You are on page 1of 44

Installation Manual

Part Number MAN-00344


Revision 001
Installation Manual

Table of Contents

List of Figures -iii

List of Tables -v

Preface P-1
1.0 About This Manual ..............................................................................................................................P-1
2.0 Quality Control ...................................................................................................................................P-1
3.0 Training ...............................................................................................................................................P-1
4.0 Prerequisites ........................................................................................................................................P-1
5.0 Product Complaints .............................................................................................................................P-1
6.0 Terms and Definitions .........................................................................................................................P-2
7.0 Warnings, Cautions, and Notes ...........................................................................................................P-3
8.0 SecurViewRT Manual Set .....................................................................................................................P-4
9.0 Using the Installation Manual ..............................................................................................................P-4
Chapter 1—General Information 1-1
1.0 Introduction .........................................................................................................................................1-1
1.1 SecurViewRT Technologist Workstation .........................................................................................1-1
1.2 SecurViewRT Manager ...................................................................................................................1-1
1.3 Multi-Workstation Configuration ...................................................................................................1-2
Chapter 2—Receiving and Unpacking 2-1
1.0 Receiving Instructions ..........................................................................................................................2-1
2.0 Unpacking ...........................................................................................................................................2-1
Chapter 3—Installing the Workstation and Manager 3-1
1.0 Installing a Stand-Alone, Client, and Manager .....................................................................................3-1
1.1 Pre-installation Considerations ......................................................................................................3-1
1.2 Installing the Hardware .................................................................................................................3-1
2.0 Initial Setup .........................................................................................................................................3-3
2.1 Pre-Startup checklist ......................................................................................................................3-3
2.2 Starting the Manager or Stand-Alone .............................................................................................3-4
Chapter 4—Configuring the SecurViewRT Workstation 4-1
1.0 System Startup .....................................................................................................................................4-1
2.0 Configuring the SecurViewRT Workstation ...........................................................................................4-1
2.1 Windows Setup .............................................................................................................................4-1
2.2 DICOM Settings—Manager and Stand-Alone ................................................................................4-1
2.3 Configuring a Multi-Workstation ...................................................................................................4-5
2.4 Calibrating the Barco Nio Display .................................................................................................4-7
3.0 Converting Stand-Alone Workstations to Clients ..................................................................................4-9
3.1 Pre-Installation Checklist—Existing Stand-alone ............................................................................4-9
3.2 Converting ....................................................................................................................................4-9

MAN-00344 i
Installation Manual

4.0 Final Checks ..................................................................................................................................... 4-12


4.1 Stand-Alone ................................................................................................................................ 4-12
4.2 Multi-Workstation ...................................................................................................................... 4-12
5.0 Other Functionality ........................................................................................................................... 4-12
5.1 Version Upgrade ........................................................................................................................ 4-12
5.2 Backup and Restore .................................................................................................................... 4-12
5.3 System Time Differences ............................................................................................................ 4-12
Chapter 5—Configuring the Barcode Scanner 5-1
1.0 Supported Type ................................................................................................................................... 5-1
2.0 Configuration ...................................................................................................................................... 5-1
2.1 Emulate Keypad ............................................................................................................................ 5-1
2.2 Beep Volume and Sound .............................................................................................................. 5-1
2.3 Changing Between Patient ID and Accession Number .................................................................. 5-2
Appendix A—Technical Characteristics A-1

Appendix B—Troubleshooting B-1


1.0 Network Problems .............................................................................................................................. B-1
1.1 Network Problems on a Client ...................................................................................................... B-1
1.2 Manager ....................................................................................................................................... B-2
2.0 Troubleshooting Table—Reviewing Module ....................................................................................... B-3
Index -i

ii MAN-00344
Installation Manual
List of Figures

List of Figures

Figure 1-1 SecurViewRT Technologist Workstation .....................................................................................1-1


Figure 1-2 SecurViewRT Multi-Workstation Overview ................................................................................1-2
Figure 3-1 System Cable Connections.........................................................................................................3-1
Figure 3-2 Display Connections (Workstation)............................................................................................3-2
Figure 3-3 Display Connections (Manager) .................................................................................................3-2
Figure 3-4 Monitor Power and Cable connections ......................................................................................3-3
Figure 4-1 Setup window to configure DICOM entities...............................................................................4-2
Figure 4-2 Printer configuration ..................................................................................................................4-3
Figure 4-3 Configuring Destinations ...........................................................................................................4-4
Figure 4-4 Multi-Workstation Tab: Configuration of System Type ...............................................................4-5
Figure 4-5 Display Settings—Display Tab ...................................................................................................4-7
Figure 4-6 Calibration—Display Calibration ...............................................................................................4-8
Figure 4-7 Display Settings—Visual Calibration..........................................................................................4-8
Figure 5-1 Enable Keypad Emulation ..........................................................................................................5-1
Figure 5-2 Beep After a Good Decode........................................................................................................5-1
Figure 5-3 Do Not Beep After a Good Decode ...........................................................................................5-1
Figure B-1 Startup Dialog .......................................................................................................................... B-1

MAN-00344 iii
Installation Manual
List of Figures

iv MAN-00344
Installation Manual
List of Tables

List of Tables

Table P-1 SecurViewRT Documents ...........................................................................................................P-4


Table 1-1 Tasks of the Nodes .....................................................................................................................1-2
Table 4-1 Dialog Elements of Workstation Integration Dialog ..................................................................4-11
Table B-1 Reviewing Module ................................................................................................................... B-3

MAN-00344 v
Installation Manual
List of Tables

vi MAN-00344
Installation Manual
Preface
About This Manual

Preface

1.0 About This Manual


This manual describes instructions for installing and configuring the SecurViewRT
Technologist Workstation.

2.0 Quality Control


• All quality control tests described must be performed at the prescribed frequencies as
required under MQSA regulations.
• To assure continued high level operation of SecurViewRT workstations, the
recommended quality control procedures must be followed.

3.0 Training
Hologic does not accept responsibility for injury or damage associated with improper or
unsafe system operation.
Service engineers must ensure that they receive training on SecurViewRT with Hologic
training programs prior to servicing the unit.
Service engineers should refer to the SecurViewRT User/Administrator Guide for directions
on how to use and service the SecurViewRT.

