Professional Documents
Culture Documents
PROCEDURE MANUAL
PREPARED BY:
NFHSI MGMT
2023
TITLE: CLINICAL CHEMISTRY PROCEDURE MANUAL DOCUMENT NUMBER: 1
AUTHOR: JOANNE TERAMOTO REVIEWED BY:
LABORATORY NAME: NFHSI, INC. APPROVED BY:
DATE CREATED: 3/1/2023 EFFECTIVE DATE:
I. GLUCOSE
SUMMARY. Measurement of glucose concentration in serum or plasma is mainly used in the
diagnosis and monitoring of treatment in diabetes mellitus. Other applications are the detection
of neonatal hypoglycemia, the exclusion of pancreatic islet cell carcinoma as well as the
evaluation of carbohydrate metabolism in various diseases.
METHOD. “GOD-PAP”: enzymatic photometric test
PRINCIPLE. Determination of glucose after enzymatic oxidation by glucose oxidase. The
colorimetric indicator is quinoneimine, which is generated from 4-aminoantipyrine and phenol
by hydrogen peroxide under the catalytic action of peroxidase (Trinder’s reaction).
SPECIMEN. See below.
Serum, heparin plasma or EDTA plasma
Separate at the latest 1h after blood collection from cellular contents.
Stability in plasma after addition of a glycolytic inhibitor (Flouride, monoiodacetate, mannose):
2 days at 20-25℃
7 days at 4-8℃
1 day at -20℃
Stability in serum (separated from cellular contents, hemlysis free) without adding a glycolytic
inhibitor:
8h at 25℃
72h at 4℃
REAGENTS. See below.
Components and Concentrations
Phosphate buffer pH7.5 250mmol/L
Phenol 5 mmol/L
4-Aminoantipyrine 0.5 mmol/L
Glucose oxidase (GOD) ≥ 10 kU/L
Peroxidase (POD) ≥ 1 kU/L
Storage Instructions and Reagent Stability
Reagent is stable up to the end of the indicated month of expiry, if stored at 2-8℃, protected
from light and contamination is avoided. Don not freeze the reagents.
NOTE: It has to be mentioned that the measurement is not influenced by occasionally ocuring
color changes, as long as the absorbance of the reagent is <0.3 at 546nm.
Reagent Preparation
Reagent is ready to use.
ASSAY PROCEDURE. Application sheets for automated systems re available on request.
Wavelength 500 nm, Hg 546 nm
Optical path 1 cm
Temperature 20-25℃/37℃
Measurement Against reagent blank
Blank Sample/Calibrator
Sample/Calibrator - 10uL
Dist. Water 10uL -
Reagent 1000uL 1000uL
Mix, incubate 20 min at 20-25℃ or 10 min at 37℃. Read absorbance against the blank within 60
min.
CALCULATION. See below.
With calibrator
Glucose (mg/dL) = A Sample x Conc. Cal. (mg/dL)
A Cal.
Conversion factor
Glucose (mg/dL) x 0.05551 = Glucose (mmol/L)
CALIBRATORS AND CONTROLS. For the calibration of automated photometric systems, DiaSys
TruCal U Calibrator is recommended. The assigned values of this calibrator have been made
traceable to the reference metho gas chromatography – isotope dilution mass spectrometry
(GC-IDMS). Glucose Standard FS may be used alternatively for calibration. For internal quality
control, DiaSys Trulab N and P controls should be assayed. Each laboratory should establish
corrective action in case of deviations in control recovery.
PERFORMANCE CHARACTERISTICS. See below.
Measuring range
The test have been developed to determine glucose concentrations within a measuring range
from 1-400 mg/dL (0.06-22.2 mmol/L). When values exceed this range samples should be
diluted 1 + 4 with NaCl solution (9g/L) and the result multiplied by 5.
Specificity/Interferences
No interference was observed by ascorbic acid up to 15 mg/dL, bilirubin up to 40 mg/dL,
hemoglobin up to 200 mg/dL and lipemia up to 2000 mg/dL triglycerides.
Sensitivity/Limit of Detection
The lower limit of detection is 1 mg/dL.
