Professional Documents
Culture Documents
Cardiovascular Engineering and Technology, Vol. 13, No. 2, April 2022 ( 2021) pp. 279–290 ENGINEERING
https://doi.org/10.1007/s13239-021-00575-2 SOCIETY
Original Article
device reproducibility and to compare diagnostic re- diac hemodynamics and peripheral pressure wave-
sults between devices in a clinical setting.30 forms. This allows us to set known changes in
A mock circulatory loop (MCL) is a useful tool for hemodynamics under a range of physiologically rele-
simulating cardiovascular hemodynamics on the vant conditions towards designing test methods for
bench. Different pathophysiologic conditions, repre- evaluating how monitoring devices can track those
senting a range of disease states and patient popula- changes. We characterized our MCL to assess the
tions, can be evaluated in a pre-clinical setting using an overall controllability across the desired functional
MCL. Academic researchers and device manufacturers range. This process includes: (1) characterizing the
have utilized MCLs for evaluating the hemodynamics system to mimic the hemodynamics over a range of
of implantable cardiovascular devices4,22 prior to clinically relevant cardiac states; (2) assessing the
conducting more expensive animal studies and clinical repeatability and stability of the pressure and flow
trials. Also, MCLs have been used for years to char- values, as well as its ability to resolve small changes in
acterize the fluid dynamics within therapeutic cardio- flow; and (3) qualitatively and quantitatively evaluat-
vascular devices, including mechanical circulatory ing the arterial pressure waveform characteristics for
support devices, heart valves, vascular conduits and three hemodynamic states. We then applied three pulse
grafts, and catheters.10–27 Further, the efficacy of using contour analysis algorithms for measuring CO to the
hemodynamic indices for diagnosing cardiovascular peripheral pressure waveforms to demonstrate the
diseases including vascular and valvular stenosis, an- potential for non-clinical characterization of hemody-
eurysms, and congenital heart defects has been evalu- namic monitors during known changes in flow and
ated using MCLs.17–2 Experimental studies using pressure using an MCL.
MCLs have led to the optimization of device designs
and the determination of appropriate performance
parameters over the full operating range of the device MATERIALS AND METHODS
in a well-controlled environment, which may not be
attainable through animal studies or clinical trials. Design of Mock Circulatory Loop
Thus, MCLs enable the evaluation of certain devices An MCL was first designed to create physiologic
earlier in their total product life cycle, leading to design pressure and flow waveforms. This system consists of
improvements and a better understanding of perfor- physical components that mimic the left ventricular
mance and limitations before more costly clinical trials. function, as well as the systemic and peripheral circu-
In recent years, MCLs have been modified to lations comprised of arteries and veins. A block dia-
incorporate peripheral vascular components for simu- gram of the MCL is shown in Fig. 1, and images of the
lating arterial pressure waveforms.42,14 Hemodynamic physical MCL are shown in the Supplementary
monitoring devices can be connected to these circuits Information as Figure S1. A description of the sys-
and algorithms can be applied28 to estimate invasive, temic loop is provided first followed by a description of
central hemodynamic parameters from measured the peripheral loop.
pressure signals that are representative of in vivo radial Deionized water is circulated through the MCL
artery pressure. Therefore, it is critical to fully char- using a piston pump [ViVitro Labs, Victoria, BC,
acterize peripheral pressure waveforms within an MCL Canada] that is fluid-coupled to a suspended silicone-
and then correlate them to the cardiac hemodynamics based mock ventricular sac. This causes the ventricular
to demonstrate the utility of using MCLs for evaluat- sac to expand and contract during each beat of the
ing hemodynamic monitoring devices. The focus of pump. The wall motion of the ventricular sac is con-
this work is to investigate if an MCL with peripheral trolled by the inputted piston waveform, beat rate, and
vascular components provides central-peripheral amplitude of piston motion. The ventricular sac mo-
pressure relationships across a range of hemodynamic tion causes the artificial (bi-leaflet mechanical) heart
states that can then be used for generating represen- valves [St. Jude Medical Regent Valve, St. Jude Med-
tative radial pressure waveforms for hemodynamic ical, Inc., St. Paul, MN], to open and close, which al-
monitoring device testing. A controllable MCL can lows fluid to be ejected unidirectionally from the left
help evaluate how well a medical device can monitor ventricle into the systemic part of the MCL during
changes in flow and pressure values. This can poten- systole. Source compliance, which simulates the left
tially expand the role of MCLs in medical device ventricular viscoelastic behavior, was controlled by
testing. manually adjusting the air volume in the source com-
We describe an MCL with peripheral vascular pliance chamber using a syringe. The baseline com-
components and provide a characterization of the pliance level was defined as complete unloading of the
central-peripheral pressure relationships. The MCL ventricular sac. Source compliance on the piston pump
offers many adjustable settings to achieve target car-
BIOMEDICAL
ENGINEERING
SOCIETY
Simulating Radial Pressure Waveforms 281
FIGURE 1. Block diagram of the mock cardiovascular flow circuit. Arrows represent the flow direction of the systemic loop (grey)
and the peripheral loop (black). The atrial, ventricular, and venous chambers contain water for circulation through the system. All
other block units can be modified to simulate different types of flows. *Pressure measurement location. °Flow measurement
location.
