Professional Documents
Culture Documents
7000DP Gseng0000
7000DP Gseng0000
Defibrillator Analyzer
Impulse 7000DP
Defibrillator/Transcutaneous Pacer Analyzer
PN 3028662
August 2007
© 2007 Fluke Corporation, All rights reserved. Printed in USA
All product names are trademarks of their respective companies.
Warranty and Product Support
Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of origi-
nal purchase OR two years if at the end of your first year you send the instrument to a Fluke Biomedical service center for
calibration. You will be charged our customary fee for such calibration. During the warranty period, we will repair or at our
option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to
Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if
the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized
Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE
EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSE-
QUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration
of instruments is not covered under the warranty.
This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since
some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages,
this limitation of liability may not apply to you. If any provision of this warranty is held invalid or unenforceable by a court or
other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
7/07
Notices
All Rights Reserved
© Copyright 2007, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into
any language without the written permission of Fluke Biomedical.
Copyright Release
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs
and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.
Technical Support
For application support or answers to technical questions, either email techservices@flukebiomedical.com or call 1-800- 648-7952 or 1-425-446-6945.
Claims
Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in their original
condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifica-
tions, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local sales representative.
Standard Terms and Conditions
Refunds and Credits
Please note that only serialized products and their accessory items (i.e., products and items bearing a distinct serial number tag) are eligible for
partial refund and/or credit. Nonserialized parts and accessory items (e.g., cables, carrying cases, auxiliary modules, etc.) are not eligible for re-
turn or refund. Only products returned within 90 days from the date of original purchase are eligible for refund/credit. In order to receive a partial re-
fund/credit of a product purchase price on a serialized product, the product must not have been damaged by the customer or by the carrier chosen by the cus-
tomer to return the goods, and the product must be returned complete (meaning with all manuals, cables, accessories, etc.) and in “as new” and resalable con-
dition. Products not returned within 90 days of purchase, or products which are not in “as new” and resalable condition, are not eligible for credit return and
will be returned to the customer. The Return Procedure (see below) must be followed to assure prompt refund/credit.
Restocking Charges
Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess of 30 days af-
ter purchase, but prior to 90 days, are subject to a minimum restocking fee of 20 %. Additional charges for damage and/or missing parts and accesso-
ries will be applied to all returns.
Return Procedure
All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an instrument to
Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its
actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper
packaging or handling.
Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging:
Use a double–walled carton of sufficient strength for the weight being shipped.
Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts.
Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.
Returns for partial refund/credit:
Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our Order Entry Group at
1-800-648-7952 or 1-425-446-6945.
Repair and calibration:
To find the nearest service center, go to www.flukebiomedical.com/service or
In the U.S.A.:
Cleveland Calibration Lab
Tel: 1-800-850-4606
Email: globalcal@flukebiomedical.com
In Asia:
Everett Calibration Lab
Tel: +425-446-6945
Email: service.international@fluke.com
Certification
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was shipped from the
factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibra-
tion standards are measured against in-house performance standards using accepted test procedures.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Bio-
medical will not be responsible for any injuries sustained due to unauthorized equipment modifications.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in
this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical for the use or reliability
of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers.
Manufacturing Location
The Impulse 6000D and 7000DP Defibrillator/Transcutaneous Pacer Analyzers are manufactured at Fluke Biomedical, 6920 Seaway Blvd.,
Everett, WA, U.S.A.
Table of Contents
Title Page
Introduction .................................................................................................................... 1
Intended Use .................................................................................................................. 1
Unpacking the Analyzer ................................................................................................. 1
Safety Information .......................................................................................................... 2
Instrument Familiarization .............................................................................................. 4
Turning the Analyzer On ................................................................................................ 7
Connecting a Defibrillator and Pacer to the Analyzer ..................................................... 7
Accessing the Analyzer Tests ........................................................................................ 8
What to Do Next ............................................................................................................. 12
Maintenance................................................................................................................... 12
Cleaning the Analyzer ............................................................................................... 12
Maintaining Peak Battery Condition........................................................................... 12
Accessories .................................................................................................................... 13
Specifications ................................................................................................................. 14
General Specifications............................................................................................... 14
Defibrillator Analyzer Specifications .......................................................................... 15
Energy Output Measurement................................................................................ 15
i
Impulse 6000D, 7000DP
Getting Started Manual
ii
List of Tables
1. Symbols................................................................................................................................. 2
2. Top-Panel Controls and Connections.................................................................................... 5
3. Rear-Panel Connections ....................................................................................................... 7
4. Accessories ........................................................................................................................... 13
iii
Impulse 6000D, 7000DP
Getting Started Manual
iv
List of Figures
v
Impulse 6000D, 7000DP
Getting Started Manual
vi
Defibrillator/Pacer Analyzer
1
Impulse 6000D, 7000DP
Getting Started Manual
Conforms to relevant Canadian and US • Use this Analyzer only in the manner
) standards
specified by the manufacturer or the
protection provided may be impaired.
