You are on page 1of 13

A study of IMRT planning parameters on planning efficiency, delivery

efficiency, and plan quality


Kathryn Mittauer
Department of Radiation Oncology, College of Medicine, University of Florida, Gainesville,
Florida 32603 and J. Crayton Pruitt Family Department of Biomedical Engineering,
University of Florida, Gainesville, Florida 32611

Bo Lu, Guanghua Yan, and Darren Kahler


Department of Radiation Oncology, College of Medicine, University of Florida, Gainesville, Florida 32603
Arun Gopal
Department of Radiation Oncology, New York-Presbyterian Hospital, Columbia University, New York,
New York 10032

Robert Amdur and Chihray Liua)


Department of Radiation Oncology, College of Medicine, University of Florida, Gainesville, Florida 32603
(Received 9 January 2013; revised 11 April 2013; accepted for publication 11 April 2013;
published 13 May 2013)
Purpose: To improve planning and delivery efficiency of head and neck IMRT without compromising
planning quality through the evaluation of inverse planning parameters.
Methods: Eleven head and neck patients with pre-existing IMRT treatment plans were selected for
this retrospective study. The Pinnacle treatment planning system (TPS) was used to compute new
treatment plans for each patient by varying the individual or the combined parameters of dose/fluence
grid resolution, minimum MU per segment, and minimum segment area. Forty-five plans per patient
were generated with the following variations: 4 dose/fluence grid resolution plans, 12 minimum seg-
ment area plans, 9 minimum MU plans, and 20 combined minimum segment area/minimum MU
plans. Each plan was evaluated and compared to others based on dose volume histograms (DVHs)
(i.e., plan quality), planning time, and delivery time. To evaluate delivery efficiency, a model was
developed that estimated the delivery time of a treatment plan, and validated through measurements
on an Elekta Synergy linear accelerator.
Results: The uncertainty (i.e., variation) of the dose-volume index due to dose calculation grid vari-
ation was as high as 8.2% (5.5 Gy in absolute dose) for planning target volumes (PTVs) and 13.3%
(2.1 Gy in absolute dose) for planning at risk volumes (PRVs). Comparison results of dose distribu-
tions indicated that smaller volumes were more susceptible to uncertainties. The grid resolution of
a 4 mm dose grid with a 2 mm fluence grid was recommended, since it can reduce the final dose
calculation time by 63% compared to the accepted standard (2 mm dose grid with a 2 mm fluence
grid resolution) while maintaining a similar level of dose-volume index variation. Threshold values
that maintained adequate plan quality (DVH results of the PTVs and PRVs remained satisfied for
their dose objectives) were 5 cm2 for minimum segment area and 5 MU for minimum MU. As the
minimum MU parameter was increased, the number of segments and delivery time were decreased.
Increasing the minimum segment area parameter decreased the plan MU, but had less of an effect on
the number of segments and delivery time. Our delivery time model predicted delivery time to within
1.8%.
Conclusions: Increasing the dose grid while maintaining a small fluence grid allows for improved
planning efficiency without compromising plan quality. Delivery efficiency can be improved by in-
creasing the minimum MU, but not the minimum segment area. However, increasing the respective
minimum MU and/or the minimum segment area to any value greater than 5 MU and 5 cm2 is not rec-
ommended because it degrades plan quality. © 2013 American Association of Physicists in Medicine.
[http://dx.doi.org/10.1118/1.4803460]

Key words: IMRT, plan quality, delivery time, treatment planning

I. INTRODUCTION popularity of IMRT for radiation therapy (RT),1 there is a


greater need for improved planning efficiency. Another need
Intensity modulated radiation therapy (IMRT) planning in the regimen of IMRT is to improve delivery efficiency
is a complex and time-consuming task involving many while maintaining the plan quality, since a long treatment
parameters that impact plan quality, treatment planning time can increase the chance of dose errors due to patient
time, and delivery time. Consequently, with the growing movement.2, 3

061704-1 Med. Phys. 40 (6), June 2013 0094-2405/2013/40(6)/061704/13/$30.00 © 2013 Am. Assoc. Phys. Med. 061704-1
061704-2 Mittauer et al.: Parametric analysis of IMRT planning 061704-2

Less “complex”4 IMRT plans allow for improved deliv- treatment delivery time can be improved. To assess delivery
ery accuracy and reduced delivery time.4 Several measures time, we have proposed a delivery time model. Similar efforts
on how to reduce the plan complexity have been proposed.4–15 include Li and Xing’s delivery time model for a Varian (Var-
These efforts focused on limiting the intensity fluctuations by ian Oncology System, Palo Alto, CA) linear accelerator.20
setting maximum intensity limits8, 9 or applying a smoothing However, no such model has been proposed at this time for
filter.4–7, 9, 10 Filtration or maximum intensity limits are ap- an Elekta (Elekta Oncology Systems, Norcross, GA) linear
plied either inside the optimization loop (i.e., concurrently accelerator. Finally, we evaluate the impact of the calculation
or postiteratively) or outside the optimization loop (i.e., prior parameters of both dose grid and fluence grid resolution on
to leaf sequencing). By limiting the intensity spikes, the plan plan quality and calculation time. We seek alternative options
complexity is reduced through the impact on the number of for both dose grid and fluence grid resolution based on plan-
MUs delivered per segment, the total number of segments, ning time considerations. The parameters found in this study
and/or the segment width. will provide a class solution that can reduce planning time.
However, there are inverse planning parameters that im-
pact the segmentation directly during IMRT optimization:
(1) the minimum segment area parameter (MSAP) limits the II. MATERIALS AND METHODS
MLC area to greater than a minimum size for each seg- II.A. Overview
ment, (2) the minimum MU parameter (MMUP) influences
the plan’s segments by limiting an individual segment’s mini- Eleven head and neck patients were chosen for this
mum MU, and (3) the maximum number of segments controls study. Plans were recomputed for each patient with varying
the total segment number for the plan. Since these parame- dose grid resolution (4 variations), minimum segment area
ters are incorporated directly into the optimization loop, plan (12 variations), minimum MU (9 variations), and the mini-
quality is not degraded with postprocessing methods, such as mum segment area and minimum MU were varied together
filtration or smoothing.9, 16 Although strategies15 on limiting (20 variations), while all other variables were held constant.
the maximum number of segments have been studied using The patients and plan variations are described in more detail
the Pinnacle treatment planning system (TPS) (Philips Radi- in Secs. II.B–II.D. An overview of this study can be seen in
ation Oncology Systems, Andover, MA), to our knowledge Table I. Plan quality, calculation time, and delivery time
initial settings for MSAP and MMUP in light of delivery effi- served as the study’s endpoints.
ciency have not been discussed.
Niemierko and Goitein first reported that a smaller grid
II.B. Patients
resolution provides superior dosimetric accuracy due to a
smaller interpolation of dose between calculation points.17 Eleven head and neck patients previously treated with
Investigators17–19 through both empirical methods18, 19 and IMRT were selected for this retrospective study. Head and
analytical models17, 18 reached a common consensus that a neck cases were chosen because of the planning complex-
dose grid resolution of 2 mm is sufficient to minimize dose ity associated with this site. As displayed in Fig. 1, all ex-
errors from a finite grid resolution. Chung et al. reported the isting treatment plans were prescribed 70 Gy to the high risk
impact on plan quality for various dose grid resolutions us- planning target volume (PTV 70) and 56 Gy to the standard
ing the Pinnacle TPS.19 However, this study did not report risk PTV (PTV 56) in 35 fractions. Since this fractionation
the cost of planning time nor did it report plan quality dif- treated both PTVs concurrently, both PTVs were optimized
ferences for critical structures. Although dose grid resolution simultaneously under the same fractionation. A summary of
has been well studied, the impact of fluence grid resolution the eleven initial treatment plans is given in Table II to show
has not been investigated at this time. Note the Pinnacle TPS that a variety of cases were selected.
allows for the user to specify these grid resolutions separately As shown in Table II, existing patient plans utilized six
for dose and fluence. or seven coplanar IMRT beams with varying gantry angles.
The purpose of this study is to recommend a set of in- Based on investigations of the appropriate number of beams
verse planning parameters for head and neck IMRT planning to use for head and neck IMRT,21 6–7 beams was considered a
that provides a relatively low planning effort without com- reasonable choice. Beam angles were chosen to best minimize
promising plan quality. Towards this goal, this study evalu- doses to planning at risk volumes (PRVs) while maximizing
ates a range of fundamental inverse planning parameters and PTV coverage. Beam energy was 6 MV for all cases.
their effect on plan quality, dose calculation time, and deliv- Patients selected included eight whole neck IMRT cases
ery time. Such a study is useful for setting an initial set of and three partial neck IMRT cases. Whole neck IMRT, illus-
parameters that allow for fast planning and efficient treatment trated in Fig. 1, was used when the gross disease extended
delivery, while allowing the user to be confident that planning into the lower neck region. For these eight patient plans, the
calculations are being performed accurately. planning isocenter was placed inside of PTV 70. For the three
In this study parameters of the MSAP and the MMUP are partial neck IMRT cases, an AP lower anterior neck (LAN)
studied both independently (i.e., varied one at a time) and de- field was added separately to treat the nodal region to 56 Gy.
pendently (i.e., varied simultaneously) in light of plan qual- A technique described by Li et al. was used in which the
ity and delivery time differences. By relaxing the MSAP and LAN field was feathered by shifting the match line 1 cm per
the MMUP, the plan complexity can be reduced; hence, the segment for a total of four segments (Fig. 2).22 On the last

Medical Physics, Vol. 40, No. 6, June 2013


061704-3 Mittauer et al.: Parametric analysis of IMRT planning 061704-3

TABLE I. Breakdown of trials for each case.

Grid resolution trials MSAP trials MMUP trials MMUSAP trials

Dose grid Fluence grid Segment Segment Segment


Trial (mm) (mm) Trial size (cm2 ) MU Trial size (cm2 ) MU Trial size (cm2 ) MU

1 2 2 5 3 3 17 4 2 26 4 3
2 3 3 6 4 3 18 4 3 27 4 5
3 4 4 7 5 3 19 4 4 28 4 7
4 4 2 8 6 3 20 4 5 29 4 10
9 7 3 21 4 6 30 6 3
10 8 3 22 4 7 31 6 5
11 9 3 23 4 10 32 6 7
12 10 3 24 4 20 33 6 10
13 16 3 25 4 35 34 8 3
14 25 3 35 8 5
15 36 3 36 8 7
16 49 3 37 8 10
38 10 3
39 10 5
40 10 7
41 10 10
42 16 3
43 16 5
44 16 7
45 16 10

Note: MSAP, min segment area parameter; MMUP, min monitor unit parameter; MMUSAP, min monitor unit and segment area parameters. (1) There are some trial
redundancies between parameters.

TABLE II. Summary of diagnosis and plan information for the 11 head and neck cases in this study.

Volume (cm3 )

Case no. Disease site No. of IMRT beams MU for 2 mm grid resolution plan PTV 70 PTV 56

1 Posterior pharyngeal wall 7 723 424.1 264.5


2 Oropharynx and nasopharynx 7 1037 420.8 950.1
3 Glottic larynx 7 560 430.7 471.7
4 R. glosso tonsiliar sulcus 7 657 464.4 436.8
5 L. true vocal cord 7 542 78.9 581
6 L. maxillary sinus 7 558 139.9 194.3
7 L. paratracheal and thyroid 6 691 99.3 358.7
8 R. oral tongue 7 + LAN 601 147.5 280.1
9 L. floor of mouth and mandible 6 755 116.9 225.5
10 R. oral tongue and floor of mouth 7 + LAN 505 167.9 251.8
11 L. base of tongue 7 + LAN 516 160.5 325.5
Statistic data Mean 649.5 241.0 394.5
Standard deviation 154.1 156.3 218.2

Note: PTV, planning target volume; LAN, lower anterior neck; PTV 70, high risk PTV; PTV 56, standard risk PTV. (1) PTV 70 volume is excluded from PTV 56 volume.

F IG . 1. Whole neck IMRT with isodose distribution for case no. 2 with display of high risk PTV 70 and standard risk PTV 56.

Medical Physics, Vol. 40, No. 6, June 2013


061704-4 Mittauer et al.: Parametric analysis of IMRT planning 061704-4

F IG . 2. AP LAN field of case no. 8 treated to 56 Gy using match line feathering in 1 cm intervals with four IMRT segments. The rightmost digitally reconstructed
radiograph (DRR) is the last segment in which the cord is blocked.

segment of the LAN field, the cord was blocked to reduce its data and experiences. Optimization objectives for organs at
dose to less than 50 Gy. The planning isocenter was placed at risk (OARs) were previously determined by organ classifica-
the junction of the LAN field with the head and neck fields, at tion, serial or parallel. Objectives for serial structures were
a depth greater than 1 cm. specified as maximum dose and volumetric maximum DVH.
Parallel structure objectives were defined as volumetric max-
imum DVH. PTV objectives were minimum dose, uniform
II.C. Treatment planning dose, and volumetric minimum DVH. Note that the objective
functions during the initial planning were customized for each
New treatment plans were generated from each of the
patient based upon the tumor size and the anatomy, which can
eleven cases by varying parameters of dose and fluence grid,
be slightly different from patient to patient. All optimization
minimum MU, and minimum segment area. Forty-five trials
objectives and contours were kept the same as the original
per patient, 495 trials in total, were reoptimized and/or re-
treatment plan when generating the 495 trials.
computed on the Pinnacle TPS (version 9.0) using direct ma-
All contour volumes were previously delineated by clin-
chine parameter optimization (DMPO) and adaptive super-
icians and were used on the initial patient treatment plan.
position/convolution for the dose calculation. The maximum
The gross tumor volume (GTV) and positive nodes were ex-
number of optimization iterations was set to 25, with intensity
panded, excluding OARs, to delineate the high risk clinical
optimization occurring within the first five iterations. Twenty-
target volume (CTV), CTV 70. A 3 mm contour expansion
five iterations was selected based on UF planning experiences
was applied to OARs and CTVs to delineate PRVs and PTVs
with DMPO as well as the Pinnacle’s historical recommen-
(Table III).
dations of using a TPS default preset of 25 iterations. After
calculation, all plans were scaled such that the dose to 95%
of PTV 70, D95%, received at least 70 Gy. Differential dose
II.D. Parameter variations
volume histograms (DVHs) were exported for plan quality re-
sults. The dose grid resolution, minimum MU, and minimum
All initial eleven cases were planned according to the pro- segment area were varied while keeping the all other param-
cess and dose objectives found in Table III. The generic dose eters constant. Note that planning parameters are TPS depen-
objectives are derived from quantitative analysis of normal dent and apply to the Pinnacle TPS. Dose grid resolution was
tissue effect in the clinic (QUANTEC) data,23 Radiation Ther- selected to evaluate the tradeoff between dose uncertainties
apy Oncology Group (RTOG) protocols, and UF historical caused by grid resolution variation and the cost of planning

TABLE III. ROIs (regions of interest) with expansions and dose goals for all cases.

ROI Expansion (mm) PTV and PRV dose goals

CTV (95% Vol > Rx) 3 95% Vol > 70 Gy (PTV70), 56 Gy (PTV56)
CTV (99% Vol > 93% Rx) 3 99% Vol > 65.1 Gy (PTV70), 46.5 Gy (PTV56)
CTV (20% Vol < 110% Rx) 3 20% Vol < 77 Gy (PTV70)
Brainstem 3 0.1 cc ≤ 55 Gy
Cochlea 3 0.1 cc ≤ 45 Gy
Cord 3 0.1 cc ≤ 50 Gy
Larynx 3 Mean ≤ 36 Gy
Parotid 3 Mean ≤ 26 Gy

Note: Contour expansion was performed after physician’s initial contour delineation.

Medical Physics, Vol. 40, No. 6, June 2013


061704-5 Mittauer et al.: Parametric analysis of IMRT planning 061704-5

time. Minimum segment area and minimum MU were se- positional error for a 1 cm2 segment would result in a 10% dif-
lected to study the effect of modulation (i.e., intensity) on plan ference in the segment area. Furthermore, the dosimetric out-
quality and treatment delivery time. put (i.e., dose/MU) for an Elekta linear accelerator (LINAC)
Default parameters were the values the parameters take fluctuates due to head scatter and loss of electronic equilib-
when not intentionally varied. Table I specifies these default rium for field sizes below 3 cm2 .27
parameters as boxed values, representing the reference plan
used for dose comparison in Sec. III. Default parameters were
2 mm (dose and fluence grid resolution), 4 cm2 (minimum II.D.3. MMUP
segment area), and 3 MU (minimum MU). We infer from pre- The MMUP influences the plan’s segments by limiting an
vious work,17–19 a 2 mm dose and fluence grid resolution as individual segment’s MU. Here, a segment cannot contribute
the established standard to measure dose accuracy. Default less than a specified MU. The MMUP was chosen to range
parameters of 4 cm2 and 3 MU were chosen empirically to from 2 to 35 MU in nine trials (Table I). Similar to Sec. II.D.2,
decrease the likeliness of mechanical inaccuracies and output the minimum limit was taken at one value below the default
fluctuations, which can arise for small MU and/or small field value of 3 MU. The maximum limit was selected when the
size deliveries.24–27 A more detailed discussion of the impact number of segments per beam approached one.
of these parameters on mechanical and dosimetric accuracy is The default MMUP was selected based on considerations
presented below. of an Elekta LINAC. Dosimetric accuracy can become com-
promised when delivering only a few MUs per segment, due
II.D.1. Dose grid and fluence grid parameter to the LINAC’s servo requiring a few MUs to stabilize the
beam. Sharpe et al. reported dosimetric output fluctuations of
From this point forward, grid resolution variation will refer the order of 2% for 1–3 MU deliveries on an Elekta LINAC.27
to the resolution variation of both dose grid and fluence grid
resolution. If dose grid resolution and fluence grid resolution
are not equal, then the dose grid resolution and the fluence II.D.4. Minimum MU and segment area
grid resolution will be noted separately. parameters (MMUSAP)
Four trials with final calculation grid resolutions of 2, Since both minimum MU and minimum segment area af-
3, and 4 mm, in addition to a 4 mm dose grid with a 2 fect the plan’s segments and modulation by different means,
mm fluence grid resolution, were computed for each patient the effect of these two parameters changing together was also
(Table I). These four trials were chosen based on the TPS re- examined. A set of 20 trials for MMUSAP (i.e., MMUP +
quirement that the dose grid resolution be a multiple of the flu- MSAP), varying the minimum MU parameter from 3 to 10
ence grid resolution. While a grid resolution of greater than 2 MU and the segment area parameter from 4 to 16 cm2 , were
mm is considered adequate for IMRT planning,17–19 any uni- computed for each patient (Table I).
form grid resolution of greater than 4 mm is considered inac- The MMUSAP parameter was included in this study to
curate in terms of dose calculation quality.18 validate that the MMUP and MSAP trends would be main-
Optimization grid resolution was kept constant as 4 mm tained when simultaneously changing both parameters. Thus,
among the four trials. Calculation times for each trial were MMUSAP trials were examined for their trend, and were not
recorded for determining planning efficiency. As previously taken to the maximum limits of 49 cm2 and 35 MU as in
noted in Sec. II.C, after the final dose calculation all plans Secs. II.D.2 and II.D.3.
were scaled by the appropriate number of monitor units to
meet D95% of the PTV 70.
II.E. Treatment delivery time model
II.D.2. MSAP II.E.1. Model derivation
The MSAP controls the plan’s modulation by limiting the Delivery time for step and shoot IMRT has been previously
MLC area to greater than a minimum size for each segment. described by Li and Xing, and serves as the basis of our de-
Twelve trials per patient of varying minimum segment area livery time model.20 Li and Xing20 previously reported an av-
from 3 to 49 cm2 were computed (Table I). To have a greater erage deviation of 2.1% with a maximum difference of 3.9%
evaluation of the parametric trend, we took the minimum pa- (16 s) between calculated and measured times on a Varian
rameter limit to one unit below our default value of 4 cm2 . TrueBeam. The Li and Xing20 model provides a valid for-
The maximum limit for MSAP was selected in consideration malism for LINACs with gridded guns and delivery methods
of the mean high risk treatment volume (Table I). The mean without a gantry position verification control point. However,
PTV 70 volume 241 cm3 (planar area 39 cm2 ) would be sat- a more detailed model is needed to account for machine-
isfied as far as coverage by a maximum segmental area of derived metrics. Also expanding the model to include fac-
49 cm2 . tors for delivery with nongridded guns, unequally spaced
The default value for MSAP was selected based on the beam geometry, and MOSAIQ Radiation Oncology Informa-
mechanical limitations and dosimetric effects of an Elekta tion System (Elekta Oncology Systems, Norcross, GA) auto-
MLC.24–26 The positional accuracy of the Elekta MLC has field-sequence (AFS) mode would allow for a more universal
been reported to be within 0.4 mm.25, 26 For example, a 1 mm model.

