You are on page 1of 12

BIOMEDICAL

Cardiovascular Engineering and Technology, Vol. 13, No. 2, April 2022 ( 2021) pp. 279–290 ENGINEERING
https://doi.org/10.1007/s13239-021-00575-2 SOCIETY

Original Article

Simulating Radial Pressure Waveforms with a Mock Circulatory Flow


Loop to Characterize Hemodynamic Monitoring Systems
ANNA PACKY,1,2 GAVIN A. D’SOUZA,1 MASOUD FARAHMAND,1 LUKE HERBERTSON,1 and CHRISTOPHER G.
SCULLY 3
1
Center for Devices and Radiological Health, Office of Science and Engineering Laboratories, U.S. Food and Drug
Administration, Silver Spring, MD, USA; 2University of Maryland, College Park, MD, USA; and 3Center for Devices and
Radiological Health, Office of Science and Engineering Laboratories, U.S. Food and Drug Administration, Bldg. 62 Rm 1129,
10903 New Hampshire Ave., Silver Spring, MD 20993, USA
(Received 22 February 2021; accepted 15 August 2021; published online 1 September 2021)

Associate Editor Igor Efimov oversaw the review of this article.

Abstract device testing applications to hemodynamic monitoring


Purpose—Mock circulatory loops (MCLs) can reproducibly systems. Properly validated MCLs can create a stable and
generate physiologically relevant pressures and flows for reproducible environment for testing patient monitoring
cardiovascular device testing. These systems have been systems over their entire operating ranges prior to clinical
extensively used to characterize the performance of thera- use.
peutic cardiac devices, but historically MCLs have had
limited use for assessing patient monitoring systems. Here, Keywords—Hemodynamic monitoring devices, Cardiac de-
we adapted an MCL to include peripheral components and vice testing, Pulse contour analysis, Cardiovascular flow
evaluated its utility for qualitative and quantitative benchtop
testing of hemodynamic monitoring devices. circuit.
Methods—An MCL was designed to simulate three physio-
logical hemodynamic states: normovolemia, cardiogenic
shock, and hyperdynamic circulation. The system was INTRODUCTION
assessed for stability in pressure and flow values over time,
repeatability, waveform morphology, and systemic-periph- Hemodynamic monitoring devices employ tech-
eral pressure relationships.
Results—For each condition, cardiac output was controlled
niques such as pulse contour analysis of radial artery
to the nearest 0.2 L/min, and flow rate and mean arterial pressure waveforms to provide continuous information
pressure remained stable and repeatable over a 60-s period about the hemodynamics of a patient in a minimally
(n = 5, standard deviation of ± 0.1 L/min invasive manner.3 Testing hemodynamic measure-
and ± 0.84 mmHg, respectively). Transfer function analyses ments has generally involved evaluating device mea-
showed that the systemic-peripheral relationships could be
adequately manipulated. The results from this MCL were
surements against more invasive reference
comparable to those from other published MCLs and measurements such as triplicate thermodilution mea-
computational simulations. However, resolving current lim- surements of cardiac output (CO) using central ca-
itations of the system would further improve its utility. Three theters, which have their own uncertainties and
pulse contour analysis algorithms were applied to the limitations31,7 and are declining in use, limiting patient
pressure and flow data from the MCL to demonstrate the
potential role of MCLs in characterizing hemodynamic
populations for testing.9,36 In addition, the use of
monitoring systems. minimally invasive, continuous monitoring devices
Conclusion—Overall, the development of robust analysis often involves characterizing rapid changes in CO and
methods in conjunction with modified MCLs can expand associated parameters to understand patient responses
to treatment, which may occur more quickly than the
Address correspondence to Christopher G. Scully, Center for time it takes to perform successive thermodilution
Devices and Radiological Health, Office of Science and Engineering measurements.1 The differences in physiological
Laboratories, U.S. Food and Drug Administration, Bldg. 62 Rm responses among patients makes it difficult to assess
1129, 10903 New Hampshire Ave., Silver Spring, MD 20993, USA.
Electronic mail: Christopher.Scully@fda.hhs.gov
279
1869-408X/22/0400-0279/0  2021 This is a U.S. government work and not under copyright protection in the U.S.; foreign copyright protection may apply
280 PACKY et al.