WARNINGS AND PRECAUTIONS. The reagent contains sodium azide (0.95 g/L) as preservative.
Do not swallow. Avoid contact with skin and mucous membranes. Please refer to the safety data
sheets and take necessary purposes, the results should always be assessed with the patients
medical history, clinical examinations and other findings.
REFERENCE RANGE AND CLINICAL INTERPRETATION. See below.
Adult : FBS 70-115 mg/dL
RBS
Elevated plasma glucose levels are expected in a variety of clinical conditions especially diabetes
mellitus, Cushing’s syndrome and hyperadrenalism. Decreased plasma glucose levels are
observed in hyper-insulinism, anti-diabetic treatment and hypoadrenalism.
II. UREA
SUMMARY. Urea contributes most of the body’s non-protein nitrogen, accounting 45% of the
total. Urea is the nitrogen-containing end product of protein catabolism. It is synthesized in the
liver, released into blood circulation and excreted by the kidneys. States associated with
elevated levels of urea in blood are referred to as hyperuremia or azotemia. Parallel
determination of urea and creatinine is performed to differentiate between pre-renal and post-
renal azotemia. Furthermore, measurement of urea in blood is a useful indicator of renal and
hepatic integrity.
Pre-renal azotemia, caused by e.g. dehydration, increased protein catabolism, cortisol
treatment or decreased renal perfusion, leads to increased urea levels, while creatinine values
remain within the reference range. In post-renal azotemias, caused by the obstruction of the
urinary tract, both urea and creatinine levels rise, but creatinine in a smaller extent. In renal
diseases urea concentrations are elevated when the glomerular filtration rate is markedly
reduced and when the protein intake is higher than 200g/day.
METHOD. “Urease-GLDH”: enzymatic UV test
PRINCIPLE. Urea reacts directly with diacetyl monoxime under strong acidic conditions to give a
yellow condensation product. The reaction is intensified by the presence of ferric ions and
thiosemicarbazide. The intense red colour formed is measured at 540nm/ yellow green filter.
SPECIMEN. See below.
Serum, plasma (no ammonium heparin), fresh urine
Dilute urine 1 + 0 with dist. water and multiply results by 51.
TruLab Urine controls must be prediluted the same way as patient samples.
Stability in serum or plasma:
7 days at 20-25℃
7 days at 4-8℃
1 year at -20℃
in urine:
2 days at 20-25℃
7 days at 4-8℃
1 month at -20℃
Freeze only once. Discard contaminated specimens.
REAGENTS. See below.
Components and Concentrations
R1: TRIS pH7.8 150 mmol/L
2-Oxoglutarate 9 mmol/L
ADP 0.75 mmol/L
Urease ≥7 kU/L
GLDH (bovine) ≥1 kU/L
R2: NADH 1.3 mmol/L
Storage Instructions and Reagent Stability
Reagents are stable up to the end of the indicated month of expiry, if stored at 2-8℃, protected
from light and contamination is avoided. Do not freeze the reagents.
Reagent Preparation
Reagents are ready to use.
ASSAY PROCEDURE. Application sheets for automated systems are available on request.
Wavelength 340 nm, Hg 334 nm, Hg 365 nm
Optical path 1 cm
Temperature 25℃/30℃/37℃
Measurement Against reagent blank; 2-point kinetic
Blank Sample/Calibrator
Sample/Calibrator - 10uL
Reagent 1 1000uL 1000uL
Mix, incubate 0-5 min., then
add:
Reagent 2 250uL 250uL
Mix, incubate for approx. 60
sec at 25℃/30℃ r approx.
30-40 sec at 37℃, then read
absorbance A1. Read
absorbance A2 exactly after
another 60 seconds.
△A=(A2-A1) Sample/Calibrator
NOTES:
1. The method is optimized for 2-point kinetic measurement. It is recommended to perform the
method only on mechanized equipment because it is difficult to incubate all samples ad the
reagent blank strictly for the same time intervals. The assay scheme may be used for adaptation
purposes or instruments with no specific adaptation sheet. The volumes may be proportionally
smaller.
2. The statement “approx. 60 sec or approx. 30-40 sec” means that the time period chosen does
not need to be exactly 60 resp. 30-40 sec. A time once chosen has to be respected exactly for all
samples, standards and the reagent blanc.