Medical Physics, Vol. 40, No. 6, June 2013


061704-6 Mittauer et al.: Parametric analysis of IMRT planning 061704-6

TABLE IV. Inputs, parameters, and outputs for treatment delivery time model.

Input (plan specific) Parameter (machine specific) Output

Name Symbol Name Symbol Name Symbol

dt G
Gantry degree spacing xG Gantry speed dx Time Ttotal
No. of beams NB Gantry lag toG
dt MLC
MLC travel distance (cm) xMLC MLC speed dx
No. of segments NSeg MLC lag + intersegmental beam-on delay toMLC
MU per fraction xMU AFS-gantry delay TCPG
dt MU
Dose rate dx
Interbeam beam-on delay toMU

Notes: AFS, auto-field-sequence mode. (1) Inputs are treatment plan dependent and correspond to Eq. (4). (2) Parameters
are machine dependent and correspond to Eqs. (1)–(4). (3) Output is the total treatment time for any step and shoot IMRT
plan.

For the success of such a model, parameters unique to the results of Eqs. (1)–(3), with the exception of the following
treatment machine must first be defined. Second, inputs from terms, because they are plan dependent: xG (the gantry degree
the patient-specific treatment plan would need to be deter- spacing between first and last beam), xMLC (the sum of the
mined. Such model (Table IV) would then use corresponding maximum leaf travel distance of each segment), and xMU (the
plan-based inputs and machine-based parameters to output the MU per fraction).
total treatment delivery time. This model relies on the following assumptions: TCPG ap-
Three independent equations can be formulated to model plies only to MOSAIQ users with AFS mode on an Elekta
the machine-based parameters from the linear relationship of LINAC, and was determined to be 4.9 s. AFS mode uses an in-
parametric incremental change and time: terbeam control point to verify the gantry position once a new
dt G G gantry angle has been reached. The MLC is able to update
TG = x + toG , (1) as the gantry rotates to the next beam, thus the first segment
dx
for each beam is excluded from the MLC speed and interseg-
mental beam-on delay; instead, the first segment of each beam
dt MLC MLC uses interbeam beam-on delay.
T MLC = x + toMLC , (2)
dx

dt MU MU II.E.2. Parameter acquisition


T MU = x + toMU , (3)
dx To assess the treatment delivery for this study, an Elekta
where x represents the displacement (e.g., gantry degree spac- Synergy LINAC was used to determine the machine-specific
ing, MLC travel distance, or the number of delivered monitor parameters of Eqs. (1)–(3). The machine parameters along
units), t0 is the lag time associated with the ramp-up (e.g., ac- with the individual plan inputs were used in the delivery time
dt model [Eq. (4)] to calculate the delivery time for all 495 treat-
celeration and/or deceleration), and dx is the inverse of the
velocity. Specifically, toMLC
is the intersegmental beam-on de- ment plans.
lay and MLC lag, and toMU is the interbeam beam-on delay. To verify the accuracy of the delivery time model, a se-
With the inclusion of the above machine-derived param- lection of 33 plans were delivered with step and shoot deliv-
eters with the plan-specific input values, the total treatment ery using MOSAIQ AFS mode. Measured times were then
delivery time for any unique IMRT plan can be determined. compared to calculated times from our model and a modified
The machine-characterized, plan-specific delivery time model Li and Xing20 model. We modified the Li and Xing20 model
for step and shoot delivery becomes to represent Elekta delivery by including the following fac-
tors from our model: intersegmental beam-on delay, inter-
dt G G
T total ∼
= (x ) + toG (N B − 1) + T CPG (N B − 1) beam beam-on delay, and AFS gantry delay.
dx As previously noted in Sec. II.E.1, the user must first char-
dt MLC MLC   acterize their machine. Details of this parameter acquisition
+ (x ) + toMLC N Seg − N B
dx will be described here. Three simple tests were developed to
characterize the LINAC’s metrics in terms of gantry, MLC,
dt MU MU
+ (x ) + toMU (N B ), (4) and dose rate. Such novel method allows for the user to eas-
dx ily parameterize their machine for delivery time. Each test
where NB is the number of beams, NSeg is the number of seg- measured a parameter’s rate of change through intervals rang-
ments, and TCPG is the gantry verification control point time. ing from the smallest measurable amount of change to the
The remaining terms of Eq. (4) are taken directly from the largest practical amount of change. The data were plotted as

Medical Physics, Vol. 40, No. 6, June 2013


061704-7 Mittauer et al.: Parametric analysis of IMRT planning 061704-7

the amount of change, x, versus time, T, and a linear fit was trials, like the other redundant trials listed in Table I, have the
performed. The three tests are as follows: same plan quality results. Evaluated dose was taken from the
DVH, using the dose criteria in Table III: mean dose (larynx,
(1) Gantry speed was measured in service mode by man-
parotid), 0.1 cc dose (spinal cord, cochleae, and brainstem),
ually rotating the gantry through various incremen-
D20% and D99% (PTV 56, PTV 70), and D95% (PTV 56).
tal angles, e.g., 10o , 20o , 70o , 90o , 180o , 200o , 340o ,
PTV 70 D95% was excluded since all plans were scaled to be
and 360o while measuring the time. No difference was
equal at D95%. Percentage differences for PRVs and PTVs
found between counterclockwise and clockwise rota-
were calculated between the evaluated DVH dose of each trial
tion, but this could be an issue with an unbalanced
and the DVH dose from its respective reference plan. These
gantry.
percentage differences will be referred to as dose differences,
(2) MLC speed was determined by delivering an IMRT
which are the dose calculation differences in DVH results that
plan designed with symmetric leaf sequences and sin-
result from changing the specified parameter.
gle direction leaf sequences. The sequences were rep-
Note for MSAP, MMUP, and MMUSAP plans, statistical
etitions of transitions from small fields to large fields,
results were eliminated for PRVs where dose was negligi-
and vice versa. Thus, both MLC extension and retrac-
ble; specifically, if the dose was less than 50% of either the
tion was accounted for. Time was calculated from the
0.1 cc or mean dose objective from Table III. These PRVs
LINAC “beam-off” indicator at the completion of one
are not statistically relevant because of the ease of achieving
segment to the delivery of the first MU on the next seg-
optimization objectives.
ment. Primary jaws and backup jaws were also tested,
and for an Elekta LINAC their speed was negligible
(i.e., faster) compared to the MLC speed.
III.A. Dose and fluence grid parameter
Note for nongridded guns there is a beam-on delay due
to the dose rate and gun ramp-up period. This beam- Dose differences among the grid resolutions of 2, 4, and
on lag time is inherently included in the above testing the 4 mm dose grid with 2 mm fluence grid resolution are
condition. However, this beam-on lag is representative displayed on the cumulative DVH (Fig. 3) for case no. 9. The
of the IMRT intersegmental beam-on delay and not the results for case no. 9 were representative of the results for all
interbeam beam-on delay. The subtle difference is that cases. The 3 mm grid resolution was eliminated from Fig. 3 to
in the former intersegmental beam-on delay the gun re- better display the maximum and minimum DVH differences
mains in active mode and in the latter interbeam beam- for grid resolution.
on delay the gun current must ramp-up from standby For PTVs and PRVs, the 4 mm dose grid with 2 mm flu-
mode to active mode. The latter contributes to a greater ence grid resolution was found to be most comparable in
time delay and is measured in the following test. Note terms of DVH to the 2 mm grid resolution in Fig. 3. The
that this test does not decouple MLC lag from acceler- 4 mm grid resolution varied the most from the 2 mm grid
ation/deceleration and intersegmental beam-on delay. resolution (Fig. 3). Specifically, Fig. 3 shows PTV 70 for the
From our observations, MLC lag is much less of a con- 4 mm grid resolution lacking a steep gradient; this represents
tribution to time compared with the beam-on delay. a breakdown of dose coverage and loss of dose homogeneity.
(3) Dose rate and interbeam beam-on delay were deter- Note that the DVH results for the 3 mm grid resolution fell
mined by delivering static fields of varying MU from
1 to 20 MU. Static fields were used to measure the
delay at the first segment of each new beam as the
gun warmed up out of standby mode, thus character-
istic of interbeam beam-on delay. Time was measured
between the “beam-on” and the “beam-off” indicator.
For gridded gun machines instantaneous dose rate is
possible because the gun remains in active mode, thus
the interbeam beam-on delay would be negligible.

