You are on page 1of 9

Analytica Chimica Acta 480 (2003) 109–117

Solid-state pH ultramicrosensor based on a tungstic oxide


film fabricated on a tungsten nanoelectrode and its
application to the study of endothelial cells
Katsunobu Yamamoto a , Guoyue Shi b,∗ , Tianshu Zhou c , Fan Xu a , Min Zhu b ,
Min Liu b , Takeshi Kato a , Ji-Ye Jin d , Litong Jin b
a Laboratory of Research and Development, Bioelectroanalytical System Inc., 1-36-6 Oshiage, Sumida-Ku, Tokyo, Japan
b Department of Chemistry, East China Normal University, Shanghai 200062, China
c Department of Environmental Science, East China Normal University, Shanghai 200062, China
d Instrumental Analysis Center, Gifu University, Yanagido, Gifu 501-1193, Japan

Received 6 June 2002; received in revised form 5 December 2002; accepted 11 December 2002

Abstract
In this paper, preparation of a novel pH ultramicrosensor and its physiological application has been discussed. A tungsten
nanoelectrode was produced by an etching method in 0.1 mol/l NaOH solution at the potential of +0.4 V (versus Ag/AgCl
reference electrode) for about 100 s and the diameters ranged from 500 to 800 nm. The pH ultramicrosensor was fabricated by
producing WO3 at W nanoelectrode surface by electrooxidation in 2.0 mol/l H2 SO4 solution between 1.0 and 2.0 V. At last,
Nafion was coated on the surface of WO3 to protect the pH ultramicrosensor. The W/WO3 pH ultramicrosensor exhibited
a good pH linear region from 2.0 to 12.0 with a super-Nernstian slope of −53.5 ± 0.5 mV/pH unit. Response times ranged
from 3 s at about pH 6.0–7.0 up to 15 s at high pH. An interference of various ions to the pH measurement was also studied in
this paper. We also studied the lifetime, stability and reproducibility of the W/WO3 pH ultramicrosensor. In order to testing
the performance of W/WO3 ultramicrosensor, we applied it to measure the extracellular pH values and a pH variation was
also given about the normal, damaged and recovery endothelial cells.
© 2002 Elsevier Science B.V. All rights reserved.
Keywords: W/WO3 pH ultramicrosensor; Endothelial cells

1. Introduction amperometric electrode [1] and potentiometric elec-


trode [2]. Among them, the most commonly used
The pH sensor is a kind of chemical sensor, which pH electrode is the glass electrode because of its
has been used widely in many fields, such as sci- high selectivity, stability and wide pH range. But the
entific, industrial, agriculture, and environmental, conventional glass electrode also has its obvious lim-
clinical and other concern. Scientists have developed itations, such as the fragility of the glass, difficulty to
many methods for the pH determination, including remodel its shape and a long time needed to reform
a hydrate layer before use. It is necessary to develop
∗ Corresponding author. Tel.: +86-21-6223-2627; many suitable pH sensors to use in both aqueous and
fax: +86-21-62232627. non-aqueous medium. Miniaturized pH sensor is also
E-mail address: iamshgy@hotmail.com (G. Shi). required in cell determination.

0003-2670/03/$ – see front matter © 2002 Elsevier Science B.V. All rights reserved.
doi:10.1016/S0003-2670(02)01602-1
110 K. Yamamoto et al. / Analytica Chimica Acta 480 (2003) 109–117

