XN-L Training &

Service & Support Global Meeting 2014


Service program comparison between
XN-L and XN-10/20          
Tadashi Hidaka
Sysmex Corporation
Contents

• Differences between XN-L and XN-10/20


• Details

XN-L Training & SSGM 2014 2


Service menu comparison
Service menu XN-L XN-10/20
Mode ✓ ✓
Aspiration Sensor ✓ ✓
Sensitivity ✓ ✓
A/D Converter Offset (Auto) ✓ ✓
A/D Converter Offset (Manual) ✓ ✓
APD Parameter Setting NA ✓
Permanent Counter ✓ ✓
Optical Axis Adjustment (rough) ✓ ✓
Optical Axis Adjustment (fine) ✓ ✓
Laser Setting ✓ ✓
Laser Power ✓ ✓

Reason
 APD is not used on XN-L

XN-L Training & SSGM 2014 3


Service menu comparison
Service menu XN-L XN-10/20
SV Test ✓ ✓
Piercer/Pipette Adjustment ✓ ✓
Tube Holder Adjustment ✓ ✓
Hand Adjustment ✓ ✓
Piercer/Pipette ✓ ✓
Hand ✓ ✓
Power ✓ ✓
Diaphragm Pump ✓ ✓
RBC Clogs ✓ ✓
LED Adjustment NA ✓

Reason
 Module concept is not used, then no need to adjust LED color
between two analyzes

XN-L Training & SSGM 2014 4


Service menu comparison
Service menus XN-L XN-10/20
Shipping1 ✓ ✓
Shipping2 ✓ ✓
Shipping3 ✓ ✓
Setting Sequence ✓ ✓
Sampler Positions Adjustment NA ✓
Sampler Power NA ✓
Blood Sensor NA ✓

Reasons
 Sampler type is different between XN-L and XN-10/20.
 XN-L does not have separate switching regulator in sampler.
 XN-L does not have Blood sensor function  

XN-L Training & SSGM 2014 5


Adjustment using service program
• Aspiration sensor
– Target value is different to
improve reproducibility
 XN-L: 13000 (Lower dilution ratio)
 XN-10/20: 6500

• Calibrator calibration
– Only XN CAL is used,
because XN-L cannot
measure PLT-F
– MCV compensation is not
implemented yet
XN-L Training & SSGM 2014 6
Service setting
• Service setting menu tree is different
XN

XN-L

Service Setting

XN-L Training & SSGM 2014 7


Service setting
• New service settings
– Laboratory only tab ON/OFF setting

XN-L Training & SSGM 2014 8


Restore
• Restore function is implemented
– “mmb” file is always saved in D:\Programfiles\Sysmex\
NVRAMfile folder
– The file copies to LOG folder when executing shutdown
– From tool bar, mmb file can be restored.

XN-L Training & SSGM 2014 9


Thank you for your kind attention.

End

XN-L Training & SSGM 2014 10