You are on page 1of 60

DxC 700 AU

Job Aid Booklet

For Training Purposes Only


These job aids are shortened versions of procedures found in the references below.
Information in the job aid is correct as of the date published. Verify you have the correct
information.

References:
 Instructions for Use DxC 700 AU Chemistry Analyzer PN B71495AC (October 2017)
 Reference Manual DxC 700 AU Chemistry Analyzer PN B71496AB (March 2017)
 DxC 700 AU Instructions for Use Addendum: Additional Instructions for Replace or Add
Reagents in MEASURE mode C07009-AB (February 2017)
 DxC 700 AU Maximum Sample Volume Limits in Sample Tubes and Cups in Racks
C02931AB (February 2017)

DxC 700 AU, Customer, Job Aid Booklet Page 1 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


DxC 700 AU Chemistry Analyzer

WARNINGS AND PRECAUTIONS

Read all product manuals and consult with Beckman Coulter-trained personnel before
attempting to operate the instrument.

Beckman Coulter, Inc. urges its customers and employees to comply with all national
health and safety standards such as the use of barrier protection. This may include,
but is not limited to, protective eyewear, gloves, suitable laboratory attire when
operating or maintaining this or any other automated laboratory equipment.

INTENTION FOR USE

This document is not intended to replace the information in your Instructions for
Use manual or Reference Manual. Information in the Instructions for Use and
Reference Manual supersedes information in any other manual.

REVISION STATUS

Version 1.07 (June 2018)


Based on DxC 700 AU software version 1.07

TRADEMARKS

DxC 700 AU Chemistry Analyzer

DxC 700 AU, Customer, Job Aid Booklet Page 2 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Table of Contents

Create a New Index......................................................................................................... 5 


Inspect the Analyzer Status............................................................................................. 6 
Check and Load Reagents .............................................................................................. 7 
Load a Fixed (Assigned) Reagent ................................................................................... 8 
Load Reagents in MEASURE Mode................................................................................ 9 
Perform Analyzer Calibration ........................................................................................ 11 
Perform Quality Controls for All Tests ........................................................................... 12 
Perform Quality Control for Selected Tests ................................................................... 13 
Sample Processing Using Racks .................................................................................. 14 
Manual Programming for Samples on Racks ................................................................ 15 
Perform Add On and Rerun Tests on Racks ................................................................. 16 
Sample Processing Using the STAT Table ................................................................... 17 
Manual Programming for Samples on the STAT Table ................................................. 18 
Perform Add On and Rerun Tests on the STAT Table .................................................. 19 
Program a Batch Order ................................................................................................. 20 
Review Patient Results from Sample Status ................................................................. 21 
Recall Patient Results ................................................................................................... 22 
Print RB/CAL/QC Reports ............................................................................................. 23 
Transfer Data to Host .................................................................................................... 24 
Program Replicates ....................................................................................................... 25 
Review Calibration Monitor ........................................................................................... 27 
Review and Print QC ..................................................................................................... 28 
Delete QC Data and Enter Comments .......................................................................... 29 
Update Calibrator Lot Concentrations ........................................................................... 30 
Configure New QC Lot .................................................................................................. 31 
Setup Name/Initials for the Maintenance Log ................................................................ 32 
Setup a Patient Sample Panel ...................................................................................... 33 
Setup a QC Panel ......................................................................................................... 34 
Edit a User Menu .......................................................................................................... 35 
Change the User (Logout) ............................................................................................. 36 
Program the Operator Name for the Maintenance Log ................................................. 37 
Daily Analyzer Inspections and Preparations ................................................................ 38 