was set before the test by adding 60 mL of air to the lar resistance is simulated using a manually controlled
working fluid which surrounds the outside of the screw compressor tubing clamp. Repeatability with the
ventricular sac. Distal to the ventricle, fluid passes screw clamp between experiments and when re-setting
through the aortic root and ascending aorta, and then the system was achieved by approximating the distance
through tubing representative of the systemic circula- between the clamp jaws using a caliper. Although this
tion, where the flow rate and aortic pressure are technique of manual adjustment is not highly accurate
recorded. Two air-filled (passive) compliance chambers in terms of repeatability, the set point for the screw
are connected to the aortic root and ascending aorta to clamp was influenced primarily by the desired flow
simulate the compliance of the respective vessels. A split between the peripheral and systemic loops. Fluid
third air-filled (passive) compliance chamber is placed from the peripheral loop circulates back into the sys-
in-line in the mock descending aorta to simulate the temic loop after the venous reservoir. Since the MCL is
systemic vascular compliance. All the air-filled com- an open-loop system consisting of a venous reservoir
pliance chambers were manually controlled using a and atrial chamber open to atmosphere (Figs. 1 and
syringe for adjusting and setting the volume of air in S1), the transient pressure pulse in the proximal and
the chamber to simulate the compliance effect. distal regions of the venous reservoir are similar. Thus,
Mechanical resistance, simulating the systemic vascu- attaching the peripheral loop after the venous reservoir
lar resistance, is applied before the systemic loop re- in the MCL (instead of the anatomically accurate
turns to the large venous reservoir using an proximal position) is not expected to alter the patho-
electronically activated ball valve. The ball valve en- physiologic hemodynamics of the peripheral circula-
ables a set amount of mechanical resistance (measured tion. Supplementary Table S1 includes details of the
as the % occlusion of the tubing cross-section) to be MCL components including tubing dimensions and
applied using a specific voltage value (0.0–10.0 Volts), material properties.
thus allowing for easy repeatability between experi- Two flow sensors and four pressure transducers
ments and when re-setting the system. Fluid from the were used in the MCL to monitor hemodynamics at
large venous reservoir is then circulated back into the the locations identified in Fig. 1. Aortic and peripheral
left atrial chamber to complete the systemic loop. loop flow rates were measured simultaneously using
The peripheral loop branches off the systemic loop ultrasound Doppler clamp-on flow sensors [ME11PXL
at the mock descending aorta proximal to the systemic (aorta) and ME5PXL (peripheral), Transonic Systems,
compliance chamber. Peripheral vascular compliance Inc., Ithaca, NY] in the mock descending aorta and in
is simulated using anatomically representative tubing the representative radial artery, respectively. The two
durometers, diameters, and lengths. Peripheral vascu- flow sensors were coupled to a flow meter [T402,
