DXC 700 Au
DXC 700 Au
References:
Instructions for Use DxC 700 AU Chemistry Analyzer PN B71495AC (October 2017)
Reference Manual DxC 700 AU Chemistry Analyzer PN B71496AB (March 2017)
DxC 700 AU Instructions for Use Addendum: Additional Instructions for Replace or Add
Reagents in MEASURE mode C07009-AB (February 2017)
DxC 700 AU Maximum Sample Volume Limits in Sample Tubes and Cups in Racks
C02931AB (February 2017)
Read all product manuals and consult with Beckman Coulter-trained personnel before
attempting to operate the instrument.
Beckman Coulter, Inc. urges its customers and employees to comply with all national
health and safety standards such as the use of barrier protection. This may include,
but is not limited to, protective eyewear, gloves, suitable laboratory attire when
operating or maintaining this or any other automated laboratory equipment.
This document is not intended to replace the information in your Instructions for
Use manual or Reference Manual. Information in the Instructions for Use and
Reference Manual supersedes information in any other manual.
REVISION STATUS
TRADEMARKS
Yes
Select STAT Status
Select STAT Pause (F2)
STAT TABLE
Select OK Yes
ROTATION LED
Wait for the STAT TABLE
blinking?
ROTATION LED to stop
blinking
No
For the correct container size (tube adapters are for either
13 mm or 16 mm diameter containers)
With the bar code facing out
With caps removed
No
The system starts analysis and the STAT TABLE ROTATION LED
begins to blink
Select Rerun (F3). The system displays the first run tests, identified by
an *, with results and flags (if any)
Select the test(s) to add on or rerun. Selected tests are highlighted in blue
(for normal analysis)
Select OK
No
The system starts analysis and the STAT TABLE ROTATION LED
begins to blink
Select the box for the desired days to start the Auto
Power On function
Select the hours and minutes for each day from the
drop down menus
Select OK
Automatic order
Yes
correct for all sample
types?
No
No
Yes
Select the Pause button
from the action button area
Yes
STAT TABLE
ROTATION LED
blinking?
Select STAT > STAT Status
Select STAT Pause (F2)
Select OK
Wait for the STAT TABLE
ROTATION LED to stop No
blinking
For the correct container size (tube adapters are for either
13 mm or 16 mm diameter containers)
With the bar code facing out
With caps removed
Order Calibrations
In Type, select the sample type
Select Auto CAL/QC Order F4. The system automatically orders calibration for the
following conditions:
- The reagent blank or calibration is expired
- The reagent blank or calibration expires soon
- There is no reagent blank or calibration data available
- The reagent blank or calibration failed
From the STAT (Calibration) dialog, select OK. Verify the order is correct. Select Type, to
verify orders of all sample types in use
No
Select CAL/QC
QC Required?
Position (F6)
Yes
Order QC
Select the STAT (QC) button
- In Type, select the sample type
- Select Select All Tests (F5), select individual tests or select Panel to select a panel,
and then select OK. Select Type to repeat for each sample type in use
- Select Save (F2)
- Select CAL/QC Position (F7). Continue with prepare RB, CAL and QC
Yes Note:
Patient samples can be
Select the Pause button STAT
interrupted to load
from the action button area TABLE ROTATION LED
calibrators and/or QC.
blinking?
Remove the patient
sample(s) and reload
them in the same
Select STAT > STAT Status positions on the STAT
Select STAT Pause (F2) table when calibration
Select OK and QC completes.
Wait for the STAT TABLE No
ROTATION LED to stop blinking
1. From the STAT Start dialog, review errors on the Error List and
perform any corrective actions, if necessary
2. Select Start
3. When the STAT TABLE ROTATION LED stops blinking, add
more calibrators and/or QC samples by performing either option
a or b listed below:
a. Remove the large STAT table cover. Pull up the two white
latches in the STAT table compartment until you hear a
clicking noise. Lift the STAT table out by using the knob.
Is this the last No
Set the next STAT table containing another load of calibrator
load of calibrators
and QC samples (calibrators must be loaded before QC) in
and/or QC
the STAT table compartment using the positioning pin as a
samples?
guide. Push down the two white latches until they make a
clicking noise to secure the STAT table. Replace the large
Yes STAT table cover
b. Open the small STAT table cover and remove all the
completed samples. Add more calibrator and QC samples
From the STAT Start dialog, review
(calibrators must be loaded before QC). Close the small
errors on the Error List and perform
STAT table cover
any corrective actions, if necessary
Select Last Table box in the dialog
4. Select STAT Start from the action button area
Select Start