Professional Documents
Culture Documents
Sieving Methods:
Guidelines for
Establishing Sieve
Analysis Procedures
5th Edition
ASTM International
100 Barr Harbor Drive
PO Box C700
West Conshohocken, PA 19428-2959
Printed in U.S.A.
Copyright © 2014 ASTM International, West Conshohocken, PA. All rights reserved.
This material may not be reproduced or copied, in whole or in part, in any printed,
mechanical, electronic, film, or other distribution and storage media, without the
written consent of the publisher.
ASTM International is not responsible, as a body, for the statements and opinions
advanced in the publication. ASTM International does not endorse any products
represented in this publication.
Printed in
Baltimore, MD
January, 2014
Foreword
This manual on test sieving methods is intended for use as a supplement
to, and not as a substitute for, the many ASTM standards relating to the
sieve analysis of materials. There has been a need for a manual that would
bring together, from many sources, proven methods for making reliable
sieve analysis to serve as a guide for the novice and a reference for the more
advanced.
Test Sieving Methods was originally compiled by W. C. Mahlig and
A. E. Reed. The latest revisions to this manual were made by a working
subcommittee, ASTM E29.01. Please contact the sponsoring committee,
E29 on Particle and Spray Characterization, for any additional information.
This is MNL 32 in ASTM International’s manual series.
Contents
1—Types of Sieves �������������������������������������������������������������������������������������������������1
1.1 ASTM Standard Test Sieves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 International Standards Organization (ISO) Test Sieves . . . . . . . . . . . . 5
1.3 ASTM and ISO Test Sieves with Enhanced Accuracy . . . . . . . . . . . . . . 10
1.4 ASTM Perforated-Plate Sieves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
1.5 ASTM Precision Electroformed Sieves . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2—Sieving Methods���������������������������������������������������������������������������������������������17
2.1 Hand Sieving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.2 Standard Mechanical Sieve Shakers . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.3 Vibratory Mechanical Sieve Shakers . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.4 Miscellaneous Mechanical Sieve Shakers . . . . . . . . . . . . . . . . . . . . . . . 19
3—Test Sieving Procedures ���������������������������������������������������������������������������������20
3.1 Sample Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.2 Sample Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3 The Sieve Stack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.4 Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.5 Weighing and Calculations���������������������������������������������������������������������22
4—Sieving Practices���������������������������������������������������������������������������������������������26
4.1 Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.2 Wet Sieving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.3 Sieve Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.4 Sieving Aids and Flow Agents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.5 Setting Up a Quality Control Program . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5—Appendix���������������������������������������������������������������������������������������������������������34
Table A-1: Suggested Bulk Volume of Test Sample for Sieve Analysis
with 8-in. and 200-mm Round Sieves�����������������������������������������34
Table A-2: Typical Bulk Densities of Various Particulate Materials . . . . . . 35
Table A-3: List of ASTM Published Standards on Sieve Analysis