DxC 700 AU, Customer, Job Aid Booklet Page 3 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Update the Maintenance Log ........................................................................................ 40 
Add Supervisor Approval to the Maintenance Log ........................................................ 41 
Add a Maintenance Procedure to the Maintenance Log................................................ 42 
Load ISE Reagents (ISE Option only) ........................................................................... 43 
Load ISE Reagents in MEASURE mode (ISE Option only)........................................... 44 
Perform ISE Clean (ISE Option only) ............................................................................ 45 
Perform ISE Calibration (ISE Option only) .................................................................... 46 
Save or Load System Parameters ................................................................................ 47 
Save Patient Data to External Media............................................................................. 48 
Setup Auto Startup ........................................................................................................ 49 
Perform an End Process ............................................................................................... 50 
Perform and Recover from an ANALYZER STOP ........................................................ 51 
Perform Analyzer Calibration from the STAT Table ..................................................... 52 
Perform Quality Control from the STAT Table ............................................................... 53 
Sample Processing Using the STAT Table with Laboratory Automation System .......... 54 
Order and Perform Analyzer Calibration and QC with Laboratory Automation System
(Multiple Loads STAT Table) ......................................................................................... 55 
ISE Calibration Failure Troubleshooting ........................................................................ 57 
Rack Jam Troubleshooting............................................................................................ 58 
Reagent Level Detection Error Troubleshooting ........................................................... 59 
RB, CAL, QC Troubleshooting ...................................................................................... 60 

DxC 700 AU, Customer, Job Aid Booklet Page 4 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Create a New Index

DxC 700 AU, Customer, Job Aid Booklet Page 5 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Inspect the Analyzer Status

DxC 700 AU, Customer, Job Aid Booklet Page 6 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Check and Load Reagents

DxC 700 AU, Customer, Job Aid Booklet Page 7 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Load a Fixed (Assigned) Reagent

DxC 700 AU, Customer, Job Aid Booklet Page 8 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Load Reagents in MEASURE Mode

DxC 700 AU, Customer, Job Aid Booklet Page 9 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Load Reagents in MEASURE Mode, continued

DxC 700 AU, Customer, Job Aid Booklet Page 10 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform Analyzer Calibration

DxC 700 AU, Customer, Job Aid Booklet Page 11 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform Quality Controls for All Tests

DxC 700 AU, Customer, Job Aid Booklet Page 12 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform Quality Control for Selected Tests

DxC 700 AU, Customer, Job Aid Booklet Page 13 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Sample Processing Using Racks

DxC 700 AU, Customer, Job Aid Booklet Page 14 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Manual Programming for Samples on Racks

DxC 700 AU, Customer, Job Aid Booklet Page 15 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform Add On and Rerun Tests on Racks

DxC 700 AU, Customer, Job Aid Booklet Page 16 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Sample Processing Using the STAT Table
Notes: This procedure START HERE
applies when:
STAT Table Settings are
set to Bar Code
Auto Rerun for the STAT
table is set to Disabled Sufficient
Multiple Loads STAT volume in No Transfer sample into a
table operation is primary sample validated low volume container
Disabled tube?

Yes
Select STAT Status
Select STAT Pause (F2)
STAT TABLE
Select OK Yes
ROTATION LED
Wait for the STAT TABLE
blinking?
ROTATION LED to stop
blinking
No

Open the small STAT table cover

Note: If STAT PAUSE (F2) was selected to load samples, then


the new samples need to be placed in any position with a
higher position number than those for the existing samples.

Place each sample in a STAT table position:

For the correct container size (tube adapters are for either
13 mm or 16 mm diameter containers)
With the bar code facing out
With caps removed

Close the small STAT table cover

Press the green STAT START button on the analyzer.

The system starts analysis and the STAT TABLE


ROTATION LED begins to blink

To monitor sample status on the analyzer:


Select RESULT
Select:
- A sample ID and then Detail to view the tests
ordered, results, and flags for the selected sample
- Realtime List to view a list of completed samples
with test results

DxC 700 AU, Customer, Job Aid Booklet Page 17 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Manual Programming for Samples on the STAT Table

