org/journal/acsodf Mini-Review
Figure 1. MR images of a rat brain with an ischemic lesion, obtained at 9.4 T with different image contrasts. (a) Proton density image (F1 ≈ F2 ≈
F3 ≈ 1), (b) T1-weighted image (F1 = 0.125; F2 ≈ F3 ≈ 1), (c) T2-weighted image (F2 = 0.135; F1 ≈ F3 ≈ 1), and (d) diffusion weighted image
(F3 = 0.464; F1 ≈ F2 ≈ 1).
Figure 2. Parametric MR mapping. Color-coded T1 (left), T2 (center), and ADC (right) parametric maps on which each pixel represents the
absolute value of the longitudinal (T1) and transverse (T2) relaxation times, or the apparent diffusion coefficient (ADC), calculated for a rat brain
with an ischemic lesion.
particular pulse sequence and set of imaging parameters; they local disturbances of the magnetic fields, etc.) that influence
play around with the extrinsic parameters to modulate the signal intensities, and contrast on MR images. Thus, direct
weighting factors and thus modulate signal intensities of comparison of images across different experimental setups is
different ROIs in an MR. complex, resulting in typical evaluation of MR images at a
Considering only the three main mechanisms of image qualitative and subjective level, usually restricted to descrip-
contrast in MRI (T1, T2 and diffusion), one can yield four main tions of relative contrast between different anatomical regions
sorts of images or combinations of these four (Figure 1): (the use of sentences like “presence of a hypo- or a
• Proton density images: (F1 ≈ F2 ≈ F3 ≈ 1) hyperintense signal in the region” are common in medical
• T1 weighted images: (F1 < 1, F2 ≈ F3 ≈ 1) literature). For quantitative and cross-setup comparison of
• T2 weighted images: (F2 < 1, F1 ≈ F3 ≈ 1) MRI data, it is crucial to extract the contributions of the
• Diffusion weighted images: (F3 < 1, F1 ≈ F2 ≈ 1) different contrast mechanisms, to provide absolute-scaled
information on tissue structure and function, independent of
Figure 1 shows the versatility of MRI and how the signal hardware and sequence related parameters.
intensity can be modulated to generate a good contrast
between regions of different intrinsic properties (lesion vs
healthy tissue). Depending on such intrinsic properties, 3. MR RELAXATION TIMES
conditioned by the biological process under study, some The quantitative nature of the MR signal can be used to extract
image modalities show higher contrast than others, or even very useful absolute measurements of biophysical parameters, if
contrast in opposite directions (hyper- vs hypointensity). one understands and decouples the contrast mechanisms that
Contrast between ROIs can be quantified by separating the contribute to the overall MR signal, which can be essentially
individual contributions of the four different contrast achieved through the use of eqs 1−4.
mechanisms (proton density, T1, T2, and diffusion) implicated Relaxation times (T1, T2, or T2*) and apparent diffusion
in the modulation of the SI. coefficients (ADC or D) are the most influential MRI
The interpretation of biomedical MR images requires prior parameters used in quantitative biomedical applications of
knowledge of the influence of the different weighting MRI. These parameters reflect the local nature and
parameters in the image under observation, the pulse sequence physicochemical properties of tissue environments that are
used for image acquisition, and the values of the corresponding directly correlated to the biological properties of such tissues
set of extrinsic parameters used. This provides a thorough (a correspondence that will be discussed later in this work).
understanding of the correlation between the corresponding The term “relaxometry” is used in MRI to describe the
signal contrast and the underlying pathophysiology of tissues process of measuring signal decay (relaxation) after excitation
under observation. For example, a clearly defined ischemic of protons by radiofrequency (RF) pulses, and the extraction
lesion (like the hyperintense region observed in the rat brain in from such measurement of absolute values for the relaxation
Figure 1c) may not be appreciable by unsuitable management times (the intrinsic properties of matter under study). Images
of the weighting factors (i.e., acquisition parameters), resulting representing the spatial distribution of the relaxation times are
in a masked lesion (Figure 1b). called relaxation maps (Figure 2). Similarly, images represent-
On top of the acquisition parameters, there are extrinsic ing the spatial distribution of ADC are called ADC maps
factors like the main magnetic field of the MRI scanner (B0), (Figure 2).