Procedures for Specific Materials or Industries�������������������������39
Table A-4: List of ASTM Published Standards on Sampling
of Particulate Materials���������������������������������������������������������������56
6—Nomenclature �������������������������������������������������������������������������������������������������62
7—References�������������������������������������������������������������������������������������������������������66
Sieve Designation +X Compliance Sieves Inspection Sieves Calibration Sieves Permissible Range of Choice
Nominal ±Y Maximum Resulting
Sieve Variation Variation Maximum Sample Maximum Sample Maximum Sample Maximum Typical
Opening for Average for Individual Openings Standard Openings Standard Openings Standard Wire
Standard Alternative (in.) Opening Opening Opening per 100 ft2 Deviation per Sieve Deviation per Sieve Deviation Diameter Min Max
BK-AST-MNL32-131022-130001.indb 2
millimetre inches millimetre millimetre millimetre millimetre
106 4.24 in. 4.24 3.12 3.99 109.99 20 – all – all – 6.3 5.4 7.2
100 4 in. 4 2.94 3.82 103.82 20 – all – all – 6.3 5.4 7.2
90 3½ in. 3.5 2.65 3.53 93.53 20 – all – all – 6.3 5.4 7.2
63 2½ in. 2.5 1.87 2.71 65.71 20 – all – all – 5.6 4.8 6.4
45 1¾ in. 1.75 1.35 2.12 47.12 20 . all – all – 4.5 3.8 5.2
37.5 1½ in. 1.5 1.13 1.85 39.35 20 1.374 all – all – 4.5 3.8 5.2
31.5 1¼ in. 1.25 0.95 1.63 33.13 20 1.066 all – all – 4 3.4 4.6
26.5 1.06 in. 1.06 0.802 1.44 27.94 20 0.869 15 0.584 all – 3.55 3 4.1
19 ¾ in. 0.750 0.579 1.13 20.13 150 0.622 15 0.418 30 0.446 3.15 2.7 3.5
5
16 /8 in. 0.625 0.490 0.99 16.99 150 0.527 15 0.354 30 0.378 3.15 2.7 3.6
13.2 0.530 in. 0.530 0.406 0.86 14.06 150 0.441 15 0.296 30 0.316 2.8 2.4 3.2
(Continued)
20/12/13 11:44 AM
Table 1—ASTM E11: Nominal Dimensions and Permissible Variations for Sieve Cloth and Compliance, Inspection, and Calibration Test Sieves
(Continued)
(1) (2) (3) (4) (5) (6) (7) (8) (9) (10) (11) (12) (13) (14) (15)
Sieve Designation +X Compliance Sieves Inspection Sieves Calibration Sieves Permissible Range of Choice
Nominal ±Y Maximum Resulting
Sieve Variation Variation Maximum Sample Maximum Sample Maximum Sample Maximum Typical
Opening for Average for Individual Openings Standard Openings Standard Openings Standard Wire
BK-AST-MNL32-131022-130001.indb 3
Standard Alternative (in.) Opening Opening Opening per 100 ft2 Deviation per Sieve Deviation per Sieve Deviation Diameter Min Max
12.5 ½ in. 0.500 0.385 0.83 13.33 150 0.421 15 0.283 30 0.302 2.5 2.1 2.9
7
11.2 /16 in. 0.438 0.346 0.77 11.97 150 0.382 15 0.256 30 0.274 2.5 2.1 2.9
3
9.5 /8 in. 0.375 0.295 0.68 10.18 150 0.330 15 0.222 30 0.237 2.24 1.9 2.6
5
8 /16 in. 0.312 0.249 0.60 8.60 150 0.284 15 0.191 30 0.204 2 1.7 2.3
Manual on Test Sieving Methods
6.7 0.265 in. 0.265 0.210 0.53 7.23 150 0.245 15 0.164 30 0.175 1.8 1.5 2.1
6.3 ¼ in. 0.250 0.197 0.51 6.81 150 0.233 15 0.157 30 0.167 1.8 1.5 2.1
5.6 No. 31/2 0.223 0.176 0.47 6.07 150 0.211 15 0.142 30 0.151 1.6 1.3 1.9
4.75 No. 4 0.187 0.150 0.41 5.16 150 0.182 15 0.123 30 0.131 1.6 1.3 1.9
4 No. 5 0.157 0.127 0.37 4.37 150 0.161 15 0.108 30 0.115 1.4 1.2 1.7
3.35 No. 6 0.132 0.107 0.32 3.67 200 0.138 20 0.097 40 0.103 1.25 1.06 1.5
2.8 No. 7 0.110 0.090 0.29 3.09 200 0.121 20 0.085 40 0.090 1.12 0.95 1.3
2.36 No. 8 0.0937 0.076 0.25 2.61 200 0.104 20 0.073 40 0.077 1 0.85 1.15
2 No. 10 0.0787 0.065 0.23 2.23 250 0.094 25 0.068 50 0.072 0.9 0.77 1.04
1.7 No. 12 0.0661 0.056 0.20 1.90 250 0.081 25 0.059 50 0.062 0.8 0.68 0.92
1.4 No. 14 0.0555 0.046 0.18 1.58 400 0.071 40 0.055 80 0.057 0.71 0.6 0.82
1.18 No. 16 0.0469 0.040 0.16 1.34 400 0.063 40 0.049 80 0.051 0.63 0.54 0.72
1 No. 18 0.0394 0.034 0.14 1.14 400 0.055 40 0.042 80 0.044 0.56 0.48 0.64
3
(Continued)
20/12/13 11:44 AM
4
Table 1—ASTM E11: Nominal Dimensions and Permissible Variations for Sieve Cloth and Compliance, Inspection, and Calibration Test Sieves