During the last decade, efficient efforts have been nanoelectrode with a tip of about 500–800 nm. Then
given on the designing non-glass pH sensors to over- WO3 film was produced on the surface of W na-
come the above-mentioned limitations of the glass noelectrode by cyclic voltammetry (CV) in 2.0 mol/l
electrode. Many related papers have been published H2 SO4 solution between 1.0 and 2.0 V at a scan rate
for pH determination since the earliest example of of 20 mV/s. The importance of endothelial cells in the
the use of a polymer film modified electrode as a pH regulation of vascular biology is being increasingly
sensor reported by Heineman et al. [3]. The modified recognized. But up to now, there are few papers to
polymers include polyaniline, poly(vinyl chloride), study the relationship between pH variation and dam-
poly(3-cyclohexyl thiophene) and other organic mate- aged endothelial cells directly with miniaturized pH
rials [4–7]. Excellent properties in chemistry, optics, sensor. Therefore, the main goal of the present study is
thermodynamics and mechanics make the sol–gel the development of pH ultramicrosensor for the mon-
technology be used in the fabrication of pH sensor itoring of pH value from cultured endothelial cells
[8–12]. Many other efficient methods for pH deter- and observes the pH variation generated from normal,
mination in various samples were also developed in damaged and recovery endothelial cells. A flavonoid
recent years [13–15]. Chinese traditional medicine was also given to the
Furthermore, metal/metal oxide pH electrode re- damaged endothelial cells and then the extracellular
ceived more and more attention with its excellent pH value was recovered to normal level.
characteristics of easy miniaturization, a long lifetime
and a strong tip. Several methods have been used to
produce metal oxide for pH sensing. They can be 2. Experimental
prepared by thermal decomposition [16], by sputter
deposition [17,18] or by electrochemical growth [19]. 2.1. Chemicals and apparatus
Yao and co-workers [20] used iridium oxide electrode
for pH measurement. Honda et al. applied Sb/Sb2 O3 Hanks balanced salt solution (HBSS) was prepared
as the pH sensor. Wipf et al. [21] also used iridium as following: NaCl 8.0 g/l, Na2 HPO4 ·2H2 O 0.06 g/l,
oxide electrode as the tip of scanning electrochemical KCl 0.4 g/l, KH2 PO4 0.06 g/l, MgSO4 ·7H2 O 0.2 g/l,
microscope (SECM). Antimony microdisk electrodes glucose 1.0 g/l, and CaCl2 ·2H2 O 0.19 g/l, pH = 7.4.
for the measurement of pH near electrode surface Nafion (1 wt.%) was purchased from Aldrich Chem-
with SECM were prepared from antimony shot by ical Company. Deoxygenated with nitrogen (above
Bard and co-workers [22]. The electrochemical be- 99.99%), HBSS was used as buffer solution in all ex-
havior of tungsten and tungsten oxide was discussed periments except real sample analysis. All aqueous
by previous work [23–25]. Wrighton and co-workers solutions were prepared with doubly distilled water.
also coated a 0.15 ␮m thick layer of polycrystalline All electrochemical measurements were performed
WO3 on Au or Pt microelectrode and studied the pH with an ALS/CHI 900Electrochemical Workstation
characteristics of WO3 [17]. (BASJ, Tokyo, Japan). W wire (10 ␮m diameter)
Along with the developing of the field of life sci- was purchased from Nilaco company of Japan.
ence, the need for miniaturization of pH sensor has Three-electrodes system included a Pt wire (0.5 mm
been recognized. Intracellular and extracellular pH diameter) as counter electrode, a modified Ag/AgCl
contribute importantly to the control of many critical wire (0.2 mm diameter) as reference electrode, and a
physiological processes. The sensor should be of very W/WO3 pH ultramicrosensor as working electrode.
small size to allow spatial resolution and to minimize Scanning electron micro-analyzer (JSM-5610LV,
distortion of the tissue, which is necessary for reliable Japan) was used to characterize the etching W nano-
physiological measurements. The pH ultramicrosen- electrode. For real sample analysis, an XDS-1 model
sors as small as a few micrometers in tip diameter have stereomicroscope (Chongqing Optical and Electrical
been constructed [5,26–29] for in vivo studying, cells Instrument Company, China) equipped with video
research or ultramicroenvironmental analysis, etc. In and a WK-2 model micromanipulator (Xi’an North-
this paper, a novel W/WO3 ultramicrosensor for pH west Optical & Electrical Company, China) was used
determination was prepared. Firstly we fabricated a W as instruments for guiding micromanipulation.
K. Yamamoto et al. / Analytica Chimica Acta 480 (2003) 109–117 111