BIOMEDICAL
ENGINEERING
SOCIETY
282 PACKY et al.
Transonic Systems, Inc., Ithaca, NY], which recorded Prior to testing, a verification step was performed to
and displayed the instantaneous flow rate (in mL/min ensure that the MCL could replicate a normal physi-
or L/min). Left ventricular and left atrial pressures ologic condition, with pressures of 120/80 mmHg and
were measured with disposable pressure transducers a CO of 5 L/min. Additionally, the fluid height levels
[Utah Medical Products Inc., Midvale, Utah] and sent were set for each compliance chamber, and any resid-
to an amplifier module [Bus 21097, ViVitro Labs, ual air removed from the flow loop. To set each
Victoria, BC, Canada]. The aortic and peripheral hemodynamic state, the resistances, beat rate, piston
pressures were measured using invasive pressure ca- waveform, and gain were first adjusted to produce the
theters [SPR-350, Millar Instruments, Houston, TX] nominal flow setting for that state. Then, MAP was set
coupled to an amplifier [PCU-2000 Pressure Control to the desired level by adjusting the systemic resistance
Unit, Millar Instruments, Houston, TX]. The periph- produced by the ball valve. Next, the gain on the pis-
eral pressure measurement was made 0.83 m distal to ton pump was manually adjusted to achieve the target
the aorta. All pressure and flow signals were acquired CO. For each hemodynamic state, all measurements
using a data acquisition system [NI-9205, National were recorded at the nominal flow for 60 seconds,
Instruments, Austin, TX] at a sampling rate of 1000 followed by data acquisition over a range of flow rates.
Hz, and an in-house software control program was Specifically, COs of ± 1 L/min of the nominal nor-
developed to modify, display, and analyze the hemo- movolemic and cardiogenic shock conditions were
dynamic signals. produced with steps of ± 0.2 L/min, and of ± 2 L/min
of the hyperdynamic state with steps of ± 0.4 L/min,
as shown in Fig. 2, by adjusting the gain of the piston
Experimental Design
pump.
The MCL was used to simulate three different
physiologic hemodynamic states for demonstrating the
Data Analysis
versatility of the system: normovolemia, cardiogenic
shock, and hyperdynamic. The representative hemo- Pressure and flow data were analyzed using MA-
dynamic states were based on average physiologic TLAB [MathWorks, Natick, MA]. A low-pass filter
ranges found in the literature.40,23 The nominal nor- with a cut-off frequency of 50 Hz was applied to the
movolemic state was defined here to have a beat rate of pressure and flow waveforms, which were subsequently
75 beats per minute (BPM), CO of 5 L/min, and mean down-sampled to 128 Hz. Individual beat cycles were
arterial pressure (MAP; defined as the time-averaged detected using a previously reported pressure pulse
aortic pressure) between 90 and 100 mmHg. The flow beat detector.35,16 All measurements were calculated
rate was varied by ± 1 L/min from the nominal flow for each beat cycle and averaged across the 60-second
for the normovolemic state. For the cardiogenic shock acquisition period. Stroke volume (SV) from the ven-
state, beat rate was set to 90 BPM, CO to 3 ± 1 L/min, tricular sac was calculated using the average flow of a
and MAP was less than 70 mmHg.38 For the hyper- 60-second acquisition period divided by the beat rate.
dynamic state, beat rate was set to 120 BPM, CO to Systemic vascular resistance (SVR) was calculated as
8 ± 2 L/min, and MAP was between 65 and 70
mmHg.28–32
The system variables, including the peripheral and
systemic resistances, beat rate, and the gain applied to
the piston waveform, were adjusted to generate the
three hemodynamic states. The systemic vascular
resistance, set using the ball valve, was empirically
derived for each hemodynamic state. The peripheral
vascular resistance was set to achieve a peripheral loop
flow rate equating to approximately 3% of the total
CO to achieve a consistent target for adjusting
peripheral resistance in each state, based on an initial
baseline position of the screw clamp (i.e., distance
between clamp jaws) used in the peripheral loop. The FIGURE 2. The flow rate was manipulated for each
hemodynamic state across a range of 61 L min21 from the
clamp position was kept constant as the flow rate was nominal flow rate (62 L min21 for the hyperdynamic state) in
varied for each hemodynamic state. This provided 0.2 L min21 increments (0.4 L min21 for the hyperdynamic
peripheral flow rates during the normovolemic state state). Each measurement step was captured for 60 seconds.
The first three steps were performed at the nominal, low, then
from ~120 to 170 mL/min, which is on the order of high flow rates done for each state. The total number of
measured radial artery volume flows.25,21 completed steps for each state was 25.
