You are on page 1of 7

Sensors and Actuators B 104 (2005) 68–74

Amperometric glucose biosensor based on immobilization


of glucose oxidase in electropolymerized o-aminophenol
film at copper-modified gold electrode
Dawei Pan, Jinhua Chen∗ , Shouzhuo Yao, Lihua Nie, Jianjun Xia, Wenyan Tao
State Key Laboratory of Chemo/Biosensing and Chemometrics, College of Chemistry and Chemical Engineering,
Hunan University, Changsha 410082, PR China
Received 3 January 2004; received in revised form 27 April 2004; accepted 27 April 2004
Available online 7 June 2004

Abstract

Amperometric glucose biosensor is developed, based on immobilization of glucose oxidase (GOD), in an electrochemically polymerized
non-conducting poly(o-aminophenol) (POAP) film at copper (Cu)-modified gold (Au) electrode. The rough surface and the ability to
electrochemically oxidize glucose of Cu nanoparticles result in the improvement of the detection limit and the increase of the maximum
response current and sensitivity. The biosensor based on Au/Cu/POAP/GOD electrode has two times lower detection limit, three times
larger maximum current and 2.5 times higher sensitivity than those of the biosensor based on Au/POAP/GOD electrode. Additionally, the
fast response time, large response current and good selectivity for ascorbic acid, uric acid and acetaminophen can also be obtained. On the
other hand, effects of electrochemical deposition time of Cu, applied potential used in the determination and electroactive compounds on
the amperometric response of the enzymatic sensor were investigated and discussed. Excellent reproducibility and stability of biosensor
are also observed.
© 2004 Elsevier B.V. All rights reserved.

Keywords: Glucose oxidase; Poly(o-aminophenol); Copper; Electrochemical polymerization; Biosensor

1. Introduction enzyme-catalysed oxidation of glucose by dissolved oxy-


gen [7]. From the viewpoint of chemical sensors, hydrogen
The development of novel biosensor for the fast, sensitive peroxide can be detected by oxidation at anodic potentials
and selective quantification of analytes of biomedical inter- (>+0.6 V versus Ag/AgCl) [8,9]. But at this relatively high
est has received considerable attention, and electrochemical potential, there may be interferences from other oxidasable
biosensors have been successfully used for quantifying species such as ascorbic acid (AA), uric acid (UA) and
these analytes [1]. Glucose is a key metabolite for living acetaminophen (AP). It results in difficulty for the quanti-
organisms, especially in the case of patients suffering di- tative analysis of the produced hydrogen peroxide and the
abetes. In general, glucose oxidase (GOD) is selected as a enzyme substrate.
model enzyme, since it is well-studied, inexpensive, stable One approach to avoid interferents developed in recent
and practically applied [2–5]. The final goal of glucose elec- decade years is related to the use of selective membranes
trochemical biosensor lies in designing high-performance that are permeable for hydrogen peroxide but non-permeable
sensors with appropriate characteristics such as sensitivity, for interference species [10]. Many of the known selective
selectivity, response time, linearity, stability and repro- membranes, such as cellulose acetate membrane, do not en-
ducibility [6]. sure a satisfactory avoidance of interferents. Therefore, seri-
Most of the electrochemical amperometric biosensors ous research activities should be taken in creating improved
for glucose are based on the electrochemical oxidation of selective membranes for selectivity biosensors. In general,
hydrogen peroxide, which is formed in the course of the the selectivity of the polymer depends on its pore size and
charges [11,12]. The non-conducting polymers, due to their
good selectivity and fast response, have attracted much inter-
∗ Corresponding author. Tel.: +86 731 8821818; fax: +86 731 8821818. est in the development of the biosensors. The glucose biosen-
E-mail address: chenjinhua@hnu.cn (J. Chen). sors based on the non-conducting poly(o-phenylenediamine)

0925-4005/$ – see front matter © 2004 Elsevier B.V. All rights reserved.
doi:10.1016/j.snb.2004.04.100
D. Pan et al. / Sensors and Actuators B 104 (2005) 68–74 69

