You are on page 1of 60

HTI37103 Radiation Protection

Lecture 13

Revision (Lecture 6-11)

Dr. Marco TAM


The ICRP Recommendations
➢ ICRP publication 103 – 2007
➢ The recommended system of radiation
protection is based upon 3 principles:
◆ Benefit of a practice must offset the
radiation detriment (Net positive
benefit)
◆ Exposures and likelihood of exposure
should be kept as low as reasonably
achievable, economic and social factors
being taken into account
◆ Dose limits should be set to ensure that no
individual faces an unacceptable risk in
normal circumstances

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Justification

A. Benefit vs Risk
B. Diagnostic efficacy
(Usually justified by referring doctor)

Presence or
Imaging procedure
Minimal radiation Optimal image(s) absence of disease
justified by
exposure produced revealed =
referring doctor
Diagnostic efficacy

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Optimization
➢ Ensure that the appropriate equipment be used

➢ The medical practitioner, the technologist or


other imaging staff select the relevant
combination of parameters, that results in:
– minimum patient exposure consistent with
acceptable image quality and the clinical purpose
of the examination (ALARA principle),
– paying particular attention to this selection for
paediatric radiology and interventional radiology

4
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Diagnostic Reference value can be used…
➢ To compare our practice (the
level of radiological risk) with
other centers.
➢ To realize if we have a
certain margin for
optimization (to improve
settings of our X-ray system
or to improve our protocols).
➢ To detect abnormal
situations with high
radiological risk for our
patients.

5
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Limits and Constraints
➢ Dose limits are one of the three principles of
protection as introduced by ICRP and BSS.
Fixed dose limits are recommended by ICRP
and often enforced by a national legal process
(Radiation Protection Legislation).

➢ Dose constraints are used in an optimization


process to guide planning. Constraints and
the importance thereof may be subject to
change to achieve the optimum solution to a
problem (Best practice guidelines).

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Dose Limits

• No dose limits for medical exposure of


the patient - it is always assumed
that the benefits for the patient
outweigh the risks
• Limits need to be applied for public
and occupational exposures.

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Radiation Protection in Hong Kong

Legislative control
– The Radiation Ordinance Chapter 303 (Hong Kong Law)

Operational guidance
– Code of Practice on Radiation Safety & Protection in HA Hospitals
(HA’s Code of Practice)

– Local Radiological Protection Rules (e.g.


Departmental Local Rules)

8
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Radiation Protection in General Radiography

• Minimizing the number of


projection, if possible.
• Avoid repeating
• Avoid artifacts
• Proper collimation
• Position in or near the center
of the image receptor
• Check for the availability of
recent radiographs/ images of
the patient taken at the same
/ different hospitals
• Periodic check DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Radiation Protection in General & Fluoroscopic Examinations

➢ In general radiography and fluoroscopy with X-


ray mobile equipment, the source-to-skin
distance should not be less than 30 cm
➢ In radiography and fluoroscopy with fixed
equipment, the skin-focus distance should not
be lower than 45 cm
➢ If less than 15 cm between skin and collimator,
possible skin exposure to electrons produced
by photon interaction with the collimator
➢ Fluoroscopy equipment without image
intensifiers must be replaced, or upgraded
with an image intensifier (1.5 – 2 mA vs 3 –
5 mA)
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
X-ray tubes
➢Standards: International Electrotechnical Commission
(IEC) and the ISO or to equivalent national standards
➢Adequate filtration (minimum 2.5 mm Al (for >70 kVp)
/ 1.5 mm Al (50-70 kVp) / 0.5 mm Al (<50 kVp) in
general radiology, minimum 2.5 mm Al for fluoroscopy
(>3mm Al with image intensification systems)),
significantly reduces the patient dose due to low
energy X-rays which do not contribute to the image
formation

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Technical Adaptations for RP in General Radiography

