You are on page 1of 41

Renatron ®

Hydraulic System
13 12

Tank  Air Vent Line


PT
11

10

5 4 Empty Tank to scribe line


Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter

 Water
3

Open Valves: 3, 6, 7, 11, 12


Jet Pump Drain
Alarm if exceeds 120 sec.
13 12

Tank  Air Vent Line

PT
11

10

5 4
Arterial Flood Tank to 507 Grams
(Black line)
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter

 Water
3

Jet Pump Drain



Open Valves:1, 11, 13 PT = Pressure Transducer

Alarm if exceeds 25 sec.


13 12

Tank  Air Vent Line


PT
11

10

5 4 Drain Tank to 247 Grams or less,


Arterial Proceed to step 5
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
Water
Jet Pump Drain
 Open Valves: 3, 4, 5, 11, 12 PT = Pressure Transducer

Alarm if exceeds 39sec.


13 12

Tank  Air Vent Line


PT
11

10

5 4 Empty Tank to scribe line


Arterial
9
15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In 8

7 Venous 6

1
6 liter
2
1 liter

Water
3

Open Valves:3, 8, 9, 11, 12 PT = Pressure Transducer


Jet Pump Drain
 Alarm if exceeds 39 sec.
13 12

Tank  Air Vent Line


PT
11

10

5 4 Set Zero and “Clean” light on


Arterial after 2 sec. Proceed to step 9
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain PT = Pressure Transducer



Open Valves: 12
13 12

Tank  Air Vent Line

PT
11

10

5 4 Draw Renalin Solution into Tank


Arterial

9 15
Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8 renalin jug

7 Venous 6

1
6 liter
2
1 liter
3
Water

Open Valves: 3, 11, 13, 15



Jet Pump Drain
PT = Pressure Transducer
Tank mass to 12g (OO, CH) or 22g (HF) proceed to step 10

Alarm if time exceeds 30 sec.

Step 10 : Energize valves 11 and 12, after 4 sec. Go to step11

Step 11 : No action, go to step 12


13 12

Tank  Air Vent Line

PT
11

10

5 4
Dilute to 507 Grams
Arterial
For HF the water will fill up to 948 g at
9 15 Step 13
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter

 Water
3

Jet Pump Drain PT = Pressure Transducer


 Open Valves:1, 11, 13
Alarm if exceeds 20sec.
13 12

Tank  Air Vent Line

PT
11

10

5 4 Flush Blood Circuit From


Arterial Arterial to Venous.
9 after 15sec go to step15
15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water
Jet Pump Drain
 PT = Pressure Transducer
13 12

Tank
507g
V  Air Vent Line
247g
PT
11

10

5 4 Draw Proportioned Renalin® Solution


Arterial to Dialysate Compartment.
9 15
When < 247g go to step 16
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


PT = Pressure Transducer
Open Valves: 3, 5, 7, 8, 9, 11, 12
Alarm if exceeds 39 sec.
13 12

Tank  Air Vent Line


PT
11

10

5 4 Reverse Ultrafilter Proportioned


Arterial Renalin® Solution through Membrane.
After 100sec go to step17
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter

 Water
3

Jet Pump Drain Open Valves: 3, 5,7, 8, 11, 12 PT = Pressure Transducer


13 12

Tank  Air Vent Line


PT
11

10

5 4 Draw Proportioned Renalin® Solution


Arterial to Blood Compartment.
>0 ml go to step18
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water
Jet Pump Drain PT = Pressure Transducer
Open Valves: 3, 5, 6, 9, 11, 12

Alarm if exceeds 90sec.
13 12

Tank  Air Vent Line


PT
11

10

5 4 Reverse Ultrafilter Proportioned


Arterial Renalin® Solution Through Membrane.
9
after 65sec go to step 19
15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter

Water
3

Jet Pump Drain


Open Valves: 3, 5,8,11, 12 PT = Pressure Transducer
13 12

Tank  Air Vent Line

PT
11

10

5 4 Refill Tank black line Reference Level


Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter

 Water
3

Jet Pump Drain


 Open Valves:1, 11, 13 PT = Pressure Transducer
13 12

Tank  Air Vent Line

PT
11

10

5 4
Fill Dialysate Compartment.
Arterial after 15sec go to step21
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter

 Water
3

Jet Pump Drain PT = Pressure Transducer


 Open Valves: 2, 4, 8, 9
13 12

Tank  Air Vent Line

PT
11

10

5 4 Fill Blood Circuit From


Arterial Arterial to Venous.
after 30sec go to step23
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3

 Water

Jet Pump Drain PT = Pressure Transducer


 Open Valves:1, 4, 7
13 12

Tank  Air Vent Line


PT
11

10

5 4 Flush Venous Manifold Gallery


Arterial and
Establish Tank Zero below scribe line
9 15 .
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
>0g go to step 24
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water
Jet Pump Drain Open Valves: 3, 6, 7, 11, 12 PT = Pressure Transducer
Alarm if exceeds 60 sec.
13 12

Tank  Air Vent Line

PT
11

10

5 4 Fill Blood Circuit


Arterial from Venous to Arterial
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter


3
 Water

Jet Pump Drain PT = Pressure Transducer


 Open Valves: 1, 5, 6
Clean Light Off
Test Light On
13 12

Tank  Air Vent Line

PT
11

10

4 Vent Manifold
5
set zero mass Reference
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


Open Valves: 4, 8, 9 PT = Pressure Transducer
13 12

Tank Air Vent Line

PT
11

10

5 4 Drain Blood Compartment,


Arterial
After 15 sec. Go to step 30
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


Open Valves: 3, 6, 9, 10, 11, 13 PT = Pressure Transducer
13 12

Tank  Air Vent Line

PT
11

10

5 4 Draw Vacuum,
Arterial >-289mmHg go to step 31
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


Valves 3,11,13 PT = Pressure Transducer

When time exceeds 30 seconds or


manifold pressure exceeds -289
mmHg proceed to the next step.
13 12

Tank Air Vent Line

PT
11

10
Sweep Blood
5 4 Compartment
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


PT = Pressure Transducer
Open Valves: 3, 6, 9, 10, 11, 13
Test Valve 10 for Magnetism
13 12

Tank  Air Vent Line

PT
11

10
Draw Vacuum
5 4
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain Valves: 3, 11, 13 PT = Pressure Transducer


When time exceeds 30 seconds or manifold
pressure exceeds -289 mmHg proceed to next step.
13 12

Tank Air Vent Line

PT
11

10
Sweep Blood Compartment
5 4
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain

Open Valves: 3, 6, 9, 10, 11, 13 PT = Pressure Transducer


13 12

Tank  Air Vent Line

PT
11

10

5 4 Vent and Stabilize


Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


PT = Pressure Transducer

Valve 12
13 12

Tank  Air Vent Line

PT
11

10

5 4 Measure Mass
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


Valves: 12

If the result is greater the or equal to the dialyzer reference Volume ,go to step 36.
Valve 12
Alarm if less than reference Volume ,Alarm
13 12

Tank  Air Vent Line


PT
11

10

5 4 Draw Vacuum
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter

 Water
3

Jet Pump Drain


Open Valves: 3, 8 ,9, 10 PT = Pressure Transducer
When pressure reaches –200mmHg
Go to next step 37
Alarm if exceeds 60 sec.
13 12

Tank  Air Vent Line


PT
11

10

5 4
Hold Vacuum
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain



Open Valves: 8, 10 PT = Pressure Transducer
After 10 sec go to step 38
13 12

Tank Air Vent Line


PT
11

10

5 4 Draw Vacuum
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter


3
 Water

Jet Pump Drain


Open Valves: 3, 8, 9, 10 PT = Pressure Transducer
When pressure reaches
–250mmHg go to next step 39
Alarm if exceeds 60sec
13 12

Tank Air Vent Line


PT
11

10
Hold Vacuum for 65 sec.
5 4
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


 PT = Pressure Transducer
Open Valves: 8, 10
Alarm if pressure changes by more than 50 mmHg
13 12

Tank  Air Vent Line

PT
11

10

5 4 Empty Tank scribe line


Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3

 Water

Jet Pump Drain


Open Valves:3, 6, 7, 8, 11, 12 PT = Pressure Transducer
13 12

Tank  Air Vent Line


PT
11

10

5
Drain Main Supply Header
4
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter

 Water
3

Jet Pump Drain


PT = Pressure Transducer
Open Valves:3, 4, 6, 7, 10

De-energize “Test” light

Energize “Disinfect” light


13 12

Tank  Air Vent Line


PT
11

10

5 4 Set Zero
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain



PT = Pressure Transducer
Open Valves:12
13 12

Tank  Air Vent Line

PT
11

10

5 Draw Renalin® Solution (OO)  27g


4
into Tank (CH)  27g
Arterial
( HF)  37.5g
9
 Renalin Line
15
Dialysate Out
Dialyzer
Dialysate In
8 renalin jug

7 Venous 6

1
6 liter
2
1 liter

 Water
3

Jet Pump Drain


PT = Pressure Transducer
Open Valves: 3, 11, 13, 15
13 12

Tank  Air Vent Line


PT
11

10

5 Vent and test for proper


4
Arterial
amount of Solution

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


 PT = Pressure Transducer
Open Valves:12
13 12

Tank  Air Vent Line

PT
11

10

5 4 Dilute to black line (507 Grams)


Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter


3
 Water

Jet Pump Drain


PT = Pressure Transducer
Open Valves:1, 11, 13
13 12

Tank  Air Vent Line

PT
11

10

5 4
Draw Proportioned Renalin® Solution
Arterial to Dialysate Compartment
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water
Jet Pump Drain
PT = Pressure Transducer
Open Valves: 3, 5, 8, 9, 11, 12
13 12

Tank  Air Vent Line

PT
11

10

5 4 Draw Proportioned Renalin® Solution


Arterial to Blood Compartment
9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water
Jet Pump Drain
Open Valves: 3, 5, 6, 9, 11, 12 PT = Pressure Transducer
13 12

Tank  Air Vent Line

PT
11

10

5 4
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


PT = Pressure Transducer
Valves: 6, 8, 11, 12
13 12

Tank  Air Vent Line

PT
11

10

5 4
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain


PT = Pressure Transducer
Valves: 4, 5, 9, 10, 12
13 12

Tank  Air Vent Line

PT
11

10

5 Hold
4
Arterial

9 15
 Renalin Line
Dialysate Out
Dialyzer
Dialysate In
8

7 Venous 6

1
6 liter
2
1 liter
3
 Water

Jet Pump Drain

PT = Pressure Transducer
Valve: 5
Auto vent until reset is pressed

You might also like