the hardware configuration (design and performance of RF Although relaxation times vary with the external magnetic
coils, magnetic field gradients, adjustment of shimming coils, field used for acquisition (see ref 2 for a practical study),2 T1
36907 https://doi.org/10.1021/acsomega.2c03549
ACS Omega 2022, 7, 36905−36917
ACS Omega http://pubs.acs.org/journal/acsodf Mini-Review
and T2 maps acquired at the same field should be absolute, time is set to at least 5.3 times the T1 value (TR≥ 5.3 T1) when
facilitating the comparison of data irrespective of the system using a flip angle of θ = 90:
and pulse sequence used for their acquisition. However, not
controlled experimental conditions (temperature of the [1 e( 5.3T1/ T1)
] sin 90 5.3
F1 = ( 5.3T1/ T1)
= [1 e ] = 0.995
subject, degree of hydration of tissues, percentage of fat or 1 cos 90[e ] (7)
presence of macromolecules in tissues, etc., all representing
biological variables that differ from one individual to another) In practice, there are several experimental approaches to
may yield different T1 and T2 values across experimental measure transverse relaxation times (T2) based on the
setups. The direct comparison of ADC absolute values across relaxation process described by eq 5 and Figure 3. The most
experimental setups is even more complex since diffusion is an common one relies on the use of a multispin echo sequence
anisotropic property, and obtained values depend on the (Figure 4), in which a series of signal echoes are acquired at
direction of the diffusion gradients applied during acquisition regular intervals of TE (ΔTE = i × TE) by using a series of N
(this is the origin of the word “apparent” in the term ADC). In consecutive “refocusing pulses” (180 deg pulses) following a
fact, it is common practice to estimate ADC values in multiple single excitation pulse (a 90° pulse). This approach is faster
directions of space (as many as 30 or more) for diffusion than acquiring a series of N single images of increasing TE
tensor imaging (DTI) acquisition schemes,3 yielding a mean value (an alternative approach to measure T2). However, it
value for ADC, termed as mean diffusivity (MD), to facilitate requires a good calibration of the 180° pulse and high B1 field
the comparison of data from different studies. homogeneity profile of excitation RF coils to avoid that poor
As we have already mentioned, relaxation times can be refocusing pulses yield in systematic errors propagated through
calculated by decoupling the different contrast mechanisms the N image series.
that contribute to the overall MR signal. For example, if one An alternative to use refocusing pulses, signal echoes can be
acquires a series of MR images with a constant value for F1 (by generated by magnetic field gradients (see ref 4 for further
fixing TR and θ) and F3 (by making diffusional gradients equal details).4 In this case, the calculated transverse relaxation times
to zero or fixing their amplitude and duration), the signal are susceptible to local magnetic field inhomogeneities, and T2
intensity in those images will only depend on the extrinsic is then termed T*2 (which can be substantially smaller than T2).
parameter TE, according to eq 1: This constant includes not only the relaxation due to the pure
T2 effects but also the influence of the so-called susceptibility
S= × F1(constant) × F 3(constant) × e( TE / T2)
= S0e( TE / T2) effects (that we will discuss later). In the case of T*2 relaxation,
(5)
a train of N echoes may also be acquired after a single
excitation pulse (multigradient echo sequence), or a series of N
This equation can be linearized as single-gradient echo images can be acquired using increasing
ln S = ln S0 (1/T2)TE (6) TE values to acquire each one of an N series of images. The
first approach is much faster, but the minimum TE difference
Thus, the transverse or spin−spin relaxation time T2 can be between consecutive steps of an echo train (ΔTE) is limited by
calculated by acquiring a series of N images of increasing echo the hardware (short ΔTE requires high gradient strengths, not
time (TE), fitting the decay of the signal intensity versus TE to always available in all scanners). Detailed information about
an exponential function, or its logarithm to a linear function the issues discussed in the previous paragraph can be found in
(Figure 3). the specialized bibliography.4−6
Likewise, estimation of longitudinal relaxation times (T1)
can be achieved by making constant the weighting factors F2
and F3, and playing around with the factor F1 (as expressed by
eq 2). In this case, F1 depends on two extrinsic parameters that
can be independently varied. Therefore, there are two
experimental procedures to determine T1 values (Figure 5).