➢It is advisable to use the highest


kVp (and the lowest mAs)
compatible with the image that
one expects to obtain. In this way
the patient dose will be minimized.
Optimization should be used to
find the proper balance between
contrast and dose
➢Short exposure times should be
used when imaging non-
cooperative patients (e.g.
Paediatric cases).
17
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Technical Adaptations for RP in General
Radiography
➢ Either 10-day rule or 28-day rule
should be applied to female patients of
reproductive age
➢ 10-day rule: Restrict X-ray examination
in the first 10 days of the menstrual
cycle, when conceptions is unlikely to
have occurred.
➢ 28-day rule: Ensure last menstrual
period within the past 28 days.
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Proper Radiological Positioning
➢ Maximize distance between x-ray tube
& patient.
➢ Minimize distance between patient &
Image Intensifier.
➢ Stand on side of the Image Intensifier

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Protective Barriers in Fluoroscopy

Primary barrier – protection


from Primary radiation
(for the patient & staff)
➢ It is a 2.0 mm Pb eq barrier
Secondary barrier – protection
from Secondary radiation
(Scattering)
(for the staff)
➢ Protection for the
technologists, etc.

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Dose Reduction in Fluoroscopy

➢ Reduce field size (collimate)


➢ Minimize field overlap.
➢ Use low pulsed fluoroscopy
➢ (7 or 3/sec)
➢ Use low frame rate
(4 or 2 or 1/sec)
➢ Avoid unnecessary runs
➢ Last image hold
➢ Screen capture

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Computed Tomography Dose Index
CT dose index (CTDI) is a standardized measure of radiation dose
output of a CT scanner which allows the user to compare radiation
output of different CT scanners. In the past, CTDI100 (measured over
a 100 mm long ionization chamber) and CTDIw (weighted average
of dose across a single slice) were used; for helical scanners in
current use, the parameter CTDIvol the most relevant one.

CTDI100 is a linear measure of dose distribution over a pencil


ionization chamber and hence does not take into consideration the
topographical variation of a human body and is therefore not in
clinical use.

CTDIw is closer to the human dose profile as compared with the


CTDI100 . It is simply:
2/3 CTDI100 (edge/ periphery) + 1/3 CTDI100 (center)

CTDIvol is obtained by dividing CTDIw by pitch factor.


DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Computed Tomography Dose Index
Volume Computed Tomography Dose Index (CTDIvol) is a
standardized parameter to measure Scanner Radiation
Output
➢ CTDIvol ≠ patient dose
➢ CTDIvol is reported in units of mGy for either a 16-cm
(for head exams) or 32-cm (for body exams) diameter
acrylic phantom
➢ For the same technique settings, the CTDIvol reported
for the 16-cm phantom is about twice that of the 32-
cm phantom
➢ The reported CTDIvol is based on measurements
made by the manufacturer in a factory setting
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Pitch

Pitch < 1 Pitch = 1 Pitch > 1


Beam Width has No overlap of Beam Some view angles are
some overlap at Width at each view angle not covered by the beam
each view angle and no view angles not width at certain table
from rotation to covered at certain table positions
rotation positions

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Acquisition Parameter Settings
Parameter Relationship to CTDIvol

Scan Mode Changes in the Scan Mode may affect CTDIvol

Exposure Time Per CTDIvol relationship to exposure time per rotation is vendor
Rotation dependent
Tube Current CTDIvol  Tube Current

Tube Potential CTDIvol  (kVp1/kVp2)n n ~ 2 to 3

Tube Current Time CTDIvol  Tube Current Time Product


Product
Effective Tube Current CTDIvol  Effective Tube Current Time Product
Time Product
Field of Measurement Changes in the Field of Measurement may affect CTDIvol

Beam Shaping Filter Changes in the Beam Shaping Filter may affect CTDIvol

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Acquisition Parameter Settings - Scan Mode
➢ In the Dynamic Scan Mode multiple acquisitions
covering the same body region are acquired.
Examples of these study types include:
◆ Perfusion Studies
◆ Bolus Tracking Studies
◆ Test Bolus Studies
➢ Dynamic Scans often have large CTDIvol values
because the scanner reports the sum of the CTDIvol
values from each rotation
➢ The reported CTDIvol is NOT skin dose or organ
dose

