You are on page 1of 7

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/265557616

A newer technique to program a semi adjustable articulator

Article  in  Journal of Pharmacy and Bioallied Sciences · July 2014


DOI: 10.4103/0975-7406.137421 · Source: PubMed

CITATIONS READS
10 3,194

6 authors, including:

Manoharan P S Jagadish Konchada


Sri Balaji Vidyapeeth University Annamalai University
55 PUBLICATIONS   92 CITATIONS    4 PUBLICATIONS   33 CITATIONS   

SEE PROFILE SEE PROFILE

Some of the authors of this publication are also working on these related projects:

Systematic Competency Oriented Education Programme for BDS interns View project

Systematic Competency Oriented Education Programme for BDS interns View project

All content following this page was uploaded by Manoharan P S on 28 June 2019.

The user has requested enhancement of the downloaded file.


Dental Science - Original Article
A newer technique to program a semi
adjustable articulator
R. Venkateshwaran, Suma Karthigeyan1, P. S. Manoharan2, Jagadish Konchada3,
Manikandan Ramaswamy4, Bhuminathan

Departments of ABSTRACT
Prosthodontics,
Introduction: The difficulty in reproducing accurate angle of condylar guidance in semi‑adjustable articulators.
Sree Balaji Dental College,
Bharath University,
Purpose: The purpose of this study was to determine the correlation between the angle of horizontal condylar
Pallikaranai, Chennai, inclination obtained on a semi‑adjustable articulator and the corresponding angle traced on a temporomandibular
1
Rajah Muthaiah Dental joint (TMJ) radiograph in completely edentulous subject. Materials and Methods: The horizontal condylar
College, Chidambaram, inclination angle was obtained in a semi‑adjustable articulator by means of height tracer (extra oral tracing
4
KSR Dental College, device) and interocclusal records to program the articulator in 21 subjects. TMJ radiograph were recorded by
Erode, Tamil Nadu, the same operator with same orthopantomogram (OPG) machine (planmeca). Tracings of inclines of articular
2
Indra Gandhi Dental eminence on the radiograph were compared with the angle obtained on a semi‑adjustable articulator. Each
College, Puducherry, measurement was made using manual methods of measuring angle. The results were subjected to the Pearson
3
Sree Sai Dental College
correlation statistical analysis (α =0.01). Results: The outline of the articular eminence in a TMJ tomogram
and Research Institute,
Srikakulam, Andra
radiographic image was identified and traced. A significant correlation was found between the horizontal
Pradesh, India condylar inclination on a semi‑adjustable and the corresponding TMJ tomogram radiographic image for both
right (R = 0.789; P = 0.001) and left (i = 0.747; P = 0.004) sides. Conclusion: The articular eminence
Address for correspondence: traced on a TMJ tomogram image represents the horizontal condylar inclination with a mean difference of
Dr. R. Venkateshwaran, 5° in 21 subjects evaluated.
E‑mail: dr.venkateshwaran@
gmail.com

Received : 30-03-14
Review completed : 30-03-14
Accepted : 09-04-14 KEY WORDS: Horizontal condylar inclination, Programming, TMJ tomogram

I n completely edentulous, radiographic image of the


slopes of articular eminence on a TMJ tomogram when
traced can be used in programming the semi‑adjustable
a perfectly harmonized occlusion. The anatomic slopes of
articular eminence are represented as the horizontal condylar
inclination in the articulator. These programming are done
articulator. by means of using protrusive and lateral inter occlusal
records.[2] Arbitrary setting of the condylar elements will
A semi‑adjustable articulator allows adjustment to replicate provide an entirely satisfactory representation of condylar
average mandibular movements Glossary of prosthodontic movements.[2] Numerous studies have shown the inconsistency
terms.[1] It is possible only if the degrees of anatomical slopes in recording and reproducing these angles in a semi‑adjustable
are accurately recorded. The exact replication of mandibular articulator.[3,4] The setting condylar inclination at a flatter
movements has been a clinical challenge in order to achieve than the average value to ensure disocclusion of the posterior
teeth during excursions.[5] Interocclusal record material used
Access this article online in programming the articulator has also shown variation up
Quick Response Code: to 21-64°.[6] The individual inclination of the eminence is
Website:
very steep or flat, guidance obtained from the average value
www.jpbsonline.org
settings may differ sufficiently to cause problems in achieving
particular clinical objectives, such as posterior disocclusion
DOI: or balanced occlusion. The use of additional aids, such as
10.4103/0975-7406.137421 dynamic registration or imaging, may resolve this potential
problem.

