You are on page 1of 1

TEST NAME REACTION FILTER MINIMUM MAXIMUM STANDARD/

ACCEPTABLE ACCEPTABLE FACTOR RANGE


ABSORBANCE FOR ABSORBANCE FOR
WORKING REAGENTS WORKING REAGENTS
GLUCOSE increasing 505 nm NA 0.3 Abs. 280-340
TRIGLYCERIDE increasing 505 nm NA 0.3 Abs. 870-1050
S
CHOLESTEROL increasing 505 nm NA 0.3 Abs. 560-697
CALCIUM- increasing 630 nm NA 1.1 Abs. 27-35
ARZANAZO
URIC ACID increasing 505 nm NA 0.3 Abs. 30-38
BILIRUBIN increasing 546 nm NA 0.2 Abs. NA
TOTAL increasing 546 nm NA 0.1 Abs. 18-24
PROTEIN
ALBUMIN increasing 630 nm NA 0.3 Abs. 3.7-4.3

CREATININE increasing 505 nm NA NA 48-62


UREA-UV decreasing 340 nm 0.8 Abs NA 800-1050
AST decreasing 340 nm 0.8 Abs NA
ALT decreasing 340 nm 0.8 Abs NA
ALKALINE increasing 405 nm NA 0.8 Abs
PHOS.
CK increasing 340 nm NA 0.8 Abs
CK-NAC increasing 340 nm NA 0.8 Abs
AMYLASE increasing 405 nm NA 0.8 Abs

Please follow below points to rectify the complaints of biochemistry reagents

 Check working reagent absorbance and match absorbance as per above chart if reagent
absorbance observed as outside range it means reagent is deteriorated
 Check standard - factor and match the factor as per above chart if factor observed as
outside range it means standard is deteriorated or contaminated