You are on page 1of 4

Tips and Tricks from Joe Flow

Rheometers: How and What Do They Measure?


To understand rheological measurements, it is often helpful to learn more about the rheometer
setup and how it works.

Rheometer design

Rotational and oscillatory rheometers are set up as


shown in Figure 1. The measuring head contains the
drive motor and encoder which measure and set the
torque, deflection angle and speed.

What is measured in rotational tests?

Let's imagine that the measuring gap contains


ketchup. The viscosity of the ketchup should be
measured with a shear-rate-controlled rotational test.
You set the shear rate profile. The viscosity is
calculated according to the viscosity law from the
quotients of the shear stress  and shear rate γ̇ .
η = τ / γ̇
As you already know, the shear stress correlates with
the torque. If you measure the ketchup with a large
To measure the rheological properties of your sample measuring system, a large torque is required because
with a rheometer requires determination of the torque the shear surface is large. If you use a small
and deflection angle of the measuring bob. Your measuring system, a smaller torque is needed to
sample provides resistance or a preset torque to the obtain the required speed.
presetting. The resulting deflection angle is then However, the viscosity of the ketchup is the same in
measured by the rheometer's encoder. The speed is both measuring systems. The rheological parameter
calculated from the deflection angle and the time. shear stress  is calculated from the torque in relation
However, as the user, you are usually more interested to the shearing surface. Therefore, the shear stress is
in the rheological parameters. How do you get these? independent of the used measuring system. It is
Rheological parameters are calculated from the easier to work with the rheological parameters.
measured values torque, deflection angle and speed The same is true for the shear rate, which correlates
using conversion factors (see Table 1). All further with the speed. If ketchup is measured in two different
parameters, such as viscosity, are not measured but measuring systems, one with a large shear gap and
calculated. one with a narrow shear gap, the same rotational
speed results in different flow velocities (or shear
Conversion rates). As with the calculation of the shear stress, for
Measurement Rheological
values factor parameter
the calculation of the rheological parameter shear
rate, the influence of the measuring system is
Torque M [mNm] CSS factor Shear stress τ [Pa] considered using a conversion factor.
Deflection angle φ
CSD factor Deformation γ [%] Viscosity values are calculated from the rheological
[μrad]
parameters which are independent of the instrument.
Speed n [min-1] CSR factor Shear rate γ̇ [s-1]
Table 1: Measuring values and rheological parameters with the
conversion factors

XRRIA012EN-B 1 www.anton-paar.com
Figure 1: Setup of a modern air-bearing rotational and oscillatory rheometer with the most important components

However, every instrument has its limits (see the Tips What is measured in oscillatory tests?
and Tricks from Joe Flow on the subject of measuring
range limits). To still be able to cover as large a
viscosity range as possible, there are a number of If jello is in the shear gap, the aim of the test is usually
measuring systems available which all have their own to investigate the viscoelastic behavior and the
conversion factor. That makes it possible to measure characterization of the undestroyed structure at rest.
almost all samples with one rheometer - from low In this case it is not very useful to carry out a
viscosity liquids to stiff solids. rotational test. Instead, oscillatory tests with set
These conversion factors are saved together with the sinusoidal oscillation of the measuring system and the
inertia, the expansion coefficients, the serial number sample are carried out. Also here, the rheometer
and the exact dimensions of the measuring system for determines the deflection of the measuring bob and
each system on the transponder chip in the measuring the required torque. However, because jello has
system axis. Using ToolmasterTM which provides fully viscoelastic properties, the response wave is time-
automatic detection of the measuring system and delayed compared to the set oscillation. This time
measuring cell, all relevant data are read out when the delay is called the phase shift δ.
measuring system is connected to the coupling of the The values measured by the rheometer (deflection
rheometer. This ensures that the software always angle, torque and phase shift) together with the
uses the correct conversion factors for the used conversion factors for the measuring system now give
measuring system for the calculation of the all necessary data to calculate the required
rheological parameters. rheological parameters such as the storage modulus
G’ or loss modulus G’’.