2.2. Preparation of the W/WO3 ultramicrosensor tremity was filled with epoxy resin in order to assure
tightness. The redundant W wire of the tip was cut off
2.2.1. Preparation of W nanoelectrode carefully under the microscopy.
A 10 ␮m W wire was connected to a copper wire
with silver conductive paint. Then it was treated in 2.2.2. Construction of W/WO3 ultramicrosensor
an etching solution (0.1 mol/l NaOH solution) with a Prior to oxidation, W nanoelectrode was ultra-
constant potential of +0.4 V. Every 10 sec, the W wire sonically cleaned with acetone, 1.0 mol/l NaOH, 1:1
was viewed under the stereomicroscope until the diam- HNO3 and double distilled water, respectively. A Pt
eter we need was received. Fig. 1A is the photograph wire as counter electrode, a modified Ag/AgCl wire
of scanning electron microscopy of W nanoelectrode. as reference electrode, the W nanoelectrode was cy-
At last, the etched W wire connected to a copper wire cled in the potential range between +1.0 and +2.0 V
was aspirated into a standard glass capillary (Esselte in 2.0 mol/l H2 SO4 at the scan rate of 20 mV/s for 20
Leitz, Germany), which was pulled in advance into cycles. The WO3 film was produced on the surface of
pipette with a tip of about 20␮m. The extremity of the W nanoelectrode. Then the W/WO3 ultramicrosensor
pipette was fixed with ethanol flame, and another ex- was immersed in 2.0 mol/l H2 SO4 solution for about
12 h.
Many substances, such as protein and enzyme exist
in cell samples. They can be absorbed on the surface of
W/WO3 pH ultramicrosensor and effect its sensitivity
and response time. So 2 ␮l 1% Nafion was drop on
the W/WO3 ultramicrosensor surface and air-dried for
10 min. Fig. 1B is the schematic presentation of the
preparation of a W/WO3 pH ultramicrosensor.
In vivo experiments were performed in HBSS un-
der constant temperature (37 ± 0.5 ◦ C) condition.
The three-electrode system (containing a Pt counter
electrode, a Ag/AgCl reference electrode and the
W/WO3 pH ultramicrosensor) was fixed together to
keep the distance of each electrode unchangeable
before it was placed as near as possible close to the
cells.

2.3. pH response measurement

The electromotive force (emf) of the W/WO3 ul-


tramicrosensor was measured as a function of pH of
the test solution. The base solution was HBSS just
as ‘Section 2.1’ mentioned to which either 1.0 mol/l
H3 PO4 or 1.0 mol/l NaOH was added dropwise to
obtain the desired pH value. The dynamic range of
this method is about pH 2.0–12.0. The emf results of
sample solutions were recorded till potential values
became stable. A commercial pH glass electrode was
also used to calibrate the W/WO3 pH ultramicroe-
sensor. A thermostat controlled the temperature of all
Fig. 1. (A) The photograph of scanning electron microscopy of
the test solutions at 37.0 ± 0.5 ◦ C. The whole appara-
tungsten nanoelectrode. (B) The schematic presentation of the tus was housed in a Faradic cage for eliminating the
W/WO3 ultramicrosensor. interferences.
112 K. Yamamoto et al. / Analytica Chimica Acta 480 (2003) 109–117

2.4. pH ultramicrosensor selectivity was immersed in the pH 7.4 buffer solution. After po-
tential stabilization, different solution of LiCl, NaCl,
The selectivity of the W/WO3 pH ultramicrosensor KCl, NH4 Cl, CaCl2 or CuCl2 prepared with pH 7.4
was test under conditions relevant to the physiologi- buffer solution, was added. The changes of emf were
cal applications. The W/WO3 ultramicrosensor, along recorded.
with the reference electrode and counter electrode,
2.5. Endothelial cells culture and pH
measurement of endothelial cells