device reproducibility and to compare diagnostic re- diac hemodynamics and peripheral pressure wave-
sults between devices in a clinical setting.30 forms. This allows us to set known changes in
A mock circulatory loop (MCL) is a useful tool for hemodynamics under a range of physiologically rele-
simulating cardiovascular hemodynamics on the vant conditions towards designing test methods for
bench. Different pathophysiologic conditions, repre- evaluating how monitoring devices can track those
senting a range of disease states and patient popula- changes. We characterized our MCL to assess the
tions, can be evaluated in a pre-clinical setting using an overall controllability across the desired functional
MCL. Academic researchers and device manufacturers range. This process includes: (1) characterizing the
have utilized MCLs for evaluating the hemodynamics system to mimic the hemodynamics over a range of
of implantable cardiovascular devices4,22 prior to clinically relevant cardiac states; (2) assessing the
conducting more expensive animal studies and clinical repeatability and stability of the pressure and flow
trials. Also, MCLs have been used for years to char- values, as well as its ability to resolve small changes in
acterize the fluid dynamics within therapeutic cardio- flow; and (3) qualitatively and quantitatively evaluat-
vascular devices, including mechanical circulatory ing the arterial pressure waveform characteristics for
support devices, heart valves, vascular conduits and three hemodynamic states. We then applied three pulse
grafts, and catheters.10–27 Further, the efficacy of using contour analysis algorithms for measuring CO to the
hemodynamic indices for diagnosing cardiovascular peripheral pressure waveforms to demonstrate the
diseases including vascular and valvular stenosis, an- potential for non-clinical characterization of hemody-
eurysms, and congenital heart defects has been evalu- namic monitors during known changes in flow and
ated using MCLs.17–2 Experimental studies using pressure using an MCL.
MCLs have led to the optimization of device designs
and the determination of appropriate performance
parameters over the full operating range of the device MATERIALS AND METHODS
in a well-controlled environment, which may not be
attainable through animal studies or clinical trials. Design of Mock Circulatory Loop
Thus, MCLs enable the evaluation of certain devices An MCL was first designed to create physiologic
earlier in their total product life cycle, leading to design pressure and flow waveforms. This system consists of
improvements and a better understanding of perfor- physical components that mimic the left ventricular
mance and limitations before more costly clinical trials. function, as well as the systemic and peripheral circu-
In recent years, MCLs have been modified to lations comprised of arteries and veins. A block dia-
incorporate peripheral vascular components for simu- gram of the MCL is shown in Fig. 1, and images of the
lating arterial pressure waveforms.42,14 Hemodynamic physical MCL are shown in the Supplementary
monitoring devices can be connected to these circuits Information as Figure S1. A description of the sys-
and algorithms can be applied28 to estimate invasive, temic loop is provided first followed by a description of
central hemodynamic parameters from measured the peripheral loop.
pressure signals that are representative of in vivo radial Deionized water is circulated through the MCL
artery pressure. Therefore, it is critical to fully char- using a piston pump [ViVitro Labs, Victoria, BC,
acterize peripheral pressure waveforms within an MCL Canada] that is fluid-coupled to a suspended silicone-
and then correlate them to the cardiac hemodynamics based mock ventricular sac. This causes the ventricular
to demonstrate the utility of using MCLs for evaluat- sac to expand and contract during each beat of the
ing hemodynamic monitoring devices. The focus of pump. The wall motion of the ventricular sac is con-
this work is to investigate if an MCL with peripheral trolled by the inputted piston waveform, beat rate, and
vascular components provides central-peripheral amplitude of piston motion. The ventricular sac mo-
pressure relationships across a range of hemodynamic tion causes the artificial (bi-leaflet mechanical) heart
states that can then be used for generating represen- valves [St. Jude Medical Regent Valve, St. Jude Med-
tative radial pressure waveforms for hemodynamic ical, Inc., St. Paul, MN], to open and close, which al-
monitoring device testing. A controllable MCL can lows fluid to be ejected unidirectionally from the left
help evaluate how well a medical device can monitor ventricle into the systemic part of the MCL during
changes in flow and pressure values. This can poten- systole. Source compliance, which simulates the left
tially expand the role of MCLs in medical device ventricular viscoelastic behavior, was controlled by
testing. manually adjusting the air volume in the source com-
We describe an MCL with peripheral vascular pliance chamber using a syringe. The baseline com-
components and provide a characterization of the pliance level was defined as complete unloading of the
central-peripheral pressure relationships. The MCL ventricular sac. Source compliance on the piston pump
offers many adjustable settings to achieve target car-
BIOMEDICAL
ENGINEERING
SOCIETY
Simulating Radial Pressure Waveforms 281

FIGURE 1. Block diagram of the mock cardiovascular flow circuit. Arrows represent the flow direction of the systemic loop (grey)
and the peripheral loop (black). The atrial, ventricular, and venous chambers contain water for circulation through the system. All
other block units can be modified to simulate different types of flows. *Pressure measurement location. °Flow measurement
location.

was set before the test by adding 60 mL of air to the lar resistance is simulated using a manually controlled
working fluid which surrounds the outside of the screw compressor tubing clamp. Repeatability with the
ventricular sac. Distal to the ventricle, fluid passes screw clamp between experiments and when re-setting
through the aortic root and ascending aorta, and then the system was achieved by approximating the distance
through tubing representative of the systemic circula- between the clamp jaws using a caliper. Although this
tion, where the flow rate and aortic pressure are technique of manual adjustment is not highly accurate
recorded. Two air-filled (passive) compliance chambers in terms of repeatability, the set point for the screw
are connected to the aortic root and ascending aorta to clamp was influenced primarily by the desired flow
simulate the compliance of the respective vessels. A split between the peripheral and systemic loops. Fluid
third air-filled (passive) compliance chamber is placed from the peripheral loop circulates back into the sys-
in-line in the mock descending aorta to simulate the temic loop after the venous reservoir. Since the MCL is
systemic vascular compliance. All the air-filled com- an open-loop system consisting of a venous reservoir
pliance chambers were manually controlled using a and atrial chamber open to atmosphere (Figs. 1 and
syringe for adjusting and setting the volume of air in S1), the transient pressure pulse in the proximal and
the chamber to simulate the compliance effect. distal regions of the venous reservoir are similar. Thus,
Mechanical resistance, simulating the systemic vascu- attaching the peripheral loop after the venous reservoir
lar resistance, is applied before the systemic loop re- in the MCL (instead of the anatomically accurate
turns to the large venous reservoir using an proximal position) is not expected to alter the patho-
electronically activated ball valve. The ball valve en- physiologic hemodynamics of the peripheral circula-
ables a set amount of mechanical resistance (measured tion. Supplementary Table S1 includes details of the
as the % occlusion of the tubing cross-section) to be MCL components including tubing dimensions and
applied using a specific voltage value (0.0–10.0 Volts), material properties.
thus allowing for easy repeatability between experi- Two flow sensors and four pressure transducers
ments and when re-setting the system. Fluid from the were used in the MCL to monitor hemodynamics at
large venous reservoir is then circulated back into the the locations identified in Fig. 1. Aortic and peripheral
left atrial chamber to complete the systemic loop. loop flow rates were measured simultaneously using
The peripheral loop branches off the systemic loop ultrasound Doppler clamp-on flow sensors [ME11PXL
at the mock descending aorta proximal to the systemic (aorta) and ME5PXL (peripheral), Transonic Systems,
compliance chamber. Peripheral vascular compliance Inc., Ithaca, NY] in the mock descending aorta and in
is simulated using anatomically representative tubing the representative radial artery, respectively. The two
durometers, diameters, and lengths. Peripheral vascu- flow sensors were coupled to a flow meter [T402,
BIOMEDICAL
ENGINEERING
SOCIETY
282 PACKY et al.

