You are on page 1of 15

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: D8473 − 22

Standard Test Method for


Determining the Biobased content of Liquid Hydrocarbon
Fuels Using Liquid Scintillation Counting with Spiked
Carbon-141
This standard is issued under the fixed designation D8473; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope included here. These are best obtained from the manufacturer
1.1 This test method covers quantitatively determining bio- of the specific instrument in use.
carbon content of liquid hydrocarbon fuels with a focus on 1.5 Pre-Requisite Requirements For Method Execution—
those produced in a typical petroleum refinery using liquid This test method uses artificial carbon-14 (14C) within the
scintillation counting (LSC). The method is designed to gen- method. Great care shall be taken to prevent laboratory
erate analogous results as Test Method D6866 Method C, for contamination of the elevated 14C. Once in the laboratory,
low quench samples, without the need of benzene synthesis. artificial 14C can contaminate a variety of laboratory surfaces
The purpose is to be able to use the produced data to report that can lead to artificially high sample biocarbon measure-
biocarbon content of refinery products to regulatory agencies ments. If vigorous cleaning attempts to remove the artificial
and monitor refinery operation. The method does not address 14
C from a laboratory are unsuccessful, instrumentation and
regulatory reporting or fuel performance. sample preparation may have to be moved to a new laboratory
1.2 The method is needed to support refinery operations away from the contamination or the laboratory may have to
when bio-feeds are co-processed with petroleum within a rely on outside third-party labs for analysis. Specific proce-
reactor with a focus on samples with 100 % biocarbon or less dural steps have been incorporated into this method that
(not for 14C labeled species). It allows refineries to report the minimize the risk of sample and lab contamination. Wipe tests
biocarbon content of refinery products to regulatory agencies and quality assurance samples can validate absence of con-
such as the Environmental Protection Agency (EPA) or Cali- tamination. In the event of contamination in the laboratory or
fornia Air Resources Board (CARB) to comply with regulatory instrument, vigorous cleaning protocols shall be implemented,
statutes such as The Renewable Fuel Standard (RFS) or Low and analysis cannot be resumed until the lab and instrument are
Carbon Fuel Standard (LCFS). free of contamination. Accepted requirements are:
1.3 This test method is applicable to any liquid fuel product, 1.5.1 Working with the elevated 14C samples in a separate
petroleum based (pure hydrocarbon), biobased (such as renew- and defined area away from the instrument and the preparation
able diesel or those that can contain oxygenates such as of any non-spiked samples.
ethanol), or blends, that contain 1 % to 100 % by mass 1.5.2 Using separate personnel to prepare the spiked
biocarbon where an instrument background can be experimen- samples and non-spiked samples.
tally determined using a sample of similar matrix that contains
1.5.3 Using separate laboratory spaces with separate HVAC
no measurable carbon-14.
systems for the handling of spiked and non-spiked samples.
1.4 This test method makes no attempt to teach the basic The use of separate fume hoods that have separate exhaust
principles of the instrumentation used although minimum ventilation satisfies this requirement.
requirements for instrument selection are referenced in Refs 1.5.4 Weekly wipe tests of 14C sample handling area(s) to
(1-11).2 However, the preparation of samples for the above test detect lab contamination.
methods is described. No details of instrument operation are
1.6 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
1
responsibility of the user of this standard to establish appro-
This test method is under the jurisdiction of ASTM Committee D02 on
Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of
priate safety, health, and environmental practices and deter-
Subcommittee D02.04.0F on Absorption Spectroscopic Methods. mine the applicability of regulatory limitations prior to use.
Current edition approved Sept. 15, 2022. Published October 2022. DOI: 1.7 This international standard was developed in accor-
10.1520/D8473-22.
2
The boldface numbers in parentheses refer to a list of references at the end of dance with internationally recognized principles on standard-
this standard. ization established in the Decision on Principles for the

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

1
D8473 − 22
Development of International Standards, Guides and Recom- 3.2.5 biofeed, n—a feedstock sourced from a plant or
mendations issued by the World Trade Organization Technical animal.
Barriers to Trade (TBT) Committee. 3.2.6 biogenic, adj—containing carbon (organic and inor-
ganic) of renewable origin like agricultural, plant, animal,
2. Referenced Documents
fungi, microorganisms, marine, or forestry materials.
2.1 ASTM Standards:3
3.2.7 biogenic carbon content, n—amount of biogenic car-
D5291 Test Methods for Instrumental Determination of
bon in the material or product as a percent of the total carbon
Carbon, Hydrogen, and Nitrogen in Petroleum Products
(TC) in the product.
and Lubricants
D6299 Practice for Applying Statistical Quality Assurance 3.2.8 carbon-14 (14C), n—naturally occurring radioactive
and Control Charting Techniques to Evaluate Analytical isotope of carbon that contains six protons and eight neutrons
Measurement System Performance with a true half-life of 5730 years.
D6300 Practice for Determination of Precision and Bias 3.2.9 cocktail, n—the solution in which samples are mixed
Data for Use in Test Methods for Petroleum Products, for measurement in an LSC.
Liquid Fuels, and Lubricants
D6708 Practice for Statistical Assessment and Improvement 3.2.10 coincidence circuit, n—a portion of the electronic
of Expected Agreement Between Two Test Methods that analysis system of an LSC which acts to reject pulses that are
Purport to Measure the Same Property of a Material not received from the two or three photomultiplier tubes (that
D6866 Test Methods for Determining the Biobased Content count the photons) within a given period of time and are
of Solid, Liquid, and Gaseous Samples Using Radiocar- necessary to rule out background interference and required for
bon Analysis any LSC used in this test method.
2.2 Other Standards: 3.2.11 coincidence threshold, n—the minimum decay en-
DIN 51637 Liquid Petroleum Products – Determination of ergy required for an LSC to detect a radioactive event; the
the Bio-Based Hydrocarbon Content in Diesel Fuels and ability to set that threshold is a requirement of any LSC used
Middle Distillates Using Liquid Scintillation Method4 in this test method.
CSN EN 16640 Bio-based products - Bio-based carbon 3.2.12 coincidence time, n—the time period used by the
content - Determination of the Bio-Based Carbon Content coincidence circuit that is used to determine if a detection event
using the Radiocarbon Method5 is counted or rejected.
3. Terminology 3.2.13 contemporary carbon, n—a direct indication of the
relative contributions of fossil carbon and “living” biospheric
3.1 The definitions of terms used in this test method are
carbon can be expressed as the fraction (or percentage) of
referenced in case the practitioner requires further information
contemporary carbon, symbol fC; this is derived from “fraction
regarding the practice of the art of isotope analysis and to
of modern” (fM) using the observed input function for atmo-
facilitate performance of these test methods.
spheric 14C over recent decades, representing the combined
3.2 Definitions: effects of fossil dilution of 14C (minor) and nuclear testing
3.2.1 background radiation, n—the radiation in the natural enhancement (major); the relation between fC and fM is
environment; this includes cosmic radiation and radionucle- necessarily a function of time; by 1985, when the particulate
otides present in the local environment, for example, materials sampling discussed in the cited reference was performed, the
of construction, metals, glass, and concrete. fM ratio had decreased to approximately 1.2 (10, 11).
3.2.2 background sample, n—sample with no detectable 14C 3.2.14 counting time, n—the total time used by the liquid
(such as a carboniferous sample that should not contain any scintillation counter to count sample 14C decays.
measurable 14C because of its geologic age).
3.2.15 counts per minute (cpm), n—the average number of
3.2.3 biobased, adj—containing organic carbon of renew- counts the liquid scintillation counter detections during analy-
able origin like agricultural, plant, animal, fungi, sis; is used to derive dpm.
microorganisms, marine, or forestry materials living in a
natural environment in equilibrium with the atmosphere. 3.2.16 decay (radioactive), n—the spontaneous transforma-
tion of one nuclide into a different nuclide or into a different
3.2.4 biobased carbon content, n—the amount of biobased
energy state of the same nuclide; the process results in a
carbon in the material or product as a percent of the total
decrease, with time, of the number of original radioactive
organic carbon (TOC) in the product.
atoms in a sample, according to the half-life of the radionu-
clide.
3
For referenced ASTM standards, visit the ASTM website, www.astm.org, or 3.2.17 delay time, n—the time the instrument waits after
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM method is started before it starts counting; allows user to delay
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. the start of analysis to allow sample photoluminescence to stop
4
Available from Deutsches Institut für Normung e.V. (DIN), Am DIN-Platz, before counting is initiated.
Burggrafenstrasse 6, 10787 Berlin, Germany, http://www.din.de.
5
Available from European Standards, Krimicka 134, 318 00 Pilsen, Czech 3.2.18 discriminator, n—an electronic circuit which distin-
Republic, VAT: CZ09567909, https://www.en-standard.eu/. guishes signal pulses according to their pulse height or energy;