(Continued)
(1) (2) (3) (4) (5) (6) (7) (8) (9) (10) (11) (12) (13) (14) (15)
850 No. 20 0.0331 29.1 127 977 400 48.76 40 37.73 80 39.36 0.5 0.43 0.58
710 No. 25 0.0278 24.7 112 822 500 42.63 50 33.82 100 35.14 0.45 0.38 0.52
BK-AST-MNL32-131022-130001.indb 4
600 No. 30 0.0234 21.2 101 701 500 38.00 50 30.14 100 31.32 0.4 0.34 0.46
500 No. 35 0.0197 18.0 89 589 600 33.23 60 26.85 120 27.82 0.315 0.27 0.36
425 No. 40 0.0165 15.5 81 506 600 29.95 60 24.20 120 25.08 0.28 0.24 0.32
355 No. 45 0.0139 13.3 72 427 800 26.45 80 21.95 160 22.64 0.224 0.19 0.26
300 No. 50 0.0117 11.5 65 365 800 23.70 80 19.66 160 20.29 0.2 0.17 0.23
250 No. 60 0.0098 9.9 58 308 800 21.02 80 17.44 160 17.99 0.16 0.13 0.19
212 No. 70 0.0083 8.7 52 264 800 18.79 80 15.59 160 16.08 0.14 0.12 0.17
180 No. 80 0.0070 7.6 47 227 1000 16.85 100 14.24 200 14.65 0.125 0.106 0.15
150 No. 100 0.0059 6.6 43 193 1000 15.30 100 12.93 200 13.30 0.1 0.085 0.115
125 No. 120 0.0049 5.8 38 163 1000 13.51 100 11.41 200 11.74 0.09 0.077 0.104
Manual on Test Sieving Methods
106 No. 140 0.0041 5.2 35 141 1000 12.39 100 10.47 200 10.77 0.071 0.06 0.082
90 No. 170 0.0035 4.6 32 122 1000 11.27 100 9.53 200 9.80 0.063 0.054 0.072
75 No. 200 0.0029 4.1 29 104 1000 10.23 100 8.64 250 9.02 0.05 0.043 0.058
63 No. 230 0.0025 3.7 26 89 1000 9.18 100 7.76 250 8.09 0.045 0.038 0.052
53 No. 270 0.0021 3.4 24 77 1000 8.44 100 7.13 250 7.44 0.036 0.031 0.041
45 No. 325 0.0017 3.1 22 67 1000 7.76 100 6.56 250 6.84 0.032 0.027 0.037
38 No. 400 0.0015 2.9 20 58 1000 7.09 100 5.99 300 6.31 0.03 0.024 0.035
32 No. 450 0.0012 2.7 18 50 1000 6.42 100 5.42 300 5.71 0.028 0.023 0.033
25 No. 500 0.0010 2.5 16 41 1000 5.71 100 4.82 300 5.06 0.025 0.021 0.029
20 No. 635 0.0008 2.3 15 35 1000 5.33 100 4.51 300 4.75 0.02 0.017 0.023
Column 3—These numbers are only approximate but are in use for reference; the sieve shall be identified by the standard designation in millimetres or micrometres.
Columns 9 and 11—See Annex A1, which specifies that all openings will be inspected for test sieves having 15 openings or less.
Columns 9 and 11—These number of sample openings are based on an 8-in. diameter test sieve.
20/12/13 11:44 AM
Manual on Test Sieving Methods 5
1 5/8 (41) HH
A
Frame height measured from top of frame to top of sieve cloth.
B
Measured 0.2 in. (5 mm) below the top of the frame.
C
FH = full height; HH = half height; IH = intermediate height.
For
For any Average Maximum Permissible
Principal Supplementary Aperture Aperture Standard Preferred Range of
Sizes Sizes Size Size Deviation Sizes Choice
75 3.09 2.22 b
6.3 7.2 5.4
(Continued)
For
For any Average Maximum Permissible
Principal Supplementary Aperture Aperture Standard Preferred Range of
Sizes Sizes Size Size Deviation Sizes Choice
(Continued)
For
For any Average Maximum Permissible
Principal Supplementary Aperture Aperture Standard Preferred Range of
Sizes Sizes Size Size Deviation Sizes Choice
For
For any Average Maximum Permissible
Principal Supplementary Aperture Aperture Standard Preferred Range of
Sizes Sizes Size Size Deviation Sizes Choice
90 90 90 32 4.6 12 63 72 54
80 30 4.3 11.3 56 64 48
75 29 4.1 10.9 50 58 43
71 28 4 10.5 50 58 43
63 63 63 26 3.7 9.9 45 52 38
(Continued)
For
For any Average Maximum Permissible
Principal Supplementary Aperture Aperture Standard Preferred Range of
Sizes Sizes Size Size Deviation Sizes Choice
56 25 3.5 9.3 40 46 34
53 24 3.4 9 36 41 31
50 23 3.3 8.7 36 41 31
45 45 45 22 3.1 8.3 32 37 27
40 21 3 7.9 32 37 27
38 20 2.9 7.7 30 35 24
32 19 2.7 6.8 28 33 23
25 16 2.5 6.1 25 29 21
20 14 2.3 5.7 20 23 17
NOTE All aperture sizes apply for plain weave. Aperture sizes of 45 μm and smaller apply also for
twilled weave. It should be noted, however, that plain and twilled weave sieves can have different
sieving characteristics.