Transonic Systems, Inc., Ithaca, NY], which recorded Prior to testing, a verification step was performed to
and displayed the instantaneous flow rate (in mL/min ensure that the MCL could replicate a normal physi-
or L/min). Left ventricular and left atrial pressures ologic condition, with pressures of 120/80 mmHg and
were measured with disposable pressure transducers a CO of 5 L/min. Additionally, the fluid height levels
[Utah Medical Products Inc., Midvale, Utah] and sent were set for each compliance chamber, and any resid-
to an amplifier module [Bus 21097, ViVitro Labs, ual air removed from the flow loop. To set each
Victoria, BC, Canada]. The aortic and peripheral hemodynamic state, the resistances, beat rate, piston
pressures were measured using invasive pressure ca- waveform, and gain were first adjusted to produce the
theters [SPR-350, Millar Instruments, Houston, TX] nominal flow setting for that state. Then, MAP was set
coupled to an amplifier [PCU-2000 Pressure Control to the desired level by adjusting the systemic resistance
Unit, Millar Instruments, Houston, TX]. The periph- produced by the ball valve. Next, the gain on the pis-
eral pressure measurement was made 0.83 m distal to ton pump was manually adjusted to achieve the target
the aorta. All pressure and flow signals were acquired CO. For each hemodynamic state, all measurements
using a data acquisition system [NI-9205, National were recorded at the nominal flow for 60 seconds,
Instruments, Austin, TX] at a sampling rate of 1000 followed by data acquisition over a range of flow rates.
Hz, and an in-house software control program was Specifically, COs of ± 1 L/min of the nominal nor-
developed to modify, display, and analyze the hemo- movolemic and cardiogenic shock conditions were
dynamic signals. produced with steps of ± 0.2 L/min, and of ± 2 L/min
of the hyperdynamic state with steps of ± 0.4 L/min,
as shown in Fig. 2, by adjusting the gain of the piston
Experimental Design
pump.
The MCL was used to simulate three different
physiologic hemodynamic states for demonstrating the
Data Analysis
versatility of the system: normovolemia, cardiogenic
shock, and hyperdynamic. The representative hemo- Pressure and flow data were analyzed using MA-
dynamic states were based on average physiologic TLAB [MathWorks, Natick, MA]. A low-pass filter
ranges found in the literature.40,23 The nominal nor- with a cut-off frequency of 50 Hz was applied to the
movolemic state was defined here to have a beat rate of pressure and flow waveforms, which were subsequently
75 beats per minute (BPM), CO of 5 L/min, and mean down-sampled to 128 Hz. Individual beat cycles were
arterial pressure (MAP; defined as the time-averaged detected using a previously reported pressure pulse
aortic pressure) between 90 and 100 mmHg. The flow beat detector.35,16 All measurements were calculated
rate was varied by ± 1 L/min from the nominal flow for each beat cycle and averaged across the 60-second
for the normovolemic state. For the cardiogenic shock acquisition period. Stroke volume (SV) from the ven-
state, beat rate was set to 90 BPM, CO to 3 ± 1 L/min, tricular sac was calculated using the average flow of a
and MAP was less than 70 mmHg.38 For the hyper- 60-second acquisition period divided by the beat rate.
dynamic state, beat rate was set to 120 BPM, CO to Systemic vascular resistance (SVR) was calculated as
8 ± 2 L/min, and MAP was between 65 and 70
mmHg.28–32
The system variables, including the peripheral and
systemic resistances, beat rate, and the gain applied to
the piston waveform, were adjusted to generate the
three hemodynamic states. The systemic vascular
resistance, set using the ball valve, was empirically
derived for each hemodynamic state. The peripheral
vascular resistance was set to achieve a peripheral loop
flow rate equating to approximately 3% of the total
CO to achieve a consistent target for adjusting
peripheral resistance in each state, based on an initial
baseline position of the screw clamp (i.e., distance
between clamp jaws) used in the peripheral loop. The FIGURE 2. The flow rate was manipulated for each
hemodynamic state across a range of 61 L min21 from the
clamp position was kept constant as the flow rate was nominal flow rate (62 L min21 for the hyperdynamic state) in
varied for each hemodynamic state. This provided 0.2 L min21 increments (0.4 L min21 for the hyperdynamic
peripheral flow rates during the normovolemic state state). Each measurement step was captured for 60 seconds.
The first three steps were performed at the nominal, low, then
from ~120 to 170 mL/min, which is on the order of high flow rates done for each state. The total number of
measured radial artery volume flows.25,21 completed steps for each state was 25.
BIOMEDICAL
ENGINEERING
SOCIETY
Simulating Radial Pressure Waveforms 283

