TOSHIBA

Multislice HELICAL CT SCANNER

Product Data
No. MPDCT0237EAB

Su per 4 Edition
APPLICATION
The Asteion ¡s a multislice Helical CT scanner that sup-
ports whole-body scanning.
The system generales 5.3 slices per second using a
Selectable Slice-thickness Multi-row Detector (SSMD).
Since it can also perform high speed reconstruction at a
máximum 4 ¡mages per second, every examination can be
processed at ultra-high speed. In combination with the
Continuous Imaging technology, ¡t is possible to quickly
perform more accurate scanning and examinations.

FEATURES
• High-speed scanning
Four slices can be acquired simultaneously in a single
rotation. In the case of a run-off study from the shoutder
to the pelvis of an adult patient, 720 mm can be scanned
within 16,5 s with 5-mm data acquisition. Tilt helicai scanning
TCOT cone-beam reconstruction, which uses the
• Long scanning range Feldkamp method to precisely compénsate for the angu-
The long scanning range of 1.800 mm and the high lation of the bearn, enables helical scanning with the
scanning speed together facilítate whole-body trauma gantry tilted in the range from 30° forward to 30° back-
examinations without repositioning of the patient. ward Angled helical acquísition is useful for many appli-
cations, ¡ncluding obtaining the correct sean plan for
• Guided mode routine brain scans. This ability makes it easy to avoid
When this function is selected. the operating procedures X-ray exposure to the patient's orbit.
are displayed on the monitor screen of the sean consolé.
Examinations can be performed by following the on- High patient throughput
screen step-by-step instructions. This enables ¡nexperi- The high cooling rate of the 4-MHU tube, which reduces
enced technologists lo handle emergency cases. the cooling time required between scans; the high scan-
ning speed, which reduces the exposure time; and the
• Excellent image quality high reconstruction speed, which reduces the patient
The system achieves a low-contrast detectability of 2 mm sean cycle time, all increase the patient throughput sig-
at 0.3% and a high-contrast resolution of 0,35 mm in the nificantly.
x. y, and z directions. The multi-tasking capabilities of the system enable
Routine multislice helicai CT examinations can be per- patient registration and protocol setting with reconstruc-
formed using slices as thin as 0.5 mm, enabling high- tion in the background and thus increase flexibility.
precision 3D and MPR images to be generated from the In addition, the system's ability to genérate ¡mages of
fine isotropic voxel data. various slice thicknesses from a single data set permits
acquisition for routine examlnations, biopsy procedures,
• Exposure dose reduction and 3D image generation to be performed in a single
The detector system has an excellent S/N ratio, which sean.
permits examinations with a reduced exposure dose. In For example, it is possible to reconstruct 5-mm images
helicai scanning, the Fíeal-EC function, which makes the for routine examinations, 1-mm images for detailed
amount of noise uniforrn for each slice, can be selected examinations, and 0.5-mm slice images for 3D image
in an eXam Plan, making the minimization of unneces- generation from a single data set acquired by helical
sary patient exposure easy. scanning with 0.5-mm slices.
Super 4 Edition

• Improvements in image analysis functions PERFORMANCE SPECIFICATIONS


During the use of multislice systerns. there may be situa-
tions where a large number of images inust be genoral- User support
ed for detailed examinations. To provide support in such • Guidod mocle: The sean operating procedures
situations, this system includos a funclion for automática^ are displayed on the monitor of
ly generating multiplanar reconstructión (MPR) images the sean consolé.
after completion of scanning by presetting the MPR gen-
eration conditions in the eXam Plan. A higtvspeed image
Sean parameters
feeding function and an auto-filming funotion that permite • Sean regions: Whole body, including head
• Sean system: 360° continuous rotate/rotate
the window level to be adjusted are also providod. These
functions are useful for image analysis. • Sean plan programming: More than 360 different sequences
can be pre-programmed.
• Sean time
• Multislice CT fluoroscopy (option)
Conventional CT fluoroscopy shows only a single slice, CT sean: 0.48 s (Partial). 0.75, 1. 1.5 (2,
but multislice CT fluoroscopy permite Iho display oí 3 and 3) s (360° )
images obtained by real-time reconslruction oí the data SCAN £ SCAN rnodc: Min. 1.8 s (rapid sequence.
from the selectable slice-thicknoss multirow detector couch-top movement 10 mm)
(SSMD). Multislice CT fluoroscopy thus significantly
improves operability in biopsy and mterventional proce-
dures.

COMPOSITION

físteion < 4-siice system >


Standard composition (Model: TSX-021B/4)
• Gantry 1
• Patient couch 1 Axial direction Longitudinal direction
• Consolé 1 sel
• Accessories Adjustable from
- Inter-unit cables I 200 mm to 1.750 mm (1.450 mm*)
- Manuals
': For the sliort patient couch versión.
- Set of phantoms
Acquisition support
Slice thickness: 0.5, 1. 2, 3, 4, 5, 8, and 10 mm
Footswitch for the patient couch
These slice thickness are implemented by stacking the
Note: The consolé dosk is not inclndetl in thn slandaul t:nn- cfata acquired in one of the following acquisition modes.
figuration.
Acquisition: 0.5 mm x 4 rows
1 mm x 4 rows
Optional items 2 mm x 4 rows
• Cerebral blood-flow analysis system 3 mm x 4 rows
(CBP-study) (CSCP-002A) 4 mm x 4 rows
• System transformer (CETF004B) 5 mm x 4 rows
• Quantltative bone minera! study system (CBM-14A)' • Gantry tilt angle: From forward 30" to backward
• Display system for dental application (CDP-07A)' 30" (in 0.5" incremente)
• FlyThrough software (CFT-03A)* Remote control from the consolé
• Multislice CT fluoroscopy (TSXF-003E) is possible.
• LCD monitor for CT fluoroscopy (15-inch type) (CMM-003E! 1 Tube position for
• DICOM storage SCP (COT-30D) scanoscopy: O", 90", 180". and 270°
• DICOM MWM (COT-32D) 1 Gantry aperture: 720 mm in diameter
• DICOM MPPS (COT-33D)
• DICOM Q/R SCP (COT-34D) Patient couch
• DICOM Q/R SCU (COT-35D) • Vertical movement
• DICOM media (CRDM-001A) System:
• Color pnnter interface (CCP-03A) Speed of vertical
• Pediatric scanning system (CHKS-002A) movement: UP: 161o24mm/s(50Hz)
• ECG-gated scanning system (CHEG-004D) 19to28mm/s(60Hz)
• Raw data storage capacity expansión kil (EISC-ÜO/A) Down: 20 to 30 mm/s
• Image data storage capacity expansión kit (EISC-008A) Approx 540 mm
• Raw data storage software_(C_RRS-0_01B)_
• No! available in the U.S.A.
MPDCT0237EAB

Mínimum couch-top
height: Approx. 310 Max. 1,750 mm/scan (for the
Máximum couch-top long paticnt couch versión)
height: Approx. 850 mrn Max 1,450 mm/scan (for the
1 Couch-top movement short patient couch versión)
System: Motor-driven or manual Up to 10 sean plans are pro-
- Speed of movement: 130 mm/s (fasl) grammable in one eXam Plan.
10 mm/s (síow) (Múltiple and/or Multi-directional
-Stroke: 2.190 mm (for the lonq palien! Helical)
couch versión) Helical sean is possible in the
1.890 mm (for (he shoit palieni range from 30'' forward to 30''
couch versión) backward (only for 16-slice
- Scannable range: 1.800 mm (for Ihe lonq palien! acquisition).
(with headrest) couch veision) The couch-top speed can be
1,500 mm (loi Ihe shorl palien! specified in the range from
couch versión) 0.83 mm/s to 53.3 mm/s.
• Step feed pitch: 0.5 lo 600 mrn in 0.5-mm
increments Setting is possible in the range
- Reproducibility: ±0.25 mm from 2.5 to 8.0 in increments of
Repoalable lo within 0.5, excluding 4.0.
±0.25 mm after GOO-rnm move- (Pitch factor: 0.625 to 2.0)
ment
470 mm Holical pitch = Couch-top movement
• Remote control from the consolé is possible. (mm/i'ol.)/nominal scanning
Note: Fhis lunction alíows the user to check the image 011 Ihe slice thickness (mm)
consolé and adjust to the couch-top withoi.il leaving ¡he CT pitch factor = Helical pltch/number of slices
consoló. Adjustment is possible in 10-nini incrementa a! scanned in a single rotation
Ihe consolé
•-Jolf?
• Load limit
Max. allowable load: 205 kg (450 Ib)
• Footswitch: Either the vertical movement or • Real-EC: Function for continuously vary-
Auto set/Auto horne (unction can ing the X-ray tube current to
be selected. ensure the optimal X-ray dose
during helical scanning
Voice-recorded instruction and sean system
(VoiceLink) Up to 4 images/s
Voice instruclions lo the patíent can be recorded eleclroni- Real-fime Helical
cally by the operator and automaticaíly played back during reconstruclion time: 12 images/s (0.083 s/image)
sean sequences as parí of the cXarn F-'lan. SureStart:
• Number Continuous
of messages: Max. 32 rnessages sean time: Max. 100 s
• Number Región of ínteres!
of seconcls: Max. 128 s for a lolal oí 32 mes- (ROÍ): Max. 3 ROIs
sages CT number
Delay time setting: The delay time between the end measuiement interval: 0.125 s (8 measurements/s)
of fhe message and the start oí Sean start delay time: Min. 3 s
scanning can be sel up lo 10 s Mean CT number. elapsed time
in incrcmenls oí 1 s.
Can be sel with a mínimum of
Helical sean 0.1-mm increments.
• X-ray tube rotation Image slice thickness: Can be sel to a valué of up to 5
speed: 0.75. 1, 1.5s/3GO times that of the sean slice
• Continuous thickness. However, the máxi-
sean time: Max. 100 s mum image slice thickness is 10
• Sean start time delay: Min. 1 s mm. (MUSCOT reconstruction)
Setting is possible in increments Reconstruction method:
of 0.1 s. MUSCOT Reconstruction (MUItiSlice COne-beam
iechnology)
TCOT reconslruction (True COne-beam Technology)
/fsteiqjn
Super 4 Eefítion

The relationships between the slice thickness and the X-ray generation
¡maging área scanned in 7.5 s (0.75 s, 10 rotations) in • X-ray exposure: Continuous
the longitudinal direction are shown for helical pilches • X-ray tube voltage: 80, 120, and 135 kV
3.0, 4.5, and 6.0. • X-ray tube current: 10 mA to 300 mA (260 mA for
135 kV) (in incrernents of 10 mA)
Image slice Setting slice Helical pitch (Note)
• X-ray tube heat capacity:4.0 MHU
thickness thickness 3.0 4.5 6.0 • X-ray tube cooling rate: Max. 864 kHU/min
0.5 mm 0.5 mm 15.0mm 22.5 mm; 30.0 mm • Focal spot size
1 mm 1 mm SO.Omm 45.0mm : 60.0 mm - IEC 60336(1993)
nominal : 0.9 mm x 0.7 mm (small)
2 mm 2 mm 60.0 mm 9O.O mm 120.0 mm
1.4 mm x 1.4 mm (large)
5 mm 5 mm 150.0 mm 225.0 mm:300.0 mm
X-ray detection
Note: The couch-top traveling dístance per rotation is shown in • Detection system: Solid-state detectors
proportion to the slice thickness
• Main detector: 788 channels x 22 elements
• Data acquisition: 788 channels x 4 rows
Dynamic sean
• Reference detector: 1 set
• Sean time: 0.75, 1, 1.53/360°
• View rate: Max. 1,200 views/s
• Programmable time: Max. 1 hour
This refers to the máximum timo
Data processing
within which a series oí scans
• Reconstruction matrix: 512 x 512 pixels
are performed following a pre-
• Picture element (pixel) size
determmed eXam Plan.
CT image: Unit: mm
Number of
programmable scans: Sean field ss s M L LL
A to *to 'to 'to *to
Píxel size
0.35 0.47 0.63 0.76 0.94
• Sean plan
-Sean interval: Min. interval: 1 s (if an interval is
- Scanogram: Unit: mm
set).
(Note that "None" (no interval) Enlargement
Standard 2x 4x
can be seleciecl.) ratio
Setting is possible in incrernents Pixel size 1.00(L) 0.71 0.50
of 0.1 s in a sean interval more (área) 0.50 (S)
than 1 s.
*: Depending on the Vari-Area or Zoom factor
Note: When a scanning mode with patient couch movement is
used, the mínimum sean interval ís increased by the time
Reconstruction filter functions
required for movement
Functions for the abdomen with BHC
• Sean start delay time: Min. 0,5s Functions for the abdomen without BHC
Setting is possible in increments - Functions for the brain with BHC
o f O . 1 s. - Functions for the brain without BHC
• Sean rate: Max. 133 scans/100 s Functions for the inner ear and bone
(0.75 s sean, 133 rotations) - Functions for the lung
• Image reconstruction - Functions for high-resolution test mode
- Number of images: Max. 4 images/scan Functions for auditory ossicles and the spine/with high-
- Image interval: Reconstruction is possible in resolution processing
¡ncrements of 0.1s. - Functions for maintenance
• Reconstruction time: Mm. 0.5 s 1 Image reconstruction time
• Real-time reconstruction Image reconstruction
time: 12 images/s (0.083 s/image) time: Up to 4 images/s
Scanoscopy: Real-time
- Real-time
reconstruction: 12 images/s (1 slice)
24 images/s (3 siices)
Note: Real-time reconstruction of 3 siices is only for CT fluo-
roscopy (option). For helícal sean, dynamíc sean, and
SureStart. real-tíme reconstructíon is for 1 slice.
MPDCT0237EAB

• Data processor (sean consolé) • Image retrieval


- Central processing - Method: On-screen menus and keyboard
unit: 32-bit microprocessor x 2 Mode: Image, series, and patient
- Memory size: 3 Gbytes • Autoview function: Software control, function key
- Magnetic disk unit: Raw data, 36 Gbytes control
Image data, 18 Gbytes Multi-frame display: Reduction/cut-off display, ROÍ
processing
Data storage • Inset scanogram display
• Magnetic disk • Selectivo related information display
- Raw data: 4,000 rotations or more • Cine display
(for 0.75-s sean) - Image display speed: Variable
- Image data: 16,000 images or more
• Magneto-optical disk Image processing
- Total storage • Scanogram processing
capadty: 4,800 MB - Slice position display (display of planned slice.
(8 x density, double-sided) preset slice, and last scanned slice)
2,600 Mbytes Anatómica! scale (display of relative position, taking
(quadruple-density. any point selected as standard as 0)
double-sided) Siice position setting
650-Mbyte single-density MODs Enlargement
are also supported. (2x, 4x for L size)
(X-series images can aíso be • CT image processing
read.) ROÍ setting and processing
- Image data: Máximum 16,000 images - ROÍ shape: Point, rectangular, polygonal.
(for an 8 x density double-sided elliptical, irregular
disk; the exact number vanes • ROÍ processing: Mean valué, standard deviation.
depending on the compression área, number of pixels
ratio) ROÍ display: 3 ROIs can be displayed on an
Raw data: Máximum 600 rotations image.
(corresponds to 2,400 slices for Size, position, rotation
0.75-s scanning) - Measurement of distance and angle between two
(for an 8 x density double-sided points
disk: the exact number varíes Profile (oblique profile also available)
depending on the compression Histogram
ratio) CT number display
- Mark display (grid display, scale display)
Image display - Volume calculation
• Display monitor: 18-inch color LCD - Enlargement (arbitrary size)
Size of display área comparable Addition/subtraction between images
to that of a 21-¡nch CRT monitor. - Band display (non-linear windowing)
Monitor rnatrix: 1,280x 1,024 - Comment and arrow insertion
Image matrix: 1,024x 1,024(max) Top/bottom, right/left, black/white reversal of image
CT number - Image filtering
- Display range: From-1,536to+8,191 - Screen save
Note: The Cl number measurement range is from -32,768 to - High-speed axial interpolation
+32,767. MultiView (Auto MPR)
Continuously variable Quantum denoising filter
Window width/level
1 Raw data processing
(adjustable at variable speed)
Three types of window settings Multi-zoorning (processing of raw data for 4 ROIs and
Preset window:
múltiple images)
can be preset for each image,
- Stack reconstruction
Window types:
- Linear and non-linear (3 types, ¡ncluding double Win- Protect/Unprotect
-- Play/Reverse reconstruction
dows)
- User-programable non-linear (6 types) - Reconstruction queue priority change
Super 4 Edition

• System management IMAGE QUALITY


- Warm-up
Noisc
Calibration data acguisition
Standard deviation: 0.5% or less
- Scanogram/CT image switchmg
Sean parameters
- eXam Plan editing
• Tube voltage: 120KV
- Examination record
• Tube current: 260 mA
• Display of exposure
• Sean time:
dose: CTDU (oí CTDh) /DI.P/ • Reconstruction
Geometnc eff f unction: FC70
Slice thickness: 10 rnm
3D color ¡mage processing Sean field S
High quality 3D images can be obtained vory quickty with Phantom: (|>24 cm water
easy operation.
8.0 !p/cm at MTF 50%
• 3D surface rendering 14,5 Ip/cm at MTF 2%
- Clipping, texture or non-texture 18 Ip/cm at cut off
• 3D volume rendering (reference)
- Máximum intensity projection (Max-IP) Sean parameters
-• Mínimum intensity projection (Min-IP) • Tube voltage:
X-ray volurne rendering • íube current:
Intensity volume rendering • Sean timc v
Shaded volume rendering (an arbilrary opacity curve Slice thickness:
can be set) Sean (¡oíd:
• Display/processing function Reconstruction
Zooming, panning, measurement (dislance, angle), functíon: FC90
annotation. cutting. drilling Phantom: 16 cm Catphan
• Cine display High-contrast resolution
• Segmentaron X-Y plañe
Partial extraction can be performed High-resolutton
•MPR rnode (FC90):
3 orthogonal planes/obligue image Standard mode
(FC30): (|)0.55 ±0.05mm
Curved MPR
Sean parameters
• Tube voltage: 120kV
Image transfer
• Tube current: 200 mA
• 100BASE-TX, 10BASE-T
• Sean time 0.75 s
• Toshiba protocol
Slice Ihickness: 2 rnm
• DICOM storage SCU
• Sean field: S (30 mm: zooming used)
• TIFF conversión
Phantom: Toshiba high-contrast measuring
phantom (acrylic/air ratio)
Filming Z-directipn
• Ethernet: Toshiba protocol Standard mode
DICOM PRINT 0.35 ±0.05 mm
(FC10):
• Sheet editing (unction using virtual film Sean parameters
• T-rnode: Related informador) ítems sueti Tube voltage: 120kV
as the palien! ñame are dis- • Tube current: 50 mA
played in (he footcr área USIIKJ a • Sean time: 0.75 s
larger font. • Slice thickness: 0.5 mm x 4
Note: To use T-mcxIe. the láser imagei musí suppori • Helical piteh: 2.5
2048 pixels x 2404 pixels for a 1 x 1 fmme. Phantom: Toshiba ladder phantom

Auto filming in eXam Plans


MPDCT0237EAB

• Low contrast X-ray Generator


detectability: 2 mm 4 mm This unit supplies stable high voltage to the X-ray tube unit.
at 0,3% al 0.3 % The high-frequency inverter method ¡s employed, resulting
Sean pararneters in a lighl and compact design. This unit is incorporated in
I he qanli y.
Tube voltage: 1 20 kV 120kV • Max. power: 42 kW
• Tube current: 250 mA 200 mA
• Sean time: 1 s 0.75 s
X-ray Tube
• Reconstruction
Hiis is a large-capacity, high-cooling-rate X-ray tube that is
function: FC41 FC4 1
ablo lo vvithstand continuous opcration as in helical scan-
• Slice thickness: 10 rniTi 10 inm .
ning
• Sean field: S S • l-leat Capacity: 4 MHU
• Surface dose: 31.1 mGy 18. 9 mGy
• Cooling Rale: Max. 864 kHU/min
Phantom: 20 cm 20 cm
Catphan Catphan Patient nniirh
The palien! couch is posilioned in front of the gantry and
Head mode supports the patient. The entire unit moves vertically and
Body modc the top moves longitudinally. In an ernergency. the couch-
top can be pulled out manually with very little effort. The
Sean pararneters
couch-top can also be lowered to a minimum height of 310
• Tube voltage: mm (at the center of the couch top) from the floor. facilitat-
• Tube current: ing transfer of the patient from a low bed or stretcher.
• Sean time: Tl'io footswitches provided on both sides of the patient
• Slice thickness: couch can be used to control couch vertical movement
without using hands.
SYSTEM COMPONENTS AND THEIR
FUNCTIONS Consolé
The consolé is provided with a hybrid keyboard, monitor.
Gantry and mouse
The scanner is composed of the gantry and Ihe palien!
• functions of the consolé
couch. The scanner uses a fan-shaped continuous X-ray
Solociion of sean parameters
beam to sean Ihe región lo be examined Transniílted X-
Scanoscope control
rays are detecled and eonverled mío elor.irioa! signáis by
Sean control
the SSMD.
Remóte conlrol of couch-top movement
The gantry includes the mam body and its support media
Remoto control of gantry tilt
nism. The X-ray tube and the SSMD are rnounted facing
Window level and window width adjustment
each other on either side of the gantry aperture. and Ihe X-
Other mouse-operated image processing functions
ray tube and detectors rotate contmuously around Ihe
aperture of the gantry. A slipring is employed to Iransmit
Magneto-optical disk unit
power between the gantry and the rotating X-ray high-voll-
The magneto-optical disk unit permits repeated recording
age generator assembly.
and rctrieval of dala, like a floppy disk dnve. It is a com-
The gantry can be tilted forward and backward in order to
pact. externally mounted, large-capacity digital storage
perform tilt scanning Three-dimensional alignment lights
device that can be used to store a large amount of data.
are provided for setting slice positions Gantry and patient
Automatic image archiving to the magneto-óptica! disk unit
couch operating controls are provided on both sides of the
can be set in eXam Plans.
front of the gantry housing. The patient gutde display indí-
cales the sean status to the operator and the palien!. The
X-ray high-voltage generator is built into the gantry. and
the systern employs a high-frequency inverler for general
ing and stabili/ing the high voltage supplied lo the X-iay
tube. The generator includes electronic eircuits for control-
ling the speed of Ihe rotating anode in Ihe X-ray tubo. Use
of a high-frequency inverter system results in high powei
output combined wilh excellent stabüity. In addition. (lio
system ¡s compact and light weighl.
Super 4 Edition

OPERATING FEATURES • Real-time helica! reconstruction mode makes it possible


to observe the images being scanned in real time at a
Patient handling and positioning máximum at 12 trames per second. This mode shows
• The couch-top can be lowered to 310 mm (at the center any shift in the slice position in real time and helps the
of the couch-top) from the floor, making ¡t easier to trans- operator to check the sean field on the image, the con-
fer the patient to and from a bed or stretcher. trast study timing, the presence of patient body motion.
• Alignment lights are provided ¡n the gantry aperture for etc. The patient can therefore be released ¡mmediately
fast and accurate patient positioning. after scanning.
• High-precision couch-top positioning in mcrements of 0.5 • The SureStart function allows the operator to start helical
mm is possible from the integrated consolé or by manual scanning at the timing of máximum enhancement ¡n con-
operation from the control panel and clear digital reaci- trast studies. SureStart monitors the sean from the start
outs are provided on the gantry. of a contrast study at a certain slice position while mea-
• The couch-top can be pulled out manually in an emer- suring the changes in CT number on the image being
gency. displayed in real time. When the contrast reaches the
predefined threshold, helical sean automatically starts.
Scanning This technique ensures optimal contrast enhancement,
• When Guided mode is selected, the operating proce- independent of individual differences in blood-flow
dures are displayed on the monitor screen of the sean speed. and at the same time minimizes the dose of con-
consolé. Examinations can be performed by following trast médium
the ¡nstructions.
• Toshiba's Scanoscope function provides a projection Data processing
image of the patient for high-precision advance planning • A variety of reconstruction algorithms are available and
of the slice positions. can be selected according to the anatómica! región to
• The longitudinal length of the scanning field for the be exarnined and the clinical objective of the study.
scanogram can be adjusted up to 1.750 mm (1.450 mm These include algorithms for the abdomen, head, bone.
for the short patient couch versión). Because the image lung, small structures, soft tissues, etc.
is reconstructed in real time, the sean can be aborted at
any time. This minimizes the patient exposure dose. Image display and processing
• The auto Índex function allows automatic incremental • Reconstructed images are automatically displayed
couch-top movement based on the slice positions deter- according to the window settings preset ¡n the eXam
mined through the scanogram. Plan.
• The eXam Plan function allows simple selection of pre- • The window save function allows the user to store an
programmed scanning parameters for routine examina- image with window settings different from the ones set in
tions, maximizing patient throughput. the eXam Plan.
• The Vari-area function allows the user to pre-select a • Filter parameters can be customized through simple on-
región of interest for zooming using raw data, permitting screen menú selections. These parameters include the
immediate post-scan analysis. Zooming using raw data number of filtering passes, matrix size. and filter coeffi-
yields higher resolution than enlarging an irnage that has cients.
already been reconstructed. • Images can be rotated and reversed either right/left,
• Multislice Helical Sean acguires raw data by rotating the top/bottom, or black/white.
X-ray tube continuously while moving the patient continu- • The Multi-frame feature allows up to 16 images to be
ously through the scanner. The volume data acquired retrieved and displayed simultaneously on the screen.
can be used to reconstruct slices at any desíred axial • The three-dimensional image display function allows
positions. This sean mode is best used for rapid patient color three-dimensional and real-time MPR images to be
scanning during a single breath-hold and for high-defini- generated from the volumetric sean data acquired by
tion three-dimensional and MPR ¡maging. helical scanning. This results in higher definition and
• The Real-EC function, which can be set in an eXam Plan. image quality than irnages reconstructed from conven-
automatically adjusts the optimal tube curren! for every tional single-slice scanning. This ¡s because helical
región, thus minimizing patient exposure. scanning provides superior data continuity along the
patient axis compared with conventional scanning.
MPDCT0237EAB

Image storage and archiving COMPLIANCE


• The system is provided with a 54 Gbyte magnetic hard
Council Directivo 93/42/EEC
disk as standard equipment, permitting the on-line stor-
Concerning Medical Devices (Medical Device Directiva)
age of approximately 16,000 images and 4.000 rotations
of raw data.
IEC: IEC 60601-2-44 (2001)
• A 4.8 Gbyte magneto-optical disk is provided as stan-
IEC 60601-2-44 Amd.1 (2002)
dard equipment. The image storage capacity of the
magneto-optical disk is approximately 16.000 images per
IEC60601- 1 (1988)
disk.
IEC 60601- 1 Amd.1 (1991)
Image filming IEC60601- 1 Amd.2(1995)
IEC 60601 •1-1 (2000)
• Filming of images can be performed manually or auto-
matically from the consolé. IEC 60601 •1-2(2001)
IEC 60601 1-3(1994)
• Automatic filming sends an entire study to the láser cam-
IEC 60601- 1-4(1996)
era. Filming is performed in background mode so that
other scanner and image processing functions can he IEC 6060t 1-4 Amd.1 (1999)
IEC 60601 2-32(1994)
performed without ¡nterruption or delay.
• When T-mode is used, related Information ilems dis-
played together with an image (surrounding the image. in DIMENSIONS AND MASS
a small font) are displayed in the footer área using a larg-
Dimensions L x W x H Mass
er font, permitting not only easier reading but also sirn- Unit
mm (in) kg(lb)
pler film management.
890 x 1,970x 1,760 1,300
Note: To use T-mode. the láser imager must support Gantry
2048 pixels x 2404 pixels for a 1 x 1 trame. (35.0 x 77.6 x 69.3) (2.866)
Long patient 2.690 x 630 x 450 450
Patient couch versión ( 105.9 x 24.8 x 17.7) (992)
Patient throughput
couch Snort pat¡ent
2.390 x 630 x 450 420
Patient throughput and cost efíectiveness were major
objectives in the design and production of the Asteion CT couch versión (94.1x24.8x 17.7) (926)
scanner. Consolé
• The system incorporales a 4-MHLJ X-ray tube with a last
820 x 450 x 700 120
cooling rate of 864 kHU/min in actual use. CPU cabinet
(32.3 x 17,7x27.6) (265)
• High-speed scans can be performed in as littlo as 0.48
second per sean Routine scans can be perforrned as 820 x 450 x 700 140
REC cabinet
quickly as 0.75 second per sean. (32.3 x 17.7x27.6) (309)
• In multisiice helical scanning, scanning can be per-
formed at a máximum 12 slices per second.
• Real-time reconstruction is possible in scanoscopy.
• CT images can be reconstructed ¡n 0.25 second for SITING REQUIREMENTS
0.75-second routine scans. Power requirements
• Ease of operation is ensureci by incorporating use of a • Pliase: Three-phase
hybrid keyboard, mouse-driven menus. and large color • Voltage: 200 V*
LCD screens. • Frequency: 50 Hz or 60 Hz
• The couch-top can be lowered very near the floor by • Une capacity: 75kVA
using the foot switch, simplifying patient transfer. • Voltage fluctuation
due to load variation: Less than 5%
• Power voltage
fluctuation: Lessthan 10%"
Please consult Toshiba in the case of other voltages or exces-
sive power fluctuation.
Represents the total voltage fluctuation due to load and power
Super 4 Edition

Grounding Mínimum área for installation


Groundinq musí be provided in accordance witli local • For the long patient couch versión: 22 m
regulations for medically used electrical equipment. • For the short patienl couch versión: 20 m'

Power distríbution board


In case of
3-phase, 200 V
In case of
3-phase. 400 V rfe
Ground resistance:
As per applicable
legal requirements.
NFB NFB NFB
150 A 50 A

Gantry Gantry control/ \é System


Ground bar O
transformer
(option)

Ambient conditions
Temperature(*1) Humidity Hea! generatíon o
Sean room
Gantry
I
O Gantry 0 Consolé
Q Patienl couch

': For the short patlent couch versión

Room layout example


Operator's room
Consolé 16'Cto28° C 40% to 80% Approx.
No 10.800kJ/hC2) Installation requirements
condensation
* 1: When the systern is not operating: 0"C to 40 C Sean room
"2' When scannlnq ís not performed. • Before installing the qantry, check the máximum perrnis-
*3: When scanning is performed continuously at Ihe máxi- sible floor load
mum rated output of the system. • The scanner emits radiation. X-ray shielding must be
provided around the sean room and the entrance ín
aocordanoe witli all local requirements and regulations,
• The ceiling shoulcl be at least 2,500 mm high to permit
ll'io use of a contras! médium injector.
• Wiring pits and ducts are required for routing cables that
conncct the various units.
MPDCT0237EAB

Operator's room Checks before bringing-in the unit


• An observation wíndow is required for rnoniloring tlie • Chock in advance the width of the corridor. the dimen-
sean room. X-ray shielding of the wíndow glass must be sions of the enlrance. and the dimensions and máximum
provided ¡n accordance with al! loeal requirements and allowable load of the stairs and elevators to ensure that it
regulations. and the bottom of the wíndow trame should is possible to bring-¡n the unit safely and without difficulty.
be 90 crn from the floor. • Mínimum external dimensions of the entrance used for
• Wiring pits and ducts are required lor routing cables that bringing-in the unit are as follows:
connect the vanous units. Width: 980 mm (38.6 in)
• The operator's room should have entráñeos tor access to Height: 1.860mm (73.2 in)
the corridor and the sean room. • The corners of corndors should be as illustrated below,
• Irlevotor mínimum load: 1.600 kg (3,527 Ib)
Cable connections between units
• Without system transformer
W >1.400 mm (55.2 in)
w wi

17

17

12

With system transformer

|~ Speaker -| -¡
[sean roomj
Ganlry k
17

12 11
\\8

/ 18

I lint, ni

11
Super 4 Edition

OUTLINE DRAWINGS

Gantry and Patient Couch

12
MPDCT0237EAB

OUTLINE DRAWINGS

°
• o
^o
o gy
o •
'i
u

_890 (35.0)
_ 360
i (14) *f" <21>
4» - 3o-
(TILTIANGLEV

' '\ 320^i -

f t

(126) (21.0)

Gantry

13
/¡stevqfí
Super 4 Edition

OUTLINE DRAWINGS

o LO o

2,190 2,187
(86.2) (86.1)

t
430* I 1,862 2.690
(17) (73.3) (105.9)

Patient Couch (for the long patient couch versión)

14
MPDCT0237EAB

OUTLINE DRAWINGS

El"

1,890 1,887
(74.4) (74.3)

430* » 1.562 2.390


(17) (61.5) (94.1)

LU
£¿ O —-
§i-n »wE S2
S
,-A'I o ^S
^O t. • ^
-±=5

Patient Couch (for the short patient couch versión)

15
MPDCT0237EAB
Super 4 Edition

OUTLINE DRAWINGS

uu

I
t /
1,600

C (63.0)

100(3.9)

Consolé

TOSHIBA MEDICAL SYSTEMS CORPOR ATIOIM

hup://www.toshibami'«.licals; su-nis.com

£t~**«* TUV Rheinlsnd

f MadeforLife -hih.i MfdHM! Svst^riK Orpnulron nii-Ms Toslnki ic.i! S\sloms Corporation Nn?ü ope'
^n.ilir>iintlvttH;oorti:^()'ít l ninlrirdsf()i Ulsalsl1, Itit-tnv
m,H|,'iiH''ntS..'íl.-'tiF.!"->ÍK>1 ISO Utíifi