You are on page 1of 12

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: D8470 − 22

Standard Practice for


Development and Implementation of Instrument
Performance Tests for Use on Multivariate Online, At-Line
and Laboratory Spectroscopic Based Analyzer Systems1
This standard is issued under the fixed designation D8470; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope calibrations via Practice D8321 to establish the performance


1.1 This practice covers basic procedures that can be used to level at the time the calibration is developed. The same tests are
develop instrument performance tests for spectroscopic based used during validation of analyzers via Practice D6122 to
online, at-line, laboratory and field analyzers. The practice is qualify the working analyzer by demonstrating comparable
intended to be applicable to Raman spectrometers and to performance.
infrared spectrophotometers operating in the near-infrared and 1.4.1 Instrument performance tests are used to requalify
mid-infrared regions. instruments after analyzer maintenance.
1.4.2 Instrument performance tests are used to qualify
1.2 This practice is not intended as a replacement for instruments in secondary analyzers to which calibrations are
specific practices, such as Practices E275, E925, E932, E958, being transferred after development on a primary analyzer.
E1421, or E1683 that exist for measuring performance of
1.5 This standard does not purport to address all of the
specific types of laboratory spectroscopic instruments. Instead,
this practice is intended to provide guidelines as to how similar safety concerns, if any, associated with its use. It is the
practices should be developed when specific practices do not responsibility of the user of this standard to establish appro-
exist for a particular instrument type, or when specific practices priate safety, health, and environmental practices and deter-
are not applicable due to sampling or safety concerns. This mine the applicability of regulatory limitations prior to use.
practice can be used to develop instrument performance tests 1.6 This international standard was developed in accor-
for on-line process spectroscopic-based analyzers. dance with internationally recognized principles on standard-
1.2.1 The performance tests described in this practice typi- ization established in the Decision on Principles for the
cally only evaluate the performance of the infrared spectropho- Development of International Standards, Guides and Recom-
tometer or Raman spectrometer part of the analyzer system, mendations issued by the World Trade Organization Technical
referred to herein as the instrument. Barriers to Trade (TBT) Committee.
1.2.2 Instrument performance tests do not typically evaluate 2. Referenced Documents
performance of analyzer sampling systems.
2.1 ASTM Standards:2
1.3 This practice describes univariate level zero and level D4175 Terminology Relating to Petroleum Products, Liquid
one tests, and multivariate level A and level B tests which can Fuels, and Lubricants
be implemented to measure instrument performance. These D6122 Practice for Validation of the Performance of Multi-
tests are designed to be used as rapid, routine checks of variate Online, At-Line, Field and Laboratory Infrared
instrument performance. They are designed to uncover mal- Spectrophotometer, and Raman Spectrometer Based Ana-
functions or other changes in instrument operation, but do not lyzer Systems
specifically diagnose or quantitatively assess the malfunction D7940 Practice for Analysis of Liquefied Natural Gas
or change. The tests are not intended for the comparison of (LNG) by Fiber-Coupled Raman Spectroscopy
instruments or analyzers of different manufacture. D8321 Practice for Development and Validation of Multi-
1.4 The instrument performance tests described in this variate Analyses for Use in Predicting Properties of
practice are used during the development of multivariate Petroleum Products, Liquid Fuels, and Lubricants based
on Spectroscopic Measurements
1
This practice is under the jurisdiction of ASTM Committee D02 on Petroleum
Products, Liquid Fuels, and Lubricants and is the direct responsibility of Subcom-
2
mittee D02.25 on Performance Assessment and Validation of Process Stream For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Analyzer Systems. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Current edition approved July 1, 2022. Published August 2022. DOI: 10.1520/ Standards volume information, refer to the standard’s Document Summary page on
D8470-22. the ASTM website.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

1
D8470 − 22
D8340 Practice for Performance-Based Qualification of probe located in the process wherein the light interacts with a
Spectroscopic Analyzer Systems process sample, and collecting the transmitted or scattered
E131 Terminology Relating to Molecular Spectroscopy light, returning it to the analyzer where it is analyzed by the
E275 Practice for Describing and Measuring Performance of instrument to produce a spectrum of the process sample.
Ultraviolet and Visible Spectrophotometers 3.2.6.1 Discussion—For some NIR and nearly all Raman
E925 Practice for Monitoring the Calibration of Ultraviolet- analyzers, the light will typically be transported to and from the
Visible Spectrophotometers whose Spectral Bandwidth analyzer using fiber optics.
does not Exceed 2 nm
3.2.7 instrument, n—for spectroscopic analyzers used in
E932 Practice for Describing and Measuring Performance of
petrochemical service, the spectrometer or spectrophotometer,
Dispersive Infrared Spectrometers
associated electronics and computer, spectrometer or spectro-
E958 Practice for Estimation of the Spectral Bandwidth of
photometer cell, and if utilized, transfer optics. D6122
Ultraviolet-Visible Spectrophotometers
E1421 Practice for Describing and Measuring Performance 3.2.8 instrument performance verification sample, n—for
of Fourier Transform Mid-Infrared (FT-MIR) Spectrom- multivariate spectroscopic analyzers used in the analysis of
eters: Level Zero and Level One Tests liquid petroleum products and fuels which employ extractive
E1683 Practice for Testing the Performance of Scanning sampling, a material representative of the product being
Raman Spectrometers analyzed which is adequately stored in sufficient quantity to be
E1840 Guide for Raman Shift Standards for Spectrometer used as a check on instrument performance; instrument perfor-
Calibration mance verification samples are used in instrument performance
E1866 Guide for Establishing Spectrophotometer Perfor- tests and as checks on calibration transfer, but the samples and
mance Tests their spectra are generally not reproducible long term D6122
3.2.8.1 Discussion—In E1866 and previous versions of
3. Terminology D6122 and this practice, an instrument performance verifica-
3.1 For terminology related to molecular spectroscopic tion sample was referred to as a test sample.
methods, refer to Terminology E131. For terminology relating 3.2.9 instrument qualification sample, n—for multivariate
to petroleum products, liquid fuels and lubricants, refer to spectroscopic analyzers used in the analysis of liquid petro-
Terminology D4175. leum products and fuels, a single pure compound, or a known,
3.2 Definitions: reproducible mixture of compounds whose spectrum is con-
3.2.1 absorbance (A), n—the logarithm to the base 10 of the stant over time such that it can be used in an instrument
reciprocal of the transmittance, (T). performance test. D6122
A 5 log10~ 1 ⁄ T ! 5 2log10T 3.2.9.1 Discussion—In E1866 and previous versions of
E131 D6122 and this practice, an instrument qualification sample
3.2.1.1 Discussion—Absorbance is a measure of the capac- was referred to as a check sample.
ity of a substance to absorb light of a specific wavelength. 3.2.10 level A test, n—for spectroscopic analyzers used in
3.2.2 action limit, n—for multivariate spectroscopic analyz- the analysis of liquid petroleum products and fuels, a pass/fail
ers used in the analysis of liquid petroleum products and fuels, instrument performance test in which the spectrum of a check
the limiting value from an instrument performance test, beyond or instrument performance verification sample is compared
which the multivariate spectroscopic analyzer is expected to against historical spectra of the same sample via a multivariate
produce potentially invalid results. D6122 analysis.
3.2.3 calibration, n—in multivariate spectroscopic 3.2.11 level B test, n—for spectroscopic analyzers used in
measurement, a process for creating a multivariate model the analysis of liquid petroleum products and fuels, a pass/fail
relating component concentrations or sample properties to instrument performance test in which the spectrum of a check
spectra for a set of known samples, referred to as calibration or instrument performance verification sample is analyzed
samples. D8321 using a multivariate model, and the results of the analysis are
3.2.4 control limits, n—limits on a control chart that are compared to historical results for prior analyses of the same
used as criteria for signaling the need for action or for judging sample.
whether a set of data does or does not indicate a state of 3.2.12 level one (1) test, n—for spectroscopic analyzers
statistical control. used in the analysis of liquid petroleum products and fuels, a
3.2.5 extractive sampling, n—for spectroscopic process simple series of measurements designed to provide quantitative
analyzers used in the analysis of liquid petroleum products and data on various aspects of spectrophotometer performance and
fuels, the process of transporting a sample from the process, information on which to base the diagnosis of problems.
performing any necessary sample conditioning and finally 3.2.13 level zero (0) test, n—for spectroscopic analyzers
presenting the sample to the analyzer so as to produce a used in the analysis of liquid petroleum products and fuels, a
spectrum of the process sample. routine check of instrument performance, which can be done in
3.2.6 in situ sampling, n—for spectroscopic process analyz- a few minutes, designed to visually detect significant changes
ers used in the analysis of liquid petroleum products and fuels, in instrument performance and provide a database to determine
the process of transporting light from the analyzer to a sample instrument performance over time.