Notes: This procedure Manually order a sample:


applies when:
STAT Table Settings are 1. Select STAT > STAT (Patient) > Test Order
set to Bar Code 2. In Type, select the sample type
Auto Rerun for the STAT
3. Enter the bar code number in Sample ID
table is set to Disabled
Multiple Loads STAT
4. Select each test individually or select Panel to select a panel, and then
table operation is select OK. Selected tests are highlighted in blue
Disabled 5. If a manual dilution was made, enter the dilution rate in the Sample
Dilution Rate
6. Select the Demographics tab to enter required patient demographic
information
7. Select Save (F2)
8. To order more samples, select the Test Order tab, repeat steps 2 to 7

Select Close (F1)


Optional: Select Pending List (F4) to view a list of samples ordered
- Select Close to close the Pending List, or
- Select the sample from the Pending List and select Jump to the
Selected Sample to view the specific sample order

Select STAT Status


Select STAT Pause (F2) Yes STAT TABLE
Select OK ROTATION
Wait for the STAT TABLE LED blinking?
ROTATION LED to stop blinking

No

Open the small STAT table cover


Load samples in the appropriate available positions with the bar codes
facing out. Press the green DIAG/TABLE ROTATION button to rotate
the STAT table to the desired position
Close the small STAT table cover

Press the green STAT START button on the analyzer.

The system starts analysis and the STAT TABLE ROTATION LED
begins to blink

DxC 700 AU, Customer, Job Aid Booklet Page 18 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform Add On and Rerun Tests on the STAT Table

Notes: This procedure applies when:


A STAT sample requires an add on or rerun test to be processed on the STAT
table using the same sample ID and index as the original sample
STAT Table Settings are set to Bar Code
Auto Rerun for the STAT table is set to Disabled
Multiple Loads STAT table operation is Disabled

Select STAT > STAT (Patient) > Test Order


Select Search by Sample ID to enter the sample ID, select Search

Order an add on or rerun test:

Select Rerun (F3). The system displays the first run tests, identified by
an *, with results and flags (if any)
Select the test(s) to add on or rerun. Selected tests are highlighted in blue
(for normal analysis)
Select OK

Optional: select Pending List (F4) to view a list of ordered samples:


Select Close to close the Pending List, or
Select the sample from the Pending List and select Jump to the
Selected Sample to view the specific sample order

Select STAT Status


Select STAT Pause (F2) Yes STAT TABLE
Select OK ROTATION LED
Wait for the STAT TABLE blinking?
ROTATION LED to stop blinking

No

Open the small STAT table cover


Load samples in the appropriate available positions with the bar codes
facing out. Press the green DIAG/TABLE ROTATION button to rotate the
STAT table to the desired position
Close the small STAT table cover

Press the green STAT START button on the analyzer

The system starts analysis and the STAT TABLE ROTATION LED
begins to blink

DxC 700 AU, Customer, Job Aid Booklet Page 19 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Program a Batch Order

DxC 700 AU, Customer, Job Aid Booklet Page 20 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Review Patient Results from Sample Status

DxC 700 AU, Customer, Job Aid Booklet Page 21 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Recall Patient Results

DxC 700 AU, Customer, Job Aid Booklet Page 22 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Print RB/CAL/QC Reports

DxC 700 AU, Customer, Job Aid Booklet Page 23 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Transfer Data to Host

DxC 700 AU, Customer, Job Aid Booklet Page 24 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Program Replicates

DxC 700 AU, Customer, Job Aid Booklet Page 25 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Program Replicates, continued

DxC 700 AU, Customer, Job Aid Booklet Page 26 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Review Calibration Monitor

DxC 700 AU, Customer, Job Aid Booklet Page 27 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Review and Print QC

DxC 700 AU, Customer, Job Aid Booklet Page 28 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Delete QC Data and Enter Comments

DxC 700 AU, Customer, Job Aid Booklet Page 29 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Update Calibrator Lot Concentrations

DxC 700 AU, Customer, Job Aid Booklet Page 30 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Configure New QC Lot

DxC 700 AU, Customer, Job Aid Booklet Page 31 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Setup Name/Initials for the Maintenance Log

DxC 700 AU, Customer, Job Aid Booklet Page 32 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Setup a Patient Sample Panel