the mean pressure drop from the aorta to the left at- namic state. The difference between the overall mini-
rium divided by CO. mum and maximum (range) values of mean aortic and
Stability of the pressure and flow values over time peripheral pressures across experiments were 76 and 74
and repeatability of the MCL were determined for each mmHg, respectively. The average SVR range was 1012
hemodynamic state. Stability was assessed by fitting a dyn s cm-5. The overall mean aortic flow rate range
linear line to the recorded pressure and flow values across experiments (difference between maximum and
over an acquisition period using the nominal level for minimum flow rate achieved) was 7.9 L/min and the
each hemodynamic state and assessing the slope. We mean peripheral flow rate range was 0.2 L/min. The
used the data from the 5 nominal positions for each gain settings were set between 19.7% and 59.4% to
hemodynamic state (see Fig. 2) to evaluate the achieve the desired flow rates for the three hemody-
repeatability by determining the variation in the pump namic states, and the ball valve position was set at
gain required to achieve the same flow rates and 62%, 64%, and 46% occlusion for the normovolemic,
pressures across the 5 replicate trials. cardiogenic shock, and hyperdynamic states, respec-
Discrete Fourier transforms were calculated for the tively. The peripheral flow was approximately 3% of
pressure signals following linear detrending to identify the CO for each hemodynamic state (Table 1), and it
the harmonics present in the generated waveforms. The did not vary substantially across the flow rate steps
relationship between aortic and peripheral pressure within each hemodynamic state (0.4, 0.5, and 0.4%
waveforms was evaluated by estimating the transfer ratio differences for the normovolemic, cardiogenic
function using a Hamming window (length of 1024 shock, and hyperdynamic states, respectively). The
samples with 512 samples overlapping). Transfer large achievable flow rate range demonstrates the
function data are reported for each hemodynamic state ability to simulate many physiologic conditions with
as mean ± standard deviation across each 60-second this MCL.
acquisition period. Flow rate was adjusted through a 22-step protocol
of 0.2 L/min increments for each hemodynamic state
(0.4 L/min for the hyperdynamic state). The mean and
Characterizing Hemodynamic Monitoring Algorithms
standard deviation of the CO at each step are shown in
Three pulse contour analysis algorithms were Fig. 4. For cardiogenic shock, we observed increased
applied to the radial pressure waveforms recorded for variability in the measured flow within each one-min-
each hemodynamic state: Liljestrand-Zander (CO = ute step as the flow rate was increased (Fig. 4d). The
k * (SBP  DBP)/(SBP + DBP) * HR), Pulse Pres- average coefficient of variation across steps for both
sure (CO = k *R (SBP  DBP) * HR), and Systolic the hyperdynamic and cardiogenic shock states was
Area (CO ¼ k  ABPðtÞdt  HR), where SBP and 2.2%, as a percentage of the mean value. The nor-
DBP refer to systolic and diastolic blood pressure, HR movolemic state, which was the middle condition in
to heart rate, and ABP to arterial blood pressure terms of CO, showed a lower average coefficient of
waveform.35 Each method requires a calibration value variation of 1.3%. The hyperdynamic state had a
‘k’. This was determined for each of the three states higher standard deviation in beat-to-beat CO values at
from the initial settings by dividing the uncalibrated each step compared to the normovolemic and cardio-
value from each algorithm by the reference flow rate, genic shock states (Fig. 4d). Overall, the maximum
and then the algorithms were applied to measure CO at coefficient of variation observed across any step for
each subsequent step in the experimental protocol de- any test condition was 5.4%, which occurred at the
scribed above and compared to the reference systemic highest flow rate of the cardiogenic shock state.
flow measurement.
Stability of Pressure and Flow Values and Repeatability
RESULTS The normovolemic and hyperdynamic states
showed a slight decrease in MAP of 1.8 mmHg and 0.6
Hemodynamic States mmHg, respectively, over the 60-second period. This
Each hemodynamic state was studied over a range decrease is small in comparison to the absolute values
of flow rates, pressures, and beat rates, Table 1. For of the MAP (~94 and 60 mmHg, respectively, Tables 1
each hemodynamic state, controllable changes to flow and 2) for those specific states. The MAP of the car-
and pressure were produced by adjusting the piston diogenic shock state was stable over the test duration.
gain, as shown in Figure 3 for the normovolemic state. Mean flow rates were stable for each condition over
Table 1 shows the full range of the system using the the 60-second duration.
lowest, nominal and highest settings of each hemody- The repeatability of the MCL at the different
hemodynamic states is characterized in Table 2. The
BIOMEDICAL
ENGINEERING
SOCIETY
284 PACKY et al.

TABLE 1. Range of the system capabilities for each hemodynamic state.

Normovolemic Cardiogenic shock Hyperdynamic

High Nominal Low High Nominal Low High Nominal Low

HR (BPM) 75 75 75 90 90 90 120 120 120


CO (L/min) 5.98 5.03 4.02 3.97 2.98 2.04 9.92 8.05 5.97
PF (L/min) 0.17 0.14 0.12 0.11 0.08 0.07 0.28 0.24 0.20
PTF (%) 2.79 2.83 2.92 2.69 2.80 3.20 2.86 2.95 3.29
SV (mL) 79.7 67.0 53.6 44.1 33.1 22.7 82.7 67.1 49.8
ABP (mmHg)
Diastolic 95.3 73.4 55.5 69.7 50.4 38.0 55.7 44.3 35.4
Mean 121.0 94.0 71.2 84.1 60.7 45.0 76.2 60.5 47.3
Systolic 152.2 119.3 90.4 103.2 75.3 56.1 108.1 84.7 65.9
PP 56.9 45.9 34.9 33.5 24.8 18.1 52.3 40.3 30.6
PBP (mmHg)
Diastolic 91.7 70.4 53.3 67.0 48.4 36.6 55.6 43.7 34.5
Mean 117.8 91.5 69.3 81.8 59.0 43.8 74.0 58.8 45.5
Systolic 156.3 122.8 93.1 108.1 78.5 58.1 113.0 88.6 67.8
PP 64.6 52.4 39.7 41.1 30.2 21.5 57.1 44.4 32.6
AtP (mmHg)
Mean 8.5 9.4 9.9 11.1 11.8 12.1 9.8 10.4 10.5
VBP (mmHg)
Mean  66.6 53.9 42.1 48.6 37.1 30.0 62.3 48.6 36.7
s
SVR dyncm 5 1505.0 1346.2 1219.1 1471.2 1312.6 1288.4 535.7 498.1 492.5

HR heart rate, CO cardiac output, PF peripheral flow, PTF percent of total flow through peripheral loop, SV stroke volume, ABP aortic blood
pressure, PP pulse pressure, PBP peripheral blood pressure, AtP atrial blood pressure, VBP ventricular blood pressure, SVR systemic
vascular resistance.

normovolemic and hyperdynamic flow rates varied by


roughly 1% for the 5 repeated trials. While the MCL is
repeatable and stable for the operating ranges tested
here, there may be increased variability in both CO
and MAP when the piston gain is further increased.
The gain values of the normovolemic and hyperdy-
namic states are nearly double that of the cardiogenic
shock state, resulting in higher standard deviations in
both CO and MAP for these two states. Repeatability
testing involved returning to the same state after
increasing and decreasing the flow. During the hyper-
dynamic trials, air was added to the compliance
chambers, which may have contributed to the vari-
ability among replicate tests.

Waveform Assessment
The systemic and peripheral pressure waveforms
resembled physiologic arterial waveforms, which in-
clude characteristics such as a steep rise during systole
FIGURE 3. 10 second segment traces for the peripheral (a) and a more prominent dicrotic notch in the peripheral
pressure and (b) flow waveforms at the nominal, low, and high
flow settings for the normovolemic state. Each setting was waveform (Fig. 5). The dicrotic notch is more evident
run for 60 seconds and the dashed vertical lines indicate that in the peripheral pressure waveforms in comparison to
the waveforms were truncated to only display the first 10 the systemic pressure waveforms, especially for the
seconds for each setting.
cardiogenic shock and hyperdynamic states. The pulse
standard deviation of the MAP across the 5 repeated morphology remains similar at each flow rate step for
trials when flow was set at the nominal value (see each state, seen in both Figs. 3 and 5. Figure 6 shows
Fig. 2) was within 1 mmHg for each state, and the the Fourier transform of the peripheral pressure
BIOMEDICAL
ENGINEERING
SOCIETY
Simulating Radial Pressure Waveforms 285

FIGURE 4. Aortic flow rates at each step for the (a) normovolemic (b) cardiogenic shock, and (c) hyperdynamic states. (d) The
mean cardiac output (CO) compared to the standard deviation of the cardiac output in L/min for each hemodynamic state. NV
normovolemic, CS cardiogenic shock, HD hyperdynamic.

