Review of Calibration Results

Type of Equipment : Standard Weight


Usage : Direct Shear Test
Machine Serial No. : -
Calibration Cert No. : MSR22120562
Date Calibrated : 8-Dec-22

Result:
Measured Value Uncertainty, Measured Value + Comply
Nom Value, g Acceptable Weight, g
(xbar) U Uncertainty, g (Yes/No)
10.00 10.00080 0.000025 10.00078 - 10.00083 9.90000 - 10.10000 Yes
20.00 20.00234 0.000018 20.00232 - 20.00236 19.80000 - 20.20000 Yes
20.00 19.99950 0.000018 19.99948 - 19.99952 19.80000 - 20.20000 Yes
50.00 49.99955 0.000046 49.99950 - 49.99960 49.50000 - 50.50000 Yes
50.00 49.99957 0.000046 49.99952 - 49.99962 49.50000 - 50.50000 Yes
100.00 100.0001 0.000120 99.99998 - 100.00022 99.00000 - 101.00000 Yes
100.00 100.0014 0.000120 100.00128 - 100.00152 99.00000 - 101.00000 Yes
100.00 100.0012 0.000120 100.00108 - 100.00132 99.00000 - 101.00000 Yes
100.00 100.0009 0.000120 100.00078 - 100.00102 99.00000 - 101.00000 Yes
100.00 100.0016 0.000120 100.00148 - 100.00172 99.00000 - 101.00000 Yes
200.00 200.0015 0.000140 200.00136 - 200.00164 198.00000 - 202.00000 Yes
200.00 200.0053 0.000140 200.00516 - 200.00544 198.00000 - 202.00000 Yes

Acceptance criteria:
Accuracy of 1% of the vertical load as per BS EN ISO 17892-10 :2018 clause 5.4.1

Prepared by: Mohd Fadhir


Laboratory Asst. Engineer

Reviewed by: Norlaila


Laboratory Manager

Date: 20th September 2023