…165 °C*** TP 17165 M TP 17 165 TP 17 166 TP 17 165 S
-10...100 °C TP Zero
RT…200 °C TP 18 200 E
RT…255 °C
RT…450 °C TP 17 450
RT…650 °C TP 17 650 M TP 17650 TP 17 650 S
RT...700 °C***
RT…850 °C TP 18 850 E
400…1300 °C TP 28 1300 E
Metrology*
Best Resolution 1 °C/1 °F 1 °C/1 °F 0.1 °C/0.1 °F 0.1 °C/0.1 °F 0.01 °C/0.01 °F 1 °C/1 °F
Stability 0.1 °C/0.18 °F 0.1 °C/0.18 °F 0.1 °C/0.18 °F 0.1 °C/0.18 °F 0.05 °C/0.09 °F 0.5 °C/0.9 °F
Accuracy 1 °C/1.8 °F 1 °C/1.8 °F <0.8 °C/1.44 °F <0.6 °C/1.08 °F <0.4 °C/0.7 °F 2 °C/3.6 °F
Testholder
Block Ø 18/28 mm 28 mm 28 mm 60 mm 28 mm 28 mm
Dry block ü ü ü ü ü ü
Infrared
Surface
Microbath
Equipment
Periodic counter ü ü ü ü ü ü
Interface ü ü
External reference connection
Internal measuring instrument
Data logger function
Controller OFF ü ü ü ü
Manual control ü ü ü ü
Set-value memory ü ü
Temperature levels ü ü
Gradient control ü ü
Accessories
Transport case ü ü ü ü ü ü
Service transport bag** ü ü ü ü
Software and data cable ü ü
calibration reference sensor

* The specifications listed are for guidance.


Detailled information can be found on the individual product pages.
** For calibrators with slim body only
*** Standard linearisation may differ from the temperature range

46
TP Premium*

TP 37 200 E
TP 17 166 S TP M 165 S TP 38 165 TP 37165 E TP 37 166 E TP 3M 165 E

TP M 255 S TP 3M 255 E
TP 17 450 S TP 17 450 S-U
TP 38 650
TP 37 700 E

0.01 °C/0.01 °F 0,01 °C/0.1 °F 0.01 °C/0.01 °F 0.01 °C/0.01 °F 0.001 °C/0.001 °F 0.001 °C/0.001 °F 0.001 °C/0.001 °F
0.05 °C/0.09 °F 0,05 °C/0.09 °F 0.05 °C/0.09 °F 0.03 °C/0.54 °F 0.05 °C/0.09 °F 0.05 °C/0.09 °F 0.05 °C/0.09 °F
<0.3 °C/0.54 °F <0,3 °C/0.54 °F <0.2 °C/0.36 °F <0.2 °C/0.36 °F <0.4 °C/0.7 °F <0.3 °C/0.54 °F <0.2 °C/0.36 °F

60 mm 60 mm 60 mm 28 mm 28 mm 60 mm 60 mm
ü ü ü ü ü ü
ü ü ü
ü ü ü
ü ü

ü ü ü ü ü ü ü
ü ü ü ü ü ü ü
ü ü ü ü ü
ü
ü ü ü ü
ü ü ü ü ü ü ü
ü ü ü
ü ü ü ü ü ü ü
ü ü ü ü ü ü ü
ü ü ü ü ü ü ü

ü ü ü ü ü ü ü
ü ü ü
ü ü ü ü ü ü ü
ü ü ü ü

TP Premium tested according to EN 61326-1, class A (industry sector)

47
Accessories
Versatile accessory range
Customised adaptations are not just limited to our calibrators. Special applications often require special accessories. For instance,
we are able to supply adapter sleeves for temperature calibrators with various drill holes for mounting several temperature sensors
or in special designs, for example for mounting sensors used in the foodstuffs industry or railway technology. Reference sensors
and adapter kits in various versions are also available.

Through close cooperation between the customer, sales and design we are able to implement solutions for nearly any application.
Installation support and customer specific system-integrations can also be implemented on request.

