SOFTWARE RELEASE (SR)

DPI620 GENII

ADVANCED MODULAR CALIBRATOR

SR0106

Copyright 2013 Druck Ltd, a Baker Hughes business


BAKER HUGHES CONFIDENTIAL. This document and the information contained herein is confidential to and the sole property of Druck
Limited and Baker Hughes. It may not be reproduced, used, disclosed, or made public without the express written consent of Druck
limited and/or Baker Hughes. The use of this document and the information contained therein may be subject to the terms of a separate
agreement with Druck limited and/or Baker Hughes. (Druck and logo are registered trademarks of Baker Hughes in the United States
and other countries. All product and company names are trademarks of their respective holders)

NON-PUBLIC Document SR0106


Page 1 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
Amendment Record (Pre-SAP)

Iss. Workflow
Date C/N No Author Amendments
No. No.

1 10 Jun 2013 N/A H.S. Bhuhi 201970 New Specification


Changes to DK401, DK420, DK421.
2 09 Jul 2013 29973 H.S. Bhuhi 203308
Added DK423.
3 26 Jul 2013 30102 H.S. Bhuhi 204791 Changes to DK420, DK421.
4 29 Oct 2013 30476 H.S. Bhuhi 209482 Changes to DK401, DK420, DK418, DK419
5 18 Dec 2013 30620 R. Imlach 211787 Changes to DK417
6 04 Feb 2013 30746 O. Talbot 213722 Corrected DK420 to read V02.00.08
Changes to DK401, DK417, DK419, DK420, DK421 &
7 13 May 2014 31293 H.S. Bhuhi 218580
DK424. Added DK431.
8 07 Oct 2014 31993 S. Ward 225082 Changes to DK420 (Upgrade to 2.02.06)
9 02 Feb 2015 32531 H.S. Bhuhi 229983 Changes to DK420 (Upgrade to 2.03.00)

Amendment Record

Rev.
Date Author Amendments
Lev.

- 13 Oct 2015 H.S.Bhuhi Changes to DK401, DK420 (Upgrade to 2.04.06) and DK0421.
A 14 Oct 2015 H.S.Bhuhi Changes to DK418, DK419, DK420 and DK0421.
B 18 Aug 2016 A.Morten Launch of DPI620G-IS. Changes to DK0420 to support DPI620G-IS variant. Updated
DK401, DK423. Introduction of DK0448, DK0447, DK0464, DK0463, DK0458, DK0459,
DK0457, DK0462 and DK0461.
Removal of DK0431.
C 15 Sep 2016 A Morten Updated DK0419, DK0420, DK0423, DK0448, DK0461.
D 23 Sep 2016 H.S.Bhuhi Updated DK0419, DK0420, DK0448.
E 18 Nov 2016 A Morten Updated DK0420
F 12 Dec 2016 A Morten Updated DK0420
G 09 Jan 2017 H.S.Bhuhi Updated DK0420
H 28 Mar 2017 A.Morten Updated DK0401, DK420, DK419, DK448
J 27 Jul 2017 A.Morten Updated DK0401, DK0420, DK0447, DK0448
K 6 Sep 2017 S Matthews Updated DK0401, DK0461 + FFBus Libraries
L 6 Dec 2017 A.Morten Updated DK0420
M 1 May 2018 A.Morten Up-issue DK0420
N 24 Jul 2018 A.Morten Up-issue DK0401, DK0420
P 26 Oct 2018 A.Morten Up-issue DK0420
R 5 Nov 2018 A.Morten Up-issue DK0420
T 14 Mar 2019 A.Morten Up-issue of DK0401, DK0420, DK0461
U 14 Jul 2019 A.Morten Up-issue of DK0420
V 21 Feb 2020 S. Smith Up-issue of DK0401, DK0417, DK0424
W 20 Apr 2020 S. Smith Up-issue of DK0418, DK0419, DK0420, DK0447, DK0448
Y 2 Jul 2020 P Sengupta Up issue of DK0420
AA 15 Jul 2020 P Sengupta Up issue of DK0420
AB 7 Sep 2020 C Gosden Up issue of DK0420 and DK0421
AC 23 Sep 2020 P Sengupta Up issue of DK0420
AD 16 Oct 2020 L Surtees Re issue of SR0106 AB - To remove changes made at revision AC
AE 09 Nov 2020 C Gosden Up issue of DK0420 and DK0421
AF 23 Dec 2020 C Gosden Up issue of DK0420