2
D8473 − 22
used to exclude extraneous radiation, background radiation, 3.2.32 pulse, n—the electrical signal result in when photos
and extraneous noise from the desired signal. are detected by the PMTs.
3.2.19 disintegrations per minute (dpm), n—the quantity of 3.2.33 quenching, n—any material that interferes with the
radioactivity; the measure dpm is derived from cpm or counts (accurate) optimal conversion of decay energy to scintillation
per minute (dpm = (cpm – cpm of background) / counting photons captured by the PMT of the LSC (a significant
efficiency); there are 2.2 × 106 dpm / µCi. interference with this test method).
3.2.20 effıciency, n—the ratio of measured observations or 3.2.33.1 chemical quenching, n—a reduction in the scintil-
counts compared to the number of decay events which oc- lation intensity seen by the PMT due to the materials present in
curred during the measurement time; expressed as a percent- the sample that interfere with the processes leading to the
age. production of light.
3.2.21 figures of merit, n—a term applied to a numerical 3.2.33.2 color quenching, n—any material that absorbs
value used to characterize the performance of a system; in generated scintillation photons.
liquid scintillation counting, specific formulas have been 3.2.34 region, n—regions of interest, also called window or
derived for quantitatively comparing certain aspects of instru- channel, or both, regarding LSC; refers to an energy level or
ment and cocktail performance and the term is frequently used subset specific to a particular isotope.
to compare efficiency and background measures (refer to Eq 1).
3.2.35 renewable, n—being readily replaced and of non-
3.2.22 fluorescence, n—the emission of light resulting from fossil origin; specifically, not of petroleum origin.
the absorption of incident radiation and persisting only as long
3.2.36 scintillation, n—the sum of all photons produced by
as the stimulation radiation is continued.
a radioactive decay event; counters used to measure this as
3.2.23 fossil carbon, n—carbon that contains no measurable described in this test method are Liquid Scintillation Counters
radiocarbon because its age is very much greater than the (LSC).
5730-year true half-life of 14C.
3.2.37 scintillation reagent, n—chemicals that absorbs de-
3.2.24 half-life, n—the time in which one half the atoms of cay energy transferred from the solvent and emits light
a particular radioactive substance disintegrate to another (photons) proportional in intensity to the decay energy.
nuclear form; the true half-life of 14C is 5730 years.
3.2.38 solvents and scintillators, n—chemicals that absorb
3.2.25 intensity, n—the amount of energy, the number of decay energy transferred from the solvent and emits light
photons, or the numbers of particles of any radiation incident (photons) proportional in intensity to the deposited energy.
upon a unit area per unit time. 3.2.39 solvent-in scintillation reagent, n—chemical(s)
3.2.26 internal standard, n—a known amount of radioactiv- which act as both a vehicle for dissolving the sample and
ity which is added to a sample to determine the counting scintillator and the location of the initial kinetic energy transfer
efficiency of that sample; the radionuclide used shall be the from the decay products to the scintillator; that is, into
same as that in the sample to be measured and have a certified excitation energy that can be converted by the scintillator into
activity. photons.
3.2.27 luminescence, n—scintillation flux of photons; can 3.2.40 specific activity (SA), n—refers to the quantity of
produce coincidences if intense; causes—luminescent reac- radioactivity per mass unit of product, that is, decays per
tions in sample, photoluminescence from bright, especially UV minute per gram of carbon (dpm/gC).
excitation. 3.2.41 standard count conditions (STDCT), n—LSC condi-
3.2.28 modern carbon, n—explicitly, 0.95 times the specific tions under which reference standards and samples are
activity of SRM 4990B (the original oxalic acid radiocarbon counted.
standard), normalized to δ13C = −19 % (10); functionally, the 3.2.42 triple to double count ratio (TDCR), n—the ratio of
fraction of modern carbon equals 0.95 times the concentration counts detected by three detectors over the number of counts
of 14C contemporaneous with 1950 wood (that is, pre- that were detected by two detectors; requires use of a liquid
atmospheric nuclear testing). To correct for the post 1950 bomb scintillation counter with three detectors.
14
C injection into the atmosphere, the fraction of modern
carbon is multiplied by a (REF) atmospheric adjustment value 3.2.43 wipe test, n—a test that is done to determine if a
representative of the excess 14C in the atmosphere at the time surface has been contaminated with 14C or any other radioac-
of measurements. tive isotope.
3.2.29 noise pulse, n—a spurious signal arising from the 4. Summary of Test Method
electronic and electrical supply of the instrument.
4.1 Fuels or their component streams, or both, are mixed
3.2.30 operational quality test (OQ), n—test done to verify with scintillation cocktail and placed into a liquid scintillation
liquid scintillation counter is working properly. counter with TDCR capabilities. The instrument is then used to
3.2.31 photomultiplier tube (PMT, pmt), n—the device in determine the number of 14C disintegrations per minute per
the LSC that counts the photons of light simultaneously at two gram of carbon in the sample to calculate its biological carbon
or three separate detectors. content. 14C spiking using a separate aliquot of the same

