NV8
Neonatal Ventilator
0197
P/N:EN-NV8-8P-20210602-V1.4
Nasal Continuous Positive Airway Pressure
For most of the premature infants, their respiratory system is under
Paw Apnea wakeup development. This will lead to irregular breath and apnea.
If not handled properly and timely, it will influence infants’brain
development. And in some severe cases, apnea will threaten infants’lives.
Paw
Nasal Intermittent Positive Pressure Ventilation
NIPPV mode can provide PIP which is higher airway pressure, and PEEP
which is relatively low airway pressure in setting frequency.
Compared with NCPAP mode, NIPPV mode can produce higher mean
airway pressure.
It will benefit the infant’s oxygenation and ventilation, reduce infant’s work
of breathing (WOB), and increase functional residual capacity (FRC).
NIPPV Waveform
Ventillation Modes
The system will switch to preset ventilation mode when an infant’s
breathing is not detected.
SNIPPV Waveform
Comen NV8 is equipped with various ventilation modes (with synchronously triggered signal display)
(including SNIPPV NIPPV NCPAP and HFNC), which enable
precise and suitable ventilation for a patient’s different stages.
Features with apnea rescue function, NV8 can handle with
infant’s apnea risk, relieving the heavy workload of clinicians.
With iFlow closed-loop pressure control system and automatic
Paw
leakage compensation system, NV8 can realize precise flow and High Flow Nasal Cannula
pressure control even in high flow ventilation. Compared with traditional oxygen therapy ventilators, NV8 features with
NV8 is designed to provide user-friendly operation experience pressure monitoring function. By this, doctors can monitor and adjust the
increase the efficiency of medical staff. infant’s airway pressure in real-time. This feature will greatly reduce the
risk of high airway pressure, which always leads to respiratory
system harm to infants.
HFNC Waveform
01
Pressure generator
Traditional CPAP device produces certain Traditional CPAP device mixes O2 and
pressure by adjusting flow rate manually. medical air mechanically, while NV9 NV8 system Traditional CPAP device
Unlike traditional CPAP device, NV8 reaches adopts digital O2/Air blender. NV9 will
to and maintains target pressure with help of automatically adjust the ratio of O2
airway pressure airway pressure
real-time near nose pressure monitoring as and Air with the help of built-in flow
well as automatic pressure compensation, sensor and flow valve, and reach to
which greatly reduces the workload of the target O2 concentration.
caregiver.
rea
l-ti
m
ef
em
tim
on
al-
ito
re
rin
g
iFlow
Flow control
Patient
module
Near-nose pressure
monitoring:
In NV8, the airway
pressure is measured at
closed-loop adjustment the near-nose end which
will provide more
accurate monitoring
of the airway
pressure.
In NV8, sensors will keep monitoring the airway pressure near the infant’s nose. The data will be feedbacked to
CPU for closed-loop adjustment.
By this system, NV8 can compensate for the leaking gas and guarantee stable pressure ventilation.
02 03
User-friendly Design
Oxygen Concentration Calibration: With the POST function, NV8 can 8”LED touch screen,
With special structure design, remind the clinicians timely when Easy to operate and observe.
Comen NV8 can conduct oxygen there is any problem with the
concentration calibration during the machine.
ventilation.