NON-PUBLIC Document SR0106


Page 2 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
Approvals
Checked Engineering Principal Engineer
N/A L. Surtees (Reviewer) S. Parkinson (Technical Reviewer)

Production Quality Product Security Leader


N/A N/A S. Parkinson (Reviewer)

NON-PUBLIC Document SR0106


Page 3 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
Contents
1 INTRODUCTION ..................................................................................................................................... 5
2 REFERENCES ........................................................................................................................................ 5
3 ABBREVIATIONS .................................................................................................................................... 5
4 SOFTWARE RELEASE ........................................................................................................................... 6
4.1 Supporting Documents and Software Items (DK) .................................................................................... 6
4.2 Release Package .................................................................................................................................... 6
4.2.1 Release Package - Data Verification ....................................................................................................... 7
4.3 Customer Release Note .......................................................................................................................... 8
4.3.1 DK0420.................................................................................................................................................... 8
4.4 How to Install the Software (DPI620-G) ................................................................................................... 8
4.4.1 Boot Loader (DK418) & Operating System (DK419) ............................................................................... 8
4.4.2 Analogue Main FPGA (DK413) ................................................................................................................ 9
4.4.3 DK420: Genii Main Application (DK0420) ................................................................................................ 9
4.4.4 DPI620G HART Processor Application (DK0417) ................................................................................. 10
4.4.5 DPI620G HART Processor Bootloader (DK0416) ................................................................................. 10
4.4.6 HART SDC625 DD Library (DK0401) .................................................................................................... 10
4.4.7 DPI620G SDC625 (HART) Application (DK0421) ................................................................................. 12
4.4.8 DPI620G GeniiFF (Fieldbus) Application (DK0423) .............................................................................. 12
4.5 How to Install the Software (DPI620G-IS) ............................................................................................. 13
4.5.1 Boot Loader (DK447) & Operating System (DK448) ............................................................................. 13
4.5.2 Analogue Main FPGA (DK458) .............................................................................................................. 13
4.5.3 DK420: Genii IS Main Application (DK0420) ......................................................................................... 13
4.5.4 DPI620G-IS HART Processor Application (DK0464) ............................................................................. 15
4.5.5 DPI620G-IS HART Processor Bootloader (DK0463) ............................................................................. 15
4.5.6 HART SDC625 DD Library (DK0401) .................................................................................................... 15
5 SOFTWARE CHECKS .......................................................................................................................... 17
5.1 Install Package Virus Scan .................................................................................................................... 17
5.2 Cyber Security & Vulnerabilities ............................................................................................................ 17
5.3 Open Source Software (OSS) Notices................................................................................................... 17

Tables
Table 1 - Reference Documents .............................................................................................................................. 5
Table 2 - Supporting Documents & Software Items (DK) Genii ................................................................................ 6
Table 3 - Supporting Documents & Software Items (DK) Genii IS ........................................................................... 6
Table 4 - Release Package Verification ................................................................................................................... 7
Table 5 - File Virus Scan Checker .......................................................................................................................... 17
Table 6 - Cyber Security Checker .......................................................................................................................... 17

NON-PUBLIC Document SR0106


Page 4 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
1 INTRODUCTION
This document details the Product Software Release (SR) for the DPI620 Genii and DPI620
Genii IS.