4.0 Prerequisites
The procedures in this manual assume that you know how to do the following operations
related to the system’s graphical user interface:
• Perform mouse operations, such as click, drag, and select
• Select from menus
• Click buttons
• Enter information in text fields
• Select options
• Select entries from drop-down lists
• Use scroll bars

5.0 Product Complaints


Any health care professional (e.g., customer or user of this system of products) who has any
complaints or has experienced any dissatisfaction in the quality, durability, reliability, safety,
effectiveness, and/or performance of this product should notify Hologic. If the device
malfunctions and may have caused or contributed to a serious injury of a patient, Hologic
should be notified immediately by telephone, fax, or written correspondence.

P/N MAN-00344 P-1


Installation Manual
Preface
Terms and Definitions

6.0 Terms and Definitions


AE Applications Entity. A DICOM term referring to the name
of an application that performs DICOM services.
Auto-Fetch The retrieval of patient data from a PACS, which is
triggered by the arrival of a new notice at a SecurViewRT
Workstation or Manager.
AWS Acquisition Workstation.
CAD Computer Aided Diagnosis/Detection. Software which
uses sophisticated image processing and image analysis
algorithms to search images for regions that show signs of
cancer.
CAD-SR CAD Structured Report.
CC, MLO Names of view positions for screening mammography.
Client A SecurViewRT workstation that is connected to a Manager

DICOM Digital Imaging and Communications in Medicine. Industry


standard for connection of medical imaging equipment.
Image A set of modality specific binary data and identifying
attributes which represents the result of an imaging
procedure performed on a patient. DICOM uniquely
identifies image objects with a globally unique identifier
(UID).
LUT A Look-Up Table (LUT) is an image processing function
that replaces one image pixel value with a different image
pixel value.
MQSA Mammography Quality Standards Act
Multi-Workstation A group of SecurViewRT workstations that share a common
database on the Manager. A group of up to four systems
that allows access to all patients at the same time on any of
the workstations.
Multi-Workstation Client A SecurViewRT workstation that is connected to a
Manager.
Operating System (OS) The basic software control system which runs all functions
of a computer.
PACS Picture Archiving and Communications System. A
computer and network system for the transfer and
archiving of digital medical images.
SCP Service Class Provider
SCU Service Class User
SecurViewRT The complete software system for the SecurViewRT
Technologist Workstation.
SecurViewRT Manager A computer that manages communication and contains a
central database within a multi-workstation configuration.

P-2 P/N MAN-00344


Installation Manual
Preface
Warnings, Cautions, and Notes

SecurViewRT Workstation The SecurViewRT Technologist Workstation is a dedicated


one monitor Softcopy Reading system with reduced
functionality compared to the SecurViewDX system with
two 5 Megapixel monitors. The SecurViewRT Technologist
Workstation works as a standalone system next to an
Acquisition workstation or as part of a Multi-Workstation.
The SecurViewRT Technologist Workstations form a
separate Multi-Workstation.
Series A set of images which have all been acquired by a single
technician for a single study with the same modality.
Stand-Alone Workstation A single SecurViewRT workstation.

Study A specific instance of a procedure on the Acquisition


Workstation, consisting of a set of images and/or series, which
has been performed on a specific patient, and is usually
identified in some locally unique way. DICOM uniquely
identifies studies with a globally unique instance UID.
UPS Uninterruptible Power Supply
View The combination of a single X-ray image and a specific set of
conditions under which the image was acquired. View is not
part of DICOM nomenclature, but in the context of DR, is
approximately synonymous with a DICOM image object.

7.0 Warnings, Cautions, and Notes


Definitions of Warnings, Cautions, and Notes used throughout this manual are as follows:

WARNING! This alerts you to procedures that you must follow precisely to
avoid causing potentially serious or fatal injury to yourself or
others.

Warning: Warnings indicate procedures that you must follow precisely to


avoid injury to yourself or others.

Caution: Cautions indicate procedures that you must follow precisely to avoid dam-
age to equipment, loss of data, or corruption of files in software applica-
tions

Note… Notes indicate important information that must be followed to ensure the
proper operation of the system.

P/N MAN-00344 P-3


Installation Manual
Preface
SecurViewRT Manual Set

8.0 SecurViewRT Manual Set


The SecurViewRT Manual set contains the following documentation:
Table P-1: SecurViewRT Documents

Title P/N Use


User/Administrator Guide MAN-00303 Technologist Workstation
Installation Manual MAN-00344 SecurViewRT Installation

9.0 Using the Installation Manual


Chapter 1—General Information. This chapter provides an overview of the
SecurViewRT system, including the Workstation and Manager, Multi-Workstation
configuration, and DICOM information.
Chapter 2—Receiving and Unpacking. This chapter contains information for
receiving and unpacking the unit.
Chapter 3—Installing the Workstation and Manager. This chapter provides
procedures for installing SecurViewRT Stand-Alone or Client Workstations and
SecurViewRT Managers. Procedures for converting Stand-Alone Workstations to
Clients are also provided.
Chapter 4—Configuring the SecurViewRT Workstation. This chapter provides
procedures for starting the system, setting up Windows and DICOM, configuring the
Workstation and Manager, and calibrating the LCD display.
Chapter 5—Configuring the Barcode Scanner. This chapter provides procedures
for configuring the barcode scanner.
Appendix A—Technical Characteristics. This section lists relevant technical
characteristics for the SecurViewRT workstation.

Appendix B—Troubleshooting. This chapter provides procedures for troubleshooting


network, image, and computer problems.

Note… At the completion of this installation, the AWS, CAD server, and/or PACS must be
configured to send to the Manager (not to individual clients). Consult the
appropriate User’s Manuals for instructions on how to configure these devices.

P-4 P/N MAN-00344


Installation Manual
Chapter 1—General Information
Introduction

Chapter 1—General Information

1.0 Introduction
1.1 SecurViewRT Technologist Workstation
The SecurViewRT Technologist Workstation is used by a physician to review images.The
Workstation consists of:
• One 3 or 5 MegaPixel Resolution Landscape Display
• Computer
• Keyboard and Mouse
• UPS

Figure 1-1: SecurViewRT Technologist Workstation

1.2 SecurViewRT Manager


• LCD display
• Computer
• Keyboard and Mouse
• UPS
This Manager acts as the DICOM entity for the multi-workstation and is responsible for
receiving DICOM images. All SecurViewRT workstations are connected via 100 Mbit/s
ethernet to a central Manager (server). The Multi-Workstation acts as one single DICOM
entity.
The Manager also provides central database access and performs image preparation for all
SecurViewRT workstations.