The endothelial cells were prepared from the sepa-


rated cells of bovine aortic by grown in gelatin-coated
culture plates at 37 ± 0.5 ◦ C. According to the method
of reference [30], cells were fed every 2 days with
RPMI 1640 medium supplemented with 15% bovine
calf serum, 100 U/l penicillin and 100 U/l strepto-
mycin. The cells were divided into two parts, one
was a normal group and the other one was a damaged
group, which was done by ischemic treatment for 15
and 30 min. Fig. 2A and B are the photos of normal
and damaged taken with a stereomicroscope (XDS-1,
Chongqing optical & electrical instrument company).
Then they were all given a flavonoid chinese tradi-
tional medicine. Fig. 2C is the photo of recovery cells
taken with the stereomicroscope.
Before electroanalytical monitoring of pH values
of endothelial cells, the slide used for cell growth on
was washed and filled with HBSS. Endothelial cells
attaching on the slide were counted under the in-
verted stereomicroscope, then placed in 4 cm diameter
petri-dishes, and kept in an incubator at 37.0 ± 0.5 ◦ C
for at least 10 min. Counted under a stereomicroscope,
the amount of the cells was about 1.0 × 106 /ml. Then
the cells were grown statically at 37 ◦ C in a humidi-
fied atmosphere of 95% air and 5% CO2 . After 6–8 h,
the cells settled to the bottom of the culture well. pH
values were recorded from the normal, damaged and
recovery endothelial cells groups.

3. Results and discussion

3.1. Characteristics of the W/WO3 pH


ultramicrosensor

Fig. 1A is a photograph of scanning electron mi-


croscopy (SEM) of tungsten nanoelectrode, which has
Fig. 2. The photos of (A) normal group (100 times); (B) damaged etched in 0.1 mol/l NaOH solution at a constant po-
group (250 times) and (C) recovery group (100 times). tential of +0.4 V. From Fig. 1A, we can see that the
K. Yamamoto et al. / Analytica Chimica Acta 480 (2003) 109–117 113

detailed diameter is about 500–800 nm. It is because also found that the pH ultramicrosensor had poor se-
that tungsten is oxidized to dissolved WO4 2− ions un- lectivity when a 10 cycle number of CV was selected.
der diffusion control by mass transport of OH− ions So the optimum condition was achieved at the thick-
as the overall reaction [23]. ness of 20 cycle number of CV ranging from +1.0 to
The electroformed species dissolve chemically in +2.0 V at the scan rate of 20 mV/s. We can see the
the electrolysis media. Therefore, we can get differ- response time is as short as 3 s at pH 6–7 and the
ent size of W electrodes through controlling different response time is more than about 15 s at extreme pH
etching time in NaOH solution. of 2 or 12 as shown in Fig. 3. It is beneficial for the
Many scientists have studied the electrochemical pH determination under the physiological condition.
characteristics of tungsten and tungsten oxide in al- It is because that WO3 undergoes the following
kaline and acidic media [17,23,31–33]. We have also reversible, proton-dependent redox reaction [17] in
produced WO3 by electrooxidation in acidic solution. buffer solution:
According to Section 2.2.2, we prepared the W/WO3
pH ultramicrosensor by cycling the W nanoelectrode WO3 + nH+ + ne− ⇔ Hn WO3
in the potential range between +1.0 and +2.0 V at the The reversible reaction changes the ratio of the
scan rate of 20 mV/s for 20 cycles in 2.0 mol/l H2 SO4 . oxide in the film, the membrane potential will also
It is obvious that the thickness of the WO3 film change subsequently.
covered on the W nanoelectrode is very important for
the pH determination. Fig. 3 is the effect of different 3.2. Emf of the W/WO3 pH ultramicrosensor
cycle number of CV to the response time of W/WO3
ultramicrosensor at different pH solution. Fig. 3 dis- The pH sensing performance of the W/WO3 ultra-
plays the W/WO3 ultramicrosensor has excellent microsensor was valuated in pH buffer solution. Fig. 4
response time at 10 or 20 cycles, and the pH ultra- shows the emf versus Ag/AgCl reference electrode
microsensor exhibits prolonged response time when of the pH ultramicrosensor and commercial glass
the cycle numbers exceed 30 cycles. Furthermore, we electrode at different pH test solutions. It proves that
the W/WO3 ultramicrosensor has a super-Nernstian
response with slope of −53.5 mV/pH unit. Linear pH

Fig. 4. An emf responses of W/WO3 ultramicrosensor and com-


mercial glass pH sensor at different pHs in HBSS with either
Fig. 3. Effect of different cycle numbers of CV on the response time 1.0 mol/l H3 PO4 or 1 mol/l NaOH added for adjustment to the
of W/WO3 ultramicrosensor at different pH solution A: 10 cycles; desired pH thermostated at 37.0 ± 0.5 ◦ C. (a) glass pH electrode;
B: 20 cycles; C: 30 cycles; D: 40 cycles. (b) W/WO3 pH ultramicrosensor.
114 K. Yamamoto et al. / Analytica Chimica Acta 480 (2003) 109–117