X Hazardous voltage
• Read the Users Manual before operating the
P Conforms to European Union directives Analyzer.
2
Defibrillator/Transcutaneous Pacer Analyzer
Safety Information
3
Impulse 6000D, 7000DP
Getting Started Manual
Instrument Familiarization
Figure 1 and Table 2 describes the top-panel controls and connections of the Analyzer .
11
2
10
3
9 275 V MAX
DEFIBRILLATOR
8 5000 Vp MAX
5 5
fak07.eps
Figure 1. Top-Panel Controls and Connections
4
Defibrillator/Transcutaneous Pacer Analyzer
Instrument Familiarization
Outputs of low-level ECG signals (RA/R, LL/F, LA/L, RL/N, V1/C1, V2/C2, V3/C3, V4/C4,
1 ECG lead connectors
V5/C5, and V6/C6).
4 Navigation buttons Cursor control buttons for navigating menus and lists.
5 Defib connectors Defibrillator connections (Shown with removable defib paddle contact plates installed).
Keys F1 through F5 are used to select from a number of selections that appear in the LCD
6 Function softkeys
display above each function softkey.
8 ECG button Opens the main menu for ECG test functions.
10 Pacer button Opens the main menu for pacer test functions (7000DP only).
11 Defibrillator button Opens the main menu for defibrillator test functions.
5
Impulse 6000D, 7000DP
Getting Started Manual
CHARGE STATUS
1 2 3 4 5
fak08.eps
Figure 2. Rear-Panel Connections
6
Defibrillator/Transcutaneous Pacer Analyzer
Turning the Analyzer On
Indicates RED while battery is charging. Indicates GREEN when the battery is fully
1 Charge Status LED
charged and the charger is still connected.
Battery Charger
2 Input connector for attaching the battery charger to the Analyzer.
connector
3 Scope output Output signal jack for displaying the defib playback wave on an oscilloscope.
4 Hi-level ECG output High-level ECG signal output jack for oscilloscope viewing.
5 Computer Port Device Port (B-style USB) for controlling the Analyzer from a PC or instrument controller.
7
Impulse 6000D, 7000DP
Getting Started Manual
W Caution
To avoid damage to the Analyzer or fak02.eps
defibrillator, do not apply defibrillator pulses Figure 4. Defib Function Menu
to the pacer inputs.
Figure 7 shows the pacer connected to either the pacer Some menu selections reveal a list of selectable items by
input jacks or the defibrillator jacks. While the pacer input displaying K to the right of the presently selected item.
jacks have a selectable load from 50 to 1500 Ω, the See Figure 5. To change the selection, press either G or
defibrillator input jacks have a fixed load of 50 Ω. H to scroll through the possible selections. Once the
desired selection appears, press the function softkey and
Figure 8 shows how to connect the ECG leads to the K disappears from the display.
Analyzer.
8
Defibrillator/Transcutaneous Pacer Analyzer
Accessing the Analyzer Tests
+
DEFIB Defibrillator/Pacer
OFF
PACER
fak11.eps
Figure 6. Defibrillator Connections
9
Impulse 6000D, 7000DP
Getting Started Manual
Defibrillator/Pacer
7000DP Only
DEFIB
OFF
PACER
50 Ω
50 - 1500 Ω Only
+
Caution
To avoid damage to the
Analyzer or defibrillator, do
not apply defibrillator pulses
to the pacer inputs.
fak10.eps
Figure 7. Pacer Connections
10
Defibrillator/Transcutaneous Pacer Analyzer
Accessing the Analyzer Tests
ECG Monitor
fak09.eps
Figure 8. ECG Connections
11
Impulse 6000D, 7000DP
Getting Started Manual
Do not pour fluid onto the Analyzer surface; To obtain the specified performance, use the
fluid seepage into the electrical circuitry may battery charger specified in this manual.
cause the Analyzer to fail.
W Caution
Do not use spray cleaners on the Analyzer;
such action may force cleaning fluid into the
Analyzer and damage electronic components.
12
Defibrillator/Transcutaneous Pacer Analyzer
Accessories
Accessories
Table 4 lists the accessories for the Analyzer. Contact your local Fluke Biomedical sales representative or go to
www.flukebiomedical.com for an up-to-date accessories list.
Table 4. Accessories
13
Impulse 6000D, 7000DP
Getting Started Manual
Specifications
General Specifications
Temperature
Operating ............................................................10 °C to 40 °C (50 °F to 104 °F)
Storage................................................................-20 °C to +60 °C (-4 °F to +140 °F)
Humidity.................................................................10 % to 90 % non-condensing
Display ...................................................................LCD display
Communications ...................................................USB device port for computer control
Modes of Operation ..............................................Manual and remote
Power .....................................................................Internal rechargeable NiMH battery pack for nine hours (typical) operation after full
charge, or the battery charger can operate the Analyzer and charge the battery
simultaneously.
Battery Charger .....................................................100 to 240 V input, 15 V/1.5 A output. For best performance, the battery charger should
be connected to a properly grounded ac receptacle.
Mechanical
Housing ...............................................................ABS Plastic
Size (H x W x L) ..................................................13 cm x 32 cm x 24 cm (5 in x 13 in x 9.5 in)
Weight .................................................................3.0 kg (6.6 lb)
Safety Standards
nd
CE .......................................................................IEC/EN61010-1 2 Edition; Pollution degree 2
CSA.....................................................................CAN/CSA- C22.2 No. 61010-1; UL61010-1
Electromagnetic Compatibility Standards (EMC)
European EMC....................................................EN61326-1
14
Defibrillator/Transcutaneous Pacer Analyzer
Specifications
Note
For Pulsed Bi-Phasic defibrillator, specified accuracy is ±(1.5 % of reading + 0.3 J) on both ranges.
Load resistance
Resistance...........................................................50 Ω
Accuracy..........................................................±1 %, non-inductive (<2 μH)
Variable external load resistance (optional) .......25, 75,100, 125, 150, or 175 Ω, All values ±1 %, non-inductive (<2 μH)
Pulse trigger level .................................................20 V
Pulse width
Range..................................................................1.0 to 50.0 ms
Accuracy..............................................................±0.1 ms
Voltage
Range..................................................................20 to 5000 V
Accuracy..............................................................±(1 % of reading + 2 V)
Current
Range..................................................................0.4 to 100.0 A
Accuracy..............................................................±(1 % of reading + 0.1 A)
15
Impulse 6000D, 7000DP
Getting Started Manual
16
Defibrillator/Transcutaneous Pacer Analyzer
Specifications
ECG Waves
ECG General
Lead configuration...............................................12-lead simulation. RA, LL, LA, RL, V1-6 with independent outputs
Lead to lead impedance ......................................1000 Ω (nominal)
Rate accuracy .....................................................±1 % of nominal
ECG Amplitudes
Reference lead ....................................................Lead 1
Settings ...............................................................0.05 to 0.45 (by 0.05) mV
0.5 to 5.0 (by 0.5) mV
Accuracy..............................................................±2 % of setting, lead I and 2 Hz square wave
For performance waves and R-wave detection, other leads are proportional to Lead I in percentage per:
Lead I ..............................................................100
Lead II .............................................................150
Lead III ............................................................50
Leads V1 through V6 ......................................100
For normal sinus waves, other leads are proportional to Lead I in percentage per:
Lead I ..............................................................100
Lead II .............................................................150
Lead III ............................................................50
17
Impulse 6000D, 7000DP
Getting Started Manual
Lead V1...........................................................24
Lead V2...........................................................48
Lead V3...........................................................100
Lead V4...........................................................120
Lead V5...........................................................112
Lead V6...........................................................80
ECG Normal Sinus
Rates...................................................................30 to 360 (by 1) BPM
ECG High Level Output (BNC Jack)
Amplitude ............................................................0.2 V/mV of Lead I amplitude
Accuracy .............................................................±5 %. 2 Hz Square Wave
Output Impedance...............................................50 Ω output impedance
ECG on Defibrillator Input Load
Same as Lead II amplitude
ECG Performance Waves
Square wave .......................................................2.0 and 0.125 Hz
Triangular wave...................................................2.0 and 2.5 Hz
Sine waves..........................................................0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz
Pulse ...................................................................30 and 60 BPM, 60 ms pulse width
R-Wave Detection
Waveform............................................................Haver-triangle
Amplitude ............................................................0.05 to 0.45 (by 0.05) V
0.5 to 5.0 (by 0.5) V
Rate.....................................................................30, 60, 80, 120, 200, and 250 BPM
Widths .................................................................8, 10, 12 ms, and 20 to 200 (by 10) ms
Accuracy .............................................................±(1 % setting + 0.2 mV)
18
Defibrillator/Transcutaneous Pacer Analyzer
Specifications
Noise Immunity
Wave ................................................................... Sine
Line Frequency....................................................50 or 60 Hz (± 0.5 Hz)
Amplitude ............................................................0.0 to 10.0 (by 0.5) mV
Accuracy..............................................................± 5%
Transvenous Pacer Pulse Simulation
Widths
Range..............................................................0.1, 0.2, 0.5, 1.0, and 2.0 ms
Accuracy..........................................................±5 % of setting
Amplitudes...........................................................0 (off) and ±2, ±4, ±6, ±8, ±10, ±12, ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, and
±700 mV
Accuracy..............................................................± (10 % of setting + 0.2 mV)
Arrhythmia Selections
Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only)
Demand...........................................................30 to 360 (by 1) BPM
Asynchronous
Non-Capture
Non-Function
Threshold (Interactive pacing simulation only) 10 to 250 (by 10) mA
Supraventricular
Atrial Fibrillation Coarse
Atrial Fibrillation fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia
19
Impulse 6000D, 7000DP
Getting Started Manual
20
Defibrillator/Transcutaneous Pacer Analyzer
Specifications
Asystole
Conduction
1° Block
2° Block Type I
2° Block Type II
3° Block
Right Bundle Branch Block RBBB
Left Bundle Branch Block LBBB
Transvenous Paced with selectable pacer spike amplitudes and widths
Atrial 80 BPM
Async 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
AV Sequential
Non-Capture
Non-Function
Selectable pacer pulse parameters for transvenous simulation. (Atrial and Ventricular channels are independently selectable):
Atrial Pacer Pulse
Width ....................................................0.1, 0.2, 0.5, 1.0, 2.0 ms
Polarity .................................................+ or -
Amplitude .............................................0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV
Ventricular Pacer Pulse
Width ....................................................0.1, 0.2, 0.5, 1.0, 2.0 ms
Polarity .................................................+ or -
Amplitude .............................................0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV
21
Impulse 6000D, 7000DP
Getting Started Manual
Measurements
Manufacturer Specific Algorithms
GE Responder (1500 & 1700)
MDE 300 (Medical Data Electronics)
Medtronic ERS/Physio Control LIFEPAK
MRL (Medical Research Laboratory/Welch Allyn)
Philips/Agilent/HP
Schiller Medical
ZOLL Medical
(plus a general purpose default algorithm selection)
Current
Range..................................................................4.00 to 250 mA
Accuracy .............................................................±(1% of reading + 0.02 mA)
22
Defibrillator/Transcutaneous Pacer Analyzer
Specifications
Pulse Rate
Range..................................................................5.0 to 800 PPM
Accuracy..............................................................±(0.5% of reading + 0.1 PPM)
Pulse Width
Range..................................................................1.00 to 100.0 ms
Accuracy..............................................................±(0.5% of reading + 0.01 ms)
Energy
Range..................................................................1 µJ to 2.00 J
Accuracy..............................................................±(4% of reading + 10 µJ)
23
Impulse 6000D, 7000DP
Getting Started Manual
Sensitivity Test
Automatic Interactive Threshold Detection
Compatible pacer rates .......................................30 to 120 PPM
ECG R wave:
Waveforms ..........................................................Square, Triangle, Sine
Width ...................................................................1 to 19 (by 1) ms
20 to 95 (by 5) ms
100 to 300 (by 25) ms
Accuracy .............................................................± 5% of setting
Amplitude ............................................................0.05 to 0.95 (by 0.05) mV
1.0 to 5.0 (by 0.5) mV
Accuracy .............................................................± 5% of setting
24