XRRIA012EN-B 2 www.anton-paar.com
What else does a rheometer measure? material has a certain compliance, which needs to be
known and compensated for. Therefore, even gap
change in the micrometer range must be detected and
The measuring temperature, which is an extremely
corrected or actively adjusted, even when the normal
important parameter, is measured in the measuring
forces are large. Especially when using a cone-plate
cells as well. Exact temperature control is very
geometry, it is very important to precisely control the
important because all rheological parameters are
measuring gap (to keep the gap constant).
greatly affected by the temperature.
In torsion tests, the measuring systems and also the
Modern rheometers simultaneously detect the axial
rheometer are affected. For example, the measuring
normal force FN. An MCR series rheometer does this
system shaft may twist by a small amount when the
using a normal force sensor integrated in the air
torque is applied because the sample in the
bearing. The normal force is also used to
measuring gap resists deflection. The challenge for
automatically determine the zero position of a cone-
the electronics is to detect the real deflection and
plate or parallel-plate measuring system. In the past,
speed of the sample from the signals coming from the
the zero position and measuring position had to be set
encoder in the measuring head. There is also a
by hand. Today, this is done for you automatically by
compensation value for this called torsional
the rheometer.
compliance. The torsional compliance is known for
If you carry out temperature tests or curing, e.g. with a each measuring system and can therefore be
parallel-plate system, the gap becomes bigger or corrected.
smaller due to thermal expansion or shrinkage of the
measuring system and sample. In this case, it is
recommended to set a constant normal force of, e.g.
FN = 0 N so that the measuring head compensates for Adjustment of parameters
the expansion or contraction using an intelligently
controlled procedure. This ensures that the measuring
At first, the rheometer does not know whether there is
gap is always filled, even if the sample shrinks and
a low viscosity, high-viscosity, ideally viscous or
that no large normal forces occur if the sample
viscoelastic sample or an elastic solid in the
expands. In both cases the measuring gap changes.
measuring gap. It is therefore important to use
However, because the MCR rheometer continuously
intelligent controllers which can adapt quickly to the
detects the measuring gap, the actual gap size at
sample conditions. The TruRate™ motor controller of
every measuring point is used for further calculations.
the MCR series rheometers controls the sample
Besides the benefits of easier handling, this opens up deformation, shear rate and shear stress exactly,
completely new measuring possibilities, such as: quickly and without overshoots so that all types of
• dynamic-mechanical thermal analysis (DMTA) on samples can be investigated precisely.
solids (torsion bars), When you set a shear rate, the rheometer controls the
• tack and penetration measurements (see the Tips speed appropriately. This is kept constant, unaffected
and Tricks from Joe Flow on this subject), by the behavior of the sample in the measuring gap.
• breaking tests (or fracture tests), as carried out by You should always remember this when interpreting
texture analyzers, the data. If your sample leaves the gap due to
• tribological measurements (wear measurements viscoelastic effects then the adaptive controller
between two media), adjusts to the set shear rate or speed. However, now
• determination of the amount of shrinkage (e.g. in that there is less material in the gap, this results in a
percent of the initial value). lower required torque. In the flow curve, you can
In the past, individual instruments were used for each usually see immediately that the shear stress function
of the mentioned test methods. Now, due to the suddenly becomes flatter or curves downwards. In the
constant development of instrument types, it is fall of the viscosity curve, however, this effect is often
possible to cover a number of measuring tasks with difficult to distinguish from shear-thinning behavior.
one instrument. As you see: modern rheometers The real-time position control TruStrainTM was
measure considerably more than just viscosity values developed for oscillatory measurements with set
and viscoelastic properties. They have become deformation. It allows these tests to be carried out
universal test instruments. without pretests and without subsequent adjustments
by the instrument control. The strain (deformation) is
adjusted in a real-time position control mode directly
The stiffness of a rheometer
on the (ideal) sine wave both within the linear
viscoelastic (LVE) range and also under LAOS
For measurements on curing materials and on solids it conditions (Large Amplitude Oscillatory Shear), i.e.
is important for the rheometer to be extremely stable, outside the LVE range. This makes oscillatory
both mechanically and thermally. However, each measurements possible at the smallest torques and

XRRIA012EN-B 3 www.anton-paar.com
deformations, even on complex liquids such as Modern controllers adapt quickly to the behavior of
suspensions, emulsions, foams and surfactant your sample and allow precise measurements, even
solutions as well as gels. These oscillatory of samples with low structural strength, without
measurements are highly accurate and without drift. destroying them prematurely.
However, despite this highly developed rheometer
technology, you, as the user should always confirm
the measurements with a critical eye and check the
Ball bearing or air bearing? What do I need?
results, as a rheometer does not detect external
sources of error, for example, gap emptying.
There are rheometers with air bearing and rheometers
with ball bearing. What does that mean? As far as
instrument handling and sample handling are
concerned, there is no difference. However, the
instruments differ significantly in the internal friction of
the motor bearing and therefore in the smallest torque
which can be measured or usefully analyzed.
Fig. 2 clearly shows the different measuring range
limits for the viscosity when using an air-bearing
rheometer and a ball bearing rheometer with the same
cylinder system.
The type of rheometer bearing also has an effect on
the type of tests which can be carried out. Oscillatory
tests usually only provide useful results when carried
out on air-bearing rheometers.
The benefits of ball-bearing rheometers include their
lower purchasing costs and lower operating costs, as
no compressed air is needed. Many applications and
investigations require a higher resolution and
oscillatory measurements. This increasingly results in
users choosing air-bearing rheometers.

Figure 2: Theoretical viscosity range of rheometers with air-


bearing and ball-bearing when using a cylinder measuring
system, shown on double logarithmic scale

Summary

A rheometer measures the torque, deflection angle


and speed with very high accuracy. All required
rheological parameters are calculated using these
measured values together with the measuring-system-
dependent conversion factors. The temperature and
normal force are also determined.
As well as being very stiff, a rheometer should also Contact Anton Paar GmbH
have an exact gap control. www.anton-paar.com/contact

XRRIA012EN-B 4 www.anton-paar.com

You might also like