You are on page 1of 4

3 Pre-Installation

3.13 System Cabling Description


# NAME POINT A POINT B LENGHT NOTE
1 Wall Stand Power Cable Wall Stand Table 15 m (50 ft) Standard
Tube Stand
2 Tube Stand Power Cable Table 11.6 m (38 ft) Standard
(Conveyor)
Wall Stand Moving Grid
3 Wall Stand Generator 12 m (40 ft) Analog
Signal Cable
Wall Stand Moving Grid
4 Wall Stand Generator 12 m (40 ft) Analog
Power Cable
Table Moving Grid Signal
5 Table Generator 12 m (40 ft) Analog
Cable
Table Moving Grid Power
6 Table Generator 12 m (40 ft) Analog
Cable
Wall Stand Detector LAN
7 Wall Stand Control Room 10 m (33 ft) Digital
Cable
Table Detector LAN
8 Table Control Room 10 m (33 ft) Digital
Cable
Table CAN Rail Encoder
9 Table 3 m (10 ft) 6 Way (Optional)
Communication Cable CAN Board
Wall Stand CAN Rail Encoder Auto Tracking
10 Wall Stand
Communication Cable CAN Board (Optional)
Tube Stand CAN Tube Stand Rail Encoder
11 0.3 m (1 ft) Standard
Communication Cable (Conveyor) CAN Board
Tube Stand Auto Tracking
12 Motor Clutch Wall Stand
(Conveyor) (Optional)
Tube Stand
13 HV Cables Generator 5 m (16 ft) Standard
(Conveyor)
Tube Stand
14 Collimator Cable Generator 5.5 m (18 ft) Standard
(Conveyor)
Tube Stand
15 Tube Rotor Cable Generator 5.5 m (18 ft) Standard
(Conveyor)
Tube Stand
16 Ground Cable Generator 5.5 m (18 ft) Standard
(Conveyor)
17 Mini-Console Cable Generator Control Room 14 m (46 ft) Digital
18 Generator Comm Cable Generator Control Room 14 m (46 ft) Digital
19 Console Comm Cable Generator Control Room 14 m (46 ft) Analog
Tube Stand
20 DAP Cable Control Room DAP (Optional)
(Conveyor)
23 Table AEC Cable Table Generator AEC (Optional)
24 Wall Stand AEC Cable Wall Stand Generator AEC (Optional)
25 Generator Power Cable Generator Mains Breaker 7 m (23 ft) Standard

OSKO INC. 3 - 18
H.V. CONNECTION ANALOG SYSTEM SIGNAL CONNECTION
DIGITAL SYSTEM SIGNAL CONNECTION AUTO-TRACKING CONNECTION
AEC & DAP CONNECTION GENERAL SYSTEM CONNECTION

You might also like