How to cite this article: Venkateshwaran R, Karthigeyan S, Manoharan PS, Konchada J, Ramaswamy M, Bhuminathan. A newer technique to program a semi
adjustable articulator. J Pharm Bioall Sci 2014;6:S135-9.

Journal of Pharmacy and Bioallied Sciences July 2014 Vol 6 Supplement 1 S135 
Venkateshwaran, et al.: A newer technique to program a semi adjustable articulator

Materials and Methods

A study was conducted at the Department of Prosthodontia,


Rajah Muthiah Dental College and Hospitals, Annamalai
University. 21 subjects requiring complete denture treatment
were chosen randomly from the outpatient department. The
inclusion criterion was patient should have good neuromuscular
coordination to produce good tracing of mandibular border
movements.
Figure 1: Centric record and protrusive record
All steps of complete denture procedures were done. A tentative
jaw relation was performed and the maxillary cast was oriented
to the articulator using a Hanau spring bow. Using the nick
and notch method a static record of centric relation was
obtained using Zinc Oxide Eugenol (ZOE) impression paste
as interocclusal recording material. Once, the cast was oriented
in articulator. Extra oral tracer height tracing device was used
in this study. The patient was trained for tracing of mandibular
border movements in the way as prescribed by Ney mandibular
excursion guide. The completely edentulous subjects were
trained for tracing of mandibular movement until a sharp arrow
point tracing was obtained. Measuring 6 mm from the apex of
the tracing a point is marked over the transparent sheet and this
will represent the protrusive position. Once this is accomplished
nick are made in the occlusal rim of both maxilla and mandible
correspondingly in canine and molar region. Then the surface
of the occlusal rim is coated with petrolatum and using the
polyvinyl siloxane material (3M CAD BITE, ESPE, Virtual CAD
bite registration material, Ivoclar, Zurich, Switzerland) as an Figure 2: Protrusive record in position to program the articulator
interocclusal record centric record is obtained [Figure 1]. Then
the patient is asked to close the mandible in a forward position
to represent the protruded position of mandible such that the
stylus is positioned on the 6 mm marking on the tracing plate.
Interocclusal record material is injected to obtain a protrusive
record. The interocclusal record material flows easily and should
be rigid on setting.

Once the protrusive and centric records have been obtained,


these records are used to program the semi‑adjustable
articulator using Lauritzen’s split cast technique [Figure 2].
With the centric record in position, the cast oriented and
articulated in exact centric position. Using the protrusive
record the articulator is programmed to determine the
horizontal condylar inclination. Once, the required horizontal
condylar inclination has been determined in an articulator.
Then the Bennett angle was obtained using the formula
L = H/8 + 12. Figure 3: Positioning light grid on patient

Then the patient is referred to Department of Oral Medicine


and Radiology for temporomandibular joint (TMJ) tomogram. the grid lines in the machine as the crossing point of the light
Radiographs were made at 70 Kv at a fixed distance in a is to be focused on the tip of the nose. In the midline, the mid
planmeca radiograph machine. Cassettes were loaded with OPG nasal and anterior nasal spine were used to vertically align the
film and the patient was positioned in the machine using head sagittal plane. The exposure for radiograph was done at 70 Kv
positioner. The positioning of the patient head was standardized and 68 Kv at 10 mA in both open mouth and closed mouth
using the positioning light grid in the machine [Figure 3]. position, respectively for increased contrast between the two
This grid was designed to be parallel to Frankfort horizontal positions. All radiographs were made by the same operator at
plane (FHP) in the patient. The anteroposterior positioning the same time and with same panaromic radiographic unit.
of the head was based on the correction scale provided by the TMJ tomogram was made of each subject with the FHP and
manufacturer. The position of the head was confirmed with upper border of radiograph parallel to the floor. Due to circular

 S136 Journal of Pharmacy and Bioallied Sciences July 2014 Vol 6 Supplement 1
Venkateshwaran, et al.: A newer technique to program a semi adjustable articulator

rotation of the radiographic recorder, this results in the upper From the descriptive analysis in the following results are arrived.
border of radiograph being parallel to the FHP image. The 21 respondents are taken into the analysis calculation of
articulator right value of the mean is 25.238, standard deviation
Once the tomograms were obtained, tracing papers were value is 9.417 and standard error is 1.721. Condylar inclination
attached to the radiograph. Tracings were made of the images in TMJ tomogram view right value of mean 30.476 and standard
of each radiograph on parchment paper along the border of deviation 7.890 in the same standard error (1.721) [Table 4].
slopes of articular eminence. Horizontal reference lines were
drawn parallel to the upper border of radiograph, contacting, Discussion
and identifying the most superior and inferior point of each
curve. The outline of each curvature and the flat reference line Among the 21 subjects, significant correlation was found
were traced. On tracing a line was drawn connecting the most between the angle obtained on the articulator and the TMJ
superior and inferior points of curvature representing the mean tomogram radiographic image. The Pearson’s correlation
curvature line. These two points were connected by a straight analysis test was conducted at a confidence interval of 99%.
line representing a mean curvature line [Figures 4 and 5]. The results show a higher correlation (r) value of articulator
Angles made by the intersection of mean curvature line and versus TMJ tomogram on the right and left as 0.789 and
the horizontal reference line (FHP) were measured. Each 0.747, respectively. This shows that the angles obtained on the
measurement was repeated twice by the operator. Data were articulator are very closely correlated to that obtained on a TMJ
subjected to Pearson’s correlation statistical analysis. tomogram radiograph. There was statistically no significant
variation in horizontal condylar inclination angle obtained from
Results both methods. The high correlation value shows that it can be
used as a reliable tool in programming the articulator. In terms
The radiographic image of the outline of articular eminence of difference in horizontal condylar inclination angle, a mean
was seen distinctly in a TMJ tomogram. The outline if this was difference of 5° was obtained between the angles obtained on
traced to obtain the angle of horizontal condylar inclination an articulator to that of the TMJ tomogram. The mean value
of the respective side and compared with the values obtained of error obtained between the left side and right side of the
after programming the articulator. Radiographic values were on articulator and TMJ tomogram was 4.7° and 5.2°, respectively.
average 5° greater than anatomic values [Table 1].
Among all TMJ imaging technique, TMJ tomogram is a
From the correlation analysis in the following results are arrived. specialized view for imaging TMJ. It represents the composite
Among the 21 subjects, significant correlation was found between sagittal inclination of the articular eminence of both sides
the angles obtained in the articulator to that of the corresponding in open mouth and closed mouth position. TMJ tomogram
radiographic image. The calculation of articulator value versus presents better image quality, as all other forms of radiographic
TMJ value of right and left in the right the correlation value (r) visualization of TMJ had overlap of the image to some extent.
is 0.789 and 0.747 at the level of 1%, respectively. The calculated The controlled radiation exposure at proper interval ensures that
value is higher than the table value and significant at 1% level. there is no image overlap and the TMJ alone is radiographed
Hence, the angle obtained in articulator versus TMJ tomogram on the film.
is very closely correlated. Hence, there is no significant variation
among the both views [Table 2]. Gilboa et al.[7] have reported that the radiographic outline of
the articular fossa and articular eminence provided an accurate
From the descriptive analysis in the following results are arrived. representation of the equivalent outline in 25 human skulls
The 21 respondents are taken into the analysis calculation of the
mean value of angle in articulator left is 25.714, standard deviation
value is 10.159 and standard error is 1.822. Condylar inclination
in TMJ tomogram view left value of mean 30.476 and standard
deviation 8.352 in the same standard error (1.822) [Table 3].

Figure 4: Tracing done on a temporomandibular joint tomogram using Figure 5: Angles obtained from temporomandibular joint tomogram
a parchment paper radiograph

Journal of Pharmacy and Bioallied Sciences July 2014 Vol 6 Supplement 1 S137 
Venkateshwaran, et al.: A newer technique to program a semi adjustable articulator

Table 1: The average angles, standard deviation between supporting the bases in order to achieve uniform simultaneous
mean curvature line and the reference line (FHP) on the contact of denture.
radiograph and in the subject
Condylar inclination Mean Standard deviation N The Gothic arch tracing is considered as the most ideal
Articulator right 25.2381 9.41756 21 method in programming the articulator. This method has
Articulator left 25.7143 10.15944 21 some minor drawbacks like, the positioning of the tracer
TMJ tomogram right 30.4762 7.89062 21 doesn’t allow the lip to be passive and hinders the rest
TMJ tomogram left 30.4762 8.35236 21 position of the mandible, excessively resorbed ridges, flabby
FHP: Frankfort horizontal plane, TMJ: Temporomandibular joint ridges, inadequate inter arch distance, large tongue. In
subjects with TMJ arthropathy, there are chances of obtaining
Table 2: The correlation between the angle of inclination in aberrant tracing that are faulty. These faulty tracings may
articulator and TMJ tomogram radiograph lead to results that do not represent the actual horizontal
condylar inclination of the patient and has no significance
Condylar inclination n r Significant at 0.01%
in programming the articulator.
Articulator right versus 21 0.789** 0.000
TMJ tomogram right
Articulator left versus 21 0.747** 0.000
The temporomandibular X‑ray film reproduces the actual
TMJ tomogram left form and angles of the osseous structure of the joint. On the
other hand, the functional registrations indicate the condylar
**Correlation is significant at the 0.01 level (one‑tailed, Pearson
correlation). TMJ: Temporomandibular joint
path angles by means of the path traversed by the condyle
along the slopes of the articular eminence. In recording
horizontal condylar inclination errors up to 21-64° has been
Table 3: Descriptive statistics among the condylar inclination observed among the same operators, technique and material.
in articulator left versus TMJ tomogram view left By examining the angle of the osseous structure of the joint
n Mean SD Standard Degree directly on the film, and making a transfer from the film to
error error
the instrument, a more specific condylar path angle can be
Condylar inclination 21 25.7143 10.15944 1.82263 4.7619 determined.
in articulator left and 21 30.4762 8.35236
condylar inclination in
TMJ tomogram view left The articular eminence inclination in the radiographic image
was traced from the most superior to the most inferior points of
TMJ: Temporomandibular joint, SD: Standard deviations
curvature representing mean inclination of curve. This may be
different from guiding inclination with approximately 4-6 mm
Table 4: Descriptive statistics among the condylar inclination of protrusion, which is clinically significant range of protrusion
in articulator right and TMJ tomogram right and condylar guidance.
n Mean SD Standard Degree
error error When dynamic extraoral tracings are used, the starting
Condylar inclination in 21 25.2381 9.41756 1.72187 5.2381 point (centric relation), individual curved path, and eccentric
articulator right and position are recorded. With static records, only the centric
condylar inclination in 21 30.4762 7.89062
relation and eccentric positions are recorded, the midpoint
TMJ tomogram view right
of the excursion produces the maximum error with a straight
SD: Standard deviation, TMJ: Temporomandibular joint condylar slot. The maximum difference between a straight
condylar path and one that has a 1/2 inch radius is 0.4 mm.
with a mean difference in inclination of 7°. Images in a TMJ A maximum condylar error of 0.4 mm produces a 0.2 mm error
tomogram provide composite sagittal representations of the at the second molar.
skeletal structures. TMJ tomogram presents better image
quality and outline of the articular eminence of the temporal The sensitivity of the Hanau Wide‑Vue 183-2 closed tract
bone without any image overlap. A variety of interocclusal semi‑adjustable articulator is of the order 5°. From the values
record materials are used in recording horizontal condylar obtained a mean difference of 5° was obtained between the
inclination, are impression plaster, polyvinyl siloxane, waxes, angles obtained on an articulator to that of the TMJ tomogram.
acrylic resin and ZOE paste. Inter occlusal record even though The mean value of error obtained between the left side of the
they are considered as the most ideal way of programming the articulator and TMJ tomogram was 4.7° and on the right side
articulator. Preti et al. have reported that average interocclusal it was 5.2°. This shows that the 5° error is minimal or almost nil
registration values of the condylar guidance inclination vary compared to other methods of recording horizontal condylar
from 21° to 64°. inclination.

Various interocclusal record material has been used in the past, The current study concluded that TMJ tomogram accurately
the difficulty in recording wax inter occlusal record was uneven represents the outline of the articular eminence and the angle
softening and uneven thickness of the recording material. of the slopes of articular eminence traced on it can be used
Boucher recommends that the centric relation should be made as an aid in setting the condylar guidance inclination on a
with minimal pressure to prevent displacement of the tissues semi‑adjustable articulator.
 S138 Journal of Pharmacy and Bioallied Sciences July 2014 Vol 6 Supplement 1
Venkateshwaran, et al.: A newer technique to program a semi adjustable articulator

Conclusion factors and clinical variation influencing the reproducibility of


interocclusal recording methods. Br Dent J 2002;192:395‑400.
3. Celar AG, Tamaki K. Accuracy of recording horizontal condylar
The radiographic outline of the articular fossa and articular inclination and Bennett angle with the Cadiax compact. J Oral Rehabil
eminence provided an accurate representation of the equivalent 2002;29:1076‑81.
outlines in 21 subjects with a mean difference in inclination 4. de Freitas A. A comparison of the radiographic and prosthetic
measurement of the sagittal path movement of the mandibular
of 5°. The image of the articular eminences in a TMJ tomogram condyle. Oral surgery, oral medicine, and oral pathology
may be used to provide an indication of the degree of inclination 1970;30:631‑8.
of the articular eminence and may be of value as an aid in setting 5. Pullinger AG, Solberg WK, Hollender L, Guichet D. Tomographic
the condylar guidance in semi‑adjustable articulators. A new analysis of mandibular condyle position in diagnostic subgroups of
temporomandibular disorders. J Prosthet Dent 1986;55:723‑9.
method for programming the articulator has been suggested 6. Pullinger AG, Hollender L, Solberg WK, Petersson A. A tomographic
in this study that is more reliable, repeatable, ease of use with study of mandibular condyle position in an asymptomatic population.
minimal drawbacks. J Prosthet Dent 1985;53:706‑13.
7. Gilboa I, Cardash HS, Kaffe I, Gross MD. Condylar guidance:
Correlation between articular morphology and panoramic
References radiographic images in dry human skulls. J Prosthet Dent
2008;99:477‑82.
1. Brady AP, McDevitt L, Stack JP, Downey D. A technique for magnetic
resonance imaging of the temporomandibular joint. Clin Radiol
1993;47:127‑33. Source of Support: Nil, Conflict of Interest: None declared.
2. Eriksson A, Ockert‑Eriksson G, Lockowandt P, Eriksson O. Clinical

Journal of Pharmacy and Bioallied Sciences July 2014 Vol 6 Supplement 1 S139 
Copyright of Journal of Pharmacy & Bioallied Sciences is the property of Medknow
Publications & Media Pvt. Ltd. and its content may not be copied or emailed to multiple sites
or posted to a listserv without the copyright holder's express written permission. However,
users may print, download, or email articles for individual use.

View publication stats

You might also like