P/N MAN-00344 1-1


Installation Manual
Chapter 1—General Information
Introduction

1.3 Multi-Workstation Configuration


1.3.1 Design
Up to four Client workstations can be configured with a SecurViewRT Manager as a
Multi-Workstation system. The general layout of the Multi-Workstation system is
displayed in Figure 1-2.

DICOM
Network
Client Client

Manager
Database

Client Client
Figure 1-2: SecurViewRT Multi-Workstation Overview

Note… A service engineer configures a Multi-Workstation system with all local


system settings.
1.3.2 Types of Nodes
The different tasks of the nodes (SecurViewRT Manager or SecurViewRT Client) are
described in Table 1-1.

Table 1-1: Tasks of the Nodes

Node Type Tasks


SecurViewRT Manager • Administration/Settings
• No reviewing
• Handles DICOM connections
• Central database manager, contains:
• DICOM data (all except pixel data)
• Configurable system settings
• User preferences
• Prepared pixel data
• Central preparation
• Central authentication
• Requires license: TWS-Server
SecurViewRT Client or • Patient browser
Stand-Alone • Administration/Settings
• Reviewing
• Query/Retrieve
• Pull prepared images from multi-workstation manager for
reviewing (Client only)
• Store copies of all CAD SRs
• Requires basic license: TWS-Client or TWS-Standalone

1-2 P/N MAN-00344


Installation Manual
Chapter 2—Receiving and Unpacking
Receiving Instructions

Chapter 2—Receiving and Unpacking

The following sections detail receiving and unpacking instructions.

1.0 Receiving Instructions


The SecurViewRT Client or Stand-Alone includes the following:
• One high resolution display
• Computer
• Keyboard and Mouse
• UPS
A SecurViewRT Manager includes the following:
• LCD display
• Computer
• Keyboard and Mouse
• UPS
Upon receipt, perform the following before opening the containers:
1. Inspect each container for damage.
2. Note any damage on the shipping manifest.
3. Notify Hologic of any external shipping damage that may have occurred.

2.0 Unpacking
WARNING! To prevent injury to personnel and/or damage to equipment, care
must be taken when uncrating the equipment.

Unpack as follows:
1. If applicable, cut the retaining straps that secure the boxes to the shock-mounted
wooden pallet. Remove the boxes from the pallet.
2. Open all boxes and verify the contents against the packing list and sales order.

Note… If there is a discrepancy between the contents and the packing list or sales
order, contact Hologic immediately. If it is necessary to repack any items for
future installation, use the original packaging materials.
3. Carefully remove shipping materials (foam padding, tie-downs, straps, shipping wrap, etc.).
4. Inspect each item for damage, then safely store them near the installation site.

Note… If shipping damage is of a concealed nature, contact the carrier as soon as


such damage is found, and request an inspection for shipping damage.
Normally, any claims for shipping damage must be completed within 15
days of receiving the shipment.

MAN-00344 2-1
Installation Manual
Chapter 2—Receiving and Unpacking
Unpacking

2-2 MAN-00344
Installation Manual
Chapter 3—Installing the Workstation and Manager
Installing a Stand-Alone, Client, and Manager

Chapter 3—Installing the Workstation and Manager

1.0 Installing a Stand-Alone, Client, and Manager


1.1 Pre-installation Considerations
This section applies to installation of new:
• Stand-Alone workstations.
• Client Workstations that are installed as part of a multi-workstation system.
Prior to beginning the installation, discuss the following with the installation site:
• An IP address and AE title for the Manager should be obtained from the PACS
administrator, and the site must be informed that an additional Ethernet port is required.
• SecurViewRT can process “For Processing” images only if they are older than five days and
considered to be prior images. For this reason, any site where SecurViewRT is to be installed
must have been generating “For Presentation” images for all new studies for at least five days.
• The SecurViewRT software now supports “auto-fetching.” This allows for automatic
querying of the PACS, upon arrival of new notices, to retrieve corresponding images.
Check with the site to determine if they want to enable this feature.
1.2 Installing the Hardware
Client Workstations are first set up as Stand-Alone Workstations using procedures in this
chapter, then are modified using the procedure in Chapter 4, Section 3.0, p. 4-9.
1.2.1 Setup of Hardware
Referring to Figure 3-1, connect the:
• Mouse to the mouse port.
• Computer keyboard to the PS2 or USB port on the rear of the computer.
• Network cable between the computer and the hospital network.
• Bar Code Scanner (not shown) to a USB port.

Figure 3-1: System Cable Connections

MAN-00344 3-1
Installation Manual
Chapter 3—Installing the Workstation and Manager
Installing a Stand-Alone, Client, and Manager

1.2.2 Cable Connections for the LCD Displays


1. Set the display in landscape position.
2. If the power and DVI cables are not installed on the workstation display, remove the
protective cover from the rear of the display. See Figure 3-2 and Figure 3-3.
3. Connect the DVI cable to the display.
4. Connect the output of the 12 VDC power supply to the display.

Note… The workstation power cable connection to the rear of the display uses an
8-pin DIN connector.
5. Connect one end of the AC power cable to the AC input of the DC power supply.
6. Reinstall the protective cover (workstation), carefully routing the power and DVI
cables out through the “chute” on the covers.
7. Connect the DVI cable to the Vid 1 connector on the computer.
Protective
Cover

12 VDC
Power Supply

Power
Cable

DVI
Cable

Figure 3-2: Display Connections (Workstation)

DVI
Cable to
Manager
PC

Power
Cable

Figure 3-3: Display Connections (Manager)

3-2 MAN-00344
Installation Manual
Chapter 3—Installing the Workstation and Manager
Initial Setup

8. Connect one end of the UPS serial cable into the UPS and the other into COM2
on the rear of the computer. See Figure 3-4.