BIOMEDICAL
ENGINEERING
SOCIETY
Simulating Radial Pressure Waveforms 283
the mean pressure drop from the aorta to the left at- namic state. The difference between the overall mini-
rium divided by CO. mum and maximum (range) values of mean aortic and
Stability of the pressure and flow values over time peripheral pressures across experiments were 76 and 74
and repeatability of the MCL were determined for each mmHg, respectively. The average SVR range was 1012
hemodynamic state. Stability was assessed by fitting a dyn s cm-5. The overall mean aortic flow rate range
linear line to the recorded pressure and flow values across experiments (difference between maximum and
over an acquisition period using the nominal level for minimum flow rate achieved) was 7.9 L/min and the
each hemodynamic state and assessing the slope. We mean peripheral flow rate range was 0.2 L/min. The
used the data from the 5 nominal positions for each gain settings were set between 19.7% and 59.4% to
hemodynamic state (see Fig. 2) to evaluate the achieve the desired flow rates for the three hemody-
repeatability by determining the variation in the pump namic states, and the ball valve position was set at
gain required to achieve the same flow rates and 62%, 64%, and 46% occlusion for the normovolemic,
pressures across the 5 replicate trials. cardiogenic shock, and hyperdynamic states, respec-
Discrete Fourier transforms were calculated for the tively. The peripheral flow was approximately 3% of
pressure signals following linear detrending to identify the CO for each hemodynamic state (Table 1), and it
the harmonics present in the generated waveforms. The did not vary substantially across the flow rate steps
relationship between aortic and peripheral pressure within each hemodynamic state (0.4, 0.5, and 0.4%
waveforms was evaluated by estimating the transfer ratio differences for the normovolemic, cardiogenic
function using a Hamming window (length of 1024 shock, and hyperdynamic states, respectively). The
samples with 512 samples overlapping). Transfer large achievable flow rate range demonstrates the
function data are reported for each hemodynamic state ability to simulate many physiologic conditions with
as mean ± standard deviation across each 60-second this MCL.
acquisition period. Flow rate was adjusted through a 22-step protocol
of 0.2 L/min increments for each hemodynamic state
(0.4 L/min for the hyperdynamic state). The mean and
Characterizing Hemodynamic Monitoring Algorithms
standard deviation of the CO at each step are shown in
Three pulse contour analysis algorithms were Fig. 4. For cardiogenic shock, we observed increased
applied to the radial pressure waveforms recorded for variability in the measured flow within each one-min-
each hemodynamic state: Liljestrand-Zander (CO = ute step as the flow rate was increased (Fig. 4d). The
k * (SBP DBP)/(SBP + DBP) * HR), Pulse Pres- average coefficient of variation across steps for both
sure (CO = k *R (SBP DBP) * HR), and Systolic the hyperdynamic and cardiogenic shock states was
Area (CO ¼ k ABPðtÞdt HR), where SBP and 2.2%, as a percentage of the mean value. The nor-
DBP refer to systolic and diastolic blood pressure, HR movolemic state, which was the middle condition in
to heart rate, and ABP to arterial blood pressure terms of CO, showed a lower average coefficient of
waveform.35 Each method requires a calibration value variation of 1.3%. The hyperdynamic state had a
‘k’. This was determined for each of the three states higher standard deviation in beat-to-beat CO values at
from the initial settings by dividing the uncalibrated each step compared to the normovolemic and cardio-
value from each algorithm by the reference flow rate, genic shock states (Fig. 4d). Overall, the maximum
and then the algorithms were applied to measure CO at coefficient of variation observed across any step for
each subsequent step in the experimental protocol de- any test condition was 5.4%, which occurred at the
scribed above and compared to the reference systemic highest flow rate of the cardiogenic shock state.
flow measurement.
Stability of Pressure and Flow Values and Repeatability
RESULTS The normovolemic and hyperdynamic states
showed a slight decrease in MAP of 1.8 mmHg and 0.6
Hemodynamic States mmHg, respectively, over the 60-second period. This
Each hemodynamic state was studied over a range decrease is small in comparison to the absolute values
of flow rates, pressures, and beat rates, Table 1. For of the MAP (~94 and 60 mmHg, respectively, Tables 1
each hemodynamic state, controllable changes to flow and 2) for those specific states. The MAP of the car-
and pressure were produced by adjusting the piston diogenic shock state was stable over the test duration.
gain, as shown in Figure 3 for the normovolemic state. Mean flow rates were stable for each condition over
Table 1 shows the full range of the system using the the 60-second duration.
lowest, nominal and highest settings of each hemody- The repeatability of the MCL at the different
hemodynamic states is characterized in Table 2. The
BIOMEDICAL
ENGINEERING
SOCIETY
284 PACKY et al.
HR heart rate, CO cardiac output, PF peripheral flow, PTF percent of total flow through peripheral loop, SV stroke volume, ABP aortic blood
pressure, PP pulse pressure, PBP peripheral blood pressure, AtP atrial blood pressure, VBP ventricular blood pressure, SVR systemic
vascular resistance.