3
D8473 − 22
sample, and prepared at a different location, is used to amplitude luminescence pulses, and employ delay time before
determine counting efficiencies of each sample. counting starts, to allow photoluminescence to fade.
6.4 Sample and scintillation cocktail volumes need to be
5. Significance and Use consistent to obtain accurate efficiency measurements. Varia-
5.1 This test method provides accurate biobased/biogenic tions in the sample to cocktail ratio will change the dilution of
carbon content results to materials whose carbon source was the quenching agents present in the sample resulting in an
directly in equilibrium with CO2 in the atmosphere at the time increase or decrease in counting efficiency if too little sample
of cessation of respiration or metabolism, such as the harvest- or too much sample is added in respect to the scintillation
ing of a crop or grass living in a field. Special considerations cocktail, respectively. Measured volumes shall not fluctuate
are needed to apply the testing method to materials originating more than 0.1 mL from the target volume to ensure data
from within artificial environments with non-natural levels of accuracy.
14
C or if the biofeed was grown over the course of several 6.5 Biological carbon contamination from outside sources,
years such as trees and contains “bomb-carbon.” Application of sources of fossil or modern carbon such as corks, paper towels,
these test methods to materials derived from CO2 uptake within plant-based rope/string or dust shall be kept away from the
artificial environments is beyond the present scope of this samples to prevent contamination of the sample.
standard.
6.6 14C Spike Standard Contamination in Samples—
5.2 This method uses LSC techniques to quantify the Precaution is needed when preparing and handling the spike
biobased content of a liquid hydrocarbon fuels using sample solution and spiked samples to prevent sample and laboratory
carbon that has been unmodified. It is designed to be able to contamination. Review 1.5 to make sure all necessary precau-
incorporate into a refinery laboratory to support biofeed and tions are being followed.
petroleum coprocessing or blending operations to determine
6.7 Sample Volatility—Addition of the sample to the vial
the biocarbon content of the intermediate or finished products.
should always be the last step. When samples are weighed, a
The test results can then be used for optimizing internal
cap needs to be tightly sealed when recording the final sample
parameters or reporting to regulatory agencies.
mass measurement and the cap should not be reopened before
5.3 The use of this method requires that a pure petroleum- measuring. Background samples that are collected and stored
based sample can be generated that has a similar matrix to each for extended periods of time shall be stored and sampled in a
product or stream to be analyzed. For example, gasoline and manner to prevent loss of volatiles.
diesel have very different matrices and will likely require the 6.8 Potassium-40 (40K)—A radioactive isotope of potassium
use of different background measurements for each. Refer to that is present in glassware. 40K predominantly decays through
10.2 for how to determine if the same background sample can beta decay to 40Ca. The beta decay LSC spectrum of 40K
be used for more than one product/stream. partially overlaps with that of 14C. Even low potassium glass
still has high enough levels of 40K to interfere with the results
6. Interferences
produced via this method. For this reason, this method requires
6.1 Sample Matrix—The sample matrix affects background the use of PTFE coated plastic vials and not glass vials to
and quenching, so backgrounds (samples with no detectable prevent this interference.
14
C) need to be measured for each stream or blend that is to be
6.9 Radon-222, see 14.4 for how to correct for radon
analyzed.
contamination.
6.2 Chemical/Color Quenching—Higher boiling samples 6.10 Bomb Pulse:
will likely contain higher levels of quenching molecules such 6.10.1 The pMC can be greater than 100 % because of the
as conjugated aromatics. The presence of these species in high continuing but diminishing effects of the 1950s nuclear testing
concentrations can require sample dilutions or prevent use of programs, which resulted in a considerable enrichment of 14C
this method. Counting efficiency measurements can be con- in the atmosphere. The decrease in 14C from the bomb testing
ducted to determine if the quench levels are too high. Counting programs has been nonlinear in the past but has been linear
efficiency is required to be above 50.0 %. since at least 2004 to present. As of 2019 the 14C activity in the
6.3 Luminescence—In LSC context, unwanted non- atmosphere has reached the 1950 level of 13.56 dpm per gram
scintillation flux of photons. Consists of single photons and, if carbon that is defined as 100 pMC. Because all sample 14C
moderate, does not markedly produce coincidences. If lumi- activities are referenced to a “prebomb” standard, and because
nescence intensity increases, probability for two photons hit- nearly all new biobased products are produced in a post-bomb
ting separate PMT’s within coincidence time increases and environment, all pMC values shall be adjusted by an atmo-
random coincidences start to occur. Probability for triple spheric correction factor (REF) to obtain the true biobased
coincidences remains near negligible though. Common causes content of the sample. The correction factor is based on the
of luminescence are luminescent reactions in sample (chemi- excess or deficiency of 14C activity in the atmosphere at the
luminescence) and photon emission after exposure to light, time the biological source was alive. A REF value of 102 pMC
especially UV illumination (photoluminescence). Precautions was determined for 2015 based on the measurements of CO2 in
to minimize luminescence include, avoid bright illumination air in a rural area in the Netherlands (Lutjewad, Groningen).
when preparing and handling samples, select triple coincidence The first version of Test Method D6866 – 04 in 2004 refer-
mode and/or set counting region (window) to exclude low- enced a value of 107.5 pMC and the Test Method D6866 – 10

4
D8473 − 22
version (2010) cited 105 pMC. These data points equate to a NOTE 1—Hidex 300 SL6 was used to conduct initial studies.
decline of 0.5 pMC per year. Therefore, on January 2 of each 7.2 Anti-coincidence systems with at least three PMTs
year, the values in Table 1 are used as REF through 2019, (multidetector systems).
reflecting the same 0.5 pMC decrease per year until 2019 when
NOTE 2—Hidex brand LSC’s are currently the only known commercial
the pre bomb pulse level was reached. For all REF values after
brand to have three detectors.6
2022, refer to the most recent version of Test Method D6866.
6.10.2 Atmospheric thermonuclear weapons testing was 7.3 Coincidence circuits.
extensive between 1952 and 1963. During this time period the 7.4 Optimized counting regions to provide very low back-
14
CO2 content in the air increased by 90 %. This means that a ground counts while maintaining counting efficiency greater
plant living in 1965 would measure about 190 pMC. Since the than 50.0 %. The optimization of counting regions shall be
signing of the testing ban in 1963 this signature declined to determined for each sample type to be analyzed. The level of
about 140 pMC by 1975, 120 pMC by 1985, and 101.5 pMC quench for a given sample type will change the optimal region
by 2016. The consequence of this effect is error in biobased of interest.
content analysis relating to when the biobased material used in
NOTE 3—Initial work found that counting over channels 100 to 400 was
the product was last actively part of a respiring/metabolizing optimal for a diesel fuel sample.
system. The error is predominant in products made from NOTE 4—Test Method D6866 Method C recommends an efficiency of
forestry products. The rings within trees each represent the 60 % or higher. As that method is using a uniform and constant matrix
previous growth season within which the previous year’s with little to no quenching present, this efficiency is more than adequate.
14 NOTE 5—This method analyzes the samples with no sample
CO2 signature was recorded. The center most ring of a tree
preparation, and the analysis of samples with higher levels of quench may
living today but planted in 1965 would be about 190 pMC be needed, but as the level of quench increases, it is more difficult to
whereas the outermost ring/bark would be 100.0 pMC. If this distinguish low biocarbon containing samples from the noise. Depending
tree is harvested and used in manufacturing a biobased product, on the expected biocarbon concentration of the sample, it may be more
the percent biobased carbon content of the product will be beneficial to dilute the sample to increase the counting efficiency as
sample dilution increases efficiency proportionally higher than it reduces
dependent on where the carbon came from within the tree and the amount of sample used, that is, a 25 % dilution will increase efficiency
would be the average pMC of each of the rings and their size by more than 25 % for high quench samples. Start running the sample with
in the portion of the tree used in the manufacturing process. a small dilution first and slowly increasing the dilution factor until the
6.10.3 Bomb carbon is readily identified in a product when measured efficiency is over 50.0 %. Some example dilutions ratios are
the product’s pMC value is greater than 6 pMC above the suggested here (mL sample+mL cocktail): 12+8, 10+10, 8+12, 5+15, and
2+18. 13.5 describes how to proceed if dilution is required.
prescribed correction factor (REF). A high value can be
predicted based on the origin of the manufacturing compo- 7.5 No single LSC is specified for this method. However,
nents. High values are typically observed in paper, cardboard, minimum counting efficiency and control of background inter-
forestry products, and forestry-derived chemicals. An exact ference is specified. Like all analytical instruments, LSCs
correction factor REF is not possible based strictly on the require study as to their specific components and counting
measured pMC value of the product. optimization. Currently Hidex6 is the only commercial source
of a three PMT LSC system that satisfies need stated in 7.2.
6.11 Variation in transmittance from vial to vial. Each PTFE
coated vial may have a slightly different transmittance through 7.6 Standardization of sample preparation is required.
it. As different vials are used to measure the background, NOTE 6—Initial work used 15 mL of sample with 5 mL of scintillation
sample, and sample counting efficiency the variations in cocktail.
transmittance from one vial to another will increase the error of 7.7 Unused 20 mL sample vials comprised of PTFE coated
the calculated results. plastic shall be used. Vials are not to be reused after sample
7. Apparatus analysis.