2
D8470 − 22
3.2.14 liquid petroleum products and fuels, n—in relation to ible with the intended instrument operation, then this practice
analyzers, any single-phase liquid material that is produced at can be used to develop practical performance tests.
a facility in the petroleum and petrochemical industries and 4.1.1 For instruments which are equipped with permanent
will be in whole or in part of a petroleum product; it is or semi-permanent sampling accessories, the test sample speci-
inclusive of biofuels, renewable fuels, blendstocks, alternative fied in a Committee E13 practice may not be compatible with
blendstocks, and additives. D8340 the instrument configuration. For example, for FT-MIR instru-
3.2.15 model variables, n—the independent variables de- ments equipped with transmittance or IRS flow cells, tests
rived from the calibration spectra which are regressed against based on putting polystyrene films into the sample position are
the calibration sample properties to produce the multivariate impractical. In such cases, this practice suggests means by
model. D6122 which the recommended test procedures can be modified by
3.2.15.1 Discussion—For MLR, the model variables would changing the test material or the location of the recommended
be the spectral intensity at the selected wavelengths or frequen- test material.
cies; for PCR or PLS, the model variables are the Principal 4.1.2 For instruments used in process measurements, the
Components or latent variables. choice of test materials may be limited due to process
contamination and safety considerations. The practice suggests
3.2.16 optical reference material, n—for spectroscopic ana-
means of developing performance tests based on materials
lyzers used in the analysis of liquid petroleum products and
which are compatible with the intended use of the analyzer.
fuels, an optical filter or other device which can be inserted into
the optical path in the spectrometer, spectrophotometer or 4.2 Tests developed using the practice are intended to allow
probe producing a spectrum which is known to be constant the user to compare the performance of an instrument on any
over time such that it can be used in place of a check or given day with prior performance, and specifically to compare
instrument performance verification sample in a performance performance during calibration of the analyzer to performance
test. during validation of the analyzer and during routine use of the
analyzer. The tests are intended to uncover malfunctions or
3.2.17 primary analyzer, n—the analyzer(s) on which cali-
other changes in instrument operation, but they are not de-
bration spectra are collected for the purpose of building a
signed to diagnose or quantitatively assess the malfunction or
multivariate model. D8321
change. The tests are not intended for the comparison of
3.2.18 Raman Shift Standard, n—a physical sample having analyzers of different manufacture.
known Raman shift characteristics used to determine the
operating wavelength of laser module and correct for any 4.3 Tests developed using this practice are also intended to
short-term or accumulated error. The position of the spectra allow the user to compare the performance of a primary
band(s) of this material can be used to calculate the operational analyzer used in development of a multivariate model to the
wavelength of the laser: performance of secondary analyzers used to apply that model
for the analysis of process or product samples.
1
λi 5 (1)
S v̄ 1
1
λs D 5. Test Conditions
5.1 When conducting the performance tests, the instrument
where λi is the incident wavelength (the operational wave-
length of the laser) in cm, λs is the measured wavelength of should be operated under the same conditions as will be in
the Raman band in cm, and v̄ is the accepted standard posi- effect during its intended use. Sufficient warm-up time should
tion of the Raman band of the reference material in be allowed before the commencement of any measurements.
wavenumbers. D7940 5.1.1 If possible, the optical configuration used for measure-
3.2.19 secondary analyzer, n—an analyzer not used in the ments of instrument performance verification and instrument
development of the multivariate model, but which will be used qualification samples should be identical to that used for
for analysis of new materials. D8321 normal operations. If identical optical configurations are not
possible, the user should recognize that the performance tests
3.2.20 spectral intensity, n—a generic term referring to may not measure the performance of the entire instrument.
either infrared absorbance or Raman scattering intensity. 5.1.2 Data collection and computation conditions should
D8321 generally be identical to those used in normal operation.
3.2.21 spectral position, n—a generic term referring to However, instrument performance tests can use data collection
either wavelength or frequency position in a spectrum. D8321 and computation conditions that are more demanding than
3.2.22 Spectrum Standard, n—light source of known spec- those used in normal operation if advantageous for instrument
trum used to standardize/calibrate the detection module and standardization.
providing correct mapping of scattered wavelength to physical 5.1.3 Spectral data used in performance tests should be date
pixel coordinates/location. Typically, an atomic emission and time stamped, and the results of the tests should be stored
source such as a neon lamp is used for this purpose. D7940 in a historical database.

4. Significance and Use 6. Materials Used for Performance Testing


4.1 If ASTM Committee E13 has not specified an appropri- 6.1 The types of materials used for performance testing will
ate test procedure for a specific type of instrument, or if the vary depending on whether the analyzer uses extractive or in
sample specified by a Committee E13 procedure is incompat- situ sampling.

3
D8470 − 22
6.1.1 For analyzers using extractive sampling, it is generally rately pipetted or weighed at ambient temperature. It is
possible to introduce an instrument qualification or instrument recommended that mixtures be independently verified for
performance verification sample into the transmission/ composition prior to use.
scattering cell. The materials used for performance testing are 6.3.3 While mixtures can be used as instrument qualifica-
chosen to be compatible with the instrument configuration, and tion samples, their spectra may be adversely affected by
to provide spectral features which are adequate for the tests temperature-sensitive interactions that may manifest them-
being performed. selves by wavelength (frequency) and spectral intensity
6.1.1.1 A liquid sample used for performance testing will changes.
generally be in the same physical state (gas, liquid, or solid) as
6.4 Instrument Performance Verification Samples—An in-
the samples to be analyzed during normal operation of the
strument performance verification sample is a process or
instrument.
product sample or a mixture of process or product samples
6.1.1.2 A liquid sample used for performance testing should
whose spectrum is expected to be constant for the time period
be physically and chemically compatible with the samples
it is used in performance testing. The instrument performance
analyzed during normal operation so as to not represent a
verification sample must be stored in bulk quantities in
safety or product quality issue if it ends up mixed into the
controlled conditions such that the material is stable over time.
process stream being analyzed.
6.1.1.3 A liquid sample used for performance testing pref- 6.4.1 Instrument Performance Verification Samples are used
erably has a spectrum that is similar to the spectra which will to conduct Level B Test (see 9.2).
be collected during normal operation. 6.4.2 Since instrument performance verification samples are
6.1.1.4 For infrared systems, a liquid sample used for often complex mixtures which cannot be synthetically
performance testing will preferably have several significant reproduced, they can only be used for performance testing for
absorbances (0.3 < absorbance < 1.0) across the spectral range limited time periods. If instrument performance verification
used for normal operation of the instrument. samples are used for this purpose, collection of historical data
6.1.1.5 In order to adequately determine the photometric on a new instrument performance verification sample should be
linearity of the infrared instrument, the peak absorbance for at initiated before previous instrument performance verification
least one absorption band of a liquid sample will preferably be samples are depleted. It is recommended that new instrument
similar to or slightly greater than the largest absorbance performance verification samples be analyzed sequentially
expected for samples measured during normal operation. with old instrument performance verification samples at least
6.1.1.6 A liquid sample used for performance testing will 15 times before they are used to replace the old instrument
typically be introduced directly into the transmission/scattering performance verification sample. The 15 analyses must be
cell and not pass through any sample conditioning system. As performed over a time period that does not exceed one month
such, it only tests the performance of the instrument and not the in duration.
total analyzer system. 6.5 Optical Reference Materials—An optical reference ma-
6.1.2 For infrared analyzers using extractive sampling, in- terial produces a spectrum which is known to be constant over
strument performance tests can be conducted by inserting an time. This material may be automatically inserted into the
optical reference material into the optical path when the optical path either within the instrument or in the fiber optic
transmission cell is empty. sample probe to allow instrument performance tests to be
6.2 For analyzers using in situ sampling, introducing an performed.
instrument qualification sample or an instrument performance 6.5.1 If an optical reference material is used routinely to
verification sample into the actual process sample probe can check or correct the spectral data collection or computation,
only be done if the probe is removed from the process. For then the same material is preferably not used for instrument
analyzers equipped with optical multiplexers, a separate chan- performance testing. If the same filter is used, then the part of
nel and probe may be used for the performance test. The the filter spectrum used in the performance testing should
performance test can be conducted on an instrument qualifica- preferably differ from that part used to check or correct the
tion sample, an instrument performance verification sample, or instrument. For example, polystyrene filters are used to stan-
on an optical reference material. It should be recognized that dardize (continuously check and correct) the wavelength scale
such tests do not verify the performance of the actual process of some dispersive NIR spectrophotometers and Raman ana-
probe. lyzers. For such systems, polystyrene filters are preferably not
6.3 Instrument Qualification Samples—Instrument qualifi- to be employed for wavelength stability performance testing. If
cation samples can be used to conduct performance tests. polystyrene filters are used, then the peaks used for wavelength
Instrument qualification samples are single pure compounds or stability testing should be different from those used for
mixtures of compounds of definite composition. standardizing the wavelength scale.
6.3.1 Instrument qualification samples are used to conduct 6.6 Spectrum Standards and Raman Shift Standards—Light
Level A tests (see 9.1). sources with known spectra are used for standardizing and
6.3.2 If mixtures are utilized as instrument qualification calibrating the frequency axis of Raman spectrometers. Atomic
samples, they must be prepared in a repeatable manner and, if emission sources such as neon lamps are used for this purpose.
stored, stored such that the mixture is stable over long periods The peak positions of a Raman Shift Standard such as toluene
of time. In preparing mixtures, components should be accu- can be used in instrument performance testing (Practice

4
D8470 − 22
D7940). Practice E1840 lists various materials that can be used spectrum may be converted to an absorption spectrum by
as Raman shift standards. taking the negative log10 before the photometric noise calcu-
lations.
6.7 For Raman systems, the spectral features used in per-
7.2.2 For double beam instruments, a 100 % line spectrum
formance testing should be in the linear range for the detector,
is measured when the two beams are both empty, both contain
above the noise level but below the saturation level. If the
empty matched cells, or both contain reference samples in
linear range of the Raman detector has not previously been
matched cells.
established, a series of mixtures should be measured at various
7.2.3 Photometric noise is measured by fitting a line to the
exposure times to determine the range over which response is
spectrum over a short spectral region centered on the test
linear.
frequency (wavelength). The region should contain at least 11
data points, preferably contains 101 data points, and should not
7. Univariate Measures of Absorbance
exceed 2 % of the spectral range. The line is subtracted from
Spectrophotometer Performance the spectral data, and the RMS noise is calculated as the square
7.1 Energy Level Tests—Energy level tests are intended to root of the mean square residual.
detect changes in the radiant power in the instrument beam. 7.2.3.1 If Ti is the transmittance at the frequency vi, then the
Decreases in energy levels may be associated with deteriora- slope, m, and intercept, b, of a line through the n data points
tion of the instrument source, with contamination or misalign- centered at test frequency v0 are given by the following:
ment of optical surfaces in the light path, or with malfunctions nΣiT i 2 ΣT i Σi ΣiT i
of the detector. m5 5 (2)
nΣi 2 2 ~ Σi ! 2 Σi 2
7.1.1 For single beam instruments where background and Σi 2 ΣT i 2 ΣiΣiT i ΣT i
sample spectra are measured separately at different times, b5
nΣi 2 2 ~ Σi ! 2
5
n
(3)
energy level tests are generally conducted on a background
spectrum. For double beam instruments where the ratio of The photometric noise is calculated as follows:
background and sample beam intensities is measured directly,
energy levels can be measured if it is possible to block the
sample beam.
Noise RMS 5 Œ Σ ~ T i 2 ~ mi 1 b !! 2
n22
(4)