The approval of this SR document authorises the Product Software Release to Manufacturing
(RTM) of all named DK software items (+Version) referenced by this SR document

2 REFERENCES
The following documents and are referenced by this SR document:

Ref Number Revision Description


1 DK0420 AE DPI620G Main Application
2 154M4766 - Release Test Report
Table 1 - Reference Documents

3 ABBREVIATIONS
AMC Advanced Modular Calibrator
DD Device Description
DK Software Item
FPGA Field Programmable Gate Array
Genii IS Genii Intrinsically Safe
HART Highway Addressable Remote Transducer
IS Intrinsically Safe
JTAG Joint Test Action Group (automated test mechanism)
OS Operating System
OSS Open Source Software
RTM Release to Manufacturing
SAP System Applications Products (an Enterprise Resource Planning
System)
SDC Smart Device Configurator
SR Software Record (document)
USB Universal Serial Bus
Wi-Fi Wireless

NON-PUBLIC Document SR0106


Page 5 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
4 SOFTWARE RELEASE

4.1 Supporting Documents and Software Items (DK)


The following table lists all the supporting documents and DK(s) which relate to this software
release. The ‘changed’ column is to indicate which items have changed since the last release.

This release comprises the following versions for the DPI620G:


SAP DK
Item Changed Document
Document Revision Software Description
In this Release Type
Name Level Version
DK0420 AE V03.20.00 Yes Software Genii Main Application
DK0419 E V02.06.00 No Software DPI620G Operating System
DK0418 B V02.02.00 No Software DPI620G Bootloader
DK0417 A V01.05.00 No Software DPI620G HART Processor Application
DK0416 - V01.00.00 No Software DPI620G HART Processor Bootloader
DK0421 D V01.06.01 No Software SDC625 (HART) Application
DK0401 H V13.00.00 No Software SDC625 HART DD Library
DK0423 B V01.05.00 No Software GeniiFF (Fieldbus) Application
DK0413 - V15.00.00 No Software DPI620G Analogue MAIN FPGA (IC43)
DK0414 - V03.00.00 No Software DPI620G Analogue ISOL CPLD (IC24)
DK0424* A V01.05.00 No Software DPI620G HART Processor Combined Image
DK0461 C V01.03.01 No Software GeniiProfi (PROFIBUS) Application
*DK0424 is not present as a real instance on the instance and its status cannot be displayed. It is a combined
binary image of DK0416 and DK0417 which is used in production to program the HART microprocessor.
Table 2 - Supporting Documents & Software Items (DK) Genii

This release comprises the following versions for the DPI620G-IS:


SAP DK
Item Changed Document
Document Revision Software Description
In this Release Type
Name Level Version
DK0420 AE V03.20.00 Yes Software Genii Main Application
DK0448 F V01.05.00 No Software DPI620G-IS Operating System
DK0447 C V01.02.00 No Software DPI620G-IS Bootloader
DK0464 - V01.01.01 No Software DPI620G-IS HART Processor Application
DK0463** - V01.00.00 No Software DPI620G-IS HART Processor Bootloader
DK0416**
DK0421 D V01.06.01 No Software SDC625 (HART) Application
DK0401 H V13.00.00 No Software SDC625 HART DD Library
DK0423 B V01.05.00 No Software GeniiFF (Fieldbus) Application
DK0458 A V03.00.00 No Software DPI620G-IS Analogue MAIN FPGA (IC43)
DK0459 - V02.00.00 No Software DPI620G-IS Analogue ISOL CPLD (IC24)
DK0457 A V03.00.00 No Software DPI620G-IS Analogue PSU Board CPLD (IC19)
DK0462* - V01.01.01 No Software DPI620G-IS HART Processor Combined Image
DK0461 C V01.03.01 No Software GeniiProfi (PROFIBUS) Application
*DK0462 is not present as a real instance on the instance and its status cannot be displayed. It is a
combined binary image of DK0463 and DK0464 which is used in production to program the HART
microprocessor. **DK0416 and DK0463 are equivalent. The function of this component is to boot the
HART processor application and either one of these is the acceptable installation on the DPI620G-IS.
Table 3 - Supporting Documents & Software Items (DK) Genii IS
4.2 Release Package
This name of the corresponding formal Product Software Release Package is:
DPI620G_RELEASE_SR0106_REV_AF.zip

NON-PUBLIC Document SR0106


Page 6 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
This package contains a copy of this document, and the folder structures which contains all the
DK Product Software folders and files required for this specific revision of SR.

4.2.1 Release Package - Data Verification


This files and folders in this release package have been verified against the DK Version
numbers and the formal data release repository using the WORD document embedded
"SR0106_AF verify.bat" script.