III. RESULTS AND DISCUSSION


The differences, correlations, and tradeoffs associated with
the IMRT planning parameters were evaluated in terms of de-
livery and planning efficiency and plan quality. We also derive
recommendations to the user.
Trial plans were compared to a specified reference plan,
as noted in Table I and Sec. II.D: using 2 mm grid resolution
F IG . 3. DVH curves of PRVs and PTVs for case no. 9 for the grid resolu-
and 3 MU with 4 cm2 for MSAP, MMUP, and MMUSAP. tion parameter. Notes: (1) medium solid curve represents 2 mm dose/fluence
Note that MSAP, MMUP, and MMUSAP reference plans had grid, thin solid curve represents 4 mm dose/fluence grid, and dashed curve
the same default IMRT parameters. Thus, these reference represents 4 mm dose grid with 2 mm fluence grid, (2) flu., fluence.

Medical Physics, Vol. 40, No. 6, June 2013


061704-8 Mittauer et al.: Parametric analysis of IMRT planning 061704-8

TABLE V. PRV and PTV dose differences for grid resolution parameter.

PRV Dmax Dmax D

Grid resolution (mm) Cochlea Serial Parallel All PRVs

3 13.3% 5.8% 0.9% 1.0%


4 7.9% 5.3% 1.5% 1.2%
4, 2 flu. 5.5% 3.5% 0.7% 0.7%

PTV Dmax Dmax D


Grid resolution (mm) PTV 70 D99% All PTVs All PTVs
3 2.8% 1.4% 0.5%
4 8.2% 3.3% 1.0%
4, 2 flu. 6.2% 1.9% 0.5%

Notes: flu., fluence; Dmax , max dose difference; Dmax , mean max dose difference; D, mean dose difference. (1) 3
and 4 mm denote matching dose/fluence grid. (2) 4, 2 flu. denotes 4 mm dose grid with 2 mm fluence grid. (3) Reference
taken at 2 mm dose/fluence grid. (4) Dmax was the cochlea and PTV 70 D99% among all cases.

between the DVH for the 4 mm dose grid with 2 mm fluence higher dose differences for serial than for parallel PRVs. Since
grid resolution and the DVH for the 4 mm grid resolution. the serial structures were defined by 0.1 cc dose objectives and
Table V lists the maximum dose difference (Dmax ), the parallel structures were typically defined by mean dose ob-
mean dose difference (D), and the mean maximum dose jectives, the same conclusion as above applies here that small
difference (Dmax ), for PRVs and PTVs. All dose differences volumetric objectives have greater dose variability due to grid
in Table V were measured as absolute percentage differences resolution.
from the 2 mm grid resolution reference plan. To best de- The greatest dose deviation for the PTV maximum dose
termine the dose variability of the grid resolution, sign con- difference in Table V was 8.2% or 5.5 Gy, for PTV 70 D99%,
vention was not used because we sought to measure the ab- 4 mm grid resolution. Thus, overall, the dose variability due
solute change of dose due to grid resolution. Note that each to grid resolution was found to be as high as 8.2% for PTVs
dose objective for PTV 56 and PTV 70 was considered sepa- and 13.3% for PRVs.
rately when averaging in the last two columns as five criteria: One anomaly in Table V was that the 3 mm grid resolu-
PTV 56 (D99%, D20%, D95%) and PTV 70 (D99%, D20%). tion PTV maximum dose difference and PTV mean maximum
The PRV maximum dose difference, being the greatest dose difference were more comparable to the 2 mm grid res-
PRV percentage difference for all cases of that grid resolu- olution than the 4 mm dose grid with 2 mm fluence grid res-
tion, was the cochlea at 13.3% or 2.1 Gy (3 mm grid reso- olution. This result was only seen for PTV 70. The anomaly
lution); 7.9% or 1.8 Gy (4 mm grid resolution); and 5.5% or is attributed to the plans being scaled to PTV 70 D95% using
1.2 Gy (4 mm dose grid with 2 mm fluence grid resolution) the nearest whole MU. Thus, whole number approximation
(Table V). This result was due to the cochlea having a 0.1 cc led to some plans being scaled higher than exactly D95%, re-
volumetric objective where dose is determined by the tail-end sulting in D99% and D20% being somewhat inconsistent for
of the DVH. As expected, small volumes with small volumet- PTV 70.
ric objectives were more susceptible to dose inaccuracies due The PRV and PTV mean dose differences remained under
to grid resolution. 1.2% for all trials (Table V), meaning the effect of grid res-
For PRVs, the mean maximum dose difference of Table V olution averaged over all structures was less than 1.2%. Note
is the mean across each maximum PRV percentage difference. there was no difference between dose objectives passing or
PRV mean maximum dose difference was calculated by deter- failing among the four grid resolutions for all cases.
mining the greatest deviation for each PRV, and then averag- The effect of grid resolution on calculation time and plan
ing the results among serial and parallel PRVs. Table V shows MU is presented in Table VI. The mean plan MU (Table VI)

TABLE VI. Planning calculation time and total plan MU for grid resolution parameter.

Calc. time Total MU for plan

Grid resolution (mm) 3 4 4, 2 flu. 3 4 4, 2 flu.

Mean − 69.8% − 84.4% − 62.7% 0.8% 1.8% 0.4%


Std. dev. 3.0% 2.0% 5.6% 0.5% 0.5% 0.4%
2 mm reference Mean Calc. Time 3.0 min Mean MU 645 MU

Note: flu., fluence grid; 2 mm, 2 mm dose/fluence grid; 3 mm, 3 mm dose/fluence grid; 4 mm, 4 mm dose/fluence grid; 4, 2 flu., 4 mm dose grid with 2 mm fluence grid.

Medical Physics, Vol. 40, No. 6, June 2013


061704-9 Mittauer et al.: Parametric analysis of IMRT planning 061704-9

increased by 0.8% (3 mm grid resolution), 1.8% (4 mm grid III.B. Minimum segment area parameter
resolution), and 0.4% (4 mm dose grid with 2 mm fluence
grid resolution). The planning time (Table VI) was reduced by Figures 4(a)–4(d) illustrate the effects on dose for PTVs
69.8% (3 mm grid resolution), 84.4% (4 mm grid resolution), and PRVs with varying MSAP and MMUP. The PTV and
and 62.7% (4 mm dose grid with 2 mm fluence grid resolu- PRV dose differences are noted separately as mean [Figs. 4(a)
tion), compared to the mean calculation time for the reference and 4(b)] and maximum dose differences [Figs. 4(c) and 4(d)]
plan of 3 min. with respect to the reference plan (Table I). Note that Fig. 4
The 4 mm dose grid with 2 mm fluence grid resolution adheres to sign convention. A positive dose difference denotes
saved considerable calculation time, about 2/3 the calculation plan quality degradation for all PRVs and for PTV D20% (i.e.,
time of the 2 mm grid resolution (Table VI). As previously PTV hotspot). On the other hand, a negative dose difference
noted, the 4 mm dose grid with 2 mm fluence grid resolu- denotes a loss of plan quality for PTV D95% and D99% (i.e.,
tion was most comparable to the 2 mm grid resolution for PTV coverage).
plan quality results (Fig. 3 and Table V). This indicates that a For MSAP there was a marked increase in the maximum
4 mm dose grid with 2 mm fluence grid resolution is a viable dose difference in Fig. 4(c) from 4% (MSAP of 5 cm2 ) to
alternative to a 2 mm grid resolution setting with considera- 11% (MSAP of 6 cm2 ). Then, dose differences remained be-
tion of dose accuracy and calculation time. low 12% for plans with MSAP < 25 cm2 . Note the anomaly
Technically, all segments are the same among plans, since at a MSAP of 3 cm2 [Fig. 4(c)]. Here, the cochlea had a large
plans were optimized as the one before the grid resolution maximum dose difference at 6%, but remained under its dose
was changed; thus, there is no effect from optimization. Dose objective.
differences are, therefore, a direct measure of the uncertain- Another measure of how MSAP changes plan quality is
ties of the dose calculation after the plan has been scaled by the threshold of MSAP that dose objectives fail. The dose
the appropriate MUs to achieve the prescription (i.e., PTV 70 to PRVs remained within the objectives for all plans with a
D95%). Since it is known that interpolation of a larger grid MSAP of <6 cm2 . For the PTV objectives, PTV 56 D95%
resolution leads to these dose differences,17–19 interpolation and D99% failed at greater than 8 and 5 cm2 , respectively.
was the cause of the dose volume index uncertainty. PTV 70 D20% and D99% objectives also failed at 25 and

F IG . 4. Mean (a) and (b) and maximum (c) and (d) percentage dose differences for all PRVs and PTVs with varying MSAP (a) and (c) and MMUP (b) and (d)
for all cases. Notes: (1) The max is the max magnitude of plan quality degradation being a negative sign for D20% and D99% and a positive sign for all other
structures. (2) Reference taken at 3 MU, 4 cm2 .

Medical Physics, Vol. 40, No. 6, June 2013


061704-10 Mittauer et al.: Parametric analysis of IMRT planning 061704-10

III.D. Minimum MU and segment area parameters


Figure 5 displays the two parameters MSAP and MMUP
changing simultaneously as designated by MMUSAP. Dose
differences were averaged over all PRVs and PTVs (D20%
only), with respect to the reference plan, as noted in Table I.
Note that Fig. 5 uses only D20% for PTVs, which denotes
the hotspot, or the maximum dose encompassing 20% of the
PTV. Figure 5 adheres to sign convention, and thus only an
increase in dose is relevant.
When simultaneously changing both MSAP and MMUP,
the trends of MSAP and MMUP (Fig. 4) were maintained for
F IG . 5. Percentage dose differences averaged over all PRVs and PTVs with
the MMUSAP (Fig. 5). Mean dose differences for MMUSAP
varying MMUSAP for all cases. Notes: (1) Only D20% is shown for PTV
results due to sign convention. (2) Reference taken at 3 MU, 4 cm2 . remained under 2.5%, but were slightly larger than the re-
spective MSAP or MMUP dose differences of Fig. 4. This
36 cm2 , respectively. Limiting the MSAP to ≤5 cm2 kept all means that as both parameters were increased individually or
PRVs and PTVs (except for PTV 56 D99%) within the dose together the dose distribution varied more (i.e., the plan be-
objectives as well as all dose deviations to under 4% (Fig. 4). came hotter) from reference 3 MU, 4 cm2 plan.

III.C. Minimum MU parameter


III.E. Delivery time model
Figure 4(b) shows a marked increase in PRV mean dose
differences for plans of MMUP > 10 MU. As the MMUP in- The machine-derived parameters of our model were com-
creases, the PTVs lose dose homogeneity. Specifically, this pared against vendor specifications. The results are tabulated
result was seen with PTV D20% for MMUP > 10 MU in Table VII along with results converted into vendor equiva-
[Figs. 4(b) and 4(d)]. lent units for ease of comparison.
As described in Sec. III.B, thresholds at which dose objec- The delivery time model [Eq. (4)] was compared and ver-
tives are maintained serve as another evaluation of the plan ified against 33 delivered plans with times that ranged from
quality. For all plans with MMUP of <7 MU, dose to the 3 to 12 min. Figure 6 shows the results of measured times
PRVs remained within the dose objectives. For the PTVs, for the 33 plans versus calculated times using our model and
dose objectives were met for plans of MMUP <20 MU for the modified Li and Xing20 model (i.e., altered for Elekta de-
PTV 56 (D95%) and PTV 70 (D20% and D99%). For PTV livery). The greatest deviations between measured and cal-
56 D99%, plans failed with MMUP ≥4 MU. A MMUP ≤ 5 culated times were 1.8% (10 s for a 9 min plan) and 5.7%
MU allowed all dose objectives to be met (except for PTV 56 (10 s for a 3 min plan) and the average deviations were 0.83%
D99%) and dose deviations to be under 4% (Fig. 4). and 1.7% for respective calculated times using our model and
Overall, plan quality degrades with increasing MSAP or the modified Li and Xing20 model.
MMUP (Fig. 4). The effect of MSAP was more marked on The Li and Xing20 model provides a valid formalism for
PRVs than on PTVs [Fig. 4(a)]. However, the effect of MMUP specific LINACs and delivery types, as previously noted in
was evident on both PRVs and PTV D20% [Fig. 4(b)]. This Sec. II.E. Our method provides new machine-derived met-
suggests that MMUP may be relaxed, if PRV dose tolerances rics and additional terms for a more detailed account of time.
and the PTV hotspot are not of concern. But if the PTV Our model also allows for factors not previously included in
hotspot is an issue, then only MSAP may be relaxed, but cau- the Li and Xing20 model such as the geometry of unequally
tion should still be used for the dose to PRVs. spaced beams, nongridded gun delay due to intersegmental

TABLE VII. Delivery time model parameters of an Elekta LINAC compared to vendor specifications.

Delivery time model parameters Comparison

Parameter Name Measured results (10−3 ) Numerical equivalent Elekta specifications


dt
MLC speed dx 7.70 min/cm 2.2 cm/s 2 cm/s
to 51.17 min 3.1 s/intersegment N/A
dt
Gantry speed dx 2.65 min/degree 57.2 s/rev 60 s/rev
to 36.06 min 2.2 s/interbeam N/A
AFS-gantry T CPG 81.67 min 4.9 s/interbeam N/A
dt
Dose rate dx 1.87 min/MU 526.3 MU/min 535 MU/mina
to 79.18 min 4.8 s/interbeam N/A

Notes: (1) Measured results from characterization testing [Eqs. (1)–(3)] of an Elekta Synergy. (2) Numerical equivalent are measured results converted into the same units
as the vendor specifications.
a
Dose rate is unique to each Elekta LINAC.

Medical Physics, Vol. 40, No. 6, June 2013


061704-11 Mittauer et al.: Parametric analysis of IMRT planning 061704-11

utilize all available segments to provide the necessary plan


conformality and PTV coverage. However, when the MSAP
was increased, similar segment shapes were created, the ef-
ficiency of generating plan conformality decreased, the MU
required to obtain PTV coverage decreased, and the overall
result resembles a 3D conformal plan, (i.e., the total MU de-
creased with an increase in the MSAP) [Fig. 7(a)]. Note that
the total delivery time was largely dependent on the number
of segments (Fig. 7), because the MLC movement is the most
time-costly parameter of the whole IMRT delivery process as
described in Sec. III.F.2.
On the other hand, by limiting the MMUP, the number
F IG . 6. Treatment delivery time model tested against 33 delivered IMRT
of segments was affected considerably. As the MMUP is in-
plans with calculated times from our model and a modified Li and Xing creased, the plan was forced to consolidate segments. This
(Ref. 20) model. Notes: (1) The modified Li and Xing (Ref. 20) model in- was because the plan attempts to maintain the same total dose
cludes two additional terms for beam-on delay and auto-field-sequence mode (i.e., similar MU) to the PTV for a given set of dose objec-
delay. (2) Measured times delivered on an Elekta LINAC. tives. The plan MU remained approximately constant only up
to 10 MMUP; beyond 10 MMUP the plan lost conformality
beam-on delay as well as interbeam beam-on delay, and a and efficiency increased (i.e., the plan became similar to a 3D
gantry control point delay due to the MOSAIQ AFS mode conformal plan) as seen in Fig. 7(b) and, as expected, the to-
to verify gantry position. Overall, our proposed delivery time tal delivery time decreased significantly with the number of
formalism provides a valid method to assess delivery time, segments.
and allows for the characterization of the user’s specific ma- From our previous results of Secs. III.B and III.C, we
chine, a novel technique. found that the effect of MMUP was more costly in terms of
plan quality than the effect of MSAP [Figs. 4(c) and 4(d)].
From the above discussion we can also conclude that the ef-
III.F. Number of segments and plan MU fect of MMUP on the number of segments was more costly to
To assess how varying the parameters of MSAP, MMUP, the plan quality than the effect of MSAP on the plan MU.
and MMUSAP affects the plan quality and delivery time, the
number of segments and plan MU were evaluated. Figure 7
III.F.2. Delivery efficiency
displays the results as percentage differences in the delivery
time, total number of segments, and plan MU, with the refer- The calculated delivery times for all cases are shown in
ence plans of Table I. Fig. 7 as percentage differences with the mean delivery time
of 9.1 min for reference plans. Delivery efficiency (i.e., treat-
ment delivery time) improved at and above 6 MU for MMUP,
III.F.1. Plan quality
but no marked improvement was found with MSAP (Fig. 7).
The factors affecting the dose distribution for MSAP can For MMUSAP, delivery time was the most efficient in com-
be seen in light of the segment shape and the plan MU. As parison to plans with the same MSAP or same MMUP, as
the MSAP is increased, the total number of segments did not expected.
change considerably [Fig. 7(a)] due to the fact that the MMUP Thus, it is the intersegmental time (i.e., MLC speed and
remained small. Therefore, the optimizer still attempted to gun ramp-up delay) that is more costly than intrasegmental

F IG . 7. Percentage difference in treatment delivery time, number of segments, and plan MU for the 3 MU, 4 cm2 reference plan versus plans of varying sizes
of MSAP (a) and MMUP (b) for all cases.

Medical Physics, Vol. 40, No. 6, June 2013


061704-12 Mittauer et al.: Parametric analysis of IMRT planning 061704-12

time (i.e., number of MU delivered). Although it was not ACKNOWLEDGMENT


seen for the cases in this study, the plan MU could affect
The authors report no conflicts of interest in conducting the
the delivery times for dose rate dependent plans. Li and Xing
research.
showed that a dose rate of 1400 MU/min compared to 400
a) Author to whom correspondence should be addressed. Electronic mail:
MU/min can improve delivery efficiency by up to 15% for the
liucr@shands.ufl.edu
same head and neck IMRT plan.20 Thus, for dose rate depen- 1 T. Bortfeld, “IMRT: A review and preview,” Phys. Med. Biol. 51, R363–
dent cases or for future delivery capabilities with an improved R379 (2006).
MLC speed design,20 treatment efficiency could be improved 2 B. Lu, J. Li, D. Kahler, G. Yan, K. Mittauer, W. Shi, P. Okunieff, and C. Liu,

with MSAP and/or MMUP. But for most current clinical “An approach for online evaluations of dose consequences caused by small
rotational setup errors in intracranial stereotactic radiation therapy,” Med.
cases only MMUP, and not MSAP, can improve delivery Phys. 38, 6203–6215 (2011).
efficiency. 3 J. Peng, D. Kahler, J. Li, S. Samant, G. Yan, R. Amdur, and C. Liu, “Char-

acterization of a real-time surface image-guided stereotactic positioning


system,” Med. Phys. 37, 5421–5433 (2010).
4 R. Mohan, M. Arnfield, S. Tong, Q. Wu, and J. Siebers, “The impact of

III.G. Treatment quality considerations fluctuations in intensity patterns on the number of monitor units and the
quality and accuracy of intensity modulated radiotherapy,” Med. Phys. 27,
A longer treatment time will not only increase the chance 1226–1237 (2000).
5 X. Sun and P. Xia, “A new smoothing procedure to reduce delivery seg-
of patient movement but also decrease the biological response
ments for static MLC-based IMRT planning,” Med. Phys. 31, 1158–1165
of the treatment. Paganetti found that the biological response (2004).
is reduced by 12% for a 30 min delivery time as compared 6 L. Ma, “Smoothing intensity-modulated treatment delivery under hardware

to a 1 min treatment.28 Increasing the beam-on time also in- constraints,” Med. Phys. 29, 2937–2945 (2002).
7 S. Spirou, N. Fourier-Bidoz, J. Yang, C. Chui, and C. Ling, “Smoothing
creases radiation leakage, which could lead to normal tissues
intensity-modulated beam profiles to improve the efficiency of delivery,”
receiving up to 20% of the prescribed dose.29 Med. Phys. 28, 2105–2112 (2001).
Another consideration with regards to the treatment quality 8 M. Coselmon, J. Moran, J. Radawski, and B. Fraass, “Improving IMRT

is the deliverability of the plan. The machine’s capabilities, delivery efficiency using intensity limits during inverse planning,” Med.
as described in Sec. II.D, must be taken into consideration Phys. 32, 1234–1245 (2005).
9 M. Matuszak, E. Larsen, and B. Fraass, “Reduction of IMRT complexity
when limiting inverse planning parameters. Dosimetric and through the use of beam modulation penalties in the objective function,”
mechanical accuracy testing can be addressed using AAPM Med. Phys. 34, 507–520 (2007).
TG 142 guidelines.30 10 M. Matuszak, E. Larsen, K. Jee, D. McShan, and B. Fraass, “Adaptive dif-

To create a more comprehensive understanding of mod- fusion smoothing: A diffusion-based method to reduce IMRT field com-
plexity,” Med. Phys. 35, 1532–1546 (2008).
ulation parameters, research is needed in the area of opti- 11 D. Craft, P. Suus, and T. Bortfield, “The tradeoff between treatment plan
mization objectives to allow for a more efficient IMRT work- quality and required number of monitor units in intensity-modulated radio-
flow. Studying the effect on plan quality and planning time therapy,” Int. J. Radiat. Oncol., Biol., Phys. 67, 1596–1605 (2007).
12 S. Webb, “A simple method to control aspects of fluence modulation in
when changing the optimization grid resolution would be a
IMRT planning,” Phys. Med. Biol. 46, N187–N195 (2001).
useful evaluation for the current clinical practice. Expand- 13 C. McGarry, C. Chinneck, M. O’Toole, J. O’Sullivan, K. Prise, and

ing the delivery time model to include sliding window de- A. Hounsell, “Assessing software upgrades, plan properties and patient
livery and even modulated arc therapy is yet another area of geometry using intensity modulated radiation therapy (IMRT) complexity
metrics,” Med. Phys. 38, 2027–2034 (2011).
interest. 14 A. McNiven, M. Sharpe, and T. Purdie, “A new metric for assessing IMRT

modulation complexity and plan deliverability,” Med. Phys. 37, 505–515


(2010).
15 E. Ludlum and P. Xia, “Comparison of IMRT planning with two-step and
IV. CONCLUSION one-step optimization: A way to simplify IMRT,” Phys. Med. Biol. 53,
807–821 (2008).
Our study has shown for a fixed number of beams and 16 B. Hårdemark, A. Liander, H. Rehbinder, and J. Loef, “P3IMRT direct ma-

gantry angles, and the greater the number of segments, the chine parameter optimization,” Pinnacle White Paper, No. 4535 983 02483,
better the plan in terms of dose distribution. However, this oc- Philips (2004).
17 A. Niemierko and M. Goitein, “The influence of the size of the grid used
curred at the expense of increased delivery time. Thus, modu- for dose calculation on the accuracy of dose estimation,” Med. Phys. 16,
lation factors such as MMUP, which largely impact the num- 239–247 (1989).
18 J. Dempsey, H. Romeijin, J. Li, D. Low, and J. Palta, “A Fourier analysis
ber of segments, greatly affected plan quality and delivery
time. Based on our results, we give the following planning of the dose grid resolution for accurate IMRT fluence map optimization,”
Med. Phys. 32, 380–388 (2005).
parameter setting recommendations: (1) 5 cm2 for MSAP, (2) 19 H. Chung, H. Jin, J. Palta, T. Suh, and S. Kim, “Dose variations with vary-
5 MU for MMUP, and (3) 4 mm dose grid with 2 mm fluence ing calculation grid size in head and neck IMRT,” Phys. Med. Biol. 51,
grid resolution for the final dose calculation settings. When 4841–4856 (2006).
20 R. Li and L. Xing, “Bridging the gap between IMRT and VMAT: Dense an-
required, delivery efficiency can be improved by increasing
gularity sampled and sparse intensity modulated radiation therapy,” Med.
either the MMUP or MMUSAP, but not the MSAP. However, Phys. 38, 4912–4919 (2011).
we do not recommend increasing these modulation parame- 21 R. Popple, J. Fiveash, and I. Brezovich, “Effect of beam number on organ-

ters beyond our recommended settings, for the sake of deliv- at-risk sparing in dynamic multileaf collimator delivery of intensity modu-
ery efficiency, due to the overall cost to plan quality. The pro- lated radiation therapy,” Med. Phys. 34, 3752–3759 (2007).
22 J. Li, C. Liu, S. Kim, R. Amdur, and J. Palta, “Matching IMRT fields with
posed delivery time model provides a valid method to assess static photon field in the treatment of head-and-neck cancer,” Med. Dosim.
delivery time. 30, 135–138 (2005).

Medical Physics, Vol. 40, No. 6, June 2013


061704-13 Mittauer et al.: Parametric analysis of IMRT planning 061704-13

23 L. Marks, E. Yorke, A. Jackson, R. Ten Haken, L. Constine, A. Eisbruch, 27 M. Sharpe, B. Miller, D. Yan, and J. Wong, “Monitor unit settings for in-
S. Bentzen, J. Nam, and J. Deasy, “Use of normal tissue complication prob- tensity modulated beams delivered using a step-and-shoot approach,” Med.
ability models in the clinic,” Int. J. Radiat. Oncol., Biol., Phys. 76, S10–S19 Phys. 27, 2719–2725 (2000).
(2010). 28 H. Paganetti, “Changes in tumor cell response due to prolonged dose de-
24 T. J. Jordan and P. C. Williams, ‘‘The design and performance char- livery times in fractionated radiation therapy,” Int. J. Radiat. Oncol., Biol.,
acteristics of a multileaf collimator,’’ Phys. Med. Biol. 39, 231–251 Phys. 63, 892–900 (2005).
(1994). 29 E. E. Klein and D. A. Low, “Interleaf leakage for 5 and 10 mm dynamic
25 C. Liu, T. Simon, C. Fox, J. Li, and J. Palta, “Multileaf collimator character- multileaf collimation systems incorporating patient motion,” Med. Phys.
istics and reliability requirements for IMRT Elekta system,” Int. J. Radiat. 28, 1703–1710 (2001).
Oncol., Biol., Phys. 71, S89–S92 (2008). 30 E. E. Klein, J. Hanley, J. E. Bayouth, F. Yin, W. Simon, S. Dresser, C.
26 T. Simon, D. Kahler, W. Simon, C. Fox, J. Li, J. Palta, and C. Liu, “An Serago, F. Aguirre, L. Ma, B. Arjomandy, and C. Liu, “Task Group 142
MLC calibration method using a detector array,” Med. Phys. 36, 4495– report: Quality assurance of medical accelerators,” Med. Phys. 36, 4197–
4503 (2009). 4212 (2009).

Medical Physics, Vol. 40, No. 6, June 2013

You might also like