In the first approach (the most common one), the flip angle, θ,
and TE are fixed for the acquisition of images of variables
repetition time (TR). Under these conditions, eq 1 becomes
Figure 3. Calculation of T2. Fitting the signal intensity decay on a
series of images acquired with different TE values (at fixed TR and θ [1 e( TR/ T1)
] sin 90
values) provides the absolute value of the transverse relaxation time S= × F 2(constant) × F 3(constant) × ( TR/ T1)
T2. 1 cos 90[e ]
( TR/ T1)
= S0[1 e ] (8)
It is crucial to notice that the fitting parameter S0 represents
Alternatively, one can opt for fixing the repetition time and
the proton density (ρ) of the pixel under analysis, and we can
vary the excitation pulse angle according to
obtain purely T2-weighted images only when the weighting
factors F1 and F3 are equal to 1. The later condition (F3 = 1) [1 e( TR / T1)
]sin
is met using no diffusional gradients (G = 0); however, strictly S= × F 2(const ant) × F 3(const ant) × ( TR / T1)
speaking, it is not possible to fulfill the (F1 = 1) condition for 1 cos [e ]
the T1 weighting factor since we would need to set a repetition ( TR / T1)
[1 e ]sin
time of TR = ∞, or a flip angle of θ = 0 (no excitation pulse, = S0 × ( TR / T1)
1 cos [e ] (9)
meaning no signal creation). Nevertheless, it is usually
assumed that when F1 ≥ 0.995 (F1 weighting factor <0.5%), Again, S0 represents the proton density only when F2 = 1,
the T1 effect on the signal can be considered negligible. and no diffusion gradients are used (G = 0 → b = 0 → F3 = 1),
Mathematically speaking, this is achieved when the repetition a situation that is only achievable when TE = 0. Like in the
36908 https://doi.org/10.1021/acsomega.2c03549
ACS Omega 2022, 7, 36905−36917
ACS Omega http://pubs.acs.org/journal/acsodf Mini-Review
Figure 4. Schematic representation of the multi-spin−echo experiment to measure T2 relaxation. Following the generation of a signal (S0) by an
excitation (90°) RF pulse, the signal decays due to defocusing of spins (process explained later in section 3). If a refocusing RF pulse (180° pulse)
is applied after a period of (TE/2), a so-called echo is generated at the same interval of time (at TE/2 from the refocusing pulse or a total TE
period from the excitation pulse). The process is repeated N times to generate N echo images. Signal decays exponentially from echo to echo with a
time constant of T2 (transverse or T2 relaxation time).
previous case, this last requirement is virtually impossible to Having disclosed the influence of the most common contrast
achieve from an experimental point of view, and it is generally mechanisms in biomedical applications of MRI (see refs 5 and
agreed that when (TE ≤ 0.005 × T2), the T2 weighing factor is 6 for details about other specific contrast mechanisms such as
negligible (F2 < 0.5%), according to magnetization transfer (MT), T1ρ, chemical exchange
saturation transfer (CEST), etc.),5,6 the advantages of
F 2 = e( TE/ T2)
= e( 0.005T2 / T2)
= 0.995 (10) quantitative imaging is more evident. The most prominent
This condition is still challenging to achieve for most advantage is the removal of influences unrelated to tissue
conventional pulse sequences, and proton densities are usually properties, like operator or MR system dependencies, differ-
assessed by using the strategy previously described through eqs ences in scan parameters, spatial variation in the magnetic field,
5 and 7, or by estimating the F1 value prior to the use of eqs 8 and image scaling. Thus, absolute quantification allows the
and 9. comparison of data across different laboratories and exper-
Situations described by eqs 8 and 9, shown in Figure 5, imental protocols, enabling longitudinal and large-scale
correspond to the so-called saturation−recovery experimental multicenter studies on patient cohorts, accounting only for
setup. There are other experimental approaches to calculate T1 intraindividual evolution and interindividual differences.
values based on the use of an inversion pulse of 180 in front of
the image acquisition schemes (inversion−recovery exper- 4. PHYSICAL MEANING OF THE RELAXATION
imental setup) or more complex approaches such as look- CONSTANTS
locker sequences, SSFP (steady-state free precession) From a physical point of view, the relaxation times T1, T2 (or
approach, etc., designed for fast acquisition of T1 values. T2*) are time constants that describe how fast an MR signal,
However, their description is out of the scope of this work, and generated by an excitation RF pulse, decays over time. Their
deeper discussion about different techniques for the determi- values are determined by the intrinsic physicochemical
nation of T1 and T2 relaxation times can be found in properties of matter under study. It is important to understand
specialized literature.5 that the longitudinal (T1) and the transverse (T2 or T2*)
Finally, with respect to the diffusional weighting factor F3, relaxation processes are independent and always occur
the apparent diffusion constants can be estimated according to simultaneously.
eqs 11 or 12, by fitting the decay of the signal intensity versus The longitudinal relaxation time (T1) is a time constant
variable b values, while keeping F1 and F2 constant (fixed TE, describing the return of observed nuclei from an excited
TR, and θ): (higher energy) state, achieved by an excitation RF pulse, to its
S= × F1(constant) × F 2(constant) × (e bD
) = S0 e bD equilibrium (lower energy state), by dissipating the received
energy to the surrounding molecular lattice (T1 is also called
(11) spin−lattice relaxation time).