Table 1
Calibration of different W/WO3 pH ultramicrosensors.
Electrode number pH range Slope (mV) R2

1 2.0–12.0 −53.5 0.9990


2 2.0–12.0 −53.8 0.9972
3 2.0–12.0 −53.4 0.9982
4 2.0–12.0 −53.0 0.9979

responses are obtained over a wide pH range of


2.0–12.0 (γ = 0.9990). It is clearly evident that the
characteristics of the W/WO3 pH ultramicrosensor
are very good and well equivalent to the performance
of a commercial glass electrode. Table 1 summarizes
the results of four different W/WO3 ultramicrosen-
sors prepared with the same method. These slopes
are within the range of −53.0 to −53.8 mV. The
following experiments were all done by the no. 1
ultramicrosensor.
The variation of the W/WO3 ultramicrosensor’s re-
sponse was determined from step changes of pH buffer
solution after injection of H3 PO4 or NaOH, as shown
in Fig. 5A and B. As can be seen, the response time
is about 4–6 s in spite of pH from 7.4 to 6.0 or from
7.4 to 9.8.

3.3. Selectivity of the W/WO3 pH ultramicrosensor

Selectivity is very important for pH determination


in biological samples. So the potentiometric selectivity
of the ultramicrosensor for hydrogen ion against ma- Fig. 5. The dependence of the emf response of the W/WO3 pH
jor interfering ions was studied using the mix-solution ultramicrosensor to an injection of (A) 1 mol/l H3 PO4 and 1 mol/l
method in this paper. Table 2 is the results of the ion NaOH into pH 7.4 buffer solution.
selectivity coefficients expressed as log KM,H . In addi-
tion, many electroactive substances, such as ascorbic
acid, urate and neurotransmitters also do not interfere ultramicrosensor is selective for pH determination
with the measurement at the level near physiological and is suitable for the following research work.
concentration. The data prove that the W/WO3 pH Some papers have reported that the Nafion film also
responds to many cations. Anson group [36] found
Table 2 that the electrode modified with redox couples incor-
Ion selectivity coefficients of no. 1 ultramicrosensor porated in Nafion coating has good responses to Li+ ,
Ions KM,H H+ , Na+ and K+ , Zn2+ , Mg2+ . Ugo et al. [37] also
Li+ 8.2
employed the Nafion coated glass carbon electrode for
Na+ 9.6 preconcentrating and detecting Fe2+ and Fe3+ cations
K+ 9.2 from aqueous solutions. So the Nafion modified elec-
NH4 + 7.4 trode is not a selective pH sensor. In this paper, we
Ca2+ 11.0 modified Nafion film on the surface of WO3 . The
Cu2+ 10.7
thickness of the Nafion film was estimated to be only
K. Yamamoto et al. / Analytica Chimica Acta 480 (2003) 109–117 115

about 0.05–0.1 ␮m. It has poor response for H+ , espe- into 12 groups, including 6 normal groups, 6 ischemia
cially in alkali solution and strong acid. But the WO3 groups that were cultured for ischemic treatment for
film that we prepared on the surface of W wire has ex- 15 and 30 min, respectively. The above-described
cellent selectivity for H+ determination and response W/WO3 pH ultramicrosensor was used to determine
in the range of pH 2.0–12.0. The similar applications pH of all the samples. Then a flavonoid medicine
of Nafion/metal oxide pH sensors have also been re- was injected into the normal and damaged groups,
ported in recent years [28,38]. which are the recovery groups. All the experiments
to monitor pH of endothelial cells were performed in
3.4. Stability, reproducibility and lifetime of the Faraday cage incubated at 37.0 ± 0.5 ◦ C.
W/WO3 pH ultramicrosensor Fig. 7 is the pH data recorded with all samples of
endothelial cells. From Fig. 7, we can see the pH of
The W/WO3 ultramicrosensor was test over a pe- normal group is about 7.4. When the cells are treated
riod of 7 days after fabrication. The particular data of ischemia for 15 min, pH value changes from 7.4
are shown as Fig. 6. Fig. 6 displays that the emf has to 6.9. The W/WO3 pH ultramicrosensor shows a
slightly variation during the period of 7 days. Further- variation of 0.5 pH. Along with the treatment of is-
more, the value only varied about 10% in 2.0 mol/l chemia for another 15 min, the pH value of the dam-
H2 SO4 solution after 1 month of storage. Emf re- aged endothelial cells decreases continuously and at
sponses have been determined in HBSS by the same last reached pH 6.2. This phenomenon is because is-
electrode and the relative standard deviation is 1.5% chemia can result in the excess production of lactic
(n = 8). This indicates that the pH ultromicrosensor acid and accumulation of protons, namely acidosis in
is stable. biology. This result is similar to the previous studies
[28,34,35].
Flavonoids are a ubiquitous group of polyphenolic
3.5. pH Measurement of normal, damaged and substances, which are present in most plants, concen-
recovery endothelial cells trating in seeds, fruit skin or peel, bark, and flowers.

