Professional Documents
Culture Documents
24
Electrochemical DNA Biosensor Based on HRP-labeled Probe 25
INTRODUCTION
EXPERIMENTAL
Hybridization buffer solution was prepared with 0.05 mol l21 NaCl
0.01 mol l21 Tris 0.001 mol l21 EDTA (pH 8.8, STE) in asepsis water and
stored in a refrigeratory. The supporting electrolyte was 0.067 mol l21
phosphate buffer solution containing 0.3 mol l21 NaCl (pH 7.0, PBS). Other
chemicals were of analytical reagent grade. All solutions were prepared
with doubly distilled water.
Apparatus
Cyclic voltammetry and amperometric analysis were carried out using a CHI
760b Electrochemical Analyzer (Chen Hua Instrument Inc., Shanghai, China).
The three-electrode system consisted of gold electrode (AuE, 1 mm diameter)
as the working electrode, a SCE as the reference electrode and a Pt foil as the
counter electrode. Cyclic voltammetric experiments were performed in
unstirred solutions. Amperometric measurements were carried out in stirred
substrate solutions with a steady-state background current first obtained
before standard H2O2 solution was added into the buffer solution. All poten-
tials were measured and reported versus the SCE. A magnetic stirrer and bar
provided convective transport.
The AuE surface was treated before modification by polishing with 0.05 mm
alumina/water slurry on a polishing cloth. The electrodes were then soaked
for 30 min in a freshly prepared piranha solution (30% H2O2/70% concen-
trated H2SO4). Subsequently, the electrodes were cleaned by cycling
between the potentials 21.0 and þ1.0 V versus SCE in 0.1 mol l21 H2SO4
solution at a scan rate of 100 mV/s for approximately 20 min until reproduci-
ble scans were observed. The electrodes were rinsed with water before self
assembly (SAM) modification.
The surface modification is illustrated in Fig. 1. An aliquot (5 ml) of probe
DNA (5.66 m mol l21) solution was pippted onto the surface of AuE. The
electrode was allowed to dry overnight for approximately 15 h (at room
28 Z. Wang et al.
Measurement
The operation of amperometric DNA sensor was based on the use of detection
DNA labeled with enzyme HRP and H2O2 as the substrate. Hydroquinone, one
of the best electron transfer mediator for HRP (Volpe et al. 1998), is chosen as
the electron transfer mediator. In the presence of HRP, H2O2 oxidizes hydro-
quinone to benzoquinone. Figure 2 shows the cyclic voltammograms
obtained with the DNA sensor in an unstirred 0.067 mol l21 PBS (pH 7.00)
with and without addition of H2O2. One couple of oxidation/reduction peak,
which represents the cyclic voltammogram of hydroquinone, was observed in
the presence of hydroquinone (Fig. 2a). But in the presence of 2.0 mmol l21
H2O2, the cyclic voltammogram displays a dramatic enhancement of the
cathodic peak current and a concomitant decrease of the anodic peak current
(Fig. 2b). These phenomena indicate that the HRP attached to the DNA
sensor surface retained high enzymatic catalytic activity and the hydroquinone
could effectively shuttled electrons from the redox center of HRP to AuE.
Figure 3 shows results obtained with the DNA sensor prepared after incu-
bating in a hybridization solution followed by amperometric determination at
2150 mV. When the amount of target DNA varied in the hybridization
solution the response current would vary. This is the basis of amperometric
assay of target DNA.
30 Z. Wang et al.
Figure 3. Curren-time curve of an incubated DNA biosensor in 0.067 mol l21 PBS
(pH 7.00) containing 1.0 mmol l21 hydroquinone recorded at 2150 mV. After the
stabilization of background current (a– b), the solution of H2O2 is added (b) and
then the current reaches an equilibrium value (c).
Self-assembled Time
The amount of probe DNA attached to the AuE surface was determined by the
self-assembled time. In order to obtain a maximum response with a minimum
self-assembled time, the AuE was self-assembled with the same amount of
probe DNA but different time at room temperature. Fig. 4 shows the current
response of the DNA sensor increased up to 15 h and then tends to change
only slightly. So, the self-assembled time of 15 h was routinely employed in
these assays.
Hybridization Time
Hybridization Temperature
An additional parameter that affected the assay was the hybridization tempera-
ture. The effect of hybridization temperature on response current was
examined from 25 to 658C (Fig. 6). It was found that the signal increases
with an increase of temperature up to 658C. At higher temperature the
dsDNA would untie, 658C has been chosen as the optimum temperature for
the hybridization of the DNA sensor.
Figure 5. Effect of hybridization time. The DNA biosensor was incubated in STE
buffer solution containing 1165 pmol l21 target DNA at 658C.
32 Z. Wang et al.
observed that the current response increases with an increase of the pH of the
substrate solution up to pH 7.00, and then it decreases at higher pH. So, pH
7.00 of the substrate solution was chosen as the optimum for the assays.
Calibration Curve
Figure 8 shows the calibration curve obtained using target DNA standards
under optimal experimental conditions. The response signal increases
linearly with the increase of the logarithm of the target DNA concentration
in the range of 1.17 10211 21.17 1027 mol l21 with a correlation coeffi-
cient value of 0.964. The linear regression equation is I ¼ 22.54 þ 3.67 log
CDNA with a detection limit of 5.85 10212.
Figure 8. Calibration curve of the DNA biosensor. The DNA biosensor was incu-
bated in SET buffer solution of pH 7.00 containning different amount of target
DNA. Other conditions are optimal experimental conditions.
Figure 9. Amperometric responses upon adding 2.0 mmol L21 H2O2 to PBS of
2325 pmol l21 random oligonucleotide (a), 1600 pmol l21 four-base-mismatched
oligonucleotide (b), and 1165 pmol l21 perfectly matched oligonucleotide (c).
34 Z. Wang et al.
Regeneration of ssDNA and removal of target DNA and detection DNA were
achieved by rinsing the electrode in hot hybridization buffer (1008C) for 5 min.
The electrode after regeneration can be used again to detect target DNA.
Reproducibility of the DNA sensor was tested by measuring 1165 pmol l21
target DNA solution. Four DNA sensors, made independently, showed the
response current values of 6.78, 6.43, 6.62 and 6.85 mA with an acceptable
variation coefficient of 2.83% (n ¼ 4).
CONCLUSIONS
REFERENCES
Evans, P.C., Gray, J., Wreghitt, T.G., Graeme, J.M., and Alexander 1998. Optimisation
of the polymerase chain reaction and dot-blot hybridisation for detecting cytomega-
lovirus DNA in urine: comparison with detection of early antigen fluorescent foci
and culture. Journal of Virological Methods, 73: 41 – 52.
Hashimoto, K., Ito, K., and Ishimori, Y. 1994. Sequence-specific gene detection with a
gold electrode modified with DNA probes and an electrochemically active dye. Anal.
Chem., 66: 3830– 3833.
Ihara, T., Nakayama, Murata, M., Nakano, K., and Maeda, M. 1997. Gene sensor using
ferrocenyl oligonucleotide. Chem. Commun., 17: 1609– 1610.
Izzo, V., Asaro, M.R., Forte, G.I., and Barbieri, R. 2000. Use of voltage gradient gel
electrophoresis in apoptotic DNA analysis. Journal of Chromatography A, 890:
371– 374.
Kievit, J. and Bruinvels, D.J. 1995. Detection of recurrence after surgery for colorectal
cancer. European Journal of Cancer, 31: 1222–1225.
Mikklesen, S.R. 1996. Electrochecmical biosensors for DNA sequence detection.
Electroanalysis, 8: 15 – 19.
Palecek, E., Fojta, M., Tomschik, M., and Wang, J. 1998. Electrochemical biosensors
for DNA hybridization and DNA damage. Biosens. Bioelectron., 13: 621– 628.
Souteyrand, E., Cloarec, J.P., Martin, J.R., and Wilson, C. Direct detection of the
hybridization of synthetic homo-oligomer DNA sequences by field effect. J. Phys.
Chem. B, 101: 2980– 2985.
Volpe, G., Compagnone, D., Drasci, R., and Palleschi, G. 1998. 3,30 ,5,50 -Tetramethyl-
benzidine as electrochemical substrate for horseradish peroxidase based enzyme
immunoassays: a comparative study. Analyst, 123: 1303– 1307.
Vossen, K.M. and Fried, M.G. 1997. Sequestration stabilizes lac repressor –DNA com-
plexesduring gel electrophoresis. Analytical Biochemistry, 245: 85 – 92.
Wang, J. 2002. Electrochemical nucleic acid biosensors. Anal. Chim. Acta, 469:
63 –71.
Wang, J., Liu, G.D., and Merkoc, A. 2003. Electrochemical coding technology for sim-
ultaneous detection of multiple DNA targets. Journal of American Chemical.
Society, 125: 3214– 3215.
Wang, J., Rivas, G., Fernandes, J.R., Paz, J., Jiang, M., and Waymire, R. 1998.
Indicator-free electrochemical DNA hybridization biosensor. Anal. Chim. Acta,
375: 197– 203.
Xie, H., Zhang, C.Y., and Gao, Z.Q. 2004. Amperometric detection of nucleic acid at
femtomolar levels with a nucleic acid/electrochemical activator bilayer on gold
electrode. Analytical Chemistry, 76: 1611– 1617.
Zhang, L.X., Gillespie, B.W., Marrs, C.F., and Foxman, B. Optimization of a
fluorescent-based phosphor imaging dot blot DNA hybridization assay to assess
E. coli virulence gene profiles. Journal of Microbiological Methods, 44: 225– 233.
Zhu, N.N., Zhang, A.P., Wang, Q.J., He, P.G., and Fang, Y.Z. 2004. Electrochemical
detection of DNA hybridization using methylene blue and electro-deposited zirconia
thin films on gold electrodes. Analytica. Chim. Acta, 510: 163– 168.