Professional Documents
Culture Documents
Jomi 2014suppl g1 2
Jomi 2014suppl g1 2
Purpose: To assess the literature on the accuracy and clinical performance of static computer-assisted
implant surgery in implant dentistry. Materials and Methods: Electronic and manual literature searches
were applied to collect information about (1) the accuracy and (2) clinical performance of static computer-
assisted implant systems. Meta-regression analysis was performed to summarize the accuracy studies.
Failure/complication rates were investigated using a generalized linear mixed model for binary outcomes and
a logit link to model implant failure rate. Results: From 2,359 articles, 14 survival and 24 accuracy studies
were included in this systematic review. Nine different static image guidance systems were involved. The
meta-analysis of the accuracy (24 clinical and preclinical studies) revealed a total mean error of 1.12 mm
(maximum of 4.5 mm) at the entry point measured in 1,530 implants and 1.39 mm at the apex (maximum
of 7.1 mm) measured in 1,465 implants. For the 14 included survival studies (total of 1,941 implants) using
static computer-assisted implant dentistry, the mean failure rate was 2.7% (0% to 10%) after an observation
period of at least 12 months. In 36.4% of the treated cases, intraoperative or prosthetic complications were
reported, which included: template fractures during the surgery, change of plan because of factors such
as limited primary implant stability, need for additional grafting procedures, prosthetic screw loosening,
prosthetic misfit, and prosthesis fracture. Conclusion: Different levels of quantity and quality of evidence
were available for static computer-assisted implant placement, with tight-fitting high implant survival rates
after only 12 months of observation in different indications achieving a variable level of accuracy. Future
long-term clinical data are necessary to identify clinical indications; detect accuracy; assess risk; and justify
additional radiation doses, effort, and costs associated with computer-assisted implant surgery. Int J Oral
Maxillofac Implants 2014;29 (suppl):25–42. doi: 10.11607/jomi.2014suppl.g1.2
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Tahmaseb et al
drill guide directing the implant osteotomy and im- a surgical one by executing computer transformation
plant insertion. Importantly, this entire process can be algorithms.5 The different computer guided systems
performed in such a way that the predicted ideal im- can also be differentiated in terms of their respective
plant position can be achieved without damaging the design for the drill guidance through the template. For
surrounding anatomical structures.4 example, some systems use surgical templates with
The various so-called guided systems incorporate sleeves of an increasing diameter, while others design
the planning of the implant positions, using a variety different drills with stops to achieve depth control.
of software tools. The resulting planned implant posi- Some systems allow a guided implant placement6–8
tions are then converted into surgical guides or loaded whereas in other systems the implants are inserted
into positioning software following a variety of meth- without using a guided device9–11 or after removal of
ods. Jung and coworkers5 have categorized many of the template. Some systems use pre-installed refer-
these into static and dynamic systems. Static systems ence points such as mini-implants8,12 while others use
are those that communicate predetermined sites using different reference markers (eg, gutta percha markers
surgical templates or implant guides in the operating on the CT imaging) or no references for performing the
field. Meanwhile, dynamic systems communicate the procedures. These variations make it extremely difficult
selected implant positions to the operative field with to draw a clear line in comparing the different systems.
visual imaging tools on a computer monitor, instead For this reason, a clear description on every system and
of rigid intraoral guides. The dynamic systems include their variation in use and precision can be beneficial to
surgical navigation and computer-aided navigation clinicians who are interested in these techniques.
technologies, and allow the surgeon to alter the surgi- Moreover, different accuracy measurement tech-
cal procedure and implant position in real time using the niques and terms have been introduced in the litera-
anatomical information available from the preoperative ture in the comparison of planned implant positions
plan and a CT or CBCT scan. Since the surgeon can see to actual inserted implants. Some use baseline crite-
an avatar of the drill in a 3D relationship to the patient’s ria such as entry or apical point while others use 3D
previously scanned anatomy during surgery, modifica- coordinates (eg, x-, y-, and z-axes), making it more
tions can be accomplished with significantly more infor- challenging to conduct a unified comparison.
mation. In essence, the navigation approach provides a The aim of this systematic review is to systematically as-
virtual surgical guidance that may be altered according sess the literature regarding the accuracy and the clinical
to the conditions encountered during surgery. performance, limitations, and complications of different
In their systematic review, Jung et al5 stated that the static techniques based on computer assisted technique
static systems have the tendency to be more accurate applications in guided surgical implant dentistry.
than the dynamic approaches. However, most of the
publications on navigation have been clinical stud-
ies, whereas the majority on static protocols has been Materials and Methods
preclinical (models or cadaver, etc), where more accu-
rate measurements are possible. The greater accuracy An electronic search on dental literature was per-
of these latter studies can be explained by the better formed with the purpose of collecting relevant data on
access, the greater visual control of the axis of the os- (1) the accuracy and (2) the clinical performance (sur-
teotomy, the lack of movement in the cadaver, and the vival) of computer-aided dental implant placement.
absence of saliva or blood in the preclinical models. A MEDLINE and EMBASE systematic search was com-
In the dynamic approach, the osteotomy and implant pleted using the following terms: dental AND (implant
insertion can be altered during the surgery. Thus the OR implants OR implantation OR implantology) AND
osteotomy has then no other guidance than the sur- (guide* OR computer*) (Table 1).
geon’s vision in the virtual model, giving the surgeon The results were limited to studies published be-
the ability to choose the implant position based on the tween January 1, 2008, and January 9, 2012, and
visual real-time anatomy. Hence, no true comparison is written in English, German, Italian, or French. These
possible between the planned and placed implant po- results were complemented by the data extracted in
sition. Due to this fact, in this systematic review, only a previous ITI consensus paper accessing the literature
the studies conducting computer-guided (static) sur- from 1966 through 2008.5 Two independent review-
gery were considered for inclusion. ers performed the article selection. In addition, hand
For computer-guided (static) surgery, one can distin- searches were performed using the reference lists of
guish different modalities regarding the procedure for the selected full-text articles and were also conducted
fabricating the drill guides such as stereolithography in the following pertinent implant-related journals;
(rapid prototyping) or the use of mechanical position- Clinical Implant Dentistry and Related Research, Clinical
ing devices that convert the radiographic template to Oral Implants Research, The International Journal of Oral
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Group 1
& Maxillofacial implants, Journal of Oral and Maxillofa- • The measurement of distances between the
cial Surgery, Journal of Periodontology, Journal of Pros- planned and actual position of the implants and/
thetic Dentistry, Implant Dentistry, and The International or implant angle deviations had to be clearly
Journal of Periodontics and Restorative Dentistry. Only described.
articles that reached consensus between the reviewers • Only the data on the position of actual inserted
were selected for data extraction. Figure 1 illustrates implants were included, whereas data were
the search strategy. excluded if the position of the osteotomy following
computer-guided surgery was provided but no
Inclusion and Exclusion Criteria actual insertion of the implants was performed.
This review includes only computer-guided (static) sur-
gery in which a CT/CBCT scan was conducted for com- Clinical studies:
puterized planning prior to the actual implant insertion.
Therefore, articles regarding dynamic computer-navi- • Clinical studies were included when based on at
gated surgery and 2D radiographic stents were exclud- least 5 patients.
ed. Meanwhile, for the accuracy and clinical performance • The follow-up period had to be at least 12 months.
studies, different inclusion criteria were considered. The reported data had to include implant survival
Accuracy studies: based on at least one of the following parameters:
clinical, radiographic, or patient-centered
• Studies with zygomatic, pterygoid, and outcomes of computer-assisted implant dentistry
orthodontic implants were excluded. in humans.
• Clinical, model, and cadaver studies were included. • Other factors such as prosthetic survival,
• The primary outcome of the studies had to be the complications, or bone-level changes were
accuracy of computer-guided implant surgery. incorporated if available.
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Tahmaseb et al
24 on accuracy 14 on survival
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Group 1
3D deviation 3D deviation
at apex at apex
on ti
evia
gle d
∠An
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Tahmaseb et al
Tahmaseb et al8 2011 Model Strain Gauges Exe-plan Partially Maxilla Mini–implant FG dig
edentulous
Vasak et al28 2011 Clinical Postop CBCT Nobel Guide Both Both Mixed + pin (mucosa/tooth) FG dig
Arisan et al29 2012 Clinical Postop CBCT SimPlant & Edentulous Both Mucosa + pin FG dig
CBCT
SimPlant & CT Edentulous Both Mucosa + pin FG dig
Behneke et al13 2012 Clinical Postop CBCT Implant 3D Partially Both Tooth FG lab
edentulous FHP lab
D’Haese et al30 2012 Clinical Postop CBCT Facilitate Edentulous Maxilla Mucosa + pin FG dig
Di Giacomo et al10 2012 Clinical Postop CBCT Implant Viewer Edentulous Both Mucosa + pin FHP dig
Maxilla Mucosa + pin FHP dig
Mandible Mucosa + pin FHP dig
Kuhl et al16 2012 Cadaver Postop CBCT coDiagnostiX Both Mandible Mixed (mucosa/tooth) FHP lab
Cadaver FG lab
Soares et al31 2012 Model Postop CBCT NeoGuide Edentulous Mandible Mucosa + pin FG dig
FHP = freehand placement; FG = fully guided implant insertion; lab = templates produced in laboratory;
dig = digitally produced templates (stereolithographic or milled); *at entry; †at apex
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Group 1
Error entry (mm) Error apex (mm) Error angle (°) Error height (mm)
Implants Flap-
(n) less x,y,z Mean SD Max Mean SD Max Mean SD Max Mean SD Max
10 Y 3D 0.8 0.3 – 0.9 0.3 – 1.8 1 – – – 1.1
21 N 3D 1.45 1.42 4.5 2.99 1.77 7.1 7.25 2.67 12.2 – – –
43 Y ° – – – – – – 3.55 1.08 – – – –
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Tahmaseb et al
Table 4 Guided Systems Used in with 15 to 279 implants in the clinical studies and 4 to
Selected Studies 145 in vitro studies. Nine computer-guided systems
were used (Table 4). Fifteen of all selected studies (and
System No. of studies using system therefore the majority of the studies) used NobelGuide.
Nobel Guide 15 The deviation at the implant entry point was re-
SimPlant 9 ported in 23 of the 24 studies while the angle and
StentCad Classic/Beyond 4/1 apex deviations were mentioned in 21 and 19 stud-
coDiagnostiX 3 ies, respectively. For comparing the data as accurately
as possible, the data were converted to 3D deviation
Exe-plan 3
when possible. This conversion resulted in suitability
Dental Slice/NeoGuide 2
of 21 studies on entry point error and 17 on implant
Implant Viewer 1.9 2 apex error for comparison. The deviations in implant
Facilitate (AstraTech) 1 height were only rarely reported and were converted
Implant 3D 1 to 3D deviation when available. Because of the limited
reporting on this parameter, the implant height devia-
tions were solely reported in the results (Table 3).
Differences in accuracy between each group were The overall average deviation at the implant entry
assessed by means of metaregression. A separate point was 1.12 mm as measured in 1,530 implants, and
analysis was performed for each of the three types of the maximum reported deviation was 4.5 mm. As deter-
deviation (eg, error at the entry point and apex, error mined for 1,465 implants, the average deviation report-
in the angle). ed at the apex was 1.39 mm, with a maximum reported
In addition, forest plots were prepared to visualize deviation of 7.1 mm. The largest number of measure-
the difference between groups. Since evidence of het- ments (1,854 implants) was performed on the implant
erogeneity was observed between the publications, angle; the average angular deviation was 3.89 degrees
the totals were calculated using random effects meta- with a maximum reported deviation of 21.16 degrees.
analysis for continuous variables. Statistically significant differences were observed
For survival, a generalized linear mixed model for when the following parameters were compared:
binary outcomes and a logit link was used to model
implant failure rate. • Study design
• Flapless versus flap approach
• Freehand versus guided implant placement
Results • Guide support
After the initial search, a total of 3,971 titles were found. No statistically significant differences were found in
Because multiple search engines were used (PubMed the testing of modalities in maxilla vs mandible, fully
& EMBASE) the duplicates needed to be filtered out, edentulous vs partially edentulous, or the guide pro-
which reduced the number of articles to 2,359 titles. duction. Table 5 shows the P values of the differences
Subsequently, 139 abstracts of relevant studies were per compared parameter.
then selected by two reviewers. After reviewers ex-
amined and discussed the abstracts, 117 publications Study Design
were selected, in consensus, for full-text evaluation. In Cadaver Studies. As measured in 390 implants, the
addition, 12 full-text articles that were selected from lowest and highest mean deviations at the entry point
the 2008 consensus statement5 were analyzed and in the cadaver studies were 0.8 and 1.5 mm, respec-
added to selected publications. The final selection as tively. The minimum and maximum measured values
based upon full-text analysis and the inclusion and ex- were 0.07 and 3.54 mm, respectively. The lowest and
clusion criteria resulted in 24 accuracy studies and 14 highest mean errors at the apex were 0.9 and 1.84 mm,
survival studies that could be used for data extraction respectively, where as the lowest and highest values
(Fig 1). were 0.12 and 3.64 mm, respectively. The lowest and
highest mean angular deviations were 1.8 and 7.9 de-
Accuracy Studies grees with the minimum and maximum values of 0.08
Twenty-four articles provided useful information on and 11.9 degrees, respectively.
the accuracy of computer-guided (static) implant sur- Model Studies. As determined in only 74 implants,
gery (Table 3). Fourteen of these articles were clinical the lowest and the highest mean deviations at the
studies, while ten were in vitro (model, cadaver) stud- entry point were 0.025 and 1.38 mm, respectively.
ies. The sample size of the groups varied considerably The maximum and minimum deviations at the entry
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Group 1
point were not often reported. The lowest reported Flapless versus Flap Approach
value was 0 whereas the maximum measured value Figure 5 illustrates a forest plot presenting significant
was 2.25 mm. The lowest and highest mean error at differences (P < .05) between cases treated with a flap-
the apex was 0.34 and 1.39 mm, whereas the minimum less protocol compared to those in which a flap was
and maximum reported values were 0.12 and 2.25 mm, raised. The flapless procedures seemed to show a sig-
respectively. The lowest mean angular deviation was nificantly better accuracy.
0.7 degrees, while the highest observed mean was 2.16
degrees. The minimum and maximum values were 0.3 Freehand versus Guided Implant Placement
and 2.16 degrees, respectively. Guided implant placement showed a statistically supe-
Clinical Studies. The majority of accepted studies rior accuracy when they are compared with freehand
were clinical studies, which assessed a total of 2,355 placement after guided osteotomy (Fig 6).
implants. The lowest and highest mean error at the
entry point was 0.15 and 1.7 mm with minimum and Guide Support
maximum values of 0 and 4.5 mm, respectively. The When all study types (clinical, in vitro) were concerned,
mean apical deviation varied from 0.28 to 2.99 mm the accuracy of mini-implant–supported guides was
with a minimum and maximum of 0.3 and 7.1 mm re- significantly higher than all other types of support,
spectively. The mean angular deviation ranged from except mucosa. Bone-supported guides showed sig-
1.49 to 8.54 degrees with a minimum and maximum of nificantly larger deviations than other types of guide
0 and 21.16 degrees, respectively. support. Tooth-supported guides tended to be slight-
A forest plot (Fig 4) shows the data for the deviation ly more accurate than mucosa or mucosa and pin–
at the point of entry, the apex, and for the angulation supported guides; however, these differences were only
based on the study design. Statistically significant differ- found in some of the compared parameters. When only
ences were observed for all three parameters in the clini- clinical studies were assessed, the accuracy of bone-
cal trials versus the model studies. Model studies showed supported guides were significantly lower in almost
a significantly better accuracy. However, the number of every compared parameter (Table 5). There were no
implants (n = 245) tested in these studies was much low- clinical studies available that investigate accuracy when
er than that (n = 1,560) used in the clinical assessments. using mini-implants.
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Tahmaseb et al
Cadaver
van Steenberghe et al20 (2002) 0.90 (0.71; 1.09)
Cadaver van Assche et al21 (2007) 1.20 (0.80; 1.60)
van Steenberghe et al20 (2002) 0.80 (0.61; 0.99) Pettersson et al6 (2010a) 1.25 (1.14; 1.36)
Van Assche et al21 (2007) 1.10 (0.70; 1.50) 1.76 (1.60; 1.92)
Pettersson et al6 (2010a) 1.06 (0.97; 1.15) Kuhl et al16 (2013)
Ruppin et al22 (2008) 1.62 (1.32; 1.92) Total 1.28 (0.92; 1.65)
Kuhl et al16 (2013) 1.55 (1.35; 1.75) Clinical
Total 1.22 (0.93; 1.51) Ozan et al18 (2009) 1.41 (1.24; 1.58)
Clinical
Ozan et al18 (2009) 1.11 (0.98; 1.24) Di Giacomo et al11 (2005) 2.99 (2.23; 3.75)
Di Giacomo et al11 (2005) 1.45 (0.84; 2.06) Ersoy et al9 (2008) 1.51 (1.31; 1.71)
Ersoy et al9 (2008) 1.22 (1.05; 1.39) Pettersson et al19 (2012b) 1.22 (1.12; 1.32)
Pettersson et al19 (2012b) 0.95 (0.86; 1.04) Vasak et al28 (2011) 1.05 (0.91; 1.19)
Platzer et al27 (2011) 0.42 (0.29; 0.55)
Vasak et al28 (2011) 0.82 (0.71; 0.93) Di Giacomo et al10 (2012) 1.79 (1.53; 2.05)
Di Giacomo et al10 (2012) 1.35 (1.19; 1.51) D’Haese et al30 (2012) 1.13 (1.01; 1.25)
D’Haese et al30 (2012) 0.91 (0.81; 1.01) Cassetta et al26 (2013) 1.91 (1.79; 2.03)
Cassetta et al26 (2013) 1.50 (1.41; 1.58) 1.44 (0.99; 1.89)
Arisan et al24 (2010) 1.22 (0.83; 1.60) Arisan et al24 (2010)
Arisan et al29 (2012) 0.78 (0.72; 0.84) Arisan et al29 (2012) 0.81 (0.74; 0.87)
Total 1.04 (0.85; 1.24) Total 1.45 (1.18; 1.73)
Model Model
Dreiseidler et al23 (2009) 0.33 (0.26; 0.41)
Tahmaseb et al7 (2010) 0.06 (0.03; 0.08) Dreiseidler et al23 (2009) 0.43 (0.34; 0.52)
Tahmaseb et al8 (2011) 0.04 (0.04; 0.04) Soares et al31 (2012) 1.39 (1.21; 1.57)
Soares et al31 (2012) 1.38 (1.19; 1.57) Viegas et al25 (2010) 0.39 (0.29; 0.49)
Viegas et al25 (2010) 0.33 (0.25; 0.41) Total 0.73 (0.27; 1.19)
Total 0.36 (0.23; 0.49)
Grand total 0.93 (0.74; 1.13) Grand total 1.29 (1.05; 1.52)
Fig 4 Forest plot illustrating mean deviation at all parameters, stratified by principle of study design.
Flap raised
Di Giacomo et al11 (2005) 1.45 (0.84; 2.06)
Ruppin et al22 (2008) 1.62 (1.32; 1.92)
Ozan et al18 (2009) 1.28 (1.03; 1.53) Flap raised
van Steenberghe et al20 (2002) 0.80 (0.61; 0.99) Di Giacomo et al11 (2005) 2.99 (2.23; 3.75)
Arisan et al24 (2010) 1.59 (1.52; 1.65) Ozan et al18 (2009) 1.57 (1.32; 1.82)
Ersoy et al9 (2008) 1.35 (1.16; 1.55) van Steenberghe et al20 (2002) 0.90 (0.71; 1.09)
Total 1.34 (1.05; 1.63) Arisan et al24 (2010) 1.90 (1.80; 2.00)
Ersoy et al9 (2008) 1.50 (1.19; 1.82)
Flapless
Total 1.69 (1.19; 2.19)
van Assche et al21 (2007) 1.10 (0.70; 1.50)
1.10 (0.92; 1.28) Flapless
Ersoy et al9 (2008)
van Assche et al (2007)
21
1.20 (0.80; 1.60)
Pettersson et al6 (2010a) 1.06 (0.97; 1.15)
Ersoy et al9 (2008) 1.40 (1.09; 1.71)
Pettersson et al19 (2012b) 0.95 (0.86; 1.04)
Pettersson et al6 (2010a) 1.25 (1.14; 1.36)
Vasak et al28 (2011) 0.82 (0.71; 0.93)
Pettersson et al19 (2012b) 1.22 (1.12; 1.32)
Platzer et al27 (2011) 0.42 (0.29; 0.55)
Vasak et al28 (2011) 1.05 (0.91; 1.19)
Di Giacomo et al10 (2012) 1.35 (1.19; 1.51) 1.79 (1.53; 2.05)
Di Giacomo et al10 (2012)
D’Haese et al30 (2012) 0.91 (0.81; 1.01) D’Haese et al30 (2012) 1.13 (1.01; 1.25)
Soares et al31 (2012) 1.38 (1.19; 1.57) Soares et al31 (2012) 1.39 (1.21; 1.57)
Arisan et al24 (2010) 1.00 (0.74; 1.26) Arisan et al24 (2010) 1.19 (0.80; 1.58)
Ozan et al18 (2009) 0.93 (0.81; 1.05) Ozan et al18 (2009) 1.26 (0.64; 1.89)
Kuhl et al16 (2013) 1.55 (1.35; 1.75) Kuhl et al16 (2013) 1.76 (1.60; 1.92)
Arisan et al29 (2012) 0.78 (0.72; 0.84) Arisan et al29 (2012) 0.81 (0.74; 0.87)
Total 1.01 (0.88; 1.15) Total 1.28 (1.09; 1.47)
Grand total 1.12 (0.95; 1.28) Grand total 1.39 (1.19; 1.59)
Fig 5 Forest plot illustrating mean deviation at all parameters, stratified by principle of the surgery (flap raised vs flapless).
Although no overall statistically significant differ- tween the two groups. However, some studies found
ences were found in the remaining comparisons for significantly better accuracy in one arch compared
guided implant placement, it may be of interest to to the other. Ozan et al17 reported significantly bet-
state them briefly: ter accuracy in the mandible when compared to the
Maxilla versus mandible: No overall significant dif- maxilla within the same study, whereas Pettersson et
ference was found for any of these parameters be- al6 observed a statistically significant higher deviation
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Group 1
Survival Studies
Cadaver
van Steenberghe et al20 (2002) 1.80 (1.18; 2.42)
Fifteen clinical studies reporting on implant survival
van Assche et al21 (2007) 1.80 (1.35; 2.25) for at least 12 months were selected (Table 6). Seven of
Pettersson et al6 (2010a) 2.64 (2.41; 2.87)
Ruppin et al22 (2008) 7.90 (6.35; 9.45) these studies reported on bone loss. The average sur-
Kuhl et al16 (2013) 3.90 (2.98; 4.81)
Total 3.32 (2.28; 4.36) vival rate of the 1,941 implants included in this review
Clinical
Ozan et al18 (2009) 4.10 (3.67; 4.53)
was 97.3%. Several studies showed longer follow-up
Di Giacomo et al11 (2005) 7.25 (6.11; 8.39) periods, while others had large dropouts at 12 months.
Ersoy et al9 (2008) 4.90 (4.42; 5.38)
Pettersson et al19 (2012b) 2.76 (2.47; 3.05) Nevertheless, when only studies reporting on survival
Nickenig et al15 (2010) 4.20 (2.96; 5.44)
Vasak et al28 (2011) 3.53 (3.16; 3.90) at 12 months were examined and further corrected
Di Giacomo et al10 (2012)
D’Haese et al30 (2012)
6.53
2.60
(5.44;
(2.24;
7.62)
2.96)
for dropouts, the survival percentage was the same:
Behneke et al13 (2012) 1.80 (1.25; 2.36) 97.3%.
Ozan et al17 (2011) 4.52 (3.39; 5.65)
Cassetta et al26 (2013) 4.79 (3.86; 5.72) In 12 out of the 14 selected studies the implants
Arisan et al24 (2010) 3.94 (3.27; 4.61)
Arisan et al29 (2012) 3.38 (3.17; 3.60) were immediately loaded. Three studies even reported
Model
Total 4.06 (3.50; 4.62) on immediate definitive prostheses. Sanna et al32 and
Dreiseidler et al (2009)
23
1.09 (0.89; 1.29) Johansson et al34 used adjustable abutments while
Soares et al31 (2012) 2.16 (1.74; 2.58)
Viegas et al25 (2010) 1.03 (0.30; 1.76) Tahmaseb et al12 achieved sufficient accuracy to install
Total 1.44 (0.68; 2.21)
Grand total 3.53 (2.98; 4.08) the final prosthesis without adjusting abutments.
Eight studies reported on prosthetic survival. The
0 2 4 6 8 average prosthetic survival rate based on these 211
Error angle prostheses was 95.5%.
There were no statistically significant differences
between the various groups (immediate loading ver-
sus delayed loading, guide support). However, a slight
difference in survival was observed between the
maxilla and mandible in favor of the maxilla (P value
Flap raised
Di Giacomo et al11 (2005) 7.25 (6.11; 8.39)
difference of .14). Table 7 demonstrates the P value
Ruppin et al22 (2008) 7.90 (6.35; 9.45) differences per reported parameter.
Ozan et al18 (2009) 4.63 (3.91; 5.35)
van Steenberghe et al20 (2002) 1.80 (1.18; 2.42)
Arisan et al24 (2010) 4.83 (4.53; 5.13) Complications
Ersoy et al9 (2008) 5.04 (4.52; 5.57)
Total 5.13 (3.86; 6.39) Within the 2,355 implants inserted in 343 treated cases,
Flapless a total of 125 cases reported complications. Although
van Assche et al (2007)
21
1.80 (1.35; 2.25)
Ersoy et al9 (2008) 4.70 (4.09; 5.31) the numbers should be interpreted with caution (not
Pettersson et al6 (2010a) 2.64 (2.41; 2.87) every study reports on all possible complications), a
Pettersson et al19 (2012b) 2.76 (2.47; 3.05)
Nickenig et al15 (2010) 4.20 (2.96; 5.44) cumulative complication rate of 36.4% per case was
Vasak et al28 (2011) 3.53 (3.16; 3.90) calculated.
Di Giacomo et al10 (2012) 6.53 (5.44; 7.62)
D’Haese et al30 (2012) 2.60 (2.24; 2.96) Eight studies recorded template fractures occurring
Soares et al31 (2012) 2.16 (1.74; 2.58) during surgery. The incidence was 3.6% (7 out of 192
Ozan et al17 (2011) 4.52 (3.39; 5.65)
Arisan et al24 (2010) 3.50 (2.83; 4.17) templates). All the fractures occurred in three of eight
Ozan et al18 (2009) 3.68 (2.11; 5.25) studies.
Kuhl et al16 (2013) 3.90 (2.98; 4.81)
Arisan et al29 (2012) 3.38 (3.17; 3.60) Ten studies reported surgical plan changes per im-
Total 3.42 (2.99; 3.85)
Grand total 3.94 (3.44; 4.45)
plant. The overall incidence was 2.0% (23 out of 1,133
implants).
0 2 4 6 8
Five studies reported on implants lost during place-
Error angle ment because of the lack of primary stability. This com-
plication was recorded as occurring two times in one
study and three times in another study.10,16 The overall
incidence was 1.3% (5 out of 383 planned implants).
These implants were not counted in the implant sur-
vival since they were not successfully inserted in the
in the mandible. Of note, these studies used different first place.
supports for the surgical guides (eg, mucosa, mucosa Ten studies recorded the occurrence of prosthesis
and pin, tooth support). fracture. The incidence was 10.19% (26 out of 238 pros-
Guide production: No overall significant differenc- theses).
es were noted between different guide production Five studies registered the occurrence of screw
types. loosening. The incidence was 2.9% (23 out of 798).
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Tahmaseb et al
Prosthetic
Author Year Study design System Implant appliance Jaw Guide support
Sanna et al32 2007 Prospective NobelGuide Nobel FB Maxilla Mucosa + pin
Balshi et al33 2008 Retrospective Nobel Guide Nobel FB Both Mucosa + pin
Johansson et al34 2009 Prospective Nobel Guide Nobel FB Maxilla Mucosa + pin
Komiyama et al14 2008 Retrospective Nobel Guide Nobel FB Both Mucosa + pin
Barter35 2010 Case series CoDiagnostiX Straumann 4 FPDs + 2 IOD Maxilla Mucosa + pin
Gillot et al36 2010 Retrospective Nobel Guide Nobel FB Maxilla Mucosa + pin
Meloni et al37 2010 Retrospective Nobel Guide Nobel FB Maxilla Mucosa + pin
Nikzad et al38 2010 Prospective SimPlant div. FPDs Mandible Tooth
Pomares39 2010 Retrospective Nobel Guide Nobel FB Both Mucosa + pin
Maxilla
Mandible
Van de Velde et al40 2010 RCT SimPlant Straumann FB Maxilla Tooth
Landázuri-Del Barrio et al41 2013 Prospective Nobel Guide Nobel FB Mandible Mucosa + pin
Abboud et al42 2012 Retrospective NobelGuide Nobel 3 FPDs + 3 FB Both Tooth or Mucosa + pin
SimPlant Ankylos 4 FPDs + 4 FB Both Tooth, bone, or mucosa
Di Giacomo et al10 2012 Prospective Implant Viewer 1.9 E-fix FB Both Mucosa + pin
Tahmaseb et al12 2012 Prospective Exe-plan Straumann FB Both Mini–implants
Maxilla
Mandible
FB = fixed full-arch bridge; FPDs = fixed partial dentures; IOD = implant-supported overdentures; NR = not reported; div = diverse;
*cumulative survival rate.
Fig 6 Forest plot illustrating mean deviation of all parameters, stratified by principle of the implant insertion (freehand placement
vs guide implant placement).
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Group 1
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Tahmaseb et al
Mucosa
Ersoy et al9 (2008) 1.70 (1.29; 2.11)
Mucosa Ozan et al18 (2009) 1.60 (1.24; 1.96)
Ersoy et al9 (2008) 1.10 (0.81; 1.39) Total 1.64 (1.37; 1.91)
Ozan et al18 (2009) 1.06 (0.85; 1.27) Mucosa + PIN
Total 1.07 (0.90; 1.25) Pettersson et al6 (2010a) 1.25 (1.14; 1.36)
Mucosa + PIN Pettersson et al19 (2012b) 1.22 (1.12; 1.32)
Pettersson et al196 (2010a) 1.06 (0.97; 1.15) Di Giacomo et al10 (2012) 1.79 (1.53; 2.05)
Pettersson et al 10(2012b) 0.95 (0.86; 1.04) D’Haese et al30 (2012) 1.13 (1.01; 1.25)
Di Giacomo et al 30 (2012) 1.35 (1.19; 1.51) Soares et al31 (2012) 1.39 (1.21; 1.57)
D’Haese et al31 (2012) 0.91 (0.81; 1.01) Arisan et al24 (2010) 1.08 (0.45; 1.70)
Soares et al24 (2012) 1.38 (1.19; 1.57) Arisan et al29 (2012) 0.81 (0.74; 0.87)
Arisan et al29 (2010) 0.96 (0.44; 1.49) Total 1.24 (1.01; 1.46)
Arisan et al (2012) 0.78 (0.72; 0.84) Bone
Total 1.05 (0.90; 1.21) van Steenberghe et al20 (2002) 0.90 (0.71; 1.09)
Bone Di Giacomo et al11 (2005) 2.99 (2.23; 3.75)
van Steenberghe et al20 (2002) 0.80 (0.61; 0.99) Ersoy et al9 (2008)) 1.60 (1.16; 2.04)
Di Giacomo et al 9 (2005)
11
1.45 (0.84; 2.06) Ozan et al18 (2009) 1.57 (1.32; 1.82)
Ersoy et al (2008) 1.30 (1.01; 1.59) Viegas et al25 (2010) 0.39 (0.29; 0.49)
Ruppin et al22 (2008) 1.62 (1.32; 1.92) Arisan et al24 (2010) 1.90 (1.80; 2.00)
Ozan et al18 (2009) 1.28 (1.03; 1.53) Total 1.52 (0.81; 2.22)
Viegas et al25 (2010) 0.33 (0.25; 0.41) Tooth
Arisan et al24 (2010) 1.59 (1.52; 1.65) van Assche et al21 (2007) 1.20 (0.80; 1.60)
Total 1.19 (0.63; 1.74) Ersoy et al9 (2008) 1.30 (1.03; 1.57)
Tooth Ozan et al18 (2009) 0.96 (0.75; 1.17)
van Assche et al219 (2007) 1.10 (0.70; 1.50) Dreiseidler et al23 (2009) 0.43 (0.34; 0.52)
Ersoy et al (2008) 1.10 (0.87; 1.33) Arisan et al24 (2010) 1.31 (0.71; 1.91)
Ozan et al18 (2009) 0.87 (0.73; 1.01) 1.01 (0.58; 1.45)
Dreiseidler et al23 (2009) 0.33 (0.26; 0.41) Total
Mini Implant
Platzer et al27 (2011) 0.42 (0.29; 0.55)
Arisan et al24 (2010) 1.05 (0.56; 1.54) van Assche et al21 (2007) 1.20 (0.80; 1.60)
Total 0.78 (0.49; 1.06) Ersoy et al9 (2008) 1.30 (1.03; 1.57)
Mini Implant Ozan et al18 (2009) 0.96 (0.75; 1.17)
Tahmaseb et al87 (2011) 0.04 (0.02; 0.06) Dreiseidler et al23 (2009) 0.43 (0.34; 0.52)
Tahmaseb et al (2010) 0.06 (0.03; 0.08) Arisan et al24 (2010) 1.31 (0.71; 1.91)
Total 0.05 (0.03; 0.06) Total 1.01 (0.58; 1.45)
Grand total 0.94 (0.73; 1.15) Grand total 1.29 (1.06; 1.52)
Fig 7 Forest plot illustrating statistical evaluation based on the guide support (tooth, bone, mucosa, and mucosa + pin support).