DxC 700 AU, Customer, Job Aid Booklet Page 33 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Setup a QC Panel

DxC 700 AU, Customer, Job Aid Booklet Page 34 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Edit a User Menu

DxC 700 AU, Customer, Job Aid Booklet Page 35 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Change the User (Logout)

DxC 700 AU, Customer, Job Aid Booklet Page 36 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Program the Operator Name for the Maintenance Log

DxC 700 AU, Customer, Job Aid Booklet Page 37 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Daily Analyzer Inspections and Preparations

DxC 700 AU, Customer, Job Aid Booklet Page 38 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Daily Analyzer Inspections and Preparations, continued

DxC 700 AU, Customer, Job Aid Booklet Page 39 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Update the Maintenance Log

DxC 700 AU, Customer, Job Aid Booklet Page 40 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Add Supervisor Approval to the Maintenance Log

DxC 700 AU, Customer, Job Aid Booklet Page 41 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Add a Maintenance Procedure to the Maintenance Log

DxC 700 AU, Customer, Job Aid Booklet Page 42 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Load ISE Reagents (ISE Option only)

DxC 700 AU, Customer, Job Aid Booklet Page 43 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Load ISE Reagents in MEASURE mode (ISE Option only)

DxC 700 AU, Customer, Job Aid Booklet Page 44 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform ISE Clean (ISE Option only)

DxC 700 AU, Customer, Job Aid Booklet Page 45 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform ISE Calibration (ISE Option only)

DxC 700 AU, Customer, Job Aid Booklet Page 46 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Save or Load System Parameters

DxC 700 AU, Customer, Job Aid Booklet Page 47 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Save Patient Data to External Media

DxC 700 AU, Customer, Job Aid Booklet Page 48 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Setup Auto Startup

Select CONFIG. > System Setup

Select Edit (F1)

Select Auto Power On Setup (F7)

Select the box for the desired days to start the Auto
Power On function

Select the hours and minutes for each day from the
drop down menus

Note: The system has to be setup for Auto Preparation


(W1 and/or Photocal) during instrument installation.

Select OK

Select Save (F1)

DxC 700 AU, Customer, Job Aid Booklet Page 49 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform an End Process

DxC 700 AU, Customer, Job Aid Booklet Page 50 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform and Recover from an ANALYZER STOP

DxC 700 AU, Customer, Job Aid Booklet Page 51 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform Analyzer Calibration from the STAT Table
Notes: This procedure applies Select STAT > STAT (Calibration)
when:
STAT Table Settings are set to
Bar Code
Auto Rerun for the STAT table is In Type, select the sample type
set to Disabled Select Auto CAL/QC Order (F4)
Multiple Loads STAT table Select OK. The system displays tests for
operation is Disabled
calibration and reagent blank in blue.

Automatic order
Yes
correct for all sample
types?

No

Note: Select QC Manually order calibration:


Same Order (F5) if
you would like to
To select a panel, select Panel. Select a panel, and select OK
process QC for the
same tests as To select a specific test, select the test from the RB or Cal. column
calibration Optional: Select Individual Order (F3) to order sequential bottles of the
same test. Make appropriate test selections, select Close to save the order
In Type, select the sample type to order reagent blank and calibration for
other sample types
Select Save (F2)

Select CAL/QC Position (F6) to view the calibrators required


Pour reagent blank and calibrators in the appropriate bar code
labeled containers
Select Close to close the Cal/QC Position dialog

Select STAT Status


Select STAT Pause (F2) STAT
Yes
Select OK TABLE
Wait for STAT TABLE ROTATION LED
ROTATION LED to stop blinking?
blinking

No

Open the small STAT table cover. Press the green


DIAG/TABLE ROTATION button to rotate the table, if necessary
Place the reagent blank (DI water) in the position labeled RB1
Load required calibrators in any available positions with bar codes facing out
Close the small STAT table cover

Select STAT Start from the action button area


Review errors on the Error List in the STAT Start dialog and perform any
corrective actions, if necessary
Select Start from the STAT Start dialog

DxC 700 AU, Customer, Job Aid Booklet Page 52 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Perform Quality Control from the STAT Table
Notes: This procedure applies Select STAT > STAT (QC)
when:
STAT Table Settings are set to
Bar Code
Auto Rerun for the STAT table is
set to Disabled
Multiple Loads STAT table In Type, select the sample type
operation is Disabled

Manually order QC:

To select a panel, select Panel. Select a panel, and select OK


To select a specific test, select the test
Optional: To order sequential bottles of the same test:
If Multiple Bottles is selected in the QC Program by Panel area, the
system automatically programs the same QC order for all bottles for the
test. No further action is required
If First Bottle Only is selected in the QC Program by Panel area, select
Individual Order (F3) to order sequential bottles of the same test.
Make the appropriate test selections, select Close to save the order
In Type, select the sample type to order QC for other sample types
Select Save (F2)

Select CAL/QC Position (F7) to view the control samples required


Note: Calibrators will also display in the CAL/QC Position dialog when
calibration is ordered.
Pour QC in appropriate bar code labeled containers
Select Close to close the CAL/QC Position dialog

Select STAT Status


Select STAT Pause (F2) STAT
Yes
Select OK TABLE
Wait for STAT TABLE ROTATION LED
ROTATION LED to stop blinking?
blinking
No

Open the small STAT table cover. Press the green


DIAG/TABLE ROTATION button to rotate the table, if necessary
Load required QC in any available positions with bar codes facing out
Close the small STAT table cover

Select STAT Start from the action button area


Review errors on the Error List in the STAT Start dialog and perform any
corrective actions, if necessary
Select Start from the STAT Start dialog

DxC 700 AU, Customer, Job Aid Booklet Page 53 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Sample Processing Using the STAT Table with
Laboratory Automation System
Notes: This procedure applies START HERE
when:
Multiple Loads STAT table
operation is Enabled
STAT Table Settings are
set to Bar Code
Auto Rerun for the STAT
Sufficient No
table is set to Disabled volume in Transfer sample into a
primary sample validated low volume container
tube?

Yes
Select the Pause button
from the action button area
Yes
STAT TABLE
ROTATION LED
blinking?
Select STAT > STAT Status
Select STAT Pause (F2)
Select OK
Wait for the STAT TABLE
ROTATION LED to stop No
blinking

Open the small STAT table cover

Note: If STAT PAUSE (F2) was selected to load samples, then


the new samples need to be placed in any position with a
higher position number than those for the existing samples.

Place each sample in a STAT table position:

For the correct container size (tube adapters are for either
13 mm or 16 mm diameter containers)
With the bar code facing out
With caps removed

Close the small STAT table cover

Press the green STAT START button on the analyzer.

The system starts analysis and the STAT TABLE


ROTATION LED begins to blink

To monitor sample status on the analyzer:


Select RESULT
Select:
- A sample ID and then Detail to view the tests
ordered, results, and flags for the selected sample
- Realtime List to view a list of completed samples
with test results

DxC 700 AU, Customer, Job Aid Booklet Page 54 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Order and Perform Analyzer Calibration and QC with Laboratory
Automation System (Multiple Loads STAT Table)
Notes:
This procedure applies to labs connected to an automation line. Calibrators and QC samples must be
processed on the STAT table using bar codes. The system provides a Multiple Loads STAT Table operation
to allow for up to 200 calibrators and 100 control samples to be processed in a single run. Use this
procedure when more than 22 calibrators and control samples are required in a single run. Perform
the procedures in the following order:
Order the Calibrations and QCs
Prepare and Load Calibrators and Control Samples on the STAT Table. Note: All calibrators must be
loaded before QC samples
Analyze the first load of Calibrators or Control Samples
Place Another Load of Calibrators or Control Samples on the STAT Table
Analyze the Last Load of Calibrators or Control Samples

All of the above procedures are included in this flowchart.