Transport trolley
A service transport trolley is available in robust aluminium Transport trolley - safely stowed!
construction to ensure safe transport to on-site locations. This
provides not only the calibrator itself a well-padded space but
also other accessories, such as adapter sleeves, tub inserts,
calibration liquids, calibration reference sensors and others.
For small calibrator types there is also the a service transport
bag in robust padded nylon.

48
Network Switch & Router
High-grade technology is needed to let data in networks rush The SIKA LAN switch creates a distributor with a total of five
quickly from one user to the next. The SIKA temperature connection ports. The transfer rate of 1,000 Mbit/s facilitates
calibrators of the TP Premium series have an Ethernet interface efficient and stable communication with all subscribers.
with RJ45 jacks that guarantee high data transmission rates To connect one or more calibrators to various pieces of terminal
and secure trouble-free, reliable work. equipment such as a PC, notebook, tablet or smartphone,
With the supplied network cable, it’s very simple to connect the a WiFi router is the best solution. The convenient connection
calibrator to, e.g., the network adapter of a PC or, through a through WiFi enables mobile access to all required calibrator
network outlet, with a cable network. data. The user can very easily monitor the current status of the
The network switch implements connecting several devices calibration process at any time.
with each other if necessary.

Calibration and testing software


The in-house calibration software application is used for •• Programmable ramp functions
temperature calibrators that are equipped with an external •• Programmable temperature cycles
interface for programming and evaluating the calibration
•• Series tests (e.g. for incoming goods inspection)
values. It can be operated easily from an external PC. The
•• Preparing the test data in graphical and in tabular form
following calibration tasks can be performed:
•• Incorporating customer data in the certificates
•• Programmable temperature gradients

49
Calibration liquids
Using a liquid calibration medium is advantageous for checking
temperature sensors with unusual shapes or dimensions. The
test item is immersed in the liquid without an insulating air gap,
resulting in direct contact between the calibrator and the test
item. The calibration liquid is chosen according to the desired
calibration temperature. The sensor lid with 5 silicone plugs
and / or a support base ensures the stable positioning of the
test items in the calibration bath. The lid reduces heat emission
over the surface of the liquid, thereby ensuring optimum
measurement results.

Tub insert
Our tub insert is the ideal solution for applications in which a
variety of liquids are used. It eliminates the time-consuming
task of exchanging the liquids and cleaning the bath. The
separate tub insert is just as leak-proof as the bath itself.

Demineralised water Silicone oil 10 CS Silicone oil 20 CS Silicone oil 50 CS


Limits 2...95 °C 35.6...203 °F -35...155 °C -31...311 °F 7...220 °C 44.6...428 °F 50...270 °C 122...518 °F
Flash point 165 °C 329 °F 230 °C 446 °F 280 °C 536 °F

Dry block - Adapter sleeves


Infrared & Surface - Adapter sleeves Dry block calibrators are designed to simplify temperature
calibration in the lab and in the field. With the help of adapter
sleeves, straight temperature sensors with almost any length
and diameter can be calibrated. The dry block covers the entire
temperature range of the calibrator with no need for changing
the calibration medium. Viscosity, flash point and outgassing
are of no concern. Every adapter sleeve can be equiped with
a single or several multi bores. Bores with diameters ranging
from 1.5 to 25.5 mm can be realised in 0.5 mm steps. Ideally, the
internal diameter of the sleeve is 0.5 mm larger than the outer
diameter of the test item.

50
Adapter Sleeves - standard configurations

Our adapter sleeves are designed for use with SIKA dry block calibrators. The sleeves are configured with various diameter bores
to accomodate industry standard temperature sensors. We provide several standard configurations for quick delivery.

Adapter sleeves Dimensions Ø 60 mm (Aluminium) Ø 28 mm (Brass alloy) Ø 18 mm (Brass alloy)


Standard Bores TP 3M Series TP 37 Series TP 18 Series
TP M Series TP 17 Series
n28mm (Messing) n2n82m8mmm(M(eMsessinsgin) g) n28mm (Mess
Weitergabe sowie Vervielfältigung dieser Dokumente, Verwertung und Mitteilung ihres Inhalts sind verboten,

n28mm (Messing) n28mm (Messing) n28mm (Messing)


tenntz-u, SGcehbardaeuncehrssmautszt. er- oder Geschmacksmustereintragung vorbehalten.

1x 3.5 mm (⅛ in)
1x 6.5 mm (¼ in)
ü ü
Snc,hadenersatz.

n281x
mm 3(,M esmsing) n28mm (Messing) n28mm (Messing) n28mm (Messing) n28mm (Mes
ute

1x 3,5mm 5m
1x 3,5mm 1x 3,5mm 1x 3,5mm 1x 3,5mm 1x 3,5mm
zo
indt.ragung vorbehalten.
nrb

1x 3.5 mm (⅛ 1in)
ee

2x 5,0mm 1x 1x66,5,5mmmm x 5,0mm 2x 5,0mm


htv

6x 6,5mm 1x 6,5mm
hlruensgeInnhavletrspfsliincd

2x 6,5mm 1x 13,5mm 1x 6.5 mm (¼ 1in)


x 6,5mm 2x 6,5mm
2x n
9,258mmmm (Messing) n28mm (Messing) 1x 9,5nm2m8mm (Messing)
1x 13.5 mm (½ in) ü n28mm (Messing)
ü 2x 9,5mm
en
lurnhganid
n,ckDsemuutsstcehrla

n2(A
n60mm 8mlummi(nMiuems)sing) n
n6600mmmm
n(2(AA
8lmluummini(nM
iuiuemms))sing) nl2u8mminmium
n60mm (A (M) essing) n60mm n
(A2l8umm
iniu(Mme) ssing) n60mm (Aluminium) n60mm (Alum
Muitwtiedie

n28mm (Messing) n218xmm3,(5Mmem


ssing) n28mm (Messing) n28mm (Messing) n28mm (Messing)
1x 3,5mm 1x 3,5mm
utnagttuentd. Z

1x 3,5mm
ea

1x 3,5mm
r aGuefsucnhgm

1x 6,5mm 1x 3.5 mm (⅛ in) 2x 5,0mm


praütceknltic. h gestatteAtl.leZuRweicdhetrehafnüdrludnegnenFavlel rdpefrlicPhate

1x 5,0mm 6x 6,5mm 1x 6,5mm


h egrets

1x 13,5mm
, GUenbtrearuncehhsmmeunsstgerru-ppoed,eK

1x 5.0 mm (³⁄₁₆ 1in)


x 6,5mm 2x 6,5mm
lircw

1x 9,5mm
ü ü 2x 9,5mm
ke
rü, cV

1x 3,5mm 1x 3,5mm 1x 6.5 mm (¼ in) 1x 3,5mm 1x 3,5mm


auesndte

SIKA Unternehmensgruppe, Kaufungen, Deutschland.

1x 5,0mm n60mm (A6lxum6i,n5imm)


um 2x 5,0mm n60mm (Alu1x
min6i,u5m)m
ieesietr nDicohktum

11xx 63,,55m mmm 1x 3,5mm 1x 9.5 mm (⅜


n60in)mm (Alu1mxin3iu,5mm)m n60mm (Aluminium) n60mm (Alu
2x 6,5mm 1x 3,5mm
n12x8m3m,5m 11exxs95s,,i5n0gm
(Mm m)m
m n28mm (M6exss6in,5gm ) m n28mm (M2 2exxss95in,,50gm m) m
m n28mm (Me1xss6in,5gm ) m
nItK-A