or its linearized version: If one considers the exited protons of a single water
ln S = ln S0 Db molecule, their return to the equilibrium state (recovery) is
(12)
basically influenced by neighboring molecules in the
Plots of S or ln S versus b are like those presented in Figure surroundings, which behave as fluctuating magnetic fields. In
3 for T2 decay. In practice, different b values are achieved by bulk water, the surrounding water molecules of the one under
using different diffusion gradient strengths, G, in eq 4 for the observation behave as “little magnets” in permanent move-
acquisition of images with different F3 weighting factor values, ment, tumbling as well at different rates (characterized by
used to extract D (or ADC) values from the fitting of different rotational correlation times, τc) (Figure 6a). These
experimental data. moving and tumbling molecules around the observed one
36909 https://doi.org/10.1021/acsomega.2c03549
ACS Omega 2022, 7, 36905−36917
ACS Omega http://pubs.acs.org/journal/acsodf Mini-Review
Figure 7. Signal lost by incoherence. Left, a sinusoidal signal of amplitude A1 (purple line) built up from three coherent sinusoidal signals of
amplitude A1/3 ((blue, red, and green lines). Middle, the amplitude of the overall signal is reduced (A2 < A1) when a signal oscillates out of phase.
Right, when all signals lose their coherence, the amplitude of the global signal is further decreased (A3 < A2 < A1).
( 1 T* ) = ( 1 T ) + (1 T ) (13)
2 2random 2fixed
Figure 9. Generation of contrast by a CA. (a) T2 decay plots for two brain regions that are easy to differentiate (A: CSF at ventricle, B: white
matter of corpus callosum) due to their different T2 values. Subtracting the two decay plots makes it easy to see how contrast between the two
tissues changes with the extrinsic parameter TE, and find the value of TE at which contrast is maximal. (b) A brain tumor (orange plot) presents a
very similar T2 value to gray matter (red plot), being virtually impossible to generate contrast between these two tissues. (c) Administration of a
contrast agent into the brain parenchyma induces a substantial reduction of the T2 value on the injected area (changes the intrinsic properties of the
tissue), generating a strong contrast where there was no contrast before.
5. BIOMEDICAL MEANING OF THE RELAXATION hemorrhage, angiogenesis, calcification in the heart (low
TIMES water content), liver and cardiac iron.1
This is only a simplistic view of some biomedical
So far, we have learned that relaxation times are susceptible to
applications of T1 and T2 relaxometry, and many different
the local environment of the nuclei that produce the MR applications can be found in the literature.
signal, reflecting their influence as shorter or longer T1, T2, and
T2* values. In practice and considering that the local 6. CONTRAST AGENTS
environment of observed protons is directly related to tissue
composition and functional status, quantitative MR relaxom- MRI is a very versatile imaging technique with multiple
mechanisms available to generate contrast. However, there are
etry allows the characterization of the nature and functional
situation on which exploitation of those mechanisms is not
status of a wide range of healthy and pathological tissues in
efficient in generating the desired contrast between tissues or
biomedicine. This section provides a simple but straightfor- ROIs because the intrinsic properties (relaxation times) of
ward interpretation of the influence of several fundamental those tissues are very similar, and extrinsic parameters cannot
physiological and pathophysiological properties of tissues on enhance their differences.
the relaxation times T1 and T2 (or T2*). When the selection of pulse sequence and tuning of extrinsic
In the absence of super-paramagnetic or paramagnetic parameters are not enough to enhance contrast, it is possible to
substances (we will discuss later the presence of the so-called use a so-called “contrast agent”. Such an agent is generally a
contrast agents), the relaxation time T1 is mostly sensitive to nanoparticle, a small molecule, a macromolecule, or a
the macromolecular content, water content, presence of supramolecular complex that alters the intrinsic properties of
dissolved oxygen, and temperature. Any difference in tissue water molecules in the tissue under its influence (alters T1 or
composition or condition that involves a difference in these T2), enabling the possibility to generate a contrast. A graphical
parameters will be reflected in the T1 value. Thus, T1 has explanation of this phenomenon is presented in Figure 9.