[13], poly(ethacridine) [14], overoxidized polyprrole [15], ric measurements were carried out under stirred condition
poly(phenol) [16] and poly(o-aminophenol) (POAP) [17] and the response current was marked with the change value
films have been reported and showed the good selectivity. between the steady-state current and background current.
Although non-conducting polymer films exhibit excellent All experiments were performed at room temperature. The
selectivity properties, their use is still a matter of concern micrographs and element composition of Au/Cu electrodes
because of their relative high detection limit and the low re- were investigated by scanning electron microscopy (SEM,
sponse current. To increase the response current, ferrocene, JSM 5600 LV, operating at 20 kV) and energy-dispersive
a satisfactory electron transfer mediator for the amperomet- X-ray spectroscopy (EDS, Vantage 4105, NORAN), respec-
ric hydrogen peroxide sensor, has been used in the biosensor tively.
based on non-conducting polymer films of phenol and its
derivatives and an increasing response current of hydrogen 2.2. Electrochemical deposition of Cu and preparation of
peroxide can be observed [18,19]. Au/Cu/POAP/GOD electrode
In this paper, an amperometric glucose biosensor based
on GOD immobilized in electrochemical polymerized POAP Prior to electrochemical deposition of Cu, the Au work-
film at Cu-modified gold (Au) electrode has been developed. ing electrode was cut to get bare fresh surface, rinsed with
o-Aminophenol (o-AP) can be electropolymerized to form double-distilled water and ethanol, and then followed by
electroinactive film when the pH of the solution is over 3 cleaning in an ultrasonic bath. The electrode was then ac-
[20]. POAP film, a non-conducting film which acted as an tivated by continuously scanning the potential between 0.0
effective barrier to protect the electrode from fouling, was and 1.5 V at 0.05 V s−1 in 1 M H2 SO4 solution and washed
successfully used in the development of hydrogen peroxide, by double-distilled water. Cu was electrochemically de-
uric and glucose biosensor [21–23]. Additionally, copper is posited by potentiostatic method in a 0.2 M acetate buffer
selected in this paper in order to increase the response current solution (pH 4.0) containing 5.0 mM CuCl2 .
because it can electrochemically oxidize glucose [24] and is The Cu-modified Au electrode was placed in 0.2 M deaer-
relatively inexpensive, commercially available in a number ated acetate buffer solution (pH 5.0) containing 200 U/ml
of forms and the method for its deposition is well established GOD + 5.0 mM o-AP monomer. The electrochemical poly-
[25,26]. The experimental conditions related to the prepara- merization of POAP/GOD film was carried out by cycling
tion and characterizations of the sensor have also been stud- in the potential range from 0.0 to 0.8 V at a scan rate of
ied in detail. The sensor exhibits excellent performances, 50 mV s−1 . For comparison, Au/POAP/GOD electrode was
such as relative low detection limit, short response time, also fabricated at bare Au electrode in the same way. All
large current density and high sensitivity. Additionally, the films used in the following experiments were grown for 15
effect of the interferents (AA, UA and AP) can be decreased cycles except as noted. The resulting enzyme electrodes were
significantly due to selective permeability of POAP film. thoroughly washed by double-distilled water and stored in
phosphate buffer solution (pH 7.0) below 4 ◦ C for future use.

2. Experimental
3. Results and discussion
2.1. Reagents and apparatus
3.1. Preparation and properties of Cu-modified Au
Glucose oxidase (EC 1.1.3.4, Type II from Aspergillus electrode
niger, 50 000 U/g) was purchased from Amresco (USA).
Glucose was obtained from ICN Biomedicals, Inc. (USA). The Cu-modified Au electrode is fabricated by applying
Glucose stock solutions were allowed to mutarotate at room 0.0 V for 180 s in a 0.2 M acetate buffer solution (pH 4.0)
temperature overnight before use. Hydrogen peroxide, o-AP containing 5.0 mM CuCl2 . The micrographs and element
(99%), copper chloride, AA, UA and AP were used as re- composition of Au/Cu electrode have been investigated by
ceived (Chemical Reagent Company of Shanghai, China). SEM and EDS, and the corresponding results are shown in
0.2 M acetate solution and 1/15 M phosphate solution were Fig. 1. From Fig. 1(a), the diameter of Cu particle is about
used as buffer solutions. All other chemicals were analytical 200 nm. Cu particles with spherical shape lie on the electrode
grade. Double-distilled water was used throughout. surface. From Fig. 1(b), EDS result indicates that Au and
All electrochemical experiments were carried out in a con- Cu are the major elements, which implies that the particles
ventional three-electrode cell controlled by CHI 660 Elec- on the Au electrode are Cu.
trochemical Work Station (Chenghua Instrument Company Cyclic voltammograms of Cu-modified Au electrode in
of Shanghai, China). A gold wire (diameter 0.2 mm) encap- 1/15 M phosphate buffer solution (pH 7.0) with and without
sulated in epoxy resin was used as working electrode. A 3 mM hydrogen peroxide are shown in Fig. 2. From Fig. 2,
platinum foil was applied as the counter electrode and a satu- a pair of Cu redox peaks at 0.28 and 0.12 V can be observed
rated calomel electrode (SCE) served as reference electrode. for the case of the absence of hydrogen peroxide. But for
All potential values given below refer to SCE. Amperomet- the case of the addition of hydrogen peroxide, the disap-
70 D. Pan et al. / Sensors and Actuators B 104 (2005) 68–74

