Systec store Approved Weight with Alibi memory

In this document we’ll use a Systec IT3 indicator, but it is also possible with other types, eg. IT1, IT6000E, etc.

The IT1/3 needs an additional module (alibi memory) for storage => ASM

Interface can be Serial or Ethernet

When ethernetis used, the indicator will need a fixed IP-adres and communicates via port 1234 by default, this can
be changed.

To perform a certified stable weighing the <RN> command needs to be send tot he <IP-adres>:1234

When the weight is STABLE, you’ll get a string like this:


<000005.06.1813:12 21 20000 0 20000kg 1 27387>

In your software you’ll have to store this info as it is stored in de ITx: date, time, weight, Ident No.

Negative weights cannot be executed!

See also IT3 technical manual chapter 20

Example with Hercules tool


Command (IT3 THE 20.03)

Field Byte Char's Description Example


RN 1 2 Command RN
Scale No. 2 1 Optional (see notes on operation with 2 scales) 1

In total: 3
Response

Field Byte Char's Description Example


Error code 1 2 00 = no error, see table error codes 00
Scale status 3 2 First digit: always 0 (scale settled) Second 00
digit: always 0 (gross positive)
Date 5 8 Date (formatted as per configuration) 05.06.18
Time 13 5 Time (formatted HH:MM) 13:12
Ident No. 18 4* Id-No. for data archive 2
(without ASM module always '0') Starts everyday at
(non-significant digits are transmitted as space 1 !!!
character)
Scale No. 22 1 See note on operation with 1 or 2 scales 1
Gross weight 23 8 Formatted as per configuration (non- significant digits 20000
are transmitted as space character)
Tare weight 31 8 Formatted as per configuration 0
(non-significant digits are transmitted as space
character)
Net weight 39 8 Formatted as per configuration (non- significant digits 20000
are transmitted as space character)
Unit 47 2 kg, g , t or lb, for g and t: kg
_ = second character is space
Tare code 49 2 PT = Preset Tare; PT
_T = Auto-tare
= scale not tared (_ = space character)
Weighing range 51 1 Weighing range, only for multiple range scales, 1
otherwise space character
Terminal No. 52 3 As entered in Service Mode, group 'Network' 001
Check sum 55 8 CRC16 checksum (non-significant digits are 27387
transmitted as space character)

Total: 62

18.2 Weight Storage (Manual IT3)


Note: The use of the long-term weight storage (data archive) requires the installation of an ASM module.
The W&M approved long-term weight storage has a capacity of approx. 1,000,000 records. A record is stored for every
completed weighing cycle. The sequence of a weighing transaction is: operating / data entry, entry in data archive,
printing and data transmission.
In the archive each record is stored with date, ident No. and gross and net weight. The Id No. is reset to 0001 with every
change of the date. To allow for a later verification of the weighing data, date and identification No. of the weighing have to
appear on the printout or must be stored together with the weight on the host computer.
The data archive can be used as an alternative to a log printer when data are processed in an EDP system. The stored
weights are read-only and cannot be deleted or changed.

Weight Storage

With printer = Y(es):


Select  Display: Show stored weights.
 Print: Print stored weights.

Memory: Display

18.2.1 Display Of Stored Weights

Enter date of weighing.

Enter ID of record to be looked up. Display of:


 Scale
 Date
 ID
 Net weight (N)
 Tare weight

Search Date 99.99.99

ID-No. 9999999999

W1 99.99.99 ID 99
9.9 kg N 9.9 kg

Previous Show previous record.

Next Show next record.

Return to step 'Search date'.

A matching record could no be found in the data archive:

Not found

Checksum error detected:


An error wasChecksum!
Error detected in the checksum of the data archive. Important note: The stored data are void!