You are on page 1of 1

HYGROMIL'S 2000

Hygrometry probe
• Continuous measurement of gas dew point
• Dew point measurement range: -80°C to +20°C at atmospheric pressure
• Great stability over time even with a low dew point
• Polymeric thin-layer sensor with automatic calibration
• Fast response time
• Insensitive towards condensation, particle contamination,oil vapours and
most of chemicals
• Alarm on dry contact adjustable on the whole measurement range
(250V - 4A maxi)
• Available output : 4 – 20 mA
• The probe can be delivered separately
• Power supply 200/260V - 50/60 Hz
43
Dew point temperature Opera ng condi ons
Measurement range -80°C…+20°C Temperature 0…+60°C
Dew point precision up to -60°C ± 2°C Relative humididty 0…100%
Dew point precision up to -60°C -60 -> -20°C Td 5s (10 s)
a gas temperature at 20°C, a flow -20 -> -60°C Td 45 s (10 min) Pressure 0…20 bar
rate higher than 1l/min and a Sample flow rate No effect
pressure of 1 bar

Probe characteris cs
Analogical output 4…20 mA 53
Operating voltage 17-35 V DC
20/28 V AC Bare probe

Sink current at 24 V DC 220 mA max.


External charge for analogical output 500 Ω max.
Sensor protection Sintered stainless steel filter
Protection level PF 65 (NEMA 4)
Storage temperature -40…+70°C
Delivered with 1,5 meter-long cable

Dimensions (mm)
Bare probe for standard exchange : 715745 ES Probe with sampling probe carrier : 715745

HYGROMIL’S 2000 V
Probe installed
in the cabinet : 715810

60
Specifications are subject to be changed without notice