Select STAT > STAT (Calibration)

Order Calibrations
In Type, select the sample type
Select Auto CAL/QC Order F4. The system automatically orders calibration for the
following conditions:
- The reagent blank or calibration is expired
- The reagent blank or calibration expires soon
- There is no reagent blank or calibration data available
- The reagent blank or calibration failed
From the STAT (Calibration) dialog, select OK. Verify the order is correct. Select Type, to
verify orders of all sample types in use

No
Select CAL/QC
QC Required?
Position (F6)

Yes

Order QC
Select the STAT (QC) button
- In Type, select the sample type
- Select Select All Tests (F5), select individual tests or select Panel to select a panel,
and then select OK. Select Type to repeat for each sample type in use
- Select Save (F2)
- Select CAL/QC Position (F7). Continue with prepare RB, CAL and QC

Prepare RB, CAL and QC


Use the CAL/QC Position list to prepare the required reagent blank, calibrator and/or QC
samples. Use the list to ensure you use the correct bar coded tubes and required volume
(dead volume of the specific container is not included) for each sample. Select Close

Continue on the next page

DxC 700 AU, Customer, Job Aid Booklet Page 55 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Order and Perform Analyzer Calibration and QC with Laboratory
Automation System (Multiple Loads STAT Table) continued
Continued from the previous page

Yes Note:
Patient samples can be
Select the Pause button STAT
interrupted to load
from the action button area TABLE ROTATION LED
calibrators and/or QC.
blinking?
Remove the patient
sample(s) and reload
them in the same
Select STAT > STAT Status positions on the STAT
Select STAT Pause (F2) table when calibration
Select OK and QC completes.
Wait for the STAT TABLE No
ROTATION LED to stop blinking

Open the upper cover


Remove the large STAT table cover
Place the first load of calibrators and/or control samples (maximum of 22) with the bar codes facing
out. Calibrators must be loaded before QC. The reagent blank must be loaded in the RB1 position
Replace the large STAT table cover
Select STAT Start from the action button area

1. From the STAT Start dialog, review errors on the Error List and
perform any corrective actions, if necessary
2. Select Start
3. When the STAT TABLE ROTATION LED stops blinking, add
more calibrators and/or QC samples by performing either option
a or b listed below:

a. Remove the large STAT table cover. Pull up the two white
latches in the STAT table compartment until you hear a
clicking noise. Lift the STAT table out by using the knob.
Is this the last No
Set the next STAT table containing another load of calibrator
load of calibrators
and QC samples (calibrators must be loaded before QC) in
and/or QC
the STAT table compartment using the positioning pin as a
samples?
guide. Push down the two white latches until they make a
clicking noise to secure the STAT table. Replace the large
Yes STAT table cover

b. Open the small STAT table cover and remove all the
completed samples. Add more calibrator and QC samples
From the STAT Start dialog, review
(calibrators must be loaded before QC). Close the small
errors on the Error List and perform
STAT table cover
any corrective actions, if necessary
Select Last Table box in the dialog
4. Select STAT Start from the action button area
Select Start

When the STAT TABLE ROTATION Yes Is this the last No


LED stops blinking the aspiration is load of calibrators Repeat starting
complete. Remove all the and/or QC at step 3.
calibrator and/or QC samples samples?

DxC 700 AU, Customer, Job Aid Booklet Page 56 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


ISE Calibration Failure Troubleshooting

DxC 700 AU, Customer, Job Aid Booklet Page 57 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Rack Jam Troubleshooting

DxC 700 AU, Customer, Job Aid Booklet Page 58 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


Reagent Level Detection Error Troubleshooting

DxC 700 AU, Customer, Job Aid Booklet Page 59 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine


RB, CAL, QC Troubleshooting

DxC 700 AU, Customer, Job Aid Booklet Page 60 of 60


Version 2.0 (June 2018) Beckman Coulter Technical Product Training

Downloaded from www.Manualslib.com manuals search engine