Waveform Assessment
The systemic and peripheral pressure waveforms
resembled physiologic arterial waveforms, which in-
clude characteristics such as a steep rise during systole
FIGURE 3. 10 second segment traces for the peripheral (a) and a more prominent dicrotic notch in the peripheral
pressure and (b) flow waveforms at the nominal, low, and high
flow settings for the normovolemic state. Each setting was waveform (Fig. 5). The dicrotic notch is more evident
run for 60 seconds and the dashed vertical lines indicate that in the peripheral pressure waveforms in comparison to
the waveforms were truncated to only display the first 10 the systemic pressure waveforms, especially for the
seconds for each setting.
cardiogenic shock and hyperdynamic states. The pulse
standard deviation of the MAP across the 5 repeated morphology remains similar at each flow rate step for
trials when flow was set at the nominal value (see each state, seen in both Figs. 3 and 5. Figure 6 shows
Fig. 2) was within 1 mmHg for each state, and the the Fourier transform of the peripheral pressure
BIOMEDICAL
ENGINEERING
SOCIETY
Simulating Radial Pressure Waveforms 285
FIGURE 4. Aortic flow rates at each step for the (a) normovolemic (b) cardiogenic shock, and (c) hyperdynamic states. (d) The
mean cardiac output (CO) compared to the standard deviation of the cardiac output in L/min for each hemodynamic state. NV
normovolemic, CS cardiogenic shock, HD hyperdynamic.
DISCUSSION
waveforms at the initial nominal step for each hemo- Hemodynamic monitoring devices can continuously
dynamic state. Information content decays at higher measure critical cardiovascular parameters by analyz-
frequencies, with a modulus of < 1 mmHg above 7 ing the peripheral pressure waveforms. However, it can
Hz. be challenging to collect adequate reference measure-
The systemic-peripheral transfer functions show ments for repeatably and precisely evaluating these
increasing amplification at the peripheral location parameters. It is also difficult to understand how dif-
from 0 to approximately 5 Hz, after which the ampli- ferent confounding factors affect the performance of
fication decreases (Fig. 7a). The amplitude varied hemodynamic monitoring devices. Non-clinical test
across hemodynamic states with the cardiogenic shock methods can potentially provide repeatable conditions
state exhibiting the highest amplitude and the hyper- for assessing safety and performance aspects of
dynamic state having the lowest amplitude at most hemodynamic monitoring devices that are difficult to
frequencies. The phase delay showed less variability determine and isolate through clinical studies. MCLs
(Fig. 7b) compared to reported aortic-radial pressure provide potential means for augmenting or reducing
transfer functions, which showed phase delays up to -8 the amount of invasive and expensive animal and
radians by 10 Hz.20 human testing that is often required for medical device
development. The focus of MCL development has
historically been for assessing cardiac devices such as
Application of Cardiac Output Algorithms
ventricular assist devices and heart valves27,37–11;
After calibration at the baseline step, the pulse therefore, the modules necessary for modeling the ra-
contour analysis algorithms generally tracked CO dial arterial pressure waveforms are not included in
BIOMEDICAL
ENGINEERING
SOCIETY
286 PACKY et al.
FIGURE 5. Pressure waveforms for the hemodynamic states. The nominal waveform for each state is a solid black line, the low is
a dashed-dotted line, and the high is a dashed line. (a) Aortic pressure at normovolemic state, (b) peripheral pressure at
normovolemic state, (c) aortic pressure at cardiogenic shock state, (d) peripheral pressure at cardiogenic shock state, (e) aortic
pressure at hyperdynamic state, and (f) peripheral pressure at hyperdynamic state.
this generation of MCLs. Recently, researchers devel- that the flow loop can produce a wide range of phys-
oped an MCL platform capable of generating radial iologic conditions to simulate different clinical sce-
arterial waveforms.42 However, the ability of MCLs to narios.
simulate radial arterial pressure remains limited. The Pulse contour analysis uses a peripherally recorded
challenges include difficulties in achieving a realistic arterial pressure waveform to continuously estimate
transfer function between systemic arterial pressure hemodynamic parameters such as CO in real-time.
