This procedure describes all the necessary steps to manage a QCV Alert on a VIDAS/ MiniVIDAS
system, at customer site.
It’s intended for use by bioMérieux Field Service Engineers.
The FSE must collect and confirm all important data during the intervention, the aim is:
To determine the appropriate root cause, the FSE must perform a troubleshooting step by step.
Depending of QCV value obtained by customer (TV1 issue, R3 issue, TV1 and R3 issue) follow the
appropriate chapter.
FSE must use the service manual procedures to repair. Depending of the root cause, apply the
appropriate procedure
o Visual inspection
o Pump seals on retainer plates
o Visual inspection of the seals (unstuck dot, crystallization, misplaced or stuck dots,
wreckage, fibers and aluminum particles)
o Visual inspection on the tray, on the door, on the shield insulate plate, on the bottom of
the frame)
o Check if doors are hard to close (possible root cause of unstuck dots)
o Perform the same steps as part : “2.1 Troubleshooting for TV1 issue”
Note: When replacing a spare part please refer to the instructions list “Qualification vs FSE
intervention.xls” present in Service Manual external documents.
FSE REPORT.xlsx
The FSE must complete the FSE report according to which value was out of range in the QCV alert.
The FSE report must be written in English and copied into the CRM case/SAP notification, or through
any other type of tool used to communicate these incidents with the GCS (i.e. email to GCS
address).
Refer to section “How to use this document” in FSE Report to have more information’s
All information from FSE visit must be provided to GCS as soon as possible and within 10 calendar
days from the complaint opening.
The FSE role is to perform a reminder of the bioMérieux recommendation to the customer, when it’s
needed.
Responsibilities reminder :
o bioMérieux cannot guarantee that reagent performances claimed into Package Inserts are
reached, if the maintenance procedures performed by our local representative or the customer
are not respected as recommended in the User Manual: