You are on page 1of 5

Microchemical Journal 146 (2019) 645–649

Contents lists available at ScienceDirect

Microchemical Journal
journal homepage: www.elsevier.com/locate/microc

Feasibility of pyrohydrolysis as a clean method for further fluorine T


determination by ISE and IC in high purity nuclear grade alumina
Lucelia Hoehnea, Rochele S. Picolotob, Michele S.P. Endersb, Gabriel T. Druzianb,
Edson I. Mullerb, Erico M.M. Floresb,

a
Universidade do Vale do Taquari – Univates, 95914-014 Lajeado, RS, Brazil
b
Departamento de Química, Universidade Federal de Santa Maria, 97105-900 Santa Maria, RS, Brazil

ARTICLE INFO ABSTRACT

Keywords: In this work, a method based on pyrohydrolysis of high purity nuclear grade alumina and subsequent F de-
Nuclear grade alumina termination by ion chromatography (IC) and ion selective electrode (ISE) was developed. After pyrohydrolysis, F
Fluorine determination was absorbed in a suitable solution (50 mmol L−1 (NH4)2CO3, 50 mmol L−1 NH4OH, and also water were
Pyrohydrolysis evaluated). Using the selected conditions, it was possible to digest a high sample mass (500 mg) in only 10 min.
Ion chromatography
The following analytical parameters were evaluated: pyrohydrolysis temperature (550 to 1050 °C), time of re-
Ion selective electrode
action (2.5 to 20 min), air flow-rate (0.1 to 0.4 L min−1), water flow-rate (0.1 to 1.0 mL min−1), and the use of
Fluorine in alumina
V2O5 (200 mg), as an auxiliary reagent. Quantitative F recoveries were obtained using air and water flow-rates
pumped through the reactor at 0.2 L min-1 and 0.5 mL min−1, respectively. Only water was required as absorbing
solution and the use of V2O5, generally recommended for inorganic matrices, was not necessary. Detection limits
of 2.0 and 3.6 μg g−1 using IC and ISE, respectively, were obtained. Accuracy was evaluated by recovery tests
and also by comparison with the results obtained by high resolution continuum source graphite furnace mole-
cular absorption spectrometry (HR-CS GF MAS). Suitable recoveries for F were obtained (95%) and no statistical
difference (95% of confidence level) was observed for the results obtained by HR-CS GF MAS. Pyrohydrolysis
allowed the F volatilization in a relatively short time avoiding the use of concentrated reagents and making
possible further determination by both analytical techniques (IC or ISE). The pyrohydrolysis method was con-
sidered suitable for further F determination, avoiding excessive dilution of digests or matrix matching ap-
proaches, making this method a suitable option for routine F determination in nuclear grade alumina samples.

1. Introduction Fluorine determination in alumina can be performed by direct


analysis using prompt-gamma activation analysis (PGAA) or high-re-
Alumina is considered one of the most important oxide ceramics due to solution continuum source graphite furnace molecular absorption
its application in several fields, such as structural, electrical, optical and spectrometry (HR-CS GF MAS). Although HR-CS GF MAS can be con-
nuclear technologies [1]. In view of its high resistance, outstanding che- sidered a suitable alternative, it requires specific reagents for molecule
mical stability, excellent wear and corrosion resistance, and distinguished formation (such as AlF, CaF, etc) in addition to sample mass being
mechanical properties even under radiation and high-temperature en- generally limited to < 1 mg [4–7]. Regarding the F determination by
vironments, alumina plays an important role for the nuclear industry [1]. PGAA, the most evident drawbacks include the difficulties of access to
The use of alumina as a nanoceramic is of great concern because it is cited nuclear reactors, lack of sensitivity, and relatively high detection limits
as the most promising material class for nuclear environments due to its (around 18 μg g−1) [8].
tolerance to radiation [2]. However, F in alumina matrices may become a Ion chromatography (IC) and ion selective electrode (ISE) techni-
corrosive species in fission reactors due to thermolysis and radiolysis re- ques have been routinely used for halogen determination, especially for
actions [2]. According to the requirements of the American Society for F [9–11]. These techniques are accessible for most of laboratories and
Testing and Materials (ASTM C method 1031-11 - Standard Specification present relatively low cost when compared to PGAA or HR-CS GF MAS.
for Nuclear Grade Aluminum Oxide Powder) [3], the maximum F content However, both IC and ISE require analytical solutions (or digests) with
allowed for this material is 50 μg g−1. low carbon content and suitable pH in order to avoid interferences [12].


Corresponding author.
E-mail address: ericommf@gmail.com (E.M.M. Flores).

https://doi.org/10.1016/j.microc.2019.01.053
Received 26 December 2018; Received in revised form 18 January 2019; Accepted 19 January 2019
Available online 22 January 2019
0026-265X/ © 2019 Elsevier B.V. All rights reserved.
L. Hoehne et al. Microchemical Journal 146 (2019) 645–649

In spite of the efficiency of methods based on closed-vessels mi- pyrohydrolysis unit and the air flow was controlled by a conventional
crowave-assisted acid digestion for several materials, it can be con- flow meter. The pyrohydrolysis unit is heated by an electrothermal
sidered as not suitable for further halogen determination due to the risk furnace (Sanchis, Porto Alegre, Brazil) with an automatic temperature
of analyte losses [8,13]. On the other hand, pyrohydrolysis has been control (maximum temperature of 1200 °C). A glass serpentine, im-
used as an alternative sample preparation method for the decomposi- mersed into an ice bath, was connected to the outlet of the quartz re-
tion of organic and inorganic materials for further halogen determi- actor to condense the gaseous products from pyrohydrolysis reaction,
nation [9,11,14–22]. Using this method, halogens are volatilized from avoiding losses of HF. The condensed solution was collected in a
matrix under heating using water vapor flow causing hydrolysis reac- polypropylene vessel containing 10 mL of a suitable absorbing solution
tions that generate the respective volatile acids (i.e., HF, HCl, HBr and for the subsequent analyte determination.
HI) [9,11,14–22]. These species can be separated, condensed and col-
lected in diluted absorbing solution such as NH4OH [23] and 2.2. Instrumentation
(NH4)2CO3 [24]. An important advantage is that absorbing solutions
are generally compatible with F determination by both ISE and IC. Fluorine determination in absorbing solutions obtained after pyr-
Another advantage is that analytes are separated from the matrix then ohydrolysis method was carried out using an ion chromatography
reducing interferences that could impair the determination by ISE and system (819 model, IC detector, Switzerland). The IC system consists of
IC [9,11,14–22]. a pump (model 833 IC liquid handling unit, Metrohm) and a con-
Based on the potential of pyrohydrolysis for inorganic materials, in ductivity detector. A column (model Metrosep A Supp 5, 150 × 4 mm
this work this sample preparation method was investigated for the first i.d., Metrohm) with packing material composed by polyvinyl alcohol
time for further F determination in nuclear grade alumina by IC and ISE with quaternary ammonium groups and a particle size of 5 μm and a
with the focus on evaluating the possibility of using well diluted solu- guard column (model Metrosep A Supp 4/5 Guard), packed with the
tion or even water for F absorption and avoiding the use of V2O5 as aid same material and particle size of the analytical column, were used. An
for F volatilization. Temperature of pyrohydrolysis, air and water flow- anion self-regeneration suppressor (model 833 suppressor unit,
rates, type and concentration of absorbing solution, use of auxiliary Metrohm) was used to reduce the conductivity of mobile phase. A so-
reagent (V2O5), and time of pyrohydrolysis reaction were investigated. lution containing 3.2 mmol L−1 Na2CO3 and 1 mmol L−1 NaHCO3 was
Accuracy was evaluated through recovery tests and also by comparison used as mobile phase and 0.7 mL min−1 and 100 μL, were used as flow-
with the values obtained by HR-CS GF MAS. rate and sample loop, respectively, according to the manufacturer re-
commendation [25]. Fluorine was also determined using a potenti-
2. Experimental ometer (781 model, Metrohm, Switzerland) equipped with a fluoride
ISE electrode (Part number: 80502.150, Metrohm) in solutions obtained
2.1. Pyrohydrolysis system after pyrohydrolysis.
For comparison of results obtained using the proposed pyrohydrolysis
A homemade pyrohydrolysis apparatus (Fig. 1) was used for alu- method, a high-resolution continuum source graphite furnace molecular
mina samples. This system is composed of a stream generator unit, a absorption spectrometer, model ContrAA 700 (Analytik Jena, Germany)
heater furnace, a pyrohydrolysis unit consisting of a tube and a boat, equipped with a manual solid sampling system model SSA 600 (Analytik
both of quartz, a condenser, a collector for the fluoride distillate. The Jena) for direct analysis of nuclear grade alumina was used. The optical
steam generator is a borosilicate Erlenmeyer-type flask containing system comprises a xenon short-arc lamp (XBO 301, GLE, Germany)
water, which was heated and connected to a flow meter and to a water operating in “hot-spot” mode as the radiation source, which emits a high-
trap via two polytetrafluoroethylene (PTFE). This water trap was used intensity continuum spectrum within the wavelength range from 180 to
to avoid the water droplets introduction to the pyrohydrolysis unit and 900 nm. The spectrometer is equipped with a prism as a pre mono-
it was made of glass and adapted to the quartz tube of the pyrohy- chromator and an echelle grade as monochromator to provide the high
drolysis unit by using a silicone stopper. A pump (Minipuls, Gilson, resolution required for HR-CS GF MAS analysis. Measurements were
Middleton, USA) was used in order to pass the water steam through the carried out at the 274.461 nm rotational molecular absorption line of AlF
with 3 pixels (central pixel ± 1). Transversely heated pyrolytically
G coated graphite tubes using platforms specially designed for solid sam-
F H pling (Analytik Jena) were used throughout. An ultra-micro balance
(Sartorius, Moldel M2P, Germany) with a resolution of 1 μg was used for
sample weighing before F determination by HR-CS GF MAS.

C 2.3. Reagents and samples

All chemicals used were of analytical grade. Solutions were pre-


E M I pared using distilled/deionized water that was further purified using a
Milli-Q (18.2 MΩ cm, Millipore, Billerica, MA). Vanadium pentoxide
free of F, generally used as accelerator auxiliary during pyrohydrolysis,
D was obtained by heating of NH4VO3 solid at 550o C in a platinum
crucible during 4 h [21]. For F determination by HR-CS GF MAS, Ba2+
B M and Al3+ solutions were prepared by serial dilutions of 1000 mg L−1 Ba
L (NO3)2 and Al(NO3)3 stock solutions, respectively. For F determination
A J by IC, Na2CO3 and NaHCO3 salts were used to prepare the mobile
phase. Glass materials or quartz apparatus used in this work were
soaked in 10% (m/v) HNO3 solution for 24 h and washed with water
Fig. 1. Pyrohydrolysis system apparatus used in this work: A) compressed air before of the used. Absorbing solutions, used for collecting the analyte,
inlet, B) flowmeter tube, C) polytetrafluoroethylene connectors, D) water vapor were prepared from NH4OH (25%) and (NH4)2CO3 salt and the con-
generation system, E) bottle to prevent backflow of water, F) quartz reactor, G) centration used for both solutions was 50 mmol L−1.
quartz boat, H) furnace, I) condenser, J) flask containing the absorbing solu- Stock standard solution of F (1000 mg L−1) was prepared by dis-
tion, L) temperature control system, and M) temperature controller of oven. solving NaF salt in water. Fluorine calibration curve in the range of 10

646
L. Hoehne et al. Microchemical Journal 146 (2019) 645–649

to 1000 μg L−1 for IC and 250 to 2500 μg L−1 for ISE were prepared by solutions presented no statistical difference (ANOVA, 95% of con-
sequential dilution of stock solution (1000 mg L−1) in water. Fluorine fidence level) with reference value (obtained by HR-CS GF MAS). The
determination by ISE was performed in presence of buffer TISAB so- use of both water and alkaline absorbing solutions allowed relative
lution. For the preparation of TISAB solution, sodium chloride (58 g), standard deviations (RSDs) lower than 7%. Therefore, water was se-
1,2-cyclohexilenediaminetetraacetic acid (4 g), and acetic acid (57 mL) lected as absorbing solution for subsequent experiments. This result was
were dissolved in water (500 mL) [11]. After, the pH was adjusted in agreement with a previous work showing that, depending on the
between 5.2 and 5.5 using a NaOH solution (20% m/v). Finally, the adjusted conditions, only water could be suitable for the quantitative
final volume was completed to 1000 mL with water [11]. Three nuclear absorption of F [9]. It is important to mention that blank values ob-
grade alumina powder samples (purity higher than 99.5%) were do- tained after pyrohydrolysis were always negligible.
nated by a supplier of nuclear grade materials (samples A, B, and C). In general, some auxiliary reagents have been used in pyrohy-
Particle size distribution of samples ranged from 53 to 210 μm. drolysis to help halogen release mainly from inorganic matrices. One of
the most used reagents for this purpose is V2O5 [14,15]. This reagent
2.4. Pyrohydrolysis method has been used due to the relatively low melting point, about 750 °C, and
suitability for several matrices [21,22,27]. Moreover, this reagent can
For the proposed method, 500 mg of alumina, previously mixed be produced in high purity from ammonium metavanadate, resulting in
with 200 mg accelerator auxiliary (V2O5, when it was used), were lower blank values, which is an important aspect to achieve low de-
weighted in a quartz boat. It was introduced into a quartz reactor, tection limits. Then, the use of V2O5 (100 to 500 mg) as aid for the
which was placed inside a furnace with a maximum temperature of volatilization of F during the hydrolysis reaction at high temperature
1050 °C. Experiments were performed using 10 min as reaction time, was evaluated. The experimental conditions used for this evaluation
temperature set at 950 °C, air flow-rate of 0.2 L min−1, and water flow- were: 10 mL of water (used as absorbing solution), sample mass of
rate pumped through the reactor set at 0.5 mL min−1. Volatilized F 500 mg, temperature of 1000 °C during 10 min, and air flow-rate of
after the pyrohydrolysis reaction was condensed and collected in a 0.2 L min−1 (F determination was performed by ISE). As no differences
vessel containing 5 mL of an absorption solution (water, 50 mmol L−1 were observed when pyrohydrolysis was carried out using or not V2O5
NH4OH, and 50 mmol L−1 (NH4)2CO3 were evaluated). After pyrohy- (results not shown), subsequent experiments were performed without
drolysis reaction, solutions were transferred to volumetric vessels, di- using V2O5.
luted with water to 25 mL and the determination of F was performed by
IC and also by ISE. 3.2. Effect of experimental conditions

2.5. Fluorine determination in alumina by HR-CS GF MAS 3.2.1. Influence of temperature and reaction time of pyrohydrolysis
In order to assure the complete F volatilization from matrix, ex-
The formation of AlF molecule in the presence of Ba allowed the F periments were performed to evaluate the effect of temperature on the
determination by HR-CS GF MAS. Previously to the determination of F proposed pyrohydrolysis method. In this way, temperatures from 550 to
by HR-CS GF MAS, graphite platforms were covered with 2.5 μg of Al 1050 °C were evaluated and the following experimental conditions were
and 10 μg of Ba by the addition of aliquots of the respective solutions of set: 10 mL of water as absorbing solution, sample mass of 500 mg, re-
Al3+ and Ba2+. Platform heating inside of graphite furnace and cooling action time of 10 min, and air flow-rate of 0.2 L min−1. According to
down before analysis of alumina samples were performed according to literature, pyrohydrolysis reaction is thermodynamically favorable at
a method described in literature [26]. After, alumina samples were temperatures higher than 500 °C [14]. However, it has already been
weighed (masses up to 0.3 mg) directly on the graphite platform. Sub- observed that F could be quantitatively released from inorganic com-
sequently, platform was inserted into the graphite furnace and the pounds only at temperatures higher than 700 °C. These results are
heating program was applied as previously described [26]. Calibration shown in Table 1.
curve (1 to 25 μg L−1) was obtained by direct pipetting aliquots of F As can be seen in Table 1, when temperatures up to 850 °C were
standard solution on the graphite platform and subsequent applying the evaluated, the F concentration was lower than 45 μg g−1 (agreement
same conditions and heating program used for powder alumina sam- around of 43% with the reference value). This result could be attributed
ples. to the non-quantitative release of analyte from alumina due to the in-
sufficient temperature. On the other hand, when temperatures of 950
3. Results and discussion and 1050 °C were evaluated, the F concentration was 108 and
107 μg g−1, respectively (agreement close to 100% with the reference
Initially, F determination in alumina sample “A” was performed by value). A relative standard deviation of 6% was obtained for tempera-
HR-CS GF MAS [26] providing a concentration for F of 109 ± 2 μg g−1 tures of 950 °C, but using 1050 °C the RSD was of 8%. In this way, a
(n = 5). This value was used as reference for further evaluations.
Table 1
3.1. Pyrohydrolysis method Influence of temperature and reaction time on F volatilization from alumina
(n = 3).
It is well known that the choice of absorbing solution is a key
Parameter F, μg g−1
parameter for developing pyrohydrolysis methods, since it has a strong
influence in the analyte recovery [12]. Particularly in case of halogen Temperature, °C 550 5.22 ± 1.51
determination the absorbing solution should be carefully optimized due 650 7.34 ± 2.81
to the risk of analyte losses. Moreover, it must be compatible with the 750 20.2 ± 4.2
850 40.3 ± 10.1
determination technique and also to be non-toxic and environmental 950 108 ± 6
friendly as for as possible. In this sense, 50 mmol L−1 (NH4)2CO3, 1050 107 ± 8
50 mmol L−1 NH4OH, and also water were evaluated as absorbing so- Time, min 2.5 20.2 ± 8.0
lutions and F determination was determined by ISE. The evaluation of 5.0 72.1 ± 21.6
10 109 ± 5
parameters was performed using alumina sample A. The follow ex-
15 109 ± 5
perimental conditions were used: volume of absorbing solution of 20 107 ± 5
10 mL, sample mass of 500 mg, temperature of 1000 °C during 10 min,
and air flow-rate of 0.2 L min−1. Obtained results using all evaluated Reference value for F obtained by HR-CS GF MAS was 109 ± 2 μg g−1.

647
L. Hoehne et al. Microchemical Journal 146 (2019) 645–649

Table 2 respectively (agreement close to 100% to reference value). In this way,


Influence of air flow-rate on F determination by ISE after air flow-rate was set at 0.2 L min−1 that was considered as enough for
pyrohydrolysis (n = 3, mean ± standard deviation). the quantitative volatilization and transport of F using the proposed
Air flow-rate, L min−1 F, μg g−1 pyrohydrolysis method and determination by ISE.

0.1 58.0 ± 3.4 3.3. Fluorine determination in alumina by IC and ISE after pyrohydrolysis
0.2 109 ± 4
0.3 108 ± 4
0.4 110 ± 5 Three nuclear grade alumina samples were analyzed using the op-
timized pyrohydrolysis conditions and F determination was carried out
Reference value for F obtained by HR-CS GF MAS was by IC and also ISE (Table 3). Experiments were performed using 500 mg
109 ± 2 μg g−1 of alumina samples, 10 mL of water (used as absorbing solution),
10 min of pyrohydrolysis reaction at 950 °C, water flow-rate of
Table 3 0.5 mL min−1, and air flow-rate of 0.2 L min−1.
Concentration of F (μg g−1) in nuclear grade alumina determined by IC and ISE According to the results showed in Table 3, the concentration of F in
after decomposition using pyrohydrolysis method (n = 3, mean ± standard two of the nuclear grade alumina samples (samples A and B) was higher
deviation).
than the maximum value allowed (50 μg g−1) for this kind of material
Sample Pyrohydrolysis method HR-CS GF MAS (provided by ASTM C1031-11). On the other hand, only sample C was
considered as suitable for nuclear applications (F concentration lower
IC ISE
than 0.5 μg g−1 by IC and 5.1 μg g−1 by ISE). However, it is important
A 107 ± 4 108 ± 5 109 ± 2 to highlight that F concentration obtained by ISE and IC after pyrohy-
B 101 ± 5 95 ± 4 96 ± 6 drolysis for samples A and B were in agreement with reference values
C < 0.5 < 5.0 < 1.5 (obtained by HR-CS GF MAS). In this sense, the proposed method can be
considered as accurate (significant difference was not observed by
comparison of pyrohydrolysis and HR-CS GF MAS results). Moreover,
temperature of 950 °C was selected as suitable temperature for F vola- the LODs achieved by pyrohydrolysis method combined to IC
tilization from the matrix. (2.0 μg g−1) or ISE (3.6 μg g−1) were suitable for the quality control of
The time of reaction is another important parameter that also alumina even better than those provided by HR-CS GF MAS, probably
should be evaluated. In general, the time of pyrohydrolysis reaction due the higher sample mass used in pyrohydrolysis system. It is im-
must be sufficient for complete F release and transport of the volatile portant to mention that there is no certified reference material for
analyte species to the absorbing solution. In this sense, the time of re- alumina with F information.
action in the proposed method was evaluated in the range of 2.5 to
20 min and the following experimental conditions were set: 10 mL of 4. Conclusion
water as absorbing solution, sample mass of 500 mg, pyrohydrolysis
temperature of 950 °C, and air flow-rate of 0.2 L min−1. The results are The pyrohydrolysis method was successfully applied to nuclear
shown in the Table 1. grade alumina for subsequent F determination by IC and ISE. The
As can be seen in Table 1, up to 5 min of pyrohydrolysis reaction proposed method provided a suitable analytical tool for quality control
were not enough for the complete F volatilization (agreement about in compliance with ASTM C1031-11. As an important advantage of the
67%). However, quantitative recoveries of F were reached after 10 min proposed method, only water was considered a suitable absorbing so-
of heating, at 950 °C. No significant difference (ANOVA, 95% of con- lution and the analyte separation from the matrix prevented inter-
fidence level) was observed by using 10 to 20 min of pyrohydrolysis ferences in IC and ISE determination. Moreover, final solution was
time. The RSD values were lower than 5% (n = 3) if the time given for suitable to be analyzed directly by IC and ISE without further dilution.
pyrohydrolysis was at least 10 min. Therefore, this time was established Statistical difference was not observed between results obtained by
for further experiments. pyrohydrolysis (F determination by IC and ISE) and HR-CS GF MAS.
Finally, due to the simplicity, low sample preparation time and as the
3.2.2. Influence of water and air flow-rates on pyrohydrolysis possibility of the use water as absorbing solutions, the pyrohydrolysis
The evaluation of the water flow-rate pumped through the reactor is method, it can be suggested for the quality control of alumina used for
important in order to provide enough water for the pyrohydrolysis re- nuclear applications.
action. In this work, this parameter was evaluated from 0.10 to
1.0 mL min−1 by using water flow-rate below or higher than Acknowledgements
0.5 mL min−1 the RSDs were higher (up to 10%). However, using a
water flow-rate of 0.5 mL min−1 the RSD was 5% and this flow-rate was The authors are grateful to CNPq and FAPERGS for supporting this
selected for further experiments allowing minimum decrease on reac- study. E.I. Muller and E. M. M. Flores have research scholarship from
tion temperature and providing a suitable F recovery (higher than CNPq, grant numbers: 304336/2015-7 and 309297-2016/8, respec-
95%). tively. Moreover, this study was financed in part by the Coordenação de
The air flow-rate is another important parameter that must be Aperfeiçoamento de Pessoal de Nível Superior - Brazil (CAPES) -
evaluated to ensure a complete transport of volatile pyrohydrolysis Finance Code 001.
products from the quartz tube reactor to the absorbing solution flask. In
the present work, experiments were carried out to evaluate the effect References
air flow-rate on F release. In this sense, alumina sample (500 mg) was
heated at 950 °C and experiments were performed using air flow-rates [1] R.J. Brook, Concise Encyclopedia of Advanced Ceramic Materials, Elsevier Science,
ranging from 0.1 to 0.4 L min−1. The results are shown in Table 2. England, UK, 2012.
[2] F.G. Ferré, A. Mairov, L. Ceseracciu, Y. Serruys, P. Trocellier, C. Baumier,
It is possible to observe (Table 2) that an air flow-rate of 0.1 L min−1 O. Kaïtasov, R. Brescia, D. Gastaldi, P. Vena, M.G. Beghi, L. Beck, K. Sridharan, F. Di
was not enough to obtain quantitative results to F (F = 58 μg g−1, Fonzo, Radiation endurance in Al2O3 nanoceramics, Sci. Rep. 6 (2016) 1–9.
agreement around 74% with reference value). On the other hand, when [3] ASTM C1031-11, Standard Specification for Nuclear-Grade Aluminum Oxide
Powder, ASTM International, West Conshohocken, PA, 2016www.astm.org.
air flow-rates of 0.2, 0.3, and 0.4 L min−1 were evaluated, the con- [4] A.R. Borges, L.L. François, B. Welz, E. Carasek, M.G.R. Vale, Determination of
centration of F was 109 ± 5, 108 ± 4, and 110 ± 5 μg g−1,

648
L. Hoehne et al. Microchemical Journal 146 (2019) 645–649

fluorine in plant materials via calcium mono-fluoride using high-resolution graphite [15] A.C.D. Newman, A simple apparatus for separating fluorine from aluminosilicates
furnace molecular absorption spectrometry with direct solid sample introduction, J. by pyrohydrolysis, Analyst 93 (1968) 827–831.
Anal. At. Spectrom. 29 (2014) 1564–1569. [16] J.G. Surak, D.J. Fisher, C.L. Burros, L.C. Bate, Apparatus for pyrohydrolytic de-
[5] P.M. Machado, S. Morés, E.R. Pereira, B. Welz, E. Carasek, J.B. Andrade, Fluorine termination of fluorides, Anal. Chem. 32 (1960) 117–119.
determination in coal using high-resolution graphite furnace molecular absorption [17] V.L. Dressler, D. Pozebon, E.L.M. Flores, J.N.G. Paniz, E.M.M. Flores,
spectrometry and direct solid sample analysis, Spectrochim. Acta B 105 (2015) Potentiometric determination of fluoride in geological and biological samples fol-
18–24. lowing pyrohydrolytic decomposition, Anal. Chim. Acta 466 (2002) 117–123.
[6] M. Resano, M. Aramendia, M.A. Belarra, High-resolution continuum source gra- [18] R.L. Clements, G.A. Sergeant, P.J. Webb, The determination of fluorine in rocks and
phite furnace atomic absorption spectrometry for direct analysis of solid samples minerals by a pyrohydrolytic method, Analyst 96 (1971) 51–54.
and complex materials: a tutorial review, J. Anal. At. Spectrom. 29 (2014) [19] C.D. Susano, J.C. White, J.E. Lee, Apparatus for pyrohydrolytic determination of
2229–2250. fluoride and other halides, Anal. Chem. 27 (1955) 453–454.
[7] H. Cadorim, J. Gois, A.R. Borges, M. Vale, B. Welz, H. Gleisner, C. Ott, [20] T.S. Nunes, C.C. Muller, P. Balestrin, A.L.H. Muller, M.F. Mesko, P.A. Mello,
Determination of fluorine in copper concentrate via high-resolution graphite fur- E.I. Muller, Determination of chlorine and sulfur in high purity flexible graphite
nace molecular absorption spectrometry and direct solid sample analysis – com- using ion chromatography (IC) and inductively coupled plasma optical emission
parison of three target molecules, Talanta 176 (2018) 178–186. spectrometry (ICP OES) after pyrohydrolysis sample preparation, Anal. Methods 7
[8] M.F. Mesko, V.C. Costa, R.S. Picoloto, C.A. Bizzi, P.A. Mello, Halogen determination (2015) 2129–2134.
in food and biological materials using plasma-based techniques: challenges and [21] C.C. Muller, T.S. Nunes, A.L.H. Muller, V.L. Dressler, E.M.M. Flores, E.I. Muller,
trends of sample preparation, J. Anal. At. Spectrom. 31 (2016) 1243–1261. Determination of boron in silicon carbide by ICP-OES and ICP-MS after sample
[9] F.G. Antes, J.S.F. Pereira, L.C. Spadoa, E.I. Muller, E.M.M. Flores, V.L. Dressler, preparation using pyrohydrolysis, J. Braz. Chem. Soc. 26 (2015) 110–116.
Fluoride determination in carbon nanotubes by ion selective electrode, J. Braz. [22] F.G. Antes, J.S.F. Pereira, M.S.P. Enders, C.M.M. Moreira, E.I. Muller,
Chem. Soc. 23 (2012) 1193–1198. E.M.M. Flores, V.L. Dressler, Pyrohydrolysis of carbon nanotubes for Br and I de-
[10] J.S.F. Pereira, F.G. Antes, L.O. Diehl, C.L. Knorr, S.R. Mortari, V.L. Dressler, termination by ICP-MS, Microchem. J. 101 (2012) 54–58.
E.M.M. Flores, Microwave-induced combustion of carbon nanotubes for further [23] L.S.F. Pereira, M.F. Pedrotti, P. Dalla Vecchia, J.S.F. Pereira, E.M.M. Flores, A
halogen determination, J. Anal. At. Spectrom. 25 (2010) 1268–1274. simple and automated sample preparation system for subsequent halogens de-
[11] V.L. Dressler, D. Pozebon, E.L.M. Flores, J.N.G. Paniz, E.M.M. Flores, Determination termination: combustion followed by pyrohydrolysis, Anal. Chim. Acta 1010 (2018)
of fluoride in coal using pyrohydrolysis for analyte separation, J. Braz. Chem. Soc. 29–36.
14 (2003) 334–338. [24] E.M.M. Flores, M.F. Mesko, D.P. Moraes, J.S.F. Pereira, P.A. Mello, J.S. Barin,
[12] P.A. Mello, J.S. Barin, F.A. Duarte, C.A. Bizzi, L.O. Diehl, E.I. Muller, E.M.M. Flores, G. Knapp, Determination of halogens in coal after digestion using the microwave-
Analytical methods for the determination of halogens in bioanalytical sciences: a induced combustion technique, Anal. Chem. 80 (2008) 1865–1870.
review, Anal. Bioanal. Chem. 405 (2013) 7615–7642. [25] Metrohm, IC Application Work: AW DE 8-0514-082002.
[13] A.L.H. Muller, C.A. Bizzi, J.S.F. Pereira, M.F. Mesko, D.P. Moraes, E.M.M. Flores, [26] E.L.M. Flores, J.S. Barin, E.M.M. Flores, V.L. Dressler, A new approach for fluorine
E.I. Muller, Bromine and chlorine determination in cigarette tobacco using micro- determination by solid sampling graphite furnace molecular absorption spectro-
wave-induced combustion and inductively coupled plasma optical emission spec- metry, Spectrochim. Acta B 62 (2007) 918–923.
trometry, J. Braz. Chem. Soc. 22 (2011) 1649–1655. [27] R.S. Picoloto, S.M. Cruz, P.A. Mello, E.I. Muller, P. Smichowski, E.M.M. Flores,
[14] J.C. Warf, W.D. Cline, R.D. Tevebaugh, Pyrohydrolysis in determination of fluoride Combining pyrohydrolysis and ICP-MS for bromine and iodine determination in
and other halides, Anal. Chem. 26 (1954) 342–346. airborne particulate matter, Microchem. J. 116 (2014) 225–229.

649

You might also like