a
In accordance with ISO 565:1990, Table 2.
Once inspection or calibration grade sieves are in use, they should periodically
be re-certified or re-verified to demonstrate continued compliance. This allows
sieves to remain in service longer with documented compliance to ASTM or ISO
requirements.
“Matched” sieves are an alternative to certified or verified sieves. Matched
sieves are selected through actual performance trials by comparing them to a “mas-
ter” set of sieves in a reference laboratory. Matched sieve products generally allow
test result variations no greater than 2 %. When “matching” sieves, it is advised that
after the matching process they may not meet ASTM or ISO measurement specifica-
tions. An inspection or calibration grade certification is recommended for compli-
ance with either ASTM or ISO.
(457 mm) square sieve plates. For openings smaller than 4 mm, 8-in. (203 mm)
round frames may also be used (Fig. 6).
In general, round-hole sieves are used only when the product specification
is based on round-hole openings. If perforated sieves and wire cloth sieves are
used in the same test, it is recommended that only square-opening sieves be used.
Results from square-opening and round-opening perforated-plate sieves are not
compatible.
BK-AST-MNL32-131022-130001.indb 13
(1) (2) (3) (4) (5) (6) (7) (8) (9) (10)
125 5 ±1.0 160 6¼ 144 to 184 3.4 0.1345 10 2.5 to 4.0
106 4¼ ±0.9 135 5¼ 122 to 155 3.4 0.1345 10 2.5 to 4.0
100G 4 ±0.9 128 5 115 to 147 3.4 0.1345 10 2.5 to 4.0
90 3½ ±0.8 111 43/8 100 to 128 2.7 0.1046 12 2.5 to 4.0
75 3 ±0.7 95 3¾ 86 to 109 2.7 0.1046 12 2.5 to 4.0
63 2½ ±0.6 80 31/8 72 to 92 2.7 0.1046 12 2.5 to 4.0
Manual on Test Sieving Methods
5 7
8.0 /16 ±0.19 11.0 /16 9.5 to 12.6 1.9 0.0747 14 1.0 to 2.0
(Continued)
20/12/13 11:44 AM
14
Table 5—ASTM E323: Nominal Dimensions, Permissible Variations, and Ranges for Perforated Plate in
Standard Sieves (Continued)
Sieve Designation and Maximum Centers Plate Thickness
Aperture SizeA Variation Preferred Permissible Preferred Permissible
of Aperture RangeC Range
StandardD Alternative Standard Alternative Standard Alternative
SizeB
mm in.E mm mm in.E mm mm inE gageF mm
BK-AST-MNL32-131022-130001.indb 14
(1) (2) (3) (4) (5) (6) (7) (8) (9) (10)
17 25
6.7 /64 ±0.17 9.9 /64 8.3 to 11.4 1.5 0.0598 16 0.8 to 1.5
3
6.3G ¼ ±0.16 9.5 /8 8.0 to 10.9 1.5 0.0598 16 0.8 to 1.5
7 11
5.6 /32 ±0.15 8.7 /32 7.2 to 10.0 1.5 0.0598 16 0.8 to 1.5
3
4.75 /16 ±0.14 6.8 ¼ 5.8 to 7.8 1.5 0.0598 16 0.8 to 1.5
5 7
4.00 /32 ±0.13 5.9 /32 5.0 to 6.8 1.5 0.0598 16 0.8 to 1.5
3
3.35 0.127(1/8) ±0.12 4.9 /16 4.2 to 5.7 1.5 0.0598 16 0.8 to 1.5
7 11
2.80 /64 ±0.11 4.4 /64 3.7 to 5.1 1.5 0.0598 16 0.8 to 1.5
3 5
2.36 /62 ±0.10 3.8 /32 3.2 to 4.4 1.5 0.0598 16 0.8 to 1.5
1
2.00 0.078 ±0.09 3.3 /8 2.8 to 3.8 1.5 0.0598 16 0.8 to 1.5
Manual on Test Sieving Methods
7
1.70 0.066 ±0.08 2.9 /64 2.5 to 3.3 0.8 0.0299 22 0.4 to 0.8
1.40 0.055 ±0.08 2.6 0.100 2.2 to 3.0 0.8 0.0299 22 0.4 to 0.8
1.18 0.045 ±0.07 2.2 0.090 1.9 to 2.5 0.8 0.0299 22 0.4 to 0.8
1.00 0.039 ±0.07 2.0 0.077 1.7 to 2.3 0.8 0.0299 22 0.4 to 0.8
A
The values shown in this table refer to both round and square apertures. In general, square-aperture perforated-sieve plates are available only in 3.35 mm and larger.