COM2

UPS Serial
Cable
Monitor
Power Cable

Computer
Power Cable

Figure 3-4: Monitor Power and Cable connections

9. Connect the power cable to the UPS.


10. Connect the computer and monitor to the UPS.
11. Connect to an Ethernet port on the facility network.

2.0 Initial Setup


2.1 Pre-Startup checklist
Prior to starting the system, review the following:
1. An IP address and AE title for the Manager should be obtained from the PACS
administrator, and the site must be informed that an additional Ethernet port is required.
2. The Manager consists of a Dell 670 computer with an LCD display. It should be located
in an area that is convenient to the SecurViewRT Client Workstations.
3. For both Manager and Stand-Alones, prior to beginning the installation, obtain the AE
titles, IP addresses, and ports for the facility PACS and DICOM printer(s) from the PACS
administrator. For Stand-Alone, record below.
Information: AE Title: ___________________
IP Address: ___________________
Ports:___________________
4. At the completion of the installation, output from the AWS, PACS, CAD server should be
directed to the Manager only, not to individual workstations.

MAN-00344 3-3
Installation Manual
Chapter 3—Installing the Workstation and Manager
Initial Setup

2.2 Starting the Manager or Stand-Alone


1. Turn on the computer. The SecurViewRT application starts.
2. Log on as the service user and select Exit to Windows.
3. From the Start menu, select Log Off scr.
4. At the Windows login prompt, log into Windows as Administrator.
5. Set the Host Name as follows:
a. Right-Click “My Computer” on the desktop.
b. Select Properties.
c. Select the “Network Identification” tab.
d. Click the Properties button.
e. Enter the AE Title obtained from the site’s PACS Administrator as the “Computer
Name” and click OK.
f. Record the AE Title (computer name) in Manager Information below. Stand-Alone
was previously recorded.
g. Close all windows.
6. Set the IP address as follows.
a. Right-click on “Network Places” on the desktop.
b. Select Properties. The “Network and Dial-up Connections” windows appears.
c. Right-click “Local Area Connection.”
d. Select Properties. The “Local Area Connection” window appears.
e. In the list of components, highlight “Internet Protocol (TCP/IP).”
f. Click Properties. The “Internet Protocol (TCP/IP)” window appears.
g. Enter the necessary information (as obtained from the PACS administrator) and click
OK.
h. Record the IP Address below.
i. Close all windows.
Manager Information: AE Title: ____________
IP Address: ___________________

3-4 MAN-00344
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation

Chapter 4—Configuring the SecurViewRT Workstation

1.0 System Startup


1. Turn on the UPS.
2. Turn on the monitor.
3. Turn on the SecurViewRT Workstation computer.
4. If necessary, contact Hologic Help Desk for passwords.

2.0 Configuring the SecurViewRT Workstation

2.1 Windows Setup


1. Log into the SecurViewRT software using a service account.
2. Click the Exit to Windows tab on the right-hand side of the monitor.
3. Double-click the TB ON shortcut.
4. Hold down the Shift button and click Start/Logoff/Yes.

Note… The Shift button must be held down during the entire sequence, until you
see the Logon prompt screen.
5. Log on to Windows as administrator.
6. Configure the following in Windows:
• User Interface Language
• Network Settings
• Regional Settings
• Host Name
7. Restart the computer.

2.2 DICOM Settings—Manager and Stand-Alone


Configuration of DICOM Settings is part of the configuration of the SecurViewRT at the
customer site.

Note… Only service users have access to DICOM settings.When entering DICOM
titles, always use capital letters.
1. Log into the SecurViewRT software using a service account.
2. Click the Administration tab on the right-hand side of the right monitor.

MAN-00344 4-1
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation
Configuring the SecurViewRT Workstation

2.2.1 Setup Tab


The Setup tab comprises four group boxes (numbered items in Figure 4-1) to
configure DICOM entities:
• This Workstation (#1): Displays all connected entities of this workstation (SCR-
System, PACS/Station, Printer). The entity entered in PACS 1 must support query/
retrieve on the image level.

Note… The PACS 1 field can only be edited if the item “Auto Fetching” in the
Settings Tab is set to “Off,” which is the default setting. See section
“Settings” in chapter “User Management” in the SecurViewRT Technologist
Workstation User/Administrator Guide.
• DICOM Entities (#2): Displays, in alphabetical order, the connected entity types
and provides tools for editing and deleting entities or sending an echo to an
entity.
• DICOM Setup (#3): Contains entries to configure an entity type by AE Title
(AET), IP Address and Port. All fields have to be filled out before a new entity
can be saved and added to the list on the left hand side.
• Additional PACS Settings (vendor dependent) (#4): Contains entries to support
configuring multiple entities and ports for Storage and Query/Retrieve.
Information can be found in the PACS documentation.

Note… Only the MagicStore PACS uses a Query/Retrieve IM port to query/retrieve


images. These settings have to be filled in only for the MagicStore PACS.

Default
Setting

Figure 4-1: Setup window to configure DICOM entities

4-2 MAN-00344
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation

2.2.2 Printer Tab


The Printer tab comprises three boxes (numbered items in Figure 4-2):
• The Printer type box (#1) displays printers with Printer type, Default Medium,
and Medium Size.
Below the list text fields display information about a selected printer from the
list. These entries can be used to edit existing values, and to add a new type to
the list above. The tool bar below contains three buttons to add, edit, or delete a
printer type.
• The Medium box (#2) displays two lists.
The list to the left list contains available mediums like Paper, Blue Film, or Clear
Film. The list to the right is used to assign default mediums to the selected
printer. Use the arrow buttons to assign (or remove) a medium to (from) a
selected printer.
• The Medium Size box (#3) displays two lists.
The list to the left contains available medium sizes. The list to the right is used to
assign exactly one medium size for the selected printer.
Use the arrow button to assign a medium size to a selected printer. An existing
value is always replaced. There are two buttons to add new medium sizes and to
delete medium sizes.

Note… If the printer to be configured is not listed, contact Hologic Support.

Figure 4-2: Printer configuration

MAN-00344 4-3
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation
Configuring the SecurViewRT Workstation

2.2.3 Destination Tab


The Destination tab comprises two boxes (numbered items in Figure 4-3) to
configure user accessible destinations for sending notices.
All entities that are configured as a PACS/Station can be a Destination.