CALCULATION. See below.
With calibrator
Urea (mg/dL) = ∆A Sample x Conc. Cal. (mg/dL)
∆A Cal.
Conversion factor
Urea (mg/dL) x 0.1665 = Urea (mmol/L)
Urea (mg/dL) x 0.467 = BUN (mg/dL)
BUN (mg/dL) x 2.14 = Urea (mg/dL)
CALIBRATORS AND CONTROLS. For the calibration of automated photometric systems, DiaSys
TruCal U calibrator is recommended. The assigned values of the calibrators have been made
traceable to NIST SRM-909 Level 1. Urea standard FS may be used alternatively for calibration.
Diasys Trulab N, P and Trulab Urine controls should be assayed for internal quality control. Each
laboratory should establish corrective action in case of deviations in control recovery.
PERFORMANCE CHARACTERISTICS. See below.
Measuring range
The test has been developed to determine urea concentrations within a measuring range from
2-300 mg/dL (0.3-50 mmol/L) in serum/plasma respectively up to 30 g/dL (5mol/L) in urine.
When values exceed this range, the samples should be diluted 1+ 2 with NaCl solution (9g/L)
and the result multiplied by 3.
Specificity/Interferences
No interference was observed by ascorbic acid up to 30 mg/dL, bilirubin up to 40 mg/dL,
hemoglobin up to 500 g/dL and lipemia up to 2000 mg/dL triglycerides. Ammonium ions
interfere; therefore, do not use ammonium heparin as anticoagulant for collection of plasma.
Sensitivity/Detection Limit
The lower limit of detection is 2 mg/dL.
WARNING AND PRECAUTIONS. The reagents contain sodium azide (0.95 g/L) as preservative.
Do not swallow. Avoid contact with skin and mucous membranes. Reagent 1 contains animal
material. Handle the product as potentially infectious according to universal precautions and
good laboratory practices. Please refer to the safety data sheets and take necessary purposes,
the results should always be assessed with the patients medical history, clinical examinations
and other findings.
REFERENCE RANGE AND CLINICAL INTERPRETATION. See below.
Adult : BUN serum/plasma
Women <50 7.94 – 20.1 mg/dL
Women >50 7.01- 18.7 mg/dL
Men <50 8.87-20.5 mg/dL
Men>50 8.41-25.7mg/dL
Urea/Creatinine ratio in serum:
25-40 (mmol/L)
20-35 (mg/dL)
Elevated serum urea levels may be due to pre-renal, renal or post-renal etiology. Pre-renal
causes could be cardiac related or due to increased protein catabolism, and dehydration. Renal
causes include glomerulonephritis, chronic nephritis, nephrotic syndrome and other kidney
disease. Post-renal causes include obstruction of the urinary tract.
Decreased serum urea levels could be due to pregnancy, intravenous infusion, low antidiuretic
hormone secretion hormone secretion, low protein intake, severe liver disease, inborn errors of
urea cycle and SIADH (Syndrome of inappropriate ADH secretion).
III. CREATININE
SUMMARY. Creatinine is a waste product excreted by the kidneys mainly by glomerular
filtration. The concentration of creatinine in plasma of a healthy individual is fairly constant,
independent from water intake, exercise and rate f urine production. Therefore increased
plasma creatinine values always indicate decreased excretion. i.e. impaired kidney function.
Creatinine clearance is a good indicator for the glomerular filtration rate (GFR) which allows
better detection of kidney diseases and monitoring of renal function. For this purpose,
creatinine is measured simultaneously in serum and urine collected over a defined time period.
METHOD. Kinetic test without depolarization according to the Jaffe method
PRINCIPLE. Creatinine forms a colored orange-red complex in an alkaline picrate solution. The
difference at fixed times during conversion is proportional to the concentration of creatinine in
the sample.
SPECIMEN. See below.
Human serum, heparin plasma or urine.
Only use suitable tubes or collection containers for specimen collection and preparation.
When using primary tubes, follow the manufacturer’s instructions.