B
This permissible variation applies to both the standard aperture sizes and the alternative aperture sizes.
C
A range of ±15 % is allowed except that in no case shall the minimum bridge (bar between apertures) be less than one half of the difference between the nominal
aperture and the preferred center.
D
These standard designations, progressing from a base of 1 mm in the ratio of approximately 4√2 to 1, correspond to the values for test sieve apertures recommended by
the International Organization for Standardization, Geneva, Switzerland.
E
Only approximately equivalent to the standard values.
F
The gage values are for carbon steel. For other materials, the gage used should be the nearest decimal equivalent of the U.S. standard gage for steel.
G
These sieves are not in the standard series but they have been included because they are in common usage.
20/12/13 11:44 AM
Manual on Test Sieving Methods 15
25 2.0 118.11 C
196.85 C
300 C
500C
sieved and coarser than 75 μm. These shakers generally agitate in an orbital, circular,
or side-to-side motion, allowing the particles to roll and retry different sieve apertures.
Wet sieving is often required for materials that are difficult to separate in a dry
state. Some sieve shakers are designed specifically for wet sieving. There are also
add-on accessories to modify sieve shakers for wet sieving operations. The applica-
tion and collection of wash water are important considerations when selecting a wet
sieving system. The system must thoroughly wet the sample while permitting agitation
and allow the wash water to be collected or processed in a way that ensures minimal
sample loss.
The volume of the test sample should be such that there is no crowding of
oversize and near-size mesh particles on any sieve surface. At the same time, the
sample must be large enough to permit a measurable amount of material to be
retained on each sieve. A rule of thumb is that there should be no more than one
or two layers of material evenly distributed across the mesh surfaces at completion
of the test. This ensures that all particles have an equal chance to be tried against
available apertures. Overloading is most likely to occur when testing materials with
a concentration of particles close to one size or when the entire sample is within a
narrow size range. A smaller, properly sampled specimen will yield more accurate
and consistent results than a larger sample that may overload one or more sieves.
An exception occurs when testing very coarse materials, in which larger samples are
required to constitute a representative portion.
A useful table of recommended sample sizes for testing with 8-in. or 200-mm
sieves is shown in the Appendix (Table A-1). Sample sizes are shown by volume. To
determine the suggested weight in grams, multiply values in columns 3 and 4 by
the bulk density of the material in grams per cubic centimetre, also shown in the
Appendix. If the particular material density is unknown, then select values for the
most similar material listed in the bulk density table. The values are a useful start-
ing point when standard test sample sizes have not been established. It is suggested
that these sample sizes be verified by the procedure outlined above before adopting
them for use.
shaker. Hand sieving of a series of sieves is also possible, but there is considerable
loss of efficiency, accuracy, and repeatability.
3.4 Testing
Most tests performed on common materials using conventional sieve shakers are
complete in 3–5 min. Materials that are difficult to separate may take 10–30 min in
some cases. Hand sieving requires shaking one sieve at a time for longer periods to
ensure acceptable accuracy. Prolonged sieving times may degrade some sensitive or
friable materials and affect test results.
To systematically determine an adequate sieving time, use the methods noted
above to create several accurate samples of the test material. Test each sample for a
different time interval, such as 3, 5, 10, 15, and 20 min, and calculate the percentage
by weight retained or passing on each sieve. Select the optimal test period where
percentages have stabilized. This same methodology can be used to fine-tune vibra-
tion, frequency, and other variable controls. For most tests, the endpoint has been
reached when an additional 1 min of sieving changes the weight on any of the sieves
by 1.0 % or less.
it represents the total percentage of the test sample coarser than the opening of
that particular sieve.