7.1.2 Energy levels should be measured at a minimum of The index i in Eq 2-4 runs from – (n – 1)/2 to (n – 1)/2 (n
two fixed frequencies (wavelengths). The frequencies (wave- must be odd). The intercept represents the transmittance at test
lengths) at which energy levels are measured should be chosen frequency v0.
to avoid interferences due to atmospheric components (for 7.2.3.2 If photometric noise is calculated on spectra, the
example, absorptions of water vapor and carbon dioxide) and spectral intensity values, Ai, are used in place of the transmit-
from interferences due to optical components (for example, tance values, Ti, in Eq 2-4. If the abscissa for the spectral data
OH absorptions in SiO2 cells and fibers). Preferably, regions is wavelength, then wavelength values, λi, are used in place of
where the background spectrum is relatively flat and slowly the frequency values, vi, in Eq 2-4. Calculations should be
varying should be used for this test. consistently performed on the same data types.
7.1.3 To minimize the effects of photometric noise on the 7.2.4 Increases in the photometric noise can indicate a
energy level measurement, it is preferable to average the misalignment of optical components, a source malfunction, or
energy over a narrow frequency (wavelength) window. a malfunction in the detector or electronics.
Typically, the intensity at five points centered on the test 7.3 Short Term Baseline Stability Test—The transmittance is
frequency are averaged. monitored at each of the test frequencies (wavelengths) used in
7.2 Photometric Noise Tests—Photometric noise is mea- the energy level and photometric noise tests. The intercept
sured at the same frequencies (wavelengths) used for the calculated in Eq 3 represents the transmittance averaged over
energy level tests. Preferably, photometric noise tests are the n points around test frequency v0. Deviation from 100 %
conducted on a 100 % line spectrum. Alternatively, photomet- transmittance is an indication of short-term baseline instability
ric noise tests may be conducted on the spectrum of an and may indicate a malfunction of the instrument.
instrument qualification or instrument performance verification 7.3.1 If the tests are conducted on absorbance spectra,
sample at regions where the spectrum is relatively flat, and the deviations from zero absorbance is used as an indication of
sample spectral intensity is minimal (<0.1). baseline instability.
7.2.1 For single beam instruments where background and 7.3.2 If photometric noise tests are conducted on the spec-
sample spectra are measured separately at different times, a trum of an instrument qualification or instrument performance
100 % line spectrum is obtained by ratioing two successive verification sample, then variations in the absorbance spectrum
background measurements to obtain a transmittance spectrum. at the test frequencies are taken as an indication of short-term
If, during normal operation of the instrument, backgrounds are baseline instability.
collected with a reference material in the optical path, then this 7.4 Optical Contamination Tests—The single beam back-
same configuration should be used for performance testing. ground scan which was used for the energy tests is examined
Photometric noise calculations are preferably done directly on for absorptions which might indicate contamination of optical
the transmittance spectrum. Alternatively, the transmittance surfaces in the beam path.

5
D8470 − 22
7.4.1 Failure to clean cell or probe windows, IRS surfaces, half-maximum (FWHM) of the peak being measured divided
etc., are the most common source of optical contamination. by the digital resolution and rounded up to the nearest odd
Frequencies (wavelengths) at which typical samples exhibit integer.
maximum absorbance should generally be examined. For 7.6.3.2 Identify the zero-crossing associated with the peak
example, for systems used in hydrocarbon analysis, the regions absorbance and compute its location by linear interpolation
where the C-H stretching vibrations occur should be examined. between the two adjacent points straddling the zero crossing.
Significant increases above a nominal background level may The zero crossing is taken as a measure of the peak position.
indicate contamination of windows and surfaces.
7.4.2 Instrument optical surfaces can be contaminated by NOTE 1—Other peak finding algorithms can be used provided that they
accurately track peak position. The procedure described in Annex A1
impurities in purge gases. For systems equipped with flow cells should be used to test peak finding algorithms to determine if they are
or circulating liquid temperature control, leaks in connecting appropriate for this application. It is the user’s responsibility to demon-
lines can expose an optical surface to contamination. Users strate that the peak finding algorithm is appropriate for monitoring
should consider possible sources of contamination and deter- instrument frequency (wavelength) stability.
mine appropriate frequencies at which absorptions would 7.7 Resolution Stability Tests—The resolution stability of
result. the instrument is monitored by measuring the bandwidths of
7.5 Purge Contamination Tests—For instruments which are several absorption peaks in the spectrum of the instrument
purged to minimize absorptions due to atmospheric qualification/instrument performance verification sample or
components, the single beam spectrum used for energy tests optical filter. At least three peaks are used for the test. If
should be checked for variations in purge quality. Frequencies possible, the peaks should be in the upper, middle and lower
(wavelengths) at which potential contaminants absorb should third of the spectral range. Variations in the measured band-
be identified, as should baseline points where contaminant widths are taken as an indication that the optical resolution of
absorption would be minimal. The absorbance for contami- the instrument is varying, suggesting a malfunction.
nants is calculated as the negative log10 of the ratio of the peak 7.7.1 The spectral intensities for peaks used in this test are
intensity to the baseline intensity. preferably in the range from 0.37 to 0.75. For peak spectral
intensity outside this range, the resolution stability measure-
7.6 Frequency (Wavelength) Stability Tests—Frequency
ment may show increased sensitivity to photometric noise.
(wavelength) stability tests are conducted by monitoring the
peak positions of several peaks across the absorption spectrum 7.7.2 Peaks used for the resolution stability test are prefer-
of the instrument qualification or instrument performance ably symmetric in shape and well resolved from neighboring
verification sample or optical filter. At least three peaks are peaks. If such peaks are not available in the spectrum of the
used for the test. If possible, the peaks should be in the upper, instrument qualification/instrument performance verification
middle, and lower third of the spectral range. sample or optical filter, the results of the resolution stability test
may be variable.
7.6.1 The absorption for peaks used in this test are prefer-
ably in the range from 0.37 to 0.75. For peak absorptions 7.7.3 It is recommended that the peak bandwidth be deter-
outside this range, the wavelength stability measurement may mined by the following steps:
show greater sensitivity to photometric noise. 7.7.3.1 Compute the second derivative of the spectrum by
7.6.2 Peaks used for the frequency stability test are prefer- applying the appropriate digital filter to the spectrum. A
ably symmetric in shape and well resolved from neighboring commonly used filter has been defined by Savitzky and Golay
peaks. If such peaks are not available in the spectrum of the (1) with corrections by Steiner, Termonia, and Deltour (2), with
instrument qualification/instrument performance verification application criteria discussed by Willson and Polo (3). The
sample or optical material, the user should be aware that latter reference discusses optimum filter parameters based upon
changes in instrument resolution will affect the measured peak the relationship between spectral bandwidth and digitization
position. interval. A cubic filter is recommended. The number of points
7.6.3 It is recommended that the peak position be deter- used in the filter should be the quotient of the FWHM of the
mined by the following steps: peak being measured divided by the digital resolution and
rounded up to the nearest odd integer.
7.6.3.1 Compute the first derivative of the spectrum by
applying the appropriate digital filter to the spectrum. A 7.7.3.2 Identify the zero crossing on each side of the peak
commonly used filter has been defined by Savitzky and Golay intensity and compute their locations by linear interpolation
(1)3 with corrections by Steiner, Termonia, and Deltour (2), between the two adjacent points straddling the zero crossings.
with application criteria discussed by Willson and Polo (3). The The difference in the frequencies of the interpolated zero
latter reference discusses optimum filter parameters based upon crossings is taken as a measure of the peak bandwidth.
the relationship between spectral bandwidth and digitization 7.8 Photometric Linearity Tests—Linearity of the instru-
interval. A cubic filter is recommended. The number of points ment response is important for quantitative applications.
used in the filter should be the quotient of the full-width-at- Unfortunately, the absolute photometric linearity cannot be
checked in a quick performance test. To do so would generally
require the use of multiple standards of known absorbance. The
3
The boldface numbers in parentheses refer to a list of references at the end of test described here is intended only to measure changes in the
this standard. photometric linearity of a instrument.

6
D8470 − 22
7.8.1 Photometric linearity is tested using the ratio of the the wavelength positions and relative intensities of the various
spectral intensity of two or more peaks in the - spectrum. One spectral features associated with the components. Each of these
peak should have a spectral intensity at or near the maximum axes is calibrated by optical means.
spectral intensity that will be used for normal operations. The
8.2 Wavelength Axis—For the wavelength axis calibration to
other peaks are preferably less intense than this maximum. If
be standardized, and thus to know which detected vibration
only two peaks are used, the second peak should be approxi-
frequencies correspond to given spectrum features, it is neces-
mately half the intensity of the first peak.
sary to determine the wavelength falling on each group of CCD
7.8.2 Linear baselines for each peak are calculated from
pixels. This is done by exposing the detector to light that is
points of minimal spectral intensity on opposite sides of the
comprised of known individual wavelengths. Typically, this is
peaks. The maximum spectral intensity for each peak is
done by arranging for light from a neon bulb to be periodically
corrected for the baseline, and the ratio of the spectral
and automatically imaged onto the detector. The emission
intensities for the two peaks is calculated. The ratio is used to
spectra of neon is comprised of a large number of individual
track changes in the photometric linearity.
wavelengths that are known to great precision, and the physics
8. Univariate Measures of Raman Spectrophotometer associated with neon light ensure wavelength stability through
Performance time and over a wide range of environmental conditions, thus
establishing standardization. By recording where each neon
8.1 During the development of the analytic method, corre- wavelength falls on the CCD, a map can be generated relating
lations are established between spectra and the sample species pixel to wavelength. Any changes in the position of a given
by taking spectra of known samples. Because of the inherent wavelength on the detector as a result of thermal or mechanical
linearity of the Raman effect, the method will correctly effects can thus be corrected by measuring the change with the
measure sample concentrations if the Raman spectra are valid. neon spectra. Ongoing instrument qualification is done analo-
Thus, the primary task to ensure analyzer calibration is to gously.
ensure the spectra are calibrated and standardized before the
8.2.1 Example Neon Spectrum—Y-axis: counts, X-axis: Ra-
analyzer is commissioned. There also needs to be a means to
man shift (cm-1) (see Fig. 1).
ensure this calibration remains valid over time by using
instrument qualification approaches. A spectrum is essentially a 8.3 Intensity Axis Calibration—Optical Path Wavelength
two-dimensional chart with the number of detected photons on Transmission Effıciency—The detection efficiency of signal
the vertical or intensity axis and the wavelength of those photons traveling from the sample, through the probe optics,
photons on the horizontal axis. The mathematical method used through the fiber, and finally, to the detection module, varies
to generate analytic results from the spectra relies on knowing with wavelength. Thus, different molecular vibrations have

