Professional Documents
Culture Documents
v
Standard Test Method for
Nitrogen in Petroleum and Petroleum Products by Boat-lnlet
Chemiluminescencel
This standard is issued under the fixed designation D5762; the number immediately following the designation inclicates the year of
original adoption or, in the case of revision. the year of last revision. A nunrber in parentheses indicates the year of last reapproval. A
superscript epsikrn (e) indicates an edikrrial change since the last fevision or reapproval.
I
Copyright by ASTM Int'l (all rights reserved); Mon Apr 19 03:46:56EDT 2010
Down loaded,/printed by
Francois Van+Dievoit 0 pursuant to License Agreement. No further reproductions authorized.
($þ osze2-oe
692mm
257mm
6mm O0 x 2mm lD
1Zmm
lZmm
35mm
25mm
50mm
-t 6Omm
6mm oD x zmm lD
Combustion tubes recommended for the two furnaces in 5.5.1 5.6 Microlitre Syringe, of 5 or 10-¡rL capacity, capable of
-1nd 5.5.2 are described in 5.3.I and 5.3.2. Other conlìgurations accurately delivering microlitre quantities.
are acceptable if precision and bias are not degraded. 5.7 Ozone Generator, to supply ozone to the detector
5.3.1 Quartz combustion tube for use with the single-zone reaction cell.
furnace is illustrated in Fig. l. A water-jacket around the inlet 5.8 Recorder (Optional), for display of chemiluminescence
section can be used to cool the boat prior to sample injection. detector signal.
5.3.2 Quaflz combustion tube for use with the two-zone
furnace is illustrated in Fig. 2. Fig. 3 illustrates a combustion 6. Reagents and Materials
tube for a two-zone furnace that is equipped with the ability to
6.1 Purin of Reagents-Reagent grade chemicals shall be
switch to a pure oxygen carier gas flow after the boat has been
used in all tests. Unless otherwise indicated, it is intended that
fully extended into the furnace (consult the instrument manual
all reagents shall conform to the specifications of the Commit-
to determine if the instrument changes to a pure oxygen canier
tee on Analytical Reagents of the American Chemical Society
gas flow after the boat is inserted). The oLrtlet end of the
where such specifications are available.-r Other grades may be
pyrolysis tube is constnrcted to hold a removable quartz insert
used, provided it is first ascertained that the reagent is of
tube. The removable quafiz insert tube is packed with an
sufficiently high purity to permit its use without lessening the
oxidation catalyst as recommended by the instrument manu-
accuracy of the determination.
factr¡rer.
6.2 Auidine, Cr3HeN, molecular weight 119.21,7.82 mass
5.4 Drier Tube, for the removal of water vapor. The reaction
(Warning-Irritant.)
7o nifro gen.
products include water vapor that shall be eliminated prìor to
measurement by the detector. This can be accomplished with a
6.3 Oxidation Ccrtcrlyst: Cupric Oxide Wire, CuO, or Plati-
rumt Ceúalyst, PrAI2O3, as recommended by the instnlment
magnesium perchlorate, Mg(ClOa)2, scrubber, a membrane
rrying tube permeation drier, or a chilled dehumidifier assem- manufacturer.
blv. 6.4 Inert Gcs-Argon or Helium only, high-purity grade
(that is, chromatographic or zero grade), 99.998 7o minimum
5.5 Furnace, Electric, held at a temperature sufficient to
purity, 5 ppm maximum moisture.
pyrolyze all of the sample and oxidize the nitrogen to NO. The
following furnace designs may be used. All furnace assemblies 6.5 Anhydror.ts Magnesium Perchlorate, Mg(ClOo)r, for
include a method for gas flow control, such as needle valves, drying products of combustion (if permeation drier or chilled
flow restrictors or mass flow controllers. Furnaces that are drier is not used). (Warning-Strong oxidizer, irritant.)
operated at temperatures below 1050"C shall be capable of 6.6 Nitrogen Stock Solution, 500 ng nitrogen/pL-
switching to 1007o oxygen carier gas flow after the boat has Accurately weigh (to the nearest 0.I mg) approximately 0.64 g
been fully extended into the furnace. of acridine into a tared 100-mL volumetric flask. Add xylene to
5.5.1 Single-zone tube furnace with temperature controller dissolve, then dilute to volume with xylene. Calculate the
nitrogen content of the stock solntion to the nearest milligram
capable of maintaining a stable furnace temperature of 1100 +
25'c. ofnitrogen per litre. This stock can be further diluted to desired
5.5.2 Two-zone tube furnace with temperature controllers
capable of maintaining the temperature of each furnace zone
independently from 950 to 1050'C (see 5.5). Or two-zone tube 3
Reugenl Clrcnùcctls, Anrcricut Chenical Soci¿ty Specilit:atlans , Anerican
furnace equipped with the ability to change to a pure oxygen Chemical Society, Washington, DC. For suggestions on the testing of reagents not
carier gas flow after the boat is fully extended in to the fumace listed by the American Chemical Society, see Anttlur Sktndar¿ls J'or Laboftttot!
Chemiculs, BDH Ltd., Poole, Dorset, U.K., and the Unite¿ Stotes Phunnacopeia
and temperatr.rre controllers capable of maintaining the tem- tutd Ntuionul Ilottnulary, U.S. Phalmacopeial Convention, Inc. (USPC), Rockville,
perature of each furnace zone independently to 950"C. MD.
Copyright by ASTM Int'l (all rights reserved); Mon Apr 19 03:46:56 EDT 2010 2
Downloaded,/printed by
Francois Van+Dievoit Q pursuant to License Agreement. No further reproductions authorized.
d$þ oszo2-os
7. Sampling shown that precision and bias are not degraded. Set the inlet
7.1 Obtain a test sample in accordance with Practice D4057 furnace temperature to 600 -, 25oC, and the outlet (catalyst)
or D4117. (Warning-Samples that are collected at tempera- temperature to 950 -+ 25oC. Set the automatic boat control as
tures below room temperature can undergo expansion at follows: I Fuc FWD 125 speed 10 time 30, 2 Fuc 285 speed 05
laboratory temperatures and rupture the container. For such time 30, 5 Fuc time 30, 6 Fuc time 90, A Fuc time 60. Refer to
samples, do not fill the container to the top. Leave sufûcient air manufacturer's instructions for a description of these settings.
space above the sample to allow room for expansion.) 8.3 Insert boat into furnace lor a minimum of 2 min to
(Warning-To minimize loss of volatile components, which remove any residual nitrogen species.
can be present in some test samples, do not uncover any longer
than necessary. Test samples should be analyzed as soon as 9. Calibration and Standardization
possible after taking from bulk supplies to prevent loss of 9.1 Prepare calibration standards containing 1, 5, 10, 50,
nitrogen or contamination due to exposure or contact with and 100 ng nitrogen/¡rl in xylene by volumetric dilution of the
sample container.) 500-ng nitrogen/¡rl nitrogen stock solution.
7.2 If lhe test sarnple is not used immediately, then thor- 9.2 Five microlitres of the matedal to be analyzed (see Note
oughly mix it in its container prior to taking a test specimen. 3) shall be quantitatively placed in the sample boat for
Some test samples require heating in order to thoroughly measurement of chemiluminescence response. There are two
homogenize. alternative injection procedures available, the volumetric and
the gravimetric procedures.
.ì. Preparation of Apparatus Norr 3-The formation of NO and NO, from oxidative combustion of
8.1 Assemble apparatus in accordance with the manufäctur- nitrogen containing hydrocarbons is dependent on combustion conditions
er's instructions. such as temperature and oxygen concentration. Injection of a constatrt
solution volume, and dilution of all test specitnens and standards with a
8.2 Adjust the oxygen flow for the ozone generator in commolr solvent, maintain consistent combustion conditions fbr test
accordance with the manufacturer's instructions. Adjust the specimens and standards.
combustion tube gas flows al'rd the pyrolysis temperature to the
recommended operating conditions using the following guide-
9.2.1 For volumetric measurement of the material by mi-
lines for each furnace type. (Warning-Ozone is extremely crolitre syringe, flush the microlitre syringe several times with
the material, discarding the flushed liquid each time. Fill the
toxic. Make sure that appropriate steps arc taken to prevent
10-¡rL syringe to the 5-¡rL level. Retract the plunger so that the
discharge of ozone within the laboratory work area.)
lower liquid meniscns falls on the 10 Eo scale mark. When
8.2.1 For the single-zone firnace without the ability to
bubbles are present within the liquid column, flush the syringe
change to apure oxygen carrier gas flow after the boat has been
and withdraw a new aliquot of the liquid. Record the volume of
ftrlly extended into the furnace, adjust the combustion tube gas
liquid in the syringe. Immediately inject the liquid into the
flows to the following values: pyrolysis oxygen, 360 -t. 36
boat, being careful to displace the last drop by touching the
ml-/min; inlet oxygen, 60 + 6 mllmin; and inert carder inlet,
edge of the boat, or the quartz wool or suitable equivalent (see
155 -r- 15 mllmin. Other gas flows may be used if it can be
shown that precision and bias are not degraded. Set the furnace
6.8) il present, with the syringe needle. After the injection,
again retract the plunger so that the lower liquid meniscus falls
temperaturc to 1100 -r 25"C. Adjust the boat drive mechanism
on the 10 7o scale mark and record the volume of liquid in the
to obtain a drive rate of 150 -t- 10 mm/min. Refer to the
syringe. The difference between the two volume readings is the
lanufacturer's instructions for descriptions of these settings.
' 8.2.2 For the two-zone fumace without the ability to change
volume of liquid injected.
to a pure oxygen carrier gas flow after the boat has been fully Nor¡ 4-An automatic sampling and injection device can be used in
extended into the furnace, adjust the combnstion tube gas flows place of the described manual injection procedure.
to the following values: combustion oxygen, 165 -f l6 ml/min; 9.2.2 For gravimetric measurement of the solution, fìll the
inlet inert carier, 85 -f 9 ml/min; and boat inert carrier, 50 + syringe as indicated in9.2.1. Weigh the microlitre syringe and
5 ml-/min. Other gas flows may be used if it can be shown that its contents, and record the mass to the nearest 0.01 mg.
precision and bias are not degraded. Set the inlet furnace Immediately inject the liquid into the boat, being careful to
temperature to 1050 -r 25oC, and the outlet furnace tempera- displace the last drop by touching the edge of the boat, or the
ture to 925 'r 25"C. Adjust the boat drive mechanism to obtain quartz wool or suitable equivalent (see 6.8) if present, with the
a drive rate of 150 -r- l0 mm/min (boat speed number 4). Refer syringe needle. After the injection, remove the syringe and
to the manufacturcr's instructions for the description of these again weigh the syringe and its contents. Record the mass to
settings. (Warning-High temperature is employed in this test the nearest 0.01 mg. The difference between the two weighings
method. Use flammable materials with care near the pyrolysis is the mass of liquid injected. The gravimetric procedure is
furnace.) more precise than the volumetric procedure, provided a bal-
8.2.3 For the two-zone furnace with the ability to change to ance with a precision of +0.01 mg is used.
a pure oxygen carrier gas flow after the boat has been fully 9.3 Activate the boat drive mechanism to inseft the boat into
extended in to the furnace, adjust the combustion tube gas the furnace. The instrument baseline should remain stable until
flows to the following values: main oxygen,400 -r- 40 ml-/min; the boat approaches the furnace and volatilization of injected
inlet argon carrieq 0.4 -+ 0.04 Llmin, and inlet oxygen carrier, material begins. After the measurement is complete, retract the
0.4 -r 0.04 Llmin. Other gas flows may be used if it can be boat. The instrument baseline should reestablish before the
Copyright by ASTM Int'l (all rights reserved); Mon Apr 19 03:46:56EDT 2010 4
Downloaded,/printed by
Francois Van+Dievoit 0 pursuant to License Agreement. No further reproductions authorized.
,($fp oszo2 - oe
boat has completely emerged from the fumace. Record the solution shall be less than the concentration of the highest
integrated chemiluminescellce response. Allow the boat to cool standard used in calibration ar-rd greater than 3 ng nitrogen/¡rl.
for at least I min before the next injection. The dilution can be performed either on a weight or volune
9.4 Calibrate the instrument using one of the following two basis.
techniques. ll.1.l For gravimetric dilution, record the mass of the test
9.4. I Perform measurements
for the calibration standards specimen and the total mass of the test specimen and solvent.
and blank using the procedure described in 9.2 and 9.3. 11.1.2 For volumetric dilution, record the mass of the test
Measure the calibration standards and blank three times each, specimen and the total volume of the test specimen and
and determine the average integrated chemiluminescence re- solvent-
sponse fbr each. Construct a curve plotting average integrated 11.2 Measure the chemiluminescence response for the test
detector response (y-axis) versus nanograms of nitrogen in- specimen solution using the procedure described in 9.2 and 9 .3 .
jected ("r-axis). 1 1.3 If the chemiluminescence response from the test speci-
9.4.2 If the system features an internal calibration routine, men solution is greater than the response from the highest
measure the calibration standards and blank three times each calibration standard used, repeat the test specimen dilution
using lhe procedure described in 9.2 and 9.3. Calibrate the described in 1 l. I using a higher dilution factor. Repeat the
analyzer in accordance with the manufacturer's instnrctions analysis procedure described in 9.2 through 9.3 on this new test
using the average of the three measurements for each standard specimen solution.
and blank. 11.4 Inspect the boat and combustion tube to verìfy com-
ì 9.5 If analyzer calibration is performed using only a subset plete combustion of the test specimen solution. Increase the
of the calibration standards listed in 9.1, the calibration residence time for the boat in the furnace if coke or soot is
standards closest in concentration to the measnred soh.rtion(s) observed on the boat. Decrease the boat drive introduction rate
must be included in the subset (that is, if the concentration of if coke or soot is observed on the exit end of the combustion
the test specimen solution is 20 ng nitrogen/¡rl, include the l0 tube. Clean any coked or sooted parts. After any cleaning or
and 50-ng nitrogen/¡rl standalds in the calibration). System adjustment, repeat instrument calibration prior to reanalysis of
performance must be checked with the calibration standards at the test specimen.
least once per day. I1.5 Measure each test specimen solution three times and
calculate the average chemiluminescence response.
10. Quality Control (QC)
l0.l Calibratiut Check Scunple(s)-A sample of known 12. Calculation
nitrogen content will be run after each calibration. The sample l2.l For analyzers calibrated using a standard curve, calcu-
can also be analyzed periodically throughout a series of late the nitrogen content of the test specimen in microgran.rs per
analyses to check the functioning of the instmment and the gram (¡-rg/g) as follows:
validity of the calibration curve. This sample can be a National (I*Y)
Institute for Standards and Technology Standard Ref'erence : (l)
Nitrogen, (Ug/g)
s;T-xE
Material (SRM) material, an acridine in xylene standard
or,
prepared to have a nitrogen value not used to calibrate the
instrument, or any other material that has been analyzed
repeatedly such that suflicient data are available to determine a
Nirrogen,(ug/g):5ffi (z)
Copyright by ASTM Int'l (all rights reserved); Mon Apr 19 03.46.56 EDT 20 l0 5
Downloaded,/printed by
Francois Van+Dievoit 0 pursuant to License Agreement. No further reproductions authorized.
.{[þ oszo2 - oe
a
Nitrogen, (Ug/g) : MXK. (3)
L4. Precision and Bias
s
14.1 Repeatability-1¡" difference between two test results
or, obtained by the same operator with the same apparatus under
: I constant operating conditions on identical test material would,
Nitrogen, (Ug/g) (4)
vxKu in the long run, iu the normal and correct operation of the test
method, exceed the following values in only one case in
where
D= density of test specimen solution, g/ml,
twenty, where x = *",ï:ïil:iil* test results.
I_ average of visual display readings of test specimen (s)
solution, ng nitrogen, 14.2 Reproducibility--:¡¡e difference between two single
Kr= gravimetric dilution fäctor, mass of test specimen/ and independent results obtained by different operators work-
mass of test specimen and solvent, g/g, ing in different laboratories on identical test material would, in
Kr= volumetlic dilution factor, mass of test specimen/ the long run, in the normal and correct operation of the test
volume of test specimen and solvent, g/ml, method, exceed the following values in only one case in
M- mass of test specimen solution injected, either mea-
X
twenty, where = the average of the two test results.
sured directly or calculated from measured volume
injected and density, V x D, mg, and R=0,266xX¡tg/g (6)
V volume of test specimen solution injected, either 14.3 Bias-A NIST SRM was analyzed by the cooperators
measured directly or calculated from measured mass participating in the repeatability and reproducibility determi-
injected and density, M/D, ¡tL. nation. This test method showed no signifìcant bias for this
sample within the repeatability of this test method.
13. Report
13.1 Report nitrogen results to two significant fìgures. State 15. Keywords
that results were obtained in accordance with Test Method 15.1 chemiluminescence; nitrogen
D5162.
a
Supporting data have been filed at ASTM Infeì'national f{eadquarters and may
be obtained by requeseing D02-1370 and D02-1507.
APPENDIX
(Nonmandatory Information)
, Xl.1 Confirm the performance of the instrument or the test X1.4 The frequency of QC testing is dependent on the
þrocedure by analyzing QC sample(s). criticality of the quality being measured, the demonstrated
stability of the testing process, and customer requirements.
X1.2 Prior to monitoring the measurement process, the user Generally, a QC sample should be analyzed each testing day
of this test method needs to determine the average value and with routine samples. The QC frequency should be increased if
control limits of the QC sample. (See Practice D6299 and a large number of samples are routinely analyzed. However,
MNL7.s) when it is demonstrated that the testing is under statistical
control, the QC testing frequency may be reduced. The QC
Xl.3 Record the QC results and analyze by control charts sample testing precision should be periodically checked against
or other statistically equivalent techniques to ascertain the the ASTM test method precision to erlsure data quality. (See
statistical control status of the total testing process. In the Practice D6299 and MNL7.'5.¡
absence of explicit requirements given in the test method, this
clause provides guidance on QC testing frequency. (See X1.5 It is recommended that, if possible, rhe type of QC
Practice D6299 and MNL7.5) Investigate any out-of-control sample that is regularly tested be representative of the material
data for root cause(s). The results of this investigation may, but routinely analyzed. An ample supply of QC sample material
not necessarily, result in instrument recalibration. should be available for the intended period of use, and it shall
be homogenous and stable under the anticipated storage
conditions.
t MNL7, Manu¿tl on Presentation of l)utu
Con.trol Churt Analysis, 6tl-t ed.. Xt.6 See Practice D62L)9 and MNLT'5 for fufther guidance
Section 3, ASTM Internationaì, W. Conshohocken, PA. on QC and control charting techniques.
Copyright by ASTM Int'l (all rights reserved); Mon Apr 19 03:46:56EDT 2010 6
Downloaded/printed by
Francois Van+Dievoit 0 pursuant to License Agreement. No further reproductions authorized.
.{$þ oszo2 - oe
SUMMARY OF CHANGES
Subcommittee D02.03 has identified the location of selected changes to this standard since the last issue
(D5762-08) that may impact the use of this standard.
ASTM lnternational takes no position respecting the validity of any patent r¡ghts asserted ¡n connection w¡th any item mentioned
¡n th¡s sta¡1dard. Users of this standard are expressly advised that determinat¡on oÍ the validity of any such patent r¡ghts, and the risk
of infringement of such r¡ghts, are entirely their own respons¡bil¡ty.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
¡fnot rev¡sed, e¡ther rcapproved or w¡thdrawn. Your comments are ¡nvited either for revision of th¡s standard or for additional standards
and should be addressed to ASTM Internationa! Headquañeß. Your comments will receive careful consideration at a meet¡ng of the
responsible techn¡cal committee, which you may attend. lf you feel that your comments have nat rece¡ved a fa¡r hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple cop¡es) of this standard may be obtained by contact¡ng ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission r¡ghts to photocopy the standard may also be secured from the ASTM website (www.astm.org/
COPYRIGHT).
Copyright by ASTM Int'l (all rights reserved); Mon Apr 19 03:46:56EDT 2010 7
Downloaded./printed by
Francois Van+Dievoit Q pursuant to License Agreement. No further reproductions authorized.