Fig. 3. Cyclic voltammograms for the electropolymerization of


POAP/GOD film in a deaerated and unstirred 5.0 mM o-AP + 200 U/ml
GOD + 0.2 mM acetate buffer solutions (pH 5.0) in the potential range
of 0.0–0.8 V at 50 mV s−1 . Numerals indicate the number of cycles.

3.2. Electrochemical co-polymerization of o-AP and GOD


on Cu-modified Au electrode

Fig. 3 presents the cyclic voltammograms for the electro-


chemical co-polymerization of 5 mM o-AP and 200 U/ml of
GOD in acetate buffer solution (pH 5.0) at the Cu-modified
Au electrode. Compared with the electrochemical polymer-
ization of o-AP [17], a reduction peak of copper appears
Fig. 1. The SEM image (a) and EDS pattern (b) of the Au/Cu electrode.
obviously at near 0.2 V at the first cycle. Furthermore,
Magnification 20 000×. with the increase of the polymerization cycles, all the peak
currents decrease. These results imply that non-conducting
POAP/GOD film has been formed on the surface of the
pearance of the copper oxidation peak can be observed. This
Cu-modified Au electrode.
implies that the chemical oxidation of copper by hydrogen
peroxide occurs. This phenomenon is similar to that reported
3.3. Influential factors on the response characteristic of
in literature [27]. Additionally, it can be observed from
the Au/Cu/POAP/GOD electrode
Fig. 2 that the oxidation of hydrogen peroxide is detected
at Cu-modified Au electrode at the potential of +0.4 V or
The amperometric response characteristics of the enzyme
higher.
electrode are affected by the deposition time of copper, ap-
plied potential used in the determination and the electroac-
tive interferents. In this paper, the effects of these factors on
the behavior of the GOD electrode have been investigated
in detail and discussed as follows.

3.3.1. Effect of deposition time of copper


In general, the response of the biosensor should be af-
fected by the loading mass of Cu nanoparticle. The effect
of the different deposition time of copper on the response
current was investigated and the corresponding results are
shown in Fig. 4. From Fig. 4, it can be observed that the re-
sponse current is affected obviously by the loading mass of
Cu. The response current increases with the increase of the
Cu loading and the maximum value can be obtained at the
deposition time of 180 s. This may result from the loading
Fig. 2. Cyclic voltammograms of the Au/Cu electrode in a 1/15 M phos-
phate buffer solution (pH 7.0) with (solid line) and without (dot line)
mass and particle size of Cu nanoparticles. When the depo-
3 mM hydrogen peroxide. The potential range is from −0.4 to 0.8 V and sition time increases, the loading mass of Cu increases and
the scan rate is 50 mV s−1 . the response current of the biosensor increases due to the
D. Pan et al. / Sensors and Actuators B 104 (2005) 68–74 71

Fig. 4. Effect of the deposition time of copper on the steady-state response


current of Au/Cu/POAP/GOD electrode in 1/15 M phosphate buffer solu-
tion (pH 7.0) containing 1 mM glucose (n = 3).