and radial pressure under various hemodynamic states Advantages of these devices include their ability to
as well as achieving a framework capable of producing identify and monitor changes in CO that are indicative
a broad range of flow rates under different hemody- of how a patient responds to therapy, or that estimate
namic conditions for assessing the performance of responsiveness such as through dynamic indices
hemodynamic monitoring systems. In this work, we including stroke volume variation.44 These applica-
adapted an MCL that was previously used only for tions measure small changes in CO over short time
cardiac device testing (e.g., ventricular assist devices, scales compared to traditional reference measure-
heart valves) to include peripheral components. The ments. In this respect, intermittent reference measure-
adapted MCL was used to generate and acquire ments can offer only limited information about the
peripheral pressure waveforms to assess hemodynamic time-varying hemodynamic behavior. Non-clinical
monitoring devices under known conditions and set testing tools such as MCLs have the potential to allow
changes to those conditions, something that cannot be us to better understand properties of monitoring sys-
feasibly performed in clinical studies. Our results show tems such as the response to a known change of CO
BIOMEDICAL
ENGINEERING
SOCIETY
Simulating Radial Pressure Waveforms 287
BIOMEDICAL
ENGINEERING
SOCIETY
290 PACKY et al.
32
Artif Organs. 44(6):E238–E250, 2020. https://doi.org/10.1 Rzheutskaya, R.E. Characteristics of hemodynamic disor-
111/aor.13636. ders in patients with severe traumatic brain injury. Crit
18
Hang, T., A. Giardini, G. Biglino, T. Conover, and R. S. Care Res Pract. 2012;2012:606179. https://doi.org/10.1155/
Figliola. In vitro validation of a multiscale patient-specific 2012/606179.
33
Norwood Palliation model. Asaio J. 62(3):317–324, 2016. Shehab, S., S. M. Allida, P. J. Newton, D. Robson, P. S.
19
Jiang, X., S. R. Srinivasan, E. Urbina, and G. S. Berenson. Macdonald, P. M. Davidson, et al. Valvular regurgitation
Hyperdynamic circulation and cardiovascular risk in chil- in a biventricular mock circulatory loop. Asaio J.
dren and adolescents: the Bogalusa Heart Study. Circula- 65(6):551–557, 2019.
34
tion. 91(4):1101–1106, 1995. Sun, J. X., A. T. Reisner, M. Saeed, T. Heldt, and R. G.
20
Karamanoglu, M., M. O’rourke, A. Avolio, and R. Kelly. Mark. The cardiac output from blood pressure algorithms
An analysis of the relationship between central aortic and trial. Crit Care Med. 37(1):72–80, 2009. https://doi.org/10.
peripheral upper limb pressure waves in man. Eur. Heart J. 1097/CCM.0b013e3181930174.
35
14(2):160–167, 1993. Sun, J, A. Reisner, M. Saeed, and R. Mark. Estimating
21
Kim, S., J. Lee, W. Kim, J. Sun, M. Kwon, and H. Kil. cardiac output from arterial blood pressure waveforms: a
Evaluation of radial and ulnar blood flow after radial ar- critical evaluation using the MIMIC II database. Comput.
tery cannulation with 20-and 22-gauge cannulae using du- Cardiol. 295–298, IEEE, 2005. https://doi.org/10.1109/CI
plex Doppler ultrasound. Anaesthesia. 67(10):1138–1145, C.2005.1588095.
36
2012. Teboul, J.-L., M. Cecconi, and T. W. Scheeren. Is there still
22
Koenig, S. C., G. M. Pantalos, K. J. Gillars, D. L. Ewert, a place for the Swan-Ganz catheter? New York: Springer,
K. N. Litwak, and S. W. Etoch. Hemodynamic and pres- 2018.
37
sure–volume responses to continuous and pulsatile ven- Trusty, P. M., M. Tree, D. Vincent, J. P. Naber, K. Maher,
tricular assist in an adult mock circulation. ASAIO J. A. P. Yoganathan, et al. In vitro examination of the Ven-
50(1):15–24, 2004. triFlo true pulse pump for failing Fontan support. Artif
23
Kroeker, E. J., and E. H. Wood. Comparison of simulta- Organs. 43(2):181–188, 2019. https://doi.org/10.1111/aor.1
neously recorded central and peripheral arterial pressure 3301.