Stability in serum/plasma:
7 days at 4-25℃
3 months at -20℃
Stability in urine:
2 days at 20-25℃
6 days at 4-8℃
6 months at -20℃
Dilute urine 1 + 49 with dist. water, multiply the result by 50. TruLab Urine controls must be
prediluted the same way as patient samples.
Only freeze once. Discard contaminated specimens.
REAGENTS. See below.
Components and Concentrations
R1: Sodium hydroxide 0.2 mol/L
R2: Picric acid 20 mmol/L
Storage and Stability
Reagents are stable up to the date of expiry indicated on the kit, if stored at 2-25℃ and
contamination is avoided. Do not freeze and protect from light.
The in-use stability of the reagent is 18 months.
Reagent Preparation
The reagents are ready to use.
ASSAY PROCEDURE. See below.
Basic settings for BioMajesty
Wavelength 505/571 nm
Temperature 37℃
Measurement Kinetic
Sample/Calibrator 5.0uL
Reagent 1 80uL
Reagent 2 20uL
Addition reagent 2 Cycle19 (286 s)
Absorbance Cycle 24/32 (354 s/464 s)
Calibration Linear
CALCULATION. See below.
With calibrator
Serum/Plasma
Creatinine [mg/dL] = ∆A Sample x Conc. Cal. [mg/dL]
∆A Cal.
Urine
Creatinine [mg/dL] = ∆A Sample x Conc. Cal. [mg/dL] x 50
∆A Cal.
Creatinine Clearance [mL/min/1.73 m2] [5]
= mg Creatinine/ 100 mL Urine x mL Urine
mg Creatinine/ 100 mL Serum x min Urine collection time
The calculated creatinine clearance refers to the average body surface of an adult (1.73 m 2).
Conversion Factor
Creatinine [mg/dL] x 88.4 = Creatinine [umol/L]
Creatinine [mg/dL] x 0.0884 = Creatinine [mmol/L]
CALIBRATORS AND CONTROLS. DiaSys TruCal U is recommended for calibration. Calibrator
values for the compensated method have been made traceable to the NIST Standard Reference
Material SRM 967 using level 1 and 2 and, therefore, to GC-IDMS. Creatinine Standard FS may
be used alternatively for calibration. Use DiaSys TruLab N and P or TruLab Urine Level 1 and 2
controls for internal quality control. Quality control must be performed after calibration. Control
intervals and limits have to be adapted to the individual requirements of each laboratory.
Results must be within the defined ranges. Follow the relevant legal requirements and
guidelines. Each laboratory should establish corrective action in case of deviations in control
recovery.
PERFORMANCE CHARACTERISTICS. See below.
Data evaluated on BioMajesty
Measuring range up to 14mg/dL
When values exceed this range samples should be diluted 1 + 1 with NaCl solution (9g/L) and
the result multiplied by 2.
Blank Sample/Calibrator
Sample/Calibrator - 10uL
Dist. water 10uL -
Reagent 1000uL 1000uL
Mix, incubate for 20 min. at 20-25℃ or for 10min. at 37℃. Read absorbance within 660 min
against reagent blank.
CALCULATION. See below.
With calibrator
Cholesterol [mg/dL] = A Sample x Conc. Cal. [mg/dL]
A Cal.
Conversion factor
Cholesterol [mg/dL] x 0.02586 = Cholesterol [mmol/L]
CALIBRATORS AND CONTROLS. For the calibration of automated photometric systems, DiaSys
TruCal U calibrator is recommended. The assigned values of the calibrators have been made
traceable to the reference method GC-IDMS. Cholesterol Standard FS may be used alternatively
for calibration. For internal quality control, DiaSys TruLab N and P or TruLab L controls should be
assayed. Each laboratory should establish corrective action in case of deviations in control
recovery.
PERFORMANCE CHARACTERISTICS. See below.
Measuring range
The test has been developed to determine cholesterol concentrations within a measuring range
from 3-750 mg/dL (0.08-19.4 mmol/L). When values exceed this range, the samples should be
diluted 1+ 4 with NaCl solution (9g/L) and the result multiplied by 5.
