+ 28 V Stepper Motors
B Probe Valve M •1-1
•1-1RBC/PLT
RBC/PLTTransducer
TransducerDrain
Drain
•1-2
•1-2RBC/PLT
RBC/PLTCount
Count
C Diluent Syringe M •1-3
SDM 1 •1-3RBC/PLT
RBC/PLTMetering
MeteringTube
TubeVent
Vent
•1-4
•1-4RBC/WBC
RBC/WBCIsolator
IsolatorVacuum
Vacuum
D Probe Rotate M MPM •1-5
•1-5RBC/WBC
RBC/WBCIsolator
IsolatorDrain
Drain
M •1-6
•1-6RBC/WBC
RBC/WBCBack
BackFlush
Flush
E Probe Up.Down
•1-7
•1-7Spare
Spare
•1-8 Spare
•1-8 Spare
F Sample Syringe M WBC
Vacuum Transducer •2-1
•2-1WBC
WBCMetering
MeteringTube
TubeVent
Vent
•2-2
•2-2Pre-Mixing
Pre-MixingCup
CupBubble
BubbleMix
Mix
Back Flush
•2-3
•2-3RBC
RBCMixing
MixingChamber
ChamberBubble
BubbleMix
Pressure
PUMP
PUMP
Mix
PUMP SDM2 •2-4 RBC/PLT Transducer Fill
Regulator
Regulator
•2-4 RBC/PLT Transducer Fill
•2-5
•2-5RBC/PLT
RBC/PLTMixing
MixingChamber
ChamberDrain
Press
+/-12V +29/+14
VAC
Drain
RBC
MPM Power
•2-6
•2-6HGB
HGBReference
Reference
Transducer •2-7
•2-7HGB
HGBSample
Sample
Pressure HGB •2-8 Spare
•2-8 Spare
Flow Cell
PUMP
RELAY
•3-1
•3-1Probe
ProbeWash
Wash(NC)
(NC)
•3-2
•3-2Probe
ProbeWash
WashDrain
DrainVacuum
Vacuum
Note: +28Vdc fans •3-3
are variable •3-3 RBC/PLTMixing
RBC/PLT MixingChamber
ChamberWash
Wash
•3-4
•3-4Pre-Mixing
Pre-MixingCup
CupWash
speed PREAMP •3-5
•3-5WBC
WBCTransducer
Wash
TransducerDrain
Drain
SDM3
•4-1
•4-1Spare
Spare
•4-2
•4-2Spare
Spare
•4-3
•4-3WBC
WBCCount
Count
•4-4
•4-4WBC
WBCTransducer
TransducerFill
Fill
•4-5
•4-5 HGB Sample/WBCMix
HGB Sample/WBC MixChamber
ChamberDrain
Drain
SDM4
100V, PreAmp
PreAmp •4-6
•4-6Lyse
LyseInlet/WBC
Inlet/WBCMix
MixChamber
ChamberDrain
Drain
Power Supply
PreAmp +12 V ,
•4-7
Filter •4-7 WBC Mix Chamber BubbleMix
WBC Mix Chamber Bubble Mix
Switching
PreAmp
AC WHT Filter •4-8
•4-8WBC
WBCMixing
MixingChamber
ChamberWash
Wash
28 V
28v
GND
DC
DCRegulator
RegulatorBoard
Board
120 V AC
•5-1
•5-1Wast
WastBottle
Bottle11Inlet
Inlet
Back Plane Power +5/+/-12V •5-2
•5-2Waste
WasteBottle
Bottle11Vacuum
Vacuum
SPM •5-3
•5-3Waste
WasteBottle
Bottle11Pres/Drain
Pres/Drain
SDM5 •5-4
•5-4Vacuum
VacuumTransfer
Transfer
SPM CCM •5-5
AC
ACSense
Sense •5-5Waste
WasteBottle
Bottle22Inlet
Inlet
Bus •5-6
AC •5-6Waste
WasteBottle
Bottle22Vacuum
Vacuum
•The AC Regulator PCB or ACRegulator
RegulatorPCB
PCB •5-7
•5-7Waste
WasteBottle
Bottle22Pres/Drain
Pres/Drain
AC Sensor Board CCM •5-8
•5-8Lyse
LyseInlet
Inlet(NC)
(NC)
Floppy, Hard Dr
Document Control Number 206316-101 ©2010 Abbott Laboratories, Abbott Park, Illinois
VOLTAGE CHECKS VP-6, VP-7
The following table lists the test points located on the PAM.
Location Function Voltage
TP5 Analog Ground N/A
TP4 +100V +100V ± 0.5V
TP7 HGB Self Test/Gain Adjust +1.84V ± 0.02V (Self Test)
VOLTAGE CHECKS
+5.2V ± 0.2V (Gain
Card Cage Back Plane Rear Side
3 LD3 +12VDC
(ANALOG) The following table lists the test points located
4 LD2 -12VDC
on the Cable Distribution Module (CDM).
(ANALOG) Location
5 LD1 +5VDC
Pin Function Voltage
(J5 and J8)
Dilution factor password – 94043
SRP GAIN VERIFICATION/ADJUSTMENT VP-09 SPM Adjustment 1 Brown Ground < 0.05V
CD1800 jump to DOS: diag menu,
press “esc” then “alt + d” : to return 2 Blank N/A N/A
– type “exit” at : Method SPM POT Target Channel Range 3 Orange Ground < 0.05V
R22 56 1 4 Yellow +14V ±0.5V
CD1800 exit to DOS: diag menu, 5.0 WBC
press “esc” then “alt + x” : to return 5 Green +28V ±1.00V
– cycle power
S RBC R27 98 1 6 Blank N/A N/A
R 5.0
On the CDM, +28V is used to initially energize
P 5.0 RER R51 See Chart solenoids and +14V is used to keep them closed.
3.31 PLT R41 136 1
Open M. Precision <2.5% <1.7% <1.2% <1.5% <6.0% PRESSURE 1 0.55 psi +/-0.0.2
Open M.Carryover [(Background(1)– Background(3)) / (Last whole blood of precision - Background(3))] X 100 PRESSURE 2 >4.0 PSI
Open Carryover Limits <1.0% <0.5% <0.8% <1.0%
VACUUM 8.25” Hg +/-0.25
3. The test runs for approximately 30 seconds. RBC Count Time 6.7 +/-0.2