A Simplified Method

to Determine Total
Sulphite Content in
Food and Beverages via
Ion Chromatography
Alyson Lanciki1 and Miguel Espinosa2, 1Metrohm AG, Herisau, Switzerland, 2Metrohm
Hispania, Madrid, Spain

Sulphite is a preservative added to a vast range of foods and beverages to


prevent browning or oxidation. Some individuals are sensitive to sulphite
additives and may experience a range of allergic reactions. Therefore, both
the U.S. Food and Drug Administration (FDA) and European Union (EU) laws
require that the presence of sulphites be declared on food labels when the
concentration exceeds 10 mg/L. Several analytical methods exist to measure
sulphite in food and beverages, however, they suffer from repeatability
issues, and can be quite cumbersome to perform. A patent has been filed
for an innovative, fast, and accurate ion chromatographic (IC) method based
on direct current (DC) mode electrochemical detection. This article explains
more.
XXXXXX/stock.adobe.com
Aliaksei/stock.adobe.com

Sulphite (SO32-) is a widely used preservative including meat, fish, legumes, soft drinks,
added to prevent browning or oxidation to wine, and beer (1,2). Sulphite is also a
a vast range of food and beverage products naturally occurring substance produced

2 Incognito 9 News 12 Lanciki and Espinosa 17 Pittcon Preview

12
20 Kelting 25 Eichman et al. 29 Training & Events 31 Staff
The Column www.chromatographyonline.com Lanciki and Espinosa

Table 1: List of solutions needed for total sulphite analysis in foodstuffs via IC Automated discrete analysis methods
Sodium hydroxide 300.0 mmol/L have been reported for sulphite analysis
Eluent Sodium acetate 300.0 mmol/L (12). These methods mainly focus the
CO2 free ultrapure water sulphite analysis on specific sample matrices
Rinsing solution Ultrapure water such as wine, beer, dried fruits, and so
on. Nevertheless, laboratories where
Formaldehyde 1.00 mol/L
Stock stabilization solution sulphite analysis is required for a wide
Sodium hydroxide 0.20 mol/L
variety of food and beverage products
Working stabilization solution Formaldehyde 1.00 mmol/L need a single, robust analytical method
(dilution 1:1000 of stock solution) Sodium hydroxide 0.20 mmol/L
to provide a solution. Methods based on
IC with conductivity detection offer a lack
during fermentation processes (2). Generally Methods of Sulphite Measurement in of selectivity combined with an extended
speaking, sulphite is an oxygen scavenger, Foodstuffs analysis time due to the need to separate
resulting in formation of either sulphate (SO42-) Traditionally, the optimized Monier‑Williams sulphite from sulphate, as well as
or sulphur dioxide (SO2) after the oxidation (OMW) Official Method 990.28 was from many other coeluting organic
process, depending on the conditions (3). used for quantification of sulphite in interferents.
Many individuals are sensitive to sulphite most matrices (7), but this method is A newer method developed by AOAC
additives and some have experienced mild to time-consuming, labour‑intensive, and (Method 990.31) shifted away from
severe allergic reactions, however, the exact has a method detection limit (MDL) at gravimetric titration to focus on the use of
mechanism behind this remains unknown, the regulatory labelling threshold. More ion-exclusion chromatography followed by
despite several hypotheses tested. Therefore, efficient and accurate methods have been electrochemical (amperometric) detection
both U.S. Food and Drug Administration tested in the meantime using stabilizers of samples after extraction in an alkaline
(FDA) and European Union (EU) laws require to overcome sulphite chemical instability medium (10). However, AOAC Method
that the presence of sulphites be declared on in aqueous solution followed by different 990.31 is not applicable to dark-coloured
food labels when the sulphite concentration detection techniques. Such techniques foods or ingredients because the sulphite
is higher than 10 mg/L (4). For certain people, are based on ion chromatography (IC) is so strongly bound. The added sulphite in
sulphite exposure in food has been reported to with conductivity detection (8,9), as well food products can bind irreversibly to the
induce asthmatic episodes (5), though further as chromatographic methods based on aldehyde groups in sugars, and therefore
tests have been performed to determine ion‑exclusion chromatography with DC the preservative effect is reduced, especially
how this affects the asthmatic population mode electrochemical detection (10), and in dark foods where the Maillard reaction
specifically, as this sensitivity may have been ion-exclusion chromatography with pulsed has taken place (13). Some types of caramel
overestimated in previous studies (6). amperometric detection (PAD) (11). coloured food additives are made with

2 9 12 17
LCTC0719 Agilent 1-Col Ad pr2f.indd 1 2019-08-02 8:03 AM
Incognito News Lanciki and Espinosa Pittcon Preview

13
20 Kelting 25 Eichman et al. 29 Training & Events 31 Staff
The Column www.chromatographyonline.com Lanciki and Espinosa

Figure 1: Red wine (diluted 1:30) – original sample 86 mg/kg (black) and sample spiked with saving time and making laboratory work Method
20 mg/kg sulphite resulting in 110 mg/kg (red), with a total recovery of sulphite calculated at 104%. much easier. The general analytical procedure developed
Background: Sulphite analysis is a consists of an alkaline extraction of the
370,0

360,0
necessity for foods and beverages to sample, then anion exchange separation via
be compliant with various food labelling IC using a high-capacity stationary phase
350,0
340,0

330,0

320,0

310,0
regulations. Most foods and beverages and a high-strength sodium hydroxide–
contain residual sulphite concentrations sodium acetate eluent, followed by DC
300,0

290,0

280,0

270,0

260,0
within a range of approximately mode electrochemical detection. Working
250,0

240,0
10–2000 mg/kg. For this article, a request electrode reconditioning was performed
nA

230,0
220,0 arrived from two different laboratories with fully automatically after each analysis using
210,0

Sulphite
200,0

190,0
the goal of measuring total sulphite in all types a combination of techniques. A recovered
of foodstuffs within 10 min per sample. The electrode surface is automatically available
180,0

170,0

160,0

150,0
140,0
first laboratory was using high performance at the beginning of every analysis, which is a
liquid chromatography (HPLC) followed by first for LC/HPLC analysis of sulphite in these
130,0

120,0

110,0

100,0

90,0
conductivity detection, where interferences types of samples.
from organics created several challenges. This Several foodstuff matrices (both solid
80,0

0,0 0,5 1,0 1,5 2,0 2,5 3,0 3,5 4,0 4,5 5,0 5,5 6,0 6,5 7,0 7,5 8,0

Time (min) method showed both a lack of selectivity and liquid) were tested against this method
and long analysis times due to the need to including chickpeas, mustard, cherries,
sulphites (Caramel Colour II / Caustic Sulphite other methods opt for disposable electrodes separate sulphite from sulphate, as well as capers, canned garlic, chili pepper, and red
Caramel, and Caramel Colour IV / Sulphite to avoid this step altogether (11). from coeluting organic interferences. The wine. Each sample was measured via IC
Ammonia Caramel) (13,14), which can cause second laboratory was using the optimized after dilution with a stabilization solution
problems for this method. Another issue Sulphite Analysis, Simplified Monier-Williams (OMW) Official Method (Table 1) in a 1:30 ratio (w/w), as well as
involves the sensitivity of the detector. After This article describes an innovative, fast, 990.28 (7). Long and tedious sample spiked with a known amount of sulphite
a few injections, fouling from contaminants and robust IC method for the determination preparation meant that two analysts could standard to determine recovery. Automatic
rapidly decreases the electrode sensitivity. of total sulphite in food and beverage complete only five sample measurements integration using peak areas was evaluated
Frequent reconditioning of the working samples based on electrochemical detection per day. Other analytical methods also with software.
electrode is necessary due to a rising (DC mode). A range of food and beverage involve demanding sample handling steps, Stabilization Solutions: Sulphite is a highly
background and baseline noise, and can be products were analyzed, with sulphite with results disadvantaged by sensitivity and unstable ion in solution. To prevent rapid
accomplished in a couple of ways. Manual recovery values near 100% in all cases recovery issues. Therefore, a robust, selective, oxidation to sulphate, it is always necessary
polishing and utilizing PAD pulse sequences under the defined method conditions. sensitive, and fast determination method to work with a stabilization solution. Fresh
are the most common choices to recondition Using a single, robust chromatographic was needed to overcome the drawbacks of solutions must be prepared daily in order to
the surface of the working electrode, while method, samples can be treated identically, conventional procedures. achieve the most accurate and reproducible

2 Incognito 9 News 12 Lanciki and Espinosa 17 Pittcon Preview

14
20 Kelting 25 Eichman et al. 29 Training & Events 31 Staff
The Column www.chromatographyonline.com Lanciki and Espinosa

Figure 2: Canned garlic – original sample 2 mg/kg (black) and sample spiked with 40 mg/kg sample vial. The sample vial was completely 930 Compact IC Flex) with a high
sulphite resulting in 42 mg/kg (red), with a total recovery of sulphite calculated at 100%. filled to avoid leaving any headspace, which capacity analytical column (Metrosep
could lead to sulphite oxidation. Carb 2 - 150/4.0 with a Metrosep Carb
Liquid Samples: 2 Guard/4.0 to protect the column [both
400.0
A 1 g aliquot of sample was measured Metrohm]) housed in a column oven to
380.0 into a tared vortex tube, then the diluted ensure a constant temperature for all
stabilization solution was added until the analyses. An amperometric detector, utilizing
nA

360.0
total weight equaled 30 g (sample dilution a reference electrode (Ag/AgCl) and working
340.0 1:30 w/w). The solution was mixed manually electrode (Au), was utilized in DC mode for
320.0 with a vortex shaker for 15 min after sulphite measurement. The detector was set
covering the tube. Approximately 3 mL to the same temperature as the column oven
300.0
of the supernatant was aspirated and the for the most accurate results.
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 sample was manually passed through a Analysis Conditions for the Measurement of
Time (min) 0.2 µm filter into a 2 mL HPLC sample vial. Total Sulphite via IC:
The sample vial was completely filled to Eluent flow: 0.5 mL/min; run time: 8 min;
results. The ultrapure water must be degassed Solid Samples: prevent further sulphite oxidation from the reconditioning of electrode: 2 min; detector
prior to eluent preparation as well as for all Solid samples should be processed by headspace. mode: DC; channel measurement: current
other working solutions and standards. The chopping and crushing until homogenized. Analysis: Eluent and stabilization solutions (nA); E(V): 300 mV; range: auto; damping:
stabilization stock solution can be prepared A 1 g portion of sample was measured were prepared from degassed ultrapure off; detector temperature: 35 °C; column
from commercial reagents, and then diluted into a tared vortex tube, then the diluted water directly before launching the analytical oven temperature: 35 °C; sample loop
1:1000 with degassed ultrapure water to stabilization solution was added until the sequence (Table 1). The sample exposure (partial loop mode): 3 µL; sample cooling
obtain a working stabilization solution total weight equaled 30 g (sample dilution to air was minimized to prevent (further) temperature: 6 °C.
(Table 1). 1:30 w/w). After covering, the solution sulphite oxidation. This includes covering the After each chromatographic
Sample Preparation: All samples and was blended in the tube for 1 min using completely filled sample vials with a septum, determination, the working electrode
standards were prepared with fresh working an Ultra-Turrax homogenizer from IKA (or which leads to higher recovery values and surface was fully reconditioned within
stabilization solution. After preparing the similar equipment), then mixed manually better reproducibility of the results. For 2 min. No manual intervention was
fresh stabilization solutions and preparing with a vortex shaker or automatically with longer sample series, it is recommended to necessary, and there is no need to buy
the sample extracts, samples were then high-throughput plant and animal tissue use an autosampler that includes a cooling disposable electrodes. In this way, each
passed through a 0.2 µm filter to prevent homogenizer. Approximately 3 mL of the function, as this can extend the stability of analysis was performed in exactly the
bacteria from influencing the results before supernatant was aspirated from below the the prepared samples to at least 24 h. same manner, making this a robust,
completely filling 2 mL sample vials covered lipid layer, and the sample manually passed Measurements were performed on an time-saving method with excellent
with a septum. through a 0.2 µm filter into a 2 mL HPLC ion chromatograph system (Metrohm reproducibility of results.

2 Incognito 9 News 12 Lanciki and Espinosa 17 Pittcon Preview

15
20 Kelting 25 Eichman et al. 29 Training & Events 31 Staff
The Column www.chromatographyonline.com Lanciki and Espinosa

Results foodstuff to be analyzed with the same Food Research 30, 1–76 (1986). Suoniemi-Kähärä, Fast and Accurate Automated
Any kind of sample can be measured with method. Since the retention time for 3. S. Taylor, R. Bush, and J. Nordlee, in Food Allergy: Method for Free Sulfite Analysis in Wine
this method without needing to change sulphite never changes with this method Adverse Reactions to Foods and Food Additives, (Application Note 71451), Thermo Fisher Scientific,
the parameters. With the newly developed configuration, a range of samples can be D. Metcalfe, H. Sampson, and R. Simon, Eds. Sunnyvale, California, USA (2014).
electrode reconditioning step, a fresh working analyzed concurrently without the need to (John Wiley & Sons, Chichester, UK, 5th ed., 2013) 13. T.A. Vollmuth, Food and Chemical Toxicology 111,
electrode surface is available after only 2 min, modify anything. pp. 361–374. 578–596 (2018).
which fulfills the customer request of a A newly developed electrode treatment 4. US Food and Drug Administration, CFR Title 21, 14. P.R. Ashurst, R. Hargitt, and F. Palmer, Soft Drink
complete total sulphite analysis within 10 min procedure enables this method to run Volume 2, Section 130.9: Sulfites in standardized and Fruit Juice Problems Solved (Woodhead
(including 8 min per chromatogram run). quickly and efficiently for all kinds of food (FDA, Silver Spring, Maryland, USA, 2019). Publishing, UK, 2nd ed., 2017), pp. 67–94.
It is important to remark that the complex matrices without fouling of the Retrieved from https://www.accessdata.fda.gov/
combination of a very high capacity column electrode surface, which would lead to scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?fr=130.9 Alyson Lanciki is currently the Scientific
and the eluent revealed an important non‑repeatable, unreliable results. Utilization 5. R. Simon, Pediatric Allergy and Immunology 3(4), Editor of Metrohm AG, a manufacturer
improvement of peak resolution between of an autosampler with cooling function 218–221 (1992). of high-precision instruments for
formaldehyde and sulphite, reducing sample extends the stability of prepared food 6. H.A. Vally, Thorax 56, 763–769 (2001). laboratory and process analysis. After a
matrix effects as well as exhibiting very samples to allow uninterrupted total sulphite 7. AOAC International, AOAC Official Method postdoctoral position in France, Alyson
precise repeatability of retention times. determination for a full 24 h. With a complete 990.28, Sulfites in Foods, Optimized joined the Metrohm family in 2013 as an
Not only is this method more robust analysis time of 10 min per sample, including Monier‑Williams Method (AOAC International, ion chromatography application specialist
than previous methods, it is also extremely the automatic reconditioning of the working Rockville, Maryland, USA, 1994). in the Netherlands at Metrohm Applikon,
sensitive to sulphite, as can be seen in electrode, this method offers the option of 8. J. Lin, Y. Zhu, W. Cheng, J. Wang, B. Wu, and J. where she eventually became the global
Figure 1 and Figure 2. Retention times for measuring total sulphite in up to 144 samples Wang, Food Science and Technology Research marketing manager for the Metrohm
these figures differ because the samples were per day. This result is a vast improvement 20(5), 1079–1085 (2014). Process Analytics brand of instruments.
analyzed in two separate laboratories with upon the current sample throughput of five 9. B.S. Liao, J.C. Sram, and D.J. Files, Journal of Miguel Espinosa is currently the Product
different setups. samples per shift (or 15 samples every 24 h) AOAC International 96(5), 1103–1108 (2013). Manager for Ion Chromatography at
at the contract laboratory, who were behind 10. AOAC International, AOAC Official Method Metrohm Hispania, located in Madrid,
Conclusions the original request to develop a new method 990.31, Sulfites in Foods and Beverages, Ion Spain. Miguel joined Metrohm in 1999
This fast and accurate IC method overcomes for total sulphite analysis in foodstuffs. Exclusion Chromatographic Method (AOAC as an application specialist for ion
several limitations encountered in total International, Rockville, Maryland, USA, 1995). chromatography.
sulphite analysis based on the different References 11. L. Chen, B. De Borba, and J. Rohrer,
analytical methodologies used to date and 1. B. Wedzicha, Chemistry of Sulphur Dioxide Determination of Total and Free Sulfite in Foods
E-mail: info@metrohm.com / ic@
is valid for any kind of food or beverage in Foods (Elsevier Applied Science Publishers, and Beverages (Application Note 54), Thermo metrohm.es
products. The simplified sample preparation Barking, UK, 1984). Fisher Scientific, Sunnyvale, California, USA (2016). Website: www.metrohm.com/es-es
technique allows both liquid and solid 2. S. Taylor, N. Higley, and R. Bush, Advances in 12. E. Naigeon, M. Rastetter, M. Klemm, and A.