catalytic property of Cu nanoparticle. However, the particle


size of Cu will increase with the increase of the deposition
time and the corresponding catalytic activity of Cu particle Fig. 6. Anodic response to the addition (indicated by the arrows) of
becomes low [28,29]. In the following experiments, 180 s 3 mM ascorbic acid at bare Au (a), Au/Cu (b) and Au/Cu/POAP/GOD (c)
was selected as the Cu deposition time. electrodes in 1/15 M phosphate buffer solution (pH 7.0). Applied potential
is 0.7 V.
3.3.2. Effect of the applied potential
Fig. 5 shows the response current of the enzyme elec-
trode in 1/15 M phosphate buffer solution (pH 7.0) contain- tential is, the easier the electroactive interferents are to be
ing 1 mM glucose as a function of the applied potential, oxidized and cause the extra response current and the easier
which ranges from 0.5 to 0.9 V with 0.05 V increments. the polymer film is to be damaged, we set the potential of
From Fig. 5, when the applied potential is below 0.7 V, 0.7 V for the oxidation operation of the enzyme electrode.
the response current increases rapidly with the increase
of the applied potential. This means that the response of 3.3.3. Effect of the interferents
the enzyme electrode in this potential range (<0.7 V) was Fig. 6 presents the response current to ascorbic acid ob-
controlled by the electrochemical oxidation of hydrogen tained at various electrodes at 0.7 V. From Fig. 6, the large
peroxide [30,31]. When the potential is over 0.7 V, the re- response current of ascorbic acid can be observed obviously
sponse current remains almost same, which is explained by at unmodified Au electrode and Cu-modified Au electrode,
the rate-limiting process of enzyme kinetics and substrate whereas at the enzyme electrode modified by POAP film, the
diffusion [30,31]. Considering that the higher applied po- response current decreases significantly. These results show
that the POAP-modified electrodes can avoid efficiently the
interference of the ascorbic acid. The similar situation can
also be observed for uric acid and acetaminophen. On the
other hand, in discussing the magnitude of the interference, it
is much more relevant to consider the magnitude of the inter-
ferents current relative to the analytical signal produced by
the analyte (glucose). The interference of electroactive com-
pounds to the glucose response was examined in the pres-
ence of their physiological normal level (0.1 mM for ascorbic
acid, 0.5 mM for uric acid and 0.1 mM for actaminophen, re-
spectively) [31] with glucose concentration of 5.6 mM. The
influence of ascorbic acid, uric acid and acetaminophen to
the glucose response was little under the testing conditions.
The ratio of IG+I to IG is 1.02 for ascorbic acid, 1.05 for uric
acid and 1.06 for acetaminophen, respectively. The sensor
Fig. 5. Effect of applied potential on the response current of Au/Cu/POAP/ based on Au/Cu/POAP/GOD electrode has practically good
GOD electrode. Steady-state currents are measured in 1/15 M phosphate selectivity. These results are due to the good selectivity of
buffer solution (pH 7.0) containing 1 mM glucose. POAP film [9,17,21].
72 D. Pan et al. / Sensors and Actuators B 104 (2005) 68–74

