Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;

Certification No: Check no LL6182


L Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG160-DN80
2.Calibration Condition
Flow Range: 1.6 ~ 250 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,15 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 80,74% 100,331 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 488,10 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000005 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 21, 2023
Qmax 250 0,04 0,01
0,7Qmax 175 -0,19 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 100 0,19 0,02
0,25Qmax 62,5 0,01 0,00
Verifier Tao Wang 0,15Qmax 37,5 0,06 0,02
License-issuing units
Inspector: Ji Yang 0,05max 12,5 -0,37 0,01
(special seal)
Approver: jinyong Liu Qmin 1,6 -1,08 0,43

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no LL6161
L Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG160-DN80
2.Calibration Condition
Flow Range: 1.6 ~ 250 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 23,12 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 82,07% 100,606 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 488,52 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000006 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 21, 2023
Qmax 250 0,01 0,00
0,7Qmax 175 -0,21 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 100 0,21 0,01
0,25Qmax 62,5 -0,06 0,00
Verifier Tao Wang 0,15Qmax 37,5 0,08 0,00
License-issuing units
Inspector: Ji Yang 0,05max 12,5 -0,34 0,01
(special seal)
Approver: jinyong Liu Qmin 1,6 -1,06 0,02

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no LL6163
L Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG160-DN80
2.Calibration Condition
Flow Range: 1.6 ~ 250 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 23,33 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 80,69% 100,569 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 488,55 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000007 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 21, 2023
Qmax 250 0,07 0,01
0,7Qmax 175 -0,19 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 100 0,18 0,01
0,25Qmax 62,5 0,00 0,00
Verifier Tao Wang 0,15Qmax 37,5 0,03 0,01
License-issuing units
Inspector: Ji Yang 0,05max 12,5 -0,35 0,00
(special seal)
Approver: jinyong Liu Qmin 1,6 -0,94 0,04

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no LL6167
L Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG160-DN80
2.Calibration Condition
Flow Range: 1.6 ~ 250 m /h
3

Ambient
Nominal pressure: 20 bar Test medium: Pure air 23,77 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 82,43% 100,438 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 488,74 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000008 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 21, 2023
Qmax 250 0,01 0,01
0,7Qmax 175 -0,20 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 100 0,20 0,01
0,25Qmax 62,5 -0,11 0,01
Verifier Tao Wang 0,15Qmax 37,5 -0,05 0,00
License-issuing units
Inspector: Ji Yang 0,05max 12,5 -0,40 0,00
(special seal)
Approver: jinyong Liu Qmin 1,6 -0,92 0,03

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no LL6173
L Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG160-DN80
2.Calibration Condition
Flow Range: 1.6 ~ 250 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,08 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 81,89% 100,318 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 488,60 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000009 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 21, 2023
Qmax 250 0,08 0,00
0,7Qmax 175 -0,16 0,02
This product is approved to leave the factory after passing the calibration. 0,4Qmax 100 0,16 0,01
0,25Qmax 62,5 -0,15 0,01
Verifier Tao Wang 0,15Qmax 37,5 -0,01 0,01
License-issuing units
Inspector: Ji Yang 0,05max 12,5 -0,38 0,01
(special seal)
Approver: jinyong Liu Qmin 1,6 -1,05 0,04

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. LL6175 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG160-DN80
2.Calibration Condition
Flow Range: 1.6 ~ 250 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,11 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 80,14% 100,290 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 488,37 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000010 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 21, 2023
Qmax 250 0,03 0,03
0,7Qmax 175 -0,19 0,02
This product is approved to leave the factory after passing the calibration. 0,4Qmax 100 0,19 0,02
0,25Qmax 62,5 -0,01 0,03
Verifier Tao Wang 0,15Qmax 37,5 -0,01 0,02
License-issuing units
Inspector: Ji Yang 0,05max 12,5 -0,41 0,01
(special seal)
Approver: jinyong Liu Qmin 1,6 -0,95 0,00

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810570 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,67℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 70,00% 100,379 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 907,17 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000011 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 09, 2023
Qmax 160 -0,33 0,00
0,7Qmax 112 -0,35 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,50 0,01
0,25Qmax 40 0,50 0,00
Verifier Tao Wang 0,15Qmax 24 0,18 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,72 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,09 0,04

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no #L Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,09℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 76,00% 100,678 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 907,04 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000012 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 06, 2023
Qmax 160 -0,29 0,00
0,7Qmax 112 -0,35 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,39 0,01
0,25Qmax 40 0,39 0,00
Verifier Tao Wang 0,15Qmax 24 0,30 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,59 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -0,70 0,00