FIG. 1 Example Neon Spectrum

7
D8470 − 22
different detection efficiencies. These efficiencies are unique to 8.3.1 – 8.3.4. This is typically done annually or at a convenient
each instrument. Thus, for a method to be valid and accurate, time when the sample point is out of service. In some cases, it
these variations shall be corrected to yield accurate relative is desired to check the intensity axis calibration without
abundance calculations. During the system installation process, removing the probe from its stream mounting. This can be
just before commissioning, reference light with a known accommodated by considering that the system components that
intensity versus wavelength profile is transmitted through the can affect intensity calibration are the probe, the fiber-optic
entire optical path by exposing it to the probe collection cables, and the detection module. The detection module can be
aperture. A spectrum is taken of this light source, which is then checked for any changes in intensity calibration by comparing
divided by the known profile, and the result is then stored as a the ratios of neon peak heights as discussed in 8.3.4. Note that
reference spectrum on the instrument being fabricated. This this is done automatically and periodically. The fiber-optic
reference spectrum is then used to correct subsequent sample cable and the probe are passive optical devices that are not
spectra to obtain spectra that are independent of the spectro- subject to wear or aging absent external modification or
graph throughput. There are several ways to produce this contamination and, thus, do not require ongoing instrument
reference light.
qualification. Any change that could affect intensity calibration
8.3.1 Broadband Light from Incandescent Bulb—The light is typically caused either by contamination of the probe
source is typically a tungsten halogen bulb with an integral window or fiber connections, or physical damage of the fiber or
diffuser to ensure uniform illumination, driven by constant
probe. Either of these two scenarios would produce noticeable
current electronics to ensure stable operation. The bulb is
effects in the Raman spectra that can be detected and acted
typically calibrated relative to a National Metrology Institute
upon. The two most likely effects are a large drop in signal or
(NMI) traceable reference source such as National Institute of
a large rise in the baseline background of the spectra, or both.
Standards and Technology (NIST) thus creating a table of
values relating the radiative intensity of the bulb/diffuser
combination at each wavelength over the spectral range of the 9. Multivariate Measures of Instrument Performance
analyzer. The bulb/diffuser combination is presented to the 9.1 Level A Tests—A Level A performance test is a pass/fail
collection aperture of the probe and calibration spectra are test that is sensitive to many of the univariate performance
taken as described above. The bulb must be recertified at the parameters discussed in Section 7. Level A tests do not identify
factory on a yearly basis. This is the preferred approach for specific failure modes, but merely indicate if the instrument
performing intensity calibration. performance is within historical bounds. In this test, the
8.3.2 Fluorescent Material—A number of materials readily spectrum of an instrument qualification sample, an instrument
fluoresce when illuminated by laser light producing a broad- performance verification sample or an optical filter is compared
band light that has a known spectral profile. NIST produces to a historical spectrum of the instrument qualification sample,
several types of glass doped with dyes that are specifically the instrument performance verification sample or the optical
engineered to calibrate Raman instruments. The glass is filter by multivariate methods (least squares fitting or a
presented to the active probe beam and a reference spectrum is
PCR/PLS model; see Practice D8321 for descriptions of PCR
taken as described above.
and PLS). This procedure can provide some information about
8.3.3 Raman Spectra of Reference Material—Instead of a specific instrument parameters, but essentially looks for devia-
broadband light source, the Raman spectrum of a known mix
tions in the residual spectrum as compared to the historical
of materials can be used as a reference. The mix should contain
residual spectra.
each of the materials, in gravimetrically established
percentages, that are intended to be measured during analyzer 9.1.1 Level A tests are generally applied on instruments
operation. The relative intensity of each band in the reference which are in use for multivariate, quantitative analysis. The
spectrum will be established by the known Raman signature of spectral range used in Level A tests should be comparable to
the material and its known molar percentage of the mix. that used in the calibration model for the analysis being
8.3.4 Spectra from Alternative Sources—Although not used performed. If the spectrum of the instrument qualification
for intensity axis calibration and standardization, the same light sample, the instrument performance verification sample or the
source used for wavelength calibration and standardization can optical filter used in the Level A test contains absorptions that
be used to validate the intensity axis. For example, each time are significantly higher than those of the typical samples being
a neon spectrum is collected, a series of ratios of the individual analyzed, then these peaks can be excluded from the Level A
peaks is calculated. The most robust practice is to generate a fit.
ratio of each peak with respect to other peaks being used for the 9.1.2 Level A Tests Using a Least Squares Method:
calibration (typically, this is eight to ten peaks). A set of these 9.1.2.1 In this Level A test, a least squares fit of the current
ratio values are then stored on the analyzer during commis- spectrum of the instrument qualification sample, instrument
sioning to be used as an instrument qualification reference. performance verification sample or optical filter is conducted
8.4 Ongoing and Field Calibration/Instrument against a historical spectrum of the same material. Baseline
Qualification—As presented in 8.2, the wavelength axis is terms may be included in the fit to compensate for variations in
calibrated and automatically validated periodically. Similarly, baseline, and scaling may be applied to compensate for
the intensity axis can be validated by periodically exposing the pathlength variations. The types of compensations (baseline or
probe to one of the reference light source options covered in pathlength) used in the fit should be similar to those employed

8
D8470 − 22
in the multivariate model used for the actual analyzer measure- done in the Level A test model should be comparable to that
ment. Methodology for calculating the least square fit is done in the multivariate calibration models being used on the
discussed by Blackburn (4) and by Antoon, Koenig, and instrument.
Koenig (5). NOTE 3—Chemometricians might refer to the analysis described in
9.1.2.2 A typical least squares model could be as follows: 9.1.3 as principal component analysis rather than principal components
regression. However, the object here is to allow the Level A test to be
g 5 a h h1a λ λ1a 1 11e (5)
developed and applied using the same chemometric software employed in
where: the development and application of the multivariate calibration model.
g = column vector containing the current spectrum of the 9.1.3.1 The PCR or PLS model is used to analyze the
instrument qualification sample, the instrument perfor- current spectrum of the instrument qualification/instrument
mance verification sample or the optical filter, performance verification or optical filter. The estimate for the
h = column vector containing the historical spectrum of the current spectrum is generated. The spectral residuals are
instrument qualification sample, the instrument perfor- calculated as the difference between the current spectrum and
mance verification sample or the optical filter, its estimate. The spectral residual can be charted to determine
λ = (v for frequency-based spectra), column vector of the if the instrument is operating within historical specifications.
wavelength axis values for spectra g and h, 9.1.3.2 PCR and PLS models can also provide information
1 = column vector of ones, on the scaling of the current spectrum. For simple models,
ah = coefficient for scaling the historical spectrum to match variation in the estimated composition from 100 % is an
the current spectrum, indication of scaling variation.
aλ = coefficient which scales λ to provide a baseline correc-
9.2 Level B Tests—Level B performance tests analyze the
tion which is linear in wavelength (or frequency),
a1 = coefficient for a baseline offset, and spectrum of an instrument qualification sample, an instrument
e = the error in the fit. performance verification sample or an optical filter against
multivariate models that are employed during the normal use of
The column vectors h, λ, and 1 are combined into a matrix the analyzer system. As such, Level B tests cannot be per-
H, and the coefficients into a matrix A. formed during the multivariate calibration of a analyzer. Level

g 5 @h λ 1# FG
ah

a1
1e 5 HA1e (6)
B tests monitor the instrument performance for deviations to
which a calibration model is sensitive. Tests on a limited
number of samples are not rigorous, but failures in these tests
are indicative that the instrument operation has changed.
9.1.2.3 The estimated coefficients â are first determined by 9.2.1 The spectrum of the instrument qualification sample,
linear least squares. the instrument performance verification sample or the optical
â 5 H 1 g 5 ~ H t H ! 21 H t g (7) filter is analyzed using the multivariate model used during
normal instrument operation. The predicted value (property or
9.1.2.4 The coefficients are then used to estimate the spec- component concentration), the Mahalanobis distance, and the
trum of the current sample ĝ. spectral residuals are again compared to historical values to
ĝ 5 Hâ (8) detect any change in the analyzer performance.
9.1.2.5 The residuals from the fit are the difference between 10. Performance Test Charts
the measured and estimated values for the data points, g – ĝ.
10.1 Performance test results should be plotted on charts
The residuals from the fit are squared and summed. The
and examined for trends. Such trend analysis may provide
resulting measure, herein referred to as the spectral residual, is
early warnings of possible analyzer problems.
used as a measure of changes in the instrument performance.
This spectral residual should be plotted on control charts. 10.2 Statistical quality control charting methods (for
example, individual value control charts, exponentially
NOTE 2—Any function of the sum of the squares of the residuals can be weighted moving average control charts, and moving range of
used, for example, the square root, or the square root of the sum divided
two control charts) can be used to detect statistically significant
by the number of points.
changes in instrument performance. However, the statistical
9.1.2.6 Additionally, the scaling and baseline coefficients control limits associated with these charts will not necessarily
can be monitored as an additional measure of instrument be useful in judging the performance test results. Instead, some
performance. performance test results are typically compared to action limits
9.1.3 Level A Tests Using a PCR or PLS Method—To as described in Section 10.
perform a Level A test using PCR or PLS, one must first 10.2.1 For some performance tests, the test results are
develop an appropriate model. A series of historical spectra for expected to trend continuously in one direction until such time
the instrument qualification/instrument performance verifica- as the analyzer is serviced. For example, the energy output of
tion sample, or the optical filter are analyzed by a PCR or PLS an infrared source is expected to decrease continuously as the
regression algorithm using 100 % for the compositional value source ages, until such time as the source is cleaned or
to generate the Level A model. Generally, only one model replaced. The decreased energy may be observed as an increase
variable should be retained in the model since all the spectra in the Level 0 photometric noise, or as an increase in the Level
are of the same material. The type of pre- or postprocessing A spectral residual. The daily change in energy, noise or