Figure 4-3: Configuring Destinations

• Possible destinations for Notices to be sent are listed in the left box (#1).
• Active (user accessible) destinations for Notices are listed in the right box (#2).

Note… Only select SecurViewDX / SecurViewRT Workstations and SecurViewDX /


SecurViewRT Managers in box #2 of Figure 4-3.

4-4 MAN-00344
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation

2.3 Configuring a Multi-Workstation


The service user is responsible for configuring the system type. Each system can be
set to a dedicated type by using the Multi-Workstation tab.
Configurable items are system type (client, manager, or stand-alone workstation as
Technologist Workstation) and manager network address.

Figure 4-4: Multi-Workstation Tab: Configuration of System Type

The multi-workstation configuration window consists of three radio buttons that


allow the service user to select the workstation type:
• Stand-Alone Workstation
The workstation is a single SecurViewRT Workstation, without a connection to a
Multi-Workstation Manager.
• Client
The Workstation is a client within a multi-workstation system. If the Multi-
Workstation Client option is selected, the Multi-Workstation manager address
(either as DNS name or as IP address) is entered in the “Manager Address” text
field. The Check button is also enabled to check for the availability of the
entered network address.
• Multi-Workstation Manager
The Workstation is the Manager in a multi-workstation system.

MAN-00344 4-5
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation
Configuring the SecurViewRT Workstation

2.3.1 Licensing Information


For each of the three workstation types, there is a separate license. The “Multi-
Workstation” tab is accessible regardless of the installed licenses. To access the
reviewing or to perform any action in the patient browser, a valid license for the
configured workstation type must be installed. See Section 3.0, p. 4-9 in this chapter
and Section 5.0 Licensing in the SecurViewRT User/Administrator Guide for specific
details on licensing procedures.
To verify the licensing information and image processing configuration while logged
into the SecurViewRT software using a service account:
1. Click the Administration tab on the right-hand side of the right monitor.
2. Click the Licensing tab.
3. Compare the license information with the information that came with the system.
4. If they do not agree, enter the correct Feature and Generation information in the
corresponding edit boxes.
5. Enter the new license number and click Check.
6. Select the Info tab and make sure that you can see Library Version and Image
Processing Version information.
2.3.2 Image Processing Configuration—Deleting Images
Images can be deleted manually or automatically.
• Manual Deletion
If images are deleted manually on a client or on the Manager, those images
are cleaned up on all clients and on the Manager, i.e., the prepared images
are deleted everywhere and the DICOM files are deleted from the manager.
• Automatic Deletion
Automatic deletion, if configured, is triggered when the hard disk level
exceeds the warning watermark level.
In a multi-workstation system, automatic deletion behavior is different for a
Manager and for a Client.
• Manager
If the manager reaches its warning watermark level, images and patient data
start auto-deleting. Those image deletions are propagated to all clients, so
the clients remove their locally stored prepared image data for those images.
• Client
If a client reaches its warning watermark level, its locally stored prepared
image data, CAD-SR, starts auto-deleting. Since this data is not removed
from the database, it is still available on the manager, and the data still
appears e.g., in the patient browser, and can be reviewed.
If this happens, the client re-retrieves the previously deleted data from the
manager. If this is the case, a message “Images are being transferred” is
displayed on the screen in the reviewing module, until all images for the
currently viewed case are locally available. Automatic deletion on one
client has no effect on other clients or on the manager.
Every client and the manager have their own individual definition of the
local watermark levels.
2.3.3 Image Processing Configuration—Image Export
Images and DICOM files are always exported locally to a CD on a client
workstation. Export of images to remote clients or to the manager is not possible.
The exported images can be saved to a CD locally on every client.

4-6 MAN-00344
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation

2.4 Calibrating the Barco Nio Display


NioWatch is the software used to calibrate the Technologist Workstation display. NioWatch
is launched using a Windows Systray (System Tray) icon or from the Windows Start menu.
1. Log in to the system as an administrator.
2. Select Exit to Windows.
3. Open NioWatch by double-clicking on the NioWatch icon in the Windows Systray.
The Display Settings window appears. The numbered icons in the upper part of the
display represent the number of display controller heads in the PC. There are three tabs:
• Display tab: You can view the name, serial number of the display, and details about:
• Lifetime: operation time including time in stand-by
• Runtime: operation time excluding time in stand-by
• Firmware version: version of internal display software
• Backlight Stabilization: status of the backlight stabilization in the display

Figure 4-5: Display Settings—Display Tab


From the Display Function drop-down menu, you can select other display functions.
If the display has been calibrated to the selected display function in the past, the
display is set according to this calibration. If a former calibration is not found,
factory default settings are automatically selected.
When you restart NioWatch, the system automatically selects the last selected
display function.
You can manually adjust luminance by adjusting the Luminance slider. Your setting is
saved when you shut down the system. If you position the slider to the right of “CAL,”
luminance is in the calibrated position as determined during the latest calibration. A
slider position to the left of the CAL mark is not in the calibrated position.

Note… To be DICOM compliant, you must select the DICOM Display function and
calibrate the display. After calibration, to maintain DICOM compliance, the
luminance must remain in the calibrated position.

MAN-00344 4-7
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation
Configuring the SecurViewRT Workstation

• Graphics Board tab: This tab displays information about the installed display
controller (serial number, driver version, and firmware version).
• Calibration tab: You can calibrate the display, revert to the latest calibration, revert to
factory default settings, and view the result of the latest calibration.
2.4.1 Manual Calibration
1. From the Display Settings window select the Calibration tab.
2. Click Calibrate. The Calibration window appears.