However, it is also acceptable to set up the specifications and report test results
on the basis of the percentage passing on each sieve. Figure 15 is a typical labora-
tory report form for recording the results of a sieve test.
Sieve analyses are often presented graphically for comparison with specifica-
tion requirements or for evaluation. Percentages retained or passing sieves not
actually used in the test can be estimated by interpolation. Likewise, the size of
an opening that would theoretically retain or pass a selected percentage can be
estimated.
The sieve analysis graph usually represents the sieve size and the percentage
retained or passing. The scale for sieve sizes may be linear or logarithmic. Because
standard sieve sizes relate to one another by powers of the fourth root of 2, loga-
rithmic scales have the advantage of representing sizes on an equally spaced scale.
For example, the distances between the #4 and #8, the #8 and #16, and the 3.4 and
3.8 in. are all the same on a logarithmic scale because openings of the larger sieves
are twice that of the smaller. The scale for percentages is usually linear, depicting
equal differences in percentage as the same distance.
Examples of the two principal types of graphs used for sieve analysis work are
shown in Figs. 15–18.
s uitable sample splitter, a user must choose a sample splitter or reducer with chute
width at least three times the size of the largest particle. Some sample splitters
allow the user to adjust the chute width for use with multiple particle sizes.
The highest precision in sample reduction and splitting is obtained through the
use of a spinning riffler or rotary divider (Fig. 21). Spinning rifflers are available to
process bulk samples from 25 mL all of the way to 51 L, producing 2 to 30 samples
in a single split. The spinning riffler greatly reduces operator error and the bias
associated with other types of sample dividing.
The accuracy of spin riffling (Table 7) was demonstrated in the 1968 M.Sc.
thesis of A. A. Khan, Bradford University. Sampling tests that were based on a
60 %/40 % mixture of coarse and fine sands produced the following comparisons:
recycling. This device was designed for testing coal, but it is well-suited for use with
other materials.
If a material is difficult to sieve because of the nature of the fines it contains,
then the fines can be removed by wet sieving, and then after drying the sample
can be tested normally. First, dry the sample to a constant weight and weigh to
the nearest 0.1 g. If the material readily mixes with water, then place the test
sample on the finest sieve and wash it back and forth with a gentle stream of
water such that there is no loss by rising dust or splashing. When the water pass-
ing through the sieve appears clear, the remaining sample should be dried to a
constant weight. Allow the sample to cool to room temperature and run a con-
ventional dry-sieving test using the mass of the original dry sample before wash-
ing as a total weight. If the material does not mix well with water, then place the
dried sample in a vessel and fill approximately three quarters full of water. Shake
contents vigorously to mix the material with the water, and then gently pour into
the sieve. The jar should be rinsed out more than once to recover all material.
The washing process can then be performed as described above. A small quantity
of sodium hexametaphosphate or other antiflocculant added to the water will
aid in dispersing the solid particles. Care must be taken not to break down the
friable materials.
which interferes with the proper fitting of the sieve to the pan and cover. Under
no circumstances force embedded particles out of the openings with a pick or any
other implement.
Occasionally wash the sieves in a warm soap and water solution to remove the
near-size particles lodged in the mesh. Brush the underside of the sieve carefully
while in the water to aid in the cleaning action. Ultrasonic cleaners are very useful
for cleaning sieves, especially the finer meshes that are sensitive to aggressive brush-
ing. Sieves should be immersed in a water and detergent solution in the ultrasonic
cleaner, which eases the removal of near-size particles. Do not use acid solutions for
cleaning sieves. The acid reduces the diameter of the wire, which enlarges the open-
ings; it loosens the weave in the wire cloth; and destroys the accuracy of the sieve.
Precision electroformed sieves require special care. Never touch the electro-
formed mesh with bare hands. Oils and acids on skin will stain and eventually
corrode the mesh. Do not brush sieves finer than 20 μm; they should be immersed
in an ultrasonic bath for cleaning. Warm water with just a few drops of mild dish
detergent is sufficient for cleaning.
All types of sieves should be examined frequently for defects in the cloth. Holes
or breaks are sometimes indicated by noticeable irregularities in the endpoint
weighing. Small holes or breaks should not be repaired, and any sieve found with
defects should be replaced. Do not continue to use a test sieve when the wire cloth
is worn, loose, or damaged.