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Group 1
Mucosa
The authors intentionally decided not to include
Ersoy et al9 (2008)
Ozan et al18 (2009)
4.90
4.51
(4.00;
(3.76;
5.80)
5.26) studies that analyzed the accuracy of computer-
Ozan et al17 (2011) 5.32 (3.42; 7.23)
Total 4.73 (4.17; 5.28) guided surgery if they only reported the position of
Mucosa + PIN
Pettersson et al6 (2010a) 2.64
2.76
(2.41;
(2.47;
2.87)
3.05)
the osteotomy, but the actual implant was not in-
Pettersson et al1910(2012b)
Di Giacomo et al30 (2012)
D’Haese et al31 (2012)
6.53
2.60
(5.44;
(2.24;
7.62)
2.96)
serted.15,18–20,22–26,29 The reason for this exclusion is
Soares et al17 (2012)
Ozan et al (2011)
2.16
3.72
(1.74;
(3.49;
2.58)
3.96) that guided drill holes after osteotomy are only an
Arisan et al24 (2010) 3.56 (2.26; 4.86)
Arisan et al29 (2012) 3.38 (3.17; 3.60) indication of the position and cannot be compared
3.25 (2.76; 3.74)
Bone Total with the actual implant positions, since implants can
van Steenberghe et al20 (2002) 1.80 (1.18; 2.42)
Di Giacomo et al119 (2005)
Ersoy et al22 (2008)
7.25
5.10
(6.11;
(4.31;
8.39)
5.89)
be inserted in a deviant position. This exclusion would
Ruppin et al18 (2008)
Ozan et al (2009)
7.90
4.63
(6.35;
(3.91;
9.45)
5.35) lead to a more meaningful comparison. Nonetheless,
Viegas et al25 (2010) 1.03 (0.30; 1.76)
Arisan et al24 (2010) 4.83 (4.53; 5.13) Table 8 shows a list of the studies assessing the accu-
Total 4.58 (3.09; 6.06)
Tooth
1.80 (1.35; 2.25)
racy based only on osteotomies.
van Assche et al219 (2007)
Ersoy et al (2008)
Ozan et al18 (2009)
4.40
2.91
(3.78;
(2.44;
5.02)
3.38) When comparing the data for the maxilla and man-
Dreiseidler et al15
23
(2009)
Nickenig et al13 (2010)
1.09
4.20
(0.89;
(2.96;
1.29)
5.44) dible, some publications reported no differences.9,29
Behneke et al24 (2012) 1.80 (1.25; 2.36)
Arisan et al (2010) 3.42 (3.22; 3.62) Ozan et al17 reported significantly better accuracy
Total 2.76 (1.74; 3.78)
Mini Implant
1.80 (1.35; 2.25)
in the mandible compared to the maxilla within the
van Assche et al219 (2007)
Ersoy et al (2008)
Ozan et al18 (2009)
4.40
2.91
(3.78;
(2.44;
5.02)
3.38) same study, while others10 reported profoundly higher
Dreiseidler et al15
23
(2009)
Nickenig et al13 (2010)
1.09
4.20
(0.89;
(2.96;
1.29)
5.44) deviations in the maxilla as well. However, Pettersson
1.80 (1.25; 2.36)
Behneke et al24 (2012)
Arisan et al (2010) 3.42 (3.22; 3.62) et al6 observed a statistically significant higher devia-
Total 2.76 (1.74; 3.78)
Grand total 3.65 (3.15; 4.14) tion in the mandible. RCTs looking to these factors in-
dividually might shed light on their impact on overall
0 2 4 6 8 precision.
Error angle Even though implant placement with a flapless ap-
proach seems to show significantly more accuracy, one
has to interpret these numbers with care. All six stud-
ies where a flap was raised reported the use of bone-
supported drill guides. The inaccuracy might thus be
related to the guide design rather than to the raising
of a flap as such.
might be caused by movements of the surgical guide As demonstrated in the studies selected in this
during implant preparation. This group suggested fur- systematic review as well as recent EAO consensus
ther improvements that could provide better stability publication on the same topic,43 different factors
of the template during surgery when unilateral bone- (teeth- versus mucosa- versus implant-supported;
supported and non–tooth-supported templates are type of guidance, etc) can play a crucial role in the
used. Moreover, SLA (computer-assisted manufacture overall success of these advanced techniques. There-
[CAM]) guides had slightly better accuracy than the fore, it would be of high importance to perform ran-
lab guides (non-CAM), although the number of cases domized clinical trials, analyzing the importance of
was significantly lower for the non-CAM group (171 vs one specific factor separately and the impact of their
1,569 implants). Furthermore, the support of guides mutual interactions. Generally it can be assumed that
has a significant impact on accuracy (Fig 7). Tahmaseb an accumulation of series of different types of errors
et al7 showed that guides supported by mini-implants can occur during the entire diagnostic and operative
provided better accuracy than all the other types of procedure leading to larger implant deviations. Finally,
support. This might be the result of the reproducibil- because of different study designs (human versus ca-
ity of the template position during the acquisition of daver or model, drill holes versus implants, or different
radiographic data and during implantation, especially evaluation methods), it is not possible to identify one
in edentulous patients. This group also used a system system as superior or inferior to others.
with a vertical control device and adjustable osteot-
omy drills, which might have improved the precision.
Moreover, for the clinical studies (Fig 8), a statistically Conclusions
significant lower accuracy was observed in the bone-
supported guides. These results could also explain As observed in this systematic review, nine different
why the flapped approaches had a much lower accu- computer-assisted (static) guided implant systems are
racy than the flapless ones (Fig 4), as the majority of described in the literature. The clinical performance
treatments where a flap was raised conducted bone- of these systems reveals a high implant survival rate
supported surgical guides. of 97.3% after 12 months of observation in different
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Tahmaseb et al
Fig 8 Forest plot illustrating statistical evaluation based on the guide support in clinical studies only (tooth, bone, mucosa, and
mucosa + pin support).
clinical situations. Furthermore, a high overall rate of curacy compared to the bone-supported guides. A
surgical and prosthetic complications and unexpected different level of evidence was stated although long-
events of 36.4% occurred at different levels of complexity. term RCTs were lacking. Long-term clinical data and
The accuracy of these systems depends on all the randomized clinical trials are necessary to detect and
cumulative and interactive errors involved, from data- understand the different factors individually and their
set acquisition to the surgical procedure. The meta- mutual interaction influencing the accuracy of these
analysis of the in vitro and in vivo studies revealed a techniques. Additionally, as it was concluded in the ITI
total mean error of 1.12 mm at the entry point and systematic review in 2008,5 no evidence yet suggest
1.39 mm at the apex. that computer-assisted surgery is superior to conven-
Furthermore, it can be stated that the tooth- and tional procedures in terms of safety, outcomes, mor-
mucosa-supported guides seem to have a better ac- bidity, or efficiency.