9
D8470 − 22
residual may be large relative to the precision with which these collected, and the analyst can examine the spectra to determine
values can be measured but have tolerable effect on the if the observed change in performance would distort the spectra
instrument results for the intended application. For such tests, sufficiently to cause misidentifications to occur.
control charts and statistically derived limits may be inappro- 11.2.3 If the normal operation of the instrument involves the
priate. An action limit for such tests needs to be determined use of spectral library searching or discriminative analysis for
from historical data or simulations as discussed in Section 10. materials identification or classification, then known (instru-
10.2.2 For some performance tests, the test results are ment qualification or instrument performance verification
expected to vary randomly about a fixed point. For example, sample) materials should be reanalyzed whenever a change in
for a properly operating instrument, the Level 0 wavelength performance is detected to determine if the change reduces the
value might be expected to vary randomly about some average discrimination power of the analysis.
value. For such tests, the statistical control charts and control 11.2.4 If the normal operation of the instruments involves
limits can be usefully employed to set initial action limits in the quantitative analysis, then quality assurance for the quantita-
absence of historical data. Such initial action limits may be tive analysis should be part of the normal operation. The
loosened if statistically significant performance changes de- significance of changes in instrument performance are judged
tected by the control charts are not found to have significant by their effects on the quality control results for the normal
effect on the instrument results for the intended application. analysis.
10.2.3 Since, for instruments used in multivariate analyses, 11.2.4.1 If the quantitative method used during normal
Level B composition or property results for instrument quali- operation is a primary method (for example, univariate cali-
fication or instrument performance verification samples are bration against quantitatively prepared or certified standards),
most directly comparable to actual analyzer results, the Level then the quality control for the quantitative method will
B composition or property estimates are most amenable to generally involve measurement of a standard and comparison
statistical control charting. Action limits for Level B compo- of the results against the known value.
sition or property estimates can be set to statistically deter- 11.2.4.2 If the quantitative method used during normal
mined control limits. operation is a secondary method (for example, univariate or
multivariate calibration against results obtained by another
11. Performance Test Action Limits primary analytical method), then the quality control for the
11.1 Spectroscopic analyses differ greatly in their sensitivity quantitative method will generally involve periodic analyses of
to various aspects of instrument performance, and each appli- an unknown by both methods and statistical comparison of the
cation differs in what constitutes an acceptable tolerance to results.
changes in the results caused by variations in instrument 11.2.4.3 If the quantitative method used during normal
performance. Although instrument performance tests are useful operation is within statistical quality control, then the results
in their own right, the user must be concerned with how for the performance tests conducted during the same time
changes in the instrument performance affect results during period should be considered an example of acceptable instru-
normal operation. Historical databases or simulations that ment performance and added to the historical database.
define acceptable performance for one application may not be 11.2.4.4 If the quantitative method used during normal
appropriate for another application. In addition, the level of operation is not within statistical quality control, the results
performance required by quantitative applications may be from the performance tests may be examples of unacceptable
changed by the updating of the calibration. instrument performance, particularly if the results from the
11.2 Setting Action Limits Based on Historical Data for performance tests are inconsistent with the historical database.
Performance Tests: Examples of unacceptable instrument performance can be used
11.2.1 Performance tests provide quantitative measures of to set action limits for future performance tests.
instrument performance. These measures can be compared to 11.2.4.5 It is strongly recommended that, at the time the
historical data for the same tests in order to recognize changes quantitative method is developed, spectra of the instrument
of instrument performance. If historical data exist, limits for qualification sample, the instrument performance verification
each test can be set and the performance can be judged against sample or the optical filter be collected along with spectra used
these limits. If historical data does not exist, it will be for calibration. Performance tests can be applied to this data to
necessary to collect them as a standard part of the instrument determine the level of performance at the time of calibration
operation, and such collection will eventually allow perfor- 11.2.4.6 Changes in analyzer performance that are detected
mance limits to be established. The collection of the historical by Level 0 tests may or may not produce a significant change
database for performance tests should be an integral part of the in the results during normal operation. Different types of
instrument operation and be continued for the life of the quantitative analyses differ significantly in their sensitivity to
instrument. various aspects of analyzer performance. By plotting the Level
11.2.2 For instruments used in qualitative analysis, what 0 test results with the quality control results for the quantitative
constitutes acceptable instrument performance may be difficult method on control charts, conditions that lead to invalid
to quantify. The expert analyst may only know that the spectra quantitative results can be identified, and action limits for each
“look wrong.” If a change in instrument performance is Level 0 test can eventually be established.
detected, spectra of known (instrument qualification or instru- 11.2.4.7 For instruments where normal operation involves
ment performance verification sample) materials can be use of a multivariate calibration model, increases in the