38
pulses during rest, exercise and tilted position in man. Circ. Vahdatpour, C., D. Collins, and S. Goldberg. Cardiogenic
Res. 3(6):623–632, 1955. shock. J. Am. Heart Assoc.8(8):e011991, 2019. https://doi.
24
Kung, E., M. Farahmand, and A. Gupta. A hybrid org/10.1161/JAHA.119.011991.
39
experimental-computational modeling framework for car- Villanueva, C., A. Albillos, J. Genescà, J. G. Abraldes, J. L.
diovascular device testing. J. Biomech. Eng. 141(5):051012, Calleja, C. Aracil, et al. Development of hyperdynamic
2019. circulation and response to b-blockers in compensated
25
Masengu, A., J. McDaid, A. P. Maxwell, and J. B. Hanko. cirrhosis with portal hypertension. Hepatology. 63(1):197–
Preoperative radial artery volume flow is predictive of 206, 2016.
40
arteriovenous fistula outcomes. J. Vasc. Surg. 63(2):429– Vincent, J.-L., A. Rhodes, A. Perel, G. S. Martin, G. Della
435, 2016. Rocca, B. Vallet, et al. Clinical review: update on hemo-
26
Meess, K., R. Izzo, M. Dryjski, R. Curl, L. Harris, M. dynamic monitoring-a consensus of 16. Crit. Care. 15(4):1–
Springer, et al. 3D Printed Abdominal Aortic Aneurysm 8, 2011.
41
Phantom for Image Guided Surgical Planning with a Pa- Wu, C., S. M. Retta, R. A. Robinson, B. A. Herman, and
tient Specific Fenestrated Endovascular Graft System. L. W. Grossman. A novel study of mechanical heart valve
SPIE Medical Imaging: SPIE, 2017. cavitation in a pressurized pulsatile duplicator. ASAIO J.
27
Pantalos, G. M., S. C. Koenig, K. J. Gillars, G. A. Girid- 55(5):445–451, 2009. https://doi.org/10.1097/MAT.0b013e
haran, and D. L. Ewert. Characterization of an adult mock 3181b4c44f.
42
circulation for testing cardiac support devices. ASAIO J. Yang, T.-H., J. U. Kim, Y.-M. Kim, J.-H. Koo, and S.-Y.
50(1):37–46, 2004. https://doi.org/10.1097/01.mat.0000104 Woo. A new blood pulsation simulator platform incorpo-
818.70726.e6. rating cardiovascular physiology for evaluating radial pulse
28
Persona, P., E. Saraceni, F. Facchin, E. Petranzan, M. waveform. J. Healthc. Eng. 2019;2019:4938063. https://doi.
Parotto, F. Baratto, et al. Pulse contour analysis of arterial org/10.1155/2019/4938063.
43
waveform in a high fidelity human patient simulator. J. Young, E., J.-F. Chen, O. Dong, S. Gao, A. Massiello, and
Clin. Monitor. Comput. 32(4):677–681, 2018. K. Fukamachi. Transcatheter heart valve with variable
29
Petrou, A., M. Granegger, M. Meboldt, and Daners M. geometric configuration. Vitro Evaluat. Artif Organs.
Schmid. A versatile hybrid mock circulation for hydraulic 35(12):1151–1159, 2011. https://doi.org/10.1111/j.1525-159
investigations of active and passive cardiovascular im- 4.2011.01331.x.
44
plants. ASAIO J. 65(5):495–502, 2019. https://doi.org/10. Zhang, Z., B. Lu, X. Sheng, and N. Jin. Accuracy of stroke
1097/mat.0000000000000851. volume variation in predicting fluid responsiveness: a sys-
30
Ramsingh, D., B. Alexander, and M. Cannesson. Clinical tematic review and meta-analysis. J. Anesth. 25(6):904–916,
review: does it matter which hemodynamic monitoring 2011.
system is used? Crit. Care. 17(2):208, 2013.
31
Robin, E., M. Costecalde, G. Lebuffe, and B. Vallet. Publisher’s Note Springer Nature remains neutral with re-
Clinical relevance of data from the pulmonary artery ca- gard to jurisdictional claims in published maps and institu-
theter. Crit. Care. 10(3):1–10, 2006.
tional affiliations.
BIOMEDICAL
ENGINEERING
SOCIETY