Fig. 8. Calibration curve of the response current of GOD biosensor to


Fig. 7. The amperometric response of Au/Cu/POAP/GOD (a) and glucose concentration in 1/15 M phosphate buffer solution (pH 7.0) at the
Au/POAP/GOD (b) electrodes in a stirred phosphate buffer solution (pH Au/Cu/POAP/GOD electrode (a) and the Au/POAP/GOD electrode (b).
7.0) upon the injection of 1 mM glucose in each step (indicated by the Applied potential is 0.7 V.
arrows). Applied potential is 0.7 V.
and the maximum response current of the biosensor based
on the Au/POAP/GOD electrode are calculated and equal
3.4. The comparison of the Au/Cu/POAP/GOD and to 22.8 mM and 0.15 mA cm−2 (0.047 ␮A), respectively.
Au/POAP/GOD electrodes for the amperometric The data of the maximum response current are similar to
determination of glucose that of GOD electrodes modified by other non-conducting
polymer films [14,16]. The sensitivity of biosensor obtained
Fig. 7 illustrates that the typical current–time plots for the from the slope of the initial linear part of the calibration is
Au/Cu/POAP/GOD and Au/POAP/GOD electrodes upon 4.7 mA M−1 cm−2 , which is similar to the reported value
the successive step addition of glucose were obtained by [31]. Compared with Au/POAP/GOD electrode, the appar-
injecting stock glucose solution. The response of these two ent Michaelis–Menten constant of the Au/Cu/POAP/GOD
biosensors to the injection of glucose is very fast (within electrode is also calculated and equal to 17.8 mM. The
5 s) due to the thin POAP film. This is also observed maximum current is 0.45 mA cm−2 (0.14 ␮A), which is
for the other non-conducting polymer-modified electrode three times larger than that at the Au/POAP/GOD elec-
[13,14]. Additionally, the response current of 1 mM glu- trode. The sensitivity of the Au/Cu/POAP/GOD electrode
cose on Au/Cu/POAP/GOD electrode is about 2.5 times is 12.6 mA M−1 cm−2 , which is almost 2.5 times larger
higher than that on the Au/POAP/GOD electrode. The than that of Au/POAP/GOD electrode. These results imply
calibration curves of the response of the Au/POAP/GOD that the biosensor based on Au/Cu/POAP/GOD electrode is
and Au/Cu/POAP/GOD electrodes for different glucose actually valuable and sensitive.
concentration at the applied potential of 0.7 V are pre-
sented in Fig. 8. From Fig. 8, the response current of the 3.5. Reproducibility, stability and sample analysis
biosensor based on Au/POAP/GOD electrode is linear with
glucose concentration up to 10 mM and then a plateau is The reproducibility of five Au/Cu/POAP/GOD electrodes
reached gradually at higher glucose concentration. The de- was estimated by the response to 1 mM glucose at the poten-
tection limit of the biosensor is 0.02 mM. Compared with tial of 0.7 V. The results reveal that the sensor has satisfied
Au/POAP/GOD biosensor, the Au/Cu/POAP/GOD biosen- reproducibility with a mean change of the response current
sor has lower detection limit (0.01 mM) and larger response of 4.1 nA and a relative standard deviation of 9.2%.
current. These may result from the following reasons: (a) The stability of enzymatic biosensor in storage conditions
the surface coverage of enzyme on the Au/Cu/POAP/GOD (phosphate buffer of pH 7.0 at 4 ◦ C) was investigated in the
electrode is higher than that on the Au/POAP/GOD elec- same phosphate buffer solution containing 2.0 mM glucose.
trode due to an increased surface by Cu particle, and (b) The corresponding result shows that 72% response current
the copper particles have the ability to oxidize the glucose. is still retained after 30 days. This result is similar to that of
However, the Au/Cu/POAP/GOD biosensor has narrower POAP-modified GOD electrode [9,17] and implies that the
linear range (up to 6 mM). Au/Cu/POAP/GOD electrode is considerably stable. Good
According to the Lineweaver–Burk form of the Michaelis– stability may be attributed to the enzyme entrapped strongly
Menten equation, the relation between the reciprocal of the in the POAP film that is stable in the neutral medium.
response current and the reciprocal of glucose concentration Human plasma samples were assayed in order to demon-
app
is obtained. The apparent Michaelis–Menten constant KM strate the practical usage of the biosensor. Fresh plasma
D. Pan et al. / Sensors and Actuators B 104 (2005) 68–74 73