10
D8470 − 22
spectral residuals that are detected by Level A tests will these spectra to simulate the expected variations in the instru-
generally reflect some change in the quantitative results pro- ment performance. The model sensitivity, for example, the
duced by the instrument. Even if the analysis result does not change in the results per unit change in a performance
change, the spectral residuals measured as part of the outlier parameter, can be estimated, and used to establish action limits
testing will generally be expected to increase. The level of for each performance parameter based on the error tolerance
increase that can be tolerated can be determined by plotting the for the application. Instrument performance parameters which
Level A test spectral residuals against analyzer results and can be modeled include wavelength (frequency) shifts, base-
determining the maximum level at which valid analysis results line shifts, changes in photometric noise, resolution changes,
are produced. and detector linearity changes. The importance of different
11.2.4.8 For instruments where normal operation involves performance parameters is both application and instrument
use of a multivariate calibration model, changes in the values type dependent. Historical data for Level 0 performance tests
produced by a Level B test are the most straightforward to are the best practice to the type response changes that should be
interpret since the values are directly comparable to the modeled for a given instrument type.
analysis results. If the analysis of the spectrum of the instru- 11.3.1.1 For example, the sensitivity of an analyzer to
ment qualification sample, the instrument performance verifi- baseline drift can be simulated by adding various baselines to
cation sample, or the optical filter is an interpolation of the a set of representative spectra, analyzing these spectra with the
model, then limits can be set directly based on the desired calibration model, and determining the change in the results as
performance of the analyzer. If the analysis of the spectrum of a function of the added baseline. The added baseline can, for
the instrument qualification sample, the instrument perfor- example, be parameterized in terms of offset, slope and
mance verification sample or the optical filter is an extrapola- curvature so that the effects of each can be determined.
tion of the model, then some care must be exercised in setting
limits since the extrapolated result may be more sensitive to 11.3.1.2 For example, the sensitivity of an analyzer to
small changes in instrument performance than analyses that are wavelength (frequency) shift can be simulated by shifting the
interpolations of the model. This is known as leverage. In this wavelength (frequency) of a set of representative spectra (see
case, initial limits should be confirmed by plotting the Level B Annex A1), analyzing these spectra with the calibration model,
results against analysis results and determining the levels at and determining the change in the results as a function of the
which valid analysis results are produced. shift. If the shift is accomplished via interpolation of the
spectra, care must be exercised that the interpolation function
NOTE 4—Any one instrument performance verification sample, instru- does not smooth or deresolve the spectra.
ment qualification sample, or optical filter only tests a small portion of the
multivariate model space and may not be sensitive to all aspects of 11.3.2 Changes in instrument performance seldom effect
instrument performance. The Level 0, A, and B tests are intended to detect only one aspect of that performance. If simulations are used to
possible instrument failure modes. Acceptable performance as measured set action limits for performance tests, it is essential that
by Level 0, A and B tests is necessary but not sufficient by themselves for multiple performance parameters by varied simultaneously.
demonstrating valid instrument performance. Comparison of analysis
results to in control, primary method laboratory values is also necessary to The magnitude of the changes to the performance parameters
demonstrate the validity of analysis results. that should be simulated are best obtained from examination of
11.3 Determining Performance Action Limits by Simulating historical data on Level 0 performance tests conducted on the
Instrument Response Changes: type of instrument being tested.
11.3.1 For instruments where normal operation involves use
of a multivariate calibration model, an alternative procedure 12. Keywords
for determining action limits for instrument performance tests 12.1 infrared spectroscopy; instrument performance; level
is to take actual, diverse, but representative spectra that are 0; molecular spectrophotometer; near-infrared; process ana-
predicted well by the model, and to mathematically modify lyzer; Raman; spectrophotometer performance tests

11
D8470 − 22

ANNEX

(Mandatory Information)

A1. PROCEDURE FOR TESTING PEAK FINDING ALGORITHMS

A1.1 Interpolation of Test Spectrum points at each end) is treated as point 0 once. The values
A1.1.1 To test a peak finding algorithm, it is necessary to Ai(i = 0.2, 0.4, 0.6 and 0.8) calculated for the repetitive
interpolate and shift a test spectrum by amounts equivalent to application of the filter are collected into 4 spectral vectors.
a fraction of the digital resolution of the spectrum. The These vectors are assigned the same frequency scales as the
following procedure can be used to interpolate and shift the original spectrum. If ∆ is the digital resolution (data point
spectrum. spacing), then the vectors A0.8, A0.6, A0.4, and A0.2 generated by
A1.1.1.1 Four adjacent points in the spectrum are selected. repetitive application of the above procedure represent spectra
The points are numbered –1, 0, 1, and 2. The spectral intensity which are shifted by –0.2∆, –0.4∆, –0.6∆, and – 0.8∆, respec-
at the four points is fit to a cubic equation. tively.
A2 1 2 4 8 C2 NOTE A1.1—If the data points are numbered –1, 0, 1, and 2 from left to

3 43
A1
A0
A 21
43 45
1 1
1 0
1 21
1 1
0 0
1 21
C1
C0
C 21
↔A 5 XC (A1.1)
right, the interpolation estimates a value for the spectrum between point 0
and point 1. The original point at 0 is replaced with this estimate. The
effect is to shift the spectrum toward the left (negative index) direction.

C2 0 0 6 0 A2 A1.2 The peak finding algorithm is applied to the original

3 43
C 5 X 21 A↔
43 4C1
C0
C 21
21 6 23 22
0 3 26 3
1 23 3 21
A1
A0
A 21
and shifted spectra. The peak positions for suitable peaks are
calculated.
A1.2.1 If the peak positions calculated by the peak finding
(A1.2) algorithm for the shifted spectra are shifted by –0.2∆, –0.4∆,
A1.1.2 The cubic equation is evaluated at four intermediate –0.6∆, and –0.8∆, respectively, relative to the peaks for the
points: original spectrum, then the peak finding algorithm can be used
A 0.8 1 0.8 0.64 0.512 C2 for tracking frequency (wavelength) stability.

3 43
A 0.6
A 0.4
A 0.2
5
1
1
1
0.6
0.4
0.2
0.36
0.16
0.04
0.216
0.064
0.008
43 4 C1
C0
C 21
(A1.3) A1.2.2 If the peak positions calculated by the peak finding
algorithm for the shifted spectra are not shifted by –0.2∆,
–0.4∆, –0.6∆, and –0.8∆, respectively, relative to the peaks for
A1.1.3 The procedure is repeated, moving the filter side- the original spectrum, then the peak finding algorithm cannot
ways so that each point in the spectrum (other than the two be used for tracking frequency (wavelength) stability.

REFERENCES

(1) Savitzky, A. and Golay, M. J. E., “Smoothing and Differentiation of (3) Willson, P. D. and Polo, S. R., “Polynomial Filters of Any Degree,”
Data by Simplified Least Squares Procedures,” Analytical Chemistry, Journal of the Optical Society of America, Vol 71, 1981, pp. 599–603.
Vol 36, pp. 1627–1639, 1964. (4) Blackburn, J. A., Analytical Chemistry, Vol 37, 1965, p. 1000.
(2) Steiner, J., Termonia, Y., and Deltour, J., “Comments on Smoothing (5) Antoon, M. K., Koenig, J. H., and Koenig, J. L., Applied
and Differentiation of Data by Simplified Least Squares Procedures,” Spectroscopy, Vol 31, 1977, p. 578.
Analytical Chemistry, Vol 44, p. 1906, 1972.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

12

You might also like