TABLE 2. Repeatability of the flow loop at each of the


hemodynamic states at the nominal flow level (Mean 6 changes for each of the three hemodynamic states
Standard Deviation). (Fig. 8). Pulse Pressure and Systolic Area methods
demonstrate reasonable tracking of changes in the
Repeatability
reference systemic flow with some undershoot/over-
Gain (%) CO (L/min) MAP (mmHg) shoot of changes that varies based on the hemody-
namic state, but the Liljestrand-Zander method is not
Normovolemic 48.1 ± 0.00 5.0 ± 0.07 93.98 ± 0.81
Cardiogenic shock 26.0 ± 0.00 3.0 ± 0.04 60.89 ± 0.15
able to track the changes.
Hyperdynamic 47.6 ± 0.88 8.1 ± 0.07 60.60 ± 0.84

DISCUSSION

waveforms at the initial nominal step for each hemo- Hemodynamic monitoring devices can continuously
dynamic state. Information content decays at higher measure critical cardiovascular parameters by analyz-
frequencies, with a modulus of < 1 mmHg above 7 ing the peripheral pressure waveforms. However, it can
Hz. be challenging to collect adequate reference measure-
The systemic-peripheral transfer functions show ments for repeatably and precisely evaluating these
increasing amplification at the peripheral location parameters. It is also difficult to understand how dif-
from 0 to approximately 5 Hz, after which the ampli- ferent confounding factors affect the performance of
fication decreases (Fig. 7a). The amplitude varied hemodynamic monitoring devices. Non-clinical test
across hemodynamic states with the cardiogenic shock methods can potentially provide repeatable conditions
state exhibiting the highest amplitude and the hyper- for assessing safety and performance aspects of
dynamic state having the lowest amplitude at most hemodynamic monitoring devices that are difficult to
frequencies. The phase delay showed less variability determine and isolate through clinical studies. MCLs
(Fig. 7b) compared to reported aortic-radial pressure provide potential means for augmenting or reducing
transfer functions, which showed phase delays up to -8 the amount of invasive and expensive animal and
radians by 10 Hz.20 human testing that is often required for medical device
development. The focus of MCL development has
historically been for assessing cardiac devices such as
Application of Cardiac Output Algorithms
ventricular assist devices and heart valves27,37–11;
After calibration at the baseline step, the pulse therefore, the modules necessary for modeling the ra-
contour analysis algorithms generally tracked CO dial arterial pressure waveforms are not included in
BIOMEDICAL
ENGINEERING
SOCIETY
286 PACKY et al.

FIGURE 5. Pressure waveforms for the hemodynamic states. The nominal waveform for each state is a solid black line, the low is
a dashed-dotted line, and the high is a dashed line. (a) Aortic pressure at normovolemic state, (b) peripheral pressure at
normovolemic state, (c) aortic pressure at cardiogenic shock state, (d) peripheral pressure at cardiogenic shock state, (e) aortic
pressure at hyperdynamic state, and (f) peripheral pressure at hyperdynamic state.

this generation of MCLs. Recently, researchers devel- that the flow loop can produce a wide range of phys-
oped an MCL platform capable of generating radial iologic conditions to simulate different clinical sce-
arterial waveforms.42 However, the ability of MCLs to narios.
simulate radial arterial pressure remains limited. The Pulse contour analysis uses a peripherally recorded
challenges include difficulties in achieving a realistic arterial pressure waveform to continuously estimate
transfer function between systemic arterial pressure hemodynamic parameters such as CO in real-time.
and radial pressure under various hemodynamic states Advantages of these devices include their ability to
as well as achieving a framework capable of producing identify and monitor changes in CO that are indicative
a broad range of flow rates under different hemody- of how a patient responds to therapy, or that estimate
namic conditions for assessing the performance of responsiveness such as through dynamic indices
hemodynamic monitoring systems. In this work, we including stroke volume variation.44 These applica-
adapted an MCL that was previously used only for tions measure small changes in CO over short time
cardiac device testing (e.g., ventricular assist devices, scales compared to traditional reference measure-
heart valves) to include peripheral components. The ments. In this respect, intermittent reference measure-
adapted MCL was used to generate and acquire ments can offer only limited information about the
peripheral pressure waveforms to assess hemodynamic time-varying hemodynamic behavior. Non-clinical
monitoring devices under known conditions and set testing tools such as MCLs have the potential to allow
changes to those conditions, something that cannot be us to better understand properties of monitoring sys-
feasibly performed in clinical studies. Our results show tems such as the response to a known change of CO

BIOMEDICAL
ENGINEERING
SOCIETY
Simulating Radial Pressure Waveforms 287

FIGURE 7.. Systemic-peripheral pressure transfer functions


for the hemodynamic states, showing (a) amplitude and (b)
phase difference. Note: step 24 of the normovloemic state was
not included due to a synchronous recording error between
the pressure and flow measurements. NV normovolemic, CS
cardiogenic shock, HD hyperdynamic.