Specificity/Interferences
No interference was observed by ascorbic acid up to 5 mg/dL, bilirubin up to 20 mg/dL,
hemoglobin up to 200 g/dL and lipemia up to 2000 mg/dL triglycerides.
Sensitivity/Detection Limit
The lower limit of detection is 3 mg/dL (0.08 mmol/L).
WARNING AND PRECAUTIONS. The reagents contain sodium azide (0.95 g/L) as preservative.
Do not swallow. Avoid contact with skin and mucous membranes. Reagent 1 contains animal
material. Handle the product as potentially infectious according to universal precautions and
good laboratory practices. Please refer to the safety data sheets and take necessary purposes,
the results should always be assessed with the patients medical history, clinical examinations
and other findings.
REFERENCE RANGE AND CLINICAL INTERPRETATION. See below.
Desirable ≤200 mg/dL
Borderline high risk 200-240 mg/dL
High risk >240 mg/dL
Serum cholesterol is increased in hypothyroidism, diabetes mellitus, nephrotic syndrome and in
various hyperlipidemias especially those causing xanthomatosis. Elevated serum cholesterol is a
serious risk factor for the development of coronary artery disease. Decreased serum cholesterol
is seen in severe hepatocellular hyperthyroidism and anemia.
The European Task Force on Coronary Prevention recommends to lower TC concentration to less
than 190 mg/dL and LDL-C to less than 115 mg/dL.
V. TOTAL PROTEIN
SUMMARY. Measurement of total protein is a useful test in a variety of disorders. Decreased
total protein concentrations can be detected in defective protein synthesis in the liver, protein
loss due to impaired kidney function, intestinal malabsorption or nutritional deficiency. Elevated
protein levels occur in chronic inflammatory disorders, liver cirrhosis and dehydration.
METHOD. Photometric test according to biuret method
PRINCIPLE. Proteins form a violet blue color complex with copper ions in alkaline solution. The
absorbance of the color is directly proportional to the concentration.
SPECIMEN. See below.
Human serum or heparin plasma
Only use suitable tubes or collection containers for specimen collection and preparation. When
using primary tubes, follow the manufacturer’s instructions.
Stability:
6 days at 20-25℃
4 weeks at 4-8℃
At least one year at -20℃
REAGENTS. See below.
Components and Concentrations
R1: Sodium hydroxide 100 mmol/L
Potassium sodium tartrate 17 mmol/L
R2: Sodium hydroxide 500 mmol/L
Potassium sodium tartrate 80 mmol/L
Potassium iodide 75 mmol/L
Copper sulphate 30 mmol/L
Storage and Stability
Reagents are stable up to date of expiry indicated on the kit, if stored at 2-25℃ and
contamination is avoided. Protect from light. The in-use stability of the reagent is 18 months.
Reagent Preparation
The reagents are ready to use.
ASSAY PROCEDURE. See below.
Wavelength 545 nm
Temperature 37℃
Measurement Endpoint
Sample/Calibrator 2.0uL
Reagent 1 80uL
Reagent 2 20 uL
Addition Reagent 2 Cycle 19 (286 s)
Absorbance 1 Cycle 17/18 (231 s/244 s)
Absorbance 2 Cycle 41/42 (586 s/600 s)
Calibration Linear
CALCULATION. See below.
With calibrator
Total Protein [g/dL] = A Sample x Conc. Cal. [g/dL]
A Cal.
CALIBRATORS AND CONTROLS. DiaSys TruCal U calibrator is recommended. The assigned values
of the calibrators have been made traceable to the biuret method. Total Protein Standard FS
may be used alternatively for calibration. Use DiaSys TruLab N and P for internal quality control.
Quality control must be performed after calibration. Control intervals and limits have to be
adapted to the individual requirements of each laboratory. Results must be within the defined
ranges. Each laboratory should establish corrective action in case of deviations in control
recovery.
PERFORMANCE CHARACTERISTICS. See below.
Data evaluated on BioMajesty
Exemplary data mentioned below may slightly differ in case of deviating measurement
conditions.
Measuring range up to 14 g/dL.
When values exceed this range, samples should be diluted 1 + 1 with NaCl solution (9 g/L) and
the result multiplied by 2.
Limit detection 0.05g/dL
Blank Sample/Calibrator
Sample/calibrator - 20uL
Dist. Water 20uL -
Reagent 1 1000uL 1000uL
Mix incubate for approx. 1
min then add:
Reagent 2 250uL 250uL
Mix, read absorbance after 1
min and start stopwatch.
Read absorbance again after
1,2 and 3 min.
Sample Start
Blank Sample/Calibrator
Sample/Calibrator - 20uL
Dist. Water 20uL -
Mono reagent 1000uL 1000uL
Mix, read absorbance after 1
min and start stopwatch.
Read absorbance again after
1,2, and 3 min.
In urine
2 days at 20-25℃
4 days at 4-8℃
3 weeks at -20℃
Add 10 mL of concentrated HCl to 24 h urine and heat the specimen to dissolve calcium
oxalate.
Only freeze once. Discard contaminated specimens.
REAGENTS. See below.
Components and Concentrations
Reagent:
Phosphate buffer pH7.5 50mmol/L
8-Hydroxyquinoline-5-sulfonic acid 5mmol/L
Arsenazo III 120umol/L
Storage and Stability
Reagents are stable up to the date of expiry indicated on the kit, if stored at 2-8℃ and
contamination is avoided. Do not freeze and protect from light.
Reagent Preparation
The reagents are ready to use.
ASSAY PROCEDURE. Applications for automated systems are available on request.
Wavelength 650nm, Hg 623nm (630-670nm)
Optical path 1cm
Temperature 20-25℃/37℃
Measurement Against reagent blank
Blank Sample/Calibrator
Sample/calibrator - 10uL
Dist. Water 10uL -
Reagent 1000uL 1000uL
Mix incubate for 5 min and read absorbance against reagent blank.
Blank Sample/Calibrator
Sample/calibrator - 10uL
Dist. Water 10uL -
Reagent 1 800uL 800uL
Mix, incubate 5 min, read
absorbance A1, then add:
Reagent 2 200uL 200uL
Mix and read absorbance A2
within 5-60 min
∆A=(A2-A1) Sample/Calibrator
CALCULATION. See below.
With calibrator
Phosphorus [mg/dL] = A Sample x Conc. Cal. [mg/dL]
A Cal.
Conversion Factor
Phosphorus [mg/dL] x 0.3229 = Phosphorus [mmol/L]
Phosphorus [mg/dL] x 3.06619 = Phosphorus [mg/dL]
CALIBRATORS AND CONTROLS. For calibration of automated photometric systems, the DiaSys
TruCal U calibrator is recommended. The assigned values of calibrators have been made
traceable to a primary phophorus standard. Phosphorus Standard FS may be used alternatively
for calibration. Each laboratory should establish corrective action in case of deviations in control
recovery.
PERFORMANCE CHARACTERISTICS. See below.
Measuring range
The test has been developed to determine phosphorus concentrations within a measuring
range from 0.2-30 mg/dL. When values exceed this range, samples should be diluted 1 + 10 with
NaCl solution (9g/L) and the result multiplied by 11.
Specificity/Interferences
No interference was observed by ascorbic acid up to 30 mg/dL, bilirubin up to 60 mg/dL,
hemoglobin up to 1000 mg/dL, lipemia up to 2000 mg/dL.
Sensitivity/Limit Detection
The lower limit of detection is 0.2 mg/dL.
WARNING AND PRECAUTIONS. See below.
1. In very rare cases, samples of patients with gammopathy might give falsified results.
2. Please refer to MSDS and take the necessary precautions for the use of laboratory reagents.
3. For professional use only.
REFERENCE RANGE AND CLINICAL INTERPRETATION. See below.
Serum
Adults 2.6-4.5 md/dL
Plasma
Concentrations of inorganic phosphate are about 0.2-0.3 mg/dL lower in heparinized plasma
than in serum.
An increase in serum phosphorus is found in chronic nephritis progressing with increase renal
failure. A moderate increase is observed in hypoparathyroidism n vitamin D excess.
A decrease in serum phosphorus is observed in rickets or osteomalacia and also in
hyperparathyroidism. Hypophosphatemia may result due to disorders of renal tubular
reabsorption.