-- EXECUTIVE STATEMENT (Data Verification) --


Reviewing Engineer C. Gosden
Date of Verify 24/12/2020
Results of the Data Verify Pass
Exception(s)

Table 4 - Release Package Verification

SR0106_AF verify.zip

NON-PUBLIC Document SR0106


Page 7 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
4.3 Customer Release Note

4.3.1 DK0420
• 4Sight2 V1.5.1 Compatibility Upgrades
• RTD-USB Interface Connection/Disconnection Resolution

4.4 How to Install the Software (DPI620-G)


4.4.1 Boot Loader (DK418) & Operating System (DK419)
File name: eboot.nb0 & nk.nb0

4.4.1.1 Install (Freescale i.MX Manufacturing Toolkit)

The Manufacturing Toolkit can be downloaded to a PC, free of charge, from FreeScale’s web
site:

http://www.freescale.com/webapp/sps/site/prod_summary.jsp?code=i.MX534&nodeId=018rH3Z
rDR988D&fpsp=1&tab=Design_Tools_Tab

under the “Programmers (Flash, etc.)” tab.

Before using the Manufacturing Toolkit a profile must be created for the DPI620G in folder
<Install Folder>\Profiles\MX53 WinCE Update\AMC. This should be based on an existing
i.MX53 profile.

When the Manufacturing Toolkit is used the first time, the Freescale i.MX53 USB BulkIO Device
Driver must be installed on the PC from folder <Install Folder>\Drivers.

1. Copy the files to the Genii profile folder:


a. Copy eboot.nb0 to <Install Folder>\Profiles\MX53 WinCE Update\AMC\files
b. Copy nk.nb0 to <Install Folder>\Profiles\MX53 WinCE Update\AMC\files
2. Connect a USB cable from Genii USB client socket to PC USB host socket
3. Ensure the DPI620G is in serial download mode:
a. Remove the power
b. Hold the button (SW1) on the Genii digital board
c. Apply the power
4. Run Manufacturing Toolkit:
a. Double click on <Install Folder>\MfgTool.exe
5. Press the Start button
6. It will take a minute to install the boot loader and operating system
7. When the status indicates Operations Complete, press the Stop button
8. Remove the power

4.4.1.2 Upgrade (DPI620G Application)

1. Copy the files to a USB memory stick:


a. Create a folder called “OS” in the root of the memory stick
b. Copy eboot.nb0 to the OS folder
c. Copy nk.nb0 to the OS folder
2. From the Genii application, take the operating system upgrade option:
a. Touch the ADVANCED button on the dashboard
b. Enter 5487
c. Select UPGRADE
d. Select OPERATING SYSTEM and confirm selection
3. Switch the power off with a long press of the power button
4. Insert the USB memory stick in the Genii USB socket
5. Switch the Genii power on
6. The new boot loader and operating system will be installed. This will take several
minutes. Wait until the Genii application is running.
7. Switch the power off with a long press of the power button

NON-PUBLIC Document SR0106


Page 8 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
8. Switch the power on.
9. The new version of boot loader and operating system will be loaded

4.4.2 Analogue Main FPGA (DK413)


File name: DK413.bin

The FPGA is pre-programmed on the PCA as delivered by the subcontract production supplier.

This FPGA may be upgraded via the Genii application upgrade option.

1. Copy the FPGA program file(s) to a folder named ‘FPGA, located at the root directory of a
USB memory stick
2. Insert the memory stick in the USB slot
3. From the Genii application dashboard, touch the ‘Advanced’ button.
4. Enter PIN number 5487
5. Select Upgrade option from the menu.
6. Select the CH1 FPGA option.
7. Select ‘Yes’ from the ‘Confirm Upgrade’ prompt.
8. Power down and restart the instrument once the ‘Software Upgrade’ message appears.
9. The FPGA will be automatically upgraded.
10. Wait until the upgrade completes and the application restarts. This will happen
automatically and may take several minutes.
11. Remove the memory stick.