approach was developed considering non-linear arte-


rial compliance-pressure relationships, which the cur-
rent design of the MCL does not address. This may be
one reason for the lack of tracking performance in the
current study.34
FIGURE 6. Frequency spectra of peripheral pressure
waveforms in the MCL for the (a) normovolemic, (b) While we presented an initial iteration of the MCL
cardiogenic shock, and (c) hyperdynamic states. here, additional optimization, refinement, and charac-
terization of the flow loop is needed to incorporate
over specified time-scales or the limit of resolution (i.e., such a loop into hemodynamic monitoring device
minimum detectable change). Thus, in this study, we testing. Historically, MCLs have been used for char-
tested the feasibility of generating and characterizing acterizing the hemodynamic performance of cardio-
different physiological waveforms on the bench using vascular devices where indices such as pressure, flow
an MCL. rate, velocity, and shear stress within the device are
We applied three different pulse contour analysis evaluated under patient- and disease-specific condi-
algorithms to the recorded peripheral pressure data to tions. For hemodynamic monitoring device testing,
demonstrate how the MCL could be used for charac- this type of characterization can be expanded to con-
terizing pressure-derived CO measurement devices. sider not only the physiologic range, but also to
For this testing, the piston beat rate and systemic encompass the limits of resolution of the system and
resistance and compliance were kept constant so that the pressure waveform properties. Pulse contour
only the ability to track changes in flow rate was tested. analysis devices estimate CO, a central flow measure-
This is in contrast to pre-clinical animal or clinical ment, based on peripheral pressure waveforms.
testing where it is not feasible to directly manipulate Therefore, the flow loop should provide physiologi-
only a single characteristic. Two of the approaches, cally relevant aortic-peripheral pressure relationships
Pulse Pressure and Systolic Area, were able to track and pressure-flow relationships. We demonstrated
changes in CO in the MCL for each of the three systemic-peripheral transfer functions from the MCL
hemodynamic states with some overshoot and similar up to ~6 Hz to aortic-radial transfer functions
undershoot that varied between the different hemody- in clinical reports.13 A lack of branching and associ-
namic states. The third approach, Liljestrand-Zander, ated wave reflections in both the aortic and peripheral
was not able to track flow changes in the MCL within waveforms could explain why the current MCL setup
each hemodynamic state. The Liljestrand-Zander does not follow the trends of clinical and computa-
BIOMEDICAL
ENGINEERING
SOCIETY
288 PACKY et al.