A sieve is considered overloaded when there is a crowding of oversize and near-
mesh particles on the sieve surface after one half or more of the finer material has
passed. The weight of the oversize material tends to wedge the near-size particles
into the openings, blocking or “blinding” them.
5. Appendix
Table A-1—Suggested Bulk Volume of Test Sample for Sieve
Analysis with 8-in. and 200-mm Round Sievesa
Standard Sieve Designation Bulk Volume of Material
Maximum Permitted
Recommended Volume on Sieve
Volume of Material on Completion of
Standard Alternate for Test Sample Sieving
1 2 3 4
25 mm 1 in. 1,800 cm 3
900 cm3
22.4 7
/8 1,600 800
19 ¾ 1,400 700
16 5
/8 1,000 500
12.5 ½ 800 400
11.2 7
/16 800 400
9.5 3
/8 600 300
8 5
/16 500 250
6.3 ¼ 400 200
5.6 No. 3 ½ 400 200
4 5 350 150
2.80 7 240 120
2 10 200 100
1.40 14 160 80
1 18 140 70
710 μm 25 120 60
500 35 100 60
355 45 80 40
250 60 70 35
180 80 60 30
125 120 50 25
90 170 40 20
63 230 35 17
45 325 30 15
38 400 25 12
a
The recommended weight of material for a sieve test sample is calculated by multiplying the bulk
volume figure in Column 3 by the particular bulk density in grams per cubic centimeter of the material,
rounded out within a tolerance of ± 25 %. If the density figure for the material being tested is not
readily available, use the factor of the nearest similar material shown in Table A-2.
Alumina 40 0.64
Bagasse 6 0.09
(Continued)
(Continued)
(Continued)
(Continued)
(Continued)
Asbestos Fiber C1121 Test Method for Turner and Newall C17
(T and N) Wet-Length Classification
of Asbestos
(Continued)
Coal Tar and Pitch D2415 Test Method for Ash in Coal Tar and D02
Pitch
(Continued)
Coke Ash D4326 Test Method for Major and Minor D05
Elements in Coal and Coke Ash By
X-Ray Fluorescence
(Continued)
(Continued)
(Continued)
(Continued)
Mineral Filler for D546 Test Method for Sieve Analysis D04
Bituminous Paving of Mineral Filler for Bituminous
Mixtures Paving Mixtures
(Continued)
Plastics and Fillers D4350 Test Method for Corrosivity Index D20
of Plastics and Fillers
Pulverized Soil D1632 Practice for Making and Curing Soil- D18
Cement Compression and Flexure
Test Specimens in the Laboratory
(Continued)
Seed Starter Mix D5539 Specification for Seed Starter Mix D18
(Continued)
Soaps and Other D502 Test Method for Particle Size of D12
Detergents Soaps and Other Detergents
(Continued)
(Continued)
Tar and Pitch D2416 Test Method for Coking Value of D02
Tar and Pitch (Modified Conradson)
Tar and Pitch D4715 Test Method for Coking Value of D02
Tar and Pitch (Alcan)
(Continued)
(Continued)
(Continued)
(Continued)
(Continued)
(Continued)
(Continued)
Waste and Soil D4547 Standard Guide for Sampling s for D34
Volatile Organic Compounds
6. NOMENCLATURE
General Terms
Bulk density Ratio of the mass of a material to its volume in a freely poured
condition.
Effective opening The size of the largest particle that will pass a screen aperture.
Nominal size The specified dimension of the opening of a sieve about which the
actual size is permitted to vary.
Open area Ratio of the total area of the openings to the total area of the screen
expressed as a percentage.
Particle size The dimension of a particle usually expressed in terms of the smallest
sieve opening through which it will pass.
Screen (a) A surface provided with openings of uniform size and shape; (b) a
machine provided with one or more screening surfaces.
Test Sieves
Certified sieve A test sieve that has been examined and certified by an authority,
accredited for the purpose, as complying with the specifications and
tolerances of the applicable standard.
Cover (lid) A cover that fits snugly over a sieve to prevent the escape of material
being sieved.
Matched sieve A test sieve that duplicates the results of another sieve within
specified limits.
Nesting pan or A pan with a nesting skirt used in a stack of sieves to permit two or
pan w/skirt more separate sieve tests to be made simultaneously, usually with a
mechanical sieve shaker.
Pan (receiver) A pan that fits snugly beneath a sieve to receive the passing fraction.