Figure 4-6: Calibration—Display Calibration


3. In the Calibration window, select Visually and click OK.
The Display Settings—Visual Calibration dialog appears.

image background

bitmap

Figure 4-7: Display Settings—Visual Calibration


4. Adjust the slider until there is almost no visible difference between the
background of the image above the slider and the bitmap inside this image.
When done, click Set point 1 of 19.
5. Repeat step 3 for point 2 of 19, 3 of 19, etc., until the Set point button
disappears, which indicates the last point is set.
6. After setting the last point, click OK.
2.4.2 Resetting to the Latest Calibration
If you want to return the display to the latest calibration:
1. From the Display Settings window, Calibration tab, click Calibrate.
2. In the Calibration window, select Use the latest calibration.
3. Click OK. The calibration and luminance reset to the latest calibrated position.
Note… When you start the system, calibration resets to the latest calibrated
position. Luminance, however, remains at the latest value.
2.4.3 Reverting to Factory Defaults
To restore factory default calibration values, in the Calibration tab dialog click Reset.

4-8 MAN-00344
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation

3.0 Converting Stand-Alone Workstations to Clients


This section applies both to a new installation of a multi-workstation system as well as to
existing stand-alone workstations which are being converted to Client Workstations with the
addition of a SecurViewRT Manager. For a new installation, first follow instructions for stand-
alone or Client Workstation Installation to set the new workstations up in stand-alone mode,
then continue with this section.

3.1 Pre-Installation Checklist—Existing Stand-alone


If this installation involves existing stand-alone Workstations which are being re-configured
as part of a Multi-Workstation system:
1. Make sure that all workstations are currently at SecurViewRT V2-0-1 or higher.
2. In order to minimize the installation time, as many cases as possible should be deleted
from each workstation using the Case Administrator account. No more than 50 cases
should remain in the patient list before the conversion is begun. The following steps
should be performed:
a. Have the users view as many cases as possible prior to the upgrade.
b. Delete all viewed cases from the system using the Case Administrator (delete)
account (not from the Service account).
c. Delete all older cases, regardless of the status. Hologic recommends that only the
last 1-2 days of unviewed cases be saved.

3.2 Converting
1. On a workstation that is to be configured as a Client, log in to the SecurViewRT
application as the service user and select Exit to Windows.
2. From the Start menu, select Log Off scr.
3. At the Windows login prompt, log into Windows as Administrator.
4. Ensure that the Windows System time matches the Manager’s Windows system time by
opening a Command prompt and typing:
net time \\<MANAGER-IP> /set
where the <MANAGER-IP> is the Manager IP recorded in a previous section.
5. Log out and log in as SCR.
6. Log in to the application as the service user.
7. Update the appropriate license keys for the Client Workstation.

Note… SecurViewRT requires new license keys. A master list of license keys is
included in the e:\mbc-scr-1\Hologic-Service\license_keys.txt file. For
each softcopy workstation, note the ID printed on the dongle (begins with
8-5E7…), then search the license_keys.txt file to locate the appropriate
keys. If the keys cannot be located, contact the Hologic Help Desk.
a. Log in to the application as the service user.
b. Select Administration
c. Select the Licensing tab.

MAN-00344 4-9
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation
Converting Stand-Alone Workstations to Clients

d. Highlight any currently installed Generation 1.0 or 2.0 license keys and select
Delete.
e. If you are converting a Stand-Alone workstation to a Client, highlight the TWS-
STANDALONE key and select Delete.

Note… If this is a new installation, then Client license keys were entered at the
factory. If this is a conversion of existing stand-alone workstations, enter the
key for TWS-CLIENT. See Section 5.0 Licensing in the SecurViewRT User/
Administrator Guide.
8. Under the Feature field, enter the following:
• Feature: TWS-CLIENT
• Generation: 3.0
• Key: 20 character string obtained from license_keys file

Note… Make sure that the Key field is a 20 character string. Append zeros to the
end if the key contains less then 20 characters.
9. Click the Check button. The system verifies the information with the dongle and adds the
key if successful.
10. If connectivity to a CAD processor is required, enter the CAD license information and
again press the Check button. If there is no CAD processor, do not enter this information.
• Feature: MBC-CAD
• Generation: 3.0
• Key: 20 character string obtained from license_keys file
11. The client workstation is initially set up as a stand-alone system. To convert from a stand-
alone system to a Client:
a. Click the “Multi-Workstation” tab.
b. Click on the “Cluster Client, Manager Address” button.
c. Enter the Manager IP (See Chapter 3, Starting the Manager or Stand-Alone, p. 3-4)
and click Check. Confirm that it connects to the Manager.
d. Click Save. Confirm that you want to save the new system type.

Note… If this is a new installation, then there is no data to transfer and you should
leave the check boxes unchecked. However, you must still perform the
Migrate step.
e. Press Migrate.
• After the migration is complete, both the Client and the Manager restart and
synchronize.
• When the system restarts, the Client synchronizes with the Manager. The length
of time it takes to synchronize depends on how much data is on the system.
f. Repeat this section for all clients in the multi-workstation system.

4-10 MAN-00344
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation

3.2.1 Dialog Elements


The dialog elements of the Workstation Integration dialog have the following
functions:

Table 4-1: Dialog Elements of Workstation Integration Dialog


Element Type Description
User List View This list view is only enabled if one or both of the “Migrate User
Assignment Data” or “Migrate Annotations” check boxes are checked. It
displays how the local users (first column) are assigned to users on
the multi-workstation manager (second column). These manager
users either exist, or they are created.
If the user already exists, their user data on the manager is
overwritten by the local data of that user. If the manager user is
“[Same],” the local user and the manager user are considered the
same, but the local user data is not transferred to the manager (i.e.,
the manager user data is kept).
If a local user is assigned to the manager user “[None],” no migration
of the local user is done. This is especially useful for the two local
users “Application” and “Service,” who are not migrated by default.
Assign to Dropdown This element is only enabled if the check box “Migrate User Data”
Manager List is checked. The list is filled with the user names of all users on the
User multi-workstation manager, in addition to the special values
“[None]” and “[Same].”
Select a user in the dropdown list, and click the button “Set.” The
local user currently selected in the list view “User Assignment” is
assigned to the selected manager user.
Set Button This dialog element is only enabled if the check box “Migrate User
Data” is checked and a user is selected in the list view “User
Assignment.”
It assigns the user currently selected in the dropdown list “Assign to
Manager User” to the user selected in the list view “User Assignment.”
New Button This dialog element is only enabled if the check box “Migrate User
Data” is checked.
Click button to open a dialog where a new user name can be entered.
The name is added to the dropdown list “Assign to Manager User.”
Migrate Button Click this button to perform the migration.
If the check box “Migrate Image Data” is checked, all local image
data is moved to the multi-workstation manager.
If the check box “Migrate User Data” is checked, a user data
migration according to the list view “User Assignment” is done. All
user names assigned in this list view, that do not exist on the
manager, are created. All user data of users that already exist on
the manager, and that are marked for migration in the list view
“User Assignment,” are overwritten by the local user data.
A message box informs that the migration has finished. From now
on the workstation has no more local data, and is integrated into
the specified multi-workstation system.
Confirm message to close the migration dialog automatically.
Cancel Button Closes the migration dialog without doing any migration.