proven value in evaluating the loss of myelin and axonal A possibility to generate contrast in tissues is the induction
alterations in brain tissue, loss of proteoglycan in cartilage, of a change in the properties of the surrounding media (pH, T,
changes in collagen and methemoglobin content in the heart, density, etc.), a process that is not always possible without
edema in the brain and heart, fat content in the liver, etc. disturbing the natural conditions that allow cells to survive and
When an exogenous gadolinium T1-reducing vascular contrast perform their functions with normality. Alternatively, we can
agent (a concept that will be discussed in detail in the next add a contrast agent in the surroundings of the observed
section) is intravenously injected, changes in T1 values and nuclei, which affects the relaxation processes and their
distribution in tissues enable assessment of tumor vessel blood characteristic relaxation times (by shortening them). Thus, a
contrast agent is an entity (small molecule, macromolecule,
flow and leakiness, breakdown of the blood−brain barrier,
supramolecular aggregates, nanoparticles, etc.) that influences
infarct or gray zone regions in the heart, and fibrosis, among
the intrinsic properties of MR active nuclei, altering the
other applications.1 contrast in the signal. It is important to mention that MR
The relaxation time T2 is sensitive to tissue composition and contrast agents are usually not active in MRI (they do not
structure, water content, and iron levels. It has been used to produce a signal), but they influence the signal produced by
assess myelin content in the brain, collagen content and other molecules (surrounding water, in most cases) instead
organization in cartilage and heart muscle, edema (inflamma- (again, we exclude from the discussion heteronuclear based
tion), hemorrhage, and iron content in the liver and heart. The imaging probes, such as 19F-containing materials, which indeed
relaxation time T2* is sensitive to deoxyhemoglobin and water generate a signal distinguished from the background by
content and has even better sensitivity than T2 to iron levels. It themselves).
has been used to assess brain function (altered deoxyhemo- In this work, we will focus on the concept of a contrast agent
globin levels), myocardial oxygenation, tumor hypoxia, as a material that alters the magnetic properties of protons, in
36912 https://doi.org/10.1021/acsomega.2c03549
ACS Omega 2022, 7, 36905−36917
ACS Omega http://pubs.acs.org/journal/acsodf Mini-Review
Figure 10. Determination of magnetic relaxivities for a CA. On top, R1 parametric maps of five tubes filled with water doped with a gadolinium-
based contrast agent at different concentrations. Bottom, characteristic linear regressions of the longitudinal (R1) and transverse (R2) relaxation
rates vs the concentration of three different materials. Two of them (Gadovist and Gd-nanogel) behave as contrast agents, reducing T1 and T2
values of water (increasing the relaxivities Ri). The third substance does not behave as a contrast agent in MR, not affecting the relaxation behavior
of water at any concentration.
1 R 2 = R 2,0 + r2[CA]
H based MRI. Materials containing other MRI active nuclei, (17)
such as 19F or 13C, that can be used as imaging probes in
heteronuclear MRI, and that are also referred as “contrast The longitudinal (r1) and transverse (r2) relaxivities of a
agents” in the literature, will not be discussed here. Other contrast agent are defined as the capacity of the agent to
options to generate contrast as the exploitation of magnet- enhance the relaxation rate constants (R1 or R2, respectively)
ization transfer mechanisms by suing the so-called CEST of water, normalized to the contrast agent’s concentration. The
contrast agents will be briefly introduced. Hyperpolarization is values of r1 and r2 are the main physicochemical parameters
another alternative to generate contrast in MRI, but the that are considered for the development of an effective
discussion of this phenomenon is beyond the scope of this magnetic resonance imaging probe (their ratio, and not just
work (see ref 7 for a review on this subject).7 their absolute values, is also important as we will discuss later).
When the action of the contrast agent results in an increase They depend on the size and chemical structure of the contrast
of the signal produced by water (by reduction of T1 times as a molecule, and on the accessibility of water molecules to the
result of a positive longitudinal relaxivity r1), they are called magnetic center.7
positive contrast agents (generally paramagnetic materials), Experimentally, the relaxivities of a contrast agent can be
while when their action results in a decrease of the signal (by estimated by preparing a series of solutions of the contrast
reduction of T2 times, as a result of a positive transverse agent at different concentrations and measuring the R1 and R2
relaxivity r2), they are called negative contrast agents (mostly (or R2*) values of those solutions using the relaxometry
super-paramagnetic materials). methods described in section 3. Then, relaxivities are
The magnitude of the signal enhancement induced by a calculated as the slope of the linear regressions of the
contrast agent (i.e., its efficiency) depends on intrinsic and relaxation rates versus the concentration of the solutions
extrinsic parameters. In general, the variation of the relaxation (Figure 10). For a very precise quantification of relaxivity
times suffered by water molecules (or any other MR detectable values, samples should be analyzed by inductively coupled
substances) in the presence of a contrast agent can be plasma (ICP)-MS or other precise analytical means to estimate
quantified by the following expressions: accurately the contrast agent concentration. It is important to
(1/T1) = (1/T1,0) + r1[CA ] remember that relaxivities depend on the external magnetic
(14)
field and temperature, when comparing absolute values from
(1/T2) = (1/T2,0) + r2[CA ] different experimental setups.
