You are on page 1of 1

1 2 3 4 5 6 7 8

H Plusteam Sterilizer : min. 5 meter


H
Drain Pipe Line Installation Sheet

Important Note: Drain line should be connected


to main drain pipe with fewest elbow connections
possible in order to reduce the flow resistance. 2" / DN 50

Increased resistance in flow reduces vacuum


135
efficiency hence affects device performance. °
G 2" / DN 50
G

45°
2" / DN 50 2" / DN 50 2" / DN 50

F F

min. 5 meter Drain flow direction.

1" / DN 25

135
° 1" / DN 25

E E

Two drain line did not collide. Flow is not Two lines collide directly. The water flow is
effected. No backflow or efficiency lose in reduced causing squeezing in junction.

45°
vacuum pump. The backflow happens.
2" / DN 50 2" / DN 50 2" / DN 50
Vacuum pump cannot operate efficiently.

D D

1 " Male / Male


Waste Hose Flexible

C C

Drain connection

Drain collector 1" Male / Male


Flexible waste hose
(140° C heat resistant)

±0.00 1" Muff

B B

Drain Line Drain Line

Temperatures : LOCATION : Ankara / TÜRKİYE PROJECT : Drain Pipe Line


General notes about infrastructure from C.S.S.D Steam Sterilizer ; Air outflow from service room: heat flow to be dissipated, temperature in
service room approx. T = 30 °C, ∆T=10 °C. Average Peak : 2.0 kW, 500 mn3/h
Floor Covering : Washer Disinfector ; Heat radiation 600 - 700 Kcal/h - W PROFORMA INVOICE : 000000 DATE : 23.02.2023
Antimicrobial flooring / Anti-static, with rounded corners.
Wall Covering : The copyright on this drawing and all enclosures is reserved to Turkuaz Biyomedikal Teknolojiler at any time. This documentation is personal and
CHECK BY : C. ALACAPINAR REVISION : 3.4
A Antimicrobial PVC hygienic wall cladding / Anti-bacterial paint
entrusted to the consignee. Do not copy this drawing or parts thereof without prior written approval by Turkuaz Biyomedikal Teknolojiler. Also it is not
allowed to make it available to or to hand it over to third parties.
TBT has various designs and several technical variants. There may be differences between the same product types and various versions. This
Tel
Fax
: (+90 850 ) 840 88 28
: (+90 312 ) 911 25 28
A
Ceiling Covering: document is intended for the use of Distributors/Dealers/Customers of TBT Inc. Reproduction in whole or in part by any other party is prohibited.
Specifications are given for reference only and are subject to change without any previous notice. Accessory/Product images are for illustrative
FORMAT : A3 DESIGNED BY : GT. MOL E-Mail : info@tbtmedical.com
Aluminum suspended ceiling grid purposes only. rev.EN 010-2023 © 2015, TBT Inc. All rights reserved. Web : www.tbtmedical.com

1 2 3 4 5 6 7 8