Sieve cloth series Sieve cloth woven to a mathematically defined set of opening
widths, wire diameters, and tolerances.
Sieve frame A rigid framework that supports the sieving medium and limits the
spread of the material being sieved.
Skirt The portion of the sieve frame that extends below the sieving surface
and nests into the next finer sieve or receiving pan.
Test Sieving
End point The stopping point in a sieve test at which further sieving fails to pass
an amount sufficient to change the result.
Gross sample A representative sample taken from a large volume of material that
is too large to use in a test sieve.
Sieve analysis The results of a sieve test showing the percentages of sample
retained on (or passing) each sieve used in the test.
Size analysis The results of dividing a sample into fractions of defined limits (see
also Sieve analysis).
Size range The limits between the smallest and the largest particle in a sample.
Test sample A representative sample that is small enough to use directly in a test
sieve or series of sieves.
Test sieving Sieving with one or more test sieves to determine the particle size
distribution of a particulate material.
Production Screening
Oversize That portion of the feed material that has failed to pass through the
openings of a screen.
Oversize in Particles in a screen undersize that are larger than the nominal point
undersize of separation.
Screen efficiency The percentage of recovery of the desired portion (usually the
undersize) from the amount available in the feed.
Undersize (fines) That portion of the feed material that has passed through the
openings of a screen.
Undersize in Particles in a screen oversize that are smaller than the nominal point
oversize of separation.
Woven-Wire Cloth
Double crimp Wire cloth woven with approximately equal corrugations in warp
and shoot to lock the wires in position.
Mesh (a) The number of apertures per unit of length; (b) in countries using
English measure, the number of openings, and fraction thereof, per
linear inch counting from the center of a wire.
Plain weave Wire cloth in which each warp wire and each shoot wire passes over
one and under the next adjacent wire in both directions.
Rectangular mesh Mesh with unequal opening widths in warp and shoot direction.
Shoot wires The wires running cross-wise of the cloth as woven (also called “shute
wires”).
Space cloth Wire cloth that is designated by the clear opening between the wires
instead of by the mesh.
Square mesh Mesh with equal opening widths in warp and shoot direction.
Twilled weave Wire cloth in which each shoot wire passes successively over two and
under two warp wires and each warp wire passes successively over
two and under two shoot wires.
Warp wires The wires running the long direction of the wire cloth as woven.
Perforated Plate
Bridge width Distance between the nearest edges of two adjacent holes in a
perforated plate.
Hole size In a perforated plate, the diameter of the round hole; width of the
square hole at its midsection; smallest width of the oblong hole.
Margin Distance between the outside edges of the outside rows of holes and
the edges of a perforated plate.
Electroformed Sieves
Supporting grid A relatively thick sheet of metal having large, uniform, square
openings to which the fine mesh is bonded for support.
7. REFERENCES
Allen, T., Particle Size Measurement, Chapman and Hall, London, 1968.
Batel, W., Einfuhrung in die Korngrossenmesstechnik, Springer-Verlag, Berlin, 1960.
Cadle, R. D., Particle Size, Reinhold, New York, 1965.
Dalla Valle, J. M., Micrometritics. The Technology of Fine Particles, 2nd ed., Pitman, New York,
1948.
Herdan, G., Small Particle Statistics, 2nd ed., Academic, New York, 1960.
Irani, R. R., and Callis, C. R., Particle Size: Measurement, Interpretation and Application, Wiley,
New York, 1963.
Khan, A. A., MSc Thesis, Bradford University, 1968.
Orr, C., Jr., Particulate Technology, Macmillan, New York, 1966.
Orr, C., Jr., and Dalla Valle, J. M., Fine Particles Measurement: Size Surface and Pore Volume,
Macmillan, New York, 1959.
Rose, H. E., Measurement of Particle Size in Very Fine Powders, Chemical, New York, 1954.
Particle Size Analysis, Society for Analytical Chemistry, London, 1967.
“Powders in Industry,” SCI Monograph 14, Society of Chemical Industry, London, 1961.
Particle Size Measurement, ASTM STP 234, ASTM International, West Conshohocken, PA,
1959.
“The Physics of Particle Size Analysis,” Br. J. Appl. Phys., Vol. 5, Suppl. 3, 1954.
“Particle Size Analysis,” Trans. Inst. Chem. Eng., Vol. 25, Suppl., 1947.
New Methods for Particle Size Determination in the Subsieve Range, ASTM STP 51, ASTM Inter-
national, West Conshohocken, PA, 1941.