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Group 1
10. Di Giacomo GA, da Silva JV, da Silva AM, Paschoal GH, Cury PR, Szarf
Mucosa (Clinical)
4.90 (4.00; 5.80) G. Accuracy and complications of computer-designed selective
Ersoy et al9 (2008)
Ozan et al18 (2009) 4.51 (3.76; 5.26) laser sintering surgical guides for flapless dental implant placement
Ozan et al17 (2011) 5.32 (3.42; 7.23) and immediate definitive prosthesis installation. J Periodontol 2012;
Total 4.73 (4.17; 5.28) 83:410–419.
Mucosa + PIN (Clinical) 11. Di Giacomo GA, Cury PR, de Araujo NS, Sendyk WR, Sendyk CL. Clin-
Pettersson et al (2012b)
19 2.76 (2.47; 3.05) ical application of stereolithographic surgical guides for implant
Di Giacomo et al10 (2012) 6.53 (5.44; 7.62) placement: Preliminary results. J Periodontol 2005;76:503–507.
D’Haese et al30 (2012) 2.60 (2.24; 2.96) 12. Tahmaseb A, De Clerck R, Aartman I, Wismeijer D. Digital protocol
Ozan et al17 (2011) 3.72 (3.49; 3.96) for reference-based guided surgery and immediate loading: A
Arisan et al24 (2010) 3.56 (2.26; 4.86) prospective clinical study. Int J Oral Maxillofac Implants 2012;27:
Arisan et al29 (2012) 3.38 (3.17; 3.60) 1258–1270.
Total 3.57 (3.00; 4.13) 13. Behneke A, Burwinkel M, Behneke N. Factors influencing transfer
Bone (Clinical) accuracy of cone beam CT-derived template-based implant place-
Di Giacomo et al11 (2005) 7.25 (6.11; 8.39) ment. Clin Oral Implants Res 2012;23:416–423.
Ersoy et al9 (2008) 5.10 (4.31; 5.89) 14. Komiyama A, Klinge B, Hultin M. Treatment outcome of imme-
Ozan et al18 (2009) 4.63 (3.91; 5.35) diately loaded implants installed in edentulous jaws following
Arisan et al24 (2010) 4.83 (4.53; 5.13) computer-assisted virtual treatment planning and flapless surgery.
Total 5.32 (4.51; 6.12) Clin Oral Implants Res 2008;19:677–685.
Tooth (Clinical) 15. Nickenig HJ, Wichmann M, Hamel J, Schlegel KA, Eitner S. Evalu-
Ersoy et al9 (2008) 4.40 (3.78; 5.02) ation of the difference in accuracy between implant placement
Ozan et al18 (2009) 2.91 (2.44; 3.38) by virtual planning data and surgical guide templates versus the
Nickenig et al15 (2010) 4.20 (2.96; 5.44) conventional free-hand method—A combined in vivo–in vitro
Behneke et al13 (2012) 1.80 (1.25; 2.36) technique using cone-beam CT (Part II). J Cranomaxillofac Surg
Arisan et al24 (2010) 3.42 (3.22; 3.62)
2010;38:488–493.
Total 3.28 (2.54; 4.03)
16. Kuhl S, Zurcher S, Mahid T, Muller-Gerbl M, Filippi A, Cattin P.
Grand total 4.08 (3.64; 4.51)
Accuracy of full guided vs half-guided implant surgery. Clin Oral
Implants Res 2013;24:763–769.
17. Ozan O, Orhan K, Turkyilmaz I. Correlation between bone density
0 2 4 6 8
and angular deviation of implants placed using CT-generated surgi-
Error angle
cal guides. J Craniofac Surg 2011;22:1755–1761.
18. Ozan O, Turkyilmaz I, Ersoy AE, McGlumphy EA, Rosenstiel SF. Clini-
cal accuracy of 3 different types of computed tomography-derived
stereolithographic surgical guides in implant placement. J Oral
Maxillofac Surg 2009;67:394–401.
19. Pettersson A, Komiyama A, Hultin M, Nasstrom K, Klinge B. Accu-
References racy of Virtually Planned and Template Guided Implant Surgery on
Edentate Patients. Clin Implant Dent Relat Res 2012;14:527–537.
20. van Steenberghe D, Naert I, Andersson M, Brajnovic I, Van Cleynen-
1. Loubele M, Bogaerts R, Van Dijck E, et al. Comparison between breugel J, Suetens P. A custom template and definitive prosthesis
effective radiation dose of CBCT and MSCT scanners for dentomax- allowing immediate implant loading in the maxilla: A clinical report.
illofacial applications. Eur J Radiol 2009;71:461–468. Int J Oral Maxillofac Implants 2002;17:663–670.
2. Guerrero ME, Jacobs R, Loubele M, Schutyser F, Suetens P, Steenber- 21. Van Assche N, van Steenberghe D, Guerrero ME, et al. Accuracy
ghe D. State-of-the-art on cone beam CT imaging for preoperative of implant placement based on pre-surgical planning of three-
planning of implant placement. Clin Oral Investig 2006;10:1–7. dimensional cone-beam images: A pilot study. J Clin Periodontol
3. Harris D, Horner K, Gröndahl K, et al. EAO guidelines for the use 2007;34:816–821.
of diagnostic imaging in implant dentistry 2011. A consensus 22. Ruppin J, Popovic A, Strauss M, Spuntrup E, Steiner A, Stoll C. Evalu-
workshop organized by the European Association for Osseointe- ation of the accuracy of three different computer-aided surgery sys-
gration at the Medical University of Warsaw. Clin Oral implants Res tems in dental implantology: Optical tracking vs stereolithographic
2012;23:1243–1253. splint systems. Clin Oral Implants Res 2008;19:709–716.
4. Widmann G, Stoffner R, Schullian P, et al. Comparison of the ac- 23. Dreiseidler T, Neugebauer J, Ritter L, et al. Accuracy of a newly
curacy of invasive and noninvasive registration methods for image- developed integrated system for dental implant planning. Clin Oral
guided oral implant surgery. Int J Oral Maxillofac Implants 2010;25: Implants Res 2009;20:1191–1199.