7.1 Low level LSCs with active shielding that can produce NOTE 7—20 mL PTFE coated plastic vials were used to keep back-
ground levels as low as possible and allow for higher sample volumes to
consistent background counts for a given sample matrix. decrease method detection limit.
7.8 The aforementioned optimizations shall be performed
TABLE 1 Percent Modern Carbon (pMC) Reference
using a suitable reference standard using the same reagents and
Year REF (pMC) counting parameters as the samples. The reference standard
2015 102.0
2016 101.5 used for optimization shall be similar to the samples to be
2017 101.0
2018 100.5
6
2019 100.0 The sole source of supply of the apparatus known to the committee at this time
2020 100.0 is HIDEX, Lemminkäisenkatu 62 FIN-20520 Turku, Finland. If you are aware of
2021 100.0 alternative suppliers, please provide this information to ASTM International
2022 100.0 Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee,1 which you may attend.

5
D8473 − 22
analyzed. The standard should have low volatility and have a responsibility of the user of this test method to establish
similar matrix to that of the samples to be analyzed, that is, if appropriate safety and health practices. It is also incumbent on
the samples are expected to contain oxygenates such as ethanol the user to conform to all the government legislation for their
the standard should also contain the same oxygenates in similar location, especially those that relate to the use of open
concentrations. radioactive source, in the performance of this test method.
Although 14C is one of the safest isotopes to work with, all
NOTE 8—Initial work used diesel samples to determine optimized
conditions as diesel is the most colored sample with the most complex government regulations shall be followed in the performance
matrix. of this test method. As this method utilizes elevated levels of
14
7.9 Counting interference concerns that shall be addressed C additional regulations and precautions may apply.
as part of specific instrument calibration and normalization
include luminance, chemical or color quench, static electricity, 10. Sampling, Test Specimens, and Test Units
random noise, temperature, and humidity variability. 10.1 When sampling from large tanks avoid using compo-
7.10 Alternate regions of interest parameters may be used nents that contain carbon-based material such as a cork or
based upon testing of 20, or more, 5 h counts of the same non-synthetic rope or string. Metal or glass shall be used
reference standard that record the raw data throughout 14C wherever possible when safe to do so.
spectral region (channels 5 to 650 for Hidex6 instrumentation). 10.2 If this method is to be used to support coprocessing of
Optimal counting conditions shall be established by maximiz- biofeeds with petroleum in a refinery or similar setting samples
ing the Figures of Merit (see Eq 1) to obtain the highest count shall be collected before coprocessing has started from each
efficiency and the lowest background. Counting efficiency of stream or product, that will be analyzed after coprocessing has
less than 50.0 % is unacceptable and can be improved by LSC started to allow for adequate backgrounds to be measured.
instrument optimization and sample/reagent compatibility or These samples serve as a background for the desired products/
shielding improvements. streams and will allow for background corrections to be made.
Figures of Merit 5 ~~ E 3 V ! 2 ⁄ bkg ! 3 k (1) One background will have to be prepared and measured weekly
(all other backgrounds on a monthly basis, measurement
where:
frequency outlined in more detail in 11.3.5) to verify laboratory
E = counting efficiency, is not contaminated with artificial 14C and to account for any
V = volume of sample, instrumental drift, thus make sure enough sample is collected
bkg = CPM measured over a specific LSC channel range,
as it may be needed for several years. More volatile samples,
and
such as gasoline boiling range material, shall be stored in a
k = 1 CPM / (% × mL)2.
cold area to prevent loss of volatile components over a long
NOTE 9—Eq 1 is from DIN 51637. period of time. The same background can be used for several
products/streams if the backgrounds of those products/streams
8. Reagents and Materials have been experimentally determined to be indistinguishable.
8.1 Unused 20 mL PTFE coated plastic vials. All vials are to To determine if backgrounds can be considered the same or not
be single use to avoid sample carryover and reducing the risk use a t-test after ten background measurements have been done
of lab contamination from artificial 14C present in the spiked and use the t-value for a 95 % confidence interval.
samples.
10.3 Samples shall be thoroughly mixed before adding to
8.2 Unquenched standard kit used to verify instrument the scintillation vial.
performance.
8.3 Internal standard kit 14C, compatible with sample 11. Calibration, Standardization, and Quality Control
matrix, used to determine counting efficiency of samples. (The 11.1 Apparatus—Operational quality of the instrument shall
14
C labeled compound used to create the spike shall be a be determined weekly to make sure the instrument is working
non-volatile compound (vapor pressure below 1 Pa at 25 °C) properly. This can be done with an unquenched standard of
such as cholesterol to minimize potential of the compound known 14C activity. Some commercial instruments have a
getting into the gas phase during sample preparation to reduce designated procedure and kits to do this.
risk of lab contamination.)
11.2 Quality Control (QC) Sample—A sample of known
8.4 Scintillation cocktail, miscible with sample matrix and
biocarbon content shall be analyzed weekly. Preferred charac-
not phase separate in the instrument.
teristics of the QC sample are:
8.5 Wipe test scintillation cocktail to dissolve wipes. (1) Similar biocarbon content to samples,
8.6 Wipe test strips designed to be dissolved in scintillation (2) Similar matrix to samples, and
cocktail. (3) Shelf stable with low volatility.
11.2.1 This QC shall be run following the same procedure
9. Hazards as samples as outlined in Section 13.
9.1 The specific safety and regulatory requirements associ- 11.2.2 The measured biocarbon content of the QC shall be
ated with radioactivity, sample preparation, and instrument control charted in accordance with Practice D6299 and moni-
operation are not addressed in this test method. It is the tored for:

6
D8473 − 22
11.2.2.1 Measurements trending away from the mean over 11.3.7 A control chart shall be kept for all background
time. If this variation is observed analysis shall be halted until values measured (each background sample type should have
reason for sustained variation has been determined and fixed. their own chart). Refer to 11.2.2 for control charting guide-
11.2.2.2 Single points measured outside 63 sigma from the lines.
mean. If this occurs rerun sample to verify result. If the rerun 11.3.7.1 The background that is run on a weekly basis is to
is also outside 63 sigma from the mean, analysis shall be aid in finding issues quickly so any issues can be addressed in
halted until reason for variation has been determined and fixed. a timely manner.
If the rerun is inside 63 sigma from the mean, record this data 11.3.8 If laboratory 14C contamination or an instrumental
issue has been identified, all data that has been collected since
point in the control chart and discard the other. Keep a separate
the issue occurred shall be reprepared and reanalyzed once the
record of when and the measured value for all points that are
issue has been rectified. If an exact date of when the issue
outside of the 63 sigma limit for reference.
started cannot be determined, all samples since the last good
11.2.3 If laboratory 14C contamination or instrumental issue background measurement shall be reprepared and reanalyzed.
has been identified, all data that has been collected since the
issue occurred shall be reprepared and reanalyzed once the 11.4 Wipe Test—A weekly wipe test is to be performed in
any location that may potentially get contaminated with the
issue has been rectified. If an exact date of when the issue
artificial 14C that is used during this analysis. The purpose of a
started cannot be determined, all samples since the last good
wipe test is to have a quick test to determine if 14C spike
QC measurement shall be reprepared and reanalyzed.
solution or any spiked samples contaminated any part of the lab
11.3 Background Sample—For each sample type to be so that cleanup can be done before the contamination is spread
analyzed, a background needs to be measured to determine too widely throughout the lab and/or contaminates any of the
proper background correction factor. samples. The most likely source of 14C contamination during
11.3.1 A background sample is a sample that contains no the use of this method is from a spill, if all method require-
measurable 14C. ments are followed with regards to 14C spike sample materials
and handling.
11.3.2 As samples are analyzed directly, this is very impor-
11.4.1 Each location where the 14C spike solution, 14C spike
tant as each sample type will likely have different types and reagent, or the 14C spiked samples are handled shall be wipe
amounts of quenching species present. For example, different tested for possible contamination. Here is a list of example
gasoline blending components (such as reformate or aklylate) locations: balance, area next to the balance, and hood where
need to have its own background unless experimentally deter- 14
C spiked samples are prepared, the knobs of any doors
mined to be identical. Refer to 10.2 for how to determine if between 14C spike sample preparation area and the instrument,
backgrounds are identical. the instrument, the instruments autosampler tray, instrument
11.3.3 To establish a background for a sample, prepare the computer keyboard, etc. As the 14C spike solution and spiked
same background sample five times using the exact same samples have a much higher concentration of 14C only a small
reagents and conditions that will be used to analyze samples spill is easily detectable by a quick wipe test.
such as sample vials, scintillation cocktail, sample to cocktail 11.4.2 Example wipe test procedure.
ratio, counting time, and region of interest. 11.4.2.1 Use dissolvable wipe test pads to wipe designated
11.3.4 The measured CPM / gC for all five backgrounds is area. Follow the wiping procedure designed specifically for
averaged together and used as the background value for wipes chosen.
samples of similar matrix. 11.4.2.2 After wiping designated area, place wipe into a
20 mL liquid scintillation vial. Make sure the lid is labeled with
11.3.5 A new background shall be measured monthly for all
the location where the wipe was used. Place one wipe directly
products/streams. In addition to the monthly measurements,
into a vial without wiping any surface and label vial as blank.
one background sample shall be run weekly to help identify if
14 11.4.2.3 Add 20 mL of scintillation cocktail designed to
C lab contamination has occurred that was not detected via
dissolve chosen wipes.
the wipe test and to verify that no short-term instrument
11.4.2.4 Wait until wipe has fully dissolved and place it into
fluctuations have occurred that would require the running of all
the liquid scintillation counter.
sample backgrounds. The background sample matrix shall
11.4.2.5 Set the instrument to count only the photons with
remain constant over the life to the sample, a sample with low triple coincidence counts over the full 14C window range, range
volatility and is stable at room temperature will simplify designation may change depending on instrument used, for one
sampling and storage requirements needed to maintain the minute.
constant sample matrix. 11.4.2.6 Control chart the results for each location and the
11.3.6 The CPM / g C value of the new background is blank, as described in Practice D6299. As the runs are so short,
averaged with the last nine values collected for that sample the standard deviation is quite high. If control charts indicate
matrix to maintain a rolling average to account for any minor that any of the wipe locations are trending upwards over the
changes in background over time. If this is near to start of course of several weeks, there is a potential that lab contami-
analysis and nine previous backgrounds have not been taken nation has occurred.
yet, average all backgrounds collected until a total of ten 11.4.2.7 If the measured CPM for all samples is below 100
backgrounds have been run. or 3× of the blank, whichever is lower, it is considered clean.

7
D8473 − 22
If the CPM is above 100 or 3× of the blank, whichever is lower, 13.2.8 Make sure the outside of each sample vial is clean,
the location may be either dirty or contaminated. Clean surface free of anything that could interfere with transmittance of light
of any dirt or dust and repeat the wipe test in that area. Keeping out of the vial such as sample or dust, by wiping the outside
all lab surfaces free of dust and dirt will help reduce the with a lint free cloth or similar before analysis.
frequency of false positive measurements. If second run 13.2.9 Backgrounds will need to be remade and measured
measured CPM is also above 100 or 3× of the blank, there is weekly/monthly to make sure same background levels are
likely 14C contamination that needs to be cleaned before any being measured or to account for any instrumental drift. See
further analysis is performed. 11.3.5 for which frequency each background sample is to be
11.4.2.8 Once contamination has been detected, cleanup run.
verification is required to be conducted by a 3rd party 13.2.10 Run the background samples using the optimized
laboratory. LSC conditions. The following are some of the main param-
11.4.2.9 If laboratory 14C contamination has been identified, eters that need to be optimized before starting routine analysis:
all data that has been collected since the issue occurred shall be analysis time, sample to cocktail ratio, coincidence time,
reprepared and reanalyzed once the contamination has been ionizer delay, and range of interest. Consult instrument manual
cleaned. If an exact date of when the issue started cannot be or instrument manufacturer if there are additional parameters
determined all samples since the last clean wipe test measure- that must be optimized. When optimizing these parameters use
ment shall be reprepared and reanalyzed. the same figures of merit question in Eq 1.
11.4.2.10 A wipe test shall be performed before any artifi- NOTE 12—Initial work used a 5 h analysis time as it was experimentally
cial 14C is handled within the laboratory to determine a determined to give a good balance of data variability while not requiring
baseline level that is to be expected from the wipe tests. a prohibitive amount of time. Shorter analysis time was found to greatly
increase data variability. A coincidence time of 35 ns was used. Ionizer
12. Conditioning delay will be highly dependent on the instrument’s environment and the
personnel prepping the samples, start with default values and increase
12.1 All non-spiked samples, QC samples, and backgrounds them if static electricity interferences is found to be a problem. Range of
shall be allowed to sit in a dark environment after it has been interest of channels 100 to 400 was determined using figures of merit
mixed with the scintillation cocktail for a minimum of 1 h calculations on a diesel sample.
before analysis is started to reduce interference from sample 13.2.11 Record the measured CPM, to the nearest
photoluminescence. If the sample is reintroduced to light, the 0.01 CPM, for each of the backgrounds measured. Calculate
timer shall be reset to 1 h before sample is analyzed. the CPM / gC and average this run with the previous runs of the
sample background sample to use when calculating biocarbon
13. Procedure content in the sample.
13.2.12 For sample types that are run infrequently that do
13.1 Operational Quality Check:
not justify running monthly background, instead of following
13.1.1 A set of unquenched standards is to be run per
13.2.7 and 13.2.9, a single background run can be used but it
manufacturer’s instruction to verify that the instrument is
shall be run immediately before or after the sample and a new
working properly.
background shall be prepared and analyzed each time that
13.2 Background Samples Preparation and Analysis: sample type is run.
13.2.1 Add optimized amount of scintillation cocktail to
13.3 14C Spike Standard Preparation—Precaution is needed
vial. Use figures of merit to determine the optimal ratio of
when preparing and handling the spike solution to prevent
sample to scintillation cocktail. (See Eq 1.)
sample and laboratory contamination. Review 1.5 to make sure
NOTE 10—Initial work used 5 mL of scintillation cocktail. all necessary precautions are being followed.
13.2.2 Set vial onto analytical balance with cap on and tare 13.3.1 Place capsules from the Internal Standard Kit 14C
the vial. into one of the sample vials. (The 14C labeled compound used
13.2.3 Remove vial cap and add optimized amount of to create the spike shall be a non-volatile compound (vapor
background sample to vial, then recap the vial. pressure below 1 Pa at 25 °C), such as cholesterol, to minimize
potential of the compound getting into the gas phase during
NOTE 11—Initial work used 15 mL of sample. sample preparation to reduce risk of lab contamination.)
13.2.4 Record the mass of the sample added to the nearest 13.3.2 Mix with a solvent that is miscible with the samples
0.1 mg and label each vial cap with what sample was placed in to be analyzed. Allow the solvent to dissolve the analyte for
each vial (DO NOT write on the side of the vial as this will 30 min before proceeding. Prepare the sample to have a high
interfere with the instrument detection). enough activity so that only 50 µL to 100 µL need to be spiked
13.2.5 Once sample is capped and properly labeled, place in in the sample for efficiency measurements. Should ideally
a dark place while other samples are being prepared. measure several thousand counts per minute (2000 cpm to
13.2.6 Repeat 13.2.1 – 13.2.5 for each background sample 8000 cpm) when spiked into sample. The use of a solvent with
needing to be measured. low volatility (vapor pressure below 1 Pa at 25 °C), such as the
13.2.7 If this is the very first-time running backgrounds, scintillation cocktail, shall be used to help prevent any of the
repeat 13.2.1 – 13.2.5 four more times to start a rolling artificial 14C molecules from being entrained into the gas phase
background average that will be continued each time back- to reduce risk of lab contamination.
grounds are remade and run. 13.3.3 Store solution in a dry dark location.

8
D8473 − 22
13.3.4 Make a new solution once depleted. detected by all three detectors. Both numbers are used to
13.4 Sample Preparation and Analysis: determine the TDCR of the spiked cocktail which is required
13.4.1 Sample CPM Measurement: for efficiency calculation.
13.4.1.1 It is advised to prepare these samples before those 13.4.2.7 Record the measured CPM and TDCR, both to the
used to determine sample counting efficiency. Doing so allows nearest 0.01, for each sample that was run.
for these samples to sit in the dark while other samples are 13.4.3 Sample Counting Effıciency Measurement Part
prepared and analyzed. 2—Precaution is needed when preparing and handling the spike
solution to prevent sample and laboratory contamination.
13.4.1.2 Add optimized amount of scintillation cocktail to
Review 1.5 to make sure all necessary precautions are being
vial, must be the same amount of scintillation cocktail used for
followed.
the background samples (13.2.1 and Note 10).
13.4.3.1 Once the 14C spiked cocktail from sample effi-
13.4.1.3 Set vial onto analytical balance with cap on and
ciency measurement part 1 has been run, take the sample out of
tare the vial.
the instrument.
13.4.1.4 Remove cap and add optimized amount of sample
13.4.3.2 Place vial with 14C spiked scintillation cocktail
to vial then recap the vial, must be the same amount used for
onto the balance, with lid on, and tare the balance (this shall be
background samples (13.2.3 and Note 11).
a different balance then the one used for non-spiked sample
13.4.1.5 Record the mass of the sample added to the nearest
preparation).
0.1 mg and label each vial cap with what sample was placed in
13.4.3.3 Remove cap and add optimized amount of sample
each vial (DO NOT write on the side of the vial as this will
to the spiked cocktail and recap the vial, this shall be the same
interfere with the instrument detection).
amount that is used for the background and sample CPM
13.4.1.6 Make sure vial cap is on tight and properly labeled,
measurements (13.2.3 and Note 11). Record the mass of the
place in a dark place while other samples are being prepared.
added sample to the nearest 0.1 mg.
13.4.1.7 Repeat 13.4.1.2 – 13.4.1.6 for each sample.
13.4.3.4 Make sure vial cap is on tight and properly labeled.
13.4.1.8 Make sure the outside of all sample vials is clean,
13.4.3.5 Run the spiked sample using the same instrument
free of anything that could interfere with transmittance of light
conditions that are used to run the samples except use the same
out of the vial such as sample or dust, by wiping the outside
time frame that was used to run the spiked cocktail in sample
with a lint free cloth or similar before analysis.
efficiency measurement part 1 (13.4.2.5 and Note 13).
13.4.1.9 Once the samples have been allowed to sit in a dark
13.4.3.6 Record the measured CPM for each sample to the
location for at least 1 h, run the samples using the same LSC
nearest 0.01 CPM.
conditions that were used to measure the background samples.
It is best to do this after doing experiments to determine sample 13.5 Dilution—Determining if dilution is required and how
counting efficiency which are described in the next two to proceed if it is.
sections. 13.5.1 Use the data that was recorded in 13.4.2.7 and
13.4.1.10 Once the samples have been run, record the 13.4.3.6 in Eq 3 (Assume CPMs = CPMb for this check).
measured CPM, to the nearest 0.01 CPM, for each sample. 13.5.1.1 If the calculated value is below 50.0 %, sample
13.4.2 Sample Counting Effıciency Measurement Part dilution is required. Proceed to 13.5.2.
1—Precaution is needed when preparing and handling the spike 13.5.1.2 If calculated value is above 50.0 %, no additional
solution to prevent sample and laboratory contamination. sample preparation is required.
Review 1.5 to make sure all necessary precautions are being 13.5.2 In an unused vial, repeat 13.4.2 and 13.4.3 using
followed. more cocktail and less sample. For example, if standard sample
13.4.2.1 To an empty vial, add 50 µL to 100 µL of the preparation uses 5 mL of cocktail and 15 mL of sample, use
prepared 14C spike standard (13.3). 8 mL of cocktail and 12 mL of sample.
13.4.2.2 Make sure the spike is delivered to the bottom of 13.5.2.1 Dilute the sample as little as possible to achieve an
the vial to help facilitate it mixing completely with the efficiency above 50.0 %.
scintillation cocktail added in the next step. 13.5.2.2 If a sample dilution ratio of 80.0 % cocktail and
13.4.2.3 Add optimized amount of scintillation cocktail to 20.0 % sample or higher is required to achieve a counting
vial, must be the same amount of scintillation cocktail used for efficiency higher than 50.0 % the sample is not compatible with
the background and sample analysis. this method and is required to be analyzed via a different
13.4.2.4 Cap and gently mix the spiked cocktail (must method such as those in Test Method D6866.
ensure that cap is not wetted while mixing), once mixed, label 13.5.3 After the new, diluted sample has been run, repeat
the top of the vial (DO NOT write on the side of the vial as this 13.5.1.
will interfere with the instrument detection). 13.5.3.1 If calculated value is still below 50.0 % more
13.4.2.5 Run the spiked cocktail with the instrument set to sample dilution is required and repeat 13.5.2.
count all beta decays over the entire 14C range for 10 min to 13.5.3.2 If calculated value is above 50.0 % proceed to
30 min. 13.5.4.
13.5.4 Prepare and analyze the sample as described in
NOTE 13—Initial work used 10 min counting time. 13.4.1 but using the same cocktail to sample ratio that was
13.4.2.6 The instrument needs to be set to collect counts that determined to have a counting efficacy above 50.0 % from
are detected by two of the three detectors along with those 13.5.3.

9
D8473 − 22
13.5.5 Prepare and analyze the appropriate background as CPMs = measured CPM of the sample, 13.4.1
described in 13.2 for the sample that required dilution using the (sample CPM measurement),
same sample solvent to cocktail ratio from 13.5.4. Cm = mass of carbon in the sample vial, Eq 2,
13.5.6 Once the efficiency, sample and background have all (CPM / gC)b = calculated CPM / gC of the sample
been measured using the same scintillation cocktail to sample background, 13.2 (background samples
ratio use the measured values to calculate biocarbon content preparation and analysis), and
using equations in Section 14. Diluting a sample will result in Ce = counting efficiency, Eq 3.
a proportionate increase in detection limit and limit of quanti- %Bio 5 DPM cm ⁄ ~ 13.56 3 ~ REF/100!! 3 100 (5)
tation as less sample is being analyzed. For example, if
normally 15 mL of sample is used and has a detection limit of where:
2 % by mass biocarbon, if the dilution requires the use of %Bio = percent biocarbon by mass expressed in units
10 mL of sample instead of the normal 15 mL the detection of grams of modern carbon in sample relative
limit for the diluted sample is 3 % by mass biocarbon. to total grams of carbon in the sample,
DPMcm = DPM of sample per gram of carbon in sample,
14. Calculations Eq 4,
13.56 = defined DPM per gram of carbon for a sample
14.1 The counts shall be compared, directly to the currently with 100 % pMC, and
defined level of 100 % by mass modern carbon. Refer to Table REF / 100 = correction factor for when the plant or animal
1 for REF correction factor for the last several years. The was alive.
inherent assumption is that all the organic components within
14.3.1 This equation only applies for when the year(s) the
the analyzed material are either fossil or present day in origin.
plant or animal was known to be living for a given biofeed and
14.2 The relative number of counts between the modern the products produced from that feed. If the plant or animal
reference and the sample is term “fraction of modern” (fM). was alive for multiple years, the carbon uptake for each year
This is most commonly referred to as percent modern carbon also needs to be known and averaged. If these are unknown, the
(pMC), the percent equivalent to fM (for example, fM = feed shall be analyzed to correct for the amount of 14C present
pMC/100). within. This will likely require the use of Test Method
D6866 Method B as most biofeeds are unable to be analyzed
14.3 Use the following equations to convert the experimen-
via this method. The exception to this is if the plant or animal
tally obtained values to percent by mass biocarbon.
was planted or born during or after 2019 as atmosphere 14C
C m 5 S m 3 %C m⁄m (2) concentrations have been constant since then.
where: 14.3.1.1 Correction for use of biofeeds that accumulated
bomb carbon over the course of more than one year (more
Cm = mass of carbon in the sample vial,
Sm = mass of sample in the sample vial, and detailed explanation in Test Method D6866).
%Cm/m = percent carbon of the sample, to be determined (1) If feed pMC = 100 – 105, use Eq 5 as written.
experimentally using methods such as CHN el- (2) If feed pMC = 106 – 122, use REF = 112.
emental analysis by D5291. (3) If feed pMC = 122 or larger, use REF = 138.
14.4 Correcting for Radon-222 Contamination:
C e 5 ~ CPM ss 2 CPM s ! ⁄ ~ CPM SC ⁄ TDCR ! (3)
14.4.1 Crude oil, especially shale oil, has the potential to
where: contain small amounts of U-238. One of the decay products is
Ce = counting efficiency, Rn-222. Rn-222 has the potential to make it through the
CPMss = measured CPM of the spiked sample, 13.4.3 refining process and be retained in the finished products. The
(sample efficiency measurement part 2), decay of Rn-222 and its subsequent products can produce
CPMs = measured CPM of the sample, 13.4.1 (sample interference in the 14C region of the LSC spectra. The half-life
CPM measurement), of Rn-222 is short, 3.82 days. The short half-life of Rn-222
CPMSC = measured CPM of spiked cocktail, 13.4.2 (sample allows for calculation to be performed to correct for its
efficiency measurement part 1), and presence. Eq 6 and Eq 7 show how to perform this correction.
TDCR = triple to double coincidence ratio, 13.4.2 (sample 14.4.2 This correction only needs to be applied to samples
efficiency measurement part 1). with Rn-222 contamination. The LSC full spectrum allows for
DPM Cm 5 ~ CPM s ⁄ C m 2 ~ CPM ⁄ g C ! b ! ⁄C e (4)
easy determination if the contamination is present. Fig. 1
shows LSC spectra with and without Rn-222 contamination.
where: CPM cor 5 Observed CPM C14 2 k ~ observed CPM 60021000
DPMCm = DPM of the sample,
2 CPM bg60021000! (6)

10
D8473 − 22

14
The orange line is the upper range of signal produced by C.

14 14
FIG. 1 LSC Spectra of a Sample with C and Rn-222 Contamination (top) and with Only C (bottom)

where: Observed CPM600 – 1000 = CPM measured over the counting


CPMcor = corrected CPM signal that can be window 600 – 1000 (If the win-
then input into equations in 14.3 dow chosen for CPMs measure-
for CPMs, ments extends above 600, this
Observed CPMC14 = CPM s for the contaminated window needs to be adjusted so it
sample over the specified region does not overlap with CPMs
of interest, window), and
k = sample type dependent correc-
tion constant (Eq 7),

