You are on page 1of 6

EMC-CAL-01

Rev. 2
CALIBRATION PROCEDURE
Page 1 OF 6

Date 15/10/2019

CALIBRATION PROCEDURE

Rev. Prepared By Approved By Reviewed By


Date Calibration Remarks
No. inspector QCM AI

Name Samar hamed Samar hamed Mahmoud Galal


2 15/10/2019
Signature
Name Samar hamed Samar hamed Mahmoud Galal
05/02/2018
1 Signature
Mohamed abd
Name Samar Hamed Abd elwahab farag
0 10/10/2016 elfatah
Signature
EMC-CAL-01

Rev. 2
CALIBRATION PROCEDURE
Page 2 OF 6

Date 15/10/2019

CONTENTS

S/N SUBJECT 0 1 2 3
1 PURPOSE 

2 REFERENCES 

3 CALIBRATION 

4 CALIBRATION WITH POSITIVE OUTCOME 

5 CALIBRATION WITH NEGATIVE 


OUTCOME
6 CERTIFICATION & MARKING 
EMC-CAL-01

Rev. 2
CALIBRATION PROCEDURE
Page 3 OF 6

Date 15/10/2019

1. PURPOSE:-
1.1. This procedure is integrated & consistence with QC Manual (Section 9)calibration control of
measuring and test equipment

1.2. This procedure describes the requirements for the calibration of measuring tools, pressure
gages and equipment either internally or externally calibrated according to National and
International standards and/or ASME requirements as appropriate
2. REFERENCES:-
2.1. ASME CODE Sec. VIII.I 2019
2.2. ASME CODE Sec. . 2019
2.3. ASME B 31.1. 2018
2.4. NBIC 2019
3. CALIBRATION:-
Generally the master instruments are calibrated by External Laboratory and the production instruments are
calibrated against the master instruments.
3.1. Measuring Tapes.
3.1.1. The 1-meter master measuring tape shall be checked by the manufacturer or by an
External Laboratory every 12 Months and shall be used for calibrating the production
measuring tapes.
3.1.2. The production measuring tapes shall be calibrated every 12 Months.
3.1.3. The maximum allowed tolerance is:
*  0.8 mm for Master Measuring Tape.
*  1 mm for each interval 1 meter and 3 mm on 3 meters Production Measuring Tape.
*  1 mm for each interval 1 meter  5 mm on 5 meters Production Measuring Tape.
*  1 mm for each interval 1 meter  10 mm on meters over than 5 meter.

3.2. UT Thickness Measuring


3.2.1. The Ultrasonic Thickness gauge shall be calibrated by an External Laboratory every 12
Months.
3.2.2. The Ultrasonic Thickness gauge shall be calibrated, against basic blocks of different
thickness where possible standardization should be confirmed by direct dimensional
measurement of material to be tested, prior to each use.
3.2.3. The basic block material shall be standardized on material having the same acoustical
velocity and attenuation as the material to be measured.

3.3. Temperature Gauge


3.3.1. The master Temperature Gauge shall be calibrated by the manufacturer or by external
laboratory every 12 Months.
EMC-CAL-01

Rev. 2
CALIBRATION PROCEDURE
Page 4 OF 6

Date 15/10/2019

3.3.2. The Production Temperature Gauges shall be calibrated by the master temperature
gauge or by master thermocouple every 6 months.
3.3.3. The maximum allowed tolerance is:
* 10°C for Master Temperature Gauge.
* 20°C for production Temperature Gauge.

3.4. Welder Portable Heaters

3.4.1. Calibration of the Thermostat of the welder portable heater shall be done every 12
month, or whenever the equipment has been subjected to electric repairs or damages,
compare the ON & OF at adjusting the selector at 100 °C against a calibrated
thermometer or thermocouple.
3.4.2. The maximum allowed tolerance is:
* 10°C for master thermometer or thermocouple.
* 20°C for welder portable heater thermostat.
3.5. Pressure Gauges
3.5.1. Master Pressure gauge shall be calibrated by external laboratory every 12 Months.
3.5.2. The Production Pressure Gauge shall be calibrated against master Pressure Gauge every
12 month.
3.5.3. The maximum allowed tolerance is :
* 2% from its corresponding value for master Pressure Gauge.
* 5% from its corresponding value for production Pressure Gauge.
3.5.4. Gauge shall not exceed the gauging limits or described in PG-100 of ASME Sec I, par
137.4.5 of B31.1, par UG-99 of ASME VIII Div 1 and sub-section 4.3 of NB23.

3.6. Lifting Power of Yoke

3.6.1. Prior to use, the Lifting power of electromagnetic yokes shall have been checked each
day of use. The magnetizing power of all yokes shall be checked whenever the yoke has
been damaged or repaired.
3.6.2. Each alternating current electromagnetic yoke shall have a lifting power of at least 10
Ib (4.5 Kg.) at the maximum Pole spacing that will be used.
3.6.3. Each weight shall be weighed with a scale from reputable manufacturer or external
laboratory and stenciled with the applicable nominal weight prior to first use. A weight
need only be verified again if damaged in manner that could have caused potential loss
of material.

3.7. Thermocouple
3.7.1.The master thermocouple shall be calibrated by the manufacturer or by External
Laboratory every 12 months.
EMC-CAL-01

Rev. 2
CALIBRATION PROCEDURE
Page 5 OF 6

Date 15/10/2019

3.7.2.The production thermocouple shall be calibrated against the master thermocouple every
12 months.
3.7.3.The maximum allowed tolerances is :

* 5°C for master thermocouple.


* 10°C for production Thermocouple.

3.8. Calipers
3.8.1.The master caliper shall be calibrated every 12 Months by External Laboratory.
3.8.2.The production Calipers shall be calibrated every 12 months.
3.8.3.The calibration shall be performed with a minimum of three readings, one near 0, one in
the middle of caliper range and one near the end.
3.8.4.The maximum allowed tolerance is:-
*  0.8 mm for master Caliper from end of scale.
*  0.5 mm for Production Caliper from end of scale.

3.9. Volt-Amp Meters

3.9.1. A master V-A meter shall be calibrated by external laboratory every 12 Months.
* The allowable tolerance is 3.5% from its Reading value.
3.9.2. The production V-A meter shall be calibrated against master V-A meter every 12
month
* The allowable tolerance is 4% from its Reading value.

3.10. Welding M/Cs Gauges


3.10.1. Calibration of welding M/Cs gauges shall be done every 12 months, or whenever the
equipment has been subjected to electric repairs or damages, compare the reading of
the Amperages at 100A, 150A, 200A & 250A with its corresponding actual values of
Volt-Amp. Master, also the voltage must be attached to the welding M/Cs.
*The maximum allowed Amperage tolerance is 10% from its Reading Value.

3.11. Densitometers

3.11.1. Densitometers shall be calibrated at least every 3 months during use as follows:
*A national standard step tablet or a step wedge calibration film, traceable to a national
standard step tablet and having at least five steps with neutral densities from at least 1.0
through 4.0, shall be used. The step wedge calibration film shall have been verified
within the last year by comparison with a national standard step tablet unless, prior to
first use, it was maintained in the original light-tight and waterproof sealed package as
supplied by the manufacturer. Step wedge calibration films may be used without
verification for one year upon opening, provided it is within the manufacturer’s stated
shelf life.
EMC-CAL-01

Rev. 2
CALIBRATION PROCEDURE
Page 6 OF 6

Date 15/10/2019

* The densitometer manufacturer’s step-by-step instructions for the operation of the


densitometer shall be followed.
* The density steps closest to 1.0, 2.0, 3.0, and 4.0 on the national standard step tablet
or step wedge calibration film shall be read
* The densitometer is acceptable if the density readings do not vary by more than ±0.05
density units from the actual density stated on the national standard step tablet or step
wedge calibration film.
3.11.2. Periodic calibration verification checks shall be performed as described in 2.11.1 at
the beginning of each shift, after 8 hr of continuous use, or after change of apertures,
whichever comes first.
3.11.3. Calibrations required by 2.11.1 shall be documented, but the actual readings for each
step do not have to be recorded. Periodic densitometer verification checks required by
2.11.2 do not have to be documented.

4. CALIBRATION WITH POSITIVE OUTCOME


4.1. The QC Calibration Inspector Certifies that the calibration was satisfactorily carried out by
updating the list of the equipment that can be used in production and applying a label on
each calibrated Equipment.
4.2. Calibrated equipment only shall be used in production.

5. CALIBRATION WITH NEGATIVE OUTCOME


In the event that any equipment is found to be out of calibration, the QC Calibration Inspector
shall attach the label “Not Calibrated- Don’t Use” on the equipment.

6. CERTIFICATION& MARKING
6.1. The certification shall include the following data:
* Equipment description and serial number.
* Description and serial of the standard used for Calibration.
* Date of calibration and operator signature.
* Date of calibration expire.
6.2. All the calibration of units listed in this procedure could be carried out by the equipment
manufacturers or by external laboratory. In this case a certificate of calibration shall be
issued by the calibrating laboratory and shall be accepted by the QCM by signature and date
such certificates. All master equipment shall be with calibration certificate accepted by QCM
by signature and date.
6.3. Equipment identification shall be traceable by permanent marking on the equipment itself
such as stenciling or reliable painting (as applicable)

You might also like