MAN-00344 4-11
Installation Manual
Chapter 4—Configuring the SecurViewRT Workstation
Final Checks

4.0 Final Checks


4.1 Stand-Alone
1. For a stand-alone configuration, log in to the application as the service user. Select
Administration.
2. Select the Backup/Restore tab. Create a backup CD for the customer by pressing the Backup
button and inserting a blank CD-R when prompted. When backup is completed, select Exit,
then Logout. Label the CD as “SecurViewRT Database Backup,” with today’s date.
4.2 Multi-Workstation
1. For a multi-workstation configuration, log in to the Manager as the service user. Select
Administration.
2. Select the Backup/Restore tab. Create a backup CD for the customer by pressing the Backup
button and inserting a blank CD-R when prompted. When backup is completed, select Exit,
then Logout. Label the CD as “SecurViewRT Database Backup,” with today’s date.
5.0 Other Functionality
5.1 Version Upgrade
For a version upgrade of the multi-workstation system, all workstations and the manager
have to be upgraded at once.
• All clients and the manager have to Exit to Windows.
• Then the version upgrade has to be performed on all computers.
• Reboot the multi-workstation system, starting with the manager.
• The upgrade procedure on every machine is the same as for a stand-alone workstation.
5.2 Backup and Restore
On multi-workstation clients, only exported images and DICOM files can be stored on a CD.
• The checkboxes for backup and restore are disabled on the clients.
• System backup and restore of the database can only be performed directly on the manager.

Note… Backup and Restore procedures are in the SecurViewRT Technologist


Workstation User/Administrator Guide.
5.3 System Time Differences
If the system time of a multi-workstation manager and the clients in a multi-workstation
environment differ, this might lead to problems.

Note… Use the Network Time Protocol to synchronize manager and client times.
See step 4 in Section 3.2, p. 4-9.

4-12 MAN-00344
Installation Manual
Chapter 5—Configuring the Barcode Scanner
Supported Type

Chapter 5—Configuring the Barcode Scanner

1.0 Supported Type


The supported Barcode Scanner is: Symbol LS1900.

2.0 Configuration
The scanner has to be configured for the same country code as the keyboard (American for
example).

2.1 Emulate Keypad


You must configure the scanner for “Emulate Keypad.” This feature allows emulation of other
country keyboard layouts not listed in the Country Selection parameter. All Microsoft
operating systems typically support this feature. When enabled, all characters are sent as
ASCII sequences over the numeric keypad. For example, ASCII A is sent as:
“ALT make” 0 6 5 “ALT Break”
To enable the Emulate Keypad feature, scan the following barcode:

Figure 5-1: Enable Keypad Emulation

2.2 Beep Volume and Sound


The beep volume and sound may be configured as desired. Use the following bar codes to
enable or disable this feature.

Figure 5-2: Beep After a Good Decode

Figure 5-3: Do Not Beep After a Good Decode

MAN-00344 5-1
Installation Manual
Chapter 5—Configuring the Barcode Scanner
Configuration

2.3 Changing Between Patient ID and Accession Number


The BarCodeScanner.ini is provided as an additional configuration file for the SCR system.
This file is located in the e:\mbc-scr folder.
This configuration file contains the following lines:
#MODE = PATIENT_ID # remove the first # of this line if Patient IDs are to be used.
# MODE = ACCESSION_NUMBER # remove the first # of this line if accession numbers
are to be used.

# do not change this line


RECEIVER=LIGHTBOX

MODE sets if accession numbers or Patient IDs are transferred/scanned.

RECEIVER sets which module gets the result of the scan. Only LIGHTBOX is currently
supported.

5-2 MAN-00344
Installation Manual
Appendix A—Technical Characteristics

Appendix A—Technical Characteristics

This section lists the relevant technical characteristics for the stand-alone or client
workstation.
Operating System Windows 2000 Professional with Service Pack
3 or greater.
Computer/Memory Dual Pentium Processors with clock rate 2
GHz or higher; 2 GBytes of 800 MHz ECC
RAM or greater.
Disk capacity >140 GB Image Storage
Archival Media CD-RW with a read rate of at least 24X.
Graphic Controller Card(s) 3 or 5 MegaPixel 8-bit data output.
Monitor 5 MegaPixel, grayscale, Landscape, 2560 x
2048 pixels
3 MegaPixel, grayscale, Landscape, 2048 x
1536 pixels
Network Interface (NIC) 10/100Base-T Ethernet
DICOM Functionality Provides the following DICOM functionality
Verification: SCU
Store: SCP
Query and Retrieve: SCU
Print: SCU
Printing User may select an image and send it to a
validated DICOM printer. The system prints
the image with all the current Window/Level
settings, zoom factor, and displayed
annotations or a diagnostic image (as
selected).

MAN-00344 A-1
Installation Manual
Appendix A—Technical Characteristics

A-2 MAN-00344
Installation Manual
Appendix B—Troubleshooting
Network Problems

Appendix B—Troubleshooting

1.0 Network Problems


Network problems can occur on the client as well as on the manager side of a multi-
workstation.

1.1 Network Problems on a Client


Network problems can occur at system startup or during normal operation.

1.1.1 Problem occurs during System Startup


If a network error occurs during system startup, the multi-workstation manager cannot be
reached. A startup dialog appears, which allows you to shut down the system or to log the
user off from the operating system.
Every five seconds. the system tries to reach the multi-workstation manager until either the
manager reconnects or the user clicks Shutdown or the Log off from Operating System
button in the system startup dialog.