our system. The MCL would better mimic the clinical


setting if a water-glycerin solution5 that more closely
mimics the viscosity of blood is used instead. A second
limitation is that while we could control the flow rate,
this needed to be done by manually adjusting a gain
knob, which limits the type of automated testing pro-
cedures that could be implemented. For example,
manual adjustments to the gain make it difficult to
have precise timing for evaluating temporal response
of hemodynamic monitoring systems. However, a
system which provides the flexibility of manipulating
the stroke volume directly is still advantageous over
animal or clinical testing, for which precise and
repeatable manipulations of pressures, flows, and
vascular properties (e.g., resistance) cannot be assured.
A third limitation is related to the mock aortic cham-
ber in our current design, which is anatomically inac-
curate causing a pressure drop of 10 mmHg due to the
presence of a 90-degree bend. For our study purposes,
we measured the aortic pressure at a location directly
distal to this bend. Future design iterations of the
MCL will need to include an anatomically accurate
mock aorta. A fourth limitation is the previous men-
tioned lack of non-linear relationship between arterial
compliance and pressure, which can affect the testing
of certain algorithms as shown in Fig. 8. These limi-
tations can be improved upon through future design
changes to the MCL. One limitation of MCLs, in
FIGURE 8. Estimated cardiac output values from three pulse general, relates to the difficulty in replicating vascular
contour analysis algorithms for the (a) normovolemic, (b) properties including compliance and resistance, and
cardiogenic shock, and (c) hyperdynamic conditions.
manipulating them over time. While testing with
benchtop MCLs provides advantages such as having
tional models past this point. This may be caused by an the ability to reproducibly test extreme operating
insufficient representation of the human anatomy in conditions in a well-controlled environment, this type
the MCL, including limited branching and viscoelastic of bench testing can be enhanced by coupling the
properties of the arteries, which could lead to damp- physical flow loops with computational models to
ened pulse wave reflections and fewer higher frequency further incorporate physiologic changes.24,18
components in the waveforms. Another study that In summary, an MCL was developed and utilized to
utilized an MCL42 defined a sophisticated representa- generate a range of physiologic conditions represented
tion of the radial pulse waveform based on the com- by three unique hemodynamic states. The MCL
parisons of early and late systolic pressure, as well as hemodynamics were adjusted using 0.2 L/min incre-
the dicrotic notch location for the average middle-aged ments to execute an experimental protocol which
person. While the analyses of the current MCL does accounts for subtle increases and decreases in CO. The
not specifically look at these factors, our analysis is MCL generated consistent and repeatable results over
done using a larger range of factors, including a wide a 60-second period and multiple trials. Using a versa-
range of hemodynamic parameters, morphology of the tile MCL, several hemodynamic measurements and
waveforms, and relationships between systemic-pe- waveform characteristics compared favorably with
ripheral pressures and flows. data published in the literature. We were able to apply
While our results showed that our MCL could multiple pulse contour analysis algorithms to pressure
simulate a wide range of physiologic conditions, was waveforms recorded from the MCL to compare the
relatively stable and repeatable, was controllable for estimated CO to reference flow values under known
small changes in flow rate, and provided pressure conditions. Future improvements to the MCL would
waveforms with realistic morphology, there are several allow for more comprehensive testing of these and
limitations to the study. One such limitation is the use other hemodynamic states, while accounting for dif-
of deionized water as the test fluid to rapidly design ferent patient populations and cardiac dysfunctions.
BIOMEDICAL
ENGINEERING
SOCIETY
Simulating Radial Pressure Waveforms 289
3
Overall, this initial study provides valuable insight Boisson, M., M. Poignard, B. Pontier, O. Mimoz, B. De-
towards development of MCLs as non-clinical testing baene, and D. Frasca. Cardiac output monitoring with
thermodilution pulse-contour analysis vs. non-invasive
tools for evaluating hemodynamic monitoring devices.
pulse-contour analysis. Anaesthesia. 74(6):735–740, 2019.
4
Broda, C., P. Smith, Y. Wang, O. Frazier, W. Dreyer, I.
Adachi, et al. Hemodynamic improvement with applica-
SUPPLEMENTARY INFORMATION tion of mechanical circulatory support in heart failure with
preserved ejection fraction in a mock circulation loop. J.
Heart Lung Transpl. 38(4):S373, 2019.
The online version contains supplementary material 5
Brookshier, K. A., and J. M. Tarbell. Evaluation of a
available at https://doi.org/10.1007/s13239-021-00575- transparent blood analog fluid: aqueous xanthan gum/
2. glycerin. Biorheology. 30(2):107–116, 1993. https://doi.org/
10.3233/bir-1993-30202.
6
Chang, T.-I., K.-H. Hsu, C.-W. Luo, J.-H. Yen, P.-C. Lu,
and C.-I. Chang. In vitro study of trileaflet polytetrafluo-
ACKNOWLEDGMENTS roethylene conduit and its valve-in-valve transformation.
The authors would like to acknowledge Dr. Sungtae Interact Cardiovasc Thorac Surg. 30(3):408–416, 2020. h
ttps://doi.org/10.1093/icvts/ivz274.
Shin for his work in developing the in-house software 7
Critchley, L. A., and J. A. Critchley. A meta-analysis of
for controlling the MCL. studies using bias and precision statistics to compare car-
diac output measurement techniques. J. Clin. Monit.
Comput. 15(2):85–91, 1999.
8
DISCLAIMER Della Rocca, G., M. G. Costa, P. Chiarandini, G. Bertossi,
M. Lugano, L. Pompei, et al. Arterial pulse cardiac output
The mention of commercial products, their sources, agreement with thermodilution in patients in hyperdy-
or their use in connection with material reported herein namic conditions. J. Cardiothorac. Vasc. Anesth.
22(5):681–687, 2008.
is not to be construed as either an actual or implied 9
Doshi, R., K. Patel, P. Patel, and P. M. Meraj. Trends in
endorsement of such products by the Department of the utilization and in-hospital mortality associated with
Health and Human Services. This article reflects the pulmonary artery catheter use for cardiogenic shock hos-
research conducted by the authors and should not be pitalizations. Indian Heart J. 70(Suppl 3):S496, 2018.
10
construed to represent FDA’s views, guidance, or D’Souza, G. A., R. K. Banerjee, and M. D. Taylor. Eval-
uation of pulmonary artery stenosis in congenital heart
policies. disease patients using functional diagnostic parameters: an
in vitro study. J Biomech. 81:58–67, 2018. https://doi.org/
10.1016/j.jbiomech.2018.09.014.
11
CONFLICT OF INTEREST Ferrari, G., C. De Lazzari, R. Mimmo, G. Tosti, D. Am-
brosi, and K. Gorczynska. A computer controlled mock
Anna Packy, Gavin D’Souza, Masoud Farahmand, circulatory system for mono-and biventricular assist device
Luke Herbertson, and Christopher Scully declare that testing. Int. J. Artif. Organs. 21(1):26–36, 1998.
12
they have no conflict of interest. Figliola, R. S., A. Giardini, T. Conover, T. A. Camp, G.
Biglino, J. Chiulli, et al. In vitro simulation and validation
DATA AVAILABILITY of the circulation with congenital heart defects. Prog
Pediatr Cardiol. 30(1–2):71–80, 2010. https://doi.org/10.10
Electronic data reported in this study will be made 16/j.ppedcard.2010.09.009.
13
available upon request. Gallagher, D., A. Adji, and M. F. O’Rourke. Validation of
the Transfer Function Technique for Generating Central
from Peripheral Upper Limb Pressure Waveform. Oxford:
CODE AVAILABILITY Oxford University Press, 2004.
14
Code used in this study will be made available upon Gehron, J., J. Zirbes, M. Bongert, S. Schäfer, M. Fiebich,
G. Krombach, et al. Development and validation of a life-
request. sized mock circulatory loop of the human circulation for
fluid-mechanical studies. Asaio J. 65(8):788–797, 2019.
15
REFERENCES Glynn, J., H. Song, B. Hull, S. Withers, J. Gelow, J. Mudd,
et al. The oregonheart total artificial heart: design and
1
Beurton, A., J.-L. Teboul, F. Gavelli, F. A. Gonzalez, V. performance on a mock circulatory loop. Artif Organs.
Girotto, L. Galarza, et al. The effects of passive leg raising 41(10):904–910, 2017. https://doi.org/10.1111/aor.12959.
16
may be detected by the plethysmographic oxygen satura- Goldberger, A. L., L. A. Amaral, L. Glass, J. M. Haus-
tion signal in critically ill patients. Crit. Care. 23(1):19, dorff, P. C. Ivanov, R. G. Mark, et al. PhysioBank,
2019. PhysioToolkit, and PhysioNet: components of a new
2
Biglino, G., A. Giardini, C. Baker, R. S. Figliola, T.-Y. research resource for complex physiologic signals. Circu-
Hsia, A. M. Taylor, et al. In vitro study of the Norwood lation. 101(23):e215–e220, 2000.
17
palliation: a patient-specific mock circulatory system. Gregory, S. D., J. P. Pauls, E. L. Wu, A. Stephens, U.
ASAIO J. 58(1):25–31, 2012. https://doi.org/10.1097/MA Steinseifer, G. Tansley, et al. An advanced mock circula-
T.0b013e3182396847. tion loop for in vitro cardiovascular device evaluation.