Table 1 [4] Y.M. Uand, T.C. Chou, Fabrication of glucose oxidase/polypyrrole


Glucose contents in human blood samples biosensor by galvanostatic method in various pH aqueous solutions,
Biosens. Bioelectron. 19 (2003) 141–147.
Sample Determined by Measured by Bias (mM)
[5] C.W. Chuang, J.S. Shih, Preparation and application of immobilized
number hospital (mM) biosensor (mM)
C60-glucose oxidase enzyme in fullerene C60-coated piezoelectric
1 3.64 3.82 +0.18 quartz crystal glucose sensor, Sens. Actuators B 81 (2001) 1–8.
2 4.27 4.55 +0.28 [6] H. Yang, T.D. Chung, Y.T. Kim, C.A. Choi, C.H. Jun, H.C. Kim,
3 5.80 6.12 +0.32 Glucose sensor using a microfabricated electrode and electropoly-
4 6.30 6.46 +0.16 merized bilayer films, Biosens. Bioelectron. 17 (2002) 251–259.
5 9.07 8.71 −0.36 [7] A.A. Karyakin, E.E. Karyakina, L. Gorton, Prussian-Blue-based am-
perometric biosensors in flow-injection analysis, Talanta 43 (1996)
samples were first analyzed in the local hospital with 1597–1606.
[8] R. Garjonyte, A. Malinauskas, Glucose biosensor based on glu-
ASCA AG-II Chemistry System, Landmark, USA. The
cose oxidase immobilized in electropolymerized polypyrrole and
samples were then re-assayed with the Au/Cu/POAP/GOD poly(o-phenylenediamine) films on a Prussian Blue-modified elec-
electrode. Plasma sample (0.5 ml) was added into 2 ml of trode, Sens. Actuators B 63 (2000) 122–128.
phosphate buffer solution (pH 7.0) and the response was [9] D.W. Pan, J.H. Chen, L.H. Nie, W.Y. Tao, S.Z. Yao, An amperometric
obtained at 0.7 V. The contents of glucose in blood can then glucose biosensor based on poly(o-aminophenol) and Prussian Blue
be calculated from the calibration curve. The results, which films at platinum electrode, Anal. Biochem. 324 (2004) 115–122.
[10] R. Garjonyte, A. Malinauskas, Amperometric glucose biosensor
are listed in Table 1, are satisfactory and agree closely with based on glucose oxidase immobilized in poly(o-phenylenediamine)
those measured by the biochemical analyzer in hospital. layer, Sens. Actuators B 56 (1999) 85–92.
[11] P.N. Bartlett, J.M. Cooper, A review of the immobilization of en-
zymes in electropolymerized films, J. Electroanal. Chem. 362 (1993)
4. Conclusions 1–12.
[12] M. Gerard, A. Chaubey, B.D. Malhotra, Application of conducting
Immobilization of GOD is introduced by the electro- polymers to biosensors, Biosens. Bioelectron. 17 (2002) 345–359.
[13] C. Malitesta, F. Palmisano, L. Torsi, P.G. Zambonin, Glucose
chemical co-polymerization of GOD and o-AP monomer
fast-response amperometric sensor based on glucose oxidase immo-
at the Cu-modified Au electrode in the weak acidic bilized in an electropolymerized poly(o-phenylenediamine), Anal.
medium. Compared with the Au/POAP/GOD sensor, the Chem. 62 (1990) 2735–2740.
Au/Cu/POAP/GOD biosensor has lower detection limit [14] J.J. Xu, H.Y. Chen, Amperometric glucose sensor based on
(0.01 mM), larger maximum current density (0.45 mA cm−2 ) glucose oxidase immobilized in electrochemically generated
and higher sensitivity (12.6 mA M−1 cm−2 ) due to the exis- poly(ethacridine), Anal. Chim. Acta 423 (2000) 101–106.
[15] C.A. Groom, J.H.T. Luong, Improvement of the selectivity of am-
tence of Cu nanoparticles. Moreover, the Au/Cu/POAP/GOD perometric biosensors by using a permselective electropolymerized
electrode also exhibits good selectivity, large response cur- film, Anal. Lett. 26 (1993) 1383–1389.
rent, fast amperometric response, good reproducibility and [16] P.N. Bartlett, D.J. Caruana, Electrochemical immobilization of en-
excellent stability. zymes. Part V∗ . Microelectrodes for the detection of glucose based
on glucose oxidase immobilized in a poly(phenol) film, Analyst 117
(1992) 1287–1292.
[17] Z.N. Zhang, H.Y. Liu, J.Q. Deng, A glucose biosensor based on im-
Acknowledgements
mobilization of glucose oxidase in electropolymerized o-aminophenol
film on platinized glassy carbon electrode, Anal. Chem. 68 (1996)
This project was financially supported by the Scientific 1632–1638.
Research Foundation for the Returned Overseas Chinese [18] Y. Nakabayashi, M. Wakuda, H. Imai, Amperometric glucose sensors
Scholars, State Education Ministry (2001-498); the National fabricated by electrochemical polymerization of phenols on carbon
Natural Science Foundation of China (No. 20275009) and paste electrodes containing ferrocene as electron transfer mediator,
Anal. Sci. 14 (1998) 1069–1076.
the foundation of the Ministry of Science and Technology
[19] Y. Nakabayashi, H. Yoshikawa, Amperometric biosensors for sensing
of China (2001-51). of hydrogen peroxide based on electron transfer between horseradish
peroxide and ferrocene as a mediator, Anal. Sci. 16 (2000) 609–613.
[20] C. Barbero, J.J. Silber, L. Sereno, Formation of a novel electroac-
References tive film by electropolymerization of ortho-aminopheol. Study of its
chemical structure and formation mechanism. Electropolymerization
[1] M.C. Rodriguez, G.A. Rivas, Amperometric glucose biosensor based of analogous compounds, J. Electroanal. Chem. 263 (1989) 333–352.
on the deposition of copper and glucose oxidase onto glassy carbon [21] M.A.V. Garcı́a, P.T. Blanco, A. Ivaska, A poly(o-aminopheol) mod-
transducer, Anal. Lett. 33 (2000) 2373–2388. ified electrode as an amperometric hydrogen peroxide biosensor,
[2] S. Yabuki, F. Mizutani, Y. Satao, Y. Hirata, Immobilization of Electrochim. Acta 43 (1998) 3533–3539.
polyglutamate–glucose oxidase onto a cysteamine-modified gold [22] E. Miland, A.J. Mı́randa Ordieres, P.T. Blanco, Poly(o-aminophenol)-
electrode, Sens. Actuators, B 91 (2003) 187–190. modified bienzyme carbon paste electrode for the detection of uric
[3] K. Ban, T. Ueki, Y. Tamada, T. Saito, S.I. Imabayashi, M. Watanabe, acid, Talanta 43 (1996) 785–796.
Electrical communication between glucose oxidase and electrodes [23] D.W. Pan, J.H. Chen, L.H. Nie, W.Y. Tao, S.Z. Yao, Amperometric
mediated by phenothiazine-labeled poly(ethylene oxide) bonded to glucose biosensor based on immobilization of glucose oxidase in
lysine residues on the enzyme surface, Anal. Chem. 75 (2003) 910– electropolymerized o-aminophenol film at Prussian Blue-modified
917. platinum electrode, Electrochim. Acta 49 (2004) 795–801.
74 D. Pan et al. / Sensors and Actuators B 104 (2005) 68–74

[24] I.G. Casella, M. Gatta, M.R. Guascito, T.R.I. Cataldi, Highly- presently employed as professor in College of Chemistry and Chemical
dispersed copper microparticles on the active gold substrate as an Engineering, Hunan University, PR China. He is mainly engaged in the
amperometric sensor for glucose, Anal. Chim. Acta 357 (1997) 63– research of chemo/biosensors, carbon nanotube and applied electrochem-
71. istry.
[25] G.A. Ragoisha, A.S. Bondarenko, Potentiodynamic electrochemical
impedance spectroscopy. Copper underpotential deposition on gold, Dawei Pan obtained Bachelor and Master’s degree from College of Chem-
Electrochem. Commun. 5 (2003) 392–395. istry and Chemical Engineering, Hunan University, PR China. He is
[26] J.M. Ortega, Electrodeposition of copper on poly(o-aminophenol) presently a PhD student in College of Chemistry and Chemical Engineer-
modified platinum electrode, Thin Solid Film 360 (2000) 159–165. ing, Hunan University, PR China. He is mainly engaged in the research
[27] M. Somasundrum, K. Kirtikara, M. Tanticharoen, Amperometric de- of biosensors and bioelectrochemistry.
termination of hydrogen peroxide by direct and catalytic reduction
at a copper electrode, Anal. Chim. Acta 319 (1996) 59–70.
[28] J.F. Zhou, J.J. Yang, Z.J. Zhang, W.M. Liu, Q.J. Xue, Study on the Shouzhuo Yao is presently employed as a professor in College of Chem-
structure and tribological properties of surface-modified Cu nanopar- istry and Chemical Engineering, Hunan University, PR China. He is
ticles, Mater. Res. Bull. 34 (1999) 1361–1367. mainly engaged in the research of analytical chemistry, piezoelectrical
[29] S.S. Joshi, S.F. Patil, V. Iyer, S. Mahumuni, Radiation induced sensors and molecular imprinting polymer.
synthesis and characterization of copper nanoparticles, Nanostruct.
Mater. 10 (1998) 1135–1144. Lihua Nie is presently employed as a senior engineer in College of
[30] S.L. Mu, H.G. Xue, B.D. Qian, Bioelectrochemical responses of the Chemistry and Chemical Engineering, Hunan University, PR China. She
polyaniline glucose oxidase electrode, J. Electrochem. Chem. 304 is mainly engaged in the research of analytical chemistry and molecular
(1991) 7–16. imprinting polymer.
[31] H. Zheng, H.G. Xue, Y.F. Zhang, Z.Q. Shen, A glucose biosen-
sor based on microporous polyacrylonitrile synthesized by single Jianjun Xia is a postgraduate student of College of Chemistry and Chem-
rare-earth catalyst, Biosens. Bioelectron. 17 (2002) 541–545. ical Engineering, Hunan University, PR China. He is mainly engaged in
the research of bioelectrochemistry.
Biographies
Wenyan Tao is a PhD student in College of Chemistry and Chemical
Jinhua Chen obtained Doctor’s degree from College of Chemistry and Engineering, Hunan University, PR China. She is mainly engaged in the
Chemical Engineering, Hunan University, PR China in 1997. He is research of biosensors and bioelectrochemistry.

You might also like