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Automatic Exposure Control
➢ Automatically adapts the Tube Current or Tube Potential
according to patient attenuation to achieve a specified
image quality
◆ Automatic adjustment of Tube Current may not occur when
Tube Potential is changed
◆ Centering the patient in the centre of the gantry is VITAL for
most AEC systems
➢ AEC aims to deliver a specified image quality across a
range of patient sizes. It tends to increase CTDIvol for
large patients and decrease it for small patients relative
to a reference patient size
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Image Quality Reference Parameter
➢ Changing the Image
Quality Reference
Parameter will affect the
CTDIvol
➢ Setting the IQR parameter
for “increased” image
quality (e.g. lower noise)
will result in more dose
➢ Setting the IQR parameter
for “decreased” image
quality (e.g., more noise)
will result in less dose

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Effective Dose of CT
➢ CT is one of the highest dose x-ray modalities used in
the hospital
◆ Head – 2 mSv
◆ Chest – 8 mSv
◆ Abdomen and pelvis – 10 mSv
◆ CT KUB – 10-15 mSv
➢ Planar chest – 0.08 mSv
➢ CT chest is 100 times the dose of planar chest
➢ Planar IVU has effective dose 1.5 – 3 mSv
➢ CT KUB is up to 10 times the dose of planar IVU

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Summary: 10 guidelines in CT

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Summary: 10 guidelines in CT

Lower Noise = Higher Patient Dose

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Summary: 10 guidelines in CT

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Summary: 10 guidelines in CT

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Summary: 10 guidelines in CT
56

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Approximate radiation dose to adults from
diagnostic nuclear medicine procedures

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Typical Effective Dose Values for Common PET/CT investigations
(Contributor: PET procedures)

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Typical Effective Dose Values for Common SPECT/CT investigations
(Contributor: SPECT procedures)

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Preparation and dispensation of radiopharmaceuticals

➢ Shields
➢ Protective clothing
➢ Tools for remote
handling of radioactive
material
➢ Containers for radioactive
waste
➢ Dose rate monitor with
alarm
➢ Contamination monitor
➢ Decontamination kit
➢ Signs, labels and records

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Administration of Radiopharmaceuticals

➢ In certain circumstances staff may need to


wear a protective lead apron.
➢ This may be necessary if staff need to be
in close contact with patients containing
greater than 800 MBq of 99mTc.
➢ Protective aprons should preferably have
a thickness of 0.5 mm lead equivalent,
comprising a separate vest and skirt
that wrap around fully, as open back
designs are not recommended.
➢ All protective clothing should be examined
at least annually.

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Administration of Radiopharmaceuticals

➢ Lead aprons provide little or no protection (i.e.


too thin) for higher energy photons and should not
be used for radionuclides such as gallium-67 or
iodine-131 (606 keV beta + 364 keV gamma) or
positron emitters (e.g. C-11, N-13, O-15, F-18).
➢ Staff leaving designated areas should remove
protective clothing, wash their hands and monitor
their hands, clothing and body as appropriate.
➢ Mobile shielding barriers may be required for
therapeutic nuclear medicine procedures using
gamma-emitting radionuclides.
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Radiation Protection issues in NM
General principles
➢ When considering the justification for a medical
exposure, the benefit is weighed against the
detriment, including radiation effects.
➢ For diagnostic procedures the potential detriment is
the risk of inducing cancer.
➢ This risk is greater in children and decreases with
age.
➢ For effective doses greater than 100 mSv the
overall lifetime risk of fatal cancer is estimated to
be 5% per Sv. (ICRP 2007b)