(15) It should be emphasized that the relaxivity of the agent and
where T1,0 and T2,0 (or T2,0*) are the relaxation times of its concentration in the tissue are not the only parameters that
observed molecules in the absence of contrast agent, r1 and r2 contribute to its efficiency. The distribution of the agent within
are the longitudinal and transverse relaxivities of the contrast the image voxel (or ROI), the proton density, and the potential
agent (intrinsic properties, in units of mM−1 s−1), and [CA] is diffusion of the agent, as well as the chemical environment of
the concentration of contrast agent (in mM). It is common to the surroundings, and of course the applied magnetic field, are
express the same relationships in terms of relaxation rates Ri = also contributors to the efficiency of signal enhancement that
(1/Ti) instead of relaxation times, as need to be considered (see ref 8 for extended explanation).8
Finally, it should be also mentioned that contrast agents for
R1 = R1,0 + r1[CA] (16) MRI should fulfill several additional requirements for
36913 https://doi.org/10.1021/acsomega.2c03549
ACS Omega 2022, 7, 36905−36917
ACS Omega http://pubs.acs.org/journal/acsodf Mini-Review
biomedical applications, including adequate relaxivity and further details on paramagnetic-based contrast agents, see
susceptibility effects, tolerance, low toxicity, stability, optimal Peter Caravan’s work.10,11
biodistribution, elimination from the body, or metabolization. 7.1.2. Agents Based on Superparamagnetic Relaxation.
The design and applications of MR contrast agents is a very Super-paramagnetic agents are based on iron oxides, mostly
active field of research that involves multiple disciplines such as magnetite (Fe3O4) or maghemite (γ-Fe2O3). These are water-
nanotechnology, biomaterial science, and biomedicine. Tuning dispersible crystals with a core diameter in the range 5−10 nm,
their structure, properties, functionality, biocompatibility, and made of several thousands of iron nuclei that are usually coated
in vivo behavior is a complex task that requires a wide range of by a polymeric shell (in the initial applications, it used to be a
skills, but in general, all of them work following the basic dextran coating, but nowadays, composition is very variable)
principles presented here. with typical sizes around 100 nm. This shell aims to stabilize
the particles, facilitate their dispersibility in water and their
7. MR CONTRAST AGENT TYPES AND APPLICATIONS biocompatibility, and allow their functionalization. When iron
We will finish this work with a brief discussion of the different ions are magnetically ordered within the crystal, the net
types of MR contrast agents and some of their biomedical magnetic moment of the whole nanoparticle is so large that it
applications. This is not a simple task since every day new MR greatly exceeds that of typical paramagnetic metal ions. This
contrast agents of different natures, mechanisms of action, and effect is called super-paramagnetism and is characterized by a
functionality are presented in the literature. Despite the large magnetic moment in an external magnetic field but no
numerous families of contrast agents, they can be classified, in remnant magnetic moment in the absence of such a field.
general, based on three different criteria: (1) the magnetic Super-paramagnetic agents strongly reduce both T1 and T2
properties of the contrast agents, (2) the in vivo relaxation times of surrounding water molecules, but their
biodistribution behavior of the contrast agent, and (3) the dominant effect is on the transverse relaxation times (T2 or
type of contrast enhancement produced by the agent (see T*2 ). Thus, super-paramagnetic contrast agents are usually
Figure 11). referred as T2 contrast agents. For further details on iron oxide-
based contrast agents, see the work of Muller and co-
workers.12
7.1.3. Agents Based on Magnetization Transfer Effects. A
novel family of contrast agents has been defined in the past
decade to introduce contrast enhancement in MR by
saturation transfer effects. Using RF pulses of appropriate
frequency, the magnetization of solid/semisolid protons
(belonging to macromolecules) can be transferred to mobile
protons of surrounding water molecules, inducing a marked
reduction in the signal intensity of mobile water molecules,
which interact with bound protons in the macromolecules.