1x 3,5mm 1x 3,5mm 1x 3,5mm


ow

1x 3,5mm
Weitergabe sowie Vervielfältigungsd

2x 6,5mm
Alle Rechte für den Fall der PateS

1x 6,5mm 1x 6,5mm 1x 3.5 mm (⅛6in) 2x 5,0mm


1x 5,0mm x 6,5mm 1 x 6,5mm
1x 13n,56m0m1mxm9(,A 5mlumminium) n60mm (Aluminium) 2 x 9 , 5 m m
1x 6,5mm 6x 6.5 mm (¼ in) n60mm (Aluminium) 2x 6n,56m0m mm (Aluminium)
n60mm (Aluminium)
1x 9n,56m0m mm (Aluminium) n60mm (Aluminium)
2x 9,5mm
n60mm (Aluminium) ü ü
is strictly forbidden
ricdhetsiagunsd

1x 3,5mm 1x 3,5mm 1 x 3 , 5 m m 1x 3,5mm 1x 3,5mm 1x 3,5mm


n60mm (Aluminium) n 6 01mx m 3 ( ,A5 m
lu m
m i n iu m ) n60mm (Aluminium) n 6 n60mm (Aluminium)
2x 5,0mm 1x 6,5mm
1x 6,5mm 1x 5,0mm 6x 6,5(m
0 m m Am luminium) 2x 5,0mm 1x 6,5mm
deeliton

n28mm12x(xM3e6,s,55smm inmg) n281x m1xm133(,M,55memsmmsing) n28m1mx1x(6M,35 e,ms5smminmg) 2x 6,5mm


sow

2x 9,5mmm 1x 3,5mm
m

1x 3.5 mm (⅛ in) 1x2x9,55m ,0mmm 2x 9,5mm


lanritlyy. in the event of patent grant, registration of a utility
Any unauthorized copying, disclosure or distribution of this document and its contents

1x 5,0mm 6x 6,5mm n28mm (Mess1ixng6) ,5mm


1x 6,5mm 2x 5.0 mm (³⁄₁₆ in) 2x 6,5mm
sterramtaionny of a utility model or design patent.

1x 9,5mm
2x 6.5 mm n60(¼ mm in)
2x 9,5mm
ü ü
lyityf.orbidden

(Aluminium)
1 x 3 , 5 m m 1x 3,5mm 1x 3,5mm 1x 3,5mm 1x 3,5m
n60mm (Aluminium) 2x 9.5 mm (⅜n in) n606m
n60mm (Aluminium)
1x 6,5mm
610xm5m,0(mAmluminium) xm6,(5Amlumminium) 2x 5,0mm 1x 6,5m
intdreicmtn

n281m 1xxm31,(35M,m5em ssming) 1x 6,5mm 2x 6,5mm


os

1x 3,5mm m 1x 13x,59m,5mmm
coilnltleenatds tis

2x 9,5mm
6x 6,5mm 1x 3,5mm 2x 5,0mm1x 3,5mm Blank sleeve 1x 6,5mm 1x 3,5mm n218xm1m0,5 (Mmemssing) 1 x 3 , 5 m m
(Zul. Abw.) 1x 5(O,0 bemrfmläche) Maßstab: 2x 6,5mm6x(Ge6w,ic5hmt)m 2x 5,0mm 1x(V6e,r5wm enm
s w

dungsbereich) (Zul. Abw.) (Oberfläche) Maßstab:


1xx 36,,55mmmm
mneyntviaonladtiiotn

m (Werkstoff, Halbzeug) 2x 9,5mm1x 3,5mm 1x 3,5mm


11xx 63,,55m mm 2
ü ü ü
(Werkstoff, Halbzeug)
en ISO 2768-m DIN ISO 1302 n60mm (Alumini2 uxm)5,0mm n 2 8 m m ( M e s s i n g ) Kmamtalogzeichnungen
11xx 95,,50m mm m (SIKA-Artikelnummer) 6x 6,5mm 2x 9,5mm 1 x 3 ,5 m m1 x 6 , 5 ISO 2768-m DIN ISO 1302
1 x 3 ,5 m m 1 x 3 ,5 m m 1 x 3 , 5 m m (SIKA-Artikelnummer)
, iG

n60m1xm (3A,l5uDmm atm 1iunxmiu6m,5) m


Nam nn2680mmmm ((MAels siinngiu) m) 2inxiu6m,5) mm
ng

me
ftthhoirsisdeodc. uA

um n60m6mx (6A,5 lummm


re

2x 5,0mm 1x 6D,5atmumm Name


utn,ge

1Bxearb6. 2,56.m 05m .2015 Seifert


Übergangns2h8üm11xxlms65e(,,M50nmmesmmØsi2ng8)mm und Übergangshülsen Ø
iceum

Gepr. 27.051.2x0159,5Smeim
2x 9,5mm Bearb. 26.05.2015 Seifert
aruafn

2x 6,5mm
ad, tpoarintd

fert Gepr. 27.05.2015 Seifert


1Nxorm13,5mm n 2 8 m m ( M e s s in g)
yf Gpraotuepn, tKg

n60mm (Aluminium) Øo6h0nme m 1x 9in,5mMms/Al


Bohrungen No r m Ø60mm in M
ou

2 x 9 , 5 m m
n a

1x 10.5 mm (⅜ in)
srs dsipsetcrifbiuctailoly
ve

1x 10,5mm
eisllerle

Dr . Si e b e r t & Kü h n Blatt n28mm (Messing)


_ Dr . Si e b e r t & Kü h n
_
thAe Ceovmepnatno

n
Gmb6H0&m ComKG(Aluminium)
1
titosn rw

(Verwendungsbereich) Gmb(Z H u&l. CAo bKw G .) ( O b e r f lä c h e ) M a ß s t ab :


St r u t h we g 7 - 9 ohne Bohrungen
ü
St r u t h we g 7 - 9 (Werkstoff, Halbzeug)
SIKA Company Group, Kaufungen, Germany

1 Bl . n218xm1m3,5 ( Mmemssing)
ah

34260 Kaufungen 34260 Kaufungen


unless specifically authorised. AnyAlvl iorlig

(Ers. f.:) n60mm (Alum(iEnrsiu .m


d.:)) K a t a l o g z e i c hn u n g e n
Any unauthorized copying, disclosuurneleo

m Name (Urspr.:) Zu s t Än d e r u n g Datum Name (Urspr.:) I S O 2 7 6 8 - m D IN I S O 1 30 2 (Ers. f.:) (Ers.


All rights reserved, particularly inSIK

1x 3,5mm 1x 3,5mm 1x 3,5mm 1x 3,5mm (SIKA-Artikelnummer)


1x 10,5mm Datum N a me
1x 5,0mm 6 x 6 , 5 m m 2 x 5 , 0 m m 1 x 6 , 5 m m
n28mm1x(M 6,e5smsming)
(Verwendungsbereich)
2x 6,5mm
(Zul. Abw.) (Oberfläche) Maßstab: (Gewicht) Bearb. 26.05.2015 Seifert
Gepr. 27.05.2015 Seifert Übergangshülsen
no2h8n1mexmB10o(M,h5ermusmnsginegn)
(Werkstoff, Halbzeug)

1x 9,5mm K a t
1x 13.5 mm
a l o
(Verwendungsbereich)
g z e i c hnu

n g e n
in)2x 9,(5Zmulm I S O 2
. Abw.)
7 6 8 - m D IN I S O 1 30 2
(Oberfläche) (SMIKaAß-sAtratibke: lnummer) (Gewicht)
No r m Ø60mm in M
n28mnm 2(8Mmemss(inMge)ssing) 1x 10,5mm Datum N a me
(Werkstoff, Halbzeug) Dr . Si e b e r t & Kü h n
KatalogzeicohhnnuengBeonhrunBgeeaISnrOb(. V2267e.6r0w85-e.2mn0d1u5nDgIsNbSeIeSirfOeeirc1t3h0)2 (Zul. Abw.) (Oberfläche) Maßstab: GmbH & Co KG (Gewicht) _
5.2015 Seifert (SI1KÜ xA-bA1re
3tik,re5lng
muma
mmenr)gshülsen Ø28(m Werm kstofu f, n
Had
St r u t h we g 7 - 9

ü
lbzeug)
n28mm (Messing) Gepr. 27D.0aK a t al o gz e i c h nu n g e n
34260 Kaufungen
n60mm (Aluminium) No r m t u m N a m e
Ø60mm in Ms/A(SlIKA-Artikelnummer)
I S O 2 7 6 8 - m Z D
u I
sN t I S O 1Ä 30
n 2
d e r u n g D a tu m N a m e ( U r s p r .: ) (Ers. f.:) (Ers
1x 3,5mm 1x 3,5mm 1x 3,5mGBmeeaprrb. . 2267..0055..22001155 SSeeiiffeerrtt ÜbergangsDhatüumlsenNaØ me
28mm uBnladtt
BeØ arb6. 20 6.m05.m2015in SeM ifes rt/Al
Dr . Si e b e r t & Kü h n
6x 6,5mm 2x 5,0mm 1x 6,5mGNm _
ohne Bohrungen
1x2x136,,55mmmm
o r m
mbH & Co KG
St r u t h we g 7 - 9
Gepr. 27.05.2015 Seifert Über1g1aBnl. gshülsen Ø28mm und
BlattØ60mm in Ms/Al
34260 Kaufungen
ohne Bohrungen Zu s t Ä n d e r u n g D a tu m N a m e
Dr . Si e b e r t & Kü h n
(U r s p r .: ) ( E rs . f .: ) No r m _ ( E rs . d .:)
ohne Bohrungen 2x 9,5mm G m b H & C o K G
1
m (Messing) 1x 10,15xmm13,5mm St r u t h we g 7 - 9
34260 Kaufungen Dr . Si e b e r t & Kü h n 1 Bl . Blatt
(VerwenZduusntgsberÄenicdhe)rung D a tu m(ZuNla. mAebw(U.)rspr.:)(Oberfläche) Maßstab(:Ers. f.:) GmbH & Co KG (Ge(wEircsh.t)d.:) _ 1
ohne Bohrungen St r u t h we g 7 - 9
n28mm (Messing) 1x 13,5mm (Werkstoff, Halbzeug) 34260 Kaufungen 1 Bl .
Katalogzeichnungen ISO 2768-mZustDIN ISÄOn1d3e0r2ung Datum Name (Urspr.:)
For other configurations - consult factory (Ers. f.:) (Ers. d.:)
(SIKA-Artikelnummer)
Datum N a me
Bearb. 26.05.2015
Gepr. 27.05.2015
Seifert
Seifert Übergangshülsen Ø28mm und
No r m Ø60mm in Ms/Al
10,5mm Dr . Si e b e r t & Kü h n Blatt 51
GmbH & Co K(GGewicht) _ 1
ohne Bohrunge(nVerwendungsbereich) (Zul. Abw.) (Oberfläche) Maßstab: St r u t h we g 7 - 9
m (Messing) 1x 13,5mm (Werkstoff, Halbzeug) 34260 Kaufungen 1 Bl .
Katalogzeichnungen ISO 2768-m DIZNuIsStO 130Ä
2 nder ung Datum Name (Urspr.:) (Ers. f.:) (Ers. d.:)
For your test equipment monitoring
TT-Scan

Resistance thermometers, thermocouples, temperature transmitters and switches must be calibrated using an instrument that
measures the output signal and displays it as a temperature.

Checking groups of temperature sensors can be automated


by extending your SIKA calibrator with a TT-Scan unit and
calibration software. Up to eight test items can be checked at
the same time with this arrangement. The configuration of the
test item type is free programmable. A reference sensor can be
connected. The TT-Scan unit has a USB port for connection to a
PC. SIKA calibration software analyses the measurement data
and presents the results in graphic or tabular form. At the same
time it automatically generates up to 8 certificates, which may
also include customer data.

52
Type TT-Scan

Properties
Possibilities to connect
Switch
Wire Wire Wire

2 2 2 2 2 2

3 3 3 3 3 3

4 4

Version Scanner device with precision measuring instrument


Measuring inputs Switchable
For up to 8 sensors
Sensor type free confi gurable
General data
Power supply 230 VAC ±10 %, 50 / 60 Hz via adapter
Power consumption Approx. 10 W
Dimensions (D x W x H) 200 x 140 + 40 x 380 mm 7.87 x 5.51 + 1.57 x 14.96 in
Weight Approx. 2.5 kg Approx. 5,51 lbs
Equipment features
32 x 4 mm/1.26 x 0.16 in connections free of thermal voltage
Connection for external calibration reference sensor
External cold junction available
Serial USB data interface, incl. USB data cable
Options
Aluminium transport case, test & calibration software, DAkkS certificate,
SIKA works certificate, external calibration reference sensors

53
Measuring inputs

Version Measuring range Tolerance


Resistance thermometer EN 60751
Pt100 2-, 3-, 4-wire -90.00 °C...850.00 °C -130.00 °F...1562.0 °F ±0.005 % full scale ±0.01 °C ±0.005 % full scale ±0.02 °F
Pt500
Pt1000
Connection possibility through 4 mm connections free of thermal voltage
Thermocouples according to DIN EN 60584 / DIN 43710
Type K NiCr-NiAl -90.00...999.99 °C -130.00...1831.9 °F ±0.007 % full scale ±0.01 °C ±0.007 % full scale ±0.02 °F
1000.0...1370.0 °C 1832.0...2498.0 °F ±0.005 % full scale ±0.1 °C ±0.005 % full scale ±0.18 °F
Type J FeCu-Ni -90.00...900.00 °C -130.00...1652.0 °F ±0.005 % full scale ±0.01 °C ±0.005 % full scale ±0.02 °F
Type N NiCrSi - NiSiMg -90.00...999.99 °C -130.00...1831.98 °F ±0.007 % full scale ±0.01 °C ±0.007 % full scale ±0.02 °F
1000.0...1370.0 °C 1832.0...2498.0 °F ±0.005 % full scale ±0.1 °C ±0.005 % full scale ±0.18 °F
Type E NiCr-CuNi -90,00...700.00 °C -130.00...1292.0 °F ±0.005 % full scale ±0.01 °C ±0.005 % full scale ±0.02 °F
Type R Pt13Rh – Pt 0.00...999.99 °C 32.00...1831.9 °F ±0.05 % full scale ±0.01 °C ±0.05 % full scale ±0.02 °F
1000.0...1760.0 °C 1832.0...3200.0 °F ±0.03 % full scale ±0.1 °C ±0.03 % full scale ±0.18 °F
Type T Cu-CuNi -90.00...400.00 °C -90.00...400.00 °F ±0.01 % full scale ±0.01 °C ±0.01 % full scale ±0.02 °F
Type B Pt30Rh-Pt6Rh 0.00...999.99 °C 32.00...1831.98 °F ±0.05 % full scale ±0.01 °C ±0.05 % full scale ±0.02 °F
1000.0...1820.0 °C 1832.0...3308.0 °F ±0.03 % full scale ±0.1 °C ±0.03 % full scale ±0.18 °F
Type S Pt10Rh-Pt 0.00...999.99 °C 32.00...1831.98 °F ±0.05 % full scale ±0.01 °C ±0.05 % full scale ±0.02 °F
1000.0...1760.0 °C 1832.0...3200.0 °F ±0.03 % full scale ±0.1 °C ±0.03 % full scale ±0.18 °F
Type L Fe-CuNi -90.00...900.00 °C -130.00...1652.0 °F ±0.005 % full scale ±0.01 °C ±0.005 % full scale ±0.02 °F
Type U Cu-CuNi 90.00...600.00 °C 194.00...1112.0 °F ±0.01 % full scale ±0.01 °C ±0.01 % full scale ±0.02 °F
Automatic comparison point compensation between 0 °C/32 °F and 60 °C/140 °F
Accuracy of the comparison point Pt100 DIN class A
Possibility of connection through 4 mm/0.16 in connections free of thermal voltage
Standard signal input
Current mA 0(4)...20 mA ±0.015 % full scale ±0.01 mA
(switchable)
Transmitter supply 24 VDC, Imax = 30 mA,
Possibility of connection through 4 mm/0.16 in connections free of thermal voltage
Temperature switch
Automatic detection of an edge change, determining the hysteresis,
Independent detection normally closed / normally open
Potential-free input contacts (Umax = 5 V, Imax = 1 mA)
Possibility of connection through 4 mm/0.16 in connections free of thermal voltage
Calibration reference sensor connection
Pt100 4-wire -90.00...850.00 °C -90.00...850.00 °F ±0.005 % full scale ±0.01 °C ±0.005 % full scale ±0.02 °F
Polynomial correctable through internal parameters or through external EEPROM inside the sensor
Possibility of connection through 7-pin built-in socket

54
Calibration reference sensors

If the sensor to be calibrated is too short to be inserted into The physical structure of the reference sensors requires that
the homogeneous temperature zone of the metal block, an different materials be joined together. The special design of the
external reference sensor can be used without any problems. joint areas prevents the occurrence of parasitic thermoelectric
This results in a small, flexible measurement zone. voltages. Thus the measurement reading is not affected by the
temperature gradients from the measurement point to the handle.
In examining the self-heating characteristics it was seen
Type TF 650-3-300 that measurement currents ≤ 1 mA are ideally suited, since
no distortion of the measurement result occurs. Here the
self-heating effect can be neglected.

Class B ± (0.3 + 0.005 |T|)


Class A ± (0.15 + 0.002 |T|)
Linearisation MH 3750
1.2

An ace of calibration 1

Particular attention is given to the physical construction to 0.8


ensure that shocks have minimal effect on the reference sensor.
Accuracy [°C]

The use of robust measuring elements in thinfilm technology 0.6

ensure standardised and reliable performance.


0.4
Intensive ageing tests are carried out at the maximum operating 0.3

temperature to examine longterm temperature stability. In 0.2 0.15

order to detect longterm effects through thermal stress, a 0.05


0
defined tempering process is carried out with a special selection
-50 0 50 100 150
of reference sensors over 300 hours. In the case of stress caused
Temperature [°C]
by thermocycling, no significant hysteresis effects were found.

Calibration reference sensor - Type TF Calibration reference sensor - Type TFEE


Pt100 without probe specific linearization in the controller for Pt100 with probe specific linearization through EEPROM in the
Series TP...S-U, TP 3... handle for TT-Scan and Series TP 38...

Technical data Technical data


Measuring range Measuring range
TF 255-3-300 -55...255 °C / -67...491 °F / TFEE 255-3-300 -55...255 °C / -67...491 °F /
sensitive area 2 mm sensitive area 0.08 in sensitive area 2 mm sensitive area 0.08 in
TF 650-3-300 -35...650 °C / -31...1202 °F / TFEE 650-3-300 -35...650 °C / -31...1202 °F /
sensitive area 5 mm sensitive area 0.2 in sensitive area 5 mm sensitive area 0.2 in
Tolerance Tolerance
Class B ±(0.3 + 0.005 |T|) ±0.05 °C between ±0.05 °F in the range
Version -35.00...199.99 °C, of -31.00...391.98 °F,
else ±0.1 °C else ±0.18 °F
Material Rust and acid-proof
Stainless steel 1.4571 Version

Robust plastic handle Material Rust and acid-proof


Stainless steel 1.4571
Immersion tube Ø 3 mm, L = 300 mm Ø 0.12 in, L = 11.81 in
Robust plastic handle
Electrical Silicon cable with 4-pin mini DIN-plug
Immersion tube Ø 3 mm, L = 300 mm Ø 0.12 in, L = 11.81 in
connection
Electrical Silicon cable with 7-pin mini DIN-plug
connection

55