BIOMEDICAL
ENGINEERING
SOCIETY
290 PACKY et al.
32
Artif Organs. 44(6):E238–E250, 2020. https://doi.org/10.1 Rzheutskaya, R.E. Characteristics of hemodynamic disor-
111/aor.13636. ders in patients with severe traumatic brain injury. Crit
18
Hang, T., A. Giardini, G. Biglino, T. Conover, and R. S. Care Res Pract. 2012;2012:606179. https://doi.org/10.1155/
Figliola. In vitro validation of a multiscale patient-specific 2012/606179.
33
Norwood Palliation model. Asaio J. 62(3):317–324, 2016. Shehab, S., S. M. Allida, P. J. Newton, D. Robson, P. S.
19
Jiang, X., S. R. Srinivasan, E. Urbina, and G. S. Berenson. Macdonald, P. M. Davidson, et al. Valvular regurgitation
Hyperdynamic circulation and cardiovascular risk in chil- in a biventricular mock circulatory loop. Asaio J.
dren and adolescents: the Bogalusa Heart Study. Circula- 65(6):551–557, 2019.
34
tion. 91(4):1101–1106, 1995. Sun, J. X., A. T. Reisner, M. Saeed, T. Heldt, and R. G.
20
Karamanoglu, M., M. O’rourke, A. Avolio, and R. Kelly. Mark. The cardiac output from blood pressure algorithms
An analysis of the relationship between central aortic and trial. Crit Care Med. 37(1):72–80, 2009. https://doi.org/10.
peripheral upper limb pressure waves in man. Eur. Heart J. 1097/CCM.0b013e3181930174.
35
14(2):160–167, 1993. Sun, J, A. Reisner, M. Saeed, and R. Mark. Estimating
21
Kim, S., J. Lee, W. Kim, J. Sun, M. Kwon, and H. Kil. cardiac output from arterial blood pressure waveforms: a
Evaluation of radial and ulnar blood flow after radial ar- critical evaluation using the MIMIC II database. Comput.
tery cannulation with 20-and 22-gauge cannulae using du- Cardiol. 295–298, IEEE, 2005. https://doi.org/10.1109/CI
plex Doppler ultrasound. Anaesthesia. 67(10):1138–1145, C.2005.1588095.
36
2012. Teboul, J.-L., M. Cecconi, and T. W. Scheeren. Is there still
22
Koenig, S. C., G. M. Pantalos, K. J. Gillars, D. L. Ewert, a place for the Swan-Ganz catheter? New York: Springer,
K. N. Litwak, and S. W. Etoch. Hemodynamic and pres- 2018.
37
sure–volume responses to continuous and pulsatile ven- Trusty, P. M., M. Tree, D. Vincent, J. P. Naber, K. Maher,
tricular assist in an adult mock circulation. ASAIO J. A. P. Yoganathan, et al. In vitro examination of the Ven-
50(1):15–24, 2004. triFlo true pulse pump for failing Fontan support. Artif
23
Kroeker, E. J., and E. H. Wood. Comparison of simulta- Organs. 43(2):181–188, 2019. https://doi.org/10.1111/aor.1
neously recorded central and peripheral arterial pressure 3301.
38
pulses during rest, exercise and tilted position in man. Circ. Vahdatpour, C., D. Collins, and S. Goldberg. Cardiogenic
Res. 3(6):623–632, 1955. shock. J. Am. Heart Assoc.8(8):e011991, 2019. https://doi.
24
Kung, E., M. Farahmand, and A. Gupta. A hybrid org/10.1161/JAHA.119.011991.
39
experimental-computational modeling framework for car- Villanueva, C., A. Albillos, J. Genescà, J. G. Abraldes, J. L.
diovascular device testing. J. Biomech. Eng. 141(5):051012, Calleja, C. Aracil, et al. Development of hyperdynamic
2019. circulation and response to b-blockers in compensated
25
Masengu, A., J. McDaid, A. P. Maxwell, and J. B. Hanko. cirrhosis with portal hypertension. Hepatology. 63(1):197–
Preoperative radial artery volume flow is predictive of 206, 2016.
40
arteriovenous fistula outcomes. J. Vasc. Surg. 63(2):429– Vincent, J.-L., A. Rhodes, A. Perel, G. S. Martin, G. Della
435, 2016. Rocca, B. Vallet, et al. Clinical review: update on hemo-
26
Meess, K., R. Izzo, M. Dryjski, R. Curl, L. Harris, M. dynamic monitoring-a consensus of 16. Crit. Care. 15(4):1–
Springer, et al. 3D Printed Abdominal Aortic Aneurysm 8, 2011.
41
Phantom for Image Guided Surgical Planning with a Pa- Wu, C., S. M. Retta, R. A. Robinson, B. A. Herman, and
tient Specific Fenestrated Endovascular Graft System. L. W. Grossman. A novel study of mechanical heart valve
SPIE Medical Imaging: SPIE, 2017. cavitation in a pressurized pulsatile duplicator. ASAIO J.
27
Pantalos, G. M., S. C. Koenig, K. J. Gillars, G. A. Girid- 55(5):445–451, 2009. https://doi.org/10.1097/MAT.0b013e
haran, and D. L. Ewert. Characterization of an adult mock 3181b4c44f.
42
circulation for testing cardiac support devices. ASAIO J. Yang, T.-H., J. U. Kim, Y.-M. Kim, J.-H. Koo, and S.-Y.
50(1):37–46, 2004. https://doi.org/10.1097/01.mat.0000104 Woo. A new blood pulsation simulator platform incorpo-
818.70726.e6. rating cardiovascular physiology for evaluating radial pulse
28
Persona, P., E. Saraceni, F. Facchin, E. Petranzan, M. waveform. J. Healthc. Eng. 2019;2019:4938063. https://doi.
Parotto, F. Baratto, et al. Pulse contour analysis of arterial org/10.1155/2019/4938063.
43
waveform in a high fidelity human patient simulator. J. Young, E., J.-F. Chen, O. Dong, S. Gao, A. Massiello, and
Clin. Monitor. Comput. 32(4):677–681, 2018. K. Fukamachi. Transcatheter heart valve with variable
29
Petrou, A., M. Granegger, M. Meboldt, and Daners M. geometric configuration. Vitro Evaluat. Artif Organs.
Schmid. A versatile hybrid mock circulation for hydraulic 35(12):1151–1159, 2011. https://doi.org/10.1111/j.1525-159
investigations of active and passive cardiovascular im- 4.2011.01331.x.
44
plants. ASAIO J. 65(5):495–502, 2019. https://doi.org/10. Zhang, Z., B. Lu, X. Sheng, and N. Jin. Accuracy of stroke
1097/mat.0000000000000851. volume variation in predicting fluid responsiveness: a sys-
30
Ramsingh, D., B. Alexander, and M. Cannesson. Clinical tematic review and meta-analysis. J. Anesth. 25(6):904–916,
review: does it matter which hemodynamic monitoring 2011.
system is used? Crit. Care. 17(2):208, 2013.
31
Robin, E., M. Costecalde, G. Lebuffe, and B. Vallet. Publisher’s Note Springer Nature remains neutral with re-
Clinical relevance of data from the pulmonary artery ca- gard to jurisdictional claims in published maps and institu-
theter. Crit. Care. 10(3):1–10, 2006.
tional affiliations.

BIOMEDICAL
ENGINEERING
SOCIETY

You might also like