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Radiation Protection issues in NM
After Radiopharmaceutical administration
➢ The period of time during which patients (and their
family and friends) should observe the restrictions
will depend on the initial external dose rate from
the patient and the rate of clearance of the
radionuclide from the body.
➢ The recommended values are based on data from
Woodings (2004) and Woodings et al (2005) using
a dose constraint of 1 mSv (0.3 mSv for travel)
for children and members of the public, and 5 mSv
for the carer/partner (ARPANSA 2002a).
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Periods of restriction, after discharge, for patients receiving
radioiodine (131I) therapy for thyroid cancer after
thyroidectomy
➢ Thyroid cancer patients may return to work the day after
being discharged from hospital.
➢ Public transport (airline, bus, or boat) can involve
people sitting close to each other and restrictions may
be required on travel of long duration.
➢ For long trips patients should be encouraged to find a
place where they can sit alone.
➢ Long distance travel immediately after administration is
not recommended due to the potential for travel sickness
and the possibility for contamination.

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Radiation Protection Issues in NM

Avoidance of Conception
➢ The period of time for which pregnancy should be
avoided is determined by the rate of clearance of the
radionuclide from the body and by the time necessary
to ensure that the underlying disease is controlled.
➢ The ICRP has recommended that a woman not
become pregnant until the potential fetal dose would
not exceed 1 mGy (ICRP 2000a).
➢ Most female patients are advised (ICRP 2004) not to
become pregnant for at least six months after
therapy with radioiodine.
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Radiation Protection Issues in NM

Avoidance of Conception
➢ Although there is no evidence that preconceptual
irradiation of males can cause any abnormality in
their offspring, it may be prudent to advise males
receiving radionuclide therapy to avoid fathering
children for a period of 4 months, which is
greater than the life of a sperm cell (ARSAC
2000).

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Dose reduction in Mammography

• Dose reduction by limiting the number of projections taken or lowering


the dose associated with each projection
• Axillary projections should only be done with radiologist request
• For routine screening, only craniocaudal and mediolateral projections
of each breast with adequate compression for uniform breast tissue
compression from the nipple to the most posterior portion

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Magnetic Field Hazards

➢ Anyone entering the MRI


scanner room will be
exposed to the magnetic
attraction of the scanner.
The closer you get to the
scanner, the stronger the
magnetic pull.
➢ Any ferromagnetic
object* brought into the
MRI scanner room will
become magnetized and
can become a Dangerous
Projectile! An object that is attracted by a magnet and
which can become magnetized itself

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Magnetic Field Hazards
Displacement and Heating Effects
of Surgical Implants of the Patients:
➢ Cardiac pacemakers
➢ Stents and Anenurysm Clips
➢ Neurostimulators
➢ Pain control pumps
➢ Penile implants
➢ Cochlear implants

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Static Field Bioeffects Summary

➢ Whole body: No effects at 2T


➢ Circulatory at 5T
➢ Extremities: No effects at 5T
➢ Discomfort; no adverse effects: 10T
➢ Conclusion: Don’t go above 2T
for head & trunk; 5T for
extremities

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Gradient Magnet Bioeffects

➢ Allows spatial encoding of the


MRI signal
➢ Also critical for a wide range of
physiologic techniques.
➢ E.g. MR angiography, diffusion
and perfusion imaging.
➢ No effects at 6 T/sec.
➢ Nerve stimulation at 20 T/sec.
➢ Don’t go above 20 T/sec.

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Quantify the Bioeffects by Radiofrequency
➢ SAR: Specific
Absorption Rate
◆Rate at which RF
energy is coupled
into tissues.
◆Units: Watts per
kilogram (W/kg)

➢ SAR - Time Product


◆Units: W-min/kg or
W-hr/kg

➢ Temperature (°C)
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Thermal Effects of Ultrasound

➢ It affects the equilibrium


between chemical reactions
➢ It is advised that tissue
temperature increases, if
applicable, should be kept below
1.5⁰C
➢ Bone and full bladder may
further increase the operational
temperature of US