This mechanism, known as magnetization transfer contrast
(MTC), can be applied in imaging applications using the so-
called CEST contrast agents. The amount of MTC will depend
on the macromolecules’ concentration, surface chemistry, and
Figure 11. Classification of contrast agents. biophysical dynamics. Thus, a CEST agent is a system
containing one or more mobile protons in slow/intermediate
exchange with water proton resonances. Thus, on irradiation of
7.1. Magnetic Properties. According to their magnetic the absorption of the exchanging protons, saturated magnet-
properties, contrast agents can be classified as follows. ization is transferred to the water signal, thereby causing a
7.1.1. Agents Based on Paramagnetic Relaxation. Most decrease in signal intensity. This approach is becoming popular
MR contrast agents used in human clinical applications are because it allows the use of specific radiofrequency pulses to
based on paramagnetic substances which are metal ions with introduce contrast, and it allows the activatable and selective
unpaired electrons in the outer orbital shells (transition and excitation of different contrast agents at the same time
lanthanide metal ions), giving rise to magnetic dipoles when (multifrequency imaging), providing the opportunity to
exposed to an external magnetic field. Since the electron’s perform multiplex detection of different contrast agents in
magnetic moment is ca. 700 times higher than the nuclear the same anatomical region. This feature is very appealing for
magnetic moment of a proton, paramagnetic ions induce large in vivo applications. For a further details on CEST-based
fluctuating magnetic fields that affect nearby protons. If the contrast agents, see the work of Aime and co-workers.13
frequency of such a fluctuation has components close to the 7.2. Biodistribution Behavior. In parallel, contrast agents
Larmor frequency of the observed nuclei, their relaxation will can be classified according to their biodistribution patterns:
be significantly enhanced, resulting in shorter relaxation times. 7.2.1. Intravascular Contrast Agents. Intravascular contrast
Many paramagnetic metal ions could be used as MR contrast agents are molecules administered intra-arterially or intra-
agents, but gadolinium or manganese ions (linked to chelating venously and have a sufficient size (in terms of molecular
molecules of different nature and functionality) are the most weight or hydrodynamic radius) to prevent their leakage from
used ones because of the seven unpaired electrons of Gd3+ ions the vascular bed to the extravascular space. Intravascular
(five for manganese), combined with long electron spin contrast agents present a half-life in blood of minutes to several
relaxation times. Paramagnetic agents affect both the T1 and hours and are eventually eliminated from the blood by the
the T2 values but are generally used to exploit their effects on Kupffer cells of the liver, spleen, and lymphatic system, or by
the longitudinal relaxation times (T1 contrast agents). For the kidneys.
36914 https://doi.org/10.1021/acsomega.2c03549
ACS Omega 2022, 7, 36905−36917
ACS Omega http://pubs.acs.org/journal/acsodf Mini-Review
■ AUTHOR INFORMATION
Corresponding Authors
biomarkers of vascular diseases for early diagnosis and monitoring
novel therapies.
Prof. Pedro Ramos-Cabrer received his Ph.D. in chemistry in 2000 at
Pedro Ramos-Cabrer − Center for Cooperative Research in the University of Santiago de Compostela, Spain. He specializes in
Biomaterials (CIC biomaGUNE), Basque Research and MRI and theranostic applications for the study of the central nervous
Technology Alliance (BRTA), 20014 Donostia-San system. His current research interests include the use of animal
Sebastián, Spain; Ikerbasque, Basque Foundation for Science, models of neurological diseases, development of imaging biomarkers
48013 Bilbao, Spain; orcid.org/0000-0003-0368-7031; of disease, and the development of drug delivery systems for the brain.
Email: pramos@cicbiomagune.es
Jesús Ruíz-Cabello − Center for Cooperative Research in
Biomaterials (CIC biomaGUNE), Basque Research and
Technology Alliance (BRTA), 20014 Donostia-San
■ ACKNOWLEDGMENTS
This work was performed under the Maria de Maeztu Units of
Sebastián, Spain; Ikerbasque, Basque Foundation for Science, Excellence Program, Grant No. MDM-2017-0720 from the
48013 Bilbao, Spain; CIBER de Enfermedades Respiratorias Ministry of Science, Innovation and Universities of Spain and
(CIBERES), 28029 Madrid, Spain; Departamento de has been funded by the Spanish Research Agency (Grants
Química en Ciencias Farmacéuticas, Universidad PID2020-118546RB-I00, PID2021-123238OB-I00, and
Complutense de Madrid, 28040 Madrid, Spain; PDC2021-121696-I00) and the Basque Government (Grant
orcid.org/0000-0001-8681-5056; Email: jruizcabello@ Elkartek KK-2021/0009). J.R.-C. received funding from the La
cicbiomagune.es Caixa Foundation (Health Research Call 2020: HR20-
36916 https://doi.org/10.1021/acsomega.2c03549
ACS Omega 2022, 7, 36905−36917
ACS Omega http://pubs.acs.org/journal/acsodf Mini-Review
00075)). P.R.-C. and J.R.-C. are financed by the Ikerbasque (20) Lee, D.-E.; Koo, H.; Sun, I.-C.; Ryu, J. H.; Kim, K.; Kwon, I. C.
Research Professors program. Multifunctional nanoparticles for multimodal imaging and therag-
nosis. Chem. Soc. Rev. 2012, 41 (7), 2656−72.
■ REFERENCES
(1) Cheng, H. L. M.; Stikov, N.; Ghugre, N. R.; Wright, G. A.
Practical medical applications of quantitative M.R. relaxometry. J.
Magn Reson Imaging 2012, 36 (4), 805−824.
(2) Lin, A. L.; Qin, Q.; Zhao, X.; Duong, T. Q. Blood longitudinal
(T1) and transverse (T2) relaxation time constants at 11. 7 T.
MAGMA 2012, 25 (3), 245−249.
(3) Kingsley, P. B. Introduction to diffusion tensor imaging
mathematics: Part II. Anisotropy, diffusion-weighting factors, and
gradient encoding schemes. Concepts Magn. Reson. 2006, 28A, 123−
154.
(4) Nessaiver, M. All You Really Need to Know about MRI Physics;
Simply Physics: Baltimore (M.D.), USA, 1997.
(5) Brown, R. W.; Cheng, Y. C. N.; Haacke, E. M.; Thompson, M.
R.; Venkatesan, R. Magnetic Resonance Imaging: Physical Principles and
Sequence Design, 2nd ed.. 2014. Wiley-Blackwell, New York, NY.
(6) Westbrook, C.; Kaut-Roth, C.; Talbot, J. MRI in Practice; Wiley-
Blackwell: New York (N.Y.), USA, 2011.
(7) Adamson, E. B.; Ludwig, K. D.; Mummy, D. G.; Fain, S. B.
Magnetic resonance imaging with hyperpolarized agents: methods
and applications. Phys. Med. Biol. 2017, 62 (13), R81−R123.
(8) Burtea, C.; Laurent, S.; Vander Elst, L.; Muller, R. N. Contrast
agents: magnetic resonance. Handb. Exp. Pharmacol. 2008, 185 (Pt
1), 135−165.
(9) Bjo̷ rnerud, A. Advanced Applications of M.R.I. Contrast Agents.
In The Physics of Magnetic Resonance Imaging. Department of Physics
University of Oslo, 2008; Chapter 12:12. Available online: http://
www.uio.no/studier/emner/matnat/fys/FYS-KJM4740/v08/
undervisningsmateriale/Compendium%20MRI%20Feb%202009.pdf.
(Accessed on 15 Feb. 2022).
(10) Caravan, P. Strategies for increasing the sensitivity of
gadolinium based MRI contrast agents. Chem. Soc. Rev. 2006, 35
(6), 512−523.
(11) Wahsner, J.; Gale, E. M.; Rodríguez-Rodríguez, A.; Caravan, P.
Chemistry of MRI Contrast Agents: Current Challenges and New
Frontiers. Chem. Rev. 2019, 119 (2), 957−1057.
(12) Liu, F.; Laurent, S.; Fattahi, H.; Vander Elst, L.; Muller, R. N.
Superparamagnetic nanosystems based on iron oxide nanoparticles for
biomedical imaging. Nanomedicine (Lond). 2011, 6 (3), 519−528.
(13) Castelli, D. D.; Terreno, E.; Longo, D.; Aime, S. Nanoparticle-
based chemical exchange saturation transfer (CEST) agents. NMR
Biomed. 2013, 26 (7), 839−849.
(14) Sanchez-Gaytan, B. L.; Fay, F.; Lobatto, M. E.; Tang, J.;
Ouimet, M.; Kim, Y. T.; van der Staay, S. E. M.; van Rijs, S. M.;
Priem, B.; Zhang, L.; et al. HDL-mimetic PLGA nanoparticle to target
atherosclerosis plaque macrophages. Bioconjug Chem. 2015, 26 (3),
443−451.
(15) Agulla, J.; Brea, D.; Campos, F.; Sobrino, T.; Argibay, B.; Al-
Soufi, W.; Blanco, M.; Castillo, J.; Ramos-Cabrer, P. In vivo
theranostics at the peri-infarct region in cerebral ischemia.
Theranostics 2014, 4 (1), 90−105.
(16) Kim, Y.; Shin, T. H.; Choi, J. S.; Yun, S.; Kim, I. S.; Song, H.-T.;
Park, K. I.; Cheon, J. T1 and T2 dual-mode MRI contrast agent for
enhancing accuracy by engineered nanomaterials. ACS Nano 2014, 8
(4), 3393−3401.
(17) Genicio, N.; Bañ obre-Lopez, M.; Gröhn, O.; Gallo, J.
Ratiometric magnetic resonance Imaging:Contrast agent design
towards better specificity and quantification. Coord. Chem. Rev.
2021, 447, No. 214150.
(18) Davies, G. L.; Kramberger, I.; Davis, J. J. Environmentally
responsive MRI contrast agents. Chem. Commun. 2013, 49 (84),
9704−21.
(19) Himmelreich, U.; Weber, R.; Ramos-Cabrer, P.; Wegener, S.;
Kandal, K.; Shapiro, E. M.; Koretsky, A. P.; Hoehn, M. Improved
stem cell MR detectability in animal models by modification of the
inhalation gas. Mol. Imaging 2005, 4 (2), 104.