Professional Documents
Culture Documents
Inline
USA: 1-800-562-7024
Canada: 1-800-655-5585
Telephone: +336 969-6909
FAX: +336 969-6375
Email: info@wagerusa.com
Homepage: http://www.wagerusa.com
Please detach this form (or make a copy) and mail to:
Warranty Registration
Robert H. Wager Co., Inc.
570 Montroyal Road
Rural Hall, NC 27045
Registration of Wager Model 6500 Smoke (Opacity) Meter
(We would appreciate very legible writing)
Name of User:
Company:
Address:
City/ZIP:
Telephone:
Fax:
Email:
Date of Purchase:
Purchased from:
Serial Number:
Optional Information:
To anticipate your needs, we would appreciate knowing more about your operation:
Robert H. Wager Co., Inc.’s full, legal warranty statement is printed on the following page. Here is
a quick overview:
Robert H. Wager Co., Inc. will make the appropriate repairs as rapidly as
possible (usually within three working days) and return the unit freight prepaid.
The level of freight service will mirror the customer’s inbound freight (i.e.,
expedited “red” or “blue” shipments will be returned the same way).
If Robert H. Wager Co., Inc. finds—at its sole discretion—that the
malfunction was due to customer abuse, misuse, attempts to repair, or
other conditions caused solely by the customer, the warranty is void. The
unit will be repaired in the normal manner and the customer is invoiced.
In order to establish the date of your purchase and match the serial
number of the unit to our records, we require the attached form “Warranty
Registration” to be filled out and returned to us.
We reserve the right to alter, amend, or adjust our warranty policy without
further notice.
If it appears within one year from date of shipment by Seller that the equipment does
not meet these express warranties and Buyer gives Seller prompt and reasonable
notice, Seller shall, at its option, either repair or replace at its expense, F.O.B. Seller’s
works, but not dismantle or reinstall, the defective parts provided, upon request such
parts are shipped freight prepaid to Seller’s works.
These warranties shall not apply if equipment is subjected to other than normal and
proper storage, handling, installation, operation and maintenance or to unauthorized
repairs or alterations. Equipment, components and accessories made by other
manufacturers are warranted only to the extent of the original manufacturer’s warranties
to Seller.
The foregoing warranty obligation of the seller shall constitute the sole and exclusive
remedy of the buyer and the sole liability of the seller, except as set forth herein and
except as to the title it is expressly agreed (a) that there is no warranty of
merchantability of any other warranty, express, implied or statutory, nor any affirmation
of fact or promises by Seller with respect to the equipment or parts or otherwise which
extend beyond the specifications mutually agreed upon in writing by Seller and Buyer,
and (b) that the Buyer acknowledges that it is purchasing the equipment solely on the
basis of the commitments of Seller expressly set forth herein, in no event shall Seller be
liable for special, indirect, or consequential damages including, without limitation,
anticipate profits.
The WAGER Model 6500 Smoke Meter is a versatile system capable of usage in many
different applications.
The meter's simple design makes it easy to obtain consistent and accurate
measurements. It provides an accurate means (± 1%) of detecting and measuring the
opacity of smoke emitted by any emissions source.
System Description
Control Unit
The control unit is operated by a membrane keypad, which consists of eight
tactile feedback push buttons.
The banana jacks (red and black) allow a 0 –1 VDC output for a chart recorder
or Data Acquisition System.
Sensing Device
The sensing device is a multi-part assembly which varies based on the individual
installation. Each sensing device consists of a Transmitting and Receiving Unit,
and conventionally an inline stack piece.
Cable
The connecting cable provides signal connection from the Control Unit to each
sensor.
Power
The meter is shipped with an AC Adapter/Charger which operates in 110V (220V
Export Version).
Design Criteria
Accuracy
The unit’s accuracy is ± 1.0% nominal ± one digit. The control unit is initially
calibrated under clear stack conditions with checks at 0% and 100% opacity.
Ambient Conditions
The unit operates in 32° to 120° F (0° to 50° C).
Reliability
Zero stability at less than 1% drift per use.
The pulsed green LED has infinite life expectancy.
All solid state electronics.
Applications
The Model 6500 Smoke Meter can be used on any emissions source, with
primary application in testing trucks, busses, cars, generators and boilers.
Functional Description
When the unit is activated, the transmitter emits a light beam at a known
intensity. As the light passes through the smoke plume, some or all the light is
diverted or absorbed.
The sensor head consists of two components, a light source (transmitter) and a
light detector (receiver).
The light detector measures the amount of light received, and compares this
with the amount of light being emitted. The difference between the two values is
opacity.
Attachment of Sensors
Each Sensor is attached to the inline stack piece by threading the female air (or
water chamber onto its mating connection on the stack piece.
Align the pins on a connector with the holes on its mating receptacle. Insert the
connector into its mating receptacle.
Lock the connector in place by screwing the locking ring clockwise until a firm
connection is made.
Press the ON/OFF button on the control unit. The unit will perform a self-
calibration.
Press the SELECT button until the desired stack size is displayed (factory default
is 5”). Press the SAVE button.
The real-time opacity measurement will appear in the upper left area of the
display.
The maximum opacity measurement will appear in the upper right area of the
display. (This value may be reset at any time by pressing the START TEST
button.)
Battery Replacement
Remove the two screws and carefully lift off the top to expose the attached
wires. Remove all four connectors from the front panel. Place the front panel
aside.
Turn the unit over. Remove the four screws holding the battery bracket in place.
Remove the two connectors from each end of the battery. Place the new battery
back and screw the holding bracket back into place over the battery.
Re-attach the four front panel connectors. Replace the front panel the right and
left screws.
Cleaning Lenses
To assure regular easy access to the lenses, we have provides two snap
closures. The lenses can be wipes with a soft cloth.
To remove soot, dip soft cloth in denatured rubbing alcohol (make sure to let it
dry) and rub gently over lenses.
In laboratory settings and heavy commercial use, please note that we have
provided connectors for an external air purge.
Connecting Cables
Connecting cables are subject to wear, based on incorrect handling, abuse, and
contact with a hot stack. Check, repair, and replace cables when their condition
affects the operation of the smoke meter.
The WAGER Model 6500 Smoke Meter has been designed to be nearly trouble
free for many years of use.
Error Messages
CABLE FAILURE DISCONNECTED?
The light source is fully powered, but no light is detected receiver. This could be
caused by several factors:
• The sensor head assembly cable is not connected to the control unit
• The light source in the sensor head assembly has failed.
• The connecting cable is malfunctioning.
Action:
Check to make sure that the cable is firmly attached to the unit.
If another unit’s sensor head assembly and cable is available, try to see if this corrects
the error.
If the message disappears, then you need to send in the sensor head assembly and/or
cable for repair.
Contact WAGER Technical Services for further instructions and/or return of the unit for
repair.
Mechanical Specifications
Size Weight
Control Unit 4” w x 8" D x 4" H 5.0 lb.
Performance Specifications
Range: 0.0% - 100.0% opacity
Warm Up Time: Negligible
Response Time/Display*: 0.45 seconds for 0-90% opacity
Response Time (Chart Recorder 0.45 seconds for 0-90% opacity
Output)
Linearity ± 1% from 0-100% opacity
Temperature Stability, Sensor Heads ± 1% from 32° - 120° F
Battery Life 40 hours; 1 hour after low battery
indication, 8 hours to full charge
Battery Life (night light on) ½ of the above
Configuration Mode
In the configuration mode, the user can change the following items in the control
unit:
Time including hour, minutes, and format (12 hour or 24 hour clock)
Date including month, day, year, and format (American or European)
Time-out battery saving feature
Reset to factory defaults
Print current configuration
To access the “Configuration Mode” turn on the unit and press ZERO and START
TEST at the same time. Press these two buttons within 2 seconds.
Set Time:
Press the SELECT button. The LED displays the following:
ZERO= set time XXXXX (your current time setting)
Press ZERO to change time format (toggles between 12 and 24 hour clock).
Press the SELECT for “minute 10’s digit mode. (12:00 a.m.)
Press the ZERO to scroll to the correct minute 10’s digit.
Press the SELECT for “minute 1’s digit mode. (12:00 a.m.)
Press the ZERO to scroll to the correct minute 1’s digit.
Press ZERO to change date format (toggles between the formats dd/mm and
mm/dd).
Press ZERO to change time out feature (toggles between enabled and disabled).
Press ZERO to reset factory defaults (as you received the unit from WAGER).
Stack Piece
Due to the nature of the installation, a stainless steel Hex bar was selected to
serve as the support structure to mount the sensors. This bar is to be inserted
horizontally through the exhaust stack, and welded in place such that the milled
slot is oriented vertically (to allow the emissions to flow freely through the
measurement light path).
Sensors
Each sensor is a sealed assembly due to the outdoor location. Each sensor is
also supplied with a thermal isolation ring to prevent direct heat transfer from the
stainless steel air chamber.
Please note, since the insulation ring is of a ceramic type product (MACOR), it is
fragile and must not be dropped or impacted.