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no #L Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,38℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 75,00% 100,686 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 907,52 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000013 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 06, 2023
Qmax 160 -0,34 0,00
0,7Qmax 112 -0,44 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,37 0,01
0,25Qmax 40 0,44 0,00
Verifier Tao Wang 0,15Qmax 24 0,16 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,74 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,11 0,06

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810550 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 23,40℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 70,00% 100,857 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 906,77 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000014 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 06, 2023
Qmax 160 -0,29 0,00
0,7Qmax 112 -0,37 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,39 0,01
0,25Qmax 40 0,39 0,00
Verifier Tao Wang 0,15Qmax 24 0,31 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,51 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -0,74 0,06

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810551 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 23,47℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 66,00% 100,884 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 907,12 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000015 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 06, 2023
Qmax 160 -0,33 0,00
0,7Qmax 112 -0,44 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,32 0,01
0,25Qmax 40 0,44 0,00
Verifier Tao Wang 0,15Qmax 24 0,21 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,71 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -0,92 0,00

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810552 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 23,66℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 66,00% 100,849 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 906,28 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000016 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 09, 2023
Qmax 160 -0,25 0,00
0,7Qmax 112 -0,29 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,31 0,01
0,25Qmax 40 0,31 0,00
Verifier Tao Wang 0,15Qmax 24 0,20 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,57 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -0,85 0,03

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810564 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,03℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 69,00% 100,358kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 907,20 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000017 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 08, 2023
Qmax 160 -0,38 0,01
0,7Qmax 112 -0,40 0,02
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,42 0,01
0,25Qmax 40 0,42 0,00
Verifier Tao Wang 0,15Qmax 24 0,22 0,01
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,74 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -0,97 0,02

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810565 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,18℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 65,00% 100,309kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 906,18 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000018 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 08, 2023
Qmax 160 -0,33 0,01
0,7Qmax 112 -0,38 0,02
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,43 0,01
0,25Qmax 40 0,43 0,00
Verifier Tao Wang 0,15Qmax 24 0,28 0,01
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,64 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,00 0,07

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810566 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,24℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 66,00% 100,265kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 908,53 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000019 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 08, 2023
Qmax 160 -0,45 0,02
0,7Qmax 112 0,65 0,02
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,65 0,01
0,25Qmax 40 0,53 0,00
Verifier Tao Wang 0,15Qmax 24 0,13 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,82 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,16 0,01

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810567 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,36℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 64,00% 100,291 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 905,82 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000020 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 08, 2023
Qmax 160 -0,28 0,02
0,7Qmax 112 -0,37 0,02
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,39 0,01
0,25Qmax 40 0,39 0,00
Verifier Tao Wang 0,15Qmax 24 0,21 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,56 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,04 0,02

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810568 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,41℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 65,00% 100,294kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 908,22 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000021 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 08, 2023
Qmax 160 -0,38 0,01
0,7Qmax 112 -0,38 0,02
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,27 0,01
0,25Qmax 40 0,38 0,00
Verifier Tao Wang 0,15Qmax 24 0,26 0,01
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,49 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -0,83 0,00

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810569 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,49℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 65,00% 100,328 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 908,11 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000022 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 09, 2023
Qmax 160 -0,39 0,00
0,7Qmax 112 -0,37 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,58 0,01
0,25Qmax 40 0,58 0,00
Verifier Tao Wang 0,15Qmax 24 0,36 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,70 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -0,95 0,00

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810571 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,86℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 74,00% 100,389 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 906,91 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000023 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 09, 2023
Qmax 160 -0,30 0,00
0,7Qmax 112 -0,36 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,45 0,01
0,25Qmax 40 0,45 0,00
Verifier Tao Wang 0,15Qmax 24 0,13 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 0,95 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -0,99 0,05

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810572 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG100-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 160 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 24,20℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 73,00% 100,344 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 906,66 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000024 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 09, 2023
Qmax 160 -0,38 0,00
0,7Qmax 112 -0,41 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 64 -0,60 0,01
0,25Qmax 40 0,60 0,00
Verifier Tao Wang 0,15Qmax 24 0,24 0,00
License-issuing units
Inspector: Ji Yang 0,05max 8 -0,69 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,27 0,01

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810216 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 22,32℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 58,00% 101,022 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1432,27 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000025 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
APr 29, 2023
Qmax 100 0,23 0,00
0,7Qmax 70 0,41 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,23 0,01
0,25Qmax 25 0,62 0,00
Verifier Tao Wang 0,15Qmax 15 0,84 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 0,03 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,54 0,08

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810211 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,54℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 45,77% 101,646 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1436,91 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000026 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
APr 27, 2023
Qmax 100 -0,22 0,02
0,7Qmax 70 -0,06 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 -0,25 0,01
0,25Qmax 25 0,22 0,01
Verifier Tao Wang 0,15Qmax 15 0,25 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,39 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,80 0,02

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810292 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,30℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 64,00% 101,283 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1435,43 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000027 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 08, 2023
Qmax 100 -0,27 0,00
0,7Qmax 70 -0,15 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 -0,27 0,01
0,25Qmax 25 0,10 0,01
Verifier Tao Wang 0,15Qmax 15 0,27 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,36 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,46 0,02

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810209 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,45℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 46,00% 101,452 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1436,35 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000028 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
APr 27, 2023
Qmax 100 -0,29 0,00
0,7Qmax 70 -0,06 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 -0,32 0,00
0,25Qmax 25 0,26 0,00
Verifier Tao Wang 0,15Qmax 15 0,32 0,01
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,46 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,54 0,00

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810481 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 25,29℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 78,00% 100,974 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1437,06 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000029 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 27, 2023
Qmax 100 -0,32 0,01
0,7Qmax 70 -0,17 0,09
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 -0,32 0,01
0,25Qmax 25 0,15 0,01
Verifier Tao Wang 0,15Qmax 15 0,32 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,38 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,16 0,01

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810207 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,29℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 46,00% 101,430 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1436,35 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000030 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
APr 27, 2023
Qmax 100 -0,36 0,01
0,7Qmax 70 -0,22 0,09
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 -0,35 0,01
0,25Qmax 25 0,06 0,01
Verifier Tao Wang 0,15Qmax 15 0,36 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,38 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,88 0,04

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810204 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,30℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 55,00% 101,900 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1436,66 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000031 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
APr 27, 2023
Qmax 100 -0,38 0,00
0,7Qmax 70 -0,16 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 -0,36 0,00
0,25Qmax 25 -0,03 0,01
Verifier Tao Wang 0,15Qmax 15 0,38 0,01
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,40 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,57 0,05

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810203 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,33℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 55,00% 101,851 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1429,57 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000032 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
APr 27, 2023
Qmax 100 0,17 0,00
0,7Qmax 70 0,31 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,24 0,00
0,25Qmax 25 0,66 0,00
Verifier Tao Wang 0,15Qmax 15 0,77 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 0,10 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,66 0,00

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810293 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,29℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 64,00% 101,357 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1437,16 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000033 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 08, 2023
Qmax 100 0,15 0,00
0,7Qmax 70 0,34 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,21 0,00
0,25Qmax 25 0,69 0,01
Verifier Tao Wang 0,15Qmax 15 0,83 0,01
License-issuing units
Inspector: Ji Yang 0,05max 5 0,05 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,73 0,03

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810201 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,30℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 50,00% 101,734 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1429,95 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000034 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
APr 27, 2023
Qmax 100 0,06 0,00
0,7Qmax 70 0,22 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,11 0,00
0,25Qmax 25 0,62 0,00
Verifier Tao Wang 0,15Qmax 15 0,90 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 0,17 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,94 0,04

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810200 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,16℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 46,00% 101,636 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1433,27 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000035 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
APr 27, 2023
Qmax 100 0,10 0,00
0,7Qmax 70 0,28 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,17 0,00
0,25Qmax 25 0,64 0,00
Verifier Tao Wang 0,15Qmax 15 0,81 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 0,02 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,52 0,06

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810199 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 20,89℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 47,00% 101,633 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1430,39 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000036 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
APr 27, 2023
Qmax 100 0,21 0,01
0,7Qmax 70 0,36 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,26 0,00
0,25Qmax 25 0,70 0,01
Verifier Tao Wang 0,15Qmax 15 0,83 0,01
License-issuing units
Inspector: Ji Yang 0,05max 5 0,08 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,92 0,02

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810128 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,85 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 43,75% 101,565 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1442,80 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000037 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 8, 2023
Qmax 100 -0.23 0,00
0,7Qmax 70 -0.23 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,13 0,00
0,25Qmax 25 0,23 0,00
Verifier Tao Wang 0,15Qmax 15 -0,06 0,01
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,46 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,90 0,01

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810202 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,91℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 38,32% 101,577 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1433,04 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000038 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 10, 2023
Qmax 65 0,13 0,00
0,7Qmax 45.5 0,01 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 26 -0,08 0,00
0,25Qmax 16.25 0,17 0,01
Verifier Tao Wang 0,15Qmax 9.75 -0,17 0,01
License-issuing units
Inspector: Ji Yang 0,05max 3.25 -0,45 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,64 0,09

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810150 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,71 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 50,17% 101,524 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1443,63 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000039 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 11, 2023
Qmax 100 -0,20 0,01
0,7Qmax 70 -0,17 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 -0,07 0,00
0,25Qmax 25 0,19 0,00
Verifier Tao Wang 0,15Qmax 15 -0,17 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,44 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,06 0,02

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no.LL3253 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 25,66℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 87,09% 100,057 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1444,48 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000040 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
Jun 19, 2023
Qmax 65 -0,14 0,01
0,7Qmax 45.5 -0,17 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 26 -0,07 0,00
0,25Qmax 16.25 0,24 0,00
Verifier Tao Wang 0,15Qmax 9.75 -0,24 0,00
License-issuing units
Inspector: Ji Yang 0,05max 3.25 -0,67 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,87 0,01

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810201 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,83℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 39,04% 101,733 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1442,77 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000041 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 9, 2023
Qmax 65 0,08 0,00
0,7Qmax 45.5 -0,08 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 26 -0,13 0,01
0,25Qmax 16.25 0,16 0,00
Verifier Tao Wang 0,15Qmax 9.75 -0,16 0,00
License-issuing units
Inspector: Ji Yang 0,05max 3.25 -0,46 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,81 0,13

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810240 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,96℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 39,40% 101,722 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1444,42 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000042 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 31, 2023
Qmax 65 -0,15 0,01
0,7Qmax 45.5 -0,21 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 26 -0,10 0,01
0,25Qmax 16.25 0,22 0,00
Verifier Tao Wang 0,15Qmax 9.75 -0,11 0,02
License-issuing units
Inspector: Ji Yang 0,05max 3.25 -0,58 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,89 0,01

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810133 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,31 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 47,26% 101,553 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1435,44 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000043 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 9, 2023
Qmax 100 0,30 0,00
0,7Qmax 70 0,27 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,59 0,00
0,25Qmax 25 0,70 0,00
Verifier Tao Wang 0,15Qmax 15 0,39 0,01
License-issuing units
Inspector: Ji Yang 0,05max 5 0,09 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,39 0,08

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810132 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,11 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 50,59% 101,621 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1441,30 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000044 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 9, 2023
Qmax 100 -0,02 0,00
0,7Qmax 70 -0,08 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,13 0,00
0,25Qmax 25 0,17 0,01
Verifier Tao Wang 0,15Qmax 15 -0,17 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,31 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,52 0,02

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810131 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 20,91 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 51,77% 101,642 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1442,38 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000045 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 9, 2023
Qmax 100 -0,12 0,00
0,7Qmax 70 -0,15 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,08 0,00
0,25Qmax 25 0,19 0,00
Verifier Tao Wang 0,15Qmax 15 -0,19 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,38 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,88 0,09

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810130 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 20,61 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 53,96% 101,665 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1443,59 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000046 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 9, 2023
Qmax 100 -0,21 0,01
0,7Qmax 70 -0,24 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,07 0,00
0,25Qmax 25 0,24 0,00
Verifier Tao Wang 0,15Qmax 15 -0,10 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,42 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,71 0,01

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810129 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,84 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 45,16% 101,613 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1442,20 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000047 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 8, 2023
Qmax 100 -0,15 0,00
0,7Qmax 70 -0,20 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,17 0,00
0,25Qmax 25 0,20 0,01
Verifier Tao Wang 0,15Qmax 15 -0,04 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,39 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,79 0,01

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. 1810127 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 21,98 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 44,12% 101,558 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1442,65 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000048 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 8, 2023
Qmax 100 -0,18 0,00
0,7Qmax 70 -0,20 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,08 0,00
0,25Qmax 25 0,20 0,01
Verifier Tao Wang 0,15Qmax 15 -0,13 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,42 0,01
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,64 0,01

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. LL7503 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG65-DN50-J
2.Calibration Condition
Flow Range: 0,65 ~ 100 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 22,33 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 44,99% 101,100 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 1441,22 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000049 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 8, 2023
Qmax 100 0,15 0,01
0,7Qmax 70 -0,20 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 40 0,08 0,00
0,25Qmax 25 0,20 0,01
Verifier Tao Wang 0,15Qmax 15 0,33 0,00
License-issuing units
Inspector: Ji Yang 0,05max 5 -0,42 0,00
(special seal)
Approver: jinyong Liu Qmin 0,65 -1,20 0,01

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. LL6217 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG160-DN80
2.Calibration Condition
Flow Range: 1.6 ~ 250 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 22,90 ℃
temperature:
Maximum Pressure of Atmospheric
4 bar Relative humidity: 62,16% 100,976 kPa
Medium: pressure:
Temperature Range of
-25 ~ 55
3. Calibration Result

Medium: High frequency instrument coefficient: 487,97 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000001 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 22, 2023
Qmax 250 0,03 0,01
0,7Qmax 175 -0,17 0,00
This product is approved to leave the factory after passing the calibration. 0,4Qmax 100 0,17 0,00
0,25Qmax 62,5 -0,11 0,02
Verifier Tao Wang 0,15Qmax 37,5 -0,03 0,01
License-issuing units
Inspector: Ji Yang 0,05max 12,5 -0,30 0,00
(special seal)
Approver: jinyong Liu Qmin 1,6 -0,58 0,00

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. LL6171 Device Uncertainty: 0.30%; k=2
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG160-DN80
2.Calibration Condition
Flow Range: 1.6 ~ 250 3
m /h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 23,98 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 82,14% 100,363 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 488,67 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000002 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 21, 2023
Qmax 250 0,03 0,00
0,7Qmax 175 -0,20 0,02
This product is approved to leave the factory after passing the calibration. 0,4Qmax 100 0,20 0,00
0,25Qmax 62,5 -0,14 0,01
Verifier Tao Wang 0,15Qmax 37,5 0,01 0,00
License-issuing units
Inspector: Ji Yang 0,05max 12,5 -0,39 0,01
(special seal)
Approver: jinyong Liu Qmin 1,6 -0,89 0,02

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. LL6169 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG160-DN80
2.Calibration Condition
Flow Range: 1.6 ~ 250 m /h
3

Ambient
Nominal pressure: 20 bar Test medium: Pure air 23,89 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 82,85% 100,409 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 488,65 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000003 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 21, 2023
Qmax 250 0,00 0,01
0,7Qmax 175 -0,22 0,02
This product is approved to leave the factory after passing the calibration. 0,4Qmax 100 0,21 0,01
0,25Qmax 62,5 -0,10 0,00
Verifier Tao Wang 0,15Qmax 37,5 -0,04 0,00
License-issuing units
Inspector: Ji Yang 0,05max 12,5 -0,47 0,01
(special seal)
Approver: jinyong Liu Qmin 1,6 -1,21 0,29

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet
Calibration Certification Device name: Sonic Nozzle Gas Flow Standard Device;
Certification No: Check no. LL6171 Device Uncertainty: 0.30%; k=2;
Product Name:
Gas Rotary Flowmeter
Calibration basis:EN12480-2018
Model ZYLG160-DN80
2.Calibration Condition
Flow Range: 1.6 ~ 250 m3/h
Ambient
Nominal pressure: 20 bar Test medium: Pure air 22,78 ℃
temperature:
Maximum Pressure Atmospheric
4 bar Relative humidity: 84,67% 100,644 kPa
of Medium: pressure:
Temperature Range
-25 55
3. Calibration Result

of Medium: High frequency instrument coefficient:
~ 488,83 1/m³
Accuracy Class: 1,0 Lowfrequency instrument coefficient: 10,00 1/m³
Factory No.: VIR-773092000004 Error Data

Flow point Flowrate Indication Error(%) Repeatability(%)


Calibration Date:
May 21, 2023
Qmax 250 0,05 0,01
0,7Qmax 175 -0,21 0,01
This product is approved to leave the factory after passing the calibration. 0,4Qmax 100 0,20 0,00
0,25Qmax 62,5 -0,09 0,00
Verifier Tao Wang 0,15Qmax 37,5 0,06 0,01
License-issuing units
Inspector: Ji Yang 0,05max 12,5 -0,36 0,00
(special seal)
Approver: jinyong Liu Qmin 1,6 -0,87 0,03

Calibration
Body Hydro Test: Qualified / No leakage Qualified
conclusion:
Functional test : Qualified / Approved data sheet