Available at www.sciencedirect.com
ScienceDirect
A R T I C L E I N F O A B S T R A C T
Article history: The use of bronchoscopes has increased in tuberculosis (TB) diagnostics to circumvent the
Received 30 July 2016 diagnostic challenges that are associated with low sputum volume and smear-negative TB.
Accepted 6 August 2016 In healthcare facilities situated in low income countries that have a high burden of TB, ade-
Available online 21 August 2016 quate decontamination of bronchoscopes is a challenge and often overlooked to save on
time and costs. This amplifies the risk of outbreaks and pseudo-outbreaks due to
Keywords: Mycobacterium tuberculosis and nontuberculosis mycobacteria. In this minireview, we review
Contaminated bronchoscopes published literature of contaminated bronchoscopes causing pseudo-outbreaks of
Pseudo-outbreaks M. tuberculosis and nontuberculosis mycobacteria in an effort to determine common
Mycobacterium tuberculosis sources, and possible mitigation strategies in low-resource settings.
Non-tuberculosis mycobacteria Ó 2016 Asian-African Society for Mycobacteriology. Production and hosting by Elsevier Ltd.
This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/
licenses/by-nc-nd/4.0/).
Bronchoscopy is an important diagnostic and therapeutic Vazque et al. [6], the highest number of contaminating inci-
tool [1,2] in both ambulatory and inpatient healthcare settings dents was attributed to bronchoscopy and gastrointestinal
[1]. In the context of pulmonary tuberculosis, bronchoalveolar endoscopy. In the United States, contaminated fiberoptic
lavage or bronchial biopsy have been proven to be essential bronchoscopes are estimated to contribute to Mycobacterium
diagnostic tools, especially for patients who are unable to tuberculosis (MTB) nosocomial infections in 460–2300 human
expectorate sufficient sputum samples [3]. However, this immunodeficiency virus infected patients annually [7].
semicritical medical device [4] has also been reported to be Additionally, pseudo-outbreaks due to environmental
a source of both pseudo-infections and infectious outbreaks microorganisms contaminating bronchoscopes have also
[5]. An indication of an improperly disinfected bronchoscope been reported [8]. However, data related to bronchoscope-
acting as a potential reservoir for contamination of both cul- associated infections and pseudo-outbreaks is underreported
tures and patients can be gauged by the fact that the biobur- [5], with a dearth of data from low-income and developing
den on bronchoscopes postwashing has been estimated to be countries.
around 6.4 104 colony forming units/mL [4]. According to a MTB, nontuberculous mycobacteria (NTM), and Pseu-
metadata analysis conducted from 1974 to 2004 by Seoane- domonas aeruginosa are the most common pathogens
* Corresponding author at: Department of Pathology and Laboratory Medicine, Aga Khan University, Stadium Road, Post Office
Box 3500, Karachi 74800, Pakistan.
E-mail address: rumina.hasan@aku.edu (R. Hasan).
Peer review under responsibility of Asian African Society for Mycobacteriology.
http://dx.doi.org/10.1016/j.ijmyco.2016.08.002
2212-5531/Ó 2016 Asian-African Society for Mycobacteriology. Production and hosting by Elsevier Ltd.
This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
Table 1 – Reported pseudo-outbreaks of nontuberculous mycobacteria and M. tuberculosis associated with bronchoscopy.
Study design Sample size (No. cases/No. of Reason for suspecting Organism Identified source of Strain similarity Refs.
bronchoscopies performed) outbreak contamination
Retrospective study 14/1270 Unique strain of M. M. cholonae subsp. abscessus Rinse water Not performed [10]
cholnea isolated
+ inconsistent culture
findings with clinical
features of patients
Retrospective study 7/16 Inconsistent culture M. cholonae, Rinse water Not performed [11]
1/16 findings with clinical M. avium Water tank
5 ( 2 0 1 6 ) 4 0 8 –4 1 1
frequency filters, automated
bronchoscope washing
machine
Retrospective 22/75 Culture isolates were M. avium, M. intercellulare Water filter, hot & cold Nested PCR [19]
inconsistent with clinical water lines + RFLP
features of patients
Prospective study 5/7 Isolation of M. gordonae in M. gordonae Tap water, water supply PFGE [20]
BAL channels
Prospective study 4/5 Recurrent cases of M. tuberculosis Contaminated suction Not performed [21]
mycobacterial cross- valve
contamination
Retrospective cohort 6/10 High incidence of M. M. tuberculosis Hole in bronchoscope RFLP [22]
study tuberculosis sheath
Retrospective cohort 2/3 No cases reported in M. tuberculosis Inadequate cleaning & Spoligotyping [23]
study hospital the previous disinfection between + IS6110-based
year, suspected patients use. AER was not RFLP
nosocomial outbreak approved
409
Note. AFB = acid-fast bacilli; BAL = bronchoalveolar lavage; M. = Mycobacterium; PCR = polymerase chain reaction; RFLP = restriction fragment length polymorphism.
410 International Journal of Mycobacteriology 5 ( 2 0 1 6 ) 4 0 8 –4 1 1
associated with transmission during bronchoscopy and be no viable environmental mycobacteria/100 mL of postflush
pseudo-outbreaks [5]. Bronchoscope contamination with water from bronchoscopes [31].
MTB is associated with extensive healthcare costs on the sys- False-positive results of MTB smears, cultures, or molecu-
tem or on patients (where out-of-pocket expenses are lar tests in bronchoscopically obtained samples are mainly
involved); false-positive cases are often investigated with due to cross-contamination owing to insufficient decontami-
repeat cultures and advanced radiological investigations, nation of bronchoscopes between use [5,32] or cross-
receive unnecessary antimycobacterial treatment, and risk contamination in the laboratory [33,34]. Given that Xpert
adverse effects of medication. An interesting study by Shim MTB/RIF (Cepheid) is a closed system, and risk of cross-
et al. [9] highlights the limitations of direct amplification tests contamination in the laboratory is lower, false-positive or
created by the presence of false-positive results even by the unexpected positive Xpert results on bronchoscope samples
presence of a few dead MTB contaminating the broncho- are likely to be a consequence of inadequate decontamination
scopes. It is therefore imperative that special attention be of the bronchoscope itself. Guidelines, however, do not focus
placed on addressing and circumventing not only false- on decontamination to ensure complete removal of DNA or
positive cultures but also false-positive molecular detection antibiotic resistance genes from bronchoscopes.
tests such as Xpert MTB/RIF (Cepheid), and molecular probe Awareness and advocacy for stringent monitoring and
assays, due to bronchoscope contamination with surveillance within the bronchoscope suite and disinfection
mycobacteria. unit including monitoring of disinfectants used [31] thus
Reports of bronchoscope contamination with NTM and becomes essential. Implementing regular training and com-
MTB in literature from 1990 to 2016 are briefly outlined in petency assessment of personnel concerned with broncho-
Table 1. While many NTM-associated outbreaks have been scope disinfection would ensure compliance with
reported, MTB-associated pseudo-outbreaks have not been recommended guidelines.
reported with equal frequency. While implementation and compliance of decontamina-
Possible factors indicated in the failure of the broncho- tion guidelines cannot be stressed enough, it is also necessary
scope decontamination process and leading to NTM infec- that proper communication be established between the clini-
tions include the design of bronchoscopes/endoscopes, an cians, bronchoscopists, and laboratories so that not only are
over-reliance on automated endoscope reprocessors, mal- pseudo-outbreaks promptly detected but a coordinated
functioning parts and damage during use, noncompliance of approach is implemented to handle postcontamination
decontaminating and handling guidelines, and use of con- responses.
taminated/nonsterile water during washing [24–26]. Reported As access to bronchoscopy, as well as its use in the diagno-
causes of bronchoscopy-associated pseudo-outbreaks further sis of TB and in particular smear-negative TB increases
include damage to the internal channel of the bronchoscope, [35–37], implementation of policies ensuring proper decon-
the ability of bacteria to form biofilms that are difficult to tamination of instruments being used as well as safety during
remove, along with inadequate cleaning with low- or the procedure achieve paramount importance. We therefore
intermediate-level disinfectants. To prevent bronchoscope- recommend that in addition to development of regional
associated pseudo-outbreaks it is thus imperative to imple- and/or national guidelines for manual bronchoscope decon-
ment standardized decontamination guidelines, such as tamination to remove mycobacteria, innovative and low-
those issued by the Centers for Disease Control and Preven- cost regional quality assurance programs be introduced to
tion [4], Association for Professionals in Infection Control ensure that specimens obtained through bronchoscopic tech-
and Epidemiology [27], Food and Drug Administration [28], niques are free of cross-contaminating mycobacterial DNA.
and European Society of Gastrointestinal Endoscopy-
European Society of Gastroenterology and Endoscopy Nurses
and Associates [29]. There are, however, differences amongst Conflicts of interest
these guidelines; for example, for recommendations on
microbiological culture surveillance of bronchoscopes and The authors have nothing to disclose.
on the frequency of culture as well as on interpretation of
data. In light of such differences, a recommendation for an
initial validation study of the protocol to be used demonstrat- R E F E R E N C E S
ing its effectiveness in particular healthcare settings would be
useful [24]. While automated endoscope reprocessors are rec-
ommended by British Society of Gastroenterology and World [1] G. Diaz-fuentes, S.K. Venkatram, Role of flexible-
Gastroenterology Organisation in advanced settings, manual bronchoscopy in pulmonary and critical care practice, in: S.P.
cleaning, and disinfection are widely carried out in Haranath, S. Razvi (Eds.), Global Perspectives on
resource-limited settings [30]. Ensuring quality control partic- Bronchoscopy, InTech Croatia, 2012, pp. 1–33.
[2] J. Dionı́sio, Diagnostic flexible bronchoscopy and accessory
ularly in resource-limited healthcare settings using manual
techniques, Rev. Port. Pneumol. 18 (2012) 99–106.
cleaning and disinfection in particular, is a considerable chal- [3] U. Kalawat, K.K. Sharma, P.N.R. Reddy, et al, Study of
lenge. Since water is identified as a major environmental bronchoalveolar lavage in clinically and radiologically
source of bronchoscope contamination, this challenge is suspected cases of pulmonary tuberculosis, Lung India 27
made all the more difficult by limited access to clean water (2010) 122–124.
in many resource-limited settings. The Health Protection [4] Centers for Disease Control and Prevention, Guidelines for
Disinfection and Sterilization in Health Care Facilities, 2008,
Surveillance Center-Ireland recommends that there should
International Journal of Mycobacteriology 5 ( 2 0 1 6 ) 4 0 8 –4 1 1 411