Figure B-1: Startup Dialog

If the network error has occurred due to a misconfiguration of the manager address on the
client side:
• The Microsoft Windows “Administrator” user can change the address using the separate
tool setmwserver.exe.
• This tool allows you to reconfigure the system type and to set the manager address.

1.1.2 Problem occurs during Normal Operation


If network problems occur, and the manager cannot be reached by a client any
more, this client is inoperable for the time of the network problem. In this case,
perform the following measures on the client (in the stated order):
1. Try to re-establish a connection to the manager after a given time-out has
expired. Time-out and number of retries are configurable in lightbox.ini on the
client using the variables MWS_RetryTimeOut (default: 1) and
MWS_RetryCount (default: 2).

P/N MAN-00344 B-1


Installation Manual
Appendix B—Troubleshooting
Network Problems

2. If the connection could not be re-established, the system displays an error


message and displays a button that allows a manual retry to re-establish the
connection to the manager:
If the user manually tries to re-establish the manager connection, and the
network problem has been removed, two cases must be considered:
• The manager has not yet cleaned up the sessions and locks of the client. In
this case, the user can continue working.
• The sessions and locks on the manager are cleaned up. In this case, the user
is logged out on the client, and all work on the current patient is lost.
If the user confirms the error message, or after the auto-logout time has expired,
the user is logged out from the system. If the network problem still exists the
next time a user wants to log in, the login fails with a message that states the
problem.
In all cases, where the user is automatically logged out due to network
problems, a message box is displayed that informs the user about the loss of
work on the current patient.

1.2 Manager
If the manager encounters a network problem concerning a client, the manager takes the
following measures (in the indicated order).
1. Tries to re-establish a client connection until a given time-out has expired (configurable
in lightbox.ini on the manager).
2. After the time-out has expired, closes all sessions and releases all locks for the
problematic client. This means that all work on the current patient gets lost for the user
of that client.

B-2 P/N MAN-00344


Installation Manual
Appendix B—Troubleshooting
Troubleshooting Table—Reviewing Module

2.0 Troubleshooting Table—Reviewing Module


Table B-1: Reviewing Module

Symptom Possible Cause Solution


Images are not displayed Images are not transferred Verify the connection to and DICOM settings on
AWS (current images).
Verify the connection to and DICOM settings on
PACS (prior images).
Verify the DICOM conformance statement of PACS
(prior images).
Images are not processed Confirm hard disk space and RAM.
Images are not stored Confirm hard disk space and RAM.
Image resolutions are not stored Possible Image Preparer failure; review the
corresponding image file (subdirectory Images)
sizes.
Confirm hard disk space and RAM.
Images are not available View images, e.g., via View Now. If the message
“Loading images” appears, wait until the images
appear and the message disappears. If the
message “Missing images” appears, resend images
from DICOM entity.
Images are not loaded Confirm hard disk space and RAM.
Verify the hardware (RAM, CPU, hard disk).
Verify the graphics cards and drivers.
Images are displayed Incomplete image transfer Verify the connection to and DICOM settings on the
incompletely SecurViewRT Diagnostic Workstation (current
images).
Verify the connection to and DICOM settings on
PACS (prior images).
Warning Message Appears Warning! Cannot load requested Reselect that patient. Use View Now to enforce the
image examination of prepared images.
Automatic Hanging of Images Standard Images (MLO and CC) For automatic hanging of standard images, DICOM
MG service class is needed.
Other images Advanced hanging has to be selected in User
preferences.
Message “Missing Images” The connection with PACS is Verify the connection to the PACS. Make sure the
does not disappear interrupted/broken. PACS is running and working properly. Try again a
View Now / Scheduling of the patient
Images are transferred only to the The SecurViewRT Technologist Workstation is not
SecurViewRT Technologist an image archive. If some images of a Patient are
Workstation, and not stored on not available from the PACS, ensure that all images
PACS. are sent to a PACS.

P/N MAN-00344 B-3


Installation Manual
Appendix B—Troubleshooting
Troubleshooting Table—Reviewing Module

B-4 P/N MAN-00344


Installation Manual
Index

Index

A F
about this manual P-1 final checks
multi-workstation 4-12
B stand-alone 4-12
bar code G
beep volume and sound 5-1
emulate keypad 5-1 glossary P-2
barcode scanner 5-1
beep volume and sound 5-1 H
C hardware setup 3-1

cable connections, monitor 3-3 I


calibrating
image export 4-6
LCD display 4-7
image processing configuration 4-6
characteristics, technical A-1
installation manual, using P-4
client 3-1
installing
workstation 3-1
client workstation 3-1
complaints, product P-1
stand-alone workstation 3-1
configuring
barcode scanner 5-1 L
image processing
deleting images 4-6 LCD display, calibrating 4-7
image export 4-6 licensing 4-6
multi-workstation 1-2, 4-5
client 4-5
M
manager 4-5 manager 1-1
standalone 4-5 manual set P-4
workstation 4-1 Manuals P-4
converting stand-alone workstations to clients 4-9 monitor cable connections 3-3
multi-workstation
D configuration 1-2
definitions P-2
deleting images 4-6
N
destination tab 4-4 network problems B-1
DICOM network problems on a client B-1
destination tab 4-1, 4-4 node types and tasks 1-2
printer tab 4-1, 4-3 normal operation problems B-1
settings 4-1
setup tab 4-1–4-2 P
E printer tab 4-3
product complaints P-1
emulate keypad 5-1

MAN-00344 i
Installation Manual
Index

Q T
quality control P-1 technical characteristics A-1
technologist workstation 1-1
R terms and definitions P-2
training P-1
receiving 2-1
troubleshooting
receiving and unpacking 2-1
manager B-2
receiving instructions 2-1
network problems B-1
S normal operation problems B-1
reviewing module B-3
SecurViewRT Manager 1-1 system startup problems B-1
Service Support tables B-3
phone number 0-i
setup of hardware 3-1 U
setup tab 4-2
unpacking 2-1
setup windows 4-1
using the installation manual P-4
specifications A-1
stand-alone 3-1 W
pre-installation 3-1
workstation 3-1 windows setup 4-1
startup 4-1 workstation
system startup 4-1 configuring 4-1
system startup problems B-1 technologist 1-1

ii MAN-00344

You might also like