4.4.3 DK420: Genii Main Application (DK0420)


Folder name: AMC

4.4.3.1 Genii application first-time installation

1. Ensure no application has been installed.


2. Copy the AMC application folder to the root of a USB memory stick.
3. Insert the memory stick in the USB slot.
4. Switch the power on.
5. The application will be automatically installed.
6. Wait until the upgrade completes and the upgraded application is running. This will
happen automatically and may take several minutes.
7. Remove the memory stick.

4.4.3.2 Genii application upgrade (by using the power button)

1. Copy the AMC application folder to the root of a USB memory stick.
2. Insert the memory stick in the USB slot.
3. Press and hold the power button on.
4. The application will be automatically upgraded.
5. Release the power button once the progress bar appears. This may take several
seconds.
6. Wait until the upgrade completes and the upgraded application is running. This will
happen automatically and may take several minutes.
7. Remove the memory stick

4.4.3.3 Genii application upgrade (from within the Genii application)

1. Copy the AMC application folder to the root of a USB memory stick
2. Insert the memory stick in the USB slot
3. From the Genii application dashboard, touch the ‘Advanced’ button.
4. Enter PIN number 5487
5. Select Upgrade option from the menu.
6. Select APPLICATION option.

NON-PUBLIC Document SR0106


Page 9 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
7. Select ‘Yes’ from the ‘Confirm Upgrade’ prompt.
8. Power down and restart the instrument once the ‘Software Upgrade’ message appears.
9. The application software will be automatically upgraded
10. Wait until the upgrade completes and the upgraded application is running. This will
happen automatically and may take several minutes.
11. Remove the memory stick.

4.4.4 DPI620G HART Processor Application (DK0417)

The HART processor application is pre-programmed on the HART modem microcontroller on


the PCA as delivered by the subcontract production supplier.

This HART modem firmware may be upgraded via the Genii application upgrade option.

NOTE: The HART Processor Application (DK0417) and the HART Processor Bootloader
(DK0416) may be upgraded at the same time, as a single process. If all the required files are
present on the upgrade folder (on the memory stick) following the process for either of the
HART processor firmware components (DK0416 or DK0417) will upgrade both.

1. Copy the HART processor firmware file(s) to a folder named ‘HART’, located at the root
directory of a USB memory stick
2. Insert the memory stick in the USB slot
3. From the Genii application dashboard, touch the ‘Advanced’ button.
4. Enter PIN number 5487
5. Select Upgrade option from the menu.
6. Select the HART DRIVER option.
7. Select ‘Yes’ from the ‘Confirm Upgrade’ prompt.
8. Power down and restart the instrument once the ‘Software Upgrade’ message appears.
9. The HART modem firmware will be automatically upgraded.
10. Wait until the upgrade completes and the application restarts. This will happen
automatically and may take several minutes.
11. Remove the memory stick.

4.4.5 DPI620G HART Processor Bootloader (DK0416)


The HART processor bootloader is pre-programmed on the HART modem microcontroller on
the PCA as delivered by the subcontract production supplier.

The HART processor bootloader may be upgraded via the Genii application upgrade option.

NOTE: The HART Processor Application (DK0417) and the HART Processor Bootloader
(DK0416) may be upgraded at the same time, as a single process. If all the required files are
present on the upgrade folder (on the memory stick) following the process for either of the
HART processor firmware components (DK0416 or DK0417) will upgrade both.

1. Copy the HART processor bootloader firmware file(s) to a folder named ‘HART’, located
at the root directory of a USB memory stick
2. Insert the memory stick in the USB slot
3. From the Genii application dashboard, touch the ‘Advanced’ button.
4. Enter PIN number 5487
5. Select Upgrade option from the menu.
6. Select the HART BOOT LOADER option.
7. Select ‘Yes’ from the ‘Confirm Upgrade’ prompt.
8. Power down and restart the instrument once the ‘Software Upgrade’ message appears.
9. The HART modem firmware will be automatically upgraded.
10. Wait until the upgrade completes and the application restarts. This will happen
automatically and may take several minutes.
11. Remove the memory stick.

4.4.6 HART SDC625 DD Library (DK0401)


• These files should be installed on an SD Card in the DPI620 Genii.