491–498. 24. Arisan V, Karabuda ZC, Ozdemir T. Accuracy of two stereolitho-
5. Jung RE, Schneider D, Ganeles J, et al. Computer technology appli- graphic guide systems for computer-aided implant placement:
cations in surgical implant dentistry: A systematic review. Int J Oral A computed tomography-based clinical comparative study.
Maxillofac Implants 2009;24(suppl):92–109. J Periodontol 2010;81:43–51.
6. Pettersson A, Kero T, Gillot L, et al. Accuracy of CAD/CAM-guided 25. Viegas VN, Dutra V, Pagnoncelli RM, de Oliveira MG. Transference
surgical template implant surgery on human cadavers: Part I. of virtual planning and planning over biomedical prototypes for
J Prosthet Dent 2010;103:334–342. dental implant placement using guided surgery. Clin Oral Implants
7. Tahmaseb A, van de Weijden JJ, Mercelis P, De Clerck R, Wismeijer D. Res 2010;21:290–295.
Parameters of passive fit using a new technique to mill implant- 26. Cassetta M, Giansanti M, Di Mambro A, Calasso S, Barbato E. Accu-
supported superstructures: An in vitro study of a novel three-di- racy of Two Stereolithographic Surgical Templates: A Retrospective
mensional force measurement-misfit method. Int J Oral Maxillofac Study. Clinical Implant Dent Relat Res 2013 Jun;15:448–459.
Implants 2010;25:247–257. 27. Platzer S, Bertha G, Heschl A, Wegscheider WA, Lorenzoni M.
8. Tahmaseb A, De Clerck R, Eckert S, Wismeijer D. Reference-based Three-Dimensional Accuracy of Guided Implant Placement: Indirect
digital concept to restore partially edentulous patients following Assessment of Clinical Outcomes. Clin Implant Dent Relat Res 2011
an immediate loading protocol: A pilot study. Int J Oral Maxillofac Dec 15 [epub ahead of print].
Implants 2011;26:707–717. 28. Vasak C, Watzak G, Gahleitner A, Strbac G, Schemper M, Zech-
9. Ersoy AE, Turkyilmaz I, Ozan O, McGlumphy EA. Reliability of im- ner W. Computed tomography-based evaluation of template
plant placement with stereolithographic surgical guides generated (NobelGuide)-guided implant positions: A prospective radiological
from computed tomography: Clinical data from 94 implants. study. Clin Oral Implants Res 2011;22:1157–1163.
J Periodontol 2008;79:1339–1345.
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.
Tahmaseb et al
29. Arisan V, Karabuda ZC, Piskin B, Ozdemir T. Conventional Multi- 38. Nikzad S, Azari A. Custom-made radiographic template, computed
Slice Computed Tomography (CT) and Cone-Beam CT (CBCT) for tomography, and computer-assisted flapless surgery for treatment
Computer-Aided Implant Placement. Part II: Reliability of Mucosa- planning in partial edentulous patients: A prospective 12-month
Supported Stereolithographic Guides. Clin Implant Dent Relat Res study. J Oral Maxillofac Surg 2010;68:1353–1359.
2012 Jan 11 [epub ahead of print]. 39. Pomares C. A retrospective study of edentulous patients rehabili-
30. D'Haese J, Van De Velde T, Elaut L, De Bruyn H. A prospective study tated according to the 'all-on-four' or the 'all-on-six' immediate
on the accuracy of mucosally supported stereolithographic surgical function concept using flapless computer-guided implant surgery.
guides in fully edentulous maxillae. Clin Implant Dent Relat Res Eur J Oral Implantol 2010;3:155–163.
2012;14:293–303. 40. Van de Velde T, Sennerby L, De Bruyn H. The clinical and radio-
31. Soares MM, Harari ND, Cardoso ES, Manso MC, Conz MB, Vidigal graphic outcome of implants placed in the posterior maxilla with
GM, Jr. An in vitro model to evaluate the accuracy of guided surgery a guided flapless approach and immediately restored with a provi-
systems. Int J Oral Maxillofac Implants 2012;27:824–831. sional rehabilitation: A randomized clinical trial. Clin Oral Implants
32. Sanna AM, Molly L, van Steenberghe D. Immediately loaded CAD- Res 2010;21:1223–1233.
CAM manufactured fixed complete dentures using flapless implant 41. Landázuri-Del Barrio RA, Cosyn J, De Paula WN, De Bruyn H,
placement procedures: A cohort study of consecutive patients. Marcantonio E Jr. A prospective study on implants installed with
J Prosthet Dent 2007;97:331–339. flapless-guided surgery using the all-on-four concept in the man-
33. Balshi SF, Wolfinger GJ, Balshi TJ. Guided implant placement and dible. Clin Oral Implants Res 2013;24:428–433.
immediate prosthesis delivery using traditional Brånemark System 42. Abboud M, Wahl G, Guirado JL, Orentlicher G. Application and suc-
abutments: A pilot study of 23 patients. Implant Dent 2008;17:1 cess of two stereolithographic surgical guide systems for implant
28–135. placement with immediate loading. Int J Oral Maxillofac Implants
34. Johansson B, Friberg B, Nilson H. Digitally planned, immediately 2012;27:634–643.
loaded dental implants with prefabricated prostheses in the recon- 43. Van Assche N, Vercruyssen M, Coucke W, Teughels W, Jacobs R,
struction of edentulous maxillae: A 1-year prospective, multicenter Quirynen M. Accuracy of computer-aided implant placement. Clin
study. Clin Implant Dent Relat Res 2009;11:194–200. Oral Implants Res 2012;23:112–123.
35. Barter S. Computer-aided implant placement in the reconstruc- 44. Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang
tion of a severely resorbed maxilla—A 5-year clinical study. Int J NP. A systematic review of the 5-year survival and complication
Periodontics Restorative Dent 2010;30:627–637. rates of implant-supported single crowns. Clin Oral Implants Res
36. Gillot L, Noharet R, Cannas B. Guided surgery and presurgical pros- 2008;19:119–130.
thesis: Preliminary results of 33 fully edentulous maxillae treated 45. Pjetursson BE, Tan K, Lang NP, Bragger U, Egger M, Zwahlen M.
in accordance with the NobelGuide protocol. Clinical Implant Relat A systematic review of the survival and complication rates of fixed
Res 2010;12(suppl 1):e104–e113. partial dentures (FPDs) after an observation period of at least
37. Meloni SM, De Riu G, Pisano M, Cattina G, Tullio A. Implant treat- 5 years. I. Implant-supported FPDs. Clin Oral Implants Res 2004;
ment software planning and guided flapless surgery with immedi- 15:625–642.
ate provisional prosthesis delivery in the fully edentulous maxilla.
A retrospective analysis of 15 consecutively treated patients. Eur J
Oral Implantol 2010;3:245–251.
© 2014 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY.
NO PART MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER.