ImageChecker® C / D CAD Server

Release Notes, v8.3


R2 Technology, Inc. R2 Technology is dedicated to product innovation
1195 W. Fremont Avenue and customer satisfaction. As such, we will notify
Sunnyvale, CA 94087 you when we introduce features or product changes
techsupport@r2tech.com that affect the way you use our products.
For Europe/South America/Asia: Questions? Contact the Technical Assistance Center.
Contact your local distributor. Tel: 866-CHECKED (243-2533) Fax: 408-481-5607

What’s New

ImageChecker C v8.3 CAD Server


ImageChecker C v8.3 is a server computer, intended EmphaSize™. All sites with v8.3 ImageChecker Mam-
specifically for the Citra Mammography Applications mography Systems can now elect to display variable-size
Suite, that uses R2’s Gold Standard CAD software to marks on R2 Technology workstations and display units.
analyze digital mammography images for display on a When the algorithm determines that a region is more
Citra Workstation. The new v8.3 software offers significant, it shows a larger CAD mark, indicating that
R2 Technology’s latest features, including EmphaSize the region should receive more attention from the radi-
CAD marks, PeerView Digital, and the choice of three ologist. EmphaSize is designed to draw the radiologist’s
algorithm thresholds, independently selectable, for calci- attention to those regions containing mammographic
fications and masses. ImageChecker C v8.3 can process features the algorithm deems to be most indicative of
images from an expanded range of digital mammogra- malignancy.
phy (FFDM) devices. ImageChecker C v8.3 CAD servers
offer: PeerView Digital is an optional feature available on
some digital review workstations that enables the radi-
Improved Algorithm Performance. The new software ologist to better understand why a region of interest was
provides an improved algorithm with higher sensitivity at marked. PeerView Digital highlights individual calcifica-
any given false-mark rate, solidifying R2 Technology’s tions and outlines the central density of each mass,
position as the market leader. The software now offers thereby helping the radiologist visualize and analyze fea-
three operating points (thresholds) when processing ana- tures that may indicate malignancy.
log (film) or digital images to accommodate differing radi-
ologist preferences. The performance of the system for RightView is an optional feature used to create ‘for
each of the three operating points as measured on presentation’ images from GE ‘for processing’ images.
R2 Technology’s film database is summarized in the fol- RightView images are high-contrast, similar to those pro-
lowing table. duced by other popular processing algorithms.

Operating Points 0 1 2 Support for Additional Digital Mammography Ven-


dors. ImageChecker Mammography Systems now ac-
Calcification Sensitivity 95% 96% 97%
cept images from additional digital mammography ven-
Mass Sensitivity 83% 88% 90% dors. The current list of vendors includes AGFA (outside
False Marks per Case 1.0 1.5 2.0 the U.S. only), Fujifilm, GE Medical Systems, Fischer Im-
Specificity 48% 35% 24% aging, Hologic, Inc., LORAD, and Siemens.

Thresholds for masses and calcifications can now be set


independently. For complete specifications, refer to the
user manual shipped with your system.

PN 14377 Rev A, July 2006 Page 1 of 2


ImageChecker® C / D CAD Server
Release Notes, v8.3

ImageChecker D v8.3 CAD Server


ImageChecker D v8.3 is R2 Technology's newest high-
end digital CAD server, functioning as a stand-alone unit
that interfaces with an expanded range of digital mam-
mography (FFDM) devices (see previous paragraph). It
offers the new features available with ImageChecker C
v8.3, including EmphaSize CAD marks, PeerView Digital,
and the choice of three operating points, independently
selectable, for calcifications and masses.

Known Issues
6707, 6712: Cases fail if processing unit 7152: Unable to delete all cases from 7275: No explanation when case ‘Failed’
is rebooted before they are transmitted to a Queue Manager (‘connection refused’) error. with incorrect output format. If DICOM cases
printer or display unit. In the event of a net- To delete cases in the Queue Manager, delete are sent to a system that does not support the
work interruption, cases will queue up for no more than 100 at a time. No fix is planned. requested DICOM Service, the case will fail
transmission on the ImageChecker Processing but no explanation will be provided. Call R2
Unit for up to 24 hours. It is not necessary to 7189: Malc marks appear differently on PU Service to check system configuration. A fix is
reboot the ImageChecker after network con- and CMU. There is a slight difference in the planned for the next scanner release.
nectivity is restored. Doing so will cause all appearance of the Malc marks between the
queued cases to fail. No workaround is avail- processing unit printouts and the display unit 7302: Vidar scanner autocalibration causes
able for this defect. No fix is planned. screens. A fix is tentatively planned for the failure of scanning case. If one scans cases
next release of CheckMate Ultra and Mam- simultaneously with scanner’s auto calibra-
7061: Multiple alerts upon transfer of molux software. tion, the case being scanned will fail. The
failed GE case. A GE case that fails to trans- default is OFF, and if it does get turned ON,
fer to the configured destination will generate 7198: ‘Unknown’ Study Time on CAD SC an R2 Technology service representative
one alert for each image in the case. There is images at PACS. R2 does not define the should reset the time at 3 AM (as an example)
no immediate plan to fix. Study Time for DigitalNow cases. No fix is using the service tool.
planned.
7084: System does not retain repeat 7325: Automatic window leveling for Digi-
cases. For digital cases, the user interface 7199: Incorrect Scan Date/Time on print- talNow needs adjustments. Automatic cali-
will display only the most recent study in each outs for reprocessed cases. PU printouts bration of Window and Level may not satisfy
series. Identical cases sent more than once generated from reprocessed images will con- your preferences. Contact R2 Service for
will cause the earlier case to be overwritten. tain a date of 1970-01-01 or 1969-12-31. manual adjustment of calibration parameters.
No fix is planned. No fix is planned. A fix is planned for future release.