11
D8473 − 22

CPMbg600 – 1000 = CPM measured for the contami- 16. Precision and Bias
nated samples background over 16.1 The repeatability and reproducibility will ultimately be
the counting window 600 – 1000 determined by an interlaboratory study in accordance with
(this data is may already be ASTM methodology. This section reports interim precision
stored in the raw data collected derived from a mini-interlaboratory study (ILS) participated by
when the background samples three laboratories.
were run, if not, new back-
16.2 Precision—Interim repeatability was calculated over
grounds need to be taken to mea-
the biocarbon ranges 0 % to 14 % by mass. The precision was
sure over the 600 – 1000 channel
developed using 14, 12, and 12 fuel samples each of gasoline,
range so this equation can be
diesel, and jet standard, respectively, by either mixing renew-
used).
able naphtha with the refinery produced gasoline or renewable
k 5 ~ CPM s t1 2 CPM s t2 ! ⁄ ~ CPM 60021000 t1 2 CPM 60021000 t2 ! diesel with the refinery produced diesel and jet fuel samples.
(7) Details of these samples can be found in Research Report
RR:D02-2038.7
where: 16.2.1 Repeatability—The difference between two results
k = sample type dependent correction con- obtained by the same operator on the same apparatus under
stant (needed for Eq 6) (k is quench constant operating conditions on identical test material would,
dependent and needs to be calculated for in the long run, in the normal and correct operation of the test
each sample type that contamination is method, exceed the following values (see Tables 2 and 3) only
observed, that is, if the contamination is in one case in twenty (95 % confidence interval).
observed in gasoline and jet fuel k would 16.2.2 Reproducibility—The breadth of the initial ILS was
need to be calculated for both sample limited and not adequate for establishing reproducibility, com-
types), pliant with Practice D6300. A full ILS will be conducted within
CPMs t1 = CPMs initially measured, five years of the method approval and publication.
CPMs t2 = CPMs measured again for exact same 16.2.3 Bias—No information can be presented on the bias of
sample (do not prepare a new one) after 3 the procedure for measuring percent biocarbon by mass be-
to 6 days (Half-life of Rn-222 is 3.82 cause no material having an accepted reference value is
days, rerunning the sample after 3-6 days available at this time.
helps determine the signal present in the
sample region that is from RN-222), 16.3 Relative Bias—A limited study to assess the relative
CPM600 – 1000 t1 = CPM measured over the counting win- bias of this method versus Test Method D6866 – 20 using AMS
dow 600 – 1000 from the same run as (accelerated mass spectrometry) for the determination of per-
CPMs t1, and cent biocarbon by mass was conducted involving one labora-
CPM600 – 1000 t2 = CPM measured over the counting win- tory running the AMS method. The lack of participating AMS
dow 600 – 1000 from the same run as labs prevents a reliable assessment of the relative bias in
CPMs t2. accordance with Practice D6708. The AMS data collected are
documented in Research Report RR:D02-2038.
15. Report
17. Keywords
15.1 Final reports shall include the following:
15.1.1 Name of testing laboratory, 17.1 biobased; bomb carbon; 14C (carbon-14); carbon dat-
15.1.2 Date of testing, ing; liquid scintillation counting; new carbon; old carbon;
15.1.3 Standard name and version, percent modern carbon
15.1.4 REF value used, and
15.1.5 Percent by mass biobased carbon content (reported to
0.1 %, raw data shall be kept containing results to the 0.01 % 7
Supporting data have been filed at ASTM International Headquarters and may
until full ILS is completed). be obtained by requesting Research Report RR:D02-2038. Contact ASTM Customer
Service at service@astm.org.
15.2 If regulatory reporting guidelines require the reporting
of results in units other than grams of modern carbon in sample TABLE 2 Interim Repeatability
per total grams of carbon in the sample (percent by mass Repeatability, r
biobased carbon content), then the conversion of the results Gasoline, jet, and diesel fuels 0.02723 (X + 22.6)A
may be required. The required conversion will have to be A
X = percentage of biocarbon, between 0 % and 13.2 %.
assessed on a case-by-case basis.

12
D8473 − 22
TABLE 3 Repeatability Calculation for Selected Biocarbon Concentrations
Biocarbon, % 0 2 4 6 8 10 12 14
Repeatability, 0.62 0.67 0.70 0.78 0.83 0.89 0.94 1.00
%

APPENDIX

(Nonmandatory Information)

X1. FIGURES

FIG. X1.1 Recommended Sample Analysis Workflow

13
D8473 − 22

FIG. X1.2 List of Routine Analyses that Need to be Conducted, Not Including Normal Sample Analysis

REFERENCES

(1) Anderson, R., and Cook, G. T., “Scintillation Cocktail Optimization Scintillation Counting,” Liquid Scintillation Counting and Organic
for 14C Dating Using the Packard 2000 CA/LL and 2260 SL,” Scintillators, Ross, H., Noakes, J. E., and Spaulding, J. D., Eds.,
Radiocarbon, Vol 33, 1991, pp. 1–7. Lewis Publishers, Chelsea, MI, 1991, pp. 501–511.
(2) Cook, G. T., Naysmith, P., Anderson, R., and Harkness, D. D., (8) Takiue, M., Natake, T., and Fujii, H., “Liquid Scintillation Radioassay
“Performance Optimization of the Packard 2000 CA/LL Liquid for Low-activity Beta Emitter Mixtures by the Method of Least
Scintillation Counter for 14C Dating,” Nucl. Geophys., Vol 4, 1990a, Squares,” J. Radioanal. Nucl. Chem. Lett., Vol 200, 1995, pp.
pp. 241–245. 247–258.
(3) Koba, M., “Improved Results Using Higher Ratios of Scintillator (9) Thomson, J., and Burns, D. A., “LSC Sample Preparation by
Solution to Benzene in Liquid Scintillation Spectrometry,” Solubilization,” Counting Solutions CS-003, Packard Instrument
Radiocarbon, Vol 42, No. 2, 2000, pp. 295–303. Company, Meriden, CT, 1996b.
(4) L’Annuniziata, M. F., “Effıciency Tracing DPM (ET-DPM) and (10) Currie, L. A., Stafford, T.W., Sheffield, A. E.,Wise, S. A., Fletcher, R.
Direct-DPM-Instrument Performance Data. Counter Intel. Tri-Carb A., Donahue, D. J., Jull, T. J. T., and Linick, T. W., “Microchemical
LSC Application,” Packard BioScience Company, Meriden, CT, 1997. and Molecular Dating,” Radiocarbon, Vol 31, No. 3, 1989, pp.
(5) L’Annuniziata, M. F., Handbook of Radioactivity Analysis, Academic 448–463.
Press, New York, 1998. (11) Currie, Lloyd A., Klinedinst, Donna B., Burch, R., Feltham, N., and
(6) Noakes, J. E., “Low-Level Liquid Scintillation Counter Array with Dorsch, R., “Authentication and Dating of Biomass Components of
Computerized Data Acquisition and Age Calculation Capabilities for Industrial Materials; Links to Sustainable Technology,” Nuclear
14C Dating,” Radiocarbon, Vol 37, No. 2, 1995, pp. 773–779. Instruments and Methods in Physics Research, Section B, Vol 172,
(7) Roessler, N., Valenta, R. J., and van Cauter, S., “Time-resolved Liquid 2000, pp. 281-287.

14
D8473 − 22
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

15

You might also like