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Non-thermal Effects of Ultrasound
➢ Cavitation
➢ Acoustic streaming
➢ The main effects of non-thermal
damage have been demonstrated in
mammalian tissues containing gas
where capillary bleeding has been
observed
➢ This potentially pertains to the neonatal
lung, intestine and also in the presence
of US microbubble contrast agents (e.g.
Sulphur hexafluoride (SonoVue) for
liver lesions, Perflexane lipid
microspheres for assessing cardiac
function)
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Special Features of RT Department

• Almost always at the basement / ground floor


• Huge machine and thick walls (e.g. 1 m concrete + 0.5 m
steel)
• Isolated from other departments in hospital
• 24 hours ventilation

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Sources of Radiation from Treatment
Machines
1. Photons
• Primary (Primary barrier)
• Leakage (Secondary barrier, maze wall, door)
• Patient scatter (Secondary barrier, door)
• Wall scatter (Secondary barrier, door)
2. Neutrons
• Primary (Primary barrier)
• Leakage (Primary barrier, secondary barrier, maze wall, door)
• Produced in laminated primary barrier (Primary barrier)
• Room and maze scattered neutrons (Door)
• Neutron induced capture gamma rays (Door)

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Linac room design

• Primary & secondary barrier


• Wall thickness
• Wall material
• Maze
• Room occupancy

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Shielding Goals

• Controlled areas = 0.1 mSv/wk (Equivalent to 5 mSv/yr for the fetus of


pregnant radiation workers, and 0.5 times of the 10 mSv/yr for adult
workers)
• Uncontrolled areas = 0.02 mSv/wk (Equivalent to 1 mSv/yr for the
public)

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Shielding Materials

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Neutron Shielding

• Photoneutron production is an issue for beams with E > 10MeV.


• More of a problem for x-ray beams than for e beams
• n production increases as E increases from 10MeV to 20 MeV,
and then remains roughly constant.
• For 15MV beam, n dose = ~0.5% of x-ray dose on treatment field, and
~0.1% of x ray dose outside the field.
• Thermal neutrons: E<0.1 MeV, after slowing by scattering events.
• Concrete barriers are sufficient for n shielding

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Maze and Door

• Doors are the biggest concern. A hydrogenous material, such as


polyethylene, is added to doors to thermalize (slow down) neutrons.
• When thermal neutrons are absorbed by nuclei, neutron capture
gamma rays are produced, with energies up to 8 MeV. Thus, additional
steel or lead shielding is necessary beneath a polyethylene slab, to
attenuate the resulting gamma radiation.
• A maze can be used to reduce the need for heavy metal shielding on
treatment room doors.

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Simulator

• Similar to the setting of a General X-ray / CT room


• Remote control room
• Lead glass window
• Sufficient barrier
• Door interlock
• Warning signs and light

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


Sealed Source
• High dose rate (HDR) brachytherapy, remote automatic machine
(afterloader) are used to deliver the sources into the patient.
• Treatment room usually incorporate with operation theatre
• C arm x ray machine
• Room design is similar to linear accelerator (Same barrier design
equations as used for external beam therapy devices with E > 500kVp)
• Additional requirements: Dose in any unrestricted area must not exceed
2 mrem (0.02 mSv) in any 1 hour.
• All barriers receive primary (source radiation is isotropic), so it makes
sense to conservatively design all barriers to be primary barriers, of the
same thickness.
• Linac vaults generally provide more than adequate shielding to house
an HDR unit
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS
Final Examination Information

• 60% of subject grade


• Must PASS to pass the whole subject
• 3/12/2022 (Sat) 12:30-14:30 SH2
• Duration: 2 hours
• Format: 60 MCQ (50%) and about 10 Structured Q (50%)
• Bring pencil for filling MCQ answer sheet
• Follow rules of examination
• Sample structured questions uploaded to Blackboard

DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS


The End!
Hope you can get good results
in final exam!
DEPARTMENT OF HEALTH TECHNOLOGY AND INFORMATICS

You might also like