Check data
Check data using control
before ☐
blood plasma
maintenance
Transducer aperture cleaned Once a year ☐
RBC Detector Draining isolation chamber Once a year ☐
Block
Clean inside. Once a year ☐
2
(First Edition)
3
(First Edition)
【Details of Adjustment Items】
Adjustment Items Specified Range Value Before Adjustment Value After Adjustment
CLOG LEVEL 100 +/- 1
ADJUSTMENT
HGB background convert 5000 +/- 200
blood aspiration sensor 5000 +/- 200
blank
Blood Asp sensor
13000 +/- 500
Sample - blank
Sensitivity Adjustment XN-CAL Target Value (Lot No. )
WDF-X Target value +/-3ch
WDF-Y Target value +/-3ch
WDF-FSC Target value +/-2h
RET-RBC-X Target value +/-3ch
RET-RBC-Y Target value +/-3ch
RET-RBC-Z Target value +/-2ch
MCV Target value +/-1fL
MPV Target value +/-3fL
HGB Target value +/- 0.2g/dL
Maintenance
Report No.
Facility Name
Instrument Name XN-350/XN-450/XN-550 Serial No.
Maintenance
Date
Name of Service
Engineer
4
(First Edition)
XN-L Series PM Standard Procedure
[Required jigs and controls, etc.]
Code Name Remarks
- XN CHECK Control blood
- XN CHECK BF Control blood for BF
- XN CAL Calibration by Calibrator
Operatio
Items to be checked Procedure
n time
6
(First Edition)
XN-L Series Periodic Replacement Parts List
7
(First Edition)
Revision History