NON-PUBLIC Document SR0106


Page 10 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
• Locate the self-extracting file called “DPI620G & DPI620GIS HART Lib2019
DK401V13.00.00.zip” from the drawing office.

• Install the microSD card in the SD card reader i.e. Drive G, H

• Double Click to run the executable file.

• Select the … box (see picture)

Select this button

Extract to SD card (G, H)

• Once the file is extracted, open the SD card drive. The files structure should appear like:

NON-PUBLIC Document SR0106


Page 11 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
• SDCARD\HCF\DDL\Library

• Remove the microSD memory card from the reader and insert into the card slot in the
DPI620 Genii battery compartment.

4.4.7 DPI620G SDC625 (HART) Application (DK0421)


The HART SDC625 application is installed with the main AMC application.

4.4.8 DPI620G GeniiFF (Fieldbus) Application (DK0423)


The GeniiFF application is installed with the main AMC application.

NON-PUBLIC Document SR0106


Page 12 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
4.5 How to Install the Software (DPI620G-IS)
4.5.1 Boot Loader (DK447) & Operating System (DK448)
File name: eboot.nb0 & nk.nb0

4.5.1.1 Install (Freescale i.MX Manufacturing Toolkit)

Follow the instructions in section 4.4.1.1

4.5.1.2 Upgrade (DPI620G-IS Application)

1. Connect Genii-IS to a PC via the USB port, where it will appear as an external device, eg.
(DPI620 (F:) ). Copy the files to the device:
a. Create a folder called “Install” in the root of the device
b. Create a folder called “OS” under “Install”
c. Copy eboot.nb0 to the OS folder
d. Copy nk.nb0 to the OS folder
2. From the Genii IS application, take the operating system upgrade option:
a. Touch the ADVANCED button on the dashboard
b. Enter 5487
c. Select UPGRADE
d. Select OPERATING SYSTEM and confirm selection
3. Switch the power off with a long press of the power button
4. Switch the Genii power on
5. The new boot loader and operating system will be installed. This will take several
minutes. Wait until the Genii application is running.
6. Switch the power off with a long press of the power button
7. Switch the power on.
8. The new version of boot loader and operating system will be loaded

4.5.2 Analogue Main FPGA (DK458)


File name: DK458.bin

The FPGA is pre-programmed on the PCA as delivered by the subcontract production supplier.

This FPGA may be upgraded via the Genii IS application upgrade option.

1. Connect Genii-IS to a PC via the USB port, where it will appear as an external device, eg.
(DPI620 (F:) ).
2. Copy the FPGA program file(s) to a folder named ‘FPGA, located in the “Install” folder on
the connected device.
3. From the Genii IS application dashboard, touch the ‘Advanced’ button.
4. Enter PIN number 5487
5. Select Upgrade option from the menu.
6. Select the CH1 FPGA option.
7. Select ‘Yes’ from the ‘Confirm Upgrade’ prompt.
8. Power down and restart the instrument once the ‘Software Upgrade’ message appears.
9. The FPGA will be automatically upgraded.
10. Wait until the upgrade completes and the application restarts. This will happen
automatically and may take several minutes.

4.5.3 DK420: Genii IS Main Application (DK0420)


Folder name: AMC

4.5.3.1 Genii IS application first-time installation

1. Switch on Genii IS. It will search for the application upgrade software and not find it as it has
not been copied onto Genii IS yet, and display a dialog with options to Cancel or Retry. Do
not select either at this stage.

NON-PUBLIC Document SR0106


Page 13 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
2. Connect Genii-IS to a PC via the USB port, where it will appear as an external device, eg.
(DPI620 (F:) ).
3. Copy the AMC application folder to the F:/Install/
4. When copy is complete, select Retry on the Genii IS.
5. The application will be automatically installed.
6. Wait until the upgrade completes and the upgraded application is running. This will happen
automatically and may take several minutes.

4.5.3.2 Genii application upgrade (by using the power button)

1. Switch on Genii IS. Connect Genii-IS to a PC via the USB port, where it will appear as an
external device, eg. (DPI620 (F:) ).
2. Copy the AMC folder to F:/Install.
3. Press and hold the power button to switch off.
4. Press and hold the power button to switch on and continue to hold until the dialog appears
asking to Force OS Upgrade - Yes or No. Select No.
5. The application will be automatically upgraded.
6. Release the power button once the progress bar appears. This may take several seconds.
7. Wait until the upgrade completes and the upgraded application is running. This will happen
automatically and may take several minutes.

4.5.3.3 Genii application upgrade (from within the Genii application)

1. Switch on Genii IS. Connect Genii-IS to a PC via the USB port, where it will appear as an
external device, eg. (DPI620 (F:) ).
2. Copy the AMC folder to F:/Install.
3. From the Genii application dashboard, touch the ‘Advanced’ button.
4. Enter PIN number 5487.
5. Select Upgrade option from the menu.
6. Select APPLICATION option.
7. Select ‘Yes’ from the ‘Confirm Upgrade’ prompt.
8. Power down and restart the instrument once the ‘Software Upgrade’ message appears.
9. The application software will be automatically upgraded
10. Wait until the upgrade completes and the upgraded application is running. This will
happen automatically and may take several minutes.

NON-PUBLIC Document SR0106


Page 14 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
4.5.4 DPI620G-IS HART Processor Application (DK0464)

The HART processor application is pre-programmed on the HART modem microcontroller on


the PCA as delivered by the subcontract production supplier.

This HART modem firmware may be upgraded via the Genii IS application upgrade option.

1. Switch on Genii IS. Connect Genii-IS to a PC via the USB port, where it will appear as an
external device, eg. (DPI620 (F:) ).
2. Copy the HART processor firmware file to a folder named ‘HART’ under F:/Install.
3. From the Genii application dashboard, touch the ‘Advanced’ button.
4. Enter PIN number 5487
5. Select Upgrade option from the menu.
6. Select the HART DRIVER option.
7. Select ‘Yes’ from the ‘Confirm Upgrade’ prompt.
8. Power down and restart the instrument once the ‘Software Upgrade’ message appears.
9. The HART modem firmware will be automatically upgraded.
10. Wait until the upgrade completes and the application restarts. This will happen
automatically and may take several minutes.

4.5.5 DPI620G-IS HART Processor Bootloader (DK0463)


The HART processor bootloader is pre-programmed on the HART modem microcontroller on
the PCA as delivered by the subcontract production supplier.

The HART processor bootloader cannot be upgraded via the Genii application upgrade option.

4.5.6 HART SDC625 DD Library (DK0401)


• These files should be installed in the internal memory in the DPI620 Genii IS.

• Locate the self-extracting file called DPI620_DD_Library_2016_1.312.exe


from the drawing office.

• Connect the Genii IS to a PC. It will appear as a Drive F, G etc.

• Double Click to run the executable file.

• Select the … box (see picture)

Select this button

Extract to Genii IS (eg. G)

• Once the file is extracted, open the drive. The files structure should appear like:

NON-PUBLIC Document SR0106


Page 15 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
• Genii IS internal memory

NON-PUBLIC Document SR0106


Page 16 of 17 UNCONTROLLED WHEN PRINTED OR TRANSMITTED
ELECTRONICALLY Revision AF
CDG/SR0106/000/27
5 SOFTWARE CHECKS

5.1 Install Package Virus Scan


The results of Virus scan of each DK software item product (the install executables) are held in
the relevant DK document. The table below summarises the results of that scan.

DK Virus PASS /
DK DK
Results Date Checked By FAIL /
Number Revision
Checked ERROR
DK420 AE Yes 23/12/2020 C. Gosden Pass
Table 5 - File Virus Scan Checker

5.2 Cyber Security & Vulnerabilities


The results of Cyber Security and OSS scan checks are held in the relevant DK document. The
table below summarises the results of that analysis.

PASS /
DK DK
Checked Date Checked By FAIL /
Number Revision
ERROR
DK420 AE Yes 23/12/2020 C. Gosden See DK
Table 6 - Cyber Security Checker

5.3 Open Source Software (OSS) Notices


There is no OSS notice for this product