The cells were cultured as Section 2.4. In terms of


the culture medium, the experimental samples divided

Fig. 7. pH measurements of (A) normal group; (B) damaged group


(ischemia for 15 min); (C) damaged group (ischemia for 30 min);
Fig. 6. The emf responses of W/WO3 pH ultramicrosensor in pH (D) normal group after giving medicine; (E) damaged group after
6.9 buffer solution. giving medicine.
116 K. Yamamoto et al. / Analytica Chimica Acta 480 (2003) 109–117

A great number of plant medicines contain flavonoids, References


which have been reported by many authors as having
antibacterial, anti-inflammatory, antiallergic, antimu- [1] M.S. Ansky, A. Pizzariello, S.S. Anska, S. Miertus, Anal.
tagenic, antiviral, antineoplastic, anti-thrombotic, and Chim. Acta 415 (2000) 151.
[2] A.K. Covington (Ed.), Ion-Selective Electrode Methodology,
vasodilatory actions [39]. vol. 1, CRC Press, Boca Raton, 1979.
In this work, the flavonoid medicine was added [3] W.R. Heineman, H.J. Wieck, A.M. Yacynych, Anal. Chem.
to the normal and damaged groups and then pH 52 (1980) 345.
was also determined after about 10 min. We find the [4] O. Dinten, U.E. Spichiger, N. Chaniotakis, P. Gehrig, B.
pH has not obvious variation of the normal groups, Rusterholz, W.E. Morf, W. Simon, Anal. Chem. 63 (1991)
596.
but the pH of damaged groups is recovered to the [5] X.J. Zhang, B. Ogorevc, J. Wang, Anal. Chim. Acta 452
normal level, from 6.2 to 7.3. It may be because (2002) 1.
the flavonoid medicine has the function of resist- [6] D. De Beer, A. Glud, E. Epping, M. Kuhl, Liminol. Oceanogr.
ing free radicals, which play a central role in many 42 (1997) 1590.
[7] R. Yuan, Y.Q. Chai, G.L. Shen, R.Q. Yu, Talanta 40 (1993)
pathological conditions and chronic diseases. At the
1255.
same time, the flavonoid medicine is effective to ex- [8] J.P. Li, T.Z. Peng, C. Fang, Anal. Chim. Acta 455 (2002) 53.
panding blood vessel, reducing blood viscosity and [9] R. Zusman, C. Rottman, M. Ottolenghi, D. Avnir, J. Noncryst.
improving anoxia endurance. The further research Solids 122 (1990) 107.
of the drug and its mechanism are being studied in [10] R. Makote, M.M. Collinson, Anal. Chim. Acta 394 (1999)
195.
our lab. [11] T.M. Butler, B.D. MacCraith, C. McDonagh, J. Noncryst.
Solids 224 (1998) 249.
[12] J. Brinker, G. Scherer, Sol–Gel Science, Academic Press,
4. Conclusion New York, 1989.
[13] P.C. Pandey, G. Singh, Talanta 55 (2001) 773.
[14] T. Sakurai, Y. Husimi, Anal. Chem. 64 (1992) 1996.
In this paper, a novel pH ultramicrosensor was de- [15] P.S. Zhao, W.J. Cai, Anal. Chim. Acta 395 (1999) 285.
veloped based on a WO3 film electrodeposited onto [16] S.M. Lin, T.C. Wen, Electrochim. Acta 39 (1994) 393.
the surface of W nanoelectrode with tip dimensions [17] M.J. Natan, T.E. Mallouk, M.S. Wrighton, J. Phys. Chem. 91
ranging from 500 to 800 nm in diameter. The prepared (1987) 648.
[18] K.G. Kreider, M.J. Tarlov, J.P. Cline Sens. Actuators B B28
W/WO3 pH ultramicrosensor exhibits a wide pH range (1995) 167.
of 2.0–12.0 and fast respond time. Comparing with [19] C. Colombo, T. Kappes, P.C. Hauser, Anal. Chim. Acta 412
conventional glass pH electrode, the W/WO3 ultrami- (2000) 69.
crosensor has a good mechanical strength, nano level [20] M. Wang, S. Yao, M. Madou, Sens. Actuators B B81 (2002)
313.
dimension of the tip, simple fabrication and conve- [21] D.O. Wipf, F.Y. Ge, T.W. Spaine, J.E. Baur, Anal. Chem. 72
nient for mass production. The W/WO3 pH ultrami- (2000) 4921.
crosensor has been applied to the pH measurement of [22] B.J.R. Horrocks, D. Schmidtke, A. Heller, A.J. Bard, Anal.
endothelial cells in this work. As a simple and quick Chem. 65 (1993) 3605.
method, the pH ultramicrosensor is hopeful to be ap- [23] P.I. Ortiz, M.L. Teijelo, M.C. Giordano, J. Electroanal. Chem.
243 (1988) 379.
plied in single cell studying. [24] B. Reichman, A.J. Bard, J. Electrochem. Soc. 126 (1979) 583.
[25] A.T. Vas’ko, in: A.J. Bard (Ed.), Encyclopedia of
Electrochemistry of the Elements, vol. XI, Marcel Dekker,
Acknowledgements New York, 1978, p. 69.
[26] C. Nicholson, J. Neurosci. Meth. 48 (1993) 199.
[27] M. Saito, H. Matsuoka, Anal. Chim. Acta 404 (2000) 223.
This work is supported by the Phosphor Project [28] S.A.M. Marzouk, S. Ufer, R.D. Buck, T.A. Johnson, L.A.
of Science & Technology of Shanghai for excellent Dunlap, W.E. Cascio, Anal. Chem. 70 (1998) 5054.
young people. And we also thank the offer grant by [29] V.V. Cosofret, M. Erodosy, T.A. Johnson, R.P. Buck, R.B.
Opening fund of State key lab of Electroanalytical Ash, M.R. Neuman, Anal. Chem. 67 (1995) 1647.
[30] S.L. Ongil, L.G. Santiago, M. Griera, J. Molpeceres, M.R.
Chemistry, Changchun Institute of Applied Chemistry, Puyol, D.R. Puyol, Life Sci. 70 (2001) 699.
Academy of Sciences of China and Bioelectroanalyt- [31] L.D. Burke, D.P. Whelan, J. Electroanal. Chem. 135 (1982)
ical System of Japan. 55.
K. Yamamoto et al. / Analytica Chimica Acta 480 (2003) 109–117 117

[32] L.D. Burke, T.A.M. Twomey, D.P. Whelan, J. Electroanal. [36] R. Naegeli, J. Redepenning, F.C. Anson, J. Phys. Chem. 90
Chem. 107 (1980) 201. (1986) 6227.
[33] B. Reichman, A.J. Bard, D. Laser, J. Electrochem. Soc. 127 [37] P. Ugo, L.M. Moretto, D. Rudello, E. Birriel, J. Chevalet,
(1980) 647. Electroanalysis 13 (2001) 661.
[34] A.A.M. Wide, G. Aksnes, Cardiovasc. Res. 29 (1995) 1. [38] P.J. Kinlen, J.E. Heider, D.E. Hubbard, Sens. Actuators B
[35] J.N. Weiss, R.C. Shein, Cardiovasc. Res. 28 (1994) B22 (1994) 13.
1125. [39] L. Alan, N.D. Miller, Alt. Med. Rev. 1 (2) (1996) 103.

You might also like