1. General
2. Precautions
4. Operating instructions
5. Maintenance
1. General
Thank you for purchasing our DeMattia Flex Cracking Tester. This operation
manual is for the parameter setting and operation instructions of the machine.
I. purpose:
leather, sole, film, rubber coated fabric and other materials at room
temperature.
Product specifications
This manual contains crucial safety guidelines and usage instructions that
Dangerous:
is not followed.
warning:
note :
This shows items that are likely to affect test results and quality.
【note】
this product.
This shows the auxiliary instructions for the operation and use of
Danger voltage
mark
This mark indicates a high voltage hazard.
Power: power key to turn on and off the power of the instrument
Set: set key, Modify settings, modify test times, and finally confirm and return
Turn on the power and press the set key. The displayed window is shown in
Once you've configured the necessary test durations, use the "Set" key to
exit the modification mode. After configuring the settings, press the "Run"
key to initiate the test, and the instrument will automatically cease operation
resume the test, press the "Run" key once more. Should you wish to erase
the test record in preparation for the next test, use the "CLR" key.
4. Operating Instructions
1. Power supply conditions: Please configure the correct power supply
dangerous
■ The permissible range for input voltage must fall within ± 10%.
!
Additionally, it is essential to ensure the machine is adequately grounded to avert
1. Preparation of the test piece: take the specified test piece according to the
provisions of the standard, and fix the test piece in the fixture;
2. Turn on the power, press the "run" key to start the machine, turn the "speed
regulation knob" to adjust to the required test speed, and then press the
"stop" key,;
3. Set the counter times according to the experimental needs;
4. Press the "run" key to test; Press the "stop" key to end the test;
5. Press the "power" key to turn off the machine and remove the test pieces;
5. Maintenance procedure
1.The instrument must be placed on a stable table before use,
4. If